Assessment of the myelin water fraction in rodent spinal cord using T2-prepared ultrashort echo time MRI

verfasst von: Tim Klasen, Cornelius Faber

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 6/2016

Einloggen, um Zugang zu erhalten

Abstract

Objective

Multi-component T2 relaxation allows for assessing the myelin water fraction in nervous tissue, providing a surrogate marker for demyelination. The assessment of the number and distribution of different T2 components for devising exact models of tissue relaxation has been limited by T2 sampling with conventional MR methods.

Materials and methods

A T2-prepared UTE sequence was used to assess multicomponent T2 relaxation at 9.4 T of fixed mouse and rat spinal cord samples and of mouse spinal cord in vivo. For in vivo scans, a cryogenically cooled probe allowed for 78-µm resolution in 1-mm slices. Voxel-wise non-negative least square analysis was used to assess the number of myelin water-associated T2 components.

Results

More than one myelin water-associated T2 component was detected in only 12 % of analyzed voxels in rat spinal cords and 6 % in mouse spinal cords, both in vivo and in vitro. However, myelin water-associated T2 values of individual voxels varied between 0.1 and 20 ms. While in fixed samples almost no components below 1 ms were identified, in vivo, these contributed 14 % of the T2 spectrum. No significant differences in MWF were observed in mouse spinal cord in vivo versus ex vivo measurements.

