16.07.2019 | Computed Tomography

Dual-layer spectral CT improves image quality of multiphasic pancreas CT in patients with pancreatic ductal adenocarcinoma

verfasst von: Yasunori Nagayama, Shota Tanoue, Taihei Inoue, Seitaro Oda, Takeshi Nakaura, Daisuke Utsunomiya, Yasuyuki Yamashita

Erschienen in: European Radiology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the image quality and optimal energies of virtual monoenergetic images (VMIs) from dual-layer spectral detector computed tomography (DLCT) in multiphasic pancreatic CT and investigate whether low-keV VMI at the portal venous phase (PVP) provides sufficient tumor conspicuity and arterial depiction relative to conventional pancreatic parenchymal phase (PPP) images.

Methods

Forty-eight patients with pancreatic ductal adenocarcinoma (PDAC) underwent contrast-enhanced DLCT during PPP and PVP. Conventional polyenergetic images (PEIs) and VMI at 40–100 keV (VMI40–100, 10-keV increments) were reconstructed at each enhancement phase. Image noise and the contrast-to-noise ratio (CNR) of the pancreas, tumors, arteries, and veins were quantified. Two radiologists independently assessed tumor conspicuity, margin delineation, image noise, sharpness of pancreatic duct, and depiction of arteries and veins on a five-point scale. Size-specific dose estimate (SSDE) was calculated.

Results

Image noise for VMI40–100 was significantly lower than that for PEI (p < 0.01). The CNR in VMI increased gradually with decreasing energy; CNRs for VMI40–60 were significantly greater than that for PEI (p < 0.01). All subjective VMI scores were maximized at VMI40, followed by VMI50–60, all of which were significantly better than of PEI (p < 0.01). Objective and subjective image qualities of VMI40–50 at the PVP were equivalent to or even better compared with conventional PPP images. No significant difference in SSDE was observed between phases (p = 0.10).

Conclusions

DLCT-VMI improved the subjective and objective image quality in multiphasic pancreatic CT for patients with PDAC. Low-keV PVP imaging may yield diagnostically adequate tumor conspicuity and arterial assessment compared with polyenergetic PPP images.

