Simultaneous electrical online estimation of changes in blood hematocrit and temperature in cardiopulmonary bypass

verfasst von: Martin Wekesa Sifuna, Daisuke Kawashima, Katsuhiro Matsuura, Hiromichi Obara, Yusuke Nakajima, Masahiro Takei

Erschienen in: Journal of Artificial Organs | Ausgabe 4/2022

Einloggen, um Zugang zu erhalten

Abstract

Two equations have been developed from multi-frequency measurements of blood impedance Zb for a simultaneous electrical online estimation of changes in blood hematocrit ΔH [%] and temperatures ΔT [K] in cardiopulmonary bypass (CPB). Zb of fixed blood volumes at varying H and T were measured by an impedance analyzer and changes in blood conductivity σb and relative permittivity εb computed. Correlation analysis were based on changes in σb with H or T at f = 1 MHz while H and T equations were developed by correlating changes in εb with H and T at dual frequencies of f = 1 MHz and f = 10 MHz which best capture blood plasma Zp and red blood cell cytoplasm Zcyt impedances respectively. Results show high correlations between σb and H (R2 = 0.987) or σb and T (R2 = 0.9959) indicating dependence of the electrical parameters of blood on its H and T. Based on computed εb, changes in blood hematocrit ΔH and temperature ΔT at a given time t are estimated as ΔH(t) = 1.7298Δεb (f = 1 MHz) – 1.0669Δεb (f = 10 MHz) and ΔT(t) = -2.186Δεb (f = 1 MHz) + 2.13Δεb (f = 10 MHz). When applied to a CPB during a canine mitral valve plasty, ΔH and ΔT had correlations of R2 = 0.9992 and R2 = 0.966 against H and T respectively as measured by conventional devices.
Literatur
1.
Zurück zum Zitat Karahan M, Sabit S, Umit K. Effects of continuous-flow left ventricular assist devices on cerebral hemodynamics. Artif Organs. 2019;44:457–64.CrossRef Karahan M, Sabit S, Umit K. Effects of continuous-flow left ventricular assist devices on cerebral hemodynamics. Artif Organs. 2019;44:457–64.CrossRef
2.
Zurück zum Zitat Al T, Ambar S, Krystina A, Sciamanna C, Pauwaa S, Macaluso G, et al. A case series of patients with left ventricular assist devices and concomitant mechanical heart valves. Artif Organs. 2020;44(10):1050–4.CrossRef Al T, Ambar S, Krystina A, Sciamanna C, Pauwaa S, Macaluso G, et al. A case series of patients with left ventricular assist devices and concomitant mechanical heart valves. Artif Organs. 2020;44(10):1050–4.CrossRef
4.
Zurück zum Zitat Hastings SM, Ku DN, Wagoner S, Maher KO, Deshpande S. Sources of circuit thrombosis in pediatric extracorporeal membrane oxygenation. ASAIO J. 2017;63(1):86–92.CrossRefPubMed Hastings SM, Ku DN, Wagoner S, Maher KO, Deshpande S. Sources of circuit thrombosis in pediatric extracorporeal membrane oxygenation. ASAIO J. 2017;63(1):86–92.CrossRefPubMed
5.
Zurück zum Zitat Berei TJ, Lilliblad MP, Wilson KJ, Garberich RF, Hryniewicz KM. Evaluation of systemic heparin versus bivalirudin in adult patients supported by extracorporeal membrane oxygenation. ASAIO J. 2018;64(5):623–9.CrossRefPubMed Berei TJ, Lilliblad MP, Wilson KJ, Garberich RF, Hryniewicz KM. Evaluation of systemic heparin versus bivalirudin in adult patients supported by extracorporeal membrane oxygenation. ASAIO J. 2018;64(5):623–9.CrossRefPubMed
7.
Zurück zum Zitat Schmied H, Kurz A, Sessler DI, Kozek S, Reiter A. Requirements during total hip arthroplasty mild. Lancet. 1995;347:289–92.CrossRef Schmied H, Kurz A, Sessler DI, Kozek S, Reiter A. Requirements during total hip arthroplasty mild. Lancet. 1995;347:289–92.CrossRef
8.
Zurück zum Zitat Kaukuntla H, Harrington D, Bilkoo I, Clutton-Brock T, Jones T, Bonser RS. Temperature monitoring during cardiopulmonary bypass - do we undercool or overheat the brain? Eur J Cardio-thoracic Surg. 2004;26(3):580–5.CrossRef Kaukuntla H, Harrington D, Bilkoo I, Clutton-Brock T, Jones T, Bonser RS. Temperature monitoring during cardiopulmonary bypass - do we undercool or overheat the brain? Eur J Cardio-thoracic Surg. 2004;26(3):580–5.CrossRef
9.
Zurück zum Zitat England JM, Waters DAW. Re-assessment of the reliability of the haematocrit. Br J Heamatology. 1972;23:247–56.CrossRef England JM, Waters DAW. Re-assessment of the reliability of the haematocrit. Br J Heamatology. 1972;23:247–56.CrossRef
10.
Zurück zum Zitat Langham GE, Maheshwari A, Contrera K, You J, Mascha E, Sessler DI. Noninvasive temperature monitoring in postanesthesia care units. Anesthesiology. 2009;111(1):90–6.CrossRefPubMed Langham GE, Maheshwari A, Contrera K, You J, Mascha E, Sessler DI. Noninvasive temperature monitoring in postanesthesia care units. Anesthesiology. 2009;111(1):90–6.CrossRefPubMed
11.
Zurück zum Zitat Engelman R, Baker RA, Likosky DS, Grigore A, Dickinson TA, Shore-Lesserson L, et al. Clinical practice guidelines for cardiopulmonary bypass - temperature management during cardiopulmonary bypass. Ann Thorac Surg. 2015;100(2):748–57.CrossRefPubMed Engelman R, Baker RA, Likosky DS, Grigore A, Dickinson TA, Shore-Lesserson L, et al. Clinical practice guidelines for cardiopulmonary bypass - temperature management during cardiopulmonary bypass. Ann Thorac Surg. 2015;100(2):748–57.CrossRefPubMed
12.
Zurück zum Zitat Sifuna MW, Baidillah R, Sapkota A, Takei M. A Cole-Cole dielectric relaxation analysis of albumin and γ -globulins for protein quantification by electrical impedance spectroscopy. Electroanalysis. 2020;32:1121–9.CrossRef Sifuna MW, Baidillah R, Sapkota A, Takei M. A Cole-Cole dielectric relaxation analysis of albumin and γ -globulins for protein quantification by electrical impedance spectroscopy. Electroanalysis. 2020;32:1121–9.CrossRef
13.
Zurück zum Zitat Sifuna MW, Baidillah MR, Kawashima D, Darma PN, Odari EO, Takei M. Determination of sensitive frequency margin for aggregated protein concentration quantification by fd-electrical impedance tomography. Measurement [Internet]. 2021;186:110–35. Sifuna MW, Baidillah MR, Kawashima D, Darma PN, Odari EO, Takei M. Determination of sensitive frequency margin for aggregated protein concentration quantification by fd-electrical impedance tomography. Measurement [Internet]. 2021;186:110–35.
14.
Zurück zum Zitat Zhbanov A, Yang S. Electrochemical impedance spectroscopy of blood for sensitive detection of blood hematocrit, sedimentation and dielectric properties. Anal Methods. 2017;9(22):3302–13.CrossRef Zhbanov A, Yang S. Electrochemical impedance spectroscopy of blood for sensitive detection of blood hematocrit, sedimentation and dielectric properties. Anal Methods. 2017;9(22):3302–13.CrossRef
15.
Zurück zum Zitat Hayashi Y, Brun MA, Machida K, Nagasawa M. Principles of dielectric blood coagulometry as a comprehensive coagulation test. Anal Chem. 2015;87(19):10072–9.CrossRefPubMed Hayashi Y, Brun MA, Machida K, Nagasawa M. Principles of dielectric blood coagulometry as a comprehensive coagulation test. Anal Chem. 2015;87(19):10072–9.CrossRefPubMed
16.
Zurück zum Zitat Ramaswamy B, Yeh YT, Zheng S. Microfluidic device and system for point-of-care blood coagulation measurement based on electrical impedance sensing. Sensors Actuators B Chem. 2013;180:21–7.CrossRef Ramaswamy B, Yeh YT, Zheng S. Microfluidic device and system for point-of-care blood coagulation measurement based on electrical impedance sensing. Sensors Actuators B Chem. 2013;180:21–7.CrossRef
17.
Zurück zum Zitat Fuse T, Sapkota A, Maruyama O, Kosaka R, Yamane T, Takei M. Analysis of the influence of volume and red blood cell concentration of a thrombus on the permittivity of blood. J Biorheol. 2015;29(1):15–8.CrossRef Fuse T, Sapkota A, Maruyama O, Kosaka R, Yamane T, Takei M. Analysis of the influence of volume and red blood cell concentration of a thrombus on the permittivity of blood. J Biorheol. 2015;29(1):15–8.CrossRef
18.
