Electrical impedance tomography during spontaneous breathing trials and after extubation in critically ill patients at high risk for extubation failure: a multicenter observational study
This study aims to assess the changes in lung aeration and ventilation during the first spontaneous breathing trial (SBT) and after extubation in a population of patients at risk of extubation failure.
Methods
We included 78 invasively ventilated patients eligible for their first SBT, conducted with low positive end-expiratory pressure (2 cm H2O) for 30 min. We acquired three 5-min electrical impedance tomography (EIT) records at baseline, soon after the beginning (SBT_0) and at the end (SBT_30) of SBT. In the case of SBT failure, ventilation was reinstituted; otherwise, the patient was extubated and two additional records were acquired soon after extubation (SB_0) and 30 min later (SB_30) during spontaneous breathing. Extubation failure was defined by the onset of post-extubation respiratory failure within 48 h after extubation. We computed the changes from baseline of end-expiratory lung impedance (∆EELI), tidal volume (∆Vt%), and the inhomogeneity index. Arterial blood was sampled for gas analysis. Data were compared between sub-groups stratified for SBT and extubation success/failure.
Results
Compared to SBT success (n = 61), SBT failure (n = 17) showed a greater reduction in ∆EELI at SBT_0 (p < 0.001) and SBT_30 (p = 0.001) and a higher inhomogeneity index at baseline (p = 0.002), SBT_0 (p = 0.003) and SBT_30 (p = 0.005). RR/Vt was not different between groups at baseline but was significantly greater at SBT_0 and SBT_30 in SBT failures, compared to SBT successes (p < 0.001 for both). No differences in ∆Vt% and arterial blood gases were observed between SBT success and failure. The ∆Vt%, ∆EELI, inhomogeneity index and arterial blood gases were not different between patients with extubation success (n = 39) and failure (n = 22) (p > 0.05 for all comparisons).
Conclusions
Compared to SBT success, SBT failure was characterized by more lung de-recruitment and inhomogeneity. Whether EIT may be useful to monitor SBT remains to be determined. No significant changes in lung ventilation, aeration or homogeneity related to extubation outcome occurred up to 30 min after extubation.
Trial registration Retrospectively registered on clinicaltrials.gov (Identifier: NCT03894332; release date 27th March 2019).
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ABGs
arterial blood gases
CPAP
continuous positive airway pressure
∆EELI
changes in milliliters from baseline of the end-expiratory lung impedance
∆Vt%
variation in percentage from baseline of the tidal volume
EELI
end-expiratory lung impedance
EIT
electrical impedance tomography
ESM
electronic supplementary materials
FiO2
inspired fraction of oxygen
ICU
intensive care unit
iMV
invasive mechanical ventilation
NIV
non-invasive ventilation
PaCO2
arterial partial pressure of carbon dioxide
PaO2
arterial partial pressure of oxygen
PEEP
positive end-expiratory pressure
PSV
pressure support ventilation
RR
respiratory rate
SB_0
first 5 min of the spontaneous breathing
SB_30
last 5 min of the spontaneous breathing
SBT_0
first 5 min after SBT initiation
SBT_30
last 5 min of the SBT
SBT
spontaneous breathing trial
SpO2
peripheral oxygen saturation
Vt
tidal volume
Background
Weaning is the whole process that leads patients to the discontinuation of mechanical ventilation and the removal of the endotracheal tube [1]. Weaning should be considered as early as possible to reduce the time spent on invasive mechanical ventilation (iMV), which is associated with morbidity and mortality [2, 3]. To assess whether patients are eligible for extubation, a spontaneous breathing trial (SBT) is often performed. The outcome of SBT is evaluated by clinical assessment and acquisition of objective parameters, such as the arterial blood gases (ABGs) and the respiratory rate (RR) to tidal volume (Vt) ratio (RR/Vt) [1]. In the case of SBT success, the patient can be extubated. However, post-extubation respiratory failure may occur, in particular in patients at increased risk [1], who may benefit from prophylactic non-invasive ventilation (NIV) application immediately after extubation [4‐9].
Electrical impedance tomography (EIT) is a non-invasive imaging technique that provides instantaneous monitoring of variations in overall lung volume and regional distribution of ventilation, as determined by variations over time in intrathoracic impedance, which is increased by air and reduced by fluids and cells. EIT allows for determining changes in end-expiratory lung impedance (EELI), a surrogate estimate of end-expiratory lung volume, assessing global and regional distribution of Vt, and obtaining indexes of the spatial distribution of ventilation, such as the global inhomogeneity index [10].
Anzeige
Because SBT failure and extubation failure are both major clinical problems, they have been repeatedly investigated, but the underlying mechanisms are not fully elucidated yet [1]. We reasoned that EIT could help comprehend the pathophysiology of SBT failure and extubation failure by providing additional insights. We, therefore, designed this observational study on patients at high risk of extubation failure, aimed at assessing variations in EELI (∆EELI) and Vt (∆Vt%), and the inhomogeneity index during SBT and after extubation.
Methods
The study was conducted from June 2015 to June 2016 in the intensive care units (ICUs) of three Italian hospitals (“Sant’Andrea” Hospital in Vercelli, “ARNAS Garibaldi” Hospital in Catania and “Infermi” Hospital in Rimini), in accordance with the principles outlined in the Declaration of Helsinki, after approval by the local Ethics Committees (Vercelli, protocol no AslVC.Rian.14.02, approved on 11/09/2014; Catania, protocol no 479/CE, approved on 30/06/2015; Rimini, protocol no 1313, approved on 18/03/2015). Written informed consent was obtained from all patients according to national regulations.
Patients
We included any ICU patient ≥ 18 years who (1) had received iMV for at least 48 h through an orotracheal tube, (2) was ready for the first SBT attempt [1] and (3) met at least one criterion for increased risk of extubation failure, as outlined in the first column of Additional file 1: Table S1 [4‐9, 11]. Patients were considered to be ready for SBT if they satisfied all of the criteria reported in the second column of Additional file 1: Table S1 [1, 4, 12]. Patients were excluded if they met one or more of the following criteria: (1) life-threatening heart arrhythmias or cardiac ischaemia, (2) pneumothorax, (3) pulmonary emphysema, (4) recent (1 week) thoracic surgery, (5) presence of chest burns, (6) pregnancy, (7) inclusion in other research protocols, (8) refusal of consent.
Inclusion and exclusion criteria were assessed daily in all patients.
Anzeige
Weaning protocol
Once able to trigger the ventilator, patients were ventilated with pressure support ventilation (PSV). Inspired fraction of oxygen (FiO2) and positive end-expiratory pressure (PEEP) were set to maintain peripheral oxygen saturation (SpO2) between 92 and 96%, while the inspiratory pressure support was titrated to generate a Vt of 6–8 ml/kg of ideal body weight with an active inspiration [13]. The SBT was conducted by setting the ventilator with 2 cm H2O in continuous positive airway pressure (CPAP) mode, with no inspiratory support for 30 min [12, 14]. SBT failure was defined by the presence of at least one of the criteria listed in the third column of Additional file 1: Table S1 [1, 15]. Patients who passed the SBT were immediately extubated and allowed to breathe spontaneously with a Venturi mask, while those who failed SBT were switched back to PSV with the same settings applied prior to SBT.
Predefined criteria for protocol interruption were the following: (1) haemodynamic instability (i.e., need for continuous infusion of dopamine or dobutamine > 5 mcg/kg/min, norepinephrine > 0.1 mcg/kg/min, or vasopressin to maintain mean arterial blood pressure > 60 mmHg), (2) life-threatening arrhythmias or electrocardiographic signs of ischemia, or (3) worsening of ABGs (i.e., onset of respiratory acidosis, as defined by arterial partial pressure of carbon dioxide (PaCO2) > 50 mmHg and pH < 7.30, or hypoxemia, as defined by arterial partial pressure of oxygen (PaO2) < 60 mmHg with a FiO2 ≥ 50%).
Data acquisition and analysis
After enrolment, a silicon 16-electrode EIT belt of proper size was placed around the patient’s chest between the 4th and 6th intercostal spaces and connected to the EIT device (PulmoVista 500; Draeger Medical GmbH, Lübeck, Germany) [16, 17]. Signal quality was determined through dedicated built-in software. The position of EIT belt was, therefore, marked on the skin with a dermographic pencil to avoid its displacement during the study period. PSV was applied with a ventilator (Infinity V500; Draeger Medical GmbH, Lübeck, Germany) connected to the EIT device through a RS232 interface.
We acquired 5-min EIT data records at baseline (during PSV), during the first (SBT_0) and the last (SBT_30) 5 min of SBT, and when the patient was extubated during spontaneous breathing immediately (SB_0) and 30 min (SB_30) after extubation.
EIT and ventilator (i.e., airway pressure, flow and volume) data were recorded at a sample of 20 Hz, downloaded and analysed off-line on a personal computer using dedicated software (EITdiag, Draeger Medical GmbH, Lübeck, Germany). Impedance changes were calibrated against the data acquired from the ventilator by the EIT device as previously described [18]. The last 3 min of each record were analysed.
From the EIT recordings, we measured Vt, respiratory rate (RR), air distribution within the lungs, and end-expiratory lung impedance (EELI), i.e., a surrogate of the end-expiratory lung volume [19]. From EIT, we also calculated (1) Vt changes from baseline, expressed as a percentage (∆Vt%), as assessed by the EIT, (2) EELI variation (∆EELI) from baseline, expressed in ml, (3) the inhomogeneity index, which describes uneven air distribution within the lung [20], and (4) the RR/Vt at baseline, SBT_0, SBT_30, SB_0 and SB_30 [15], as derived from EIT measurements.
where DI is the differential impedance, DIxy refers to the DI of pixels in a defined lung region and DIlung represents the DI of the whole lung [20].
ABGs were obtained at baseline, SBT_30 and SB_30.
Anzeige
Extubation failure was defined by the occurrence of post-extubation respiratory failure, as characterized by one or more of the criteria outlined in the fourth column of Additional file 1: Table S1, in the 48 h following extubation [1, 21]. Patients who failed extubation underwent ‘‘rescue’’ CPAP by means of a high-flow (> 45 l/min) generator (Flow-Meter, StarMed, Mirandola, Modena, Italy) in the case of normocapnic hypoxemia (i.e., PaO2/FiO2 < 200 mmHg), or NIV, as delivered through NIV modality, in the case of respiratory acidosis (pH < 7.35 and PaCO2 > 50 mmHg). Patients were re-intubated and iMV reinstituted if one or more of the criteria displayed in the fifth column of Additional file 1: Table S1 occurred [21].
Statistical analysis
Due to the lack of published data at the time of the study design, no power analysis could be performed. Therefore, we arbitrarily planned to enrol a sample of 80 consecutive patients. Because of the lack of normal distribution of data, as assessed by the Shapiro–Wilk test, the data were expressed as the median [25th–75th interquartile range]. Data were compared between subgroups of patients (i.e., SBT success versus failure; extubation success versus failure) using the Mann–Whitney U-test. Thresholds for statistical significance were adjusted with Bonferroni correction for multiple comparisons. Categorical data were compared by Fisher’s exact test; for this comparison, a two-sided p < 0.05 was considered significant.
Receiving operating curves (ROC) were obtained for RR/Vt and PaO2/FiO2 and EIT data. Cut-off values were obtained calculating the Youden index. The area under the curve (AUC), sensitivity, specificity, positive (LR +) and negative (LR −) likelihood ratios were also calculated.