Key Points

• Low-keV VMI from DLCT yields better subjective and objective image quality of multiphasic pancreas CT in comparison with conventional PEI for the assessment of pancreatic ductal adenocarcinoma.
• Tumor conspicuity and depiction of peripancreatic vasculature were maximized at VMI 40 without an increase in the image noise.
• Low-keV VMI of the portal venous phase provides sufficient tumor conspicuity and arterial depiction, potentially allowing the early detection and local staging of PDAC on routine abdominal CT performed for various clinical indications.
Literatur
1.
Zurück zum Zitat Siegel RL, Miller KD, Jemal A (2019) Cancer statistics, 2019. CA Cancer J Clin 69:7–34CrossRef Siegel RL, Miller KD, Jemal A (2019) Cancer statistics, 2019. CA Cancer J Clin 69:7–34CrossRef
2.
Zurück zum Zitat Konstantinidis IT, Warshaw AL, Allen JN et al (2013) Pancreatic ductal adenocarcinoma: is there a survival difference for R1 resections versus locally advanced unresectable tumors? What is a “true” R0 resection? Ann Surg 257:731–736CrossRef Konstantinidis IT, Warshaw AL, Allen JN et al (2013) Pancreatic ductal adenocarcinoma: is there a survival difference for R1 resections versus locally advanced unresectable tumors? What is a “true” R0 resection? Ann Surg 257:731–736CrossRef
3.
Zurück zum Zitat Somers I, Bipat S (2017) Contrast-enhanced CT in determining resectability in patients with pancreatic carcinoma: a meta-analysis of the positive predictive values of CT. Eur Radiol 27:3408–3435CrossRef Somers I, Bipat S (2017) Contrast-enhanced CT in determining resectability in patients with pancreatic carcinoma: a meta-analysis of the positive predictive values of CT. Eur Radiol 27:3408–3435CrossRef
4.
Zurück zum Zitat Al-Hawary MM, Francis IR, Chari ST et al (2014) Pancreatic ductal adenocarcinoma radiology reporting template: consensus statement of the Society of Abdominal Radiology and the American Pancreatic Association. Radiology 270:248–260CrossRef Al-Hawary MM, Francis IR, Chari ST et al (2014) Pancreatic ductal adenocarcinoma radiology reporting template: consensus statement of the Society of Abdominal Radiology and the American Pancreatic Association. Radiology 270:248–260CrossRef
5.
Zurück zum Zitat Zaky AM, Wolfgang CL, Weiss MJ, Javed AA, Fishman EK, Zaheer A (2017) Tumor-vessel relationships in pancreatic ductal adenocarcinoma at multidetector CT: different classification systems and their influence on treatment planning. Radiographics 37:93–112CrossRef Zaky AM, Wolfgang CL, Weiss MJ, Javed AA, Fishman EK, Zaheer A (2017) Tumor-vessel relationships in pancreatic ductal adenocarcinoma at multidetector CT: different classification systems and their influence on treatment planning. Radiographics 37:93–112CrossRef
6.
Zurück zum Zitat Marchegiani G, Todaro V, Boninsegna E et al (2018) Surgery after FOLFIRINOX treatment for locally advanced and borderline resectable pancreatic cancer: increase in tumour attenuation on CT correlates with R0 resection. Eur Radiol 28:4265–4273CrossRef Marchegiani G, Todaro V, Boninsegna E et al (2018) Surgery after FOLFIRINOX treatment for locally advanced and borderline resectable pancreatic cancer: increase in tumour attenuation on CT correlates with R0 resection. Eur Radiol 28:4265–4273CrossRef
7.
Zurück zum Zitat Lu DS, Vedantham S, Krasny RM, Kadell B, Berger WL, Reber HA (1996) Two-phase helical CT for pancreatic tumors: pancreatic versus hepatic phase enhancement of tumor, pancreas, and vascular structures. Radiology 199:697–701CrossRef Lu DS, Vedantham S, Krasny RM, Kadell B, Berger WL, Reber HA (1996) Two-phase helical CT for pancreatic tumors: pancreatic versus hepatic phase enhancement of tumor, pancreas, and vascular structures. Radiology 199:697–701CrossRef
8.
Zurück zum Zitat Ichikawa T, Erturk SM, Sou H et al (2006) MDCT of pancreatic adenocarcinoma: optimal imaging phases and multiplanar reformatted imaging. AJR Am J Roentgenol 187:1513–1520CrossRef Ichikawa T, Erturk SM, Sou H et al (2006) MDCT of pancreatic adenocarcinoma: optimal imaging phases and multiplanar reformatted imaging. AJR Am J Roentgenol 187:1513–1520CrossRef
9.
Zurück zum Zitat Prokesch RW, Chow LC, Beaulieu CF, Bammer R, Jeffrey RB Jr (2002) Isoattenuating pancreatic adenocarcinoma at multi-detector row CT: secondary signs. Radiology 224:764–768CrossRef Prokesch RW, Chow LC, Beaulieu CF, Bammer R, Jeffrey RB Jr (2002) Isoattenuating pancreatic adenocarcinoma at multi-detector row CT: secondary signs. Radiology 224:764–768CrossRef
10.
Zurück zum Zitat Ishigami K, Yoshimitsu K, Irie H et al (2009) Diagnostic value of the delayed phase image for iso-attenuating pancreatic carcinomas in the pancreatic parenchymal phase on multidetector computed tomography. Eur J Radiol 69:139–146CrossRef Ishigami K, Yoshimitsu K, Irie H et al (2009) Diagnostic value of the delayed phase image for iso-attenuating pancreatic carcinomas in the pancreatic parenchymal phase on multidetector computed tomography. Eur J Radiol 69:139–146CrossRef
11.
Zurück zum Zitat Falck C, Galanski M, Shin H-o (2010) Informatics in radiology: sliding-thin-slab averaging for improved depiction of low-contrast lesions with radiation dose savings at thin-section CT. Radiographics 30:317–326CrossRef Falck C, Galanski M, Shin H-o (2010) Informatics in radiology: sliding-thin-slab averaging for improved depiction of low-contrast lesions with radiation dose savings at thin-section CT. Radiographics 30:317–326CrossRef
12.
Zurück zum Zitat Yanaga Y, Awai K, Nakayama Y et al (2007) Pancreas: patient body weight tailored contrast material injection protocol versus fixed dose protocol at dynamic CT. Radiology 245:475–482CrossRef Yanaga Y, Awai K, Nakayama Y et al (2007) Pancreas: patient body weight tailored contrast material injection protocol versus fixed dose protocol at dynamic CT. Radiology 245:475–482CrossRef
13.
Zurück zum Zitat Kondo H, Kanematsu M, Goshima S et al (2007) MDCT of the pancreas: optimizing scanning delay with a bolus-tracking technique for pancreatic, peripancreatic vascular, and hepatic contrast enhancement. AJR Am J Roentgenol 188:751–756CrossRef Kondo H, Kanematsu M, Goshima S et al (2007) MDCT of the pancreas: optimizing scanning delay with a bolus-tracking technique for pancreatic, peripancreatic vascular, and hepatic contrast enhancement. AJR Am J Roentgenol 188:751–756CrossRef
14.
Zurück zum Zitat Marin D, Nelson RC, Barnhart H et al (2010) Detection of pancreatic tumors, image quality, and radiation dose during the pancreatic parenchymal phase: effect of a low-tube-voltage, high-tube-current CT technique--preliminary results. Radiology 256:450–459CrossRef Marin D, Nelson RC, Barnhart H et al (2010) Detection of pancreatic tumors, image quality, and radiation dose during the pancreatic parenchymal phase: effect of a low-tube-voltage, high-tube-current CT technique--preliminary results. Radiology 256:450–459CrossRef
15.
Zurück zum Zitat Loizou L, Albiin N, Leidner B et al (2016) Multidetector CT of pancreatic ductal adenocarcinoma: effect of tube voltage and iodine load on tumour conspicuity and image quality. Eur Radiol 26:4021–4029CrossRef Loizou L, Albiin N, Leidner B et al (2016) Multidetector CT of pancreatic ductal adenocarcinoma: effect of tube voltage and iodine load on tumour conspicuity and image quality. Eur Radiol 26:4021–4029CrossRef
16.
Zurück zum Zitat Lin X-Z, Machida H, RT IT et al (2014) CT of the pancreas: comparison of image quality and pancreatic duct depiction among model-based iterative, adaptive statistical iterative, and filtered back projection reconstruction techniques. Abdom Imaging 39:497–505CrossRef Lin X-Z, Machida H, RT IT et al (2014) CT of the pancreas: comparison of image quality and pancreatic duct depiction among model-based iterative, adaptive statistical iterative, and filtered back projection reconstruction techniques. Abdom Imaging 39:497–505CrossRef
17.
Zurück zum Zitat Yamamura S, Oda S, Utsunomiya D et al (2013) Dynamic computed tomography of locally advanced pancreatic cancer: effect of low tube voltage and a hybrid iterative reconstruction algorithm on image quality. J Comput Assist Tomogr 37:790–796CrossRef Yamamura S, Oda S, Utsunomiya D et al (2013) Dynamic computed tomography of locally advanced pancreatic cancer: effect of low tube voltage and a hybrid iterative reconstruction algorithm on image quality. J Comput Assist Tomogr 37:790–796CrossRef
18.
Zurück zum Zitat Patel BN, Thomas JV, Lockhart ME, Berland LL, Morgan DE (2013) Single-source dual-energy spectral multidetector CT of pancreatic adenocarcinoma: optimization of energy level viewing significantly increases lesion contrast. Clin Radiol 68:148–154CrossRef Patel BN, Thomas JV, Lockhart ME, Berland LL, Morgan DE (2013) Single-source dual-energy spectral multidetector CT of pancreatic adenocarcinoma: optimization of energy level viewing significantly increases lesion contrast. Clin Radiol 68:148–154CrossRef
19.
Zurück zum Zitat Frellesen C, Fessler F, Hardie AD et al (2015) Dual-energy CT of the pancreas: improved carcinoma-to-pancreas contrast with a noise-optimized monoenergetic reconstruction algorithm. Eur J Radiol 84:2052–2058CrossRef Frellesen C, Fessler F, Hardie AD et al (2015) Dual-energy CT of the pancreas: improved carcinoma-to-pancreas contrast with a noise-optimized monoenergetic reconstruction algorithm. Eur J Radiol 84:2052–2058CrossRef
20.
Zurück zum Zitat Hardie AD, Picard MM, Camp ER et al (2015) Application of an advanced image-based virtual Monoenergetic reconstruction of dual source dual-energy CT data at low keV increases image quality for routine pancreas imaging. J Comput Assist Tomogr 39:716–720CrossRef Hardie AD, Picard MM, Camp ER et al (2015) Application of an advanced image-based virtual Monoenergetic reconstruction of dual source dual-energy CT data at low keV increases image quality for routine pancreas imaging. J Comput Assist Tomogr 39:716–720CrossRef
21.
Zurück zum Zitat McNamara MM, Little MD, Alexander LF, Carroll LV, Beasley TM, Morgan DE (2015) Multireader evaluation of lesion conspicuity in small pancreatic adenocarcinomas: complimentary value of iodine material density and low keV simulated monoenergetic images using multiphasic rapid kVp-switching dual energy CT. Abdom Imaging 40:1230–1240CrossRef McNamara MM, Little MD, Alexander LF, Carroll LV, Beasley TM, Morgan DE (2015) Multireader evaluation of lesion conspicuity in small pancreatic adenocarcinomas: complimentary value of iodine material density and low keV simulated monoenergetic images using multiphasic rapid kVp-switching dual energy CT. Abdom Imaging 40:1230–1240CrossRef
22.
Zurück zum Zitat Bellini D, Gupta S, Ramirez-Giraldo JC et al (2017) Use of a noise optimized monoenergetic algorithm for patient-size independent selection of an optimal energy level during dual-energy CT of the pancreas. J Comput Assist Tomogr 41:39–47CrossRef Bellini D, Gupta S, Ramirez-Giraldo JC et al (2017) Use of a noise optimized monoenergetic algorithm for patient-size independent selection of an optimal energy level during dual-energy CT of the pancreas. J Comput Assist Tomogr 41:39–47CrossRef
23.
