Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects

verfasst von: Yoshitake Kato, Shunsuke Nakakura, Naoko Matsuo, Kayo Yoshitomi, Marina Handa, Hitoshi Tabuchi, Yoshiaki Kiuchi

Erschienen in: International Ophthalmology | Ausgabe 2/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the inter-device agreement among the Goldmann applanation tonometer (GAT), iCare and Icare PRO rebound tonometers, non-contact tonometer (NCT), and Tonopen XL tonometer.

Methods

Sixty healthy elderly subjects were enrolled. The intraocular pressure (IOP) in each subject’s right eye was measured thrice using each of the five tonometers. Intra-device agreement was evaluated by calculating intraclass correlation coefficients (ICCs). Inter-device agreement was evaluated by ICC and Bland–Altman analyses.

Results

ICCs for intra-device agreement for each tonometer were >0.8. IOP as measured by iCare (mean ± SD, 11.6 ± 2.5 mmHg) was significantly lower (p < 0.05) than that measured by GAT (14.0 ± 2.8 mmHg), NCT (13.6 ± 2.5 mmHg), Tonopen XL (13.7 ± 4.1 mmHg), and Icare PRO (12.6 ± 2.2 mmHg; Bonferroni test). There was no significant difference in mean IOP among GAT, NCT, and Tonopen XL. Regarding inter-device agreement, ICC was lower between Tonopen XL and other tonometers (all ICCs < 0.4). However, ICCs of GAT, iCare, Icare PRO, and NCT showed good agreement (0.576–0.700). The Bland–Altman analysis revealed that the width of the 95% limits of agreement was larger between the Tonopen XL and the other tonometers ranged from 14.94 to 16.47 mmHg. Among the other tonometers, however, the widths of 95% limits of agreement ranged from 7.91 to 9.24 mmHg.

Conclusion

There was good inter-device agreement among GAT, rebound tonometers, and NCT. Tonopen XL shows the worst agreement with the other tonometers; therefore, we should pay attention to its’ respective IOP.