Results
We enrolled 80 consecutive patients, as planned. Two patients, erroneously enrolled with no risk criteria for post-extubation respiratory failure, were excluded from the data analysis. The patients’ flow of study enrolment is shown in Additional file 1: Figure S1, while the protocol flow is depicted in Fig. 1. None of the patients interrupted the study protocol. Table 1 shows the patients’ characteristics at ICU admission and immediately before SBT, and the reasons for ICU admission and risk factors for post-extubation respiratory failure, for the overall population and separately for the SBT and extubation success and failure subgroups.
Table 1
Patient’s characteristics at ICU admission, risk factors for post-extubation respiratory failure, the reason for ICU admission and characteristics before SBT performance in the overall population and subgroups
Overall
SBT success
(n = 61)
SBT failure
(n = 17)
p value
Extubation success (n = 39)
Extubation failure (n = 22)
p value
Characteristics at ICU admission
Weight (kg)
75 [65; 83]
75 [65; 81]
75 [68; 95]
0.458
70 [62; 80]
80 [70; 86]
0.128
Height (cm)
170 [165; 173]
170 [165; 175]
170 [163; 170]
0.286
170 [165; 175]
170 [160; 172]
0.203
SAPS-II
52 [37; 67]
50 [34; 66]
59 [41; 68]
0.188
44 [31; 55]
63 [48; 83]
< 0.001
SOFA
6 [4; 9]
6 [4; 9]
4 [3; 7]
0.049
6 [4; 9]
7 [4; 10]
0.790
Risk factors for post-extubation respiratory failure
PaCO2 > 45 mmHg at SBT n (%)
30 (38)
22 (36)
8 (47)
0.416
12 (31)
10 (45)
0.279
Chronic respiratory disease n (%)
17 (22)
12 (20)
5 (29)
0.510
7 (18)
5 (23)
0.741
Chronic heart failure n (%)
17 (22)
12 (20)
5 (29)
0.510
9 (23)
3 (14)
0.510
Upper airway stridor n (%)
2 (3)
2 (3)
0 (0)
> 0.999
2 (5)
0 (0)
0.531
Age ≥ 65 years n (%)
48 (62)
38 (62)
10 (59)
0.786
24 (62)
14 (64)
> 0.999
Cardiac failure as reason of iMV n (%)
10 (13)
8 (13)
2 (12)
> 0.999
6 (15)
2 (9)
0.699
APACHE-II score > 12 at extubation n (%)
29 (37)
24 (39)
5 (29)
0.575
14 (36)
10 (45)
0.587
ARF requiring > 72 h of iMV n (%)
44 (56)
30 (49)
14 (82)
0.025
19 (49)
11 (48)
> 0.999
BMI > 35 kg/m2n (%)
11 (14)
5 (8)
6 (35)
0.011
3 (8)
2 (9)
> 0.999
Neuromuscular disease n (%)
9 (12)
4 (7)
5 (29)
0.020
3 (8)
1 (5)
> 0.999
Reason for ICU admission
Acute heart failure n (%)
10 (13)
6 (10)
4 (23)
0.212
0 (0)
6 (27)
0.013
Acute on Chronic respiratory failure n (%)
17 (22)
12 (20)
5 (29)
0.507
7 (18)
5 (23)
0.742
Sepsis/septic shock n (%)
21 (27)
20 (33)
1 (6)
0.031
14 (36)
6 (27)
0.577
Trauma n (%)
10 (13)
7 (11)
3 (18)
0.682
7 (18)
0 (0)
0.042
Pneumonia n (%)
20
16 (26)
4 (24)
> 0.999
11 (28)
5 (23)
0.766
Characteristics before the SBT
PEEP (cmH2O)
5.0 [5.0; 5.1]
5.0 [5.0; 5.1]
5.0 [5.0; 5.0]
0.168
5.0 [5.0; 5.1]
5.0 [5.0; 5.1]
0.953
Inspiratory support (cmH2O)
10 [8; 11]
10 [8; 10]
10 [9; 11]
0.531
9 [8; 11]
10 [10; 10]
0.118
pH
7.44 [7.41; 7.47]
7.44 [7.40; 7.48]
7.46 [7.42; 7.48]
0.546
7.45 [7.41; 7.48]
7.42 [7.40; 7.45]
0.397
PaCO2 (mmHg)
39.3 [36.3; 45.0]
39.3 [36.0; 45.0]
39.5 [38.2; 44.4]
0.392
38.5 [34.0; 44.3]
39.3 [38.0; 47.3]
0.176
PaO2/FiO2 (mmHg)
268 [230; 310]
274 [241; 306]
248 [205; 324]
0.347
274 [243; 303]
270 [226; 313]
0.932
SBT: Spontaneous breathing trial; ICU: intensive care unit; SAPS-II: simplified acute physiology score II; SOFA: sequential organ failure assessment; PaCO2: arterial partial pressure of carbon dioxide; iMV: invasive mechanical ventilation; APACHE-II: acute physiology and chronic health evaluation II; ARF: acute respiratory failure; BMI: body mass index; PEEP: positive-end expiratory pressure; PaO2/FiO2: ratio between arterial partial pressure and inspired fraction of oxygen
×
Anzeige
SBT success and failure
Seventeen patients (21.8%) failed SBT and were returned to iMV. At baseline, PEEP values were similar between groups (5.0 [4.9; 5.1] cm H2O and 5.0 [5.0; 5.0] cm H2O for SBT success and failure, respectively; p = 0.168). Table 2 reports ABGs and RR/Vt separately for SBT success and failure. ABGs were not significantly different between groups at both baseline and SBT_30. RR/Vt was also not different between the groups at baseline but was significantly greater at SBT_0 and SBT_30 in SBT failures compared to SBT success (p < 0.001 for both).
Table 2
Arterial blood gases and the rapid shallow breathing index (RR/Vt) in patients with SBT success and failure
SBT success (n = 61)
SBT failure (n = 17)
Baseline
SBT_0
SBT_30
Baseline
SBT_0
SBT_30
SBT success
vs. failure
pH
7.44 [7.40; 7.48]
//
7.43 [7.39; 7.47]
7.46 [7.42; 7.48]
//
7.41 [7.38; 7.48]
p = 0.546a
p = 0.762c
PaCO2 (mmHg)
39.3 [36.0; 45.0]
//
40.0 [35.2; 46.0]
39.5 [38.2; 44.4]
//
41.0 [38.3; 53.5]
p = 0.392a
p = 0.194c
PaO2/FiO2 (mmHg)
274 [241; 306]
//
253 [221; 291]
248 [205; 324]
//
188 [156; 237]
p = 0.347a
p = 0.034c
RR/Vt (breaths/min/L)
35 [25; 57]
57 [38; 84]
60 [33; 92]
46 [26; 64]
102 [70; 135]
103 [75; 127]
p = 0.389a
p < 0.001b
p < 0.001c
SBT, Spontaneous Breathing Trial; SBT_0, first 5 min after the beginning of the SBT; SBT_30, last 5 min of the SBT; PaCO2, arterial partial pressure of carbon dioxide; PaO2/FiO2, arterial partial pressure to inspired fraction of oxygen ratio; RR/Vt, respiratory rate to tidal volume ration
//Data not present
aComparison between groups within baseline
bComparison between groups within SBT_0
cComparison between groups within SBT_30. According to Bonferroni correction, the threshold for statistical significance is p < 0.025 for pH, PaCO2 and PaO2/FiO2, while p < 0.017 for RR/Vt
Examples of EIT images from three representative patients, two succeeding (panels A and C) and one failing (panel B) SBT, are depicted in Fig. 2; ∆EELI and the inhomogeneity index are shown at baseline, SBT_0 and SBT_30 (see figure legend for detailed information). Table 3 displays the values of ∆Vt%, ∆EELI and inhomogeneity index in SBT success and failure, respectively. The ∆Vt% was no different at baseline (p > 0.999), SBT_0 (p = 0.079) and SBT_30 (p = 0.054). Compared to SBT successes, SBT failures showed a greater reduction in ∆EELI at SBT_0 (p < 0.001) and SBT_30 (p = 0.001), and a higher inhomogeneity index at baseline (p = 0.002), SBT_0 (p = 0.003) and SBT_30 (p = 0.005).
Table 3
EIT data in patients with SBT success and failure
SBT success (n = 61)
SBT failure (n = 17)
Success vs. failure
Baseline
SBT_0
SBT_30
Baseline
SBT_0
SBT_30
∆Vt%
0 [0; 0]
− 16.4 [− 37.0; − 2.5]
− 12.9 [− 35.0; 5.0]
0 [0; 0]
− 27.6 [− 51.0; − 16.0]
− 22.9 [− 50.6; − 13.3]
p > 0.999a
p = 0.079b
p = 0.054c
∆EELI (ml)
0 [0; 0]
− 117 [− 240; 21]
− 103 [− 292; 62]
0 [0; 0]
− 456 [− 934; − 162]
− 333 [− 1375; − 157]
p > 0.999a
p < 0.001b
p = 0.001c
Inhomogeneity index
51 [44; 61]
56 [48; 71]
57 [46; 70]
65 [54; 87]
89 [61; 105]
90 [62; 101]
p = 0.002a
p = 0.003b
p = 0.005c
SBT: Spontaneous Breathing Trial; SBT_0: first 5 min after the beginning of the SBT; SBT_30: last 5 min of the SBT; ∆Vt%: change from baseline of the tidal volume in percentage; ∆EELI: change from baseline of the end-expiratory lung impedance
aComparison between groups within baseline
bComparison between groups within SBT_0
cComparison between groups within SBT_30. According to Bonferroni correction, the threshold for statistical significance is p < 0.017
×
The AUC, sensitivity, specificity, LR + and LR − of ∆Vt%, ∆EELI, inhomogeneity index, RR/Vt and PaO2/FiO2 for SBT failure prediction are shown in Additional file 1: Table S2.
Extubation success and failure
Twenty-two of 61 (36.1%) patients extubated after SBT success met criteria for extubation failure: 15 for ABG worsening and respiratory distress and 7 for non-respiratory reasons (5 for neurological deterioration, one for inability to remove secretions and one for haemodynamic instability). Eleven patients were successfully treated by CPAP or NIV, while 11 required endotracheal intubation, 3 immediately and 8 after an unsuccessful attempt of CPAP or NIV.
Anzeige
As shown in Additional file 1: Table S3, there were no differences between extubation success and failure in pH (p = 0.397 at baseline; p = 0.453 at SBT_30; p = 0.179 at SB_30), PaCO2 (p = 0.176 at baseline; p = 0.467 at SBT_30; p = 0.757 at SB_30) and PaO2/FiO2 (p = 0.932 at baseline; p = 0.636 at SBT_30; p = 0.139 at SB_30). RR/VT was also not different among trials (p = 0.712 at baseline; p = 0.704 at SBT_0; p = 0.597 at SBT_30; p = 0.059 at SB_0; p = 0.185 at SB_30).
Table 4 reports the ∆Vt%, ∆EELI and inhomogeneity index for extubation success and failure. No differences were observed for any of these variables between extubation success and failure.
Table 4
EIT data in patients with extubation success and failure
Baseline
SBT_0
SBT_30
SB_0
SB_30
∆Vt%
Extubation success (n = 39)
0 [0; 0]
− 12 [− 31; 0]
− 9 [− 29; 7]
− 3 [− 16; 18]
− 12 [− 34; 19]
Extubation failure (n = 22)
0 [0; 0]
− 25 [− 42; − 10]
− 29 [− 45; − 11]
− 21 [− 39; 1]
− 22 [− 46; − 11]
Extubation success vs. failure
p > 0.999
p = 0.117
p = 0.024
p = 0.014
p = 0.061
∆EELI (ml)
Extubation success (n = 39)
0 [0; 0]
− 125 [− 237; 23]
− 78 [− 250; 115]
− 187 [− 488; 82]
− 60 [− 371; 256]
Extubation failure (n = 22)
0 [0; 0]
− 112 [− 316; 17]
− 194 [− 335; − 59]
− 236 [− 438; 139]
− 290 [− 537; 152]
Extubation success vs. failure
p > 0.999
p = 0.671
p = 0.253
p > 0.999
p = 0.132
Inhomogeneity index
Extubation success (n = 39)
46 [42; 63]
53 [44; 68]
56 [44; 70]
52 [45; 66]
53 [45; 69]
Extubation failure (n = 22)
53 [48; 60]
65 [55; 80]
62 [54; 71]
64 [50; 79]
66 [55; 74]
Extubation success vs. failure
p = 0.166
p = 0.025
p = 0.132
p = 0.049
p = 0.029
SBT: Spontaneous Breathing Trial; SBT_0: first 5 min after the beginning of the SBT; SBT_30: last 5 min of the SBT; SB_0: first 5 min of the spontaneous breathing; SB_30: last 5 min of the spontaneous breathing; ∆Vt%: change from baseline of the tidal volume in percentage; ∆EELI: change from baseline of the end-expiratory lung impedance
According to Bonferroni correction, the threshold for statistical significance is p < 0.017
The AUC, sensitivity, specificity, LR + and LR − of ∆Vt%, ∆EELI, inhomogeneity index, RR/Vt and PaO2/FiO2 for extubation failure prediction are shown in Additional file 1: Table S4. In addition, as reported in Additional file 1: Tables S5 and S6, we found no difference in any EIT parameters between patients succeeding and failing NIV and between patients with post-extubation respiratory failure due to respiratory and non-respiratory causes.
Discussion
In this physiological study aimed at describing, in a general population of critically ill patients at the first SBT attempt, changes in lung aeration, ventilation distribution and inhomogeneity occurring during the weaning process, we found the following: (1) compared to SBT success, SBT failure is characterized by more lung de-recruitment, as suggested by the higher loss in ∆EELI and by the concomitant PaO2/FiO2 worsening, and greater lung inhomogeneity, as indicated by higher inhomogeneity index value; and (2) no EIT variables were significantly different between patients succeeding and failing extubation.
All patients considered ready-to-wean were considered eligible for inclusion in this study, and we did not focus only on patients with prolonged weaning, who represent only approximately 15% of the overall ICU population [1]. We found an SBT failure rate of 21% which was not different from those previously reported [22‐24], while in keeping with Ferrer et al. [6], the rate of extubation failure in our population at high risk of post-extubation respiratory failure amounted to 35%. Finally, consistent with recently published data in patients recovering from hypoxemic acute respiratory failure [25], 50% of all patients experiencing post-extubation respiratory failure required reintubation, 42% of whom after failing CPAP or NIV.
The application of EIT during weaning has been recently reported only in small populations of patients characterized by prolonged weaning or prolonged mechanical ventilation [26‐28]. These studies report that, compared to patients undergoing a successful SBT, patients experiencing SBT failure are characterized by greater loss of EELI at the end of the SBT. While confirming this finding in a less selected patient population, we find that the loss of EELI occurs after only 5 min.
In the present study, the inhomogeneity index appears to be greater in SBT failure, as opposed to SBT success, which is also consistent with previously published data. In patients who had received prolonged iMV, Zhao et al. [27] also found that weaning was more likely to be successful when the ventilation distribution was more homogeneous at decreasing support levels. In 31 patients with prolonged weaning, Bickenbach et al. [26] found the inhomogeneity index to be greater in SBT failures, as opposed to successes. In our study, the inhomogeneity index was also significantly higher at baseline, while RR/Vt showed significant differences only in the course of SBT. In addition, after 5 min of SBT, we observed a remarkable increase in the inhomogeneity index, without important changes at SBT completion.
We also investigated potential variations in EIT parameters related to the outcome of extubation but found no differences in any of the EIT parameters considered between patients who succeeded and failed extubation. This finding may suggest, on the one hand, that the process leading to failure takes more than the 30-min period we observed in the present study, while on the other hand, that factors other than alterations of ventilation and homogeneity intervene in the pathophysiology of post-extubation respiratory failure. Indeed, in 7 patients (32%), the primary causes of extubation were either unmanageable secretions or haemodynamic instability or neurologic deterioration. In addition, we failed to observe differences in any EIT parameters between patients who succeeded and failed NIV as rescue treatment of overt post-extubation respiratory failure.
Our study has some points of strength. First, in contrast to previous investigations [26‐28], this is a multicentre study. Second, to reduce the risk of bias, the data analysis was performed in only one centre by a single investigator unaware of SBT and extubation outcomes. Third, we were extremely careful in maintaining the position of the EIT belt throughout the study period by marking the skin with a dermographic pencil, to prevent the risk of belt displacement, which is a well-known source of bias [10].
Our study also includes some limitations. We did not calculate the sample size because of the lack of published data at the time this study was designed and initiated. We did not register the trial prior to patient enrolment because the registration was not mandatory for observational studies when the study was initiated. We did not record the amount of fluids administered in the course of the study protocol. Fluid overload alters the measurement of EELI, mimicking a reduction in end-expiratory lung volume. One study in pigs showed that a rapid infusion (33 ml/kg/h) of fluids caused a significant reduction in overall relative impedance [29]. In another animal study, after a rapid (approximately 6 min) infusion of 500 ml of crystalloids the authors observed a decrease in EELI, corresponding to a mean virtual reduction in end-expiratory lung volume of 227 ml [30]. Conversely, a sudden increase in urine output up to 1200 ml consequent to a bolus of diuretics [31] or a loss of volume during haemodialysis [32] may determine EELI to increase. It should be noted, however, that a stable cardiovascular condition was a prerequisite for SBT eligibility in our study, which makes it highly unlikely that our patients received rapid liquid infusion or removal. In addition, none of the patients enrolled required haemodialysis or continuous veno-venous haemofiltration.
Conclusions
SBT failure is overall characterized early in the course of SBT by a greater reduction in EELI reflecting a fall in end-expiratory lung volume and more inhomogeneity of ventilation distribution, as opposed to patients succeeding SBT, but our data do not identify clear-cut thresholds. Whether EIT may be a useful adjunctive tool to monitor SBT remains to be determined by specifically designed clinical studies. No significant changes in lung aeration, ventilation distribution or homogeneity related to extubation outcome occurs up to 30 min after extubation.
Acknowledgements
None.
Ethical committee and Consent to participate
The study was approved by the local Ethics Committees (Vercelli, protocol no AslVC.Rian.14.02, approved on 11/09/2014; Catania, protocol no 479/CE, approved on 30/06/2015; Rimini, protocol no 1313, approved on 18/03/2015). Written informed consent was obtained from each participant before inclusion in the study, according to the local regulations and principles outlined in the Helsinki Declaration.
Consent for publication
All patients gave consent for data publication according to national regulations. All authors read the final manuscript and approved the submitted version.
Competing interests
Dr. Navalesi’s research laboratory has received equipment and grants from Maquet Critical Care, Draeger and Intersurgical S.p.A. He also received honoraria/speaking fees from Maquet Critical Care, Orionpharma, Philips, Resmed, MSD and Novartis. Dr. Navalesi contributed to the development of the helmet Next, whose licence for patent belongs to Intersurgical S.P.A., and receives royalties for that invention. Dr. Longhini and Dr. Navalesi contributed to the development of a new device not discussed in the present study whose patent is in progress (European Patent application number EP20170199831). The remaining authors have disclosed that they do not have any Competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electrical impedance tomography during spontaneous breathing trials and after extubation in critically ill patients at high risk for extubation failure: a multicenter observational study
verfasst von
Federico Longhini Jessica Maugeri Cristina Andreoni Chiara Ronco Andrea Bruni Eugenio Garofalo Corrado Pelaia Camilla Cavicchi Sergio Pintaudi Paolo Navalesi
Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.
Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.
Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.
Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.
Update AINS
Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.