Effective dose estimation for pediatric upper gastrointestinal examinations using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology

verfasst von: Brent Emigh, Christopher L. Gordon, Bairbre L. Connolly, Michelle Falkiner, Karen E. Thomas

Erschienen in: Pediatric Radiology | Ausgabe 9/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

There is a need for updated radiation dose estimates in pediatric fluoroscopy given the routine use of new dose-saving technologies and increased radiation safety awareness in pediatric imaging.

Objective

To estimate effective doses for standardized pediatric upper gastrointestinal (UGI) examinations at our institute using direct dose measurement, as well as provide dose-area product (DAP) to effective dose conversion factors to be used for the estimation of UGI effective doses for boys and girls up to 10 years of age at other centers.

Materials and methods

Metal oxide semiconductor field-effect transistor (MOSFET) dosimeters were placed within four anthropomorphic phantoms representing children ≤10 years of age and exposed to mock UGI examinations using exposures much greater than used clinically to minimize measurement error. Measured effective dose was calculated using ICRP 103 weights and scaled to our institution’s standardized clinical UGI (3.6-min fluoroscopy, four spot exposures and four examination beam projections) as determined from patient logs. Results were compared to Monte Carlo simulations and related to fluoroscope-displayed DAP.

Results

Measured effective doses for standardized pediatric UGI examinations in our institute ranged from 0.35 to 0.79 mSv in girls and were 3–8% lower for boys. Simulation-derived and measured effective doses were in agreement (percentage differences <19%, T > 0.18). DAP-to-effective dose conversion factors ranged from 6.5 ×10−4 mSv per Gy-cm2 to 4.3 × 10−3 mSv per Gy-cm2 for girls and were similarly lower for boys.

Conclusion

Using modern fluoroscopy equipment, the effective dose associated with the UGI examination in children ≤10 years at our institute is < 1 mSv. Estimations of effective dose associated with pediatric UGI examinations can be made for children up to the age of 10 using the DAP-normalized conversion factors provided in this study. These estimates can be further refined to reflect individual hospital examination protocols through the use of direct organ dose measurement using MOSFETs, which were shown to agree with Monte Carlo simulated doses.
Literatur
1.
Zurück zum Zitat Council NR (2006) Health risks of exposure to low level of ionizing radiation. BEIR VII. National Academic Press, Washington, DC Council NR (2006) Health risks of exposure to low level of ionizing radiation. BEIR VII. National Academic Press, Washington, DC
2.
Zurück zum Zitat International Commission on Radiological Protection (ICRP) (2007) The 2007 Recommendations of the ICRP, Annals of the ICRP. Publication 103. Elsevier, Amsterdam International Commission on Radiological Protection (ICRP) (2007) The 2007 Recommendations of the ICRP, Annals of the ICRP. Publication 103. Elsevier, Amsterdam
3.
Zurück zum Zitat Geleijns J, Broerse JJ, Chandie Shaw MP et al (1998) A comparison of patient dose for examinations of the upper gastrointestinal tract at 11 conventional and digital X-ray units in The Netherlands. Br J Radiol 71:745–753PubMed Geleijns J, Broerse JJ, Chandie Shaw MP et al (1998) A comparison of patient dose for examinations of the upper gastrointestinal tract at 11 conventional and digital X-ray units in The Netherlands. Br J Radiol 71:745–753PubMed
4.
Zurück zum Zitat Hart D, Wall BF (1994) Estimation of effective dose from dose-area product measurements for barium meals and barium enemas. Br J Radiol 67:485–489PubMedCrossRef Hart D, Wall BF (1994) Estimation of effective dose from dose-area product measurements for barium meals and barium enemas. Br J Radiol 67:485–489PubMedCrossRef
5.
Zurück zum Zitat Ruiz-Cruces R, Ruiz F, Perez-Martinez M et al (2000) Patient dose from barium procedures. Br J Radiol 73:752–761PubMed Ruiz-Cruces R, Ruiz F, Perez-Martinez M et al (2000) Patient dose from barium procedures. Br J Radiol 73:752–761PubMed
6.
Zurück zum Zitat Dimitriadis A, Gialousis G, Makri T et al (2011) Monte Carlo estimation of radiation doses during paediatric barium meal and cystourethrography examinations. Phys Med Biol 56:367–382PubMedCrossRef Dimitriadis A, Gialousis G, Makri T et al (2011) Monte Carlo estimation of radiation doses during paediatric barium meal and cystourethrography examinations. Phys Med Biol 56:367–382PubMedCrossRef
7.
Zurück zum Zitat Strauss KJ (2006) Pediatric interventional radiography equipment: safety considerations. Pediatr Radiol 36:126–135PubMedCrossRef Strauss KJ (2006) Pediatric interventional radiography equipment: safety considerations. Pediatr Radiol 36:126–135PubMedCrossRef
8.
Zurück zum Zitat Hiorns MP, Saini A, Marsden PJ (2006) A review of current local dose-area product levels for paediatric fluoroscopy in a tertiary referral centre compared with national standards. Why are they so different? Br J Radiol 79:326–330PubMedCrossRef Hiorns MP, Saini A, Marsden PJ (2006) A review of current local dose-area product levels for paediatric fluoroscopy in a tertiary referral centre compared with national standards. Why are they so different? Br J Radiol 79:326–330PubMedCrossRef
9.
Zurück zum Zitat Peet D, Pryor MD (1999) Evaluation of a MOSFET radiation sensor for the measurement of entrance surface dose in diagnostic radiology. Br J Radiol 72:562–568PubMed Peet D, Pryor MD (1999) Evaluation of a MOSFET radiation sensor for the measurement of entrance surface dose in diagnostic radiology. Br J Radiol 72:562–568PubMed
10.
Zurück zum Zitat Yoshizumi T, Goodman PC, Frush DP et al (2007) Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: comparison with thermoluminescent dosimetry. AJR Am J Roentgenol 188:1332–1336PubMedCrossRef Yoshizumi T, Goodman PC, Frush DP et al (2007) Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: comparison with thermoluminescent dosimetry. AJR Am J Roentgenol 188:1332–1336PubMedCrossRef
11.
Zurück zum Zitat Glennie D, Connolly BL, Gordon C (2008) Entrance skin dose measured with MOSFETs in children undergoing interventional radiology procedures. Pediatr Radiol 38:1180–1187PubMedCrossRef Glennie D, Connolly BL, Gordon C (2008) Entrance skin dose measured with MOSFETs in children undergoing interventional radiology procedures. Pediatr Radiol 38:1180–1187PubMedCrossRef
12.
Zurück zum Zitat Lee R, Thomas K, Connolly B et al (2009) Effective dose estimation for pediatric voiding cystourethrography using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology. Pediatr Radiol 39:608–615PubMedCrossRef Lee R, Thomas K, Connolly B et al (2009) Effective dose estimation for pediatric voiding cystourethrography using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology. Pediatr Radiol 39:608–615PubMedCrossRef
13.
Zurück zum Zitat Miksys N, Gordon CL, Thomas K et al (2010) Estimating effective dose to pediatric patients undergoing interventional radiology procedures using anthropomorphic phantoms and MOSFET dosimeters. AJR Am J Roentgenol 194:1315–1322PubMedCrossRef Miksys N, Gordon CL, Thomas K et al (2010) Estimating effective dose to pediatric patients undergoing interventional radiology procedures using anthropomorphic phantoms and MOSFET dosimeters. AJR Am J Roentgenol 194:1315–1322PubMedCrossRef
14.
Zurück zum Zitat Gaca A, Jaffe TA, Delaney S et al (2008) Radiation doses from small-bowel follow-through and abdomen/pelvis MDCT in pediatric Crohn disease. Pediatr Radiol 38:285–291PubMedCrossRef Gaca A, Jaffe TA, Delaney S et al (2008) Radiation doses from small-bowel follow-through and abdomen/pelvis MDCT in pediatric Crohn disease. Pediatr Radiol 38:285–291PubMedCrossRef
15.
Zurück zum Zitat Hollingsworth C, Yoshizumi TT, Frush DP et al (2007) Pediatric cardiac-gated CT angiography: assessment of radiation dose. AJR Am J Roentgenol 189:12–18PubMedCrossRef Hollingsworth C, Yoshizumi TT, Frush DP et al (2007) Pediatric cardiac-gated CT angiography: assessment of radiation dose. AJR Am J Roentgenol 189:12–18PubMedCrossRef
17.
Zurück zum Zitat Cygler J, Saoudi A, Perry G et al (2006) Feasibility study of using MOSFET detector for in vivo dosimetry during permanent low-dose-rate prostate implants. Radiother Oncol 80:296–301PubMedCrossRef Cygler J, Saoudi A, Perry G et al (2006) Feasibility study of using MOSFET detector for in vivo dosimetry during permanent low-dose-rate prostate implants. Radiother Oncol 80:296–301PubMedCrossRef
18.
Zurück zum Zitat Dong SL, Chu TC, Lan GY et al (2002) Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF:Mg, Cu, P thermoluminescence dosimeters for use in diagnostic radiology. Appl Radiat Isotopes 57:883–891CrossRef Dong SL, Chu TC, Lan GY et al (2002) Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF:Mg, Cu, P thermoluminescence dosimeters for use in diagnostic radiology. Appl Radiat Isotopes 57:883–891CrossRef
19.
Zurück zum Zitat Sessions J, Roshau JN, Tressler MA et al (2002) Comparisons of point and average organ dose within an anthropomorphic physical phantom and a computational model of the newborn patient. Med Phys 29:1080–1089PubMedCrossRef Sessions J, Roshau JN, Tressler MA et al (2002) Comparisons of point and average organ dose within an anthropomorphic physical phantom and a computational model of the newborn patient. Med Phys 29:1080–1089PubMedCrossRef
20.
Zurück zum Zitat Jones A, Pazik FD, Hintenlang DE et al (2005) MOSFET dosimeter depth-dose measurements in heterogeneous tissue-equivalent phantoms at diagnostic X-ray energies. Med Phys 32:3209–3213PubMedCrossRef Jones A, Pazik FD, Hintenlang DE et al (2005) MOSFET dosimeter depth-dose measurements in heterogeneous tissue-equivalent phantoms at diagnostic X-ray energies. Med Phys 32:3209–3213PubMedCrossRef
21.
Zurück zum Zitat Cristy M (1981) Active bone marrow distribution as a function of age in humans. Phys Med Biol 26:389–400PubMedCrossRef Cristy M (1981) Active bone marrow distribution as a function of age in humans. Phys Med Biol 26:389–400PubMedCrossRef
22.
Zurück zum Zitat Cristy M (1980) Mathematical phantoms representing children of various ages for use in estimating internal dose. In: Commission UNR (ed), Oak Ridge, TN Cristy M (1980) Mathematical phantoms representing children of various ages for use in estimating internal dose. In: Commission UNR (ed), Oak Ridge, TN
23.
Zurück zum Zitat Ward V, Strauss KJ, Barnewolt CE et al (2008) Pediatric radiation exposure and effective dose reduction during voiding cystourethrography. Radiology 249:1002–1009PubMedCrossRef Ward V, Strauss KJ, Barnewolt CE et al (2008) Pediatric radiation exposure and effective dose reduction during voiding cystourethrography. Radiology 249:1002–1009PubMedCrossRef
24.
Zurück zum Zitat Newman B, John S, Goske M et al (2011) Pause and pulse: radiation dose in pediatric fluoroscopy. Pediatr Rev 32:e83–e90PubMedCrossRef Newman B, John S, Goske M et al (2011) Pause and pulse: radiation dose in pediatric fluoroscopy. Pediatr Rev 32:e83–e90PubMedCrossRef
25.
Zurück zum Zitat Staton R, Williams JL, Arreola MM et al (2007) Organ and effective doses in infants undergoing upper gastrointestinal (UGI) fluoroscopic examination. Med Phys 34:703–710PubMedCrossRef Staton R, Williams JL, Arreola MM et al (2007) Organ and effective doses in infants undergoing upper gastrointestinal (UGI) fluoroscopic examination. Med Phys 34:703–710PubMedCrossRef
26.
Zurück zum Zitat Damilakis J, Stratakis J, Raissaki M et al (2006) Normalized dose data for upper gastrointestinal tract contrast studies performed to infants. Med Phys 33:1033–1040PubMedCrossRef Damilakis J, Stratakis J, Raissaki M et al (2006) Normalized dose data for upper gastrointestinal tract contrast studies performed to infants. Med Phys 33:1033–1040PubMedCrossRef
27.
Zurück zum Zitat Board NRP (2002) Doses to patients from medical X-ray examinations in the UK—2000 review. National Radiological Protection Board, Chilton Board NRP (2002) Doses to patients from medical X-ray examinations in the UK—2000 review. National Radiological Protection Board, Chilton
28.
Zurück zum Zitat Delichas M, Hatziionannou K, Papanastassiou E et al (2004) Radiation doses to patients undergoing barium meal and barium enema examinations. Radiat Prot Dosim 109:243–247CrossRef Delichas M, Hatziionannou K, Papanastassiou E et al (2004) Radiation doses to patients undergoing barium meal and barium enema examinations. Radiat Prot Dosim 109:243–247CrossRef
29.
Zurück zum Zitat Hall E (2002) Lessons we have learned from our children: cancer risks from diagnostic radiology. Pediatr Radiol 32:700–706PubMedCrossRef Hall E (2002) Lessons we have learned from our children: cancer risks from diagnostic radiology. Pediatr Radiol 32:700–706PubMedCrossRef
30.
Zurück zum Zitat Perisinakis K, Raissaki M, Damilakis J et al (2006) Fluoroscopy-controlled voiding cystourethrography in infants and children: are the radiation risks trivial? Eur Radiol 16:846–851PubMedCrossRef Perisinakis K, Raissaki M, Damilakis J et al (2006) Fluoroscopy-controlled voiding cystourethrography in infants and children: are the radiation risks trivial? Eur Radiol 16:846–851PubMedCrossRef
31.
Zurück zum Zitat Rosenstein M, Suleiman OH, Burkhart RL et al (1992) Handbook of selected tissue doses for the upper gastrointestinal fluoroscopic examination. In: Health CfDaR (ed), Rockville, MD Rosenstein M, Suleiman OH, Burkhart RL et al (1992) Handbook of selected tissue doses for the upper gastrointestinal fluoroscopic examination. In: Health CfDaR (ed), Rockville, MD
Metadaten
Titel
Effective dose estimation for pediatric upper gastrointestinal examinations using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology
verfasst von
Brent Emigh
Christopher L. Gordon
Bairbre L. Connolly
Michelle Falkiner
Karen E. Thomas
Publikationsdatum
01.09.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 9/2013
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-013-2674-5

Weitere Artikel der Ausgabe 9/2013

Pediatric Radiology 9/2013 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.