Zurück zum Zitat Bhosale P, Le O, Balachandran A, Fox P, Paulson E, Tamm E (2015) Quantitative and qualitative comparison of single-source dual-energy computed tomography and 120-kVp computed tomography for the assessment of pancreatic ductal adenocarcinoma. J Comput Assist Tomogr 39:907–913CrossRef Bhosale P, Le O, Balachandran A, Fox P, Paulson E, Tamm E (2015) Quantitative and qualitative comparison of single-source dual-energy computed tomography and 120-kVp computed tomography for the assessment of pancreatic ductal adenocarcinoma. J Comput Assist Tomogr 39:907–913CrossRef
25.
Zurück zum Zitat Kalisz K, Rassouli N, Dhanantwari A, Jordan D, Rajiah P (2018) Noise characteristics of virtual monoenergetic images from a novel detector-based spectral CT scanner. Eur J Radiol 98:118–125CrossRef Kalisz K, Rassouli N, Dhanantwari A, Jordan D, Rajiah P (2018) Noise characteristics of virtual monoenergetic images from a novel detector-based spectral CT scanner. Eur J Radiol 98:118–125CrossRef
26.
Zurück zum Zitat Nagayama Y, Nakaura T, Oda S et al (2018) Dual-layer DECT for multiphasic hepatic CT with 50 percent iodine load: a matched-pair comparison with a 120 kVp protocol. Eur Radiol 28:1719–1730CrossRef Nagayama Y, Nakaura T, Oda S et al (2018) Dual-layer DECT for multiphasic hepatic CT with 50 percent iodine load: a matched-pair comparison with a 120 kVp protocol. Eur Radiol 28:1719–1730CrossRef
27.
Zurück zum Zitat Sellerer T, Noel PB, Patino M et al (2018) Dual-energy CT: a phantom comparison of different platforms for abdominal imaging. Eur Radiol 28:2745–2755CrossRef Sellerer T, Noel PB, Patino M et al (2018) Dual-energy CT: a phantom comparison of different platforms for abdominal imaging. Eur Radiol 28:2745–2755CrossRef
28.
Zurück zum Zitat Christner JA, Braun NN, Jacobsen MC, Carter RE, Kofler JM, McCollough CH (2012) Size-specific dose estimates for adult patients at CT of the torso. Radiology 265:841–847CrossRef Christner JA, Braun NN, Jacobsen MC, Carter RE, Kofler JM, McCollough CH (2012) Size-specific dose estimates for adult patients at CT of the torso. Radiology 265:841–847CrossRef
29.
Zurück zum Zitat Hickethier T, Iuga AI, Lennartz S et al (2018) Virtual monoenergetic images from a novel dual-layer spectral detector computed tomography scanner in portal venous phase: adjusted window settings depending on assessment focus are essential for image interpretation. J Comput Assist Tomogr 42:350–356CrossRef Hickethier T, Iuga AI, Lennartz S et al (2018) Virtual monoenergetic images from a novel dual-layer spectral detector computed tomography scanner in portal venous phase: adjusted window settings depending on assessment focus are essential for image interpretation. J Comput Assist Tomogr 42:350–356CrossRef
30.
Zurück zum Zitat Almeida RR, Lo GC, Patino M, Bizzo B, Canellas R, Sahani DV (2018) Advances in pancreatic CT imaging. AJR Am J Roentgenol 211:52–66CrossRef Almeida RR, Lo GC, Patino M, Bizzo B, Canellas R, Sahani DV (2018) Advances in pancreatic CT imaging. AJR Am J Roentgenol 211:52–66CrossRef
31.
Zurück zum Zitat Sakabe D, Funama Y, Taguchi K et al (2018) Image quality characteristics for virtual monoenergetic images using dual-layer spectral detector CT: comparison with conventional tube-voltage images. Phys Med 49:5–10CrossRef Sakabe D, Funama Y, Taguchi K et al (2018) Image quality characteristics for virtual monoenergetic images using dual-layer spectral detector CT: comparison with conventional tube-voltage images. Phys Med 49:5–10CrossRef
32.
Zurück zum Zitat Nagayama Y, Iyama A, Oda S et al (2019) Dual-layer dual-energy computed tomography for the assessment of hypovascular hepatic metastases: impact of closing k-edge on image quality and lesion detectability. Eur Radiol 29:2837–2847CrossRef Nagayama Y, Iyama A, Oda S et al (2019) Dual-layer dual-energy computed tomography for the assessment of hypovascular hepatic metastases: impact of closing k-edge on image quality and lesion detectability. Eur Radiol 29:2837–2847CrossRef
33.
