Clinical evaluation of left ventricular function and morphology using an accelerated k-t sensitivity encoding method in cardiovascular magnetic resonance

verfasst von: Antonildes Nascimento Assuncao-Jr, Roberto Nery Dantas-Jr, Renata Margarida do Val, Priscilla Gianotto, Angela dos Santos Marin, Mark Golden, Marco Antonio Gutierrez, Jose Rodrigues Parga, Cesar Higa Nomura

Erschienen in: Insights into Imaging | Ausgabe 1/2019

Abstract

Objectives

To provide clinical validation of a recent 2D SENSE-based accelerated cardiovascular magnetic resonance (CMR) sequence (accelerated k-t SENSE), investigating whether this technique accurately quantifies left ventricle (LV) volumes, function, and mass as compared to 2D cine steady-state free precession (2D-SSFP).

Methods

Healthy volunteers (n = 16) and consecutive heart failure patients (n = 26) were scanned using a 1.5 T MRI system. Two LV short axis (SA) stacks were acquired: (1) accelerated k-t SENSE (5–6 breath-holds; temporal/spatial resolution: 37 ms/1.82 × 1.87 mm; acceleration factor = 4) and (2) standard 2D-SSFP (10–12 breath-holds; temporal/spatial resolution: 49 ms/1.67 × 1.87 mm, parallel imaging). Ascending aorta phase-contrast was performed on all volunteers as a reference to compare LV stroke volumes (LVSV) and validate the sequences. An image quality score for SA images was used, with lower scores indicating better quality (from 0 to 18).

Results

There was a high agreement between accelerated k-t SENSE and 2D-SSFP for LV measurements: bias (limits of agreement) of 2.4% (− 5.4% to 10.1%), 6.9 mL/m2 (− 4.7 to 18.6 mL/m2), − 1.5 (− 8.3 to 5.2 mL/m2), and − 0.2 g/m2 (− 11.9 to 12.3 g/m2) for LV ejection fraction, end-diastolic volume index, end-systolic volume index, and mass index, respectively. LVSV by accelerated k-t SENSE presented good agreement with aortic flow. Interobserver and intraobserver variabilities for all LV parameters were also high.

Conclusion

The accelerated k-t SENSE CMR sequence is clinically feasible and accurately quantifies LV volumes, function, and mass, with short acquisition time and good image quality.
Hinweise
Antonildes N. Assuncao-Jr and Roberto N. Dantas Jr contributed equally to this work.
Antonildes N. Assuncao-Jr and Roberto N. Dantas Jr are joint first authors.
Abkürzungen
2D-SSFP
2D cine steady-state free precession
BLAST
Broad-use linear acquisition speed-up technique
BSA
Body surface area
CMR
Cardiovascular magnetic resonance
LV
Left ventricle
LVEDVI
Left ventricle end-diastolic volume index
LVEF
Left ventricle ejection fraction
LVESVI
Left ventricle end-systolic volume index
LVMI
Left ventricle mass index
LVSV
Left ventricle stroke volume
NYHA
New York Heart Association
SA
Short axis
SENSE
Sensitivity encoding

Key points

  • CMR is considered the non-invasive gold standard method for ventricular functional measurements, accomplished mainly through 2D-SSFP cine images
  • Accelerated imaging may favor patients with impaired breath-hold or frequent arrhythmia
  • Recently developed accelerated CMR cine sequences speed up image acquisition either in temporal, spatial (k-space), or even both domains simultaneously (accelerated k-t SENSE)
  • The accelerated k-t SENSE CMR sequence is clinically feasible and accurate for ventricular functional assessment, with short acquisition time and good image quality

