Impact of network performance on remote robotic-assisted endovascular interventions in porcine model

verfasst von: Peter Legeza, Gavin W. Britz, Alpesh Shah, Kalyna Sconzert, John-Michael Sungur, Ponraj Chinnadurai, Kavya Sinha, Alan B. Lumsden

Erschienen in: Journal of Robotic Surgery | Ausgabe 1/2022

Abstract

Remote robotic-assisted endovascular interventions require real-time control of the robotic system to conduct precise device navigation. The delay (latency) between the input command and the catheter response can be affected by factors such as network speed and distance. This study evaluated the effect of network latency on robotic-assisted endovascular navigation in three vascular beds using in-vivo experimental model. Three operators performed femoral, carotid, and coronary endovascular robotic navigation blinded from the hybrid room with the prototype remote-enabled CorPath GRX system in a porcine model. Navigation was performed to different targets with randomly assigned network latencies from 0 to 1000 ms. Outcome measurements included navigation success, navigation time, perceived lag (1 = imperceptible, 5 = too long), and procedural impact scored by the operators (1 = no impact, 5 = unacceptable). Robotic-assisted remote endovascular navigation was successful in all 65 cases (9 femoral, 38 external carotid, 18 coronary). Guidewire times were not significantly different across the simulated network latency times. Compared to 0 ms added latency, both the procedural impact and perceived lag scores were significantly higher when the added latency was 400 ms or greater (< 0.01). Remote endovascular intervention was feasible in all studied anatomic regions. Network latency of 400 ms or above is perceptible, although acceptable to operators, which suggests that remote robotic-assisted femoral, carotid or coronary arterial interventions should be performed with network latency below 400 ms to provide seamless remote device control.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Robotic-assisted endovascular interventions have multiple advantages compared to conventional, manually performed interventions. They play a significant role in reducing occupational hazards for the operator, such as radiation exposure and orthopedic complications due to heavy lead aprons [1, 2]. With the use of telecommunication, robotic-assisted interventions can be performed from great geographical distances. This method could increase access to care for patients with limited access to appropriate health services, due to geographic barriers, or could allow remote proctoring for certain procedures.
The feasibility of percutaneous coronary intervention (PCI) from long geographic distances has been previously described in both animal models and humans [35], but tele-robotic peripheral and neurovascular intervention has not been demonstrated yet.
During remote robotic-assisted navigation, the operators use joysticks to manipulate the endovascular devices. Real-time response of the system is required to carry out precise navigation between the remote and local sites, which highly depends on network performance. The effect of network latency during coronary artery navigation and PCI has been demonstrated, but no data is available on its effect, when robotically navigating in the peripheral arterial system. This preclinical study aims to evaluate the effect of network latency on robotic-assisted endovascular navigation in coronary, lower extremity and extracranial arteries using an in vivo experimental model.

Materials and methods

The robotic system

The CorPath GRX (Corindus, A Siemens Healthineers Company, Waltham, Massachusetts) is a robotic system, FDA-cleared for coronary and peripheral vascular interventions and CE-marked for neurovascular interventions. The robot allows the operator to deliver and navigate guidewires, rapid exchange devices and guide catheters via joystick control. The system is compatible with commercially available 0.014-inch guidewires and standard rapid exchange (RX) balloons and stent delivery systems. The bedside component includes a single-use sterile cassette, which advances, retracts, and rotates the guidewire and the guide catheter, advances and retracts the RX device. This cassette is attached to a robotic drive that can advance and retract the guide catheter with its forward and backward movements. With the use of the advanced cassette, the system is also capable of manipulating microcatheters. The robotic drive is supported by the extended reach arm, which is mounted on the bedside rail.
The operator sits in a radiation-shielded interventional cockpit and navigates the devices via the control console. The control console consists of a touchscreen and three joysticks for RX device, guidewire and guide catheter manipulation.
The present study utilized a remote prototype modification of the CorPath GRX system, which allowed the physical separation of the remote and local nodes, while connected via the institutional network. The robotic control unit and drive were connected to a target computer (Mobile RT, Speedgoat, Inc., Natick, MA) that each utilized a grandmaster clock and global positioning system antenna for synchronization of the control unit with the robotic drive. In this study the in vivo model, robotic arm, and bedside technician were located in a hybrid interventional suite (“local”), while the operators were navigating the endovascular interventional devices from the control room (“remote”), facing away from the hybrid room (Fig. 1).
The remote workstation includes monitors that display angiographic images, reference images, and hemodynamic data, similar to a typical CorPath GRX case. A telepresence system (LifeSize, Austin, TX) was used to communicate via live audiovisual streams of the operating room, operational field, and remote workstation.
Network latency was simulated by delaying robotic commands in the remote prototype modification of the CorPath GRX System. Simulated latencies ranged from 0 to 1,000 ms and were additive relative to the low (but nonzero) native latency of the institutional network.

Procedural details

The procedure was performed in a domestic cross, female swine weighing 49.1 kg. The animal procedure was conducted under a protocol approved by the Institutional Animal Care and Use Committee.
The animal underwent standard anesthesia procedure by a clinical veterinarian and the access site was prepared and draped. Right common femoral access was gained, and a 6 French sheath (Destination; Terumo Interventional Systems, Somerset, NJ) was placed in the contralateral common femoral artery under fluoroscopy guidance (Zeego; Siemens Healthineers, Malvern, PA). Femoral arterial robotic navigation was performed with a 0.014″ (BMW; Abbott Vascular, Santa Clara, CA) or 0.018″ wire (V18; Boston Scientific, Marlborough, MA) and a 0.027″ microcatheter (Renegade; Boston Scientific, Marlborough, MA). For the external carotid arterial approach, a 7F Envoy guide catheter (90 cm) (Codman, Raynham, MA) was placed in the ostium of the left common carotid artery with a 0.014″ wire (Synchro2, Stryker Neurovascular, Fremont, CA) and a Renegade Hi-Flo microcatheter (0.027″ ID, 135 cm). For coronary arterial navigation, a Q3.5 guide catheter (Boston Scientific, Marlborough, MA) was placed in the ostium of the left coronary artery with a 0.014″ wire.
Angiography was performed in each vascular bed after the sheath was placed in the ostium of the target vessel, and based on the angiography, navigational targets were defined. These targets were a distal deep femoral branch, lingual artery, branch of the facial artery, and the diagonal branch of the left anterior descending coronary artery.

Navigational tasks

A vascular surgeon, a neurosurgeon, and an interventional cardiologist participated, and performed femoral, external carotid and coronary arterial navigation, respectively. Each operator has completed > 20 endovascular robotic cases in both ex-vivo and in-vivo cases. Their task was to robotically navigate the guidewire from the tip of the guide catheter to the preselected targets. After the navigation to a target was completed, the wire was retrieved in the tip of the guide catheter, and the run was repeated.
During different runs, randomly assigned robotic command delays (latencies) were added to the system ranging from 0 to 1000 ms (0, 150, 250, 400, 600, 1000 ms). LAN connection includes an intrinsic command and image delay, which was incremental to the injected latency times. Operators were blinded to latency times. The first four runs were performed with 0 ms added latency, then random latencies were added to each run, up to 19 runs total per navigational target.
After finishing the endovascular navigational tasks, devices were removed, the access site was closed, and the animal was euthanized.

Data analysis

The success of guidewire navigation was defined as the ability to navigate the guidewire from the tip of the guide catheter to the preselected target. For each run, guidewire times were measured. After every run, operators scored the perceived latency and the procedural impact on a scale from 1 to 5. Perceived latency was scored by the operators according to the following scale: 1 = imperceptible; 2 = noticeable but minor; 3 = noticeable; 4 = noticeable and major; 5 = too long. Procedural impact was scored by the interventionalist according to the following scale: 1 = no impact; 2 = minor impact (acceptable performance); 3 = noticeable impact (loss in efficiency, successful outcome); 4 = significant degradation (can complete, but not desired); 5 = unacceptable to complete. Data were analyzed across the three vascular beds, and also separately by making two groups: peripheral (femoral and external carotid) and coronary.
Categorical data are presented as count (n) and percentage (%), continuous variables are summarized as mean and standard deviation (SD). A Kruskal–Wallis test was conducted to determine if wiring times, perceived latency scores and procedural impact scores were different for among the injected latencies. P for trend was obtained from a Wilcoxon-type test for trend to evaluate the trend in perceived latency and procedural impact scores. Results were considered significant when p-value was < 0.05. Post-hoc tests for comparing added latency of 0 ms with added latencies up to 1000 ms were performed with the Man-Whitney test. STATA statistical software (StataCorp LP, College Station, TX, USA) was used for statistical analysis.

Results

Procedural success and guidewire navigation time

All 65 robotic-assisted guidewire navigation attempts with the added latencies from 0 to 1000 ms were successful (9 femoral, 13.8%; 38 external carotid, 58.5%; and 18 coronary, 27.7%). Mean wire navigation time for femoral, carotid target no. 1, carotid target no. 2 and coronary target were 131 ± 84.25, 26.26 ± 29.66, 104.9 ± 84.25 and 70.22 ± 65.18 s, respectively. Across the range of latencies studied, there was no significant difference and no trend observed in navigational times within each vascular bed.

Procedural impact and perceived latency scores

There was a significant trend for higher procedural impact and perceived latency scores across the three vascular beds with the increasing latencies (p = 0.006 and p = 0.002, respectively) (Fig. 2). Across the three vascular beds, a significant difference was shown in the distribution of procedural impact (p = 0.048) and the perceived latency (p = 0.038) scores. When analyzing the scores between the peripheral and the coronary navigation, a tendency of higher scores was seen with the longer added latencies but the difference was not significant (Fig. 3).
For both the procedural impact and perceived lag scores, post-hoc analysis showed no significant differences when comparing the scores between the baseline latency of 0 ms to 150 and 250 ms, but the scores were significantly different when the added latency was 400 ms or greater (p < 0.05)(Tables 1, 2).
Table 1
Mean procedural impact scores with different added command latencies
Added latency (ms)
Procedural impact score
p-value
0
1
N/A
150
1.07 ± 0.7
0.548
250
1.11 ± 0.33
0.55
400
1.55 ± 0.93
0.03
600
1.8 ± 1.23
0.01
1000
1.67 ± 1
0.02
The scores with added latencies were compared to scores with 0 ms added latency
Table 2
Mean perceived latency scores with different added command latencies
Added latency (ms)
Perceived latency score
p-value
0
1
N/A
150
1.14 ± 0.1
0.32
250
1.33 ± 0.5
0.13
400
1.91 ± 1.04
 < 0.01
600
1.9 ± 1.45
0.03
1000
2 ± 1.41
0.02
The scores with added latencies were compared to scores with 0 ms added latency

Discussion

Remote robotic assistance in endovascular surgery allows the operator to perform the intervention even while the operator is physically in a separate location from the patient. Although locally performed robotic-assisted endovascular procedures have been introduced to interventional cardiology, peripheral, and neurovascular surgery [1, 68], previous tele-robotic studies only investigate the feasibility and characteristics of percutaneous coronary interventions [3, 5].
The operating team has to overcome several challenges when performing remote endovascular interventions, including the quality of network connection. Previous studies have shown that high-quality network connection is required for the safe navigation of the surgical devices in real-time, and to provide stable audiovisual connection between the local and remote site. A tele-robotic surgical study determined an acceptable limit of 330 ms time delay [9]. Madder et al. came to the conclusion that latencies during tele-robotic coronary navigation were perceptible for the operator at 400 ms or above [3].
This study investigated the feasibility of remote, robotic-assisted endovascular navigation in femoral, external carotid, and coronary arteries, and assessed the effect of network performance on the success of endovascular navigation. The setting of the study simulated an environment of remote intervention with operators who were connected to the hybrid room through a telepresence system.
Since the operator’s tasks only involved navigation, the arteries selected required complex wire and catheter manipulation by the operator. The 100% success rate of navigation to these preselected targets, even with the added command latencies, is a promising result in terms of precise robotic control. Our current knowledge of robotic-assisted peripheral vascular interventions is based on simple anatomic configurations [7]; however, the successful navigation to all targets in this study suggests that future applications of the robot may include peripheral cases, where complex navigation is needed.
Guidewire navigation time did not show a significant difference when analyzed across the different latency times and different anatomical regions, which may be due to the physician’s ability to compensate for a delay of up to 1000 ms. No data were collected on latencies above 1000 ms, thus we do not know if latencies outside of the tested range influence the guidewire navigational times or not.
Compared to 0 ms added latency, a significant increase in the procedural impact scores at 400, 600 and 1000 ms latency were shown. At these latency values the operators did experience the impact of the network latency during navigation. However, this increase in the scores only indicated a minor effect on their performance of the procedure (Table 1). The perceived latency scores were significantly higher at 400 ms and above. Based on the scores given by the operators, the perception of the latency at these latency values was “noticeable but minor” (Table 2).
When estimating the threshold of the network latency for remote robotic control, even minor noticeable changes in the network quality should be considered. According to the findings of this study the suggested value of seamless robotic control should be below 400 ms.
It is important to outline that our results are only showing the effect of network performance on navigation, and not ballooning or stent deployment. These procedures may require even more harmony between the local and remote sites, so acceptable network latency values may be lower during these maneuvers. Further studies are needed to evaluate the impact of network performance on robotic-assisted interventional procedures.
The limitations of this study are the single porcine model and the repetitions of the same tasks multiple consecutive times. This study only evaluated wire, microcatheter, and guide catheter navigation, but no balloon angioplasty or stenting was performed. The number of navigational runs per targets were too low; therefore, the statistical power was not high enough to see significant results, when analyzing the targets separately. Besides the guidewire navigation time, no other objective measurements were made, including fluoroscopy time and radiation doses.
This study is an important first step in understanding the parameters which are influencing the performance of remote endovascular interventions.

Conclusion

Robotic-assisted femoral, external carotid and coronary navigation are feasible under remote conditions. No influence was demonstrated on the time of guidewire navigation to the target across the range of tested latencies (0–1000 ms). Operators reported a minor impact on their performance with latencies of 400 ms or above. Network latency of 400 ms and above was reported to be perceptible but acceptable. These results suggests that remote robotic-assisted femoral, carotid, or coronary arterial interventions should be performed with network latency below 400 ms to provide seamless remote device control.

Data availibility

Available upon request.

Acknowledgements

The authors would like to thank Rebecca Barnes, Daryl Schulz and Dr. Leslie Jenkins for their help during the study.

Compliance with ethical standards

Conflict of interest

Kalyna Sconzert is an employee of Corindus, a Siemens Healthineers Company; John-Michael Sungur is an employee of Corindus, a Siemens Healthineers Company; Ponraj Chinnadurai is an employee of Siemens Medical Solutions USA Inc; Peter Legeza, Gavin W. Britz, Alpesh Shah, Kavya Shina and Alan B. Lumsden declare that they have no conflict of interest.

Ethical approval

The animal procedure was conducted under a protocol approved by the Institutional Animal Care and Use Committee.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Urologie

Kombi-Abonnement

Mit e.Med Urologie erhalten Sie Zugang zu den urologischen CME-Fortbildungen und Premium-Inhalten der urologischen Fachzeitschriften.