Zurück zum Zitat Grosse Hokamp N, Obmann VC, Kessner R et al (2018) Improved visualization of hypodense liver lesions in virtual monoenergetic images from spectral detector CT: proof of concept in a 3D-printed phantom and evaluation in 74 patients. Eur J Radiol 109:114–123CrossRef Grosse Hokamp N, Obmann VC, Kessner R et al (2018) Improved visualization of hypodense liver lesions in virtual monoenergetic images from spectral detector CT: proof of concept in a 3D-printed phantom and evaluation in 74 patients. Eur J Radiol 109:114–123CrossRef
34.
Zurück zum Zitat Nagayama Y, Nakaura T, Oda S et al (2018) Dual-layer detector CT of chest, abdomen, and pelvis with a one-third iodine dose: image quality, radiation dose, and optimal monoenergetic settings. Clin Radiol 73:1058.e1021–1058.e1029 Nagayama Y, Nakaura T, Oda S et al (2018) Dual-layer detector CT of chest, abdomen, and pelvis with a one-third iodine dose: image quality, radiation dose, and optimal monoenergetic settings. Clin Radiol 73:1058.e1021–1058.e1029
35.
Zurück zum Zitat Brook OR, Gourtsoyianni S, Brook A, Siewert B, Kent T, Raptopoulos V (2013) Split-bolus spectral multidetector CT of the pancreas: assessment of radiation dose and tumor conspicuity. Radiology 269:139–148CrossRef Brook OR, Gourtsoyianni S, Brook A, Siewert B, Kent T, Raptopoulos V (2013) Split-bolus spectral multidetector CT of the pancreas: assessment of radiation dose and tumor conspicuity. Radiology 269:139–148CrossRef
36.
Zurück zum Zitat Hickethier T, Byrtus J, Hauger M et al (2018) Utilization of virtual mono-energetic images (MonoE) derived from a dual-layer spectral detector CT (SDCT) for the assessment of abdominal arteries in venous contrast phase scans. Eur J Radiol 99:28–33CrossRef Hickethier T, Byrtus J, Hauger M et al (2018) Utilization of virtual mono-energetic images (MonoE) derived from a dual-layer spectral detector CT (SDCT) for the assessment of abdominal arteries in venous contrast phase scans. Eur J Radiol 99:28–33CrossRef
37.
Zurück zum Zitat Lohofer FK, Kaissis GA, Koster FL et al (2018) Improved detection rates and treatment planning of head and neck cancer using dual-layer spectral CT. Eur Radiol 28:4925–4931CrossRef Lohofer FK, Kaissis GA, Koster FL et al (2018) Improved detection rates and treatment planning of head and neck cancer using dual-layer spectral CT. Eur Radiol 28:4925–4931CrossRef
38.
Zurück zum Zitat Zopfs D, Lennartz S, Laukamp K et al (2018) Improved depiction of atherosclerotic carotid artery stenosis in virtual monoenergetic reconstructions of venous phase dual-layer computed tomography in comparison to polyenergetic reconstructions. Eur J Radiol 100:36–42CrossRef Zopfs D, Lennartz S, Laukamp K et al (2018) Improved depiction of atherosclerotic carotid artery stenosis in virtual monoenergetic reconstructions of venous phase dual-layer computed tomography in comparison to polyenergetic reconstructions. Eur J Radiol 100:36–42CrossRef
39.
Zurück zum Zitat Fletcher JG, Wiersema MJ, Farrell MA et al (2003) Pancreatic malignancy: value of arterial, pancreatic, and hepatic phase imaging with multi-detector row CT. Radiology 229:81–90CrossRef Fletcher JG, Wiersema MJ, Farrell MA et al (2003) Pancreatic malignancy: value of arterial, pancreatic, and hepatic phase imaging with multi-detector row CT. Radiology 229:81–90CrossRef
Metadaten
Titel
Dual-layer spectral CT improves image quality of multiphasic pancreas CT in patients with pancreatic ductal adenocarcinoma
verfasst von
Yasunori Nagayama
Shota Tanoue
Taihei Inoue
Seitaro Oda
Takeshi Nakaura
Daisuke Utsunomiya
Yasuyuki Yamashita
Publikationsdatum
16.07.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 1/2020
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06337-y

Weitere Artikel der Ausgabe 1/2020

European Radiology 1/2020 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.