User-friendly freehand ultrasound calibration using Lego bricks and automatic registration

verfasst von: Yiming Xiao, Charles Xiao Bo Yan, Simon Drouin, Dante De Nigris, Anna Kochanowska, D. Louis Collins

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 9/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

As an inexpensive, noninvasive, and portable clinical imaging modality, ultrasound (US) has been widely employed in many interventional procedures for monitoring potential tissue deformation, surgical tool placement, and locating surgical targets. The application requires the spatial mapping between 2D US images and 3D coordinates of the patient. Although positions of the devices (i.e., ultrasound transducer) and the patient can be easily recorded by a motion tracking system, the spatial relationship between the US image and the tracker attached to the US transducer needs to be estimated through an US calibration procedure. Previously, various calibration techniques have been proposed, where a spatial transformation is computed to match the coordinates of corresponding features in a physical phantom and those seen in the US scans. However, most of these methods are difficult to use for novel users.

Methods

We proposed an ultrasound calibration method by constructing a phantom from simple Lego bricks and applying an automated multi-slice 2D–3D registration scheme without volumetric reconstruction. The method was validated for its calibration accuracy and reproducibility.

Results

Our method yields a calibration accuracy of \(1.23\pm 0.26\) mm and a calibration reproducibility of 1.29 mm.

Conclusion

We have proposed a robust, inexpensive, and easy-to-use ultrasound calibration method.
Literatur
1.
Zurück zum Zitat Machi J, Oishi AJ, Furumoto NL, Oishi RH (2004) Intraoperative ultrasound. Surg Clin N Am 84:1085–1111CrossRefPubMed Machi J, Oishi AJ, Furumoto NL, Oishi RH (2004) Intraoperative ultrasound. Surg Clin N Am 84:1085–1111CrossRefPubMed
2.
Zurück zum Zitat Boctor E, deOliveira M, Choti M, Ghanem R, Taylor R, Hager G, Fichtinger G (2006) Ultrasound monitoring of tissue ablation via deformation model and shape priors. Med Image Comput Comput Assist Interv 9:405–412PubMed Boctor E, deOliveira M, Choti M, Ghanem R, Taylor R, Hager G, Fichtinger G (2006) Ultrasound monitoring of tissue ablation via deformation model and shape priors. Med Image Comput Comput Assist Interv 9:405–412PubMed
3.
Zurück zum Zitat Hammoud MA, Ligon BL, ElSouki R, Shi WM, Schomer DF, Sawaya R (1996) Use of intraoperative ultrasound for localizing tumors and determining the extent of resection: a comparative study with magnetic resonance imaging. J Neurosurg 84:737–741CrossRefPubMed Hammoud MA, Ligon BL, ElSouki R, Shi WM, Schomer DF, Sawaya R (1996) Use of intraoperative ultrasound for localizing tumors and determining the extent of resection: a comparative study with magnetic resonance imaging. J Neurosurg 84:737–741CrossRefPubMed
4.
Zurück zum Zitat Charboneau JW, Reading CC, Welch TJ (1990) Ct and sonographically guided needle-biopsy—current techniques and new innovations. Am J Roentgenol 154:1–10CrossRef Charboneau JW, Reading CC, Welch TJ (1990) Ct and sonographically guided needle-biopsy—current techniques and new innovations. Am J Roentgenol 154:1–10CrossRef
5.
Zurück zum Zitat Comeau RM, Sadikot AF, Fenster A, Peters TM (2000) Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery. Med Phys 27:787–800CrossRefPubMed Comeau RM, Sadikot AF, Fenster A, Peters TM (2000) Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery. Med Phys 27:787–800CrossRefPubMed
6.
Zurück zum Zitat Keles GE, Lamborn KR, Berger MS (2003) Coregistration accuracy and detection of brain shift using intraoperative sononavigation during resection of hemispheric tumors. Neurosurgery 53:556–562CrossRefPubMed Keles GE, Lamborn KR, Berger MS (2003) Coregistration accuracy and detection of brain shift using intraoperative sononavigation during resection of hemispheric tumors. Neurosurgery 53:556–562CrossRefPubMed
7.
Zurück zum Zitat Mercier L, Lango T, Lindseth F, Collins DL (2005) A review of calibration techniques for freehand 3-D ultrasound systems. Ultrasound Med Biol 31:449–471CrossRefPubMed Mercier L, Lango T, Lindseth F, Collins DL (2005) A review of calibration techniques for freehand 3-D ultrasound systems. Ultrasound Med Biol 31:449–471CrossRefPubMed
8.
Zurück zum Zitat Detmer PR, Bashein G, Hodges T, Beach KW, Filer EP, Burns DH, Jr DE (1994) Strandness, 3D ultrasonic image feature localization based on magnetic scanhead tracking—in-vitro calibration and validation. Ultrasound Med Biol 20:923–936CrossRefPubMed Detmer PR, Bashein G, Hodges T, Beach KW, Filer EP, Burns DH, Jr DE (1994) Strandness, 3D ultrasonic image feature localization based on magnetic scanhead tracking—in-vitro calibration and validation. Ultrasound Med Biol 20:923–936CrossRefPubMed
9.
Zurück zum Zitat Trobaugh JW, Trobaugh DJ, Richard WD (1994) Three-dimensional imaging with stereotactic ultrasonography. Comput Med Imaging Gr 18:315–323CrossRef Trobaugh JW, Trobaugh DJ, Richard WD (1994) Three-dimensional imaging with stereotactic ultrasonography. Comput Med Imaging Gr 18:315–323CrossRef
10.
Zurück zum Zitat Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3D freehand ultrasound. Ultrasound Med Biol 24:855–869CrossRefPubMed Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3D freehand ultrasound. Ultrasound Med Biol 24:855–869CrossRefPubMed
11.
