A narrative commentary about interoperability in medical devices and data used in diabetes therapy from an academic EU/UK/US perspective

verfasst von: Johan Jendle, Peter Adolfsson, Pratik Choudhary, Klemen Dovc, Alexander Fleming, David C. Klonoff, Julia K. Mader, Nick Oliver, Jennifer L. Sherr, Jan Šoupal, Lutz Heinemann

Erschienen in: Diabetologia | Ausgabe 2/2024

Abstract

People living with diabetes have many medical devices available to assist with disease management. A critical aspect that must be considered is how systems for continuous glucose monitoring and insulin pumps communicate with each other and how the data generated by these devices can be downloaded, integrated, presented and used. Not only is interoperability associated with practical challenges, but also devices must adhere to all aspects of regulatory and legal frameworks. Key issues around interoperability in terms of data ownership, privacy and the limitations of interoperability include where the responsibility/liability for device and data interoperability lies and the need for standard data-sharing protocols to allow the seamless integration of data from different sources. There is a need for standardised protocols for the open and transparent handling of data and secure integration of data into electronic health records. Here, we discuss the current status of interoperability in medical devices and data used in diabetes therapy, as well as regulatory and legal issues surrounding both device and data interoperability, focusing on Europe (including the UK) and the USA. We also discuss a potential future landscape in which a clear and transparent framework for interoperability and data handling also fulfils the needs of people living with diabetes and healthcare professionals.

Graphical Abstract

Hinweise

Supplementary Information

The online version contains a slideset of the figures for download available at https://​doi.​org/​10.​1007/​s00125-023-06049-5.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ACE
Alternate controller-enabled
AID
Automated insulin delivery
API
Application programming interface
CGM
Continuous glucose monitoring
EMR
Electronic medical record
EMA
European Medicines Agency
FDA
US Food and Drug Administration
GDPR
General Data Protection Regulation
HCP
Healthcare professional
iCGM
Integrated continuous glucose monitoring
iCoDE
Integration of Continuous Glucose Monitoring Data into the Electronic Health Record
IEEE
Institute of Electrical and Electronics Engineers
IFU
Instructions for use
MDR
Medical Device Regulation
PLWD
People living with diabetes
RWE
Real-world evidence
VDC
Virtual diabetes clinic

Introduction

Many different medical devices are available to support people living with diabetes (PLWD) with diabetes management. However, to get the most benefit from these devices it is important that they communicate effectively with each other (device interoperability) and that the data generated are seamlessly integrated as well as viewed across platforms (data interoperability) (Figs 1 and 2). Data are invaluable for PLWD and their healthcare professionals (HCPs); however, issues related to data generation, transfer, storage, access and use must be considered. Several factors currently limit the full potential of the data generated by these medical products. Some are hurdles regarding costs and regulatory policies; however, many are related to data sharing. One major question in this respect is related to data ownership, which might look different in different countries. HCPs and, to a greater extent, device users are end-users and the data generated are crucial to population research and large-scale real-world evidence (RWE) studies.
A key objective in enabling data interoperability is to have a standard application programming interface (API), which would allow the integration of data from different devices into a site’s/organisation’s electronic medical record (EMR) system, with a single point of entry for the system [1] (Fig. 1). An open API enables software developers to handle data from multiple sources, for example for use in virtual diabetes clinics (VDCs). However, a VDC could require multiple data platforms and multiple sign-ins, which would be time-consuming and hinder efficient workflows. Consequently, a simplified process enabling the use and integration of data transfer from different products into digital platforms would reduce the burden for both PLWD and their HCPs, with automated or semi-automated data transfer making the process smoother [1]. The seamless transmission and interpretation of data could also be made more accessible through agreement on a minimum dataset and standardisation of interoperability between medical devices, device-specific software, EMR systems, apps and HCP interface portals [2]. Open access to data, while complying with the General Data Protection Regulation (GDPR), which came into effect in the EU in 2018, would also support data analysis, data interpretation and display of data in a clear and instructive manner [3]. Ongoing initiatives such as the integration of Continuous Glucose Monitoring Data into the Electronic Health Record (iCoDE) project are currently focusing on providing a data standard and workflow framework for integrating continuous glucose monitoring (CGM) data into EMRs to enable these data to be readily accessed and used in routine clinical care [4]. iCoDE is based on the US Health Insurance Portability and Accountability Act of 1996. A European version of iCoDE would need to be compliant with the GDPR. In addition to use by HCPs, manufacturers also use device-specific data for proprietary data management systems and quality control, as required by post-marketing surveillance requirements set out in the 2017 EU Medical Device Regulation (MDR) [5], wherein the manufacturer is responsible for the information presented by the proprietary data management system (not for third-party interpretation).
This article provides an overview of critical issues around interoperability in medical devices and data used in diabetes therapy, including data ownership, privacy, security, interoperability limitations and the influence of interoperability on the device marketplace. It also discusses the main challenges that must be addressed for device and data interoperability to generate a seamless, secure process that complies with local regulatory and legal requirements in Europe (including the UK) and the USA. For definitions related to device and data operability, see the textbox (‘Device and data interoperability: definitions’). This article is not intended to be a systematic (scientific) review or a consensus paper about interoperability. In a sense, it is more an academic commentary, mainly from an EU/UK/US perspective. We are fully aware of the fact that interoperability is a complex topic that undergoes rapid changes. One limitation of our approach is that we are a group of academics with a focus on the clinical point of view; consideration from other stakeholders is also needed.

