Digital radiography: optimization of image quality and dose using multi-frequency software

verfasst von: H. Precht, O. Gerke, K. Rosendahl, A. Tingberg, D. Waaler

Erschienen in: Pediatric Radiology | Ausgabe 9/2012

Einloggen, um Zugang zu erhalten

Abstract

Background

New developments in processing of digital radiographs (DR), including multi-frequency processing (MFP), allow optimization of image quality and radiation dose. This is particularly promising in children as they are believed to be more sensitive to ionizing radiation than adults.

Objective

To examine whether the use of MFP software reduces the radiation dose without compromising quality at DR of the femur in 5-year-old-equivalent anthropomorphic and technical phantoms.

Materials and methods

A total of 110 images of an anthropomorphic phantom were imaged on a DR system (Canon DR with CXDI-50 C detector and MLT[S] software) and analyzed by three pediatric radiologists using Visual Grading Analysis. In addition, 3,500 images taken of a technical contrast-detail phantom (CDRAD 2.0) provide an objective image-quality assessment.

Results

Optimal image-quality was maintained at a dose reduction of 61% with MLT(S) optimized images. Even for images of diagnostic quality, MLT(S) provided a dose reduction of 88% as compared to the reference image. Software impact on image quality was found significant for dose (mAs), dynamic range dark region and frequency band.

