Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CT

verfasst von: Daniele Marin, Jose J. Pratts-Emanuelli, Achille Mileto, Daniela B. Husarik, Mustafa R. Bashir, Rendon C. Nelson, Daniel T. Boll

Erschienen in: European Radiology | Ausgabe 3/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To assess the impact of patient habitus, acquisition parameters, detector efficiencies, and reconstruction techniques on the accuracy of iodine quantification using dual-source dual-energy CT (DECT).

Materials and methods

Two phantoms simulating small and large patients contained 20 iodine solutions mimicking vascular and parenchymal enhancement from saline isodensity to 400 HU and 30 iodine solutions simulating enhancement of the urinary collecting system from 400 to 2,000 HU. DECT acquisition (80/140 kVp and 100/140 kVp) was performed using two DECT systems equipped with standard and integrated electronics detector technologies. DECT raw datasets were reconstructed using filtered backprojection (FBP), and iterative reconstruction (SAFIRE I/V).

Results

Accuracy for iodine quantification was significantly higher for the small compared to the large phantoms (9.2 % ± 7.5 vs. 24.3 % ± 26.1, P = 0.0001), the integrated compared to the conventional detectors (14.8 % ± 20.6 vs. 18.8 % ± 20.4, respectively; P = 0.006), and SAFIRE V compared to SAFIRE I and FBP reconstructions (15.2 % ± 18.1 vs. 16.1 % ± 17.6 and 18.9 % ± 20.4, respectively; P ≤ 0.003). A significant synergism was observed when the most effective detector and reconstruction techniques were combined with habitus-adapted dual-energy pairs.

Conclusion

In a second-generation dual-source DECT system, the accuracy of iodine quantification can be substantially improved by an optimal choice and combination of acquisition parameters, detector, and reconstruction techniques.

Key Points

• Iodine quantification techniques are not immune to error
• Systematic deviations between the measured and true iodine concentrations exist
• Acquisition parameters, detector efficiencies, and reconstruction techniques impact accuracy of iodine quantification
Literatur
1.
Zurück zum Zitat Chiro GD, Brooks RA, Kessler RM et al (1979) Tissue signatures with dual-energy computed tomography. Radiology 131:521–523CrossRefPubMed Chiro GD, Brooks RA, Kessler RM et al (1979) Tissue signatures with dual-energy computed tomography. Radiology 131:521–523CrossRefPubMed
2.
Zurück zum Zitat Cann CE, Gamsu G, Birnberg FA, Webb WR (1982) Quantification of calcium in solitary pulmonary nodules using single- and dual-energy CT. Radiology 145:493–496CrossRefPubMed Cann CE, Gamsu G, Birnberg FA, Webb WR (1982) Quantification of calcium in solitary pulmonary nodules using single- and dual-energy CT. Radiology 145:493–496CrossRefPubMed
3.
Zurück zum Zitat Genant HK, Boyd D (1977) Quantitative bone mineral analysis using dual energy computed tomography. Invest Radiol 12:545–551CrossRefPubMed Genant HK, Boyd D (1977) Quantitative bone mineral analysis using dual energy computed tomography. Invest Radiol 12:545–551CrossRefPubMed
4.
Zurück zum Zitat Schmidt-Bindert G, Henzler T, Chu TQ et al (2012) Functional imaging of lung cancer using dual energy CT: how does iodine related attenuation correlate with standardized uptake value of 18FDG-PET-CT? Eur Radiol 22:93–103CrossRef Schmidt-Bindert G, Henzler T, Chu TQ et al (2012) Functional imaging of lung cancer using dual energy CT: how does iodine related attenuation correlate with standardized uptake value of 18FDG-PET-CT? Eur Radiol 22:93–103CrossRef
5.
Zurück zum Zitat Apfaltrer P, Meyer M, Meier C et al (2012) Contrast-enhanced dual-energy CT of gastrointestinal stromal tumors. Is iodine-related attenuation a potential indicator of tumor response? Invest Radiol 47:65–70CrossRefPubMed Apfaltrer P, Meyer M, Meier C et al (2012) Contrast-enhanced dual-energy CT of gastrointestinal stromal tumors. Is iodine-related attenuation a potential indicator of tumor response? Invest Radiol 47:65–70CrossRefPubMed
6.
Zurück zum Zitat Ascenti G, Mileto A, Krauss B et al (2013) Distinguishing enhancing from nonenhancing renal masses with dual-source dual-energy CT: iodine quantification versus standard enhancement measurements. Eur Radiol 23:2288–2295CrossRefPubMed Ascenti G, Mileto A, Krauss B et al (2013) Distinguishing enhancing from nonenhancing renal masses with dual-source dual-energy CT: iodine quantification versus standard enhancement measurements. Eur Radiol 23:2288–2295CrossRefPubMed
7.
Zurück zum Zitat Chandarana H, Megibow AJ, Cohen BA et al (2011) Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. AJR Am J Roentgenol 196:W693–W700CrossRefPubMed Chandarana H, Megibow AJ, Cohen BA et al (2011) Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. AJR Am J Roentgenol 196:W693–W700CrossRefPubMed
8.
Zurück zum Zitat Koonce JD, Vliegenthart R, Schoepf UJ et al (2014) Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model. Eur Radiol 24:512–518CrossRefPubMedCentralPubMed Koonce JD, Vliegenthart R, Schoepf UJ et al (2014) Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model. Eur Radiol 24:512–518CrossRefPubMedCentralPubMed
9.
Zurück zum Zitat Feuerlein S, Heye TJ, Bashir MR, Boll DT (2012) Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Invest Radiol 47:656–661CrossRefPubMed Feuerlein S, Heye TJ, Bashir MR, Boll DT (2012) Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Invest Radiol 47:656–661CrossRefPubMed
11.
Zurück zum Zitat Duan X, Wang J, Leng S, Schmidt B, Allmendinger T, Grant K, Flohr T, McCollough CH (2013) Electronic noise in CT detectors: Impact on image noise and artifacts. AJR Am J Roentgenol 201:W626–632CrossRefPubMed Duan X, Wang J, Leng S, Schmidt B, Allmendinger T, Grant K, Flohr T, McCollough CH (2013) Electronic noise in CT detectors: Impact on image noise and artifacts. AJR Am J Roentgenol 201:W626–632CrossRefPubMed
12.
Zurück zum Zitat Primak AN, Giraldo JC, Eusemann CD et al (2010) Dual-source dual-energy CT with additional tin filtration: dose and image quality evaluation in phantoms and in vivo. AJR Am J Roentgenol 195:1164–1174CrossRefPubMedCentralPubMed Primak AN, Giraldo JC, Eusemann CD et al (2010) Dual-source dual-energy CT with additional tin filtration: dose and image quality evaluation in phantoms and in vivo. AJR Am J Roentgenol 195:1164–1174CrossRefPubMedCentralPubMed
13.
Zurück zum Zitat Pontana F, Pagniez J, Duhamel A et al (2013) Reduced-dose low-voltage chest CT angiography with sinogram-affirmed iterative reconstruction versus standard-dose filtered back projection. Radiology 267:609–618CrossRefPubMed Pontana F, Pagniez J, Duhamel A et al (2013) Reduced-dose low-voltage chest CT angiography with sinogram-affirmed iterative reconstruction versus standard-dose filtered back projection. Radiology 267:609–618CrossRefPubMed
14.
Zurück zum Zitat Li Y, Shi G, Wang S, Wang S, Wu R (2013) Iodine quantification with dual-energy CT: phantom study and preliminary experience with VX2 residual tumour in rabbits after radiofrequency ablation. Br J Radiol 86:20130143CrossRefPubMedCentralPubMed Li Y, Shi G, Wang S, Wang S, Wu R (2013) Iodine quantification with dual-energy CT: phantom study and preliminary experience with VX2 residual tumour in rabbits after radiofrequency ablation. Br J Radiol 86:20130143CrossRefPubMedCentralPubMed
15.
Zurück zum Zitat Morsbach F, Bickelhaupt S, Rätzer S, Schmidt B, Alkadhi H (2014) Integrated circuit detector technology in abdominal CT: added value in obese patients. AJR Am J Roentgenol 202:368–374CrossRefPubMed Morsbach F, Bickelhaupt S, Rätzer S, Schmidt B, Alkadhi H (2014) Integrated circuit detector technology in abdominal CT: added value in obese patients. AJR Am J Roentgenol 202:368–374CrossRefPubMed
16.
Zurück zum Zitat Primak AN, Ramirez Giraldo JC, Liu X, Yu L, McCollough CH (2009) Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration. Med Phys 36:1359–1369CrossRefPubMedCentralPubMed Primak AN, Ramirez Giraldo JC, Liu X, Yu L, McCollough CH (2009) Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration. Med Phys 36:1359–1369CrossRefPubMedCentralPubMed
18.
Zurück zum Zitat Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15CrossRefPubMed Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15CrossRefPubMed
Metadaten
Titel
Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CT
verfasst von
Daniele Marin
Jose J. Pratts-Emanuelli
Achille Mileto
Daniela B. Husarik
Mustafa R. Bashir
Rendon C. Nelson
Daniel T. Boll
Publikationsdatum
01.03.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 3/2015
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3447-8

Weitere Artikel der Ausgabe 3/2015

European Radiology 3/2015 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.