Limitations of interoperability

Device interoperability is particularly vulnerable to limitations such as cybersecurity, quality and liability in terms of a system’s intended purpose. While the intended purpose of CGM systems is glucose monitoring and that of insulin pumps is insulin delivery, the intended use of automated insulin delivery (AID) systems is diabetes management as driven by an AID algorithm, the accuracy of insulin delivery and input by the CGM system. In the USA, the US Food and Drug Administration (FDA) tends to regulate end-to-end systems but differences in sensor accuracy, algorithms or insulin delivery, in particular, may impact clinical outcomes. On a similar note, the efficacy and safety of a system’s performance is based on the engineering specifications of the individual components and not on the performance in general of the system.
From the user perspective, using a combination of different products means that there may be some gaps in customer care that arise from interoperability considerations, as well as restricted or piecemeal clinical and technical support for HCPs. Additionally, the use of different products means that there is the potential for discrepancies within a model of open data handling with different devices from different manufacturers. For example, a user may have indicated to the manufacturer of a given CGM system that a third party may use their data. However, the user may not have granted permission for the pump manufacturer to use their data or may initially grant permission to both manufacturers and subsequently withdraw permission from one manufacturer but not the other. In this situation, it is necessary to consider how the CGM information may be passed from one manufacturer to the other. Another potential legal concern around combined products is that, in the eventuality of a malfunction, who is liable and how is this determined?
Given the limited computational power of AID systems, the checking of data integrity, imputation of missing data and reconciliation of outliers in data by these systems are limited. Hence, the coordination of such activities by various devices becomes another challenge for the interoperability of medical devices. Potential limitations from the regulatory perspective concerning AID system approval include that some markets have defined diabetes-specific operability standards but have not yet created the approval process for the system in its intended use. At the same time, new AID technologies are bringing new concepts regarding a ‘hands-off’ approach to glucose control. Therefore, it is likely that at least some of the challenges for users and HCPs in terms of coping with the volume of data and the provision of therapy recommendations by HCPs will be overcome by design.
Overview
There are substantial differences across jurisdictions concerning the regulatory and legal frameworks for medical products and devices. The regulation of medical devices in the USA (where the FDA is the single national agency responsible for device approval) and the EU (where different notified bodies are designated and contracted for device review and approval) differ considerably in terms of requirements and organisational structure. There are differences across jurisdictions in terms of how data/RWE can be used. Companies may be subject to specific laws depending on where the company headquarters are located, where the diabetes technology devices are manufactured and where the servers for data storage are located. For example, the legal frameworks for data protection differ between Europe and the USA [6, 7].
Regulatory frameworks in the EU/UK/USA
The EU does not have an interoperable diabetes device pathway comparable to that in the USA. Instead, in the EU, the manufacturer specifies the intended purpose, technical specifications, indications and limits of use for its product supported by clinical evidence, which are detailed in the instructions/information for users and the technical documentation. Within the EU, the European Medicines Agency (EMA) is responsible for the regulation of some medicines but not for the regulation of medical devices; instead, the competent authorities are the regulators of medical devices and accredit notified bodies in Europe to assess manufacturers [8].
Since May 2017, a new regulatory framework for medical devices (MDR) has been in place in the EU [5], which represents a significant change in how they are regulated [9]. The implementation of the MDR commenced in May 2021 and, from May 2024, all medical devices placed on the market will have to conform to the MDR. Medical devices are evaluated and those that conform to the EU MDR are allowed to enter the market by the independent notified bodies. The MDR requires the manufacturer to provide technical documentation and perform clinical evaluation. Traditionally, medical devices, but not necessarily diabetes-related products, have reached the market sooner in the EU than in the USA. However, implementation of the MDR may reduce the differences in data requirements and marketing approval times between the EU and the USA. In particular, the MDR has new, stricter classification rules for AID systems (e.g. Rule 22, ‘Active therapeutic devices with an integrated or incorporated diagnostic function that significantly determines the patient management’) with a higher risk classification (class III) and is more complex than the modular US approach. In addition, unlike the USA, there are no standard technical specifications for any of the different devices within AID systems. Furthermore, no clear direction or guidance exists on how notified bodies should practically apply the definition of AID systems and the higher risk classification through their reviews and issuance of certificates.
Combining products from different manufacturers
The manufacturer of a system component defines its interoperability with other components. In the area of diabetes, for example, this results in the availability of AID system components intended to be combined only with other specified system components (e.g. from the same manufacturer), as well as the availability of system components intended to be used with a wider range of components (e.g. from different manufacturers). In instances where components from different manufacturers are combined, interoperability must be explicitly demonstrated (i.e. similar to the FDA’s interoperability provision). Key considerations for combined products include establishing who is responsible for the combined product in terms of liability and warranty and how to demonstrate the safety and efficacy of combined devices in accordance with the MDR. With regard to AID systems, other components might be added for continuous monitoring of physical activity or ketone bodies.
Liability in Europe
According to the EU Product Liability Directive, the manufacturer of a device is liable for the device and should ensure that it is working according to the product’s specifications, as formalised in the CE (Conformité Européenne) mark and the instructions for use (IFU) (per EU Council Directive 85/374/EEC of 1985 on the approximation of the laws, regulations and administrative provisions of the Member States concerning liability for defective products [10]. The labelling or IFU is defined in the MDR as the ‘information provided by the manufacturer to inform the user of a device’s intended purpose and proper use, and of any precautions to be taken’, and the manufacturer is defined as the ‘natural or legal person who manufactures or fully refurbishes a device or has a device designed, manufactured or fully refurbished, and markets that device under its name or trademark’ [5]. However, in terms of liability, the PLWD who uses the device also has a degree of responsibility in that they are expected to use the device according to the IFU provided by the manufacturer. Therefore, these instructions need to be clear, transparent and understandable. Some manufacturers provide specific warnings to intended users in case of misuse or modifications of a device, such as ‘modifying the devices can cause serious injury, interfere with your ability to operate the device and void your warranty’ [11, 12].
The EU’s position on interoperability should also be considered in the context of digital health, wherein the EU acknowledges the need for digital tools to support PLWD-centred care. However, many discussions about digital health and future directions are also ongoing within the EU, and the use of any digital tools should also consider the current data privacy protection guidelines (GDPR) and the corresponding data privacy provisions of the UK Data Protection Act 2018, which is the UK’s version of the GDPR [13]. Any exchange of information from PLWD will require a European EMR to be established, which in turn will require data interoperability specifications to be met, and only when data can be assessed in a standardised manner can the data generated by diabetes technology systems be successfully integrated into EMRs. However, a key component of the GDPR is the ‘right to be forgotten’ [14], which raises the questions of what happens to an individual’s data if that individual wishes to stop using a given AID system and whether this information can be appropriately deleted. Data are supposed to be stored on servers located in the EU only. From our point of view, all data generated in the context of this therapy are owned by the individual PLWD and not the manufacturer of a given device. PLWD should therefore have access to the data stored by the manufacturer and the opportunity to have these data deleted.
In Europe, several legislative initiatives have been designed to both ease and regulate access to data; these include the Artificial Intelligence Act [15], the Data Act [16] and the creation of the European Health Data Space [17]. The European Health Data Space, specifically, once formally adopted, will provide a health-specific ecosystem comprising rules, common standards and infrastructures. It will also include a governance framework that aims to empower individuals through increased digital access to and control of their electronic personal health data. This approach can work at both a national level and an EU-wide level and will support the free movement of data. Moreover, it will also foster a genuine single market for EMR systems, relevant medical devices and high-risk artificial intelligence systems. Finally, the availability of information in all EU languages, in accordance with EU regulations, will enable data searching by all individuals across the EU, providing a consistent, trustworthy and efficient set-up for using health data for research, innovation, policymaking and regulatory activities.
Regulatory framework on interoperability in the USA
The FDA has been highly supportive of the development of diabetes technology devices, beginning with detailed guidance introduced in 2012 [18]. This guidance covered CGM systems, primary endpoints that can be used to determine safety and effectiveness, indications for use, and clinical study progression and design. Moreover, the 2012 guidance explicitly notes that the intention is to apply the ‘least burdensome’ approach to investigating and developing AID systems and making them available to PLWD. The FDA has also encouraged interoperability among available insulin pumps, CGM systems and other diabetes devices, with the specification in 2018 of an integrated CGM (iCGM) regulatory pathway for CGM systems [19]. In 2019 the FDA authorised the first interoperable insulin pump, also known as an alternate controller-enabled (ACE) insulin pump, which would allow PLWD to customise treatment through their diabetes management devices. This designation meant that an ACE insulin pump could be used with different components that make up diabetes therapy systems [20]. In 2023 the FDA approved the first interoperable automated glycaemic controller app [21]. Connected diabetes devices allow PLWD to tailor their diabetes management according to their preferences for devices.
Additionally, the FDA is also able to grant clearance for different device enhancements and new products, which is designed to ease the burden on clinical practices and allow for greater PLWD-centred care and self-management. Such features can also help PLWD to make better decisions about how and when to treat their diabetes, reduce some of the challenges associated with glucose management and reduce the burden of living with diabetes.

