Skip to main content
Erschienen in: European Radiology 12/2010

01.12.2010 | Magnetic Resonance

Comparison of a new whole-body continuous-table-movement protocol versus a standard whole-body MR protocol for the assessment of multiple myeloma

verfasst von: S. Weckbach, H. J. Michaely, A. Stemmer, S. O. Schoenberg, D. J. Dinter

Erschienen in: European Radiology | Ausgabe 12/2010

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate a whole body (WB) continuous-table-movement (CTM) MR protocol for the assessment of multiple myeloma (MM) in comparison to a step-by-step WB protocol.

Methods

Eighteen patients with MM were examined at 1.5T using a WB CTM protocol (axial T2-w fs BLADE, T1-w GRE sequence) and a step-by-step WB protocol including coronal/sagittal T1-w SE and STIR sequences as reference. Protocol time was assessed. Image quality, artefacts, liver/spleen assessability, and the ability to depict bone marrow lesions less than or greater than 1 cm as well as diffuse infiltration and soft tissue lesions were rated. Potential changes in the Durie and Salmon Plus stage and the detectability of complications were assessed.

Results

Mean protocol time was 6:38 min (CTM) compared to 24:32 min (standard). Image quality was comparable. Artefacts were more prominent using the CTM protocol (P = 0.0039). Organ assessability was better using the CTM protocol (P < 0.001). Depiction of bone marrow and soft tissue lesions was identical without a staging shift. Vertebral fractures were not detected using the CTM protocol.

Conclusions

The new protocol allows a higher patient throughput and facilitates the depiction of extramedullary lesions. However, as long as vertebral fractures are not detectable, the protocol cannot be safely used for clinical routine without the acquisition of an additional sagittal sequence.
Literatur
3.
Zurück zum Zitat Durie BG (2006) The role of anatomic and functional staging in myeloma: description of Durie/Salmon Plus staging system. Eur J Cancer 42:1539–1543CrossRefPubMed Durie BG (2006) The role of anatomic and functional staging in myeloma: description of Durie/Salmon Plus staging system. Eur J Cancer 42:1539–1543CrossRefPubMed
4.
Zurück zum Zitat Baur A, Stabler A, Nagel D et al (2002) Magnetic resonance imaging as a supplement for the clinical staging system of Durie and Salmon? Cancer 95:1334–1345CrossRefPubMed Baur A, Stabler A, Nagel D et al (2002) Magnetic resonance imaging as a supplement for the clinical staging system of Durie and Salmon? Cancer 95:1334–1345CrossRefPubMed
5.
Zurück zum Zitat Baur-Melnyk A, Buhmann S, Durr HR, Reiser M (2005) Role of MRI for the diagnosis and prognosis of multiple myeloma. Eur J Radiol 55:56–63CrossRefPubMed Baur-Melnyk A, Buhmann S, Durr HR, Reiser M (2005) Role of MRI for the diagnosis and prognosis of multiple myeloma. Eur J Radiol 55:56–63CrossRefPubMed
6.
Zurück zum Zitat Bauerle T, Hillengass J, Fechtner K et al (2009) Multiple myeloma and monoclonal gammopathy of undetermined significance: importance of whole-body versus spinal MR imaging. Radiology 252:477–485CrossRefPubMed Bauerle T, Hillengass J, Fechtner K et al (2009) Multiple myeloma and monoclonal gammopathy of undetermined significance: importance of whole-body versus spinal MR imaging. Radiology 252:477–485CrossRefPubMed
7.
8.
Zurück zum Zitat Baur-Melnyk A, Buhmann S, Becker C et al (2008) Whole-body MRI versus whole-body MDCT for staging of multiple myeloma. AJR Am J Roentgenol 190:1097–1104CrossRefPubMed Baur-Melnyk A, Buhmann S, Becker C et al (2008) Whole-body MRI versus whole-body MDCT for staging of multiple myeloma. AJR Am J Roentgenol 190:1097–1104CrossRefPubMed
9.
Zurück zum Zitat Shortt CP, Gleeson TG, Breen KA et al (2009) Whole-body MRI versus PET in assessment of multiple myeloma disease activity. AJR Am J Roentgenol 192:980–986CrossRefPubMed Shortt CP, Gleeson TG, Breen KA et al (2009) Whole-body MRI versus PET in assessment of multiple myeloma disease activity. AJR Am J Roentgenol 192:980–986CrossRefPubMed
10.
Zurück zum Zitat Weininger M, Lauterbach B, Knop S et al (2009) Whole-body MRI of multiple myeloma: comparison of different MRI sequences in assessment of different growth patterns. Eur J Radiol 69:339–345CrossRefPubMed Weininger M, Lauterbach B, Knop S et al (2009) Whole-body MRI of multiple myeloma: comparison of different MRI sequences in assessment of different growth patterns. Eur J Radiol 69:339–345CrossRefPubMed
11.
Zurück zum Zitat Ghanem N, Lohrmann C, Engelhardt M et al (2006) Whole-body MRI in the detection of bone marrow infiltration in patients with plasma cell neoplasms in comparison to the radiological skeletal survey. Eur Radiol 16:1005–1014CrossRefPubMed Ghanem N, Lohrmann C, Engelhardt M et al (2006) Whole-body MRI in the detection of bone marrow infiltration in patients with plasma cell neoplasms in comparison to the radiological skeletal survey. Eur Radiol 16:1005–1014CrossRefPubMed
12.
Zurück zum Zitat Zenge MO, Ladd ME, Quick HH (2009) Novel reconstruction method for three-dimensional axial continuously moving table whole-body magnetic resonance imaging featuring autocalibrated parallel imaging GRAPPA. Magn Reson Med 61:867–873CrossRefPubMed Zenge MO, Ladd ME, Quick HH (2009) Novel reconstruction method for three-dimensional axial continuously moving table whole-body magnetic resonance imaging featuring autocalibrated parallel imaging GRAPPA. Magn Reson Med 61:867–873CrossRefPubMed
13.
Zurück zum Zitat Kruger DG, Riederer SJ, Polzin JA, Madhuranthakam AJ, Hu HH, Glockner JF (2005) Dual-velocity continuously moving table acquisition for contrast-enhanced peripheral magnetic resonance angiography. Magn Reson Med 53:110–117CrossRefPubMed Kruger DG, Riederer SJ, Polzin JA, Madhuranthakam AJ, Hu HH, Glockner JF (2005) Dual-velocity continuously moving table acquisition for contrast-enhanced peripheral magnetic resonance angiography. Magn Reson Med 53:110–117CrossRefPubMed
14.
