Variability between the trial and final implant measurements during the sensor-guided total knee arthroplasty

verfasst von: Jong-Keun Kim, Do Weon Lee, Du Hyun Ro, Hyuk-Soo Han, Myung Chul Lee

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 8/2022

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Compartmental load-sensing technology has been used in the attempt to achieve optimal soft tissue balance during total knee arthroplasty (TKA). This study was conducted to investigate the validity of such use of intraoperative sensing technology.

Methods

Ninety-three knees scheduled to undergo total knee arthroplasty for knee osteoarthritis with a tibial sensor were prospectively enrolled. Measurements were divided into three groups according to the three different time points of intraoperative load testing: group Trial (with the trial components), group Final (with the definitive cemented implants and an open joint capsule), and group Closed (with the definitive cemented implants and a closed joint capsule). Load measurements and component rotational alignments were documented at 10°, 30°, 45°, 90°, and 120° of flexion for all three groups, and compared. One year postoperatively, the joint line obliquity angle was obtained radiographically in the valgus and varus stress views at 10° and 30° flexion to evaluate the clinical instability. The Knee Society, Hospital for Special Surgery, and Western Ontario McMaster Universities Osteoarthritis Index scores were used to determine functional outcomes. The correlations of the above outcomes with intraoperative load were evaluated.

Results

There were significant differences in medial and lateral loads at all flexion angles (except at a 120° lateral load) between group Trial and group Final (p < 0.05). Tibial trays were internally rotated to a significantly higher degree in group Final than in group Trial (p = 0.010). The lateral compartmental load significantly decreased after patellar inversion (p = 0.037). There were no correlations of intraoperative load with clinical instability and functional outcomes.

Conclusion

Significant variability was observed between the trial and final implant measurements and intraoperative sensing data were not correlated with instability or functional outcomes over a 1-year period. Therefore, intraoperative sensor technology provides limited feedback and clinical efficacy in the adjustment of the soft tissue balance during TKA.

Level of evidence

Level II.
Literatur
1.
Zurück zum Zitat Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW (1988) Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol 15:1833–1840PubMed Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW (1988) Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol 15:1833–1840PubMed
2.
Zurück zum Zitat Chow J, Wang K, Elson L, Anderson C, Roche M (2017) Effects of cementing on ligament balance during total knee arthroplasty. Orthopedics 40:e455–e459PubMed Chow J, Wang K, Elson L, Anderson C, Roche M (2017) Effects of cementing on ligament balance during total knee arthroplasty. Orthopedics 40:e455–e459PubMed
3.
Zurück zum Zitat Geller JA, Lakra A, Murtaugh T (2017) The Use of electronic sensor device to augment ligament balancing leads to a lower rate of arthrofibrosis after total knee arthroplasty. J Arthroplasty 32:1502–1504CrossRef Geller JA, Lakra A, Murtaugh T (2017) The Use of electronic sensor device to augment ligament balancing leads to a lower rate of arthrofibrosis after total knee arthroplasty. J Arthroplasty 32:1502–1504CrossRef
4.
Zurück zum Zitat Ghirardelli S, Bala A, Peretti G, Antonini G, Indelli PF (2019) Intraoperative sensing technology to achieve balance in primary total knee arthroplasty: a review of the literature. JBJS Rev 7:e4CrossRef Ghirardelli S, Bala A, Peretti G, Antonini G, Indelli PF (2019) Intraoperative sensing technology to achieve balance in primary total knee arthroplasty: a review of the literature. JBJS Rev 7:e4CrossRef
5.
Zurück zum Zitat Gustke KA (2014) Soft-tissue and alignment correction: the use of smart trials in total knee replacement. Bone Joint J 96-b:78–83CrossRef Gustke KA (2014) Soft-tissue and alignment correction: the use of smart trials in total knee replacement. Bone Joint J 96-b:78–83CrossRef
6.
Zurück zum Zitat Gustke KA, Golladay GJ, Roche MW, Elson LC, Anderson CR (2014) A new method for defining balance: promising short-term clinical outcomes of sensor-guided TKA. J Arthroplasty 29:955–960CrossRef Gustke KA, Golladay GJ, Roche MW, Elson LC, Anderson CR (2014) A new method for defining balance: promising short-term clinical outcomes of sensor-guided TKA. J Arthroplasty 29:955–960CrossRef
7.
Zurück zum Zitat Gustke KA, Golladay GJ, Roche MW, Elson LC, Anderson CR (2014) Primary TKA patients with quantifiably balanced soft-tissue achieve significant clinical gains sooner than unbalanced patients. Adv Orthop 2014:628695CrossRef Gustke KA, Golladay GJ, Roche MW, Elson LC, Anderson CR (2014) Primary TKA patients with quantifiably balanced soft-tissue achieve significant clinical gains sooner than unbalanced patients. Adv Orthop 2014:628695CrossRef
8.
Zurück zum Zitat Han HS, Lee MC (2017) Cementing technique affects the rate of femoral component loosening after high flexion total knee arthroplasty. Knee 24:1435–1441CrossRef Han HS, Lee MC (2017) Cementing technique affects the rate of femoral component loosening after high flexion total knee arthroplasty. Knee 24:1435–1441CrossRef
9.
Zurück zum Zitat Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the knee society clinical rating system. Clin Orthop Relat Res 1:13–14 Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the knee society clinical rating system. Clin Orthop Relat Res 1:13–14
10.
Zurück zum Zitat Kim JK, Ro DH, Lee HJ, Park JY, Han HS, Lee MC (2020) Efficacy of systemic steroid use given one day after total knee arthroplasty for pain and nausea: a randomized controlled study. J Arthroplasty 35:69–75CrossRef Kim JK, Ro DH, Lee HJ, Park JY, Han HS, Lee MC (2020) Efficacy of systemic steroid use given one day after total knee arthroplasty for pain and nausea: a randomized controlled study. J Arthroplasty 35:69–75CrossRef
11.
Zurück zum Zitat Livermore AT, Erickson JA, Blackburn B, Peters CL (2020) Does the sequential addition of accelerometer-based navigation and sensor-guided ligament balancing improve outcomes in TKA? Bone Joint J 102-b:24–30CrossRef Livermore AT, Erickson JA, Blackburn B, Peters CL (2020) Does the sequential addition of accelerometer-based navigation and sensor-guided ligament balancing improve outcomes in TKA? Bone Joint J 102-b:24–30CrossRef
12.
Zurück zum Zitat MacDessi SJ, Cohen DA, Wood JA, Diwan AD, Harris IA (2021) Does the use of intraoperative pressure sensors for knee balancing in total knee arthroplasty improve clinical outcomes? A comparative study with a minimum two-year follow-up. J Arthroplasty 36:514–519CrossRef MacDessi SJ, Cohen DA, Wood JA, Diwan AD, Harris IA (2021) Does the use of intraoperative pressure sensors for knee balancing in total knee arthroplasty improve clinical outcomes? A comparative study with a minimum two-year follow-up. J Arthroplasty 36:514–519CrossRef
13.