Cybersecurity

In 2022, the Institute of Electrical and Electronics Engineers (IEEE) Standards Association in the USA completed IEEE 2621, a standard for wireless diabetes device security assurance. This is the first and only medical device cybersecurity standard that comprehensively evaluates performance claims by manufacturers. It is focused on wireless diabetes devices and supports interoperability. This standard includes a certification programme built with industry consensus to ensure that IEEE-certified products continuously demonstrate conformance to the standard [22]. The standard was officially recognised by the FDA in 2022 [23]. No such standards exist in the EU.

Data ownership, privacy and security

Even though HCPs often use the data generated by medical devices and, while these are invaluable in terms of generating RWE, it is important to remember that the data are owned by the PLWD, and it is the PLWD who ultimately determines the extent to which the data are shared and can be used. That is, it is the user who consents to whether the data can be used for personal health management, research, product development or marketing. In some cases, and from a legal perspective, some may argue that questions from companies regarding data use are not straightforward, nor are the potential usages of data asked about separately. Some AID devices send data directly via a Subscriber Identification Module (SIM) card without the PLWD doing anything to the upload device; there is one AID system on the market that requires the user to be notified. To ensure essential understanding, data use questions should explicitly state the purpose and potential area of data usage. Furthermore, PLWD should be allowed to respond 'yes' or 'no' to each question. Currently, companies may ask for data for research purposes; however, no ethical approval is needed for such data use.
For data emanating from single or combined devices, security, privacy and data protection are of fundamental importance, particularly with the trend among manufacturers to use the data in RWE studies. The use of RWE is gaining increasing recognition in both clinical and political decision-making (by policymakers, regulators and payers), given that RWE studies often involve large datasets and include data from all people who have consented for their data to be used, which usually represents a more diverse cohort than those who meet inclusion criteria for clinical trials [24]. The advantage of RWE studies is that PLWD who are ineligible for clinical trials because of the inclusion/exclusion criteria (age, diagnosis of dementia, HbA1c cut-off values, prone to hypoglycaemia, presence of comorbidities) are included in such datasets. Such studies therefore provide valuable insight into the effectiveness of devices in a real-world environment. However, a key question regarding RWE relates to whether device users know how their data will be used. Users are likely to indicate a willingness to donate their data for research purposes, but it should be noted that the consent should be accompanied by information on data usage. Ethics committee approvals for large datasets that ensure privacy and anonymity are not commonly required for such studies.
Other concerns regarding data privacy relate to the potential use of data by entities such as insurance companies or courts of law. For example, if data from individuals were accessible to insurance companies, would these companies be able to modify coverage? Additionally, could users refuse to share their data with HCPs if CGM data are identifiable? What would be the implications of such a refusal to share data?

Impact of device interoperability on the device marketplace

Consumer and market forces influence interoperability; for example, a key incentive for component interoperability is the user’s ability to select specific components based on merit or reviews/opinions expressed online or on social media. Similar to other industries, the user may choose between different systems, supporting the need for different components to be available. While AID systems are considered medical devices first and foremost, they are also devices that the user has to interact with frequently during the day. In this way, AID systems may also be considered consumer devices. Therefore, providing users with a choice of components that best suits their lifestyles and wishes may help with ongoing engagement and use of the system to achieve therapeutic goals. This choice can be considered analogous to assembling a desktop computer with a choice of components compared with buying a fully assembled desktop. PLWD may want to choose a pump, an algorithm and a CGM based on their preferences, or a fully assembled system. It should be noted that neither option has been shown to be better than the other.
Interoperability between the different components needed to establish an AID system is a key hurdle for individuals who build their own systems. Such open-source AID systems are not addressed in detail here; however, their existence is in part due to the requirement for interoperability from PLWD. In the last few years a number of articles have been published that support some of the fundamental considerations presented here [25, 26].

Summary and conclusions

Transparent guidelines/standards developed using a stepwise approach should help ensure that innovations can be approved while also accounting for data ownership, safety of the devices used, and data safety and privacy. An open and transparent standard for data handling remains to be established, with the solution needing to fulfil the requirements of the reality of routine clinical practice. Only when data can be assessed in a standardised manner can the data generated by AID systems be integrated into EMRs. There is a need for an international committee on worldwide interoperability regulations; this role could be performed by an initiative of the WHO/IDF.
Through the combined efforts of many stakeholders (including legislative, regulatory and clinical stakeholders and medical device manufacturers), considerable progress has been made in formulating practical solutions for data and device interoperability in medical devices and data used in diabetes therapy. However, there is a need for professional diabetes associations and those directly involved in the care of PLWD to improve their understanding of the requirement to consent to analysis of their data. Furthermore, it is also important to make PLWD aware of how their data may be used with different degrees of anonymity and security, for example by hospitals, insurance companies and law enforcement agencies.
Finally, given the rapid pace of development of technology used in the management of PLWD, a regular update on data and device interoperability appears advisable.

Acknowledgements

The authors would like to thank O. Cohen and S. de Portu (Medtronic) for their administrative support.

Funding

Open access funding provided by Örebro University. Medtronic commissioned and funded medical writing support. J. Smith-Palmer (Covalence Research, UK) produced the first draft of this article. The authors received no honoraria.

Authors’ relationships and activities

JJ has been a lecturer/member of the scientific advisory boards at the following companies: Abbott, AstraZeneca, Boehringer Ingelheim, Eli Lilly, Medtronic, Nordic InfuCare, NovoNordisk and Sanofi. PA has received speaker’s honoraria from Dexcom, Eli Lilly, Insulet, NovoNordisk, Sanofi and Tandem. He has received consulting fees from and/or held advisory positions with Dexcom, Eli Lilly, Medtronic, NovoNordisk and Roche. PC has received personal fees from Medtronic, Abbott, Dexcom, Lilly, Sanofi, NovoNordisk, Glooko, DreaMed and MannKind. AF is the Executive Chairman of Kinexum, which advises multiple health product companies, including multiple insulin manufacturers. Relevant device companies advised include Abbott, Biolinq, CMC Magnetics, Diabeloop, Hagar, Know Labs, Modular Medical, SFC Fluidics and Surf Bio. He was formerly the Group Leader of the Division of Metabolism and Endocrine Drug Products at the US Food and Drug Administration. DCK is a consultant for Better Therapeutics, EOFlow, Integrity, Lifecare, Nevro, Novo Nordisk, Rockley Photonics and Sanofi. JKM is a member of the advisory boards of Abbott Diabetes Care, Becton Dickinson, Boehringer Ingelheim, Eli Lilly, Embecta, Medtronic, NovoNordisk, Roche Diabetes Care, Sanofi-Aventis and Viatris and has received speaker honoraria from A. Menarini Diagnostics, Abbott Diabetes Care, AstraZeneca, Boehringer Ingelheim, Dexcom, Eli Lilly, Medtrust, MSD, NovoNordisk, Roche Diabetes Care, Sanofi, Servier and Ypsomed. She is a shareholder of decide Clinical Software GmbH and elyte diagnostics, where she also serves as Chief Medical Officer. KD served on the advisory boards of Novo Nordisk, Pfizer and Sanofi. KD has also received honoraria for participation in the speaker’s bureaus of Abbott, Eli Lilly, Medtronic, NovoNordisk and Pfizer. NO has participated on the advisory boards of Medtronic, Roche Diabetes Care and Dexcom. He has received speaker fees from Dexcom and research funding from Medtronic Diabetes, Roche Diabetes Care and Dexcom. JLS has conducted clinical trials for Eli Lilly, Insulet, Medtronic and Provention Bio and has received in-kind support for research studies from Dexcom and Medtronic. She has consulted for and/or received speaker honoraria from Eli Lilly, Lexicon, Insulet, Medtronic and Zealand. She has been a member of the advisory boards of Bigfoot Biomedical, Cecelia Health, Eli Lilly, Insulet, StartUp Health’s T1D Moonshot, the T1D Fund and Vertex. JŠ has received speaker honoraria from and consulted for Abbott, Dexcom, Eli Lilly, Medtronic, NovoNordisk and Roche. LH holds shares in the Profil Institute for Metabolic Research, Neuss, Germany. LH is also a consultant for embecta, Lifecare and Roche Diabetes Care.

Contribution statement

All authors were responsible for drafting the article and reviewing it critically for important intellectual content. All authors approved the version to be published.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

e.Med Allgemeinmedizin

Kombi-Abonnement

Mit e.Med Allgemeinmedizin erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der allgemeinmedizinischen Zeitschriften, inklusive einer gedruckten Allgemeinmedizin-Zeitschrift Ihrer Wahl.

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
11.
Zurück zum Zitat Sherr JL, Heinemann L, Fleming GA et al (2023) Automated insulin delivery: benefits, challenges, and recommendations. A Consensus Report of the Joint Diabetes Technology Working Group of the European Association for the Study of Diabetes and the American Diabetes Association. Diabetologia 66(1):3–22. https://doi.org/10.1007/s00125-022-05744-zCrossRefPubMed Sherr JL, Heinemann L, Fleming GA et al (2023) Automated insulin delivery: benefits, challenges, and recommendations. A Consensus Report of the Joint Diabetes Technology Working Group of the European Association for the Study of Diabetes and the American Diabetes Association. Diabetologia 66(1):3–22. https://​doi.​org/​10.​1007/​s00125-022-05744-zCrossRefPubMed
32.
Zurück zum Zitat Adolfsson P (2022) Smart futures: creating a digital ecosystem for diabetes. A digital dawn for diabetes: how data can transform patient care (Industry Symposium oral presentation, supported by Novo Nordisk). 15th International Conference on Advanced Technologies & Treatments for Diabetes, 27–30 April, Barcelona, Spain. 2022 Adolfsson P (2022) Smart futures: creating a digital ecosystem for diabetes. A digital dawn for diabetes: how data can transform patient care (Industry Symposium oral presentation, supported by Novo Nordisk). 15th International Conference on Advanced Technologies & Treatments for Diabetes, 27–30 April, Barcelona, Spain. 2022
Metadaten
Titel
A narrative commentary about interoperability in medical devices and data used in diabetes therapy from an academic EU/UK/US perspective
verfasst von
Johan Jendle
Peter Adolfsson
Pratik Choudhary
Klemen Dovc
Alexander Fleming
David C. Klonoff
Julia K. Mader
Nick Oliver
Jennifer L. Sherr
Jan Šoupal
Lutz Heinemann
Publikationsdatum
02.12.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Diabetologia / Ausgabe 2/2024
Print ISSN: 0012-186X
Elektronische ISSN: 1432-0428
DOI
https://doi.org/10.1007/s00125-023-06049-5

Weitere Artikel der Ausgabe 2/2024

Diabetologia 2/2024 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

RAS-Blocker bei Hyperkaliämie möglichst nicht sofort absetzen

14.05.2024 Hyperkaliämie Nachrichten

Bei ausgeprägter Nierenfunktionsstörung steigen unter der Einnahme von Renin-Angiotensin-System(RAS)-Hemmstoffen nicht selten die Serumkaliumspiegel. Was in diesem Fall zu tun ist, erklärte Prof. Jürgen Floege beim diesjährigen Allgemeinmedizin-Update-Seminar.

Gestationsdiabetes: In der zweiten Schwangerschaft folgenreicher als in der ersten

13.05.2024 Gestationsdiabetes Nachrichten

Das Risiko, nach einem Gestationsdiabetes einen Typ-2-Diabetes zu entwickeln, hängt nicht nur von der Zahl, sondern auch von der Reihenfolge der betroffenen Schwangerschaften ab.

Labor, CT-Anthropometrie zeigen Risiko für Pankreaskrebs

13.05.2024 Pankreaskarzinom Nachrichten

Gerade bei aggressiven Malignomen wie dem duktalen Adenokarzinom des Pankreas könnte Früherkennung die Therapiechancen verbessern. Noch jedoch klafft hier eine Lücke. Ein Studienteam hat einen Weg gesucht, sie zu schließen.

Battle of Experts: Sport vs. Spritze bei Adipositas und Typ-2-Diabetes

11.05.2024 DDG-Jahrestagung 2024 Kongressbericht

Im Battle of Experts traten zwei Experten auf dem Diabeteskongress gegeneinander an: Die eine vertrat die Auffassung „Sport statt Spritze“ bei Adipositas und Typ-2-Diabetes, der andere forderte „Spritze statt Sport!“ Am Ende waren sie sich aber einig: Die Kombination aus beidem erzielt die besten Ergebnisse.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.