Zurück zum Zitat Asakura Y, Sapkota A, Maruyama O, Kosaka R, Yamane T, Takei M. Relative permittivity measurement during the thrombus formation process using the dielectric relaxation method for various hematocrit values. J Artif Organs. 2015;18(4):346–53.CrossRefPubMed Asakura Y, Sapkota A, Maruyama O, Kosaka R, Yamane T, Takei M. Relative permittivity measurement during the thrombus formation process using the dielectric relaxation method for various hematocrit values. J Artif Organs. 2015;18(4):346–53.CrossRefPubMed
19.
Zurück zum Zitat Cha K, Faris RG, Brown EF, Wilmore DW. An electronic method for rapid measurement of haematocrit in blood samples. Physiol Meas. 1994;15:129–37.CrossRefPubMed Cha K, Faris RG, Brown EF, Wilmore DW. An electronic method for rapid measurement of haematocrit in blood samples. Physiol Meas. 1994;15:129–37.CrossRefPubMed
20.
Zurück zum Zitat Haines DE, Verow AF. Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium. Circulation. 1990;82(3):1034–8.CrossRefPubMed Haines DE, Verow AF. Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium. Circulation. 1990;82(3):1034–8.CrossRefPubMed
21.
Zurück zum Zitat Hoetink AE, Faes TJC, Visser KR, Heethaar RM. On the flow dependency of the electrical conductivity of blood. IEEE Trans Biomed Eng. 2004;51(7):1251–61.CrossRefPubMed Hoetink AE, Faes TJC, Visser KR, Heethaar RM. On the flow dependency of the electrical conductivity of blood. IEEE Trans Biomed Eng. 2004;51(7):1251–61.CrossRefPubMed
22.
Zurück zum Zitat Sifuna MW, Koishi M, Uemura T, Tatekawa H, Haneishi H, Sapkota A, et al. Connector sensors for permittivity-based thrombus monitoring in extracorporeal life support. J Artif Organs. 2020;24(1):15–21.CrossRefPubMed Sifuna MW, Koishi M, Uemura T, Tatekawa H, Haneishi H, Sapkota A, et al. Connector sensors for permittivity-based thrombus monitoring in extracorporeal life support. J Artif Organs. 2020;24(1):15–21.CrossRefPubMed
23.
Zurück zum Zitat Zhao T, Jacobson B, Ribbe T. Triple-frequency method for measuring blood impedance. Physiol Meas. 1993;14:145–56.CrossRefPubMed Zhao T, Jacobson B, Ribbe T. Triple-frequency method for measuring blood impedance. Physiol Meas. 1993;14:145–56.CrossRefPubMed
24.
Zurück zum Zitat Nguyen Huu D, Kikuchi D, Maruyama O, Sapkota A, Takei M. Cole-Cole analysis of thrombus formation in an extracorporeal blood flow circulation using electrical measurement. Flow Meas Instrum. 2017;53:172–9.CrossRef Nguyen Huu D, Kikuchi D, Maruyama O, Sapkota A, Takei M. Cole-Cole analysis of thrombus formation in an extracorporeal blood flow circulation using electrical measurement. Flow Meas Instrum. 2017;53:172–9.CrossRef
25.
Zurück zum Zitat Li J, Sapkota A, Kikuchi D, Sakota D, Maruyama O, Takei M. Red blood cells aggregability measurement of coagulating blood in extracorporeal circulation system with multiple-frequency electrical impedance spectroscopy. Biosens Bioelectron. 2018;112:79–85.CrossRefPubMed Li J, Sapkota A, Kikuchi D, Sakota D, Maruyama O, Takei M. Red blood cells aggregability measurement of coagulating blood in extracorporeal circulation system with multiple-frequency electrical impedance spectroscopy. Biosens Bioelectron. 2018;112:79–85.CrossRefPubMed
26.
Zurück zum Zitat Zhbanov A, Yang S. Effects of aggregation on blood sedimentation and conductivity. PLoS One. 2015;10(6):1–25.CrossRef Zhbanov A, Yang S. Effects of aggregation on blood sedimentation and conductivity. PLoS One. 2015;10(6):1–25.CrossRef
27.
Zurück zum Zitat Zhang A, Kawashima D, Obara H, Takei M. Analysis of influence of fibrinogen concentration on blood dielectproperties by ghz electrical impedance spectroscopy. J Biorheol. 2020;34(1):38–46.CrossRef Zhang A, Kawashima D, Obara H, Takei M. Analysis of influence of fibrinogen concentration on blood dielectproperties by ghz electrical impedance spectroscopy. J Biorheol. 2020;34(1):38–46.CrossRef
28.
