01.12.2015 | Research article | Ausgabe 1/2015 Open Access

BMC Anesthesiology 1/2015

Comparisons of clinical performance of guardian laryngeal mask with laryngeal mask airway ProSeal

Zeitschrift:
BMC Anesthesiology > Ausgabe 1/2015
Autoren:
Ajay Kumar Pajiyar, Zhiting Wen, Haiyun Wang, Lin Ma, Lumin Miao, Guolin Wang
Wichtige Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​s12871-015-0039-3) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

AKP helped design the study, conduct the study, analyze the data and write the initial draft of the manuscript. ZW recruited the patients, performed many measurement, data collection and analysis. LMM recruited the patient, data collection and analysis. LM recruited the patient, collected data and analyzed them, and prepared the manuscript. HW designed the study, collected and analyzed the data, and revised the manuscript. GW designed the study, collected and analyzed the data, and revised the manuscript. All authors read and approved the final manuscript.

Abstract

Background

The Guardian Laryngeal Mask Airway (G-LMA) is a new silicone-based single-use extraglottic device with the drainage port and a cuff pilot valve with pressure indicator. The aim of this study is to compare the clinical performance of this laryngeal mask airway with ProSeal laryngeal mask airway (P-LMA).

Methods

In this prospective randomized study, we included adult patients with ASA grading I and II scheduled for elective surgery requiring supine position under total intravenous anesthesia. The patients were randomly allocated to two groups, 40 in each. G-LMA and P-LMA were used in groups G and P respectively. The cuff of each device was air inflated to 60 cmH2O. The primary outcome was to compare the airway sealing pressure and the secondary outcome was to compare the efficacy and safety of these two devices with respect to insertion success, insertion time, ease of insertion, volume of air for cuff inflation to 60 cmH2O, intracuff pressure measurement, gastric tube insertion attempt, gastric tube insertion time, Fiberoptic laryngeal view, and postoperative pharyngolaryngeal morbidity.

Results

The airway sealing pressure at 60cmH2O cuff pressure was significantly greater in G-LMA than P-LMA (p = 0.04).The first successful attempt of both groups were comparable (p = 1.000). Insertion time was significantly shorter in G-LMA than P-LMA (p < 0.0001). The first successful attempt for the gastric tube insertion in both groups was comparable (p = 0.431). Gastric tube insertion time was less in G-LMA than in P-LMA (p < 0.0001). The volume of air for cuff inflation to 60 cmH2O was more in G-LMA than in P-LMA (<0.0001). The intracuff pressure measurement at 30, 60, 90 and 120 minutes were comparable (p = 0.823, 0.182, 0.870, 0.658).We did not find differences in ease of insertion (p = 0.60); Fiber-optic positions of airway devices were comparable (p = 0.83). In addition, blood staining (p = 1.00), sore throat and dysphagia at 1, 2 and 24 hour (p = 1.00) were comparable in both groups.

Conclusion

The Guardian laryngeal mask airway was associated with high airway sealing pressure with a quicker insertion of the device as well as gastric tube.

Trial registration

Clinical Trial.gov Identifier: NCT02063516. Date: June 2013
Zusatzmaterial
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 1/2015

BMC Anesthesiology 1/2015 Zur Ausgabe

Neu im Fachgebiet AINS

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update AINS und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise