Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) Board of Trustees Task Force on Standardized Post Processing
verfasst von:
Jeanette Schulz-Menger, David A Bluemke, Jens Bremerich, Scott D Flamm, Mark A Fogel, Matthias G Friedrich, Raymond J Kim, Florian von Knobelsdorff-Brenkenhoff, Christopher M Kramer, Dudley J Pennell, Sven Plein, Eike Nagel
With mounting data on its accuracy and prognostic value, cardiovascular magnetic resonance (CMR) is becoming an increasingly important diagnostic tool with growing utility in clinical routine. Given its versatility and wide range of quantitative parameters, however, agreement on specific standards for the interpretation and post-processing of CMR studies is required to ensure consistent quality and reproducibility of CMR reports. This document addresses this need by providing consensus recommendations developed by the Task Force for Post Processing of the Society for Cardiovascular MR (SCMR). The aim of the task force is to recommend requirements and standards for image interpretation and post processing enabling qualitative and quantitative evaluation of CMR images. Furthermore, pitfalls of CMR image analysis are discussed where appropriate.
The online version of this article (doi:10.1186/1532-429X-15-35) contains supplementary material, which is available to authorized users.
Competing interests
The authors declare that they have no competing interests.
Authors’ contributions
JSM wrote paragraphs, edited manuscript, corresponding author. DAB, JB, SDF, MAF, MGF, RJK, FvKB, CMK, DJP, SP and EN wrote paragraphs, edited manuscript. All authors read and approved the final manuscript.
Abkürzungen
bSSFP
Balanced steady state free precession
CAD
Coronary artery disease
CMR
Cardiovascular magnetic resonance
FWHM
Full width half maximum
LGE
Late Gadolinium enhancement
LV
Left ventricle
MIP
Maximum intensity projection
MPR
Multiplanar reconstruction
PSIR
Phase sensitive inversion recovery
ROI
Region of interest
RV
Right ventricle
SD
Standard deviation
SI
Signal intensity
STIR
Short-tau inversion recovery
TE
Echo time
TI
Inversion time
TR
Repetition time
TSE
Turbo spin echo
VENC
Velocity encoding
VR
Volume rendering.
Preamble
With mounting data on its accuracy and prognostic value, cardiovascular magnetic resonance (CMR) is becoming an increasingly important diagnostic tool with growing utility in clinical routine. Given its versatility and wide range of quantitative parameters, however, agreement on specific standards for the image interpretation and post processing of CMR studies is required to ensure consistent quality and reproducibility of CMR reports. This document addresses this need by providing consensus recommendations developed by the Task Force for Post Processing of the Society for Cardiovascular MR (SCMR). The aim of the task force is to recommend requirements and standards for image interpretation and post processing enabling qualitative and quantitative evaluation of CMR images. Furthermore, pitfalls of CMR image analysis are discussed where appropriate.
The Task Force is aware that for many of the recommendations, the body of evidence is limited. Thus, this document represents expert consensus providing guidance based on the best available evidence at present as endorsed by the SCMR. As CMR undergoes rapid development, updated recommendations for image acquisition, interpretation and post processing are needed regularly and will be provided by online appendices when needed and updated Task Force papers in due course.
Anzeige
The recommendations are considered for the application of CMR in clinical routine. For some applications, quantification is considered as providing added information but is not mandatory (e.g perfusion imaging), whereas for others quantification is required for all clinical reports (e.g. T2* assessment in iron overload). In general the intention of this task force is to describe, in which scenarios quantitative analysis should be performed and how it is performed.
The recommendations respect societal recommendations for structured reporting of cardiovascular imaging studies in general (ACCF / ACR / AHA / ASE / ASNC / HRS / NASCI / RSNA / SAIP / SCAI / SCCT / SCMR) [1] and in specific for CMR studies (SCMR) [2]. These recommendations will be reviewed and updated regularly and updates made available on the SCMR website.
The recommendations do not supersede clinical judgment regarding the contents of individual interpretation of imaging studies.
The Task Force made every effort to avoid conflicts of interests and, where present, to disclose potential conflicts.
Anzeige
General recommendations
The recommendations listed in this section apply to the acquisition and post processing of all CMR data. CMR studies should be performed for recommended indications [3], respecting published appropriateness criteria [4] and the recently published societal CMR expert consensus document [5]. Any analysis should be performed using uncompressed or lossless compressed DICOM (Digital Imaging and Communications in Medicine) source images. Readers should have adequate training and clinical experience. The identity and responsibility of the reader should be appropriately documented in the report. Data acquisition should conform to the recommendations of SCMR [6].
Furthermore, the reader of clinical data is also responsible for the use of adequate post processing hardware and software. The general requirements include
Workstation and screen of adequate specification and resolution (as per the specifications of the post-processing software)
Post processing software with regulatory approval for use in patients, ideally providing the following tools:
Ability to view all short-axis cines in a single display
Ability to perform endocardial and epicardial contour tracings on short-axis cines
Ability to correct for atrioventricular annular location from the long-axis slice onto the most basal left ventricular (LV) short-axis location in contour tracings
Cross references for confirmation of slice position
Ability to compare cine, late gadolinium enhancement (LGE) and/or perfusion images from the same location simultaneously
Ability to compare short- and long-axis images of the same region simultaneously
Ability to compare images of the approximate same location on the current and prior study simultaneously for longitudinal studies
Ability to perform (semi-) quantitative signal intensity (SI) analysis
Ability to perform standardized segmentation of the myocardium according to the model of the American Heart Association (AHA) [7]
Ability to use baseline-correction or comparison to a phantom for flow measurements can be helpful
Ability to manually correct heart rate, weight, body surface area
Regarding evaluation of angiography the software should provide the following tools:
Subtraction of post-contrast from pre-contrast 3D datasets
3D multiplanar and maximum intensity projection (MIP) capabilities
Volume rendering and surface shaded reconstructions optional for reporting but not mandatory for quantitative analysis
Quantitative diameter analysis based on non-subtracted 3D-MR angiography (MRA)
MIP reconstruction based on non-subtracted or subtracted 3D-MRA datasets
Left ventricular chamber assessment
1. Visual analysis
a)
Before analyzing the details, review all cines in cine mode, validate observations from one plane with the others, and check for artifacts.
b)
Dynamic evaluation of global LV function: Interpretation of both ventricular chambers, in concert with extracardiac structures including assessment for hemodynamic interaction between the two chambers (e.g., shunts, evidence of constrictive physiology).
c)
Assessment of LV function from a global and segmental perspective. Segmental wall motion is based on segmental wall thickening during systole. Wall motion is categorized as: hyperkinetic, normokinetic, hypokinetic, akinetic, dyskinetic.
d)
In presence of segmental wall motion abnormalities, use of standard LV segmentation nomenclature corresponding to the supplying coronary artery territories is recommended [2, 7].
2. Quantitative analysis
a)
General recommendations
i.
Calculated parameters: LV end-diastolic volume, LV end-systolic volume, LV ejection fraction, LV stroke volume, cardiac output, LV mass, and body-surface area indexed values of all except ejection fraction. The parameters quantified may vary depending on the clinical need.
ii.
Evaluation of the stack of short axis images with computer-aided analysis packages.
iii.
Contours of endocardial and epicardial borders at end-diastole and end-systole (Figure 1).
iv.
Epicardial borders should be drawn on the middle of the chemical shift artifact line (when present).
v.
The LV end-diastolic image should be chosen as the image with the largest LV blood volume. For its identification, the full image stack has to be evaluated and one phase has to be identified as end-diastole for all the short axis locations.
vi.
The LV end-systolic image should be chosen as the image with the smallest LV blood volume. For its identification, the full image stack has to be evaluated and one phase has to be identified as end-systole for all the short axis locations.
vii.
Deviations may occur and extra care should be taken in the setting of LV dyssynchrony or severe mitral regurgitation. Aortic valve closure defines end-systole.
viii.
Automatic contour delineation algorithms must be checked for appropriateness by the reader.
b)
LV volumes
i)
Papillary muscles are myocardial tissue and thus ideally should be included with the myocardium. Because not all evaluation tools allow for their inclusion without manual drawing of contours, they are however often included in the volume in clinical practice, which is acceptable. Reference ranges that use the same approach should be used and the inclusion or exclusion of papillary muscles should be mentioned in the report (Figure 1) [8‐10].
ii)
Outflow tract: The LV outflow tract is included as part of the LV blood volume. When aortic valve cusps are identified on the basal slice(s) the contour is drawn to include the outflow tract to the level of the aortic valve cusps.
iii)
Basal descent: As a result of systolic motion of the mitral valve toward the apex (basal descent) care must be taken with the one or two most basal slices. A slice that contains blood volume at end-diastole may include only left atrium (LA) without LV blood volume at end-systole. The LA can be identified when less than 50% of the blood volume is surrounded by myocardium and the blood volume cavity is seen to be expanding during systole. Some software packages automatically adjust for systolic atrioventricular ring descent using cross-referencing from long-axis locations.
c)
LV mass
i)
Calculation: difference between the total epicardial volume (sum of epicardial cross-sectional areas multiplied by the sum of the slice thickness and interslice gap) minus the total endocardial volume (sum of endocardial cross-sectional areas multiplied by the sum of the slice thickness and interslice gap), which is then multiplied by the specific density of myocardium (1.05 g/ml).
ii)
Papillary muscles: Papillary muscles are myocardial tissue and thus ideally should be included with the myocardium, and this is particularly relevant in diseases with LV hypertrophy. However, readers may decide to exclude trabecular tissue and papillary muscles from the myocardial mass. Reference ranges that use the same approach should be used and the inclusion / exclusion of papillary should be specifically mentioned in the report (Figure 1) [8‐10].
iii)
Basal descent and apex: When the most basal slice contains only a small crescent of basal lateral myocardium and no discernable ventricular blood pool, an epicardial contour for the visible myocardium is included for LV mass only. Similarly, when the most apical slice contains only a circle of myocardium without cavitary blood pool, an epicardial contour without an endocardial contour should be drawn for LV mass calculations.
d)
Rapid quantitative analysis
i)
A rapid quantitative analysis can also be performed using rotational long axis views (e.g. 2- and 4-chamber views). In cases without expected significant regional variation of wall motion, this technique allows for faster evaluation and is not limited by problems related to basal descent. When the area-length method is used, with either a single long-axis view or a bi-plane approach, specific mention of the analysis technique should be made in the report.
Single long-axis equation: LV volume = 0.85 × (LV-area)2/ LV-length. This is typically performed using a 4-chamber view with calculations of LV volume obtained on both end-diastolic and end-systolic phases. LV-area is the planimetered area of the LV cavity from an endocardial contour with the base drawn as a straight line through the medial and lateral aspects of the mitral annulus. LV-length is the linear dimension from the midpoint of the mitral annular line to the apical tip of the endocardial contour.
