A method for evaluating luminance non-uniformity of displays by use of a commercially available digital camera

verfasst von: Keishin Kawamoto, Shogo Tokurei, Junji Morishita

Erschienen in: Radiological Physics and Technology | Ausgabe 4/2017

Einloggen, um Zugang zu erhalten

Abstract

The luminance uniformity of liquid–crystal displays (LCDs) deteriorates with their prolonged use. In this paper, we present a method for evaluating the degree of luminance non-uniformity of LCDs with the use of a commercially available digital camera. In this study, seven monochrome LCDs, which were used during various operating times ranging from 5000 to 25,000 h, were evaluated with use of a camera. The maximum luminance deviation (MLD) was measured on the two-dimensional (2D) images obtained with the camera. In addition, an index of the luminance non-uniformity was calculated as the ratio of the area exhibiting luminance non-uniformity to the area of the entire LCD screen. We determined the area with the luminance non-uniformity by setting the allowable luminance deviation as the judgment criterion to evaluate the degree of luminance non-uniformity. The MLD values were less than 20% for all conditions, and they varied depending on the locations of the luminance measurement. The area ratios of the luminance non-uniformity based on 2D luminance distributions tended to increase with the duration of use of LCDs, and they indicated the degree of luminance non-uniformity of the LCDs regardless of the measurement locations. Our approach of using a commercially available digital camera showed its potential usefulness for providing more detailed and consistent evaluations of the degree of luminance non-uniformity of LCDs based on the 2D luminance distributions.
Literatur
2.
Zurück zum Zitat DIN6868-57. Image quality assurance in X-ray diagnostics. Part 57: acceptance testing for image display devices. DIN6868-57. Image quality assurance in X-ray diagnostics. Part 57: acceptance testing for image display devices.
4.
Zurück zum Zitat Kagadis GC, Flannigan AW, Krupinski EA, Nagy PG, Katsanos K, Diamantopoulos A, Langer SG. Medical imaging displays and their use in image interpretation. RadioGraphics. 2014;33(1):275–90.CrossRef Kagadis GC, Flannigan AW, Krupinski EA, Nagy PG, Katsanos K, Diamantopoulos A, Langer SG. Medical imaging displays and their use in image interpretation. RadioGraphics. 2014;33(1):275–90.CrossRef
5.
Zurück zum Zitat Madsack B, Walz M, Weisser G. Abnahme und Konstanzprüfung an Bildwiedergabesystemen—was ändert sich mit der neuen DIN V 6868-157. Radiopraxis. 2014;7:195–210.CrossRef Madsack B, Walz M, Weisser G. Abnahme und Konstanzprüfung an Bildwiedergabesystemen—was ändert sich mit der neuen DIN V 6868-157. Radiopraxis. 2014;7:195–210.CrossRef
6.
Zurück zum Zitat Takeda Y, Yano H, Ohira T, Aoyama N, Uetsuki T, Fukumasa O. The effect of mercury on damage to cold cathode fluorescent lamp electrodes. J Light Vis Environ. 2011;35(2):101–7.CrossRef Takeda Y, Yano H, Ohira T, Aoyama N, Uetsuki T, Fukumasa O. The effect of mercury on damage to cold cathode fluorescent lamp electrodes. J Light Vis Environ. 2011;35(2):101–7.CrossRef
7.
Zurück zum Zitat Hellén-Halme K, Hellén-Halme B, Wenzel A. The effect of aging on luminance of standard liquid crystal display (LCD) monitors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;112(2):237–42.CrossRefPubMed Hellén-Halme K, Hellén-Halme B, Wenzel A. The effect of aging on luminance of standard liquid crystal display (LCD) monitors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;112(2):237–42.CrossRefPubMed
8.
Zurück zum Zitat Yano Y, Morishita J, Akamine H, Hashimoto N, Nakamura Y, Yabuuchi H. Effect of fluorescent lighting on the color of liquid-crystal displays. Radiol Phys Technol. 2014;7:367–72.CrossRefPubMed Yano Y, Morishita J, Akamine H, Hashimoto N, Nakamura Y, Yabuuchi H. Effect of fluorescent lighting on the color of liquid-crystal displays. Radiol Phys Technol. 2014;7:367–72.CrossRefPubMed
