Effect of Voxel Size on 3D Micro-CT Analysis of Cortical Bone Porosity

verfasst von: David Cooper, Andrei Turinsky, Christoph Sensen, Benedikt Hallgrimsson

Erschienen in: Calcified Tissue International | Ausgabe 3/2007

Einloggen, um Zugang zu erhalten

Abstract

This study examines the impact of voxel size on 3D micro-CT analysis of human cortical bone porosity. The study is based on computed microtomography scans of 10 human anterior femoral midshaft specimens acquired at 5, 10, and 15 μm voxel sizes. Artificial voxel sizes (10, 20, and 40 μm) were generated from the smallest scan voxel size (5 μm) in order to compare actual scanning with artificial degradation, a method employed in other similar studies. Canal volume fraction (CaV/TV), canal surface to volume ratio (CaS/CaV), mean canal diameter (CaDm), mean canal separation (CaSp), canal number (CaN), degree of anisotropy (DA), and canal connectivity density (CaConnD) were calculated from matching volumes of interest for all datasets. Qualitatively, the clarity of the actual scan datasets deteriorated rapidly as voxel size increased. In contrast, within the artificially generated datasets, the clarity of cortical pores was better maintained until the largest voxel size (40 μm). Mean absolute percent error values, correlation coefficients, and paired t-tests revealed a pattern of increasing, and generally significant, differences between the smallest and progressively larger voxel sizes (both scanned and artificial). Relative to the actual scans, however, the artificial datasets were less sensitive to changing voxel size. These findings indicated that subtle changes in voxel size, within the range examined, have a considerable effect on human cortical porosity structural parameters. Additionally, the use of artificially increased voxel sizes should be viewed with caution as they may not reflect what can actually be obtained by scanning.
Literatur
1.
Zurück zum Zitat Cooper DM, Matyas JR, Katzenberg MA, Hallgrimsson B (2004) Comparison of microcomputed tomographic and microradiographic measurements of cortical bone porosity. Calcif Tissue Int 74:437–447PubMedCrossRef Cooper DM, Matyas JR, Katzenberg MA, Hallgrimsson B (2004) Comparison of microcomputed tomographic and microradiographic measurements of cortical bone porosity. Calcif Tissue Int 74:437–447PubMedCrossRef
2.
Zurück zum Zitat Cooper DM, Turinsky AL, Sensen CW, Hallgrimsson B (2003) Quantitative 3D analysis of the canal network in cortical bone by micro-computed tomography. Anat Rec 274B:169–179CrossRef Cooper DM, Turinsky AL, Sensen CW, Hallgrimsson B (2003) Quantitative 3D analysis of the canal network in cortical bone by micro-computed tomography. Anat Rec 274B:169–179CrossRef
3.
Zurück zum Zitat Basillais A, Chappard C, Bensamoun S, Brunet-Imbault B, Ho Ba Tho M, Benhamou C (2004) Three Dimensional Characterization of Cortical Bone Porosity on Microcomputed Tomography Images: A New Method of Evaluation. J Bone Miner Res 19:s112–113 Basillais A, Chappard C, Bensamoun S, Brunet-Imbault B, Ho Ba Tho M, Benhamou C (2004) Three Dimensional Characterization of Cortical Bone Porosity on Microcomputed Tomography Images: A New Method of Evaluation. J Bone Miner Res 19:s112–113
4.
Zurück zum Zitat Bousson V, Peyrin F, Bergot C, Hausard M, Sautet A, Laredo JD (2004) Cortical bone in the human femoral neck: three-dimensional appearance and porosity using synchrotron radiation. J Bone Miner Res 19:794–801PubMedCrossRef Bousson V, Peyrin F, Bergot C, Hausard M, Sautet A, Laredo JD (2004) Cortical bone in the human femoral neck: three-dimensional appearance and porosity using synchrotron radiation. J Bone Miner Res 19:794–801PubMedCrossRef
5.
Zurück zum Zitat Jones AC, Milthorpe B, Averdunk H, Limaye A, Senden TJ, Sakellariou A, Sheppard AP, Sok RM, Knackstedt MA, Brandwood A, Rohner D, Hutmacher DW (2004) Analysis of 3D bone ingrowth into polymer scaffolds via micro-computed tomography imaging. Biomaterials 25:4947–4954PubMedCrossRef Jones AC, Milthorpe B, Averdunk H, Limaye A, Senden TJ, Sakellariou A, Sheppard AP, Sok RM, Knackstedt MA, Brandwood A, Rohner D, Hutmacher DW (2004) Analysis of 3D bone ingrowth into polymer scaffolds via micro-computed tomography imaging. Biomaterials 25:4947–4954PubMedCrossRef
6.
Zurück zum Zitat Wachter NJ, Augat P, Krischak GD, Mentzel M, Kinzl L, Claes L (2001a) Prediction of cortical bone porosity in vitro by microcomputed tomography. Calcif Tissue Int 68:38–42CrossRef Wachter NJ, Augat P, Krischak GD, Mentzel M, Kinzl L, Claes L (2001a) Prediction of cortical bone porosity in vitro by microcomputed tomography. Calcif Tissue Int 68:38–42CrossRef
7.
Zurück zum Zitat Wachter NJ, Augat P, Krischak GD, Sarkar MR, Mentzel M, Kinzl L, Claes L (2001b) Prediction of strength of cortical bone in vitro by microcomputed tomography. Clin Biomech (Bristol, Avon) 16:252–256CrossRef Wachter NJ, Augat P, Krischak GD, Sarkar MR, Mentzel M, Kinzl L, Claes L (2001b) Prediction of strength of cortical bone in vitro by microcomputed tomography. Clin Biomech (Bristol, Avon) 16:252–256CrossRef