Conclusion

Voxel-wise analysis methods using relaxation models with one myelin water-associated T2 component are appropriate for assessing myelin content of nervous tissue.
Literatur
1.
Zurück zum Zitat McDonald WI, Compston A, Edan G, Goodkin D, Hartung HP, Lublin FD, McFarland HF, Paty DW, Polman CH, Reingold SC, Sandberg-Wollheim M, Sibley W, Thompson A, van den Noort S, Weinshenker BY, Wolinsky JS (2001) Recommended diagnostic criteria for multiple sclerosis: guidelines from the International Panel on the diagnosis of multiple sclerosis. Ann Neurol 50(1):121–127CrossRefPubMed McDonald WI, Compston A, Edan G, Goodkin D, Hartung HP, Lublin FD, McFarland HF, Paty DW, Polman CH, Reingold SC, Sandberg-Wollheim M, Sibley W, Thompson A, van den Noort S, Weinshenker BY, Wolinsky JS (2001) Recommended diagnostic criteria for multiple sclerosis: guidelines from the International Panel on the diagnosis of multiple sclerosis. Ann Neurol 50(1):121–127CrossRefPubMed
2.
Zurück zum Zitat Tofts PS, Kermode AG (1991) Measurement of the blood-brain barrier permeability and leakage space using dynamic MR imaging. 1. Fundamental concepts. Magn Reson Med 17(2):357–367CrossRefPubMed Tofts PS, Kermode AG (1991) Measurement of the blood-brain barrier permeability and leakage space using dynamic MR imaging. 1. Fundamental concepts. Magn Reson Med 17(2):357–367CrossRefPubMed
3.
Zurück zum Zitat Ingrisch M, Sourbron S, Morhard D, Ertl-Wagner B, Kumpfel T, Hohlfeld R, Reiser M, Glaser C (2012) Quantification of perfusion and permeability in multiple sclerosis: dynamic contrast-enhanced MRI in 3D at 3T. Invest Radiol 47(4):252–258CrossRefPubMed Ingrisch M, Sourbron S, Morhard D, Ertl-Wagner B, Kumpfel T, Hohlfeld R, Reiser M, Glaser C (2012) Quantification of perfusion and permeability in multiple sclerosis: dynamic contrast-enhanced MRI in 3D at 3T. Invest Radiol 47(4):252–258CrossRefPubMed
4.
Zurück zum Zitat Cramer SP, Larsson HB (2014) Accurate determination of blood-brain barrier permeability using dynamic contrast-enhanced T1-weighted MRI: a simulation and in vivo study on healthy subjects and multiple sclerosis patients. J Cereb Blood Flow Metab 34(10):1655–1665CrossRefPubMedPubMedCentral Cramer SP, Larsson HB (2014) Accurate determination of blood-brain barrier permeability using dynamic contrast-enhanced T1-weighted MRI: a simulation and in vivo study on healthy subjects and multiple sclerosis patients. J Cereb Blood Flow Metab 34(10):1655–1665CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat MacKay A, Laule C, Vavasour I, Bjarnason T, Kolind S, Madler B (2006) Insights into brain microstructure from the T2 distribution. Magn Reson Imaging 24(4):515–525CrossRefPubMed MacKay A, Laule C, Vavasour I, Bjarnason T, Kolind S, Madler B (2006) Insights into brain microstructure from the T2 distribution. Magn Reson Imaging 24(4):515–525CrossRefPubMed
6.
Zurück zum Zitat Laule C, Vavasour IM, Kolind SH, Li DK, Traboulsee TL, Moore GR, MacKay AL (2007) Magnetic resonance imaging of myelin. Neurotherapeutics 4(3):460–484CrossRefPubMed Laule C, Vavasour IM, Kolind SH, Li DK, Traboulsee TL, Moore GR, MacKay AL (2007) Magnetic resonance imaging of myelin. Neurotherapeutics 4(3):460–484CrossRefPubMed
7.
Zurück zum Zitat Alonso-Ortiz E, Levesque IR, Pike GB (2015) MRI-based myelin water imaging: a technical review. Magn Reson Med 73(1):70–81CrossRefPubMed Alonso-Ortiz E, Levesque IR, Pike GB (2015) MRI-based myelin water imaging: a technical review. Magn Reson Med 73(1):70–81CrossRefPubMed
8.
Zurück zum Zitat Laule C, Vavasour IM, Moore GR, Oger J, Li DK, Paty DW, MacKay AL (2004) Water content and myelin water fraction in multiple sclerosis. A T2 relaxation study. J Neurol 251(3):284–293CrossRefPubMed Laule C, Vavasour IM, Moore GR, Oger J, Li DK, Paty DW, MacKay AL (2004) Water content and myelin water fraction in multiple sclerosis. A T2 relaxation study. J Neurol 251(3):284–293CrossRefPubMed
9.
Zurück zum Zitat Oh J, Han ET, Lee MC, Nelson SJ, Pelletier D (2007) Multislice brain myelin water fractions at 3T in multiple sclerosis. J Neuroimaging 17(2):156–163CrossRefPubMed Oh J, Han ET, Lee MC, Nelson SJ, Pelletier D (2007) Multislice brain myelin water fractions at 3T in multiple sclerosis. J Neuroimaging 17(2):156–163CrossRefPubMed