Zurück zum Zitat Kruger DG, Riederer SJ, Rossman PJ, Mostardi PM, Madhuranthakam AJ, Hu HH (2005) Recovery of phase inconsistencies in continuously moving table extended field of view magnetic resonance imaging acquisitions. Magn Reson Med 54:712–717CrossRefPubMed Kruger DG, Riederer SJ, Rossman PJ, Mostardi PM, Madhuranthakam AJ, Hu HH (2005) Recovery of phase inconsistencies in continuously moving table extended field of view magnetic resonance imaging acquisitions. Magn Reson Med 54:712–717CrossRefPubMed
15.
Zurück zum Zitat Zenge MO, Ladd ME, Vogt FM, Brauck K, Barkhausen J, Quick HH (2005) Whole-body magnetic resonance imaging featuring moving table continuous data acquisition with high-precision position feedback. Magn Reson Med 54:707–711CrossRefPubMed Zenge MO, Ladd ME, Vogt FM, Brauck K, Barkhausen J, Quick HH (2005) Whole-body magnetic resonance imaging featuring moving table continuous data acquisition with high-precision position feedback. Magn Reson Med 54:707–711CrossRefPubMed
16.
Zurück zum Zitat Fautz HP, Kannengiesser SA (2006) Sliding multislice (SMS): a new technique for minimum FOV usage in axial continuously moving-table acquisitions. Magn Reson Med 55:363–370CrossRefPubMed Fautz HP, Kannengiesser SA (2006) Sliding multislice (SMS): a new technique for minimum FOV usage in axial continuously moving-table acquisitions. Magn Reson Med 55:363–370CrossRefPubMed
17.
Zurück zum Zitat Aldefeld B, Bornert P, Keupp J (2006) Continuously moving table 3D MRI with lateral frequency-encoding direction. Magn Reson Med 55:1210–1216CrossRefPubMed Aldefeld B, Bornert P, Keupp J (2006) Continuously moving table 3D MRI with lateral frequency-encoding direction. Magn Reson Med 55:1210–1216CrossRefPubMed
18.
Zurück zum Zitat Bornert P, Aldefeld B (2008) Principles of whole-body continuously-moving-table MRI. J Magn Reson Imaging 28:1–12CrossRefPubMed Bornert P, Aldefeld B (2008) Principles of whole-body continuously-moving-table MRI. J Magn Reson Imaging 28:1–12CrossRefPubMed
19.
Zurück zum Zitat Kramer H, Zenge M, Schmitt P, Glaser C, Reiser MF, Herrmann KA (2008) Peripheral magnetic resonance angiography (MRA) with continuous table movement at 3.0T: initial experience compared with step-by-step MRA. Invest Radiol 43:627–634CrossRefPubMed Kramer H, Zenge M, Schmitt P, Glaser C, Reiser MF, Herrmann KA (2008) Peripheral magnetic resonance angiography (MRA) with continuous table movement at 3.0T: initial experience compared with step-by-step MRA. Invest Radiol 43:627–634CrossRefPubMed
20.
Zurück zum Zitat Vogt FM, Zenge MO, Ladd ME et al (2007) Peripheral vascular disease: comparison of continuous MR angiography and conventional MR angiography-pilot study. Radiology 243:229–238CrossRefPubMed Vogt FM, Zenge MO, Ladd ME et al (2007) Peripheral vascular disease: comparison of continuous MR angiography and conventional MR angiography-pilot study. Radiology 243:229–238CrossRefPubMed
21.
Zurück zum Zitat Zenge MO, Vogt FM, Brauck K et al (2006) High-resolution continuously acquired peripheral MR angiography featuring partial parallel imaging GRAPPA. Magn Reson Med 56:859–865CrossRefPubMed Zenge MO, Vogt FM, Brauck K et al (2006) High-resolution continuously acquired peripheral MR angiography featuring partial parallel imaging GRAPPA. Magn Reson Med 56:859–865CrossRefPubMed
22.
Zurück zum Zitat Kruger DG, Riederer SJ, Grimm RC, Rossman PJ (2002) Continuously moving table data acquisition method for long FOV contrast-enhanced MRA and whole-body MRI. Magn Reson Med 47:224–231CrossRefPubMed Kruger DG, Riederer SJ, Grimm RC, Rossman PJ (2002) Continuously moving table data acquisition method for long FOV contrast-enhanced MRA and whole-body MRI. Magn Reson Med 47:224–231CrossRefPubMed
23.
Zurück zum Zitat Brauck K, Zenge MO, Vogt FM et al (2008) Feasibility of whole-body MR with T2-and T1-weighted real-time steady-state free precession sequences during continuous table movement to depict metastases. Radiology 246:910–916CrossRefPubMed Brauck K, Zenge MO, Vogt FM et al (2008) Feasibility of whole-body MR with T2-and T1-weighted real-time steady-state free precession sequences during continuous table movement to depict metastases. Radiology 246:910–916CrossRefPubMed
24.
Zurück zum Zitat Baumann T, Ludwig U, Pache G et al (2009) Continuously moving table MRI with sliding multislice for rectal cancer staging: image quality and lesion detection. Eur J Radiol 73(3):579–587CrossRefPubMed Baumann T, Ludwig U, Pache G et al (2009) Continuously moving table MRI with sliding multislice for rectal cancer staging: image quality and lesion detection. Eur J Radiol 73(3):579–587CrossRefPubMed
25.