9.
Zurück zum Zitat Akamine H, Morishita J, Matsuyama M, Nakamura Y, Hashimoto N, Toyofuku F. Chromaticity and correlated color temperature of the white point in medical liquid-crystal displays. Med Phys. 2012;39(8):5127–35.CrossRefPubMed Akamine H, Morishita J, Matsuyama M, Nakamura Y, Hashimoto N, Toyofuku F. Chromaticity and correlated color temperature of the white point in medical liquid-crystal displays. Med Phys. 2012;39(8):5127–35.CrossRefPubMed
10.
Zurück zum Zitat Roehring H, Du X, Fan J. NPS assessment of color medical displays using a monochromatic CCD camera. Proc SPIE. 2012;8318. doi:10.1117/12.966331. Roehring H, Du X, Fan J. NPS assessment of color medical displays using a monochromatic CCD camera. Proc SPIE. 2012;8318. doi:10.​1117/​12.​966331.
11.
Zurück zum Zitat Yamazaki A, Marini-Bettolo C, Kodera Y, Badano A. Sharpness and noise characteristics of a half-mirror stereoscopic display. J Soc Inf Disp. 2014;22(3):170–6.CrossRef Yamazaki A, Marini-Bettolo C, Kodera Y, Badano A. Sharpness and noise characteristics of a half-mirror stereoscopic display. J Soc Inf Disp. 2014;22(3):170–6.CrossRef
12.
Zurück zum Zitat Ichikawa K, Kodera Y, Nishimura A, Hasegawa M, Kimura N, Takemura A, Matsubara K. Analysis method of noise power spectrum for medical monochrome liquid crystal displays. Radiol Phys Technol. 2008;1(2):201–7.CrossRefPubMed Ichikawa K, Kodera Y, Nishimura A, Hasegawa M, Kimura N, Takemura A, Matsubara K. Analysis method of noise power spectrum for medical monochrome liquid crystal displays. Radiol Phys Technol. 2008;1(2):201–7.CrossRefPubMed
13.
Zurück zum Zitat Tokurei S, Morishita J. A method for evaluating image quality of monochrome and color displays based on luminance by use of a commercially available color digital camera. Med Phys. 2015;42(8):4773–82.CrossRefPubMed Tokurei S, Morishita J. A method for evaluating image quality of monochrome and color displays based on luminance by use of a commercially available color digital camera. Med Phys. 2015;42(8):4773–82.CrossRefPubMed
14.
Zurück zum Zitat National Electrical Manufacturers Association. Digital imaging and communications in medicine (DICOM) part 14: grayscale standard display function. NEMA PS 3.14-2001. VA: National Manufacturers Association. 2007. http://www.nema.org/stds/ps3-14.cfm National Electrical Manufacturers Association. Digital imaging and communications in medicine (DICOM) part 14: grayscale standard display function. NEMA PS 3.14-2001. VA: National Manufacturers Association. 2007. http://​www.​nema.​org/​stds/​ps3-14.​cfm
15.
Zurück zum Zitat Wyszecki G, Stiles WS. Color Science. Hoboken: Wiley; 2000. p. 155–6. Wyszecki G, Stiles WS. Color Science. Hoboken: Wiley; 2000. p. 155–6.
16.
Zurück zum Zitat Kobayashi S, Mikoshiba S, Lim S. LCD backlights. Hoboken: Wiley; 2009. p. 5–6.CrossRef Kobayashi S, Mikoshiba S, Lim S. LCD backlights. Hoboken: Wiley; 2009. p. 5–6.CrossRef
17.
Zurück zum Zitat Ide T, Murata H, Taira Y, Mizuta H, Suzuki M, Noguchi M, Katsu Y. A novel dot-pattern generation to improve luminance uniformity of LCD backlight. JSID. 2003;11(4):659–66.CrossRef Ide T, Murata H, Taira Y, Mizuta H, Suzuki M, Noguchi M, Katsu Y. A novel dot-pattern generation to improve luminance uniformity of LCD backlight. JSID. 2003;11(4):659–66.CrossRef
Metadaten
Titel
A method for evaluating luminance non-uniformity of displays by use of a commercially available digital camera
verfasst von
Keishin Kawamoto
Shogo Tokurei
Junji Morishita
Publikationsdatum
16.08.2017
Verlag
Springer Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 4/2017
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-017-0412-7

Weitere Artikel der Ausgabe 4/2017

Radiological Physics and Technology 4/2017 Zur Ausgabe

Acknowledgments

Acknowledgment

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.