Bi-plane equation: LV volume = 0.85 × (LV-area1 x LV-area2)/ LV-length. Here, both 4-chamber (LV-area1) and 2-chamber [or vertical] (LV-area2) long axis views are used to calculate both end-diastolic and end-systolic volumes, similar to the single long-axis equation.
e)
Cavity diameter and LV wall thickness can be obtained similar to echocardiography using two CMR approaches [12, 14]:
i)
Basal short axis slice: immediately basal to the tips of the papillary muscles;
ii)
3-chamber view: in the LV minor axis plane at the mitral chordae level basal to the tips of the papillary muscles.
iii)
Both approaches have good reproducibility. The 3-chamber view is most comparable to data obtained with echocardiography.
f)
Research:
i)
Quantitative evaluation of LV dynamics (e.g. strain, rotation, time-to-peak velocity) is feasible by several imaging techniques (e.g. tagging, DENSE, tissue phase mapping, cine) and requires specific post-processing software. As research applications are evolving and consensus evidence is being accumulated, the Task Force chooses to refrain from making a dedicated statement at this time.
×
Right ventricular chamber assessment
1. Visual analysis
a)
Before analyzing the details, review all cines in cine mode, validate observations from one plane with the others, and check for artifacts and reliability
b)
Assessment of global and regional right ventricular (RV) function (septal wall, free wall), where appropriate. Categorization should be noted as: hyperkinetic, normokinetic, hypokinetic, akinetic, dyskinetic.
c)
Assessment of both ventricular chambers for hemodynamic interaction (i.e. constrictive physiology)
2. Quantitative analysis
a)
General recommendations
i)
Calculated parameters: RV end-diastolic volume, RV end-systolic volume, RV ejection fraction, RV stroke volume, cardiac output, and body-surface area indexed values of all except ejection fraction. Similar to the LV, the parameters quantified may vary depending on the clinical need [15].
ii)
The contiguous stack of short axis images or transaxial cine images is evaluated with computer-aided analysis packages (Figure 2) [16, 17].
iii)
Transaxial stack of cines covering the RV enable best identification of the tricuspid valve plane.
iv)
Endocardial borders are contoured at end-diastole and end-systole (Figure 2).
v)
The RV end-diastolic image should be chosen as the image with the largest RV blood volume. For its identification, the full image stack has to be evaluated and one phase has to be identified as end-diastole for all short / transaxial locations.
vi)
RV end-systolic image should be chosen as the image with the smallest RV blood volume. For its identification, the full image stack has to be evaluated and one phase has to be identified as end-systole for all short / transaxial locations.
vii)
As for the LV, it may be necessary to review all image slices in the stack to define end-systole.
viii)
The pulmonary valve may be visualized, and contours are included just up to, but not superior to this level.
ix)
Trabeculations of the RV are ignored and a smooth endocardial border is drawn to improve reader reproducibility.
b)
RV volumes
i)
Total volumes are taken as the sum of volumes from individual 2D slices, accounting for any interslice gap and slice thickness. RV trabeculae and papillary muscles are typically included in RV volumes.
c)
RV mass
i)
Usually not quantified in routine assessment.
d)
Confirmation of results
i)
If no intra- or extracardiac shunts are present, the RV and LV stroke volumes should be nearly equal (small differences are seen as a result of bronchial artery supply). Since the LV stroke volume is more reliably determined than the RV stroke volume, the LV data can be used to validate RV data.
×
Post processing of myocardial perfusion imaging
1. Visual analysis
a)
For most clinical indications, visual analysis of myocardial perfusion CMR images is appropriate
b)
Work-flow:
i)
Display rest and stress perfusion images side-by-side. If possible also display corresponding LGE images.
ii)
Adjust window and level: The aim of image adjustment is to set a maximal window width without “overspilling” of the LV cavity signal into the myocardium. Ensure that myocardium before contrast arrival is nearly black and that signal in the RV and LV cavities is bright grey rather than a pure white. Correct level and window settings may require review of both pre- and peak contrast images.
iii)
Apply the same contrast, brightness and window settings to all images of the dynamic series.
iv)
Review series as cines and/or by scrolling through individual images. Some software packages allow the display of only dynamic images during the first myocardial passage.
v)
The key diagnostic feature is contrast arrival and first passage through the LV myocardium.
vi)
Visual analysis allows a comparison between regions to identify relative hypoperfusion. Comparison can be made between endocardial and epicardial regions, between segments of the same slice or between slices.
c)
Compare rest and stress images to identify inducible perfusion defects and artifacts. Note that unlike nuclear perfusion methods, in CMR the finding of a “fixed perfusion defect” on rest and stress perfusion images is not the preferred method to identify myocardial scar. Instead, scar should be identified from LGE images.
d)
Criteria for an inducible perfusion defect (Figure 3a):
i)
Occurs first when contrast arrives in LV myocardium
ii)
Persists beyond peak myocardial enhancement and for several RR intervals (usually >4)
iii)
Is more than one pixel wide
iv)
Is usually most prominent in the subendocardial portion of the myocardium
v)
Often manifests as a transmural gradient across the wall thickness of the segment involved: densest in the endocardium and gradually becoming less dense towards the epicardium
vi)
Over time, defect regresses towards the subendocardium
vii)
Is present at stress but not at rest
viii)
Conforms to the distribution territory of one or more coronary arteries.
e)
Interpret location and extent of inducible perfusion defect(s) using AHA segment model [2, 7].
i)
Estimate number of segments involved
ii)
Comment on transmurality of perfusion defect
iii)
Indicate extent of perfusion defect relative to scar on LGE
f)
Pitfalls of visual analysis
i)
Dark banding artifacts (Figure 3b): A common source of false-positive reports are subendocardial dark banding artifacts [18]. These artifacts
typically occur first and are most prominent when contrast arrives in the LV blood pool, i.e. before contrast arrival in the LV myocardium, depending on applied sequence
lead to a reduction in signal compared with baseline myocardial signal (whereas a true perfusion defect will always show an increase in signal compared with the baseline even if this increase is small). These subtle differences can be hard to appreciate visually. It can therefore be helpful to draw a region of interest (ROI) around the suspected artifact and display its signal-intensity-time profile.
persist only transiently before the peak myocardial contrast enhancement, often for less than approximately 6 RR intervals
appear predominantly in the phase-encoding direction
are approximately one pixel wide
Dark banding present at stress and at rest with no corresponding scar on LGE images is also indicative of an artifact [19]. Note however that differences in heart rate and baseline contrast can change the appearance and presence of dark banding between stress and rest perfusion images. Absence of dark banding at rest with typical dark banding at stress should therefore not on its own be considered diagnostic for an inducible perfusion defect.
ii)
Multi-vessel disease: as visual analysis is a relative assessment of perfusion within an imaged section of the heart, the presence of balanced multi-vessel disease can result in most or all of the imaged section appearing hypoperfused. This can lead to false-negative readings and needs to be considered in relevant clinical circumstances. On visual analysis, a clear endocardial to epicardial signal gradient may be seen in multi-vessel disease [20]. Quantitative analysis of the dynamic perfusion data may be of further help to detect globally reduced myocardial perfusion reserve in multi-vessel disease.
iii)
Microvascular disease: Diseases that affect the myocardial microvasculature (e.g. diabetes mellitus, systemic hypertension) may lead to concentric reduction in perfusion [21‐24]. This can lead to apparent false-positive readings relative to angiographic methods and needs to be considered in relevant clinical circumstances. Features suggesting microvascular disease are the presence of concentric LV hypertrophy and a concentric, often subendocardial perfusion defect crossing coronary territories. Differentiation from multi-vessel disease can be challenging.
iv)
When performing a stress first-rest second perfusion protocol, caution is needed when a defect over a region of infarction may be misinterpreted as reversible. Contrasting the extent and location of reversibility by perfusion imaging with LGE is important to avoid over-calling of segmental ischemia [25].
v)
If vasodilator stress during data acquisition was inadequate, visual analysis may lead to false-negative interpretation [26]. Quantitative analysis of the dynamic perfusion data may be of further help to detect globally reduced myocardial perfusion reserve in case of inadequate vasodilator stress.
vi)
Signal intensity may vary depending on the distance of myocardium from the surface coil and may lead to misinterpretation if not considered in the analysis. These problems are less likely if acquisition is corrected for coil sensitivity.
×
Anzeige
2. Research tools / Quantitative analysis
a)
Objective description of SI change in myocardial perfusion CMR studies can be performed. Several methods have been described for this purpose. In clinical practice, these are rarely required, but they may supplement visual analysis for example in suspected multi-vessel disease or suspected inadequate response to vasodilator stress. Quantitative analysis is also frequently used in research studies.
b)
Requirements:
i)
Validation and definition of a normal range with the specific pulse sequence and contrast regime used for data acquisition. If only a comparison between regions of the same study is made, establishing a normal range is less relevant.
ii)
Consideration of contrast dosage at time of acquisition (high doses are more likely to lead to saturation effects in particular of the arterial input function).
c)
Semi-quantitative analysis:
i)
Analysis methods that describe characteristics of the SI profile of myocardial perfusion CMR studies without estimating myocardial blood flow are typically referred to as “semi-quantitative analysis methods”.
ii)
Work-flow:
Select an image from the dynamic series with good contrast between all cardiac compartments (some post-processing tools generate an average image of the series).
Outline LV endocardial and epicardial contours on this image (manual or automated) (Figure 3c).
Propagate contours to all other dynamic images.
Avoid myocardial fat in the ROI
Correct contour position for through-plane motion (some analysis packages register images prior to contours being outlined).
Depending on the type of analysis to be performed, place a separate ROI in the LV blood pool. Preferably the basal slice is used. Exclude papillary muscles from the ROI.
Select a reference point in the LV myocardium for segmentation(usually the RV insertion point).
Segment LV myocardium according to AHA classification [7].
Generate SI / time profiles for myocardial segments +/- LV blood pool.
Consider generating division into endocardial and epicardial layers and repeat analysis.
iii)
Frequently used semi-quantitative analysis methods (see [27] for detailed review):
Maximal upslope of the myocardial SI profile, may be normalized to LV upslope [28]
Ratio of stress/rest ratios for the above (often referred to as “myocardial perfusion reserve index”) [30]
The upslope integral (area under the signal intensity-time curve) [31].
iv)
Limitations of semi-quantitative analysis methods:
SI may vary according to distance from coil
No absolute measurement of myocardial blood flow derived
d)
Quantitative analysis
i)
Analysis methods that process the SI profile of myocardial perfusion CMR studies to derive estimates of myocardial blood flow are typically referred to as “quantitative analysis methods”. See [27] for review.
ii)
Requirements:
It is a prerequisite for reliable quantification that data acquisition used an appropriate pulse sequence and contrast regime.
An “input function” for analysis of the myocardial tissue response can be derived from the ROI in the LV blood pool. In order to reduce saturation effects in the blood pool, a “dual-bolus” myocardial perfusion regime may be used, in which the input function is derived from a pre-bolus with a small contrast agent concentration.
The contrast dose and delivery need to be chosen to minimize saturation effects also the myocardium and typically contrast doses required for reliable quantitative analysis are lower than those optimized for visual analysis in clinical routine.
A temporal resolution of 1-2 RR intervals is required.
iii)
Work-flow:
Typically, the same source data as for semi-quantitative analysis are used
Further post-processing may then take place on the same or a separate off-line workstation.
iv)
Several analysis methods have been described, including:
Post processing of late gadolinium enhancement studies
1. Visual assessment
a)
For most clinical indications, visual assessment of LGE images is sufficient
b)
Work-flow:
i)
Modify image window and level so that:
Noise is still detectable (nulled myocardium should not be a single image intensity)
LGE regions are not clipped (LGE regions should not be a single image intensity)
ii)
Note if normal myocardium has a faint “etched” appearance (darkest at the border with slightly higher image intensity centrally), this signifies an inversion time that was set too short and will lead to underestimation of the true extent of LGE (Figure 4). In general, an inversion time that is slightly too long is preferred to one that is slightly too short [34].
c)
Criteria for presence of LGE
i)
High SI area that may be as bright as the LV blood pool
ii)
Rule out artifacts (see below)
d)
Assess pattern of LGE
i)
Coronary artery disease (CAD) type: Should involve the subendocardium and be consistent with a coronary artery perfusion territory
ii)
Non-CAD-type: Usually spares the subendocardium and is limited to the mid-wall or epicardium, although non-CAD-type should be considered if subendocardial involvement is global [35, 36].
e)
Interpret location and extent using AHA 17-segment model [7].
i)
Comparison of LGE images should be made with cine and perfusion images (if the latter are obtained) to correctly categorize ischemia and viability [25]
ii)
Estimate average transmural extent of LGE within each segment (0%, 1-25%, 26-50%, 51-75%, 76-100%) [34].
iii)
In patients with acute myocardial infarction, include subendocardial and mid-myocardial hypoenhanced, no-reflow zones as part of infarct size
f)
Pitfalls
i)
Check for artifacts
Verify regions with LGE in at least one other orthogonal plane and/or in the same plane obtain a second image after changing the direction of readout
Bright ghosting artifacts can result from poor ECG gating, poor breath-holding, and long T1 species in the imaging plane (e.g. cerebrospinal fluid, pleural effusion, gastric fluid, etc.) [37]
ii)
On non-PSIR (phase sensitive inversion recovery) images, tissue with long T1 (regions below the zero-crossing) may appear enhanced [34, 38].
iii)
Occasionally, it can be difficult to distinguish no-reflow zones or mural thrombus from viable myocardium. Post-contrast cine imaging may be helpful in this regard.
iv)
In case of reduced contrast, the interpretation of additional sequences may be necessary.
×
2. Quantitative analysis
a)
Quantitative analysis is primarily performed to measure LGE extent and/or “grey-zone” extent for research purposes. Subjective visual assessment is still a prerequisite to identify poor nulling, artifacts, no-reflow zones, etc, and to draw endocardial and epicardial borders.
b)
Multiple different methods of delineating LGE extent are described in the literature, including: manual planimetry, the “n”-SD technique, and the full width half maximum (FWHM) technique (see 3) [39‐42].
c)
As the research applications are evolving and consensus evidence is being accumulated, the Task Force chooses to refrain from making a dedicated statement at this time regarding the optimal method for quantitative assessment
Anzeige
3. Research tools / Quantitative analysis
a)
Quantification of LGE extent:
i)
Manual planimetry:
Outline endocardial and epicardial borders
Manual planimetry of LGE regions in each slice
Summation of LGE areas
Multiplication of total LGE area with slice thickness plus interslice gap as well as specific gravity of myocardium provides the approximate LGE weight, which can be used to calculate the ratio of LGE to normal myocardium
considered subjective
ii)
The “n”-SD technique:
Outlining of endocardial and epicardial borders for the myocardial ROI.
Selection of a normal “remote” (dark) region ROI within the myocardium to define the reference SI (mean and standard deviation, SD). This subjective approach can affect measurements.
Is susceptible to spatial variations in surface coil sensitivity.
Selection of a threshold between normal myocardium and LGE. The relative SNR of scar tissue versus normal myocardium can vary dependent on contrast agent type, dose and time after injection, field strength, type of sequence and other variables including the underlying injury itself. As such, there is no cutoff value, which works for all situations and usually manual tracing is performed as the standard of truth. But (semi-)automated thresholding may improve reproducibility after adequate standardization. As a starting point for semiautomatic thresholding we recommend n+5SD for infarction and n+3SD for myocarditis.
The presence of LGE within the myocardium is then determined automatically
requires manual corrections to include no-reflow zones and to exclude artifacts and LV blood pool (errors in the endocardial contour)
iii)
FWHM technique:
Outlining of endocardial and epicardial borders for the myocardial ROI
Uses the full width of the myocardial ROI SI histogram at half the maximal signal within the scar as the threshold between normal myocardium and LGE
Determination whether LGE is present or not, and, if LGE is present, selection of a ROI that includes the “maximum” signal. This subjective selection can affect measurements.
Is also susceptible to spatial variations in surface coil sensitivity, albeit perhaps less so than the “n”-SD technique [40].
Considered more reproducible than the n-SD technique [42].