10.
Zurück zum Zitat Stewart WA, MacKay AL, Whittall KP, Moore GR, Paty DW (1993) Spin-spin relaxation in experimental allergic encephalomyelitis. Analysis of CPMG data using a non-linear least squares method and linear inverse theory. Magn Reson Med 29(6):767–775CrossRefPubMed Stewart WA, MacKay AL, Whittall KP, Moore GR, Paty DW (1993) Spin-spin relaxation in experimental allergic encephalomyelitis. Analysis of CPMG data using a non-linear least squares method and linear inverse theory. Magn Reson Med 29(6):767–775CrossRefPubMed
11.
Zurück zum Zitat Gareau PJ, Rutt BK, Bowen CV, Karlik SJ, Mitchell JR (1999) In vivo measurements of multi-component T2 relaxation behaviour in guinea pig brain. Magn Reson Imaging 17(9):1319–1325CrossRefPubMed Gareau PJ, Rutt BK, Bowen CV, Karlik SJ, Mitchell JR (1999) In vivo measurements of multi-component T2 relaxation behaviour in guinea pig brain. Magn Reson Imaging 17(9):1319–1325CrossRefPubMed
12.
Zurück zum Zitat Webb S, Munro CA, Midha R, Stanisz GJ (2003) Is multicomponent T2 a good measure of myelin content in peripheral nerve? Magn Reson Med 49(4):638–645CrossRefPubMed Webb S, Munro CA, Midha R, Stanisz GJ (2003) Is multicomponent T2 a good measure of myelin content in peripheral nerve? Magn Reson Med 49(4):638–645CrossRefPubMed
13.
Zurück zum Zitat Bjarnason TA, Vavasour IM, Chia CL, MacKay AL (2005) Characterization of the NMR behavior of white matter in bovine brain. Magn Reson Med 54(5):1072–1081CrossRefPubMed Bjarnason TA, Vavasour IM, Chia CL, MacKay AL (2005) Characterization of the NMR behavior of white matter in bovine brain. Magn Reson Med 54(5):1072–1081CrossRefPubMed
14.
Zurück zum Zitat Kozlowski P, Liu J, Yung AC, Tetzlaff W (2008) High-resolution myelin water measurements in rat spinal cord. Magn Reson Med 59(4):796–802CrossRefPubMed Kozlowski P, Liu J, Yung AC, Tetzlaff W (2008) High-resolution myelin water measurements in rat spinal cord. Magn Reson Med 59(4):796–802CrossRefPubMed
15.
Zurück zum Zitat McCreary CR, Bjarnason TA, Skihar V, Mitchell JR, Yong VW, Dunn JF (2009) Multiexponential T2 and magnetization transfer MRI of demyelination and remyelination in murine spinal cord. Neuroimage 45(4):1173–1182CrossRefPubMedPubMedCentral McCreary CR, Bjarnason TA, Skihar V, Mitchell JR, Yong VW, Dunn JF (2009) Multiexponential T2 and magnetization transfer MRI of demyelination and remyelination in murine spinal cord. Neuroimage 45(4):1173–1182CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Kozlowski P, Rosicka P, Liu J, Yung AC, Tetzlaff W (2014) In vivo longitudinal myelin water imaging in rat spinal cord following dorsal column transection injury. Magn Reson Imaging 32(3):250–258CrossRefPubMed Kozlowski P, Rosicka P, Liu J, Yung AC, Tetzlaff W (2014) In vivo longitudinal myelin water imaging in rat spinal cord following dorsal column transection injury. Magn Reson Imaging 32(3):250–258CrossRefPubMed
17.
Zurück zum Zitat Vasilescu V, Katona E, Simplaceanu V, Demco D (1978) Water compartments in the myelinated nerve. III. Pulsed NMR results. Experientia 34(11):1443–1444CrossRefPubMed Vasilescu V, Katona E, Simplaceanu V, Demco D (1978) Water compartments in the myelinated nerve. III. Pulsed NMR results. Experientia 34(11):1443–1444CrossRefPubMed
18.
Zurück zum Zitat Dula AN, Gochberg DF, Valentine HL, Valentine WM, Does MD (2010) Multiexponential T2, magnetization transfer, and quantitative histology in white matter tracts of rat spinal cord. Magn Reson Med 63(4):902–909CrossRefPubMedPubMedCentral Dula AN, Gochberg DF, Valentine HL, Valentine WM, Does MD (2010) Multiexponential T2, magnetization transfer, and quantitative histology in white matter tracts of rat spinal cord. Magn Reson Med 63(4):902–909CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Harkins KD, Valentine WM, Gochberg DF, Does MD (2013) In-vivo multi-exponential T2, magnetization transfer and quantitative histology in a rat model of intramyelinic edema. Neuroimage Clin 2:810–817CrossRefPubMedPubMedCentral Harkins KD, Valentine WM, Gochberg DF, Does MD (2013) In-vivo multi-exponential T2, magnetization transfer and quantitative histology in a rat model of intramyelinic