Comparison of performance metrics with digital 2D versus tomosynthesis mammography in the diagnostic setting

verfasst von: Manisha Bahl, Sarah Mercaldo, Charmi A. Vijapura, Anne Marie McCarthy, Constance D. Lehman

Erschienen in: European Radiology | Ausgabe 2/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To compare performance metrics between digital 2D mammography (DM) and digital breast tomosynthesis (DBT) in the diagnostic setting.

Methods

Consecutive diagnostic examinations from August 2008 to February 2011 (DM group) and from January 2013 to July 2015 (DM/DBT group) were reviewed. Core biopsy and surgical pathology results within 365 days after the mammogram were collected. Performance metrics, including cancer detection rate (CDR), abnormal interpretation rate (AIR), positive predictive value (PPV) 2, PPV3, sensitivity, and specificity were calculated. Multivariable logistic regression models were fit to compare performance metrics in the DM and DM/DBT groups while adjusting for clinical covariates.

Results

A total of 22,883 mammograms were performed before DBT integration (DM group), and 22,824 mammograms were performed after complete DBT integration (DM/DBT group). After adjusting for multiple variables, the CDR was similar in both groups (38.2 per 1,000 examinations in the DM/DBT group versus 31.3 per 1,000 examinations in the DM group, p = 0.14); however, a higher proportion of cancers were invasive rather than in situ in the DM/DBT group [83.7% (731/873) versus 72.3% (518/716), p < 0.01]. The AIR was lower in the DM/DBT group (p < 0.01), and PPV2, PPV3, and specificity were higher in the DM/DBT group (all p = 0.01 or p < 0.01).

Conclusions

Complete integration of DBT into the diagnostic setting is associated with improved diagnostic performance. Increased utilization of DBT may thus result in better patient outcomes and lead to a shift in the benchmarks that have been established for DM.