Zurück zum Zitat Baumann T, Ludwig U, Pache G et al (2008) Detection of pulmonary nodules with move-during-scan magnetic resonance imaging using a free-breathing turbo inversion recovery magnitude sequence. Invest Radiol 43:359–367CrossRefPubMed Baumann T, Ludwig U, Pache G et al (2008) Detection of pulmonary nodules with move-during-scan magnetic resonance imaging using a free-breathing turbo inversion recovery magnitude sequence. Invest Radiol 43:359–367CrossRefPubMed
26.
Zurück zum Zitat Sommer G, Schaefer AO, Baumann T, Ludwig UA, Fautz HP (2008) Sliding multislice MRI for abdominal staging of patients with pelvic malignancies: a pilot study. J Magn Reson Imaging 27:666–672CrossRefPubMed Sommer G, Schaefer AO, Baumann T, Ludwig UA, Fautz HP (2008) Sliding multislice MRI for abdominal staging of patients with pelvic malignancies: a pilot study. J Magn Reson Imaging 27:666–672CrossRefPubMed
27.
Zurück zum Zitat Bley TA, Tittelbach-Helmrich D, Baumann T et al (2007) Sliding multislice MRI for abdominal staging of rectal gastrointestinal stromal tumours. In Vivo 21:891–894PubMed Bley TA, Tittelbach-Helmrich D, Baumann T et al (2007) Sliding multislice MRI for abdominal staging of rectal gastrointestinal stromal tumours. In Vivo 21:891–894PubMed
28.
Zurück zum Zitat Schaefer JF, Schlemmer HP (2006) Total-body MR-imaging in oncology. Eur Radiol 16:2000–2015CrossRefPubMed Schaefer JF, Schlemmer HP (2006) Total-body MR-imaging in oncology. Eur Radiol 16:2000–2015CrossRefPubMed
29.
Zurück zum Zitat Johnson KM, Leavitt GD, Kayser HW (1997) Total-body MR imaging in as little as 18 seconds. Radiology 202:262–267PubMed Johnson KM, Leavitt GD, Kayser HW (1997) Total-body MR imaging in as little as 18 seconds. Radiology 202:262–267PubMed
30.
Zurück zum Zitat Ludwig U, Sommer G, Zaitsev M, Ghanem N, Hennig J, Fautz HP (2006) 2D axial moving table acquisitions with dynamic slice adaptation. Magn Reson Med 55:423–430CrossRefPubMed Ludwig U, Sommer G, Zaitsev M, Ghanem N, Hennig J, Fautz HP (2006) 2D axial moving table acquisitions with dynamic slice adaptation. Magn Reson Med 55:423–430CrossRefPubMed
31.
Zurück zum Zitat Daffner RH, Lupetin AR, Dash N, Deeb ZL, Sefczek RJ, Schapiro RL (1986) MRI in the detection of malignant infiltration of bone marrow. AJR Am J Roentgenol 146:353–358PubMed Daffner RH, Lupetin AR, Dash N, Deeb ZL, Sefczek RJ, Schapiro RL (1986) MRI in the detection of malignant infiltration of bone marrow. AJR Am J Roentgenol 146:353–358PubMed
32.
Zurück zum Zitat Dinter DJ, Neff WK, Klaus J et al (2008) Comparison of whole-body MR imaging and conventional X-ray examination in patients with multiple myeloma and implications for therapy. Ann Hematol 88:457–464CrossRefPubMed Dinter DJ, Neff WK, Klaus J et al (2008) Comparison of whole-body MR imaging and conventional X-ray examination in patients with multiple myeloma and implications for therapy. Ann Hematol 88:457–464CrossRefPubMed
33.
34.
35.
Zurück zum Zitat Kyle RA, Rajkumar SV (2009) Treatment of multiple myeloma: a comprehensive review. Clin Lymphoma Myeloma 9:278–288CrossRefPubMed Kyle RA, Rajkumar SV (2009) Treatment of multiple myeloma: a comprehensive review. Clin Lymphoma Myeloma 9:278–288CrossRefPubMed
36.
Zurück zum Zitat Barkhausen J, Quick HH, Lauenstein T et al (2001) Whole-body MR imaging in 30 seconds with real-time true FISP and a continuously rolling table platform: feasibility study. Radiology 220:252–256PubMed Barkhausen J, Quick HH, Lauenstein T et al (2001) Whole-body MR imaging in 30 seconds with real-time true FISP and a continuously rolling table platform: feasibility study. Radiology 220:252–256PubMed
37.
Zurück zum Zitat Varettoni M, Corso A, Pica G, Mangiacavalli S, Pascutto C, Lazzarino M (2009) Incidence, presenting features and outcome of extramedullary disease in multiple myeloma: a longitudinal study on 1003 consecutive patients. Ann Oncol 21:325–330CrossRefPubMed Varettoni M, Corso A, Pica G, Mangiacavalli S, Pascutto C, Lazzarino M (2009) Incidence, presenting features and outcome of extramedullary disease in multiple myeloma: a longitudinal study on 1003 consecutive patients. Ann Oncol 21:325–330CrossRefPubMed
Metadaten
Titel
Comparison of a new whole-body continuous-table-movement protocol versus a standard whole-body MR protocol for the assessment of multiple myeloma
verfasst von
S. Weckbach
H. J. Michaely
A. Stemmer
S. O. Schoenberg
D. J. Dinter
Publikationsdatum
01.12.2010
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 12/2010
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-010-1865-9

Weitere Artikel der Ausgabe 12/2010

European Radiology 12/2010 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.