edema. Neuroimage Clin 2:810–817CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Horch RA, Gore JC, Does MD (2011) Origins of the ultrashort-T2 1H NMR signals in myelinated nerve: a direct measure of myelin content? Magn Reson Med 66(1):24–31CrossRefPubMedPubMedCentral Horch RA, Gore JC, Does MD (2011) Origins of the ultrashort-T2 1H NMR signals in myelinated nerve: a direct measure of myelin content? Magn Reson Med 66(1):24–31CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat MacKay A, Whittall K, Adler J, Li D, Paty D, Graeb D (1994) In vivo visualization of myelin water in brain by magnetic resonance. Magn Reson Med 31(6):673–677CrossRefPubMed MacKay A, Whittall K, Adler J, Li D, Paty D, Graeb D (1994) In vivo visualization of myelin water in brain by magnetic resonance. Magn Reson Med 31(6):673–677CrossRefPubMed
22.
Zurück zum Zitat Whittall KP, MacKay AL, Graeb DA, Nugent RA, Li DK, Paty DW (1997) In vivo measurement of T2 distributions and water contents in normal human brain. Magn Reson Med 37(1):34–43CrossRefPubMed Whittall KP, MacKay AL, Graeb DA, Nugent RA, Li DK, Paty DW (1997) In vivo measurement of T2 distributions and water contents in normal human brain. Magn Reson Med 37(1):34–43CrossRefPubMed
23.
Zurück zum Zitat Menon RS, Rusinko MS, Allen PS (1992) Proton relaxation studies of water compartmentalization in a model neurological system. Magn Reson Med 28(2):264–274CrossRefPubMed Menon RS, Rusinko MS, Allen PS (1992) Proton relaxation studies of water compartmentalization in a model neurological system. Magn Reson Med 28(2):264–274CrossRefPubMed
24.
Zurück zum Zitat Nguyen TD, Wisnieff C, Cooper MA, Kumar D, Raj A, Spincemaille P, Wang Y, Vartanian T, Gauthier SA (2012) T2 prep three-dimensional spiral imaging with efficient whole brain coverage for myelin water quantification at 1.5 tesla. Magn Reson Med 67(3):614–621CrossRefPubMed Nguyen TD, Wisnieff C, Cooper MA, Kumar D, Raj A, Spincemaille P, Wang Y, Vartanian T, Gauthier SA (2012) T2 prep three-dimensional spiral imaging with efficient whole brain coverage for myelin water quantification at 1.5 tesla. Magn Reson Med 67(3):614–621CrossRefPubMed
25.
Zurück zum Zitat Wilhelm MJ, Ong HH, Wehrli SL, Li C, Tsai PH, Hackney DB, Wehrli FW (2012) Direct magnetic resonance detection of myelin and prospects for quantitative imaging of myelin density. Proc Natl Acad Sci USA 109(24):9605–9610CrossRefPubMedPubMedCentral Wilhelm MJ, Ong HH, Wehrli SL, Li C, Tsai PH, Hackney DB, Wehrli FW (2012) Direct magnetic resonance detection of myelin and prospects for quantitative imaging of myelin density. Proc Natl Acad Sci USA 109(24):9605–9610CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Prasloski T, Rauscher A, MacKay AL, Hodgson M, Vavasour IM, Laule C, Madler B (2012) Rapid whole cerebrum myelin water imaging using a 3D GRASE sequence. Neuroimage 63(1):533–539CrossRefPubMed Prasloski T, Rauscher A, MacKay AL, Hodgson M, Vavasour IM, Laule C, Madler B (2012) Rapid whole cerebrum myelin water imaging using a 3D GRASE sequence. Neuroimage 63(1):533–539CrossRefPubMed
27.
Zurück zum Zitat Deoni SC, Rutt BK, Arun T, Pierpaoli C, Jones DK (2008) Gleaning multicomponent T1 and T2 information from steady-state imaging data. Magn Reson Med 60(6):1372–1387CrossRefPubMed Deoni SC, Rutt BK, Arun T, Pierpaoli C, Jones DK (2008) Gleaning multicomponent T1 and T2 information from steady-state imaging data. Magn Reson Med 60(6):1372–1387CrossRefPubMed
28.
Zurück zum Zitat Lankford CL, Does MD (2013) On the inherent precision of mcDESPOT. Magn Reson Med 69(1):127–136CrossRefPubMed Lankford CL, Does MD (2013) On the inherent precision of mcDESPOT. Magn Reson Med 69(1):127–136CrossRefPubMed
29.
Zurück zum Zitat Zhang J, Kolind SH, Laule C, MacKay AL (2015) How does magnetization transfer influence mcDESPOT results? Magn Reson Med 74(5):1327–1335CrossRefPubMed Zhang J, Kolind SH, Laule C, MacKay AL (2015) How does magnetization transfer influence mcDESPOT results? Magn Reson Med 74(5):1327–1335CrossRefPubMed
30.
