Open Access 14.05.2019 | Computed Tomography

High sensitivity and specificity of 4D-CTA in the detection of cranial arteriovenous shunts

verfasst von: Matthijs in ’t Veld, Rolf Fronczek, Marlise P. dos Santos, Marianne A. A. van Walderveen, Frederick J. A. Meijer, Peter W. A. Willems

Erschienen in: European Radiology | Ausgabe 11/2019

Abstract

Purpose

In a prospective cohort study, we evaluated the diagnostic accuracy of time-resolved CT angiography (4D-CTA) compared to digital subtraction angiography (DSA) for detecting cranial arteriovenous shunts.

Material and methods

Patients were enrolled if a DSA had been ordered querying either a dural arteriovenous fistula (dAVF) or a cerebral arteriovenous malformation (bAVM). After enrolment, both a DSA and a 4D-CTA were performed. Both studies were evaluated using a standardized form. If a dAVF or bAVM was found, its classification, angioarchitectural details, and treatment options were recorded.

Results

Ninety-eight patients were enrolled and 76 full datasets were acquired. DSA demonstrated a shunting lesion in 28 out of 76 cases (prevalence 37%). 4D-CTA demonstrated all but two of these lesions (sensitivity of 93%) and produced one false positive (specificity of 98%). These numbers yielded a positive predictive value (PPV) of 96% and a negative predictive value (NPV) of 96%. Significant doubt regarding the 4D-CTA diagnosis was reported in 6.6% of all cases and both false-negative 4D-CTA results were characterized by such doubt.

Conclusions

4D-CTA has very high sensitivity and specificity for the detection of intracranial arteriovenous shunts. Based on these results, 4D-CTA may replace DSA imaging as a first modality in the diagnostic workup in a large number of patients suspected of a cranial dAVF or bAVM, especially if there is no doubt regarding the 4D-CTA diagnosis.

Key Points

• 4D-CTA was shown to have a high diagnostic accuracy and is an appropriate, less invasive replacement for DSA as a diagnostic tool for cranial arteriovenous shunts in the majority of suspected cases.
• Doubt regarding the 4D-CTA result should prompt additional DSA imaging, as it is associated with false negatives.
• False-positive 4D-CTA results are rare, but do exist.
Hinweise
Matthijs in ’t Veld and Rolf Fronczek contributed equally to this work.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
4D-CTA
Time-resolved CTA
bAVM
Brain arteriovenous malformation
dAVF
Dural arteriovenous fistula
DSA
Digital subtraction angiography
ECA
External carotid artery
ICA
Internal carotid artery
MRA
MR angiography
SM
Spetzler-Martin
VA
Vertebral artery

Introduction

Intracranial arteriovenous malformations (AVMs) represent abnormal connections between arteries and veins. They may be divided into brain (pial) AVMs (bAVMs) and dural AVMs depending on whether the feeders are pial or dural arteries. bAVMs may be further divided into nidus-type AVMs, if there is a tangle of abnormal vessels, and arteriovenous fistulas (AVFs), if there is no intervening network [1, 2]. Since dural AVMs do not have a true nidus, they may also be called dural AVFs (dAVFs). dAVFs may drain solely through venous sinuses in which case they do not represent a risk of hemorrhage. They may also drain through or reflux into pial veins in which case they do represent a risk of hemorrhage [3, 4]. bAVMs and dAVFs may be asymptomatic or present with a wide variety of symptoms such as focal neurological deficits, headache, bruit, or seizures. Angiographically, all of these lesions are characterized by early venous filling caused by the lack of passage of blood through a normal capillary network.
Although computed tomographic angiography (CTA) and magnetic resonance angiography (MRA) can demonstrate many characteristics of an arteriovenous shunt, digital subtraction angiography (DSA) is the gold standard to prove the existence of the shunt and distinguish it from mimicking lesions [57]. Furthermore, DSA can identify the feeding arteries and venous draining patterns [8]. Unfortunately, DSA is time-consuming, expensive, and not without risks, due to its invasive nature. Although DSA may be considered to be a safe diagnostic procedure in experienced hands, it carries a risk of (silent) embolic events, especially in patients with vascular risk factors [9]. Serious complications are rare, but can result in transient or permanent neurological damage [1012].
Whole-head time-resolved CTA (4D-CTA) allows visualization of blood flow dynamics in cranial vessels with the first pass of an intravenous contrast bolus. 4D-CTA is less expensive and less time-consuming than conventional DSA [13, 14]. Furthermore, it is non-invasive and carries no risk of thromboembolic complications. Earlier small studies of selected cases suggested a role for 4D-CTA in the diagnostic workup, treatment planning, and follow-up of cranial dAVFs and bAVMs [1518]. The current study aims to determine the true diagnostic value of 4D-CTA for such lesions in a non-selected patient population.

