Combined quantitative analysis of attenuation corrected and non-corrected myocadial perfusion SPECT: Method development and clinical validation

verfasst von: Yuan Xu, PhD, Mathews Fish, MD, James Gerlach, CNMT, Mark Lemley, BSc, Daniel S. Berman, MD, Guido Germano, PhD, Piotr J. Slomka, PhD

Erschienen in: Journal of Nuclear Cardiology | Ausgabe 4/2010

Einloggen, um Zugang zu erhalten

Abstract

Background

Attenuation corrected myocardial perfusion SPECT (AC-MPS) has been demonstrated to improve the specificity of detecting coronary artery disease (CAD) by visual analysis which utilizes both non-corrected (NC) and AC data. However, the combined automated quantification of NC and AC-MPS has not been previously described. We aimed to develop a combined quantitative analysis from AC and NC data to improve the accuracy of automated detection of CAD from AC-MPS.

Methods

Stress total perfusion deficit (TPD) values were generated by standard analysis for NC (NC-TPD), AC (AC-TPD) and by combined NC-AC analysis (NA-TPD), in which the hypoperfusion severity in each polar map location was defined as the average of AC and NC severity computed by comparison with separate AC and NC normal limits. Ischemic TPD was also calculated as the difference between stress TPD and rest TPD for each measure. Stress/rest Tc-99m sestamibi MPS studies in 650 patients with correlating coronary angiography and in 345 patients with a low-likelihood (LLk) of CAD were used to assess diagnostic performance of combined NC-AC analysis.

Results

NA-TPD had a higher receiver-operator-characteristic area under the curve (ROC-AUC) (0.87) than NC-TPD (0.85; P < .01) or AC-TPD (0.85; P < .01) for detection of stenosis ≥70% in angiographic group. It also had higher specificity (75%) vs NC-TPD (65%; P < .0001), or AC-TPD (70%; P = .016). In LLk group, the normalcy rate of NA-TPD (95%) was higher than for NC-TPD (90%; P < .01) and similar to AC-TPD (94%; P = NS). NA-TPD had higher ROC-AUC than that for 17-segment expert visual scoring of stress scans in angiographic group (0.84; P = .01), comparable accuracy (81%) and similar normalcy rates (95% vs 97%; P = NS). Ischemic TPD by combined NC-AC analysis had higher ROC-AUC than that for any ischemic measure. Similar to stress NA-TPD, it also obtained the similar performance results as compared with ischemic TPD based on NC or AC and higher sensitivity (89% vs 85%; P = .0295) as compared with ischemic visual score in angiographic group.

Conclusion

Combined NC-AC MPS quantification using either stress or ischemic TPD shows significant improvements for ROC-AUC and specificity of MPS in the detection of CAD compared with standard NC-MPS or AC-MPS and comparable performance to expert visual scoring. This technique may lead to an enhancement in a fully automated quantification for the perfusion analysis by AC-MPS.
Literatur
1.
Zurück zum Zitat Masood Y, Liu YH, Depuey G, et al. Clinical validation of SPECT attenuation correction using x-ray computed tomography-derived attenuation maps: Multicenter clinical trial with angiographic correlation. J Nucl Cardiol 2005;12(6):676-86.CrossRefPubMed Masood Y, Liu YH, Depuey G, et al. Clinical validation of SPECT attenuation correction using x-ray computed tomography-derived attenuation maps: Multicenter clinical trial with angiographic correlation. J Nucl Cardiol 2005;12(6):676-86.CrossRefPubMed
2.
Zurück zum Zitat Links JM, Becker LC, Rigo P, et al. Combined corrections for attenuation, depth-dependent blur, and motion in cardiac SPECT: A multicenter trial. J Nucl Cardiol 2000;7(5):414-25.CrossRefPubMed Links JM, Becker LC, Rigo P, et al. Combined corrections for attenuation, depth-dependent blur, and motion in cardiac SPECT: A multicenter trial. J Nucl Cardiol 2000;7(5):414-25.CrossRefPubMed
3.
Zurück zum Zitat Heller GV, Links J, Bateman TM, et al. American Society of Nuclear Cardiology and Society of Nuclear Medicine joint position statement: Attenuation correction of myocardial perfusion SPECT scintigraphy. J Nucl Cardiol 2004;11(2):229-30.CrossRefPubMed Heller GV, Links J, Bateman TM, et al. American Society of Nuclear Cardiology and Society of Nuclear Medicine joint position statement: Attenuation correction of myocardial perfusion SPECT scintigraphy. J Nucl Cardiol 2004;11(2):229-30.CrossRefPubMed
4.
Zurück zum Zitat Ficaro EP, Fessler JA, Shreve PD, Kritzman JN, Rose PA, Corbett JR. Simultaneous transmission/emission myocardial perfusion tomography. Diagnostic accuracy of attenuation-corrected 99mTc-sestamibi single-photon emission computed tomography. Circulation 1996;93(3):463-73.PubMed Ficaro EP, Fessler JA, Shreve PD, Kritzman JN, Rose PA, Corbett JR. Simultaneous transmission/emission myocardial perfusion tomography. Diagnostic accuracy of attenuation-corrected 99mTc-sestamibi single-photon emission computed tomography. Circulation 1996;93(3):463-73.PubMed
5.
Zurück zum Zitat Kluge R, Sattler B, Seese A, Knapp WH. Attenuation correction by simultaneous emission-transmission myocardial single-photon emission tomography using a technetium-99m-labelled radiotracer: Impact on diagnostic accuracy. Eur J Nucl Med 1997;24(9):1107-14.PubMed Kluge R, Sattler B, Seese A, Knapp WH. Attenuation correction by simultaneous emission-transmission myocardial single-photon emission tomography using a technetium-99m-labelled radiotracer: Impact on diagnostic accuracy. Eur J Nucl Med 1997;24(9):1107-14.PubMed
6.
Zurück zum Zitat Gallowitsch HJ, Sykora J, Mikosch P, et al. Attenuation-corrected thallium-201 single-photon emission tomography using a gadolinium-153 moving line source: Clinical value and the impact of attenuation correction on the extent and severity of perfusion abnormalities. Eur J Nucl Med 1998;25(3):220-8.CrossRefPubMed Gallowitsch HJ, Sykora J, Mikosch P, et al. Attenuation-corrected thallium-201 single-photon emission tomography using a gadolinium-153 moving line source: Clinical value and the impact of attenuation correction on the extent and severity of perfusion abnormalities. Eur J Nucl Med 1998;25(3):220-8.CrossRefPubMed
7.
Zurück zum Zitat Hendel RC, Berman DS, Cullom SJ, et al. Multicenter clinical trial to evaluate the efficacy of correction for photon attenuation and scatter in SPECT myocardial perfusion imaging. Circulation 1999;99(21):2742-9.PubMed Hendel RC, Berman DS, Cullom SJ, et al. Multicenter clinical trial to evaluate the efficacy of correction for photon attenuation and scatter in SPECT myocardial perfusion imaging. Circulation 1999;99(21):2742-9.PubMed
8.
Zurück zum Zitat Links JM, DePuey EG, Taillefer R, Becker LC. Attenuation, correction, gating synergistically improve the diagnostic accuracy of myocardial perfusion SPECT. J Nucl Cardiol 2002;9(2):183-7.CrossRefPubMed Links JM, DePuey EG, Taillefer R, Becker LC. Attenuation, correction, gating synergistically improve the diagnostic accuracy of myocardial perfusion SPECT. J Nucl Cardiol 2002;9(2):183-7.CrossRefPubMed
9.
Zurück zum Zitat Shotwell M, Singh BM, Fortman C, Bauman BD, Lukes J, Gerson MC. Improved coronary disease detection with quantitative attenuation-corrected Tl-201 images. J Nucl Cardiol 2002;9(1):52-62.CrossRefPubMed Shotwell M, Singh BM, Fortman C, Bauman BD, Lukes J, Gerson MC. Improved coronary disease detection with quantitative attenuation-corrected Tl-201 images. J Nucl Cardiol 2002;9(1):52-62.CrossRefPubMed
10.
Zurück zum Zitat Grossman GB, Garcia EV, Bateman TM, et al. Quantitative Tc-99m sestamibi attenuation-corrected SPECT: Development and multicenter trial validation of myocardial perfusion stress gender-independent normal database in an obese population. J Nucl Cardiol 2004;11(3):263-72.CrossRefPubMed Grossman GB, Garcia EV, Bateman TM, et al. Quantitative Tc-99m sestamibi attenuation-corrected SPECT: Development and multicenter trial validation of myocardial perfusion stress gender-independent normal database in an obese population. J Nucl Cardiol 2004;11(3):263-72.CrossRefPubMed
11.