Zurück zum Zitat Bouhrara M, Spencer RG (2016) Improved determination of the myelin water fraction in human brain using magnetic resonance imaging through Bayesian analysis of mcDESPOT. Neuroimage 127:456–471CrossRefPubMed Bouhrara M, Spencer RG (2016) Improved determination of the myelin water fraction in human brain using magnetic resonance imaging through Bayesian analysis of mcDESPOT. Neuroimage 127:456–471CrossRefPubMed
31.
Zurück zum Zitat Bouhrara M, Spencer RG (2015) Incorporation of nonzero echo times in the SPGR and bSSFP signal models used in mcDESPOT. Magn Reson Med 74(5):1227–1235CrossRefPubMedPubMedCentral Bouhrara M, Spencer RG (2015) Incorporation of nonzero echo times in the SPGR and bSSFP signal models used in mcDESPOT. Magn Reson Med 74(5):1227–1235CrossRefPubMedPubMedCentral
32.
Zurück zum Zitat Deoni SC, Matthews L, Kolind SH (2013) One component? Two components? Three? The effect of including a nonexchanging “free” water component in multicomponent driven equilibrium single pulse observation of T1 and T2. Magn Reson Med 70(1):147–154CrossRefPubMed Deoni SC, Matthews L, Kolind SH (2013) One component? Two components? Three? The effect of including a nonexchanging “free” water component in multicomponent driven equilibrium single pulse observation of T1 and T2. Magn Reson Med 70(1):147–154CrossRefPubMed
33.
Zurück zum Zitat Kirsch S, Schad LR (2011) Single-slice mapping of ultrashort T(2). J Magn Reson 210(1):133–136CrossRefPubMed Kirsch S, Schad LR (2011) Single-slice mapping of ultrashort T(2). J Magn Reson 210(1):133–136CrossRefPubMed
34.
Zurück zum Zitat Whittall KP, Mackay AL (1989) Quantitative interpretation of nmr relaxation data. J Magn Reson 84(1):134–152 Whittall KP, Mackay AL (1989) Quantitative interpretation of nmr relaxation data. J Magn Reson 84(1):134–152
35.
Zurück zum Zitat Does MD, Gore JC (2002) Compartmental study of T(1) and T(2) in rat brain and trigeminal nerve in vivo. Magn Reson Med 47(2):274–283CrossRefPubMed Does MD, Gore JC (2002) Compartmental study of T(1) and T(2) in rat brain and trigeminal nerve in vivo. Magn Reson Med 47(2):274–283CrossRefPubMed
36.
Zurück zum Zitat Oh S-H, Bilello M, Schindler M, Markowitz CE, Detre JA, Lee J (2013) Direct visualization of short transverse relaxation time component (ViSTa). Neuroimage 83:485–492CrossRefPubMed Oh S-H, Bilello M, Schindler M, Markowitz CE, Detre JA, Lee J (2013) Direct visualization of short transverse relaxation time component (ViSTa). Neuroimage 83:485–492CrossRefPubMed
37.
Zurück zum Zitat Du J, Sheth V, He Q, Carl M, Chen J, Corey-Bloom J, Bydder GM (2014) Measurement of T1 of the ultrashort T2* components in white matter of the brain at 3T. PLoS One 9(8):e103296CrossRefPubMedPubMedCentral Du J, Sheth V, He Q, Carl M, Chen J, Corey-Bloom J, Bydder GM (2014) Measurement of T1 of the ultrashort T2* components in white matter of the brain at 3T. PLoS One 9(8):e103296CrossRefPubMedPubMedCentral
38.
Zurück zum Zitat Du J, Ma G, Li S, Carl M, Szeverenyi NM, VandenBerg S, Corey-Bloom J, Bydder GM (2014) Ultrashort echo time (UTE) magnetic resonance imaging of the short T2 components in white matter of the brain using a clinical 3T scanner. Neuroimage 87:32–41CrossRefPubMed Du J, Ma G, Li S, Carl M, Szeverenyi NM, VandenBerg S, Corey-Bloom J, Bydder GM (2014) Ultrashort echo time (UTE) magnetic resonance imaging of the short T2 components in white matter of the brain using a clinical 3T scanner. Neuroimage 87:32–41CrossRefPubMed
39.
40.
Zurück zum Zitat Shepherd TM, Thelwall PE, Stanisz GJ, Blackband SJ (2009) Aldehyde fixative solutions alter the water relaxation and diffusion properties of nervous tissue. Magn Reson Med 62(1):26–34CrossRefPubMedPubMedCentral Shepherd TM, Thelwall PE, Stanisz GJ, Blackband SJ (2009) Aldehyde fixative solutions alter the water relaxation and diffusion properties of nervous tissue. Magn Reson Med 62(1):26–34CrossRefPubMedPubMedCentral
Metadaten
Titel
Assessment of the myelin water fraction in rodent spinal cord using T2-prepared ultrashort echo time MRI
verfasst von
Tim Klasen
Cornelius Faber
Publikationsdatum
09.07.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 6/2016
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-016-0579-7

Weitere Artikel der Ausgabe 6/2016

Magnetic Resonance Materials in Physics, Biology and Medicine 6/2016 Zur Ausgabe

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.