Key Points

• Integration of tomosynthesis into the diagnostic setting is associated with improved performance.
• A higher proportion of cancers are invasive rather than in situ with digital breast tomosynthesis.
• Increased utilization of tomosynthesis may lead to a shift in established benchmarks.
Literatur
1.
Zurück zum Zitat Hooley RJ, Durand MA, Philpotts LE (2017) Advances in digital breast tomosynthesis. AJR Am J Roentgenol 208(2):256–266CrossRef Hooley RJ, Durand MA, Philpotts LE (2017) Advances in digital breast tomosynthesis. AJR Am J Roentgenol 208(2):256–266CrossRef
2.
Zurück zum Zitat Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14(7):583–589CrossRef Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14(7):583–589CrossRef
3.
Zurück zum Zitat Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267(1):47–56CrossRef Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267(1):47–56CrossRef
4.
Zurück zum Zitat Friedewald SM, Rafferty EA, Rose SL et al (2014) Breast cancer screening using tomosynthesis in combination with digital mammography. JAMA 311(24):2499–2507CrossRef Friedewald SM, Rafferty EA, Rose SL et al (2014) Breast cancer screening using tomosynthesis in combination with digital mammography. JAMA 311(24):2499–2507CrossRef
5.
Zurück zum Zitat McCarthy AM, Kontos D, Synnestvedt M et al (2014) Screening outcomes following implementation of digital breast tomosynthesis in a general-population screening program. J Natl Cancer Inst 106(11) McCarthy AM, Kontos D, Synnestvedt M et al (2014) Screening outcomes following implementation of digital breast tomosynthesis in a general-population screening program. J Natl Cancer Inst 106(11)
6.
Zurück zum Zitat Sharpe RE Jr, Venkataraman S, Phillips J et al (2016) Increased cancer detection rate and variations in the recall rate resulting from implementation of 3D digital breast tomosynthesis into a population-based screening program. Radiology 278(3):698–706CrossRef Sharpe RE Jr, Venkataraman S, Phillips J et al (2016) Increased cancer detection rate and variations in the recall rate resulting from implementation of 3D digital breast tomosynthesis into a population-based screening program. Radiology 278(3):698–706CrossRef
7.
Zurück zum Zitat Raghu M, Durand MA, Andrejeva L et al (2016) Tomosynthesis in the diagnostic setting: changing rates of BI-RADS final assessment over time. Radiology 281(1):54–61CrossRef Raghu M, Durand MA, Andrejeva L et al (2016) Tomosynthesis in the diagnostic setting: changing rates of BI-RADS final assessment over time. Radiology 281(1):54–61CrossRef
8.
Zurück zum Zitat Sickles EA, D'Orsi CJ, Bassett LW et al (2013) ACR BI-RADS mammography. In: ACR BI-RADS atlas, breast imaging reporting and data system. American College of Radiology, Reston, VA Sickles EA, D'Orsi CJ, Bassett LW et al (2013) ACR BI-RADS mammography. In: ACR BI-RADS atlas, breast imaging reporting and data system. American College of Radiology, Reston, VA
9.
Zurück zum Zitat Rose SL, Tidwell AL, Bujnoch LJ, Kushwaha AC, Nordmann AS, Sexton R Jr (2013) Implementation of breast tomosynthesis in a routine screening practice: an observational study. AJR Am J Roentgenol 200(6):1401–1408CrossRef Rose SL, Tidwell AL, Bujnoch LJ, Kushwaha AC, Nordmann AS, Sexton R Jr (2013) Implementation of breast tomosynthesis in a routine screening practice: an observational study. AJR Am J Roentgenol 200(6):1401–1408CrossRef
10.
Zurück zum Zitat Bahl M, Gaffney S, McCarthy AM, Lowry KP, Dang PA, Lehman CD (2017) Breast cancer characteristics associated with 2D digital mammography versus digital breast tomosynthesis for screening-detected and interval cancers. Radiology 287(1):49–57CrossRef Bahl M, Gaffney S, McCarthy AM, Lowry KP, Dang PA, Lehman CD (2017) Breast cancer characteristics associated with 2D digital mammography versus digital breast tomosynthesis for screening-detected and interval cancers. Radiology 287(1):49–57CrossRef
11.
Zurück zum Zitat Sprague BL, Arao RF, Miglioretti DL et al (2017) National performance benchmarks for modern diagnostic digital mammography: update from the Breast Cancer Surveillance Consortium. Radiology 283(1):59–69CrossRef Sprague BL, Arao RF, Miglioretti DL et al (2017) National performance benchmarks for modern diagnostic digital mammography: update from the Breast Cancer Surveillance Consortium. Radiology 283(1):59–69CrossRef
12.
Zurück zum Zitat Bahl M, Lamb LR, Lehman CD (2017) Pathologic outcomes of architectural distortion on digital 2D versus tomosynthesis mammography. AJR Am J Roentgenol 209(5):1162–1167CrossRef Bahl M, Lamb LR, Lehman CD (2017) Pathologic outcomes of architectural distortion on digital 2D versus tomosynthesis mammography. AJR Am J Roentgenol 209(5):1162–1167CrossRef
13.
Zurück zum Zitat Poplack SP, Tosteson TD, Kogel CA, Nagy HM (2007) Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. AJR Am J Roentgenol 189(3):616–623CrossRef Poplack SP, Tosteson TD, Kogel CA, Nagy HM (2007) Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. AJR Am J Roentgenol 189(3):616–623CrossRef
14.
Zurück zum Zitat Gennaro G, Toledano A, di Maggio C et al (2010) Digital breast tomosynthesis versus digital mammography: a clinical performance study. Eur Radiol 20(7):1545–1553CrossRef Gennaro G, Toledano A, di Maggio C et al (2010) Digital breast tomosynthesis versus digital mammography: a clinical performance study. Eur Radiol 20(7):1545–1553CrossRef