Material and methods

Patient data and data collection

The study protocol was approved by our institutional research ethics board in April 2010 and was conducted according to the principles of the Declaration of Helsinki.
At the two participating institutions (Leiden University Medical Centre and Radboud University Medical Centre), all patients were prospectively recruited from October 2010 until January 2016 in a tertiary care setting. Patients were included if they were 18 years or older, the presence of a dAVF or bAVM was suspected, and a DSA was requested to confirm this diagnosis. Exclusion criteria were a previous diagnosis of dAVF or bAVM with time-resolved imaging, current treatment for diabetes mellitus, renal failure (baseline eGFR < 50 ml/min), or a known allergy to iodinated contrast agents. After obtaining informed consent, patients underwent both DSA and 4D-CTA imaging.

DSA examination

All DSA examinations were performed using standard biplane fluoroscopy (Infinix®, Toshiba Medical Systems; or Allura Xper®, Philips Healthcare). A routine angiographic protocol was performed and generally consisted of bilateral injections of the internal and external carotid arteries and at least one vertebral artery. During each injection, anterior-posterior (AP) and lateral projections were obtained at 6 frames/s.

4D-CTA examination

4D-CTA examination was performed using an Aquilion ONE multidetector CT scanner (Toshiba Medical Systems), equipped with 320 × 0.5 mm detector rows covering 16 cm of volume per rotation. Imaging was performed in a manner previously described [1416, 19].
Although acquisition protocols could vary, the study protocol required a minimum rotation speed of 1 Hz and a maximum reconstruction interval of 1 s during the continuous phase, i.e., the phase of continuous acquisition during which the contrast bolus traveled from arterial to venous vasculature. The highest rotation speed and shortest reconstruction interval used were 2 Hz and 0.3 s, respectively [20].
Typically, the entire study would be as follows. An intravenous infusion of 60 ml non-ionic contrast was followed by 29 ml of saline. Subsequently, the following volumes were acquired: a mask volume, two volumes before contrast medium could reach the cranial vessels, the continuous phase as detailed above, and three volumes during venous “wash-out.” The mask volume was subtracted from all subsequent volumes which were then stored in DICOM files.
Depending on the reconstruction interval during the continuous phase, a total of 22 to 51 volumes were reconstructed. These volumes were used to reconstruct time-resolved (arterial to venous) maximum intensity projection series in AP, lateral, and several oblique viewing angles. Typically, the oblique reconstructions were projected at 30°, 45°, and 60°. Two series were added in the arterial phase (without temporal resolution) in which the volume rotated around an axial axis (“spin”) and a lateral axis (“tumble”), again using maximum intensity projections. Finally, a slice stack was generated using time-averaged intensities. If necessary, additional post-processing options were available, but were generally not requested.

Angiographic evaluation

One reader at each participating institution read the DSA and 4D-CTA images, locally. This was an experienced neuroradiologist (FJAM) at one institution and an experienced neurointerventionalist (PWAW) at the other. A standard score sheet was used to score all relevant aspects of the imaging. To blind the readers for the DSA result at the time of 4D-CTA evaluation, the 4D-CTA was always reviewed before the DSA. The protocol allowed this to be done after both studies had already been acquired, as long as the reader was not aware of the DSA result.
The scoring sheet consisted of the following items: the presence of a vascular lesion (“dAVF,” “bAVM,” or “other”); the certainty of the diagnosis using a 3-point scale (“absolutely certain,” “sufficiently certain,” or “significant doubt”); the overall quality of the images in case of 4D-CTA, using a 3-point scale (“good,” “moderate,” “poor”); the classification according to either Borden for dAVFs or Spetzler-Martin for bAVMs; the type of fistulous area in case of a dAVF (“focal” or “diffuse”) and the types of feeders that could be identified. The latter assessment consisted of “posterior circulation” and/or “anterior circulation” for bAVMs and “external carotid” and/or “internal carotid” and/or “vertebrobasilar” for dAVFs. Finally, the reader was asked to record whether the 4D-CTA study would have been sufficient to determine the best treatment strategy and whether DSA would have been required, additionally, prior to such treatment.
With regard to the first item on the scoring sheet, i.e., the diagnosis, pial AVFs were classified as bAVMs, since the distinction between a very small nidus (micro-AVM) and a pial AVF may be difficult on imaging and their hemodynamic and clinical characteristics are very similar.

Interobserver analysis

Following the initial analysis, a representative subset of 30 4D-CTA records was selected. These were independently evaluated by a third experienced neurointerventional radiologist (MAAVW) in exactly the same way as the initial analysis. This radiologist was blinded for both the prior 4D-CTA assessments and the DSA results.