8.
Zurück zum Zitat Schneider P, Stampanoni M, Donahue L, Muller R (2004) Bone Vasculature as a New Marker for Bone Quality Varies within the Femoral Diaphysis of Two Genetically Distinct Mouse Strains. J Bone Miner Res 19:s241 Schneider P, Stampanoni M, Donahue L, Muller R (2004) Bone Vasculature as a New Marker for Bone Quality Varies within the Femoral Diaphysis of Two Genetically Distinct Mouse Strains. J Bone Miner Res 19:s241
9.
Zurück zum Zitat Matsumoto T, Yoshino M, Asano T, Uesugi K, Todoh M, Tanaka M (2006) Monochromatic synchrotron radiation {micro}CT reveals disuse-mediated canal network rarefaction in cortical bone of growing rat tibiae. J Appl Physiol 100:274–280PubMedCrossRef Matsumoto T, Yoshino M, Asano T, Uesugi K, Todoh M, Tanaka M (2006) Monochromatic synchrotron radiation {micro}CT reveals disuse-mediated canal network rarefaction in cortical bone of growing rat tibiae. J Appl Physiol 100:274–280PubMedCrossRef
10.
Zurück zum Zitat Muller R, Koller B, Hildebrand T, Laib A, Gianolini S, Ruegsegger P (1996) Resolution dependency of microstructural properties of cancellous bone based on three-dimensional mu-tomography. Technol Health Care 4:113–119PubMed Muller R, Koller B, Hildebrand T, Laib A, Gianolini S, Ruegsegger P (1996) Resolution dependency of microstructural properties of cancellous bone based on three-dimensional mu-tomography. Technol Health Care 4:113–119PubMed
11.
Zurück zum Zitat Peyrin F, Salome M, Cloetens P, Laval-Jeantet AM, Ritman E, Ruegsegger P (1998) Micro-CT examinations of trabecular bone samples at different resolutions: 14, 7 and 2 micron level. Technol Health Care 6:391–401PubMed Peyrin F, Salome M, Cloetens P, Laval-Jeantet AM, Ritman E, Ruegsegger P (1998) Micro-CT examinations of trabecular bone samples at different resolutions: 14, 7 and 2 micron level. Technol Health Care 6:391–401PubMed
12.
Zurück zum Zitat Kothari M, Keaveny TM, Lin JC, Newitt DC, Genant HK, Majumdar S (1998) Impact of spatial resolution on the prediction of trabecular architecture parameters. Bone 22:437–443PubMedCrossRef Kothari M, Keaveny TM, Lin JC, Newitt DC, Genant HK, Majumdar S (1998) Impact of spatial resolution on the prediction of trabecular architecture parameters. Bone 22:437–443PubMedCrossRef
13.
Zurück zum Zitat Tabor Z (2004) Analysis of the influence of image resolution on the discriminating power of trabecular bone architectural parameters. Bone 34:170–179PubMedCrossRef Tabor Z (2004) Analysis of the influence of image resolution on the discriminating power of trabecular bone architectural parameters. Bone 34:170–179PubMedCrossRef
14.
Zurück zum Zitat Majumdar S, Newitt D, Mathur A, Osman D, Gies A, Chiu E, Lotz J, Kinney J, Genant H (1996) Magnetic resonance imaging of trabecular bone structure in the distal radius: relationship with X-ray tomographic microscopy and biomechanics. Osteoporos Int 6:376–385PubMedCrossRef Majumdar S, Newitt D, Mathur A, Osman D, Gies A, Chiu E, Lotz J, Kinney J, Genant H (1996) Magnetic resonance imaging of trabecular bone structure in the distal radius: relationship with X-ray tomographic microscopy and biomechanics. Osteoporos Int 6:376–385PubMedCrossRef
15.
Zurück zum Zitat Vieth V, Link TM, Lotter A, Persigehl T, Newitt D, Heindel W, Majumdar S (2001) Does the trabecular bone structure depicted by high-resolution MRI of the calcaneus reflect the true bone structure? Invest Radiol 36:210–217PubMedCrossRef Vieth V, Link TM, Lotter A, Persigehl T, Newitt D, Heindel W, Majumdar S (2001) Does the trabecular bone structure depicted by high-resolution MRI of the calcaneus reflect the true bone structure? Invest Radiol 36:210–217PubMedCrossRef
16.
Zurück zum Zitat Pothuaud L, Laib A, Levitz P, Benhamou CL, Majumdar S (2002) Three-dimensionalline skeleton graph analysis of high-resolution magnetic resonance images: a validation study from 34–microm-resolution microcomputed tomography. J Bone Miner Res 17:1883–1895PubMedCrossRef Pothuaud L, Laib A, Levitz P, Benhamou CL, Majumdar S (2002) Three-dimensionalline skeleton graph analysis of high-resolution magnetic resonance images: a validation study from 34–microm-resolution microcomputed tomography. J Bone Miner Res 17:1883–1895PubMedCrossRef
17.
Zurück zum Zitat Last D, Peyrin F, Guillot G (2005) Accuracy of 3D MR microscopy for trabecular bone assessment: a comparative study on calcaneus samples using 3D synchrotron radiation microtomography. Magma 18:26–34PubMedCrossRef Last D, Peyrin F, Guillot G (2005) Accuracy of 3D MR microscopy for trabecular bone assessment: a comparative study on calcaneus samples using 3D synchrotron radiation microtomography. Magma 18:26–34PubMedCrossRef
18.
Zurück zum Zitat Kim DG, Christopherson GT, Dong XN, Fyhrie DP, Yeni YN (2004) The effect of microcomputed tomography scanning and reconstruction voxel size on the accuracy of stereological measurements in human cancellous bone. Bone 35:1375–1382PubMedCrossRef Kim DG, Christopherson GT, Dong XN, Fyhrