Image intensifier distortion influences a surgeon’s ability to aim guidewires during orthopaedic procedures

verfasst von: Thomas R. Ward, Ben Schwarz, Brian T. N. Le, Geoffrey C. S. Smith, Robert B. Molnar, Paul N. Smith

Erschienen in: Skeletal Radiology | Ausgabe 9/2019

Einloggen, um Zugang zu erhalten

Abstract

Objective

Accurate insertion of a guidewire under image intensifier guidance is a fundamental skill required by orthopaedic surgeons. This study investigated how image intensifier distortion, which is composed of pin-cushion and sigmoidal components, changed the apparent trajectory of a guidewire, and the resulting deviation between the intended and actual guidewire tip position.

Materials and methods

Intraoperative image intensifier images for 220 consecutive patients with hip fractures were retrospectively corrected for distortion using a global polynomial method. The deviation between the intended and actual guidewire tip positions was calculated. Additional distortion parameters were tested using an image intensifier produced by a different manufacturer, and a flat-panel c-arm.

Results

Deviation was approximately 1 cm if the guidewire was aimed from the extremity of the image and almost 0 if the entry point was only 20% from the centre (p < 0.001). The direction of deviation was different for left and right hips, with average deviations measuring 3 mm proximal and 5 mm distal respectively (p < 0.001). The flat-panel c-arm almost completely eliminated distortion.