Zurück zum Zitat Manning WA, Ghosh K, Blain A, Longstaff L, Deehan DJ (2017) Tibiofemoral forces for the native and post-arthroplasty knee: relationship to maximal laxity through a functional arc of motion. Knee Surg Sports Traumatol Arthrosc 25:1669–1677CrossRef Manning WA, Ghosh K, Blain A, Longstaff L, Deehan DJ (2017) Tibiofemoral forces for the native and post-arthroplasty knee: relationship to maximal laxity through a functional arc of motion. Knee Surg Sports Traumatol Arthrosc 25:1669–1677CrossRef
14.
Zurück zum Zitat McClelland JA, Webster KE, Ramteke AA, Feller JA (2017) Total knee arthroplasty with computer-assisted navigation more closely replicates normal knee biomechanics than conventional surgery. Knee 24:651–656CrossRef McClelland JA, Webster KE, Ramteke AA, Feller JA (2017) Total knee arthroplasty with computer-assisted navigation more closely replicates normal knee biomechanics than conventional surgery. Knee 24:651–656CrossRef
16.
Zurück zum Zitat Nodzo SR, Franceschini V, Gonzalez Della Valle A (2017) Intraoperative load-sensing variability during cemented, posterior-stabilized total knee arthroplasty. J Arthroplasty 32:66–70CrossRef Nodzo SR, Franceschini V, Gonzalez Della Valle A (2017) Intraoperative load-sensing variability during cemented, posterior-stabilized total knee arthroplasty. J Arthroplasty 32:66–70CrossRef
17.
Zurück zum Zitat Ranawat CS, Shine JJ (1973) Duo-condylar total knee arthroplasty. Clin Orthop Relat Res 94:185–195CrossRef Ranawat CS, Shine JJ (1973) Duo-condylar total knee arthroplasty. Clin Orthop Relat Res 94:185–195CrossRef
18.
Zurück zum Zitat Roche MW, Elson LC, Anderson CR (2015) A novel technique using sensor-based technology to evaluate tibial tray rotation. Orthopedics 38:e217-222CrossRef Roche MW, Elson LC, Anderson CR (2015) A novel technique using sensor-based technology to evaluate tibial tray rotation. Orthopedics 38:e217-222CrossRef
19.
Zurück zum Zitat Schnaser E, Lee YY, Boettner F, Gonzalez Della Valle A (2015) The position of the patella and extensor mechanism affects intraoperative compartmental loads during total knee arthroplasty: a pilot study using intraoperative sensing to guide soft tissue balance. J Arthroplasty 30:1348-1353.e1343CrossRef Schnaser E, Lee YY, Boettner F, Gonzalez Della Valle A (2015) The position of the patella and extensor mechanism affects intraoperative compartmental loads during total knee arthroplasty: a pilot study using intraoperative sensing to guide soft tissue balance. J Arthroplasty 30:1348-1353.e1343CrossRef
20.
Zurück zum Zitat Sculco P, Gruskay J, Nodzo S, Carrol K, Shanaghan K, Haas S et al (2018) The role of the tourniquet and patella position on the compartmental loads during sensor-assisted total knee arthroplasty. J Arthroplasty 33:S121-s125CrossRef Sculco P, Gruskay J, Nodzo S, Carrol K, Shanaghan K, Haas S et al (2018) The role of the tourniquet and patella position on the compartmental loads during sensor-assisted total knee arthroplasty. J Arthroplasty 33:S121-s125CrossRef
21.
Zurück zum Zitat Siddiqi A, Smith T, McPhilemy JJ, Ranawat AS, Sculco PK, Chen AF (2020) Soft-tissue balancing technology for total knee arthroplasty. JBJS Rev 8:e0050CrossRef Siddiqi A, Smith T, McPhilemy JJ, Ranawat AS, Sculco PK, Chen AF (2020) Soft-tissue balancing technology for total knee arthroplasty. JBJS Rev 8:e0050CrossRef
22.
Zurück zum Zitat Song SJ, Kang SG, Lee YJ, Kim KI, Park CH (2019) An intraoperative load sensor did not improve the early postoperative results of posterior-stabilized TKA for osteoarthritis with varus deformities. Knee Surg Sports Traumatol Arthrosc 27:1671–1679CrossRef Song SJ, Kang SG, Lee YJ, Kim KI, Park CH (2019) An intraoperative load sensor did not improve the early postoperative results of posterior-stabilized TKA for osteoarthritis with varus deformities. Knee Surg Sports Traumatol Arthrosc 27:1671–1679CrossRef
23.
Zurück zum Zitat van der Linde JA, Leong AKL (2019) Does bone cement influence soft tissue balancing in total knee arthroplasty? Knee 26:405–409CrossRef van der Linde JA, Leong AKL (2019) Does bone cement influence soft tissue balancing in total knee arthroplasty? Knee 26:405–409CrossRef
Metadaten
Titel
Variability between the trial and final implant measurements during the sensor-guided total knee arthroplasty
verfasst von
Jong-Keun Kim
Do Weon Lee
Du Hyun Ro
Hyuk-Soo Han
Myung Chul Lee
Publikationsdatum
04.01.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 8/2022
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-021-06795-2

Weitere Artikel der Ausgabe 8/2022

Knee Surgery, Sports Traumatology, Arthroscopy 8/2022 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Lever-Sign-Test hilft beim Verdacht auf Kreuzbandriss

15.05.2024 Vordere Kreuzbandruptur Nachrichten

Mit dem Hebelzeichen-Test lässt sich offenbar recht zuverlässig feststellen, ob ein vorderes Kreuzband gerissen ist. In einer Metaanalyse war die Vorhersagekraft vor allem bei positivem Testergebnis hoch.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.