Zurück zum Zitat Thompson RC, Heller GV, Johnson LL, et al. Value of attenuation correction on ECG-gated SPECT myocardial perfusion imaging related to body mass index. J Nucl Cardiol 2005;12(2):195-202.CrossRefPubMed Thompson RC, Heller GV, Johnson LL, et al. Value of attenuation correction on ECG-gated SPECT myocardial perfusion imaging related to body mass index. J Nucl Cardiol 2005;12(2):195-202.CrossRefPubMed
12.
Zurück zum Zitat Wolak A, Slomka PJ, Fish MB, Lorenzo S, Berman DS, Germano G. Quantitative diagnostic performance of myocardial perfusion SPECT with attenuation correction in women. J Nucl Med 2008;49(6):915-22.CrossRefPubMed Wolak A, Slomka PJ, Fish MB, Lorenzo S, Berman DS, Germano G. Quantitative diagnostic performance of myocardial perfusion SPECT with attenuation correction in women. J Nucl Med 2008;49(6):915-22.CrossRefPubMed
13.
Zurück zum Zitat Wolak A, Slomka PJ, Fish MB, et al. Automated analysis of attenuation-corrected SPECT: Comparison of two state-of-the-art software packages [abstract]. J Nucl Cardiol 2008;14(suppl):S110. Wolak A, Slomka PJ, Fish MB, et al. Automated analysis of attenuation-corrected SPECT: Comparison of two state-of-the-art software packages [abstract]. J Nucl Cardiol 2008;14(suppl):S110.
14.
Zurück zum Zitat Diamond GA, Forrester JS. Analysis of probability as an aid in the clinical diagnosis of coronary-artery disease. New Engl J Med 1979;300(24):1350-8.PubMedCrossRef Diamond GA, Forrester JS. Analysis of probability as an aid in the clinical diagnosis of coronary-artery disease. New Engl J Med 1979;300(24):1350-8.PubMedCrossRef
15.
Zurück zum Zitat Slomka PJ, Fish MB, Lorenzo S, et al. Simplified normal limits and automated quantitative assessment for attenuation-corrected myocardial perfusion SPECT. J Nucl Cardiol 2006;13(5):642-51.CrossRefPubMed Slomka PJ, Fish MB, Lorenzo S, et al. Simplified normal limits and automated quantitative assessment for attenuation-corrected myocardial perfusion SPECT. J Nucl Cardiol 2006;13(5):642-51.CrossRefPubMed
16.
Zurück zum Zitat Nishina H, Slomka PJ, Abidov A, et al. Combined supine and prone quantitative myocardial perfusion SPECT: Method development and clinical validation in patients with no known coronary artery disease. J Nucl Med 2006;47(1):51-8.PubMed Nishina H, Slomka PJ, Abidov A, et al. Combined supine and prone quantitative myocardial perfusion SPECT: Method development and clinical validation in patients with no known coronary artery disease. J Nucl Med 2006;47(1):51-8.PubMed
17.
Zurück zum Zitat Germano G, Kiat H, Kavanagh PB, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36(11):2138-47.PubMed Germano G, Kiat H, Kavanagh PB, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36(11):2138-47.PubMed
18.
Zurück zum Zitat Slomka PJ, Nishina H, Berman DS, et al. Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005;12(1):66-77.CrossRefPubMed Slomka PJ, Nishina H, Berman DS, et al. Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005;12(1):66-77.CrossRefPubMed
19.
Zurück zum Zitat Delong ER, Delong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves a nonparametric approach. Biometrics 1988;44(3):837-46.CrossRefPubMed Delong ER, Delong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves a nonparametric approach. Biometrics 1988;44(3):837-46.CrossRefPubMed
20.
Zurück zum Zitat Berman DS, Kang X, Slomka PJ, et al. Underestimation of extent of ischemia by gated SPECT myocardial perfusion imaging in patients with left main coronary artery disease. J Nucl Cardiol 2007;14(4):521-8.CrossRefPubMed Berman DS, Kang X, Slomka PJ, et al. Underestimation of extent of ischemia by gated SPECT myocardial perfusion imaging in patients with left main coronary artery disease. J Nucl Cardiol 2007;14(4):521-8.CrossRefPubMed
21.
Zurück zum Zitat Okuda K, Nakajima K, Motomura N, et al. Attenuation correction of myocardial SPECT by scatter-photopeak window method in normal subjects. Ann Nucl Med 2009;23(5):501-6.CrossRefPubMed Okuda K, Nakajima K, Motomura N, et al. Attenuation correction of myocardial SPECT by scatter-photopeak window method in normal subjects. Ann Nucl Med 2009;23(5):501-6.CrossRefPubMed
22.
Zurück zum Zitat Xu Y, Kavanagh P, Fish M, et al. Automated quality control for segmentation of myocardial perfusion SPECT. J Nucl Med 2009;50(9):1418-26.CrossRefPubMed Xu Y, Kavanagh P, Fish M, et al. Automated quality control for segmentation of myocardial perfusion SPECT. J Nucl Med 2009;50(9):1418-26.CrossRefPubMed
23.
Zurück zum Zitat Bateman TM, Heller GV, McGhie AI, et al. Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol 2009;16(5):726-35.CrossRefPubMed Bateman TM, Heller GV, McGhie AI, et al. Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol 2009;16(5):726-35.CrossRefPubMed
Metadaten
Titel
Combined quantitative analysis of attenuation corrected and non-corrected myocadial perfusion SPECT: Method development and clinical validation
verfasst von
Yuan Xu, PhD
Mathews Fish, MD
James Gerlach, CNMT
Mark Lemley, BSc
Daniel S. Berman, MD
Guido Germano, PhD
Piotr J. Slomka, PhD
Publikationsdatum
01.08.2010
Verlag
Springer-Verlag
Erschienen in
Journal of Nuclear Cardiology / Ausgabe 4/2010
Print ISSN: 1071-3581
Elektronische ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-010-9220-8

Weitere Artikel der Ausgabe 4/2010

Journal of Nuclear Cardiology 4/2010 Zur Ausgabe

Advances in Nonnuclear Imaging Technologies

Molecular imaging with contrast enhanced ultrasound

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.