Zurück zum Zitat Tran AK, Kawashima D, Sugarawa M, Obara H, Okeyo KO, Takei M. Development of a noise elimination electrical impedance spectroscopy system for single cell identification. Sens Bio-Sensing Res. 2020;30:100381.CrossRef Tran AK, Kawashima D, Sugarawa M, Obara H, Okeyo KO, Takei M. Development of a noise elimination electrical impedance spectroscopy system for single cell identification. Sens Bio-Sensing Res. 2020;30:100381.CrossRef
29.
Zurück zum Zitat Tran AK, Sapkota A, Wen J, Li J, Takei M. Linear relationship between cytoplasm resistance and hemoglobin in red blood cell hemolysis by electrical impedance spectroscopy and eight-parameter equivalent circuit. Biosens Bioelectron. 2018;119:103–9.CrossRefPubMed Tran AK, Sapkota A, Wen J, Li J, Takei M. Linear relationship between cytoplasm resistance and hemoglobin in red blood cell hemolysis by electrical impedance spectroscopy and eight-parameter equivalent circuit. Biosens Bioelectron. 2018;119:103–9.CrossRefPubMed
30.
Zurück zum Zitat Pethig R. Protein-water interactions determined by dielectric methods. Annu Rev Phys Chem. 1992;43(1):177–205.CrossRefPubMed Pethig R. Protein-water interactions determined by dielectric methods. Annu Rev Phys Chem. 1992;43(1):177–205.CrossRefPubMed
31.
Zurück zum Zitat Zhao T. Electrical impedance and haematocrit of human blood with various anticoagulants. Physiol Meas. 1993;14:299.CrossRefPubMed Zhao T. Electrical impedance and haematocrit of human blood with various anticoagulants. Physiol Meas. 1993;14:299.CrossRefPubMed
32.
Zurück zum Zitat Baskurt OK, Uyuklu M, Meiselman HJ. Time course of electrical impedance during red blood cell aggregation in a glass tube: comparison with light transmittance. IEEE Trans Biomed Eng. 2010;57(4):969–78.CrossRefPubMed Baskurt OK, Uyuklu M, Meiselman HJ. Time course of electrical impedance during red blood cell aggregation in a glass tube: comparison with light transmittance. IEEE Trans Biomed Eng. 2010;57(4):969–78.CrossRefPubMed
33.
Zurück zum Zitat Cametti C, Marchetti S, Gambi CMC, Onori G. Dielectric relaxation spectroscopy of lysozyme aqueous solutions: analysis of the δ-dispersion and the contribution of the hydration water. J Phys Chem B. 2011;115(21):7144–53.CrossRefPubMed Cametti C, Marchetti S, Gambi CMC, Onori G. Dielectric relaxation spectroscopy of lysozyme aqueous solutions: analysis of the δ-dispersion and the contribution of the hydration water. J Phys Chem B. 2011;115(21):7144–53.CrossRefPubMed
34.
Zurück zum Zitat Pradhan R, Mitra A, Das S. Impedimetric characterization of human blood using three-electrode based ECIS devices. J Electr Bioimpedance [Internet]. 2012;3(1):12–9.CrossRef Pradhan R, Mitra A, Das S. Impedimetric characterization of human blood using three-electrode based ECIS devices. J Electr Bioimpedance [Internet]. 2012;3(1):12–9.CrossRef
35.
Zurück zum Zitat Asami K. Cell Electrofusion in centrifuged erythrocyte pellets assessed by dielectric spectroscopy. J Membr Biol. 2016;249(1–2):31–9.CrossRefPubMed Asami K. Cell Electrofusion in centrifuged erythrocyte pellets assessed by dielectric spectroscopy. J Membr Biol. 2016;249(1–2):31–9.CrossRefPubMed
Metadaten
Titel
Simultaneous electrical online estimation of changes in blood hematocrit and temperature in cardiopulmonary bypass
verfasst von
Martin Wekesa Sifuna
Daisuke Kawashima
Katsuhiro Matsuura
Hiromichi Obara
Yusuke Nakajima
Masahiro Takei
Publikationsdatum
07.03.2022
Verlag
Springer Nature Singapore
Erschienen in
Journal of Artificial Organs / Ausgabe 4/2022
Print ISSN: 1434-7229
Elektronische ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-022-01320-9

Weitere Artikel der Ausgabe 4/2022

Journal of Artificial Organs 4/2022 Zur Ausgabe

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Deutlich weniger Infektionen: Wundprotektoren schützen!

08.05.2024 Postoperative Wundinfektion Nachrichten

Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.

Chirurginnen und Chirurgen sind stark suizidgefährdet

07.05.2024 Suizid Nachrichten

Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.