Since technique assumes a bright LGE core, may be less accurate than the “n”-SD technique if LGE is patchy or grey [43]
Requires manual corrections to include no-reflow zones and to exclude artifacts and LV blood pool (errors in the endocardial contour)
may be helpful in identifying diffuse myocardial fibrosis
may provide a quantitative assessment of the extent of fibrosis
multiple sequences and imaging protocols are described in the literature
as the research application(s) are evolving and consensus evidence is being accumulated, the Task Force chooses to refrain from making a specific recommendation at this time
Post processing of T2-weighted imaging
1. Visual analysis
a)
The visual analysis should aim for detecting or excluding regions with significant SI increase, indicating increased free water content.
b)
Qualitative, visual analysis of myocardial SI may be sufficient for diseases with regional injury to the myocardium such as acute coronary syndromes/infarction (Figure 5), early stages of myocarditis and sarcoidosis.
c)
Work-flow:
i)
Identify and display appropriate image(s)
ii)
Modify image contrast and brightness to minimize SI in a background noise area (noise should still be detectable) and to reduce the maximal SI displayed to an area with the highest SI without allowing for “over-shining” with erroneous display of pixels as white
iii)
Check for artifacts
d)
Criteria for edema:
i)
Clearly detectable high SI area
ii)
Respecting anatomical borders
iii)
Following an expected regional distribution pattern (mainly subendocardial, transmural, mainly subepicardial, focal)
iv)
Verifiable in two perpendicular views
e)
High SI areas suggestive of myocardial edema should be compared to
i)
regional function
ii)
other tissue pathology such as irreversible injury: scar/fibrosis, infiltration
f)
Pitfalls of visual analysis:
i)
Surface coil reception field inhomogeneity: The uneven distribution of the sensitivity of the receiver in surface coil may lead to false low SI in segments most distant to the coil surface or false high SI in segments closest to the coil surface, especially in dark-blood triple-inversion recovery spin echo (STIR, TIRM) images. Therefore, the body coil or a reliable and accurate correction algorithm should be used to ensure a homogeneous signal reception.
ii)
Low SI artifacts: Arrhythmia or through-plane motion of myocardium may cause artifacts, making large areas appear with false low SI, especially in dark-blood triple-inversion recovery spin echo images.
iii)
High SI artifacts: In dark-blood triple-inversion recovery spin echo images, slow flowing blood may lead to insufficient flow suppression and results in high SI, which may be confused with myocardial edema.
×
2. Semi-quantitative analysis
a)
Because low SI artifacts can lead to SI distribution patterns similar to extensive myocardial edema, a mere visual analysis may lead to incorrect results. SI quantification with reference regions is much less sensitive to these errors and therefore is recommended.
b)
Requirements:
i)
Tested normal values for SI values or ratios.
c)
Work-flow
i)
Global SI analysis:
Outline LV endocardial and epicardial contours.
For the T2 SI ratio, draw the contour for a ROI in a large area of the skeletal muscle closest to the heart and to the center of the reception field of the coil (for short axis views preferably in the M. serratus anterior).
ii)
Regional SI analysis:
Draw the contour for a ROI in the affected area and divide the SI by that of the skeletal muscle.
iii)
While a cut-off of 1.9 can be used for dark-blood triple-inversion recovery spin echo [50], a locally established value is recommended, because SI and ratio values may vary between sequence settings (especially echo time (TE)) and scanner models. For these images, a color-coded map, based on the parametric calculation and display of myocardial pixels with a SI ratio of 2 or higher, can also be used.
can account for spatial variations in coil sensitivity
multiple sequences and imaging protocols are described in the literature
may provide quantitative measurements of edema
as research application(s) are evolving and consensus evidence is being accumulated, the Task Force chooses to refrain from making a dedicated statement at this time
Post processing of T2* imaging
1. Visual analysis
T2* imaging always requires a quantitative analysis. Visual analysis is used to ensure optimal image quality, which is the most important factor in the accuracy of data analysis.
2. Quantitative analysis
a)
Evaluation of T2* always requires a quantitative analysis using software with regulatory approval for T2* evaluation in patients
b)
Full thickness ROI defined of the ventricular septum
i)
Take care to avoid blood pool and proximal blood vessels
ii)
Septal ROI avoids susceptibility artifact from tissue interfaces
c)
Mean myocardial SI from the ROI is plotted against TE (Figure 6)
i)
SI falls with increasing TE
ii)
A mono-exponential curve is fitted to the data
iii)
The time for the decay of SI falls (shorter T2*) with increasing iron burden
iv)
In heavily iron overloaded patients, SI for higher TEs may fall below background noise causing the curve to plateau and underestimating T2*.
v)
This can be compensated for by:
Truncating the curve by removing later echo times (Figure 6e) [52, 53]
This issue is not significant when using the double inversion recovery (black blood) sequence [54]
T2* <10ms indicates increased risk of development of heart failure [57]
e)
CMR assessment of T2* at 3T for assessment of iron overload cardiomyopathy cannot be recommended at this time. T2* shortens with increasing field strength making assessment of severe iron overload more problematic, and there is a lack of clinical verification.