Complications of DSA

For all cases with complete datasets, the hospital records were screened for complications relating to the DSA.

Statistical analysis

Sensitivity, specificity, and positive and negative predictive values for 4D-CTA were calculated using SPSS® software (version 23.0, IBM Corp.), with DSA as the golden standard. For the interobserver analysis, kappa values were calculated for the diagnosis (“no diagnosis,” “dAVF,” or “bAVM”), Borden classification, and Spetzler-Martin classification.

Results

Inclusion

Between October 2010 and January 2016, 98 patients with a clinical suspicion of a cranial arteriovenous shunt were included. Seventy-six complete datasets, containing both a DSA and a 4D-CTA, were obtained in 75 patients. The average interval (± SD) between both studies was 32.6 days (± 43.7). In 71.1% of the cases, the second study was acquired within 4 weeks of the first. One patient was enrolled twice on two unrelated occasions, first due to hemorrhage and 3 months later due to tinnitus. This yielded two complete datasets which were analyzed independently (see Fig. 1 for the patient inclusion details). Table 1 shows the baseline characteristics and presenting symptoms of these patients.
Table 1
Patient characteristics
Characteristic
 
Number
Percentage
Age (mean ± SD)
 
53.1 ± 12.6
 
Sex
Male
35
46.1
Female
41
53.9
Presenting symptom
Hemorrhage
12
15.6
Focal neurologic deficits
2
2.6
Tinnitus
41
53.2
Bruit
6
7.8
Seizure
3
3.9
Headache
4
5.2
Other symptoms
8
11.7
SD, standard deviation

Prior imaging

Prior imaging was performed in 61 cases (80.3%), consisting of CT (17), contrast-enhanced CT (6), CTA (13), MR (20), contrast-enhanced MR (20), and/or MRA (28). In 35 cases (57.4%), this prior imaging was reported to have contributed to the suspicion of an arteriovenous shunt (dAVF or bAVM). Since we aimed to determine the value of 4D-CTA in a realistic setting, this prior imaging was available to the person reading the DSA or 4D-CTA.

Diagnostic value of 4D-CTA

The imaging results are listed in Table 2. DSA demonstrated an arteriovenous shunt in 28 out of 76 datasets (prevalence of 36.8%). This was a bAVM in 9 and a dAVF in 19 cases. 4D-CTA demonstrated all but two of these shunting lesions. Of the 9 bAVM patients, 8 were correctly diagnosed (88.9%) and one was reported normal by 4D-CTA (false negative). Of the 19 dAVF patients, 17 were correctly diagnosed (89.5%), one was reported as a bAVM (misclassified), and one was reported normal by 4D-CTA (false negative). Of the 48 patients where DSA did not demonstrate a shunt, 47 were correctly negative (97.9%) and one was incorrectly diagnosed with a Borden type 1 dAVF with 4D-CTA (false positive). The resulting 4D-CTA test characteristics are shown in Table 3.
Table 2
Diagnosis found by 4D-CTA
Diagnosis by 4DCTA
No.
Diagnosis by DSA
No.
AVM
9
Confirmed AVM
8
  
dAVF
1
dAVF
18
Confirmed dAVF
17
  
No diagnosis
1
Other diagnosis
8
Confirmed “Other diagnosis”
8
No diagnosis
41
Confirmed “No diagnosis”
37
  
AVM
1
  
dAVF
1
  
Other diagnosis
2
 
n = 76
 
n = 76
4D-CTA, time-resolved computer tomography angiography; DSA, digital subtraction angiography; dAVF, dural arteriovenous fistula; AVM, arteriovenous malformation
Table 3
Sensitivity, specificity, positive predictive value, and negative predictive value of 4D-CTA in diagnosing cerebral shunting lesions
  
