11.03.2023 | Original Article

Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study

verfasst von: Masashi Takakura, Tasuku Fujii, Tomoya Taniguchi, Shogo Suzuki, Kimitoshi Nishiwaki

Erschienen in: Journal of Anesthesia | Ausgabe 3/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The estimated continuous cardiac output (esCCO) system was recently developed as a noninvasive hemodynamic monitoring alternative to the thermodilution cardiac output (TDCO). However, the accuracy of continuous cardiac output measurements by the esCCO system compared to TDCO under different respiratory conditions remains unclear. This prospective study aimed to assess the clinical accuracy of the esCCO system by continuously measuring the esCCO and TDCO.

Methods

Forty patients who had undergone cardiac surgery with a pulmonary artery catheter were enrolled. We compared the esCCO with TDCO from mechanical ventilation to spontaneous respiration through extubation. Patients undergoing cardiac pacing during esCCO measurement, those receiving treatment with an intra-aortic balloon pump, and those with measurement errors or missing data were excluded. In total, 23 patients were included. Agreement between the esCCO and TDCO measurements was evaluated using Bland–Altman analysis with a 20 min moving average of the esCCO.

Results

The paired esCCO and TDCO measurements (939 points before extubation and 1112 points after extubation) were compared. The respective bias and standard deviation (SD) values were 0.13 L/min and 0.60 L/min before extubation, and − 0.48 L/min and 0.78 L/min after extubation. There was a significant difference in bias before and after extubation (P < 0.001); the SD before and after extubation was not significant (P = 0.315). The percentage errors were 25.1% before extubation and 29.6% after extubation, which is the criterion for acceptance of a new technique.

Conclusion

The accuracy of the esCCO system is clinically acceptable to that of TDCO under mechanical ventilation and spontaneous respiration.
Literatur
1.
Zurück zum Zitat Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ. A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol. 1971;27:392–6.CrossRefPubMed Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ. A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol. 1971;27:392–6.CrossRefPubMed
2.
Zurück zum Zitat Connors AF Jr, McCaffree DR, Gray BA. Evaluation of right-heart catheterization in the critically ill patient without acute myocardial infarction. N Engl J Med. 1983;308:263–7.CrossRefPubMed Connors AF Jr, McCaffree DR, Gray BA. Evaluation of right-heart catheterization in the critically ill patient without acute myocardial infarction. N Engl J Med. 1983;308:263–7.CrossRefPubMed
3.
Zurück zum Zitat Streisand JB, Clark NJ, Pace NL. Pulmonary arterial catheterization before anesthesia in patients undergoing cardiac surgery. Placement of the pulmonary arterial catheter before anesthesia for cardiac surgery: safe, intelligent, and appropriate use of invasive hemodynamic monitoring. J Clin Monit. 1985;1:193–7.CrossRefPubMed Streisand JB, Clark NJ, Pace NL. Pulmonary arterial catheterization before anesthesia in patients undergoing cardiac surgery. Placement of the pulmonary arterial catheter before anesthesia for cardiac surgery: safe, intelligent, and appropriate use of invasive hemodynamic monitoring. J Clin Monit. 1985;1:193–7.CrossRefPubMed
4.
Zurück zum Zitat Sharma J, Bhise M, Singh A, Mehta Y, Trehan N. Hemodynamic measurements after cardiac surgery: transesophageal Doppler versus pulmonary artery catheter. J Cardiothorac Vasc Anesth. 2005;19:746–50.CrossRefPubMed Sharma J, Bhise M, Singh A, Mehta Y, Trehan N. Hemodynamic measurements after cardiac surgery: transesophageal Doppler versus pulmonary artery catheter. J Cardiothorac Vasc Anesth. 2005;19:746–50.CrossRefPubMed
5.
Zurück zum Zitat Rajaram SS, Desai NK, Kalra A, Gajera M, Cavanaugh SK, Brampton W, Young D, Harvey S, Rowan K. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev. 2013;2013:CD003408.PubMedPubMedCentral Rajaram SS, Desai NK, Kalra A, Gajera M, Cavanaugh SK, Brampton W, Young D, Harvey S, Rowan K. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev. 2013;2013:CD003408.PubMedPubMedCentral
6.
Zurück zum Zitat Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003;348:5–14.CrossRefPubMed Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003;348:5–14.CrossRefPubMed
7.
Zurück zum Zitat Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K, PAC-Man Study Collaboration. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet. 2005;366:472–7.CrossRefPubMed Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K, PAC-Man Study Collaboration. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet. 2005;366:472–7.CrossRefPubMed
8.
Zurück zum Zitat Bossert T, Gummert JF, Bittner HB, Barten M, Walther T, Falk V, Mohr FW. Swan-Ganz catheter-induced severe complications in cardiac surgery: right ventricular perforation, knotting, and rupture of a pulmonary artery. J Card Surg. 2006;21:292–5.CrossRefPubMed Bossert T, Gummert JF, Bittner HB, Barten M, Walther T, Falk V, Mohr FW. Swan-Ganz catheter-induced severe complications in cardiac surgery: right ventricular perforation, knotting, and rupture of a pulmonary artery. J Card Surg. 2006;21:292–5.CrossRefPubMed
9.
Zurück zum Zitat Xu Y, Stout LC. Complete step section microscopic study of a Swan-Ganz catheter-related pulmonary artery rupture: a frequently lethal complication that to our knowledge has not had a comprehensive microscopic examination: case report and literature review. Cardiovasc Pathol. 2017;31:17–9.CrossRefPubMed Xu Y, Stout LC. Complete step section microscopic study of a Swan-Ganz catheter-related pulmonary artery rupture: a frequently lethal complication that to our knowledge has not had a comprehensive microscopic examination: case report and literature review. Cardiovasc Pathol. 2017;31:17–9.CrossRefPubMed
10.
Zurück zum Zitat Procaccini B, Clementi G. Pulmonary artery catheterization in 9071 cardiac surgery patients: a review of complications (in Italian). Ital Heart J Suppl. 2004;5:891–9.PubMed Procaccini B, Clementi G. Pulmonary artery catheterization in 9071 cardiac surgery patients: a review of complications (in Italian). Ital Heart J Suppl. 2004;5:891–9.PubMed
11.
Zurück zum Zitat Schwann NM, Hillel Z, Hoeft A, Barash P, Möhnle P, Miao Y, Mangano DT. Lack of effectiveness of the pulmonary artery catheter in cardiac surgery. Anesth Analg. 2011;113:994–1002.CrossRefPubMed Schwann NM, Hillel Z, Hoeft A, Barash P, Möhnle P, Miao Y, Mangano DT. Lack of effectiveness of the pulmonary artery catheter in cardiac surgery. Anesth Analg. 2011;113:994–1002.CrossRefPubMed
12.
Zurück zum Zitat Barone JE, Tucker JB, Rassias D, Corvo PR. Routine perioperative pulmonary artery catheterization has no effect on rate of complications in vascular surgery: a meta-analysis. Am Surg. 2001;67:674–9.CrossRefPubMed Barone JE, Tucker JB, Rassias D, Corvo PR. Routine perioperative pulmonary artery catheterization has no effect on rate of complications in vascular surgery: a meta-analysis. Am Surg. 2001;67:674–9.CrossRefPubMed
13.
Zurück zum Zitat Marik PE. Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth. 2013;27:121–34.CrossRefPubMed Marik PE. Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth. 2013;27:121–34.CrossRefPubMed
14.
Zurück zum Zitat Ball TR, Tricinella AP, Kimbrough BA, Luna S, Gloyna DF, Villamaria FJ, Culp WC Jr. Accuracy of noninvasive estimated continuous cardiac output (esCCO) compared to thermodilution cardiac output: a pilot study in cardiac patients. J Cardiothorac Vasc Anesth. 2013;27:1128–32.CrossRefPubMed Ball TR, Tricinella AP, Kimbrough BA, Luna S, Gloyna DF, Villamaria FJ, Culp WC Jr. Accuracy of noninvasive estimated continuous cardiac output (esCCO) compared to thermodilution cardiac output: a pilot study in cardiac patients. J Cardiothorac Vasc Anesth. 2013;27:1128–32.CrossRefPubMed
15.
Zurück zum Zitat Yamada T, Tsutsui M, Sugo Y, Sato T, Akazawa T, Sato N, Yamashita K, Ishihara H, Takeda J. Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output. Anesth Analg. 2012;115:82–7.CrossRefPubMed Yamada T, Tsutsui M, Sugo Y, Sato T, Akazawa T, Sato N, Yamashita K, Ishihara H, Takeda J. Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output. Anesth Analg. 2012;115:82–7.CrossRefPubMed
16.
Zurück zum Zitat Tsutsui M, Araki Y, Masui K, Kazama T, Sugo Y, Archer TL, Manecke GR Jr. Pulse wave transit time measurements of cardiac output in patients undergoing partial hepatectomy: a comparison of the esCCO system with thermodilution. Anesth Analg. 2013;117:1307–12.CrossRefPubMed Tsutsui M, Araki Y, Masui K, Kazama T, Sugo Y, Archer TL, Manecke GR Jr. Pulse wave transit time measurements of cardiac output in patients undergoing partial hepatectomy: a comparison of the esCCO system with thermodilution. Anesth Analg. 2013;117:1307–12.CrossRefPubMed
17.
Zurück zum Zitat Xue H, Ruan M, Cheng Y. A fixed length adaptive moving average filter-based synchrophasor measurement algorithm for P class PMUs. Energies. 2019;12:4168.CrossRef Xue H, Ruan M, Cheng Y. A fixed length adaptive moving average filter-based synchrophasor measurement algorithm for P class PMUs. Energies. 2019;12:4168.CrossRef
18.
Zurück zum Zitat Aranda M, Mihm FG, Garrett S, Mihm MN, Pearl RG. Continuous cardiac output catheters: delay in in vitro response time after controlled flow changes. Anesthesiology. 1998;89:1592–5.PubMed Aranda M, Mihm FG, Garrett S, Mihm MN, Pearl RG. Continuous cardiac output catheters: delay in in vitro response time after controlled flow changes. Anesthesiology. 1998;89:1592–5.PubMed
19.
Zurück zum Zitat Siegel LC, Hennessy MM, Pearl RG. Delayed time response of the continuous cardiac output pulmonary artery catheter. Anesth Analg. 1996;83:1173–7.CrossRefPubMed Siegel LC, Hennessy MM, Pearl RG. Delayed time response of the continuous cardiac output pulmonary artery catheter. Anesth Analg. 1996;83:1173–7.CrossRefPubMed
20.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
21.
Zurück zum Zitat Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–60.CrossRefPubMed Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–60.CrossRefPubMed
22.
Zurück zum Zitat Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRefPubMed Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRefPubMed
23.
Zurück zum Zitat Kirkeby-Garstad I, Trønnes H, Stenseth R, Sellevold OF, Aadahl P, Skogvoll E. The precision of pulmonary artery catheter bolus thermodilution cardiac output measurements varies with the clinical situation. J Cardiothorac Vasc Anesth. 2015;29:881–8.CrossRefPubMed Kirkeby-Garstad I, Trønnes H, Stenseth R, Sellevold OF, Aadahl P, Skogvoll E. The precision of pulmonary artery catheter bolus thermodilution cardiac output measurements varies with the clinical situation. J Cardiothorac Vasc Anesth. 2015;29:881–8.CrossRefPubMed
24.
Zurück zum Zitat Kubitz JC, Kemming GI, Schultheib G, Starke J, Podtschaske A, Goetz AE, Reuter DA. The influence of cardiac preload and positive end-expiratory pressure on the pre-ejection period. Physiol Meas. 2005;26:1033–8.CrossRefPubMed Kubitz JC, Kemming GI, Schultheib G, Starke J, Podtschaske A, Goetz AE, Reuter DA. The influence of cardiac preload and positive end-expiratory pressure on the pre-ejection period. Physiol Meas. 2005;26:1033–8.CrossRefPubMed
25.
Zurück zum Zitat Sugo Y, Ukawa T, Takeda S, Ishihara H, Kazama T, Takeda J. A novel continuous cardiac output monitor based on pulse wave transit time. Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:2853–6.PubMed Sugo Y, Ukawa T, Takeda S, Ishihara H, Kazama T, Takeda J. A novel continuous cardiac output monitor based on pulse wave transit time. Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:2853–6.PubMed
Metadaten
Titel
Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study
verfasst von
Masashi Takakura
Tasuku Fujii
Tomoya Taniguchi
Shogo Suzuki
Kimitoshi Nishiwaki
Publikationsdatum
11.03.2023
Verlag
Springer Nature Singapore
Erschienen in
Journal of Anesthesia / Ausgabe 3/2023
Print ISSN: 0913-8668
Elektronische ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-023-03176-6

Weitere Artikel der Ausgabe 3/2023

Journal of Anesthesia 3/2023 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Häufigste Gründe für Brustschmerzen bei Kindern

06.05.2024 Pädiatrische Diagnostik Nachrichten

Akute Brustschmerzen sind ein Alarmsymptom par exellence, schließlich sind manche Auslöser lebensbedrohlich. Auch Kinder klagen oft über Schmerzen in der Brust. Ein Studienteam ist den Ursachen nachgegangen.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.