Conclusion

By optimizing image processing parameters, a significant dose reduction is possible without significant loss of image quality.
Literatur
1.
Zurück zum Zitat ICRP (2004) Managing patient dose in digital radiology. ICRP Publication 93. Ann. ICRP 34:1–73 ICRP (2004) Managing patient dose in digital radiology. ICRP Publication 93. Ann. ICRP 34:1–73
2.
Zurück zum Zitat Reiner BI, Siegel EL, Siddiqui K et al (2006) Quality assurance: the missing link. Radiology 238:13–16PubMedCrossRef Reiner BI, Siegel EL, Siddiqui K et al (2006) Quality assurance: the missing link. Radiology 238:13–16PubMedCrossRef
3.
Zurück zum Zitat Seeram E (2011) Digital radiography, an introduction. Delmar Cengage Learning, New York Seeram E (2011) Digital radiography, an introduction. Delmar Cengage Learning, New York
4.
Zurück zum Zitat Gonzales RC, Woods RE (2008) Digital image processing. Prentice Hall (3), New Jersey Gonzales RC, Woods RE (2008) Digital image processing. Prentice Hall (3), New Jersey
5.
Zurück zum Zitat Canon Inc. (2008) CXDI Image Processing Software MLT(S) User’s Manual. Japan Canon Inc. (2008) CXDI Image Processing Software MLT(S) User’s Manual. Japan
6.
Zurück zum Zitat Beutel J, Sonka M, Fitzpatrick JM (2004) Handbook of medical imaging, vol. 2; Medical image processing and analysis. SPIE, Washington Beutel J, Sonka M, Fitzpatrick JM (2004) Handbook of medical imaging, vol. 2; Medical image processing and analysis. SPIE, Washington
7.
Zurück zum Zitat Bushberg JT, Seibert JA, Leidholdt EM et al (2002) The essential physics of medical imaging. Lippincott Williams & Wilkins (2), Philadelphia Bushberg JT, Seibert JA, Leidholdt EM et al (2002) The essential physics of medical imaging. Lippincott Williams & Wilkins (2), Philadelphia
9.
Zurück zum Zitat Huda W (2004) Assessment of the problem: pediatric doses in screen-film and digital radiography. Pediatr Radiol 34(3):S173–S182PubMedCrossRef Huda W (2004) Assessment of the problem: pediatric doses in screen-film and digital radiography. Pediatr Radiol 34(3):S173–S182PubMedCrossRef
10.
Zurück zum Zitat European Commission (1996) European guidelines on quality criteria for diagnostic radiographic images in pediatrics. Luxemborg European Commission (1996) European guidelines on quality criteria for diagnostic radiographic images in pediatrics. Luxemborg
11.
Zurück zum Zitat Danish Society of Radiology (2006) Vejledning vedrørende Radiologiske Procedure Danish Society of Radiology (2006) Vejledning vedrørende Radiologiske Procedure
12.
Zurück zum Zitat Rapp-Bernhardt U, Bernhardt T, Lenzen H et al (2005) Experimental evaluation of a portable indirect flat-panel detector for the pediatric chest: comparison with storage phosphor radiography at different exposures by using a chest phantom. Radiology 237:485–491PubMedCrossRef Rapp-Bernhardt U, Bernhardt T, Lenzen H et al (2005) Experimental evaluation of a portable indirect flat-panel detector for the pediatric chest: comparison with storage phosphor radiography at different exposures by using a chest phantom. Radiology 237:485–491PubMedCrossRef
13.
Zurück zum Zitat Ludwig K, Ahlers K, Wormanns D et al (2005) Lumbar spine radiography: digital flat-panel detector versus screen-film and storage-phosphor systems in monkeys as a pediatric model. Radiology 229:140–154CrossRef Ludwig K, Ahlers K, Wormanns D et al (2005) Lumbar spine radiography: digital flat-panel detector versus screen-film and storage-phosphor systems in monkeys as a pediatric model. Radiology 229:140–154CrossRef
14.
Zurück zum Zitat Hansson H, Båth M, Håkansson M et al (2005) An optimization strategy in a digital environment applied to neonatal chest imaging. Radiat Protect Dosim 114:278–285CrossRef Hansson H, Båth M, Håkansson M et al (2005) An optimization strategy in a digital environment applied to neonatal chest imaging. Radiat Protect Dosim 114:278–285CrossRef
15.
Zurück zum Zitat Uffmann M, Schaefer-Prokop C, Neitzel U et al (2005) Skeletal applications for flat-panel versus storage-phosphor radiography: effect of exposure on detection of low-contrast details. Radiology 231:506–514CrossRef Uffmann M, Schaefer-Prokop C, Neitzel U et al (2005) Skeletal applications for flat-panel versus storage-phosphor radiography: effect of exposure on detection of low-contrast details. Radiology 231:506–514CrossRef
16.
Zurück zum Zitat Artinis (2006) Manual—contrast-detail phantom. Artinis CDRAD type 2.0, Zetten Artinis (2006) Manual—contrast-detail phantom. Artinis CDRAD type 2.0, Zetten
17.
Zurück zum Zitat International Commission on Radiation units and Measurements (1989) Tissue substitutes in radiation dosimetry and measurement. ICRU Report 44, Maryland International Commission on Radiation units and Measurements (1989) Tissue substitutes in radiation dosimetry and measurement. ICRU Report 44, Maryland
19.
Zurück zum Zitat Tingberg A, Sjöström D (2005) Optimisation of image plate radiography with respect to tube voltage. Radiat Protect Dosim 114:1–3CrossRef Tingberg A, Sjöström D (2005) Optimisation of image plate radiography with respect to tube voltage. Radiat Protect Dosim 114:1–3CrossRef
20.