DSA
 
Shunt present
Shunt absent
4DCTA
Test positive
26
1
27
Test negative
2
47
49
  
28
48
76
Sensitivity 92.9%, positive predictive value 96.3%
Specificity 97.9%, negative predictive value 95.9%
4D-CTA, time-resolved computer tomography angiography; DSA, digital subtraction angiography
Results of 4D-CTA and DSA regarding further details of the arteriovenous shunts are provided in Table 4. Although 4D-CTA agreed with DSA on Borden and Spetzler-Martin classification, there was less agreement regarding further details such as feeding arteries and delineation of the fistulous area.
Table 4
Characteristics of the shunting lesions confirmed by 4D-CTA
Diagnosis
Characteristics
DSA (golden standard)
Correctly identified by 4D-CTA (%)
dAVF
Number of dAVFs
19
17a
Borden  I
10
10(100%)
     II
2
2 (100%)
     III
7
5 (71%)
Feeding branches
 ECA
10
7 (70%)
 ICA
1
1 (100%)
 VA
1
1 (100%)
 Multiple feeding territories
7
0 (0%)
Fistulous area
 Focal area
12
6 (50%)
 Diffuse area
7
3 (43%)
AVM
Number of AVMs
9
8b
SM  size < 3 cm
9
8 (89%)
    Eloquent
    Non-eloquent
5
4
4 (80%)
4 (100%)
    Superficial drainage
    Deep drainage
4
5
3 (75%)
5 (100%)
Feeding branches
 Anterior circulation
5
4 (80%)
 Posterior circulation
2
1 (50%)
 Both
1
0 (0%)
a4D-CTA missed a dAVF in one patient and one dAVF was misclassified
b4D-CTA missed an AVM in one patient
dAVF dural arteriovenous fistula, AVM arteriovenous malformation, DSA digital subtraction angiography, 4D-CTA time-resolved computer tomography angiography, ECA external carotid artery, ICA internal carotid artery, VA vertebral artery, SM Spetzler-Martin
With regard to the locations of the 19 dAVFs, 13 were located at the level of the transverse or sigmoid sinus, one at the level of the torcular herophili, one at the free margin of the tentorium cerebelli, one in the parietal region, draining directly into a cortical vein, one on the cerebellar convexity. The final two were cavernous sinus dAVFs.
In 10 cases, DSA showed an abnormality other than a shunt: two dural sinus stenoses, three developmental venous anomalies, two arterial aneurysms, one venous aneurysm, one calcifying tumor, and one internal carotid artery dissection. Eight of these (80%) were also correctly reported on 4D-CTA and two were not reported: one developmental venous anomaly and one venous aneurysm.

Certainty and image quality in 4D-CTA

The image quality was reported to be “good” in 55 (73.7%), “moderate” in 15 (19.7%), and “poor” in 5 cases (6.6%). Comparing DSA with 4D-CTA, the level of certainty regarding the diagnosis was as follows: 72 versus 22 for “absolutely certain,” 4 versus 49 for “sufficiently certain,” and 0 versus 5 for “significant doubt.”
For the two cases with a false-negative 4D-CTA result, the image quality was reported to be “good” in one and “moderate” in the other and the level of certainty regarding the 4D-CTA diagnosis was “significant doubt” in both. In other words, out of five 4D-CTAs with significant doubt, two diagnoses were indeed false negative (40%). Thus, if we removed all cases with significant doubt about the diagnosis (6.6% of the 4D-CTAs), the diagnostic value of the 4D-CTA would increase, with a sensitivity of 100%. The reasons for reduced image quality were not addressed in the score sheets but, generally, consist of motion artifacts and or poor contrast bolus timing.

Treatment strategy

In 16 of the 27 cases (59.3%) where 4D-CTA demonstrated a bAVM or dAVF, the 4D-CTA study was thought to be sufficient to determine the best applicable treatment strategy. In the other 11 cases, additional DSA was still deemed necessary for the treatment planning.

Complications of DSA

Ten patients (13.2%) reported temporary subjective symptoms such as slightly blurred vision (6), headaches (3), or a combination of the two (1).
Two patients (2.6%) experienced transient neurological signs: one mild ataxia and one hemiparesis. Both recovered quickly and completely.
Eight cases (10.5%) suffered from an inguinal hematoma. In four of these patients, Angio Seal (St. Jude Medical) had been used as closure system.
No permanent deficits, systemic complications, or mortality were recorded and none of the patients required readmission after discharge.

Details of the four cases that were incorrectly classified by 4D-CTA

Case 1 (false positive, Fig. 2): the 4D-CTA of a 52-year-old female, presenting with tinnitus, suggested a Borden-type I dAVF. 4D-CTA image quality was good with the diagnosis being considered sufficiently certain. It showed early venous filling in the late arterial phase in the right hemisphere which was not reproduced by DSA.
Case 2 (misclassification, Fig. 3): the 4D-CTA of a 59-year-old male, presenting with headache, was of poor image quality and showed early venous filling and appeared to demonstrate a bAVM, as an abnormal vascular pattern suggested the presence of a nidus. The 4D-CTA diagnosis was considered to be sufficiently certain. However, DSA demonstrated a Borden-type I dAVF.
Case 3 (false negative, Fig. 4): the 4D-CTA of a 29-year-old female, presenting with hemorrhage, was of good image quality and did not demonstrate an abnormality, although significant doubt was noted. DSA demonstrated a micro-AVM (< 3 cm) with multiple draining veins.
Case 4 (false negative, Fig. 5): the 4D-CTA of a 31-year-old male, presenting with a seizure, was of moderate image quality and could not demonstrate an AV shunt. Significant doubt was noted and DSA subsequently demonstrated a Borden-type III dAVF.