Zurück zum Zitat Najafi M, Afsham N, Abolmaesumi P, Rohling R (2015) A closed-form differential formulation for ultrasound spatial calibration: single wall phantom. Ultrasound Med Biol 41:1079–1094CrossRefPubMed Najafi M, Afsham N, Abolmaesumi P, Rohling R (2015) A closed-form differential formulation for ultrasound spatial calibration: single wall phantom. Ultrasound Med Biol 41:1079–1094CrossRefPubMed
12.
Zurück zum Zitat Najafi M, Afsham N, Abolmaesumi P, Rohling R (2014) A closed-form differential formulation for ultrasound spatial calibration: multi-wedge phantom. Ultrasound Med Biol 40:2231–2243CrossRefPubMed Najafi M, Afsham N, Abolmaesumi P, Rohling R (2014) A closed-form differential formulation for ultrasound spatial calibration: multi-wedge phantom. Ultrasound Med Biol 40:2231–2243CrossRefPubMed
13.
Zurück zum Zitat Brown RA (1979) Stereotactic head frame for use with Ct body scanners. Investig Radiol 14:401–401CrossRef Brown RA (1979) Stereotactic head frame for use with Ct body scanners. Investig Radiol 14:401–401CrossRef
14.
Zurück zum Zitat Comeau RM, Fenster A, Peters TM, (1998) Integrated MR and ultrasound imaging for improved image guidance in neurosurgery. In: Proceedings of SPIE, The International Society for Optical Engineering, pp 747–754 Comeau RM, Fenster A, Peters TM, (1998) Integrated MR and ultrasound imaging for improved image guidance in neurosurgery. In: Proceedings of SPIE, The International Society for Optical Engineering, pp 747–754
15.
Zurück zum Zitat Lindseth F, Tangen GA, Lango T, Bang J (2003) Probe calibration for freehand 3D ultrasound. Ultrasound Med Biol 29:1607–1623CrossRefPubMed Lindseth F, Tangen GA, Lango T, Bang J (2003) Probe calibration for freehand 3D ultrasound. Ultrasound Med Biol 29:1607–1623CrossRefPubMed
16.
Zurück zum Zitat Pagoulatos N, Haynor DR, Kim Y (2001) A fast calibration method for 3D tracking of ultrasound images using a spatial localizer. Ultrasound Med Biol 27:1219–1229CrossRefPubMed Pagoulatos N, Haynor DR, Kim Y (2001) A fast calibration method for 3D tracking of ultrasound images using a spatial localizer. Ultrasound Med Biol 27:1219–1229CrossRefPubMed
17.
Zurück zum Zitat Boctor E, Viswanathan A, Choti M, Taylor RH, Fichtinger G, Hager G (2004) A novel closed form solution for ultrasound calibration. In: 2nd IEEE international symposium on biomedical imaging: macro to nano, Vols 1 and 2. pp 527–530, 2004 Boctor E, Viswanathan A, Choti M, Taylor RH, Fichtinger G, Hager G (2004) A novel closed form solution for ultrasound calibration. In: 2nd IEEE international symposium on biomedical imaging: macro to nano, Vols 1 and 2. pp 527–530, 2004
18.
Zurück zum Zitat Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35:79–93CrossRefPubMed Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35:79–93CrossRefPubMed
19.
Zurück zum Zitat Hsu PW, Prager RW, Gee AH, Treece GM (2008) Real-time freehand 3D ultrasound calibration. Ultrasound Med Biol 34:239–251CrossRefPubMed Hsu PW, Prager RW, Gee AH, Treece GM (2008) Real-time freehand 3D ultrasound calibration. Ultrasound Med Biol 34:239–251CrossRefPubMed
20.
Zurück zum Zitat Hsu PW, Treece GM, Prager RW, Houghton NE, Gee AH (2008) Comparison of freehand 3D ultrasound calibration techniques using a stylus. Ultrasound Med Biol 34:1610–1621CrossRefPubMed Hsu PW, Treece GM, Prager RW, Houghton NE, Gee AH (2008) Comparison of freehand 3D ultrasound calibration techniques using a stylus. Ultrasound Med Biol 34:1610–1621CrossRefPubMed
21.
Zurück zum Zitat Melvaer EL, Morken K, Samset E (2012) A motion constrained cross-wire phantom for tracked 2D ultrasound calibration. Int J Comput Assist Radiol Surg 7:611–620CrossRefPubMed Melvaer EL, Morken K, Samset E (2012) A motion constrained cross-wire phantom for tracked 2D ultrasound calibration. Int J Comput Assist Radiol Surg 7:611–620CrossRefPubMed
22.
Zurück zum Zitat Barratt DC, Penney GP, Chan CS, Slomczykowski M, Carter TJ, Edwards PJ, Hawkes DJ (2006) Self-calibrating 3D-ultrasound-based bone registration for minimally invasive orthopedic surgery. IEEE Trans Med Imaging 25:312–323CrossRefPubMed Barratt DC, Penney GP, Chan CS, Slomczykowski M, Carter TJ, Edwards PJ, Hawkes DJ (2006) Self-calibrating 3D-ultrasound-based bone registration for minimally invasive orthopedic surgery. IEEE Trans Med Imaging 25:312–323CrossRefPubMed
23.
Zurück zum Zitat Blackall JM, Rueckert D, Maurer CR, Penney GP, Hill DLG, Hawkes DJ (2000) An image registration approach to automated calibration for freehand 3D ultrasound. Med Image Comput Comput-Assist Interv—Miccai 1935:462–471 Blackall JM, Rueckert D, Maurer CR, Penney GP, Hill DLG, Hawkes DJ (2000) An image registration approach to automated calibration for freehand 3D ultrasound. Med Image Comput Comput-Assist Interv—Miccai 1935:462–471
24.
Zurück zum Zitat Mercier L, Del Maestro RF, Petrecca K, Kochanowska A, Drouin S, Yan CX, Janke AL, Chen SJ, Collins DL (2011) New prototype neuronavigation system based on preoperative imaging and intraoperative freehand ultrasound: system description and validation. Int J Comput Assist Radiol Surg 6:507–522CrossRefPubMed Mercier L, Del Maestro RF, Petrecca K, Kochanowska A, Drouin S, Yan CX, Janke AL, Chen SJ, Collins DL (2011) New prototype neuronavigation system based on preoperative imaging and intraoperative freehand ultrasound: system description and validation. Int J Comput Assist Radiol Surg 6:507–522CrossRefPubMed
25.
Zurück zum Zitat Poon TC, Rohling RN (2005) Comparison of calibration methods for spatial tracking of a 3D ultrasound probe. Ultrasound Med Biol 31:1095–1108CrossRefPubMed Poon TC, Rohling RN (2005) Comparison of calibration methods for spatial tracking of a 3D ultrasound probe. Ultrasound Med Biol 31:1095–1108CrossRefPubMed
26.
Zurück zum Zitat Caramanos Z, Fonov VS, Francis SJ, Narayanan S, Pike GB, Collins DL, Arnold DL (2010) Gradient distortions in MRI: characterizing and correcting for their effects on SIENA-generated measures of brain volume change. Neuroimage 49:1601–1611CrossRefPubMed Caramanos Z, Fonov VS, Francis SJ, Narayanan S, Pike GB, Collins DL, Arnold DL (2010) Gradient distortions in MRI: characterizing and correcting for their effects on SIENA-generated measures of brain volume change. Neuroimage 49:1601–1611CrossRefPubMed
27.
Zurück zum Zitat Soehl M, Walsh R, Rankin A, Lasso A, Fichtinger G (2014) Tracked ultrasound calibration studies with a phantom made of LEGO bricks. Med Imaging, Image-Guided Proced, Robot Interv, Modeling 9036:2014 Soehl M, Walsh R, Rankin A, Lasso A, Fichtinger G (2014) Tracked ultrasound calibration studies with a phantom made of LEGO bricks. Med Imaging, Image-Guided Proced, Robot Interv, Modeling 9036:2014
28.
Zurück zum Zitat Yan CXB, Goulet B, Tampieri D, Collins DL (2012) Ultrasound-CT registration of vertebrae without reconstruction. Int J Comput Assist Radiol Surg 7:901–909CrossRefPubMed Yan CXB, Goulet B, Tampieri D, Collins DL (2012) Ultrasound-CT registration of vertebrae without reconstruction. Int J Comput Assist Radiol Surg 7:901–909CrossRefPubMed
29.
Zurück zum Zitat Cleary K, Anderson J, Brazaitis M, Devey G, DiGioia A, Freedman M, Gronemeyer D, Lathan C, Lemke H, Long D, Mun SK, Taylor R (2000) Final report of the technical requirements for image-guided spine procedures workshop, April 17–20, 1999, Ellicott City, MA, USA. Comput Aided Surg 5:180–215PubMed Cleary K, Anderson J, Brazaitis M, Devey G, DiGioia A, Freedman M, Gronemeyer D, Lathan C, Lemke H, Long D, Mun SK, Taylor R (2000) Final report of the technical requirements for image-guided spine procedures workshop, April 17–20, 1999, Ellicott City, MA, USA. Comput Aided Surg 5:180–215PubMed
Metadaten
Titel
User-friendly freehand ultrasound calibration using Lego bricks and automatic registration
verfasst von
Yiming Xiao
Charles Xiao Bo Yan
Simon Drouin
Dante De Nigris
Anna Kochanowska
D. Louis Collins
Publikationsdatum
01.09.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 9/2016
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-016-1368-5

Weitere Artikel der Ausgabe 9/2016

International Journal of Computer Assisted Radiology and Surgery 9/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.