Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry

verfasst von: Annette Lösch, Alexander Scheuerle, Volker Rupp, Gerd Auffarth, Matthias Becker

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 4/2005

Einloggen, um Zugang zu erhalten

Abstract

Background

A recently developed digital tonometer for transpalpebral intraocular pressure (IOP) measurement, distributed by Corneal, Inc., allows the noninvasive measurement of IOP for screening purposes.

Method

We measured the IOP of 218 eyes in 109 patients of the Interdisciplinary Uveitis Center of the University of Heidelberg with intact corneal epithelium. IOPs were measured first with the TGDc-01 tonometer, and then by means of Goldmann tonometry. IOPs were recorded by two independent examiners. The mean of three measurements obtained with the TGDc-01 was taken, whereas only one measurement was performed with the Goldmann tonometer.

Results

The mean difference between the TGDc-01 and Goldmann measurements was 3.7 mmHg. The standard deviation of the differences was ±4.06 mmHg. Thus measurements acquired with the TGDc-01 may range 4.4 mmHg above or 11.8 mmHg below the values given by Goldmann tonometry.

Conclusion

The IOP values obtained with the TGDc-01 were in poor agreement with Goldmann tonometry. We found a higher variation as well as a bias towards lower IOP values with the TGDc-01. It is a question of clinical judgement as to how far these deviating measurements can be accepted for screening purposes. Because the IOPs obtained with the TGDc-01 are generally lower and less accurate than those obtained with the Goldmann tonometer we believe that the TGDc-01 is not a reliable tool for IOP measurement in clinical routine.
Literatur
1.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
2.
Zurück zum Zitat Digital mobile tonometer for measurement of intraocular pressure through the eyelid: TGDc-01 “PRA”. Instruction booklet, Rjazan State Toolmaking Plant, 32 Kaljaeva Street, 390000 Rjazan, Russia Digital mobile tonometer for measurement of intraocular pressure through the eyelid: TGDc-01 “PRA”. Instruction booklet, Rjazan State Toolmaking Plant, 32 Kaljaeva Street, 390000 Rjazan, Russia
3.
Zurück zum Zitat Draeger J, Stodtmeister R (2003) Ein “neues” Tonometer? Augenarzt 37:318–319 Draeger J, Stodtmeister R (2003) Ein “neues” Tonometer? Augenarzt 37:318–319
4.
Zurück zum Zitat Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J (2000) Comparing methods of clinical measurements: reporting standards for Bland and Altman analysis. Anesth Analg 90:593–602PubMed Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J (2000) Comparing methods of clinical measurements: reporting standards for Bland and Altman analysis. Anesth Analg 90:593–602PubMed
5.
Zurück zum Zitat Martin M, Pache M, Flammer J (2003) Der portable Tonometer TGDc-01 und die Goldmann- und Lufttonometrie im Vergleich - eine Pilotstudie. In: Auffarth GU, Welt R, Demeler U (eds) Transactions, 17th Congress of the German Society for Intraocular Lens Implantation and Refractive Surgery (DGII). Biermann, Cologne, pp 126–130 Martin M, Pache M, Flammer J (2003) Der portable Tonometer TGDc-01 und die Goldmann- und Lufttonometrie im Vergleich - eine Pilotstudie. In: Auffarth GU, Welt R, Demeler U (eds) Transactions, 17th Congress of the German Society for Intraocular Lens Implantation and Refractive Surgery (DGII). Biermann, Cologne, pp 126–130
Metadaten
Titel
Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry
verfasst von
Annette Lösch
Alexander Scheuerle
Volker Rupp
Gerd Auffarth
Matthias Becker
Publikationsdatum
01.04.2005
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 4/2005
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-004-0971-2

Weitere Artikel der Ausgabe 4/2005

Graefe's Archive for Clinical and Experimental Ophthalmology 4/2005 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.