Interobserver analysis

Thirty cases were selected to represent the entire study population: 21 negative cases, 6 dAVFs, and 3 bAVMs. The 4D-CTAs from these cases were independently read a second time. Regarding diagnosis (no diagnosis, dAVF, and bAVM), the observers agreed in 29 cases (97%, kappa = 0.925). The one case reported differently had a Borden-type III dAVF that was correctly reported by one reader and missed by the other. Both readers reported the image quality to be poor and both reported significant doubt regarding the diagnosis with 4D-CTA. This was due to movement artifacts and poor timing of the contrast bolus.
Regarding Borden and Spetzler-Martin classifications, both observers agreed in all cases (kappa = 1.00).

Discussion

In this prospective cohort study, 4D-CTA proved to have high diagnostic accuracy for detecting cranial arteriovenous shunts, supporting its use as a non-invasive alternative to DSA. We demonstrated that 4D-CTA has high sensitivity (93%) and specificity (98%) for detecting shunting lesions and that all false-negative results were characterized by significant doubt regarding the diagnosis.
Since all arteriovenous shunts lead to early venous filling, it is this feature that is looked for in the angiographic images to determine whether a shunt exists. Once the early venous filling is detected, features such as appearance of a nidus, the location of the shunt, and the apparent feeders will help to determine whether the lesion matches a dAVF or a bAVM. Although 4D-CTA misdiagnosed over 10% of the AV shunts (3 out of 28; 10.7%), in one of these, a shunt was found on 4D-CTA but erroneously diagnosed as a bAVM instead of a dAVF. The other two were truly false negative but read with significant doubt. Significant doubt was recorded in 6.6% of cases and should prompt additional imaging. In other words, if 4D-CTA is used as a diagnostic tool when an AV shunt is suspected, there is a small chance (6.6%) that doubt regarding the diagnosis persists and DSA is still warranted. In the remaining majority, all lesions will be found, but with a very small chance (1.4%) that 4D-CTA will not interpret the type of shunt correctly.
Although in the lesions detected with 4D-CTA, 4D-CTA agreed with DSA regarding Borden or Spetzler-Martin classification, DSA is more accurate in the evaluation of angioarchitectural details. This is due to selective vessel injection and a higher spatial and temporal resolution with DSA. Although 4D-CTA resolution may improve with future developments, non-selective vessel opacification is a fundamental characteristic of this modality and will, thus, always hamper image interpretation to some extent. On the other hand, 4D-CTA adds time-resolved cross-sectional imaging and also visualizes non-vascular tissues, which may prove beneficial during surgery, especially as it may be used in frameless stereotaxy. In earlier studies, we demonstrated these differences between 4D-CTA and DSA in patients known to have a dAVF or bAVM [1518]. Other authors also reported on the use of 4D-CTA in both spinal and cranial arteriovenous shunts [21, 22]. However, in all of these studies, only patients who were already diagnosed with such a lesion were examined. The present study not only corroborates those findings, but also demonstrates the diagnostic capabilities of 4D-CTA in a patient population suspected, rather than known, to have a bAVM or dAVF. Thus, this is the first demonstration of the true value of 4D-CTA as a diagnostic tool.
Based on these results, in centers with similar imaging technology available, an additional DSA could be reserved for cases when there is significant doubt regarding the 4D-CTA diagnosis (6.6% of all cases) or when a dAVF or bAVM has been found by 4D-CTA, to further determine its precise angioarchitectural details and allow treatment planning. In our population, this would have resulted in the avoidance of a DSA in almost two-thirds (62%) of cases. An even higher number of cases would not require a DSA if it is omitted when treatment is not being considered, e.g., based on the lesion’s classification or other clinical details. Finally, endovascular treatment will allow DSA visualization at the beginning of the procedure, again sparing the patient an additional diagnostic DSA session. This reduction in DSA studies is important as 4D-CTA is less invasive, less expensive, and less time-consuming and does not carry the risk of neurological complications.
Despite the high sensitivity and specificity, two shunting lesions were not detected by 4D-CTA: one dAVF and one bAVM (false-negative 4D-CTA results). Not having discovered such a lesion might have had great consequences for the patients involved, as both lesions represented a risk of intracranial hemorrhage. Index of suspicion was high in both cases, as the dAVF presented with hemorrhage and the bAVM with a neurological symptom (seizure). Using the above reasoning, both would have required an additional DSA due to significant doubt regarding the 4D-CTA diagnosis. However, one could also argue that a case with such a high index of suspicion deserves a DSA, rather than or supplementary to a 4D-CTA, anyway.
Furthermore, a high interobserver agreement was found regarding the presence of a dAVF or bAVM and its classification. The single case with disagreement between the two readers again underlines the need for additional imaging when image quality is poor and significant doubt regarding the diagnosis persists.
With regard to radiation dose, it is not easy to compare the dose of 4D-CTA with that of DSA, since the dose received during these studies depends on many variables. In an earlier publication, we reported the radiation dose of diagnostic cerebral DSA to lie between 7.9 and 9.1 mSv [16]. As presented in our methods, different 4D-CTA acquisition protocols were used in our study with the tradition dose varying between 5.2 and 11.5 mSv. This depended mostly on the rotation speed during the continuous acquisition phase.
There are several limitations to the present study. Although we intended to include an unselected population suspected of having a dAVF or bAVM, this did not appear entirely feasible. The protocol dictated that the 4D-CTA and DSA were not performed on the same day due to the cumulative contrast burden. Thus, in emergency cases, inclusion was sometimes thought to introduce an unacceptable delay in diagnosis and treatment. This may have resulted in a skewed distribution of presenting symptoms, with less intracranial hemorrhages and more non-emergency presentations such as pulsatile tinnitus. Furthermore, the protocol did not require any non-dynamic imaging, such as single-phase CTA or MRI/MRA, to be acquired prior to inclusion. Therefore, we cannot draw conclusions regarding the value of 4D-CTA versus such modalities. Moreover, the protocol excluded patients with prior time-resolved imaging, since we were interested in newly suspected cases. Thus, any comparison between 4D-CTA and such time-resolved modalities, e.g., trMRA, also lies beyond the scope of this study. Finally, it would be of interest to use 4D-CTA in other patient categories as well, e.g., as repeat imaging modality in patients with a wait-and-scan policy or to check for therapy efficacy after surgery or endovascular treatment. Although such use of 4D-CTA appears attractive, our results cannot be extrapolated to these patient categories.