Zurück zum Zitat Almén A, Tingberg A, Mattsson S et al (2000) The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria. Br J Radiol 73:1192–1199PubMed Almén A, Tingberg A, Mattsson S et al (2000) The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria. Br J Radiol 73:1192–1199PubMed
21.
Zurück zum Zitat Sund P, Båth M, Kheddache S et al (2004) Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography. Eur Radiol 14(1):143–150 Sund P, Båth M, Kheddache S et al (2004) Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography. Eur Radiol 14(1):143–150
22.
Zurück zum Zitat Månsson LG (2000) Methods for the evaluation of image quality: a review. Radiat Protect Dosim 90:89–99CrossRef Månsson LG (2000) Methods for the evaluation of image quality: a review. Radiat Protect Dosim 90:89–99CrossRef
23.
Zurück zum Zitat Bontrager KL (2002) Textbook of radiographic positioning and related anatomi. Bontrager Publishing (4), Arizona Bontrager KL (2002) Textbook of radiographic positioning and related anatomi. Bontrager Publishing (4), Arizona
24.
Zurück zum Zitat Lanhede B, Båth M, Kherddache S et al (2002) The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteria. Br J Radiol 75:38–49PubMed Lanhede B, Båth M, Kherddache S et al (2002) The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteria. Br J Radiol 75:38–49PubMed
25.
Zurück zum Zitat Brennan PC, Johnston D (2002) Irish X-ray departments demonstrate varying levels of adherence to European guidelines on good radiographic technique. Br J Radiol 75:243–248PubMed Brennan PC, Johnston D (2002) Irish X-ray departments demonstrate varying levels of adherence to European guidelines on good radiographic technique. Br J Radiol 75:243–248PubMed
26.
Zurück zum Zitat Rainford LA, Al-Qattan E, McFadden S et al (2006) CEC analysis of radiological images produced in Europe and Asia. Radiography 13:202–209CrossRef Rainford LA, Al-Qattan E, McFadden S et al (2006) CEC analysis of radiological images produced in Europe and Asia. Radiography 13:202–209CrossRef
27.
Zurück zum Zitat Larsen ALS (2006) Videnskab og forskning: en lærebog for professionsuddannelser. Gads Forlag (2):145 Larsen ALS (2006) Videnskab og forskning: en lærebog for professionsuddannelser. Gads Forlag (2):145
28.
Zurück zum Zitat Polit DF, Beck CT (2008) Nursing research generating and assessing evidence for nurcing practice. Lippincott Williams & Wilkins (8), Philadelphia Polit DF, Beck CT (2008) Nursing research generating and assessing evidence for nurcing practice. Lippincott Williams & Wilkins (8), Philadelphia
29.
Zurück zum Zitat Norrman E (2007) Optimisation of radiographic imaging by means of factorial experiments. Doctoral Dissertation, University of Örebro, Sweden Norrman E (2007) Optimisation of radiographic imaging by means of factorial experiments. Doctoral Dissertation, University of Örebro, Sweden
30.
Zurück zum Zitat Fleiss JL (1971) Measuring nominal scale agreement among many raters. Psychol Bull 76:378–382CrossRef Fleiss JL (1971) Measuring nominal scale agreement among many raters. Psychol Bull 76:378–382CrossRef
31.
Zurück zum Zitat Efron B (1979) Bootstrap methods: another look at the jackknife. Ann Stat 7:1–26CrossRef Efron B (1979) Bootstrap methods: another look at the jackknife. Ann Stat 7:1–26CrossRef
32.
Zurück zum Zitat Miller DP (2004) Obtain robust confidence intervals for any statistic. SAS Institute Inc. Proceedings of the Twenty-Ninth Annual SAS (R) User Group International Conference. SAS Institute Inc, North Carolina Miller DP (2004) Obtain robust confidence intervals for any statistic. SAS Institute Inc. Proceedings of the Twenty-Ninth Annual SAS (R) User Group International Conference. SAS Institute Inc, North Carolina
33.
Zurück zum Zitat Canon Inc. (2008) Multiobjective frequency processing function manual—MLT(S) edition. Japan Canon Inc. (2008) Multiobjective frequency processing function manual—MLT(S) edition. Japan
34.
Zurück zum Zitat Beutel J, Kundel HL, Van Metter RL (2000) Handbook of medical imaging, vol. 1; Physics and Psychophysics. SPIE, Washington Beutel J, Kundel HL, Van Metter RL (2000) Handbook of medical imaging, vol. 1; Physics and Psychophysics. SPIE, Washington
35.
Zurück zum Zitat Crevret S (2006) Statistical methods for dose-finding experiments. Wiley, EnglandCrossRef Crevret S (2006) Statistical methods for dose-finding experiments. Wiley, EnglandCrossRef
36.
Zurück zum Zitat Eubank RL (1999) Nonparametric regression and spline smoothing. CRC press, New York Eubank RL (1999) Nonparametric regression and spline smoothing. CRC press, New York
37.
Zurück zum Zitat Härdle WK, Müller M, Sperlich S et al (2004) Nonparametric and semiparametric models. Springer, BerlinCrossRef Härdle WK, Müller M, Sperlich S et al (2004) Nonparametric and semiparametric models. Springer, BerlinCrossRef
Metadaten
Titel
Digital radiography: optimization of image quality and dose using multi-frequency software
verfasst von
H. Precht
O. Gerke
K. Rosendahl
A. Tingberg
D. Waaler
Publikationsdatum
01.09.2012
Verlag
Springer-Verlag
Erschienen in
Pediatric Radiology / Ausgabe 9/2012
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-012-2385-3

Weitere Artikel der Ausgabe 9/2012

Pediatric Radiology 9/2012 Zur Ausgabe

Hermes

Hermes

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.