Clinical trial registration

Japan Clinical Trials Register; number: UMIN000011544.
Literatur
1.
Zurück zum Zitat Goldmann H (1955) Un nouveau tonometre d’applanation. Bull Soc Ophtalmol Fr 67:474–478 Goldmann H (1955) Un nouveau tonometre d’applanation. Bull Soc Ophtalmol Fr 67:474–478
2.
Zurück zum Zitat Stamper RL (2011) A history of intraocular pressure and its measurement. Optom Vis Sci 88:16–28CrossRef Stamper RL (2011) A history of intraocular pressure and its measurement. Optom Vis Sci 88:16–28CrossRef
3.
Zurück zum Zitat Sahin A, Basmak H, Niyaz L, Yildirim N (2007) Reproducibility and tolerability of the Icare rebound tonometer in school children. J Glaucoma 16:185–188CrossRefPubMed Sahin A, Basmak H, Niyaz L, Yildirim N (2007) Reproducibility and tolerability of the Icare rebound tonometer in school children. J Glaucoma 16:185–188CrossRefPubMed
4.
Zurück zum Zitat Grigorian F, Grigorian AP, Olitsky SE (2012) The use of the Icare tonometer reduced the need for anesthesia to measure intraocular pressure in children. J AAPOS 16:508–510CrossRefPubMed Grigorian F, Grigorian AP, Olitsky SE (2012) The use of the Icare tonometer reduced the need for anesthesia to measure intraocular pressure in children. J AAPOS 16:508–510CrossRefPubMed
5.
Zurück zum Zitat Nakakura S, Mori E, Yamamoto M et al (2015) Intradevice and interdevice agreement between a rebound tonometer, Icare PRO, and the Tonopen XL and Kowa Hand-held applanation tonometer when used in the sitting and supine position. J Glaucoma 24:515–521CrossRefPubMed Nakakura S, Mori E, Yamamoto M et al (2015) Intradevice and interdevice agreement between a rebound tonometer, Icare PRO, and the Tonopen XL and Kowa Hand-held applanation tonometer when used in the sitting and supine position. J Glaucoma 24:515–521CrossRefPubMed
6.
Zurück zum Zitat Scuderi GL, Cascone NC, Regine F et al (2011) Validity and limits of the rebound tonometer (Icare®): clinical study. Eur J Ophthalmol 21:251–257CrossRefPubMed Scuderi GL, Cascone NC, Regine F et al (2011) Validity and limits of the rebound tonometer (Icare®): clinical study. Eur J Ophthalmol 21:251–257CrossRefPubMed
7.
Zurück zum Zitat Nakamura M, Darhad U, Tatsumi Y et al (2006) Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. Am J Ophthalmol 142:332–334CrossRefPubMed Nakamura M, Darhad U, Tatsumi Y et al (2006) Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. Am J Ophthalmol 142:332–334CrossRefPubMed
8.
Zurück zum Zitat Poostchi A, Mitchell R, Nicholas S et al (2009) The Icare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 37:687–691CrossRefPubMed Poostchi A, Mitchell R, Nicholas S et al (2009) The Icare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 37:687–691CrossRefPubMed
9.
Zurück zum Zitat Feng CS, Jin KW, Yi K et al (2015) Comparison of intraocular pressure measurements obtained by rebound, noncontact, and Goldmann applanation tonometry in children. Am J Ophthalmol 160:937–943CrossRefPubMed Feng CS, Jin KW, Yi K et al (2015) Comparison of intraocular pressure measurements obtained by rebound, noncontact, and Goldmann applanation tonometry in children. Am J Ophthalmol 160:937–943CrossRefPubMed
10.
Zurück zum Zitat Tamçelik N, Atalay E, Cicik E et al (2015) Comparability of Icare Pro rebound tonometer with Goldmann applanation and noncontact tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res 54:18–25CrossRefPubMed Tamçelik N, Atalay E, Cicik E et al (2015) Comparability of Icare Pro rebound tonometer with Goldmann applanation and noncontact tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res 54:18–25CrossRefPubMed
11.
Zurück zum Zitat Lee TE, Yoo C, Hwang JY et al (2015) Comparison of intraocular pressure measurements between Icare Pro rebound tonometer and Tono-Pen XL tonometer in supine and lateral decubitus body positions. Curr Eye Res 40:923–929CrossRefPubMed Lee TE, Yoo C, Hwang JY et al (2015) Comparison of intraocular pressure measurements between Icare Pro rebound tonometer and Tono-Pen XL tonometer in supine and lateral decubitus body positions. Curr Eye Res 40:923–929CrossRefPubMed
12.
Zurück zum Zitat Güler M, Bilak Ş, Bilgin B et al (2015) Comparison of intraocular pressure measurements obtained by Icare PRO rebound tonometer, Tomey FT-1000 noncontact tonometer, and Goldmann applanation tonometer in healthy subjects. J Glaucoma 24:613–618CrossRefPubMed Güler M, Bilak Ş, Bilgin B et al (2015) Comparison of intraocular pressure measurements obtained by Icare PRO rebound tonometer, Tomey FT-1000 noncontact tonometer, and Goldmann applanation tonometer in healthy subjects. J Glaucoma 24:613–618CrossRefPubMed
13.
Zurück zum Zitat Nakakura S, Mori E, Yamamoto M et al (2013) Intraocular pressure of supine patients using four portable tonometers. Optom Vis Sci 90:700–706CrossRefPubMed Nakakura S, Mori E, Yamamoto M et al (2013) Intraocular pressure of supine patients using four portable tonometers. Optom Vis Sci 90:700–706CrossRefPubMed
14.
Zurück zum Zitat McGraw KO, Wong SP (1996) Forming inferences about some intraclass coefficients. Psychol Methods 1:30–46CrossRef McGraw KO, Wong SP (1996) Forming inferences about some intraclass coefficients. Psychol Methods 1:30–46CrossRef
15.
Zurück zum Zitat Cook JA, Botello AP, Elders A, Surveillance of Ocular Hypertension Study Group et al (2012) Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology 119:1552–1557CrossRefPubMed Cook JA, Botello AP, Elders A, Surveillance of Ocular Hypertension Study Group et al (2012) Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology 119:1552–1557CrossRefPubMed
16.
Zurück zum Zitat Poostchi A, Mitchell R, Nicholas S et al (2009) The Icare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 37:687–691CrossRefPubMed Poostchi A, Mitchell R, Nicholas S et al (2009) The Icare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 37:687–691CrossRefPubMed
17.
Zurück zum Zitat Sakamoto M, Kanamori A, Fujihara M et al (2014) Assessment of IcareONE rebound tonometer for self-measuring intraocular pressure. Acta Ophthalmol 92:243–248CrossRefPubMed Sakamoto M, Kanamori A, Fujihara M et al (2014) Assessment of IcareONE rebound tonometer for self-measuring intraocular pressure. Acta Ophthalmol 92:243–248CrossRefPubMed
18.
Zurück zum Zitat Moreno-Montañés J, Martínez-de-la-Casa JM, Sabater AL et al (2015) Clinical evaluation of the new rebound tonometers Icare PRO and Icare ONE compared with the Goldmann tonometer. J Glaucoma 24:527–532CrossRefPubMed Moreno-Montañés J, Martínez-de-la-Casa JM, Sabater AL et al (2015) Clinical evaluation of the new rebound tonometers Icare PRO and Icare ONE compared with the Goldmann tonometer. J Glaucoma 24:527–532CrossRefPubMed
19.
Zurück zum Zitat Noguchi A, Nakakura S, Fujio Y et al (2016) A pilot evaluation assessing the ease of use and accuracy of the new self/home-tonometer IcareHOME in healthy young subjects. J Glaucoma 25(10):835–841CrossRefPubMed Noguchi A, Nakakura S, Fujio Y et al (2016) A pilot evaluation assessing the ease of use and accuracy of the new self/home-tonometer IcareHOME in healthy young subjects. J Glaucoma 25(10):835–841CrossRefPubMed
20.
Zurück zum Zitat Chihara E (2008) Assessment of true intraocular pressure: the gap between theory and practical data. Surv Ophthalmol 53:203–218CrossRefPubMed Chihara E (2008) Assessment of true intraocular pressure: the gap between theory and practical data. Surv Ophthalmol 53:203–218CrossRefPubMed
21.
Zurück zum Zitat Mark HH, Robbins KP, Mark TL (2002) Axial length in applanation tonometry. J Cataract Refract Surg 28:504–506CrossRefPubMed Mark HH, Robbins KP, Mark TL (2002) Axial length in applanation tonometry. J Cataract Refract Surg 28:504–506CrossRefPubMed
22.
Zurück zum Zitat Wong TY, Klein BE, Klein R et al (2003) Refractive errors, intraocular pressure, and glaucoma in a white population. Ophthalmology 110:211–217CrossRefPubMed Wong TY, Klein BE, Klein R et al (2003) Refractive errors, intraocular pressure, and glaucoma in a white population. Ophthalmology 110:211–217CrossRefPubMed
23.
Zurück zum Zitat Asaoka R, Nakakura S, Tabuchi H et al (2015) The relationship between Corvis ST tonometry measured corneal parameters and intraocular pressure, corneal thickness and corneal curvature. PLoS ONE 10:e0140385CrossRefPubMedPubMedCentral Asaoka R, Nakakura S, Tabuchi H et al (2015) The relationship between Corvis ST tonometry measured corneal parameters and intraocular pressure, corneal thickness and corneal curvature. PLoS ONE 10:e0140385CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Kling S, Marcos S (2013) Contributing factors to corneal deformation in air puff measurements. Invest Ophthalmol Vis Sci 54:5078–5085CrossRefPubMed Kling S, Marcos S (2013) Contributing factors to corneal deformation in air puff measurements. Invest Ophthalmol Vis Sci 54:5078–5085CrossRefPubMed
25.
Zurück zum Zitat Strobbe E, Cellini M, Barbaresi U et al (2014) Influence of age and gender on corneal biomechanical properties in a healthy Italian population. Cornea 33:968–972CrossRefPubMed Strobbe E, Cellini M, Barbaresi U et al (2014) Influence of age and gender on corneal biomechanical properties in a healthy Italian population. Cornea 33:968–972CrossRefPubMed
26.
Zurück zum Zitat Tian L, Wang D, Wu Y et al (2016) Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmol 94:e317–e324CrossRefPubMed Tian L, Wang D, Wu Y et al (2016) Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmol 94:e317–e324CrossRefPubMed
27.
Zurück zum Zitat Shin J, Lee JW, Kim EA, Caprioli J (2015) The effect of corneal biomechanical properties on rebound tonometer in patients with normal-tension glaucoma. Am J Ophthalmol 159:144–154CrossRefPubMed Shin J, Lee JW, Kim EA, Caprioli J (2015) The effect of corneal biomechanical properties on rebound tonometer in patients with normal-tension glaucoma. Am J Ophthalmol 159:144–154CrossRefPubMed
28.
Zurück zum Zitat Kim NR, Kim CY, Kim H et al (2011) Comparison of Goldmann applanation tonometer, noncontact tonometer, and TonoPen XL for intraocular pressure measurement in different types of glaucomatous, ocular hypertensive, and normal eyes. Curr Eye Res 36:295–300CrossRefPubMed Kim NR, Kim CY, Kim H et al (2011) Comparison of Goldmann applanation tonometer, noncontact tonometer, and TonoPen XL for intraocular pressure measurement in different types of glaucomatous, ocular hypertensive, and normal eyes. Curr Eye Res 36:295–300CrossRefPubMed
29.
Zurück zum Zitat Iwase A, Suzuki Y, Araie M et al (2004) The prevalence of primary open-angle glaucoma in Japanese: the Tajimi Study. Ophthalmology 111:1641–1648PubMed Iwase A, Suzuki Y, Araie M et al (2004) The prevalence of primary open-angle glaucoma in Japanese: the Tajimi Study. Ophthalmology 111:1641–1648PubMed
Metadaten
Titel
Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects
verfasst von
Yoshitake Kato
Shunsuke Nakakura
Naoko Matsuo
Kayo Yoshitomi
Marina Handa
Hitoshi Tabuchi
Yoshiaki Kiuchi
Publikationsdatum
09.04.2017
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 2/2018
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-017-0518-2

Weitere Artikel der Ausgabe 2/2018

International Ophthalmology 2/2018 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

CME: Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Wird ein Glaukom chirurgisch behandelt, ist die anschließende Wundheilung von entscheidender Bedeutung. In diesem CME-Kurs lernen Sie, welche Pathomechanismen der Vernarbung zugrunde liegen, wie perioperativ therapiert und Operationsversagen frühzeitig erkannt werden kann.

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.