Introduction

Cardiovascular magnetic resonance (CMR) imaging is the non-invasive gold standard modality for quantification of left ventricle (LV) function, volumes, and mass [1]. For these purposes, a 2D cine steady-state free precession (2D-SSFP)-based sequence has been widely used in clinical practice, with high reproducibility and accuracy [13]. However, 2D-SSFP requires multiple breath-holds, resulting in prolonged exam duration, and representing a challenge for patients with heart failure symptoms. In addition to that, slice misregistration can be a result of inconsistent breath holding, hampering volumetric estimation [46]. Thus, the chosen imaging sequence for determined populations is crucial for delivering fast and accurate diagnostic information. In response to these issues, there have been considerable attempts to accelerate cine sequences, either in temporal, spatial (k-space), or even both domains simultaneously, without considerably reducing spatial or temporal resolutions [79].
Accelerated CMR sequences, such as k-t BLAST (broad-use linear acquisition speed-up technique; single receiver coil) and k-t SENSE (sensitivity encoding; multiple receiver coils) consider spatiotemporal correlations throughout the image and rely on undersampling and signal overlap (aliasing) recovery through computational algorithms [7, 8]. Signal correlations and coil sensitivity estimates are obtained in an initial low-resolution acquisition (training stage), and this information is used afterwards for image reconstruction, allowing acceleration by sparsely sampling k-space over time in the main acquisition stage [9].
Recently, a new accelerated 2D k-t SENSE-based cine sequence (accelerated k-t SENSE) was developed to accelerate image acquisition, without performing the training stage [10]. Signal correlations and coil sensitivity estimates are extracted from the acquisition stage data itself during post processing, allowing faster image acquisition [10]. Our goal in the present study is to clinically investigate whether this newly developed accelerated k-t SENSE sequence accurately quantifies LV volumes, function, and mass as compared to the 2D-SSFP cine with acceptable image quality.

Materials and methods

Subjects

Healthy volunteers (n = 16) and consecutive heart failure patients (n = 26) were prospectively enrolled in this study. Exclusion criteria were age under 18 years old, contraindications to CMR (e.g., CMR-incompatible devices, metallic bodies in the eye, intracranial metal clips), irregular heart rhythms, severely impaired breath-hold capacity, claustrophobia, and pregnancy. All patients were clinically referred to CMR assessment of LV volumes, function, and mass. Indications for CMR in patients included non-ischemic (n = 20) and ischemic cardiomyopathies (n = 6). This study was approved by the Institutional Ethics Committee and all subjects provided written informed consent.

Accelerated k-t SENSE technique

The rationale and technical details of the acceleration sequence have been described elsewhere [10]. Briefly, accelerated k-t SENSE consists of a k-t SENSE-based SSFP cine sequence that uses no training stage (a previous additional low resolution scan used only for obtaining reference signals) for defining spatiotemporal signal correlations and k-t coil sensitivity maps, thus allowing several fold acceleration. These signal correlations and sensitivity maps are estimated from the acquired image itself during post-processing. As a result, image acquisition is accelerated through undersampling of k-space over time in the main acquisition stage, which provides all correlations needed for final image reconstruction. Therefore, acceleration occurs in both spatial and temporal directions, with partially sampled data, allowing optimization of scan timing.

MRI protocol

All subjects were prospectively scanned using a 1.5T MRI system (Canon Vantage Titan, Canon Medical System Corporation, Japan). In addition to the standard protocol regarding the clinical indication, two short axis (SA) stacks fully covering both ventricles were acquired and their duration was measured: prospective ECG-triggered accelerated k-t SENSE cine (two slices/breath-hold), and a standard 2D-SSFP cine with parallel imaging and retrospective ECG triggering (one slice/breath-hold).
In order to validate the left ventricle stroke volume (LVSV) obtained by accelerated k-t SENSE, a retrospective phase-contrast flow measurement in the ascending aorta (immediately above sinotubular junction) was performed as a reference in all healthy subjects (VENC 200 cm/s, matrix 256 × 88, temporal/spatial resolution 46 ms/1.48 × 4.31 mm). All cine sequences included standard shimming (64 × 64 mm matrix, voxel size 6.25 × 6.25 mm) and use of view sharing. Imaging parameters are described in Table 1.
Table 1
Imaging parameters
 
2D-SSFP
Accelerated k-t SENSE
ECG triggering
Retrospective
Prospective
TE/TR (ms)
1.7/3.4
1.7/3.4
FOV (mm)
320 × 360
320 × 360
Image matrix
192 × 192
176 × 192
Spatial resolution (mm)
1.67 × 1.87
1.82 × 1.87
Temporal resolution (ms)
49
37
Slice thickness/spacing (mm)
10/0
10/0
Flip angle (°)
60
60
Bandwidth (Hz/pixel)
977
1302
Cardiac phases (n)
25
32
View sharing (segments)
14
11
Breath-holds (n)
10–12
5–6
Slices per breath-hold
1
2
Acceleration factor
Parallel imaginga
4x
Acquisition duration (s)
161 ± 25
67 ± 14
TE echo time, TR repetition time, FOV field of view, 2D-SSFP 2D steady-state free precession
aParallel imaging intrinsic acceleration

Assessment of LV function and morphology

First, all images were de-identified and digitally stored. Then, for quantitative measurements, two experienced cardiovascular imaging specialists (> 5 years of training) analyzed the SA stacks—twice by one reader (time interval between readings of 1 week), using a commercially available software Medis Suite 3.0 (Medis, Leiden, The Netherlands).
All endocardial and epicardial contours were manually drawn as recommended by the Society for Cardiovascular Magnetic Resonance Guidelines [11], covering LV from basal to apical slices, including both papillary muscles and trabeculations as part of the LV cavity. For basal slices, contour was carefully drawn to include the LV outflow tract to the level of the aortic valve cusps, and left atrium was recognized when less than 50% of the blood volume was surrounded by ventricular myocardium [11].
LV volumes (end-diastolic—LVEDV and end-systolic—LVESV), ejection fraction (LVEF), and mass, and their respective indexes corrected for the body surface area, were calculated using the Simpson method. End-diastole phases were chosen as those with the maximum volume in a mid-ventricular slice to provide more consistent estimations of the LV volumes [12]. Quantitative wall motion per-segment analysis (American Heart Association—16-segment model) was also performed, using a centerline method applied to the endocardial and epicardial contours at end-diastole and end-systole [13].
For aortic flow quantification on the healthy volunteers group, the borders of the ascending aorta were traced to include only its cavity. For optimal results, image plane was properly centered and aligned. Aliasing was double checked and, if occurred, velocity encoding sensitivity was set accordingly. The final aortic flow included the estimated coronary arteries flow (mL/beat), calculated as 0.8 × LV mass (g) / heart rate (beats/min) [14].

Image quality

Quality assessment of cine CMR SA images was performed based on 11 qualitative criteria [15]. This assessment yields a score for LV coverage, presence of artifacts (wrap around, ghosts, metallic and shimming artifacts, image blurring/mis-triggering), signal loss, correct orientation of stack, and adequate gap between slices, assigning individual scores that range from 0 to 18 (the lower the score, the better the image quality) (Table 2).

Statistical analysis

Continuous variables are shown as mean ± standard deviation, and were compared using paired t test or paired Wilcoxon test, as appropriate. Categorical variables were expressed as frequencies (percentages). Agreement of LV parameters on the accelerated k-t SENSE and 2D-SSFP as well as agreement between volumetrically determined LVSV and aortic flow were assessed by Bland-Altman analysis. The same analysis was performed to investigate interobserver and intraobserver agreement. Repeatability coefficients (RC), two times the SDs of the differences between the two measurements, were also calculated for each LV parameter. All statistical analyses were performed using the software R 3.4.3 (The R Foundation for Statistical Computing, Vienna, Austria) and a p value < 0.05 was considered statistically significant.
Table 2
Image quality of accelerated k-t SENSE and 2D-SSFP method
 
2D-SSFP
k-t
Score 0
Score 1
Score 2
Score 3
Score 0
Score 1
Score 2
Score 3
Image blurring/mis triggering
35
5
2
0
30
5
5
2
Shimming
40
1
1
0
37
0
2
3
Ghosts
42
0
0
0
36
6
0
0
Correct LV long axes
        
LV coverage
42
0
0
0
42
0
0
0
Metallic artifact
42
0
0
0
42
0
0
0
Orientation of stack
42
0
0
0
42
0
0
0
Signal loss
42
0
0
0
42
0
0
0
Slice thickness/Gap
42
0
0
0
42
0
0
0
Wrap around
42
0
0
0
42
0
0
0
Total scorea
0.33 ± 0.64
0.98 ± 1.25
2D-SSFP 2D steady-state free precession, LV left ventricle
amean ± SD

Results

Study population, acquisition duration, and image quality

Characteristics of the study population are displayed in Table 3. Most patients were referred to investigate non-ischemic cardiomyopathies and presented mean LVEF of 52 ± 21% (p = 0.05), with larger LV volumes than volunteers (p values< 0.05). One patient presented infrequent premature ventricular contractions and all individuals had sinus rhythm. The accelerated k-t SENSE reduced by nearly 60% acquisition duration of short-axis images compared to 2D-SSFP (67 ± 14 s versus 161 ± 25 s, p < 0.001), with less breath-holds (Table 1).
Table 3
Study population
 
Patients (n = 26)
Volunteers (n = 16)
p value
Demographics
 Age, years
53 ± 13
43 ± 14
< 0.001
 Male, n (%)
17 (65)
10 (63)
0.91
 BMI, kg/m2
29 ± 6
26 ± 3
0.004
 BSA, m2
1.91 ± 0.27
1.83 ± 0.24
0.33
 HR, bpm
68 ± 17
63 ± 17
0.73
 PVC, n (%)
1 (4%)
Cardiovascular risk factors
 Hypertension, n (%)
17 (65)
1 (6)
 Diabetes, n (%)
8 (31)
 Hypercholesterolemia, n (%)
11 (42)
 Smoking, n (%)
10 (39)
Diagnosis
 HCM, n (%)
7 (27)
 Ischemic heart disease, n (%)
6 (24)
 Chagas heart disease, n (%)
3 (11)
 DCM, n (%)
3 (11)
 Other cardiomyopathies, n (%)
7 (27)
CMR findingsa
 LVEDVI, mL/m2
92 ± 50
66 ± 10
0.01
 LVESVI, mL/m2
52 ± 53
26 ± 10
0.04
 LVMI, g/m2
80 ± 28
54 ± 10
< 0.001
 LVSV, mL
76 ± 25
75 ± 15
0.90
 LVEF, %
52 ± 21
62 ± 5
0.05
BMI body index mass, BSA body surface area, HR heart rate, PVC premature ventricular contraction, HCM hypertrophic cardiomyopathy, DCM dilated cardiomyopathy, CMR cardiovascular magnetic resonance, LVEDVI left ventricle end-diastolic volume index, LVESVI left ventricle end-systolic volume index, LVMI left ventricle mass index, LVSV left ventricle stroke volume, LVEF left ventricle ejection fraction
Plus-minus values are means ± SD
aLV measurements obtained from 2D-SSFP cine
Mean quality score resulted in high image quality for both methods, yet 2D-SSFP performed slightly better (0.33 ± 0.64 versus 0.98 ± 1.25, p < 0.001) (Table 2). Five accelerated k-t SENSE acquisitions (12%) had score ≥ 3 due to either image blurring/mis-triggering or shimming artifacts (Table 2), but they did not preclude adequate recognition of endocardial/epicardial borders (Fig. 1).

Agreement between accelerated k-t SENSE and 2D-SSFP cine

There was a strong correlation between accelerated k-t SENSE and 2D-SSFP for quantification of LV measurements, with high agreement. Mean difference or bias (limits of agreement, LOA) were 2.4% (− 5.4% to 10.1%), 6.9 mL/m2 (− 4.7 to 18.6 mL/m2), − 1.5 (− 8.3 to 5.2 mL/m2), and − 0.2 g/m2 (LOA − 11.9 to 12.3 g/m2) for LVEF, left ventricle end-diastolic volume index (LVEDVI), left ventricle end-systolic volume index (LVESVI), and left ventricle mass index (LVMI), respectively (Fig. 2). Regional myocardial wall motion analysis also showed good correlation (0.63 ≤ correlation coefficient ≤ 0.87, all p values < 0.001) and agreement (− 0.74 ≤ bias ≤ 0.06) between accelerated k-t SENSE and 2D-SSFP (Fig. 3). Interobserver and intraobserver variabilities of accelerated k-t SENSE ranged from 0.1 mL/m2 to − 5.2 g/m2 and from 0.6 mL/m2 to − 4.5 g/m2, respectively for LVEDVI and LVMI. RC ranged from 6% (LVEF) to 14 g/m2 (LVMI) (Table 4).
Table 4
Interobserver and intraobserver reproducibility of accelerated k-t SENSE
 
Interobserver
Intraobserver
Agreement bias (LOA)
Correlation r, p value
RC
Agreement bias (LOA)
Correlation r, p value
RC
LVEF, %
− 2.6 (− 8.2 to 3.6)
0.98, p < 0.001
6
1.7 (− 6.5 to 9.8)
0.96, p < 0.001
8
LVEDVI, mL/m2
0.1 (− 12.2 to 12.2)
0.98, p < 0.001
12
0.6 (− 10.8 to 12.1)
0.95, p < 0.001
12
LVESVI, mL/m2
2.2 (− 7.7 to 9.9)
0.98, p < 0.001
10
0.7 (− 9.9 to 8.5)
0.98, p < 0.001
9
LVMI, g/m2
− 5.2 (− 19.3 to 8.9)
0.95, p < 0.001
14
− 4.5 (− 16.5 to 7.5)
0.98, p < 0.001
12
LOA limits of agreement, RC repeatability coefficient, LVEF left ventricle ejection fraction, LVEDVI left ventricle end-diastolic volume index, LVESVI left ventricle end-systolic volume index, LVMI left ventricle mass index

Validation of accelerated k-t SENSE against aortic flow

The quantification of LVSV by the accelerated k-t SENSE showed a strong correlation and good agreement with aortic flow, slightly underestimating LVSV by 0.58 mL (Fig. 4).

Discussion

In this study, the accelerated k-t SENSE cine (no training stage approach) was compared to 2D-SSFP for the quantification of LV measurements, with high agreement. The accelerated sequence required half the breath-holds (5 to 6 vs 10 to 12) to cover the entire LV (reducing acquisition time by 60%), and yielded an excellent image quality in 88% of subjects, with accurate assessment of LV function, volumes, and mass as compared to the 2D-SSFP cine.
Agreement between accelerated and standard cines was high for LVEF, with only a slightly wider variability (− 4.1% to 4.3%) [3, 16]. Additionally, accelerated k-t SENSE was quite robust regarding reproducibility for LVEF (RC 6% and 8% for inter- and intraobserver, respectively), closely matching the interobserver reproducibility of 2D-SSFP cine (6% and 12%) [17, 18]. Accelerated k-t SENSE images also provided comparable results in the quantitative regional myocardial wall motion analysis compared to 2D-SSFP.
Likewise, a relevant agreement between the techniques was achieved for LV volumes and mass. Accelerated k-t SENSE promoted a small myocardial mass overestimation and LVEDVI underestimation in a similar fashion of what was found when studying agreement of a 3D k-t BLAST technique with 2D-SSFP cines (despite differences between BLAST and SENSE based sequences) [8]. Other studies regarding spatiotemporal acceleration techniques found similar results, either with compressed sensing or SENSE [19, 20]. Accelerated k-t SENSE had also high accuracy for calculating LVSV using as reference the aortic flow, with a very small overestimation of 0.58 mL. This result is in line with the reported overestimation of LVSV by accelerated sequences in aortic valves [20, 21].
Although the vast majority of patients had excellent image quality, accelerated k-t SENSE yielded, on average, a higher image quality score when compared to the 2D SSFP cine. Indeed, accelerated sequences may favor blurring or residual aliasing artifacts, degrading image quality [22, 23]. However, even in patients with artifacts, images were reasonable for adequate recognition of myocardial borders.
It is remarkable that, despite not reducing acquisition time comparable to neither compressed sensing [20] nor whole heart 3D cine (about 80%) [23], accelerated k-t SENSE reached 60% faster cine acquisition. However, these two other techniques present high computational burden for image reconstruction (precluding immediate assessment of quality or planning next steps) and reporting time (to up 30 min), making accelerated k-t SENSE advantageous.
Some further limitations must be outlined. Firstly, although we included heart failure patients, they had no limiting symptoms and/or significant arrhythmias given the study design. Therefore, additional studies are planned to confirm whether this expressive acquisition time reduction also benefits subjects with more limited breath-hold capabilities. Secondly, despite the possibility of using a wider range of acceleration factors by the accelerated k-t SENSE, faster acquisition (> 4×) accuracy still needs evaluation regarding reliability, as higher acceleration factors with k-t SENSE produced poorer image quality and agreement in previous experience [9]. Finally, retrospective accelerated k-t SENSE was not fully developed by the completion of this work, and may represent further achievements in function assessment.
In conclusion, the accelerated k-t SENSE is feasible and can accurately quantify LV volumes, function, and mass, with good image quality and considerable shortening of acquisition time.