Weitere Produktempfehlungen anzeigen
Literatur
1.
Zurück zum Zitat Weisz G, Metzger DC, Caputo RP et al (2013) Safety and feasibility of robotic percutaneous coronary intervention: PRECISE (percutaneous robotically-enhanced coronary intervention) study. J Am Coll Cardiol 61:1596–1600CrossRef Weisz G, Metzger DC, Caputo RP et al (2013) Safety and feasibility of robotic percutaneous coronary intervention: PRECISE (percutaneous robotically-enhanced coronary intervention) study. J Am Coll Cardiol 61:1596–1600CrossRef
2.
Zurück zum Zitat Madder RD, VanOosterhout S, Mulder A et al (2017) Impact of robotics and a suspended lead suit on physician radiation exposure during percutaneous coronary intervention. Cardiovasc Revascularization Med 18:190–196CrossRef Madder RD, VanOosterhout S, Mulder A et al (2017) Impact of robotics and a suspended lead suit on physician radiation exposure during percutaneous coronary intervention. Cardiovasc Revascularization Med 18:190–196CrossRef
3.
Zurück zum Zitat Madder RD, VanOosterhout S, Mulder A et al (2019) Network latency and long-distance robotic telestenting: exploring the potential impact of network delays on telestenting performance. Catheter Cardiovasc Interv 95:914–919CrossRef Madder RD, VanOosterhout S, Mulder A et al (2019) Network latency and long-distance robotic telestenting: exploring the potential impact of network delays on telestenting performance. Catheter Cardiovasc Interv 95:914–919CrossRef
4.
Zurück zum Zitat Madder RD, VanOosterhout S, Mulder A et al (2019) Feasibility of robotic telestenting over long geographic distances: a pre-clinical ex vivo and in vivo study. EuroIntervention 15:510–512CrossRef Madder RD, VanOosterhout S, Mulder A et al (2019) Feasibility of robotic telestenting over long geographic distances: a pre-clinical ex vivo and in vivo study. EuroIntervention 15:510–512CrossRef
5.
Zurück zum Zitat Patel TM, Shah SC, Pancholy SB (2019) Long distance tele-robotic-assisted percutaneous coronary intervention: a report of first-in-human experience. EClinicalMedicine 14:53–58CrossRef Patel TM, Shah SC, Pancholy SB (2019) Long distance tele-robotic-assisted percutaneous coronary intervention: a report of first-in-human experience. EClinicalMedicine 14:53–58CrossRef
6.
Zurück zum Zitat Caputo R, Lesser A, Simons A (2018) Feasibility of robotic percutaneous renal artery revascularization. J Am Coll Cardiol Intv 8:S35–S36CrossRef Caputo R, Lesser A, Simons A (2018) Feasibility of robotic percutaneous renal artery revascularization. J Am Coll Cardiol Intv 8:S35–S36CrossRef
7.
Zurück zum Zitat Mahmud E, Schmid F, Kalmar P et al (2016) Feasibility and safety of robotic peripheral vascular interventions: results of the RAPID trial. J Am Coll Cardiol Intv 9:2058–2064CrossRef Mahmud E, Schmid F, Kalmar P et al (2016) Feasibility and safety of robotic peripheral vascular interventions: results of the RAPID trial. J Am Coll Cardiol Intv 9:2058–2064CrossRef
8.
Zurück zum Zitat Pereira VM, Cancelliere NM, Nicholson P et al (2020) First-in- human, robotic-assisted neuroendovascular intervention. J Neurointervent Surg 12:338–340CrossRef Pereira VM, Cancelliere NM, Nicholson P et al (2020) First-in- human, robotic-assisted neuroendovascular intervention. J Neurointervent Surg 12:338–340CrossRef
9.
Zurück zum Zitat Marescaux J, Leroy J, Gagner M et al (2001) Transatlantic robot-assisted telesurgery. Nature 413:379–380CrossRef Marescaux J, Leroy J, Gagner M et al (2001) Transatlantic robot-assisted telesurgery. Nature 413:379–380CrossRef
Metadaten
Titel
Impact of network performance on remote robotic-assisted endovascular interventions in porcine model
verfasst von
Peter Legeza
Gavin W. Britz
Alpesh Shah
Kalyna Sconzert
John-Michael Sungur
Ponraj Chinnadurai
Kavya Sinha
Alan B. Lumsden
Publikationsdatum
07.02.2021
Verlag
Springer London
Erschienen in
Journal of Robotic Surgery / Ausgabe 1/2022
Print ISSN: 1863-2483
Elektronische ISSN: 1863-2491
DOI
https://doi.org/10.1007/s11701-021-01196-6

Weitere Artikel der Ausgabe 1/2022

Journal of Robotic Surgery 1/2022 Zur Ausgabe

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Deutlich weniger Infektionen: Wundprotektoren schützen!

08.05.2024 Postoperative Wundinfektion Nachrichten

Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.

Chirurginnen und Chirurgen sind stark suizidgefährdet

07.05.2024 Suizid Nachrichten

Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.