×
Flow image interpretation and post processing
1. Visual analysis
a)
Appropriately aligned acquisitions of cines and stacks of cines can give valuable information on flow in relation to adjacent structures, notably on the directions, time courses and approximate dimensions of jets resulting from valve regurgitation, stenoses or shunts. Such information can be important in assessing the credibility of measurements of flow, which may be subject to several possible sources of error. Gradient echo cines differ somewhat from SSFP in terms of degrees of signal augmentation or reduction attributable to flow effects. Of note, SSFP can provide clear delineation between the relatively bright signal from voxels aligned within the coherent core of a jet, and low signal from the shear layer that bounds such a jet core. In- or through-plane phase contrast flow velocity acquisitions can also provide visual information on the directions, dimensions and time courses of flow; it can also image morphology, which can yield a clue to the etiology of an abnormal jet (e.g. imaging accelerated flow jet of a coarctation or of valve insufficiency, or flow direction across an atrial septal defect or Fontan fenestration (Figure 7)) [58, 59].
b)
Pitfalls:
i)
Flow appearances on both cine and phase encoded acquisitions are highly dependent on image location and orientation, especially in the case of jet flow
ii)
If the range of velocity encoding (VENC) is set too high, visualization of the jet may not be obtained. If it is set too low, a mosaic pattern on the images will be visualized [60].
iii)
If slice thickness is too large on in-plane velocity mapping, the higher velocities will be “averaged out” with the lower velocities and stationary tissue; jets and flow may not be visualized correctly.
iv)
If the annulus of valves is very dynamic or the imaging plane is not set correctly, the valve morphology may not be visualized.
v)
If imaging in the presence of metal containing devices, signal loss may be present as artifact and interpretation must proceed with caution.
vi)
The TE should be as low as possible for increased accuracy, especially with high velocity turbulent jets; this should be kept to 3.5 ms or lower [61].
×
2. Quantitative analysis
a)
Work-flow:
i)
Load phase and magnitude images into software. Window the magnitude and phase images to the appropriate brightness and contrast so that the borders of the ROI are sharp.
ii)
Examine the images to ensure the quality is sufficient and that the VENC was not exceeded, or there was little contrast (ie the VENC was too high).
iii)
Trace the borders of the vessel of interest on each phase and magnitude image so that only the cavity of the vessel is included (Figure 8); make sure the noise outside the vessel is not included. Check that this is performed correctly on the magnitude images always keeping in mind that it is the phase images that contain the encoded information.
iv)
Baseline-correction or comparison to a phantom for flow measurements may be considered [62, 63].
v)
Directly calculated parameters include: antegrade volume, retrograde volume, peak velocity and mean velocity
vi)
Derived parameters include:
Net volume [ml| = antegrade volume - retrograde volume
Cardiac output (liters/min = (net volume [ml] x heart rate [beats/minute])/1000) and cardiac index (cardiac output/BSA) when integrating heart rate and body surface area
Regional flow to both lungs by measuring cardiac output in each branch pulmonary artery (eg percentage of flow to the right lung = (right pulmonary artery flow / right pulmonary artery flow + left pulmonary artery flow) × 100).
Regurgitant volumes of the atrioventricular valves may be obtained by either of 2 methods: A) direct measurement of diastolic flow across the valve and subtraction of systolic forward flow across the associated semilunar valve or B) measurement of stroke volume using cine CMR and subtraction of forward flow across the associated semilunar valve.
b)
Pitfalls:
i)
On the phase images, the area of flow may be slightly larger than the area of the magnitude images.
ii)
If the VENC is exceeded, most software packages allow for moving/changing the “dynamic range” of the images so that the VENC is not exceeded. For example, if the peak velocity in the aorta is 175 cm/s and the VENC was set at 150 cm/s, the dynamic range is between -150 cm/s and +150 cm/s (i.e. 300 cm/s). This may be moved to -100 cm/s and +200 cm/s to account for this accelerated velocity. This will be demonstrated on the graph of the velocity where the phase in which the VENC is exceeded does not “clip” (appears to go the wrong way) after correction.
iii)
In general, the area that exceeds the VENC in the ROI is in the center of the vessel and not at the edges; if at the edges, it is usually (but not always) outside the vessel.
iv)
If imaging in the presence of devices, signal loss may be present as artifact and interpretation must proceed with caution.
v)
When measuring peak velocity, some software packages will determine the peak velocity in one pixel in the ROI whereas others may take the peak velocity of the average of a few adjacent pixels in the ROI. By reporting the peak velocity in a single pixel, noise may make this measurement inaccurate. By reporting this as an average of a few adjacent pixels, noise is less of an issue, however, the true “peak velocity” may be higher than the reported value. These factors must be kept in mind and interpretation may need to be adapted to the measurement technique used.
vi)
When attempting to measure peak velocity using through plane velocity mapping along a vessel, interpretation should be tempered by the notion that this parameter may be an underestimate as the true peak velocity lies somewhere along the vena contracta; the through plane velocity map may not have been obtained at the level of the true peak velocity. If the vena contracta is itself narrow or ill defined, jet velocity mapping is unlikely to be possible
vii)
Peak velocity is only minimally affected by small background phase offsets, while volume measurements can be dramatically affected by even a small background phase offset due to the cumulative aspect of integration overspace (within the ROI) and time (over the cardiac cycle). Dilatation of a great vessel tends to increase error of this type [64].
viii)
Orientation of the image plane perpendicular to flow direction can have a significant impact on peak velocity measurement, while not significantly affecting volume flow [65].
×
3. Research tools
a)
4D flow: The utility of this approach is the subject of ongoing research.
b)
Real time velocity mapping: The utility and post processing algorithm best applied to this approach is the subject of ongoing research.
Post processing of angiography of thoracic aorta, pulmonary arteries and veins
1. Visual analysis
a)
Maximum intensity projections (MIP) for first review of 3D data and for demonstration purposes (Figure 9A). Volume rendered (VR) techniques may be used for demonstration purposes, but not for detailed analysis.
Screen for atypical insertion and small accessory veins.
e)
Coronary arteries:
i)
Coronary MRA may play a role in assessment of congenital anomalies but not usually in the context of ischemic heart disease. The course of coronary arteries is best assessed on source images, multiplanar reconstructions (MPR) or targeted MIP reconstructions.