15.
Zurück zum Zitat Hakim CM, Chough DM, Ganott MA, Sumkin JH, Zuley ML, Gur D (2010) Digital breast tomosynthesis in the diagnostic environment: a subjective side-by-side review. AJR Am J Roentgenol 195(2):W172–W176CrossRef Hakim CM, Chough DM, Ganott MA, Sumkin JH, Zuley ML, Gur D (2010) Digital breast tomosynthesis in the diagnostic environment: a subjective side-by-side review. AJR Am J Roentgenol 195(2):W172–W176CrossRef
16.
Zurück zum Zitat Noroozian M, Hadjiiski L, Rahnama-Moghadam S et al (2012) Digital breast tomosynthesis is comparable to mammographic spot views for mass characterization. Radiology 262(1):61–68CrossRef Noroozian M, Hadjiiski L, Rahnama-Moghadam S et al (2012) Digital breast tomosynthesis is comparable to mammographic spot views for mass characterization. Radiology 262(1):61–68CrossRef
17.
Zurück zum Zitat Tagliafico A, Astengo D, Cavagnetto F et al (2012) One-to-one comparison between digital spot compression view and digital breast tomosynthesis. Eur Radiol 22(3):539–544CrossRef Tagliafico A, Astengo D, Cavagnetto F et al (2012) One-to-one comparison between digital spot compression view and digital breast tomosynthesis. Eur Radiol 22(3):539–544CrossRef
18.
Zurück zum Zitat Brandt KR, Craig DA, Hoskins TL et al (2013) Can digital breast tomosynthesis replace conventional diagnostic mammography views for screening recalls without calcifications? A comparison study in a simulated clinical setting. AJR Am J Roentgenol 200(2):291–298CrossRef Brandt KR, Craig DA, Hoskins TL et al (2013) Can digital breast tomosynthesis replace conventional diagnostic mammography views for screening recalls without calcifications? A comparison study in a simulated clinical setting. AJR Am J Roentgenol 200(2):291–298CrossRef
19.
Zurück zum Zitat Waldherr C, Cerny P, Altermatt HJ et al (2013) Value of one-view breast tomosynthesis versus two-view mammography in diagnostic workup of women with clinical signs and symptoms and in women recalled from screening. AJR Am J Roentgenol 200(1):226–231CrossRef Waldherr C, Cerny P, Altermatt HJ et al (2013) Value of one-view breast tomosynthesis versus two-view mammography in diagnostic workup of women with clinical signs and symptoms and in women recalled from screening. AJR Am J Roentgenol 200(1):226–231CrossRef
20.
Zurück zum Zitat Zuley ML, Bandos AI, Ganott MA et al (2013) Digital breast tomosynthesis versus supplemental diagnostic mammographic views for evaluation of noncalcified breast lesions. Radiology 266(1):89–95CrossRef Zuley ML, Bandos AI, Ganott MA et al (2013) Digital breast tomosynthesis versus supplemental diagnostic mammographic views for evaluation of noncalcified breast lesions. Radiology 266(1):89–95CrossRef
21.
Zurück zum Zitat Lei J, Yang P, Zhang L, Wang Y, Yang K (2014) Diagnostic accuracy of digital breast tomosynthesis versus digital mammography for benign and malignant lesions in breasts: a meta-analysis. Eur Radiol 24(3):595–602CrossRef Lei J, Yang P, Zhang L, Wang Y, Yang K (2014) Diagnostic accuracy of digital breast tomosynthesis versus digital mammography for benign and malignant lesions in breasts: a meta-analysis. Eur Radiol 24(3):595–602CrossRef
22.
Zurück zum Zitat Peppard HR, Nicholson BE, Rochman CM, Merchant JK, Mayo RC 3rd, Harvey JA (2015) Digital breast tomosynthesis in the diagnostic setting: indications and clinical applications. Radiographics 35(4):975–990CrossRef Peppard HR, Nicholson BE, Rochman CM, Merchant JK, Mayo RC 3rd, Harvey JA (2015) Digital breast tomosynthesis in the diagnostic setting: indications and clinical applications. Radiographics 35(4):975–990CrossRef
23.
Zurück zum Zitat Poplack S (2017) Breast tomosynthesis: clinical evidence. Radiol Clin North Am 55(3):475–492CrossRef Poplack S (2017) Breast tomosynthesis: clinical evidence. Radiol Clin North Am 55(3):475–492CrossRef
24.
Zurück zum Zitat Spangler ML, Zuley ML, Sumkin JH et al (2011) Detection and classification of calcifications on digital breast tomosynthesis and 2D digital mammography: a comparison. AJR Am J Roentgenol 196(2):320–324CrossRef Spangler ML, Zuley ML, Sumkin JH et al (2011) Detection and classification of calcifications on digital breast tomosynthesis and 2D digital mammography: a comparison. AJR Am J Roentgenol 196(2):320–324CrossRef
25.
Zurück zum Zitat Svahn TM, Houssami N, Sechopoulos I, Mattsson S (2015) Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography. Breast 24(2):93–99CrossRef Svahn TM, Houssami N, Sechopoulos I, Mattsson S (2015) Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography. Breast 24(2):93–99CrossRef
26.
Zurück zum Zitat Gennaro G, Bernardi D, Houssami N (2018) Radiation dose with digital breast tomosynthesis compared to digital mammography: per-view analysis. Eur Radiol 28(2):573–581CrossRef Gennaro G, Bernardi D, Houssami N (2018) Radiation dose with digital breast tomosynthesis compared to digital mammography: per-view analysis. Eur Radiol 28(2):573–581CrossRef
Metadaten
Titel
Comparison of performance metrics with digital 2D versus tomosynthesis mammography in the diagnostic setting
verfasst von
Manisha Bahl
Sarah Mercaldo
Charmi A. Vijapura
Anne Marie McCarthy
Constance D. Lehman
Publikationsdatum
02.07.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5596-7

Weitere Artikel der Ausgabe 2/2019

European Radiology 2/2019 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.