Conclusions

Image intensifier distortion significantly altered the intended trajectory of a guidewire, with guidewires aimed from the image periphery more affected than guidewires aimed from the centre. Furthermore, for right hips, guidewires should be aimed distal to their intended position, and for left hips they should be aimed proximal to achieve their desired position. The flat-panel c-arm eliminated the effect of distortion; hence, it may be preferable if precision in guidewire positioning is vital.
Literatur
1.
Zurück zum Zitat Froelich JM, Milbrandt JC, Novicoff WM, Saleh KJ, Allan DG. Surgical simulators and hip fractures: a role in residency training? J Surg Educ. 2011;68(4):298–302.CrossRefPubMed Froelich JM, Milbrandt JC, Novicoff WM, Saleh KJ, Allan DG. Surgical simulators and hip fractures: a role in residency training? J Surg Educ. 2011;68(4):298–302.CrossRefPubMed
2.
Zurück zum Zitat Altintas B, Biber R, Bail HJ. The learning curve of proximal femoral nailing. Acta Orthop Traumatol Turc. 2014;48(4):396–400.CrossRefPubMed Altintas B, Biber R, Bail HJ. The learning curve of proximal femoral nailing. Acta Orthop Traumatol Turc. 2014;48(4):396–400.CrossRefPubMed
3.
Zurück zum Zitat Mayman D, Vasarhelyi EM, Long W, Ellis RE, Rudan J, Pichora DR. Computer-assisted guidewire insertion for hip fracture fixation. J Orthop Trauma. 2005;19(9):610–5.CrossRefPubMed Mayman D, Vasarhelyi EM, Long W, Ellis RE, Rudan J, Pichora DR. Computer-assisted guidewire insertion for hip fracture fixation. J Orthop Trauma. 2005;19(9):610–5.CrossRefPubMed
4.
Zurück zum Zitat Bjorgul K, Novicoff WM, Saleh KJ. Learning curves in hip fracture surgery. Int Orthop. 2011;35(1):113–9.CrossRefPubMed Bjorgul K, Novicoff WM, Saleh KJ. Learning curves in hip fracture surgery. Int Orthop. 2011;35(1):113–9.CrossRefPubMed
5.
Zurück zum Zitat Gronenschild E. Correction for geometric image distortion in the x-ray imaging chain: local technique versus global technique. Med Phys. 1999;26(12):2602–16.CrossRefPubMed Gronenschild E. Correction for geometric image distortion in the x-ray imaging chain: local technique versus global technique. Med Phys. 1999;26(12):2602–16.CrossRefPubMed
6.
Zurück zum Zitat Gronenschild E. The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain. Med Phys. 1997;24(12):1875–88.CrossRefPubMed Gronenschild E. The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain. Med Phys. 1997;24(12):1875–88.CrossRefPubMed
7.
Zurück zum Zitat Wallace WA, Johnson F. Detection and correction of geometrical distortion in x-ray fluoroscopic images. J Biomech. 1981;14(2):123–5.CrossRefPubMed Wallace WA, Johnson F. Detection and correction of geometrical distortion in x-ray fluoroscopic images. J Biomech. 1981;14(2):123–5.CrossRefPubMed
8.
Zurück zum Zitat Baltzopoulos V. A videofluoroscopy method for optical distortion correction and measurement of knee-joint kinematics. Clin Biomech (Bristol, Avon). 1995;10(2):85–92.CrossRef Baltzopoulos V. A videofluoroscopy method for optical distortion correction and measurement of knee-joint kinematics. Clin Biomech (Bristol, Avon). 1995;10(2):85–92.CrossRef
9.
Zurück zum Zitat Fantozzi S, Cappello A, Leardini A. A global method based on thin-plate splines for correction of geometric distortion: an application to fluoroscopic images. Med Phys. 2003;30(2):124–31.CrossRefPubMed Fantozzi S, Cappello A, Leardini A. A global method based on thin-plate splines for correction of geometric distortion: an application to fluoroscopic images. Med Phys. 2003;30(2):124–31.CrossRefPubMed
10.
Zurück zum Zitat Nickoloff EL. AAPM/RSNA physics tutorial for residents: physics of flat-panel fluoroscopy systems: survey of modern fluoroscopy imaging: flat-panel detectors versus image intensifiers and more. Radiographics. 2011;31(2):591–602.CrossRefPubMed Nickoloff EL. AAPM/RSNA physics tutorial for residents: physics of flat-panel fluoroscopy systems: survey of modern fluoroscopy imaging: flat-panel detectors versus image intensifiers and more. Radiographics. 2011;31(2):591–602.CrossRefPubMed
11.
Zurück zum Zitat Wong JM, Bewsher S, Yew J, Bucknill A, de Steiger R. Fluoroscopically assisted computer navigation enables accurate percutaneous screw placement for pelvic and acetabular fracture fixation. Injury. 2015;46(6):1064–8.CrossRefPubMed Wong JM, Bewsher S, Yew J, Bucknill A, de Steiger R. Fluoroscopically assisted computer navigation enables accurate percutaneous screw placement for pelvic and acetabular fracture fixation. Injury. 2015;46(6):1064–8.CrossRefPubMed
12.
Zurück zum Zitat Muller MC, Belei P, Pennekamp PH, Kabir K, Wirtz DC, Burger C, et al. Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study. Int Orthop. 2012;36(7):1463–9.CrossRefPubMedPubMedCentral Muller MC, Belei P, Pennekamp PH, Kabir K, Wirtz DC, Burger C, et al. Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study. Int Orthop. 2012;36(7):1463–9.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Gutierrez LF, Ozturk C, McVeigh ER, Lederman RJ. A practical global distortion correction method for an image intensifier based x-ray fluoroscopy system. Med Phys. 2008;35(3):997–1007.CrossRefPubMedPubMedCentral Gutierrez LF, Ozturk C, McVeigh ER, Lederman RJ. A practical global distortion correction method for an image intensifier based x-ray fluoroscopy system. Med Phys. 2008;35(3):997–1007.CrossRefPubMedPubMedCentral
Metadaten
Titel
Image intensifier distortion influences a surgeon’s ability to aim guidewires during orthopaedic procedures
verfasst von
Thomas R. Ward
Ben Schwarz
Brian T. N. Le
Geoffrey C. S. Smith
Robert B. Molnar
Paul N. Smith
Publikationsdatum
21.02.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 9/2019
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-019-03171-7

Weitere Artikel der Ausgabe 9/2019

Skeletal Radiology 9/2019 Zur Ausgabe

Browser’s Notes

Browser’s notes

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.