Tensile strength and failure load of sutures for robotic surgery

verfasst von: Ahmad Abiri, Omeed Paydar, Anna Tao, Megan LaRocca, Kang Liu, Bradley Genovese, Robert Candler, Warren S. Grundfest, Erik P. Dutson

Erschienen in: Surgical Endoscopy | Ausgabe 8/2017

Einloggen, um Zugang zu erhalten

Abstract

Background

Robotic surgical platforms have seen increased use among minimally invasive gastrointestinal surgeons (von Fraunhofer et al. in J Biomed Mater Res 19(5):595–600, 1985. doi:10.​1002/​jbm.​820190511). However, these systems still suffer from lack of haptic feedback, which results in exertion of excessive force, often leading to suture failures (Barbash et al. in Ann Surg 259(1):1–6, 2014. doi:10.​1097/​SLA.​0b013e3182a5c8b8​). This work catalogs tensile strength and failure load among commonly used sutures in an effort to prevent robotic surgical consoles from exceeding identified thresholds. Trials were thus conducted on common sutures varying in material type, gauge size, rate of pulling force, and method of applied force.

Methods

Polydioxanone, Silk, Vicryl, and Prolene, gauges 5-0 to 1-0, were pulled till failure using a commercial mechanical testing system. 2-0 and 3-0 sutures were further tested for the effect of pull rate on failure load at rates of 50, 200, and 400 mm/min. 3-0 sutures were also pulled till failure using a da Vinci robotic surgical system in unlooped, looped, and at the needle body arrangements.

Results

Generally, Vicryl and PDS sutures had the highest mechanical strength (47–179 kN/cm2), while Silk had the lowest (40–106 kN/cm2). Larger diameter sutures withstand higher total force, but finer gauges consistently show higher force per unit area. The difference between material types becomes increasingly significant as the diameters decrease. Comparisons of identical suture materials and gauges show 27–50% improvement in the tensile strength over data obtained in 1985 (Ballantyne in Surg Endosc Other Interv Tech 16(10):1389–1402, 2002. doi:10.​1007/​s00464-001-8283-7). No significant differences were observed when sutures were pulled at different rates. Reduction in suture strength appeared to be strongly affected by the technique used to manipulate the suture.

Conclusions

Availability of suture tensile strength and failure load data will help define software safety protocols for alerting a surgeon prior to suture failure during robotic surgery. Awareness of suture strength weakening with direct instrument manipulation may lead to the development of better techniques to further reduce intraoperative suture breakage.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
2.
3.
Zurück zum Zitat Ballantyne GH (2002) Robotic surgery, telerobotic surgery, telepresence, and telementoring: review of early clinical results. Surg Endosc Other Interv Tech 16(10):1389–1402. doi:10.1007/s00464-001-8283-7 CrossRef Ballantyne GH (2002) Robotic surgery, telerobotic surgery, telepresence, and telementoring: review of early clinical results. Surg Endosc Other Interv Tech 16(10):1389–1402. doi:10.​1007/​s00464-001-8283-7 CrossRef
11.
Zurück zum Zitat Marucci DD, Shakeshaft AJ, Cartmill JA, Cox MR, Adams SG, Martin CJ (2000) Grasper trauma during laparoscopic cholecystectomy. Aust N Z J Surg 70(8):578–581CrossRefPubMed Marucci DD, Shakeshaft AJ, Cartmill JA, Cox MR, Adams SG, Martin CJ (2000) Grasper trauma during laparoscopic cholecystectomy. Aust N Z J Surg 70(8):578–581CrossRefPubMed
13.
Zurück zum Zitat Heijnsdijk EAM, van der Voort M, de Visser H, Dankelman J, Gouma DJ (2003) Inter- and intraindividual variabilities of perforation forces of human and pig bowel tissue. Surg Endosc 17(12):1923–1926. doi:10.1007/s00464-003-9002-3 CrossRefPubMed Heijnsdijk EAM, van der Voort M, de Visser H, Dankelman J, Gouma DJ (2003) Inter- and intraindividual variabilities of perforation forces of human and pig bowel tissue. Surg Endosc 17(12):1923–1926. doi:10.​1007/​s00464-003-9002-3 CrossRefPubMed
14.
Zurück zum Zitat Akinbiyi T, Reiley CE, Saha S et al (2006) Dynamic augmented reality for sensory substitution in robot-assisted surgical systems. Annu Int Conf IEEE Eng Med Biol Proc. doi:10.1109/IEMBS.2006.259707 Akinbiyi T, Reiley CE, Saha S et al (2006) Dynamic augmented reality for sensory substitution in robot-assisted surgical systems. Annu Int Conf IEEE Eng Med Biol Proc. doi:10.​1109/​IEMBS.​2006.​259707
18.
Zurück zum Zitat Cundy TP, Gattas NE, Yang G-Z, Darzi A, Najmaldin AS (2014) Experience related factors compensate for haptic loss in robot-assisted laparoscopic surgery. J Endourol. doi:10.1089/end.2013.0671 PubMed Cundy TP, Gattas NE, Yang G-Z, Darzi A, Najmaldin AS (2014) Experience related factors compensate for haptic loss in robot-assisted laparoscopic surgery. J Endourol. doi:10.​1089/​end.​2013.​0671 PubMed
21.
Zurück zum Zitat Paydar OH, Wottawa CR, Fan RE et al (2012) Fabrication of a thin-film capacitive force sensor array for tactile feedback in robotic surgery. In: Proceedings of the annual international conference of the IEEE engineering in medicine and biology society, EMBS, pp 2355–2358. doi:10.1109/EMBC.2012.6346436 Paydar OH, Wottawa CR, Fan RE et al (2012) Fabrication of a thin-film capacitive force sensor array for tactile feedback in robotic surgery. In: Proceedings of the annual international conference of the IEEE engineering in medicine and biology society, EMBS, pp 2355–2358. doi:10.​1109/​EMBC.​2012.​6346436
22.
Zurück zum Zitat Karabulut R, Sonmez K, Turkyilmaz Z, Bagbanci B, Basaklar AC, Kale N (2010) An in vitro and in vivo evaluation of tensile strength and durability of seven suture materials in various pH and different conditions: an experimental study in rats. Indian J Surg 72(5):386–390. doi:10.1007/s12262-010-0158-5 CrossRefPubMedPubMedCentral Karabulut R, Sonmez K, Turkyilmaz Z, Bagbanci B, Basaklar AC, Kale N (2010) An in vitro and in vivo evaluation of tensile strength and durability of seven suture materials in various pH and different conditions: an experimental study in rats. Indian J Surg 72(5):386–390. doi:10.​1007/​s12262-010-0158-5 CrossRefPubMedPubMedCentral
Metadaten
Titel
Tensile strength and failure load of sutures for robotic surgery
verfasst von
Ahmad Abiri
Omeed Paydar
Anna Tao
Megan LaRocca
Kang Liu
Bradley Genovese
Robert Candler
Warren S. Grundfest
Erik P. Dutson
Publikationsdatum
07.12.2016
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 8/2017
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-016-5356-1

Weitere Artikel der Ausgabe 8/2017

Surgical Endoscopy 8/2017 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.