Conclusion

4D-CTA is a less invasive dynamic imaging tool than DSA, with high sensitivity, specificity, and interobserver reliability in the detection and classification of intracranial arteriovenous shunts. Although, in rare occasions, false negatives may occur, in our population, these cases were all characterized by significant doubt regarding the 4D-CTA diagnosis. Taking this into account, 4D-CTA could have replaced DSA in the majority of our study population. Our study provides evidence that 4D-CTA is an alternative to DSA as a first modality in the diagnostic workup of patients suspected of having a dAVF or bAVM.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is Dr. Peter W.A. Willems.

Conflict of interest

The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Statistics and biometry

No complex statistical methods were necessary for this paper.
Written informed consent was obtained from all subjects (patients) in this study.

Ethical approval

Institutional Review Board approval was obtained.

Study subjects or cohorts overlap

Data of 50 patients with pulsatile tinnitus have been previously reported in our previous paper (In ‘t Veld M, Fronczek R, de Laat JA, Kunst HPM, Meijer FJA, Willems PWA (2018) The Incidence of Cranial Arteriovenous Shunts in Patients With Pulsatile Tinnitus: A Prospective Observational Study. Otol Neurotol 39:648–653). The 4D-CTA results of those cases were not reported and neither were the results of the remaining cases, presenting with other symptoms than pulsatile tinnitus.

Methodology

• prospective
• observational
• multicenter study
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Lasjaunias P, Chiu M, ter Brugge K, Tolia A, Hurth M, Bernstein M (1986) Neurological manifestations of intracranial dural arteriovenous malformations. J Neurosurg 64:724–730CrossRef Lasjaunias P, Chiu M, ter Brugge K, Tolia A, Hurth M, Bernstein M (1986) Neurological manifestations of intracranial dural arteriovenous malformations. J Neurosurg 64:724–730CrossRef
2.
Zurück zum Zitat Lasjaunias P (1997) Pial arteriovenous malformation. In: Vascular diseases in neonates, infants and children: interventional neuroradiology management. Springer-Verlag, Berlin, Heidelberg, pp 203–319 Lasjaunias P (1997) Pial arteriovenous malformation. In: Vascular diseases in neonates, infants and children: interventional neuroradiology management. Springer-Verlag, Berlin, Heidelberg, pp 203–319
3.
Zurück zum Zitat van Dijk JM, terBrugge KG, Willinsky RA, Wallace MC (2002) Clinical course of cranial dural arteriovenous fistulas with long-term persistent cortical venous reflux. Stroke 33:1233–1236CrossRef van Dijk JM, terBrugge KG, Willinsky RA, Wallace MC (2002) Clinical course of cranial dural arteriovenous fistulas with long-term persistent cortical venous reflux. Stroke 33:1233–1236CrossRef
4.
Zurück zum Zitat Borden JA, Wu JK, Shucart WA (1995) A proposed classification for spinal and cranial dural arteriovenous fistulous malformations and implications for treatment. J Neurosurg 82:166–179CrossRef Borden JA, Wu JK, Shucart WA (1995) A proposed classification for spinal and cranial dural arteriovenous fistulous malformations and implications for treatment. J Neurosurg 82:166–179CrossRef
5.
Zurück zum Zitat Krings T, Hans F (2004) New developments in MRA: time-resolved MRA. Neuroradiology 46(Suppl 2):s214–s222CrossRef Krings T, Hans F (2004) New developments in MRA: time-resolved MRA. Neuroradiology 46(Suppl 2):s214–s222CrossRef
6.
Zurück zum Zitat Siebert E, Bohner G, Dewey M et al (2009) 320-slice CT neuroimaging: initial clinical experience and image quality evaluation. Br J Radiol 82:561–570CrossRef Siebert E, Bohner G, Dewey M et al (2009) 320-slice CT neuroimaging: initial clinical experience and image quality evaluation. Br J Radiol 82:561–570CrossRef
7.
Zurück zum Zitat Geibprasert S, Pongpech S, Jiarakongmun P, Shroff MM, Armstrong DC, Krings T (2010) Radiologic assessment of brain arteriovenous malformations: what clinicians need to know. Radiographics 30:483–501CrossRef Geibprasert S, Pongpech S, Jiarakongmun P, Shroff MM, Armstrong DC, Krings T (2010) Radiologic assessment of brain arteriovenous malformations: what clinicians need to know. Radiographics 30:483–501CrossRef
8.
Zurück zum Zitat Krings T, Hans FJ, Geibprasert S, Terbrugge K (2010) Partial “targeted” embolisation of brain arteriovenous malformations. Eur Radiol 20:2723–2731CrossRef Krings T, Hans FJ, Geibprasert S, Terbrugge K (2010) Partial “targeted” embolisation of brain arteriovenous malformations. Eur Radiol 20:2723–2731CrossRef
9.
Zurück zum Zitat Bendszus M, Koltzenburg M, Burger R, Warmuth-Metz M, Hofmann E, Solymosi L (1999) Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet 354:1594–1597CrossRef Bendszus M, Koltzenburg M, Burger R, Warmuth-Metz M, Hofmann E, Solymosi L (1999) Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study. Lancet 354:1594–1597CrossRef
10.
Zurück zum Zitat Cloft HJ, Joseph GJ, Dion JE (1999) Risk of cerebral angiography in patients with subarachnoid hemorrhage, cerebral aneurysm, and arteriovenous malformation: a meta-analysis. Stroke 30:317–320CrossRef Cloft HJ, Joseph GJ, Dion JE (1999) Risk of cerebral angiography in patients with subarachnoid hemorrhage, cerebral aneurysm, and arteriovenous malformation: a meta-analysis. Stroke 30:317–320CrossRef
11.
Zurück zum Zitat Kaufmann TJ, Huston J 3rd, Mandrekar JN, Schleck CD, Thielen KR, Kallmes DF (2007) Complications of diagnostic cerebral angiography: evaluation of 19,826 consecutive patients. Radiology 243:812–819CrossRef Kaufmann TJ, Huston J 3rd, Mandrekar JN, Schleck CD, Thielen KR, Kallmes DF (2007) Complications of diagnostic cerebral angiography: evaluation of 19,826 consecutive patients. Radiology 243:812–819CrossRef
12.
Zurück zum Zitat Willinsky RA, Taylor SM, TerBrugge K, Farb RI, Tomlinson G, Montanera W (2003) Neurologic complications of cerebral angiography: prospective analysis of 2,899 procedures and review of the literature. Radiology 227:522–528CrossRef Willinsky RA, Taylor SM, TerBrugge K, Farb RI, Tomlinson G, Montanera W (2003) Neurologic complications of cerebral angiography: prospective analysis of 2,899 procedures and review of the literature. Radiology 227:522–528CrossRef
13.