Funding

This work received research support from Canon Medical Systems do Brasil.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
2.
Zurück zum Zitat Lorenz CH, Walker ES, Morgan VL, Klein SS, Graham TP Jr (1999) Normal human right and left ventricular mass, systolic function, and gender differences by cine magnetic resonance imaging. J Cardiovasc Magn Reson 1:7–21CrossRef Lorenz CH, Walker ES, Morgan VL, Klein SS, Graham TP Jr (1999) Normal human right and left ventricular mass, systolic function, and gender differences by cine magnetic resonance imaging. J Cardiovasc Magn Reson 1:7–21CrossRef
3.
Zurück zum Zitat Grothues F, Smith GC, Moon JC et al (2002) Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol 90:29–34CrossRef Grothues F, Smith GC, Moon JC et al (2002) Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol 90:29–34CrossRef
13.
Zurück zum Zitat Holman ER, Buller VG, de Roos A et al (1997) Detection and quantification of dysfunctional myocardium by magnetic resonance imaging. A new three-dimensional method for quantitative wall-thickening analysis. Circulation 95:924–931CrossRef Holman ER, Buller VG, de Roos A et al (1997) Detection and quantification of dysfunctional myocardium by magnetic resonance imaging. A new three-dimensional method for quantitative wall-thickening analysis. Circulation 95:924–931CrossRef
16.
Zurück zum Zitat Danilouchkine MG, Westenberg JJ, de Roos A, Reiber JH, Lelieveldt BP (2005) Operator induced variability in cardiovascular MR: left ventricular measurements and their reproducibility. J Cardiovasc Magn Reson 7:447–457CrossRef Danilouchkine MG, Westenberg JJ, de Roos A, Reiber JH, Lelieveldt BP (2005) Operator induced variability in cardiovascular MR: left ventricular measurements and their reproducibility. J Cardiovasc Magn Reson 7:447–457CrossRef
Metadaten
Titel
Clinical evaluation of left ventricular function and morphology using an accelerated k-t sensitivity encoding method in cardiovascular magnetic resonance
verfasst von
Antonildes Nascimento Assuncao-Jr
Roberto Nery Dantas-Jr
Renata Margarida do Val
Priscilla Gianotto
Angela dos Santos Marin
Mark Golden
Marco Antonio Gutierrez
Jose Rodrigues Parga
Cesar Higa Nomura
Publikationsdatum
01.12.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Insights into Imaging / Ausgabe 1/2019
Elektronische ISSN: 1869-4101
DOI
https://doi.org/10.1186/s13244-019-0750-6

Weitere Artikel der Ausgabe 1/2019

Insights into Imaging 1/2019 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.