×
2. Quantitative analysis
a)
Aorta:
i)
Multilevel measurements of aortic diameters on double oblique multiplanar images perpendicular to blood flow at standardized levels (Figure 10) [66]. Measurements should be obtained at diastole if possible.
ii)
Inner diameter (lumen width). In the presence of wall thickening (e.g. thrombus or intramural hematoma) outer diameter including vessel walls should also be reported.
iii)
Diameters of sinuses or sinotubular junction may not be measured on ungated images since motion artifacts can lead to blurring and may result in diameter under- or overestimation. These require ECG gated acquisitions, either from 3-dimensional SSFP acquired in late diastole, or from a contiguous stack of cines aligned to transect the axis of the aortic root. Consistent methods of acquisition and measurement are essential for the evaluation of any change over time, for example by the measurement in late diastole, of all three sinus-commissure dimensions of the aortic root, which may dilate asymmetrically.
iv)
Standardized report including table of diameters.
b)
Pulmonary artery:
i)
The widest inner diameter is measured perpendicular to the long axis of the main pulmonary artery at the level of the pulmonary bifurcation in transaxial slices
c)
Pulmonary veins:
i)
MPR of pulmonary veins perpendicular to blood flow for diameter measurements.
×
Acknowledgements
Jeanette Schulz-Menger: Institutional research support from Siemens Healthcare ,Circle Cardiovascular Imaging Inc and Medis. Consultant for Bayer Healthcare.
Christopher M. Kramer: Dr. Kramer acknowledges research support from Siemens Healthcare
Florian von Knobelsdorff-Brenkenhoff: Dr. von Knobelsdorff-Brenkenhoff acknowledges research support from the Else Kröner-Fresenius-Stiftung, Germany
Scott D. Flamm: Dr. Flamm acknowledges institutional grant/research support from Siemens Medical Solutions and Philips Healthcare, and acts as a Consultant/Advisory Board Member for Bayer Healthcare and Circle Cardiovascular, Inc.
Scott D. Flamm: Consultant for Bayer Healthcare. Institutional research support from Siemens Healthcare and Philips Healthcare.
David A. Bluemke: ---
Sven Plein: Dr. Plein acknowledges research support from Philips Healthcare
Raymond J. Kim: 1) Dr. RJ Kim is an inventor on a United States patent, owned by Northwestern University, concerning the use of contrast-enhanced MRI to detect myocardial viability. 2) Dr. RJ Kim has an education grant from Siemens Medical Solutions. 3) Dr. RJ Kim is a co-founder of HeartIT, LLC
Matthias G. Friedrich: Dr. Friedrich is partially funded by the Canadian Foundation for Innovation and the Fonds de Recherche Santé Québec
Matthias G. Friedrich: Matthias G. Friedrich is on the board of directors and shareholder of Circle Cardiovascular Imaging Inc., the manufacturer of a CMR-post processing and evaluation software
Dudley J. Pennell: This work was supported by the NIHR Cardiovascular Biomedical Research Unit of Royal Brompton & Harefield NHS Foundation Trust and Imperial College.
Dudley J. Pennell: Consultant Siemens, Apotex Novartis. Director and stockholder, Cardiovascular Imaging Solutions.
Mark A. Fogel: Two NIH RO1 grants, grant from Siemens to develop functional fetal cardiac MR, grant from Edwards Life Sciences - CMR Core lab for COMPASSION trial and Kereos - Medical Monitor for P19 imaging agent.
Jens Bremerich: ---
Eike Nagel: Dr. Nagel acknowledges financial support from the Department of Health via the National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award to Guy’s & St Thomas’ NHS Foundation Trust in partnership with King’s College London and King’s College Hospital NHS Foundation Trust; the Centre of Excellence in Medical Engineering funded by the Wellcome Trust and EPSRC under grant number WT 088641/Z/09/Z; Funded by the British Heart Foundation award RE/08/003. Grant support from Philips HealthcareGrant support from Bayer Healthcare.
* Endorsed by the board of trustees of the Society for Cardiovascular Magnetic Resonance:
Andrew Arai, MD ; Albert de Roos ; Orlando Simonetti, PhD; Victor A. Ferrari, MD ; Scott D. Flamm, MD, MBA ; David A. Bluemke, MD, PhD; Jens Bremerich, MD; Ralph Gentry, RT, R, MR, CT ; Frederick H. Epstein, PhD; Raymond Kwong, MD; Edward T. Martin, MD; Dudley J. Pennell, MD; Sven Plein, MD, PhD; Andrew J. Powell, MD; Jeanette Schulz-Menger, MD; Joseph Selvanayagam, MBBS, FRACP; Matthias Stuber, PhD.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Competing interests
The authors declare that they have no competing interests.
Authors’ contributions
JSM wrote paragraphs, edited manuscript, corresponding author. DAB, JB, SDF, MAF, MGF, RJK, FvKB, CMK, DJP, SP and EN wrote paragraphs, edited manuscript. All authors read and approved the final manuscript.
Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) Board of Trustees Task Force on Standardized Post Processing
verfasst von
Jeanette Schulz-Menger David A Bluemke Jens Bremerich Scott D Flamm Mark A Fogel Matthias G Friedrich Raymond J Kim Florian von Knobelsdorff-Brenkenhoff Christopher M Kramer Dudley J Pennell Sven Plein Eike Nagel
Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.
Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.
Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.
Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.
Update Radiologie
Bestellen Sie unseren Fach-Newsletterund bleiben Sie gut informiert.