Zurück zum Zitat Klingebiel R, Siebert E, Diekmann S et al (2009) 4-D imaging in cerebrovascular disorders by using 320-slice CT: feasibility and preliminary clinical experience. Acad Radiol 16:123–129CrossRef Klingebiel R, Siebert E, Diekmann S et al (2009) 4-D imaging in cerebrovascular disorders by using 320-slice CT: feasibility and preliminary clinical experience. Acad Radiol 16:123–129CrossRef
14.
Zurück zum Zitat Brouwer PA, Bosman T, van Walderveen MA, Krings T, Leroux AA, Willems PW (2010) Dynamic 320-section CT angiography in cranial arteriovenous shunting lesions. AJNR Am J Neuroradiol 31:767–770CrossRef Brouwer PA, Bosman T, van Walderveen MA, Krings T, Leroux AA, Willems PW (2010) Dynamic 320-section CT angiography in cranial arteriovenous shunting lesions. AJNR Am J Neuroradiol 31:767–770CrossRef
15.
Zurück zum Zitat Willems PW, Brouwer PA, Barfett JJ, terBrugge KG, Krings T (2011) Detection and classification of cranial dural arteriovenous fistulas using 4D-CT angiography: initial experience. AJNR Am J Neuroradiol 32:49–53CrossRef Willems PW, Brouwer PA, Barfett JJ, terBrugge KG, Krings T (2011) Detection and classification of cranial dural arteriovenous fistulas using 4D-CT angiography: initial experience. AJNR Am J Neuroradiol 32:49–53CrossRef
16.
Zurück zum Zitat Willems PW, Taeshineetanakul P, Schenk B, Brouwer PA, Terbrugge KG, Krings T (2012) The use of 4D-CTA in the diagnostic work-up of brain arteriovenous malformations. Neuroradiology 54:123–131CrossRef Willems PW, Taeshineetanakul P, Schenk B, Brouwer PA, Terbrugge KG, Krings T (2012) The use of 4D-CTA in the diagnostic work-up of brain arteriovenous malformations. Neuroradiology 54:123–131CrossRef
17.
Zurück zum Zitat Biswas S, Chandran A, Radon M et al (2015) Accuracy of four-dimensional CT angiography in detection and characterisation of arteriovenous malformations and dural arteriovenous fistulas. Neuroradiol J 28:376–384CrossRef Biswas S, Chandran A, Radon M et al (2015) Accuracy of four-dimensional CT angiography in detection and characterisation of arteriovenous malformations and dural arteriovenous fistulas. Neuroradiol J 28:376–384CrossRef
18.
Zurück zum Zitat Singh R, Gupta V, Ahuja C, Khandelwal N (2018) Time resolved computed tomography angiography in the evaluation of brain arteriovenous malformation: a feasibility study. Neuroradiol J 31:230–234CrossRef Singh R, Gupta V, Ahuja C, Khandelwal N (2018) Time resolved computed tomography angiography in the evaluation of brain arteriovenous malformation: a feasibility study. Neuroradiol J 31:230–234CrossRef
19.
Zurück zum Zitat Salomon EJ, Barfett J, Willems PW, Geibprasert S, Bacigaluppi S, Krings T (2009) Dynamic CT angiography and CT perfusion employing a 320-detector row CT: protocol and current clinical applications. Klin Neuroradiol 19:187–196CrossRef Salomon EJ, Barfett J, Willems PW, Geibprasert S, Bacigaluppi S, Krings T (2009) Dynamic CT angiography and CT perfusion employing a 320-detector row CT: protocol and current clinical applications. Klin Neuroradiol 19:187–196CrossRef
20.
Zurück zum Zitat Hoogenboom TC, van Beurden RM, van Teylingen B, Schenk B, Willems PW (2012) Optimization of the reconstruction interval in neurovascular 4D-CTA imaging. A technical note. Interv Neuroradiol 18:377–379CrossRef Hoogenboom TC, van Beurden RM, van Teylingen B, Schenk B, Willems PW (2012) Optimization of the reconstruction interval in neurovascular 4D-CTA imaging. A technical note. Interv Neuroradiol 18:377–379CrossRef
21.
Zurück zum Zitat Yamaguchi S, Takemoto K, Takeda M et al (2017) The position and role of four-dimensional computed tomography angiography in the diagnosis and treatment of spinal arteriovenous fistulas. World Neurosurg 103:611–619CrossRef Yamaguchi S, Takemoto K, Takeda M et al (2017) The position and role of four-dimensional computed tomography angiography in the diagnosis and treatment of spinal arteriovenous fistulas. World Neurosurg 103:611–619CrossRef
22.
Zurück zum Zitat Wang H, Ye X, Gao X, Zhou S, Lin Z (2014) The diagnosis of arteriovenous malformations by 4D-CTA: a clinical study. J Neuroradiol 41:117–123CrossRef Wang H, Ye X, Gao X, Zhou S, Lin Z (2014) The diagnosis of arteriovenous malformations by 4D-CTA: a clinical study. J Neuroradiol 41:117–123CrossRef
Metadaten
Titel
High sensitivity and specificity of 4D-CTA in the detection of cranial arteriovenous shunts
verfasst von
Matthijs in ’t Veld
Rolf Fronczek
Marlise P. dos Santos
Marianne A. A. van Walderveen
Frederick J. A. Meijer
Peter W. A. Willems
Publikationsdatum
14.05.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 11/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06234-4

Weitere Artikel der Ausgabe 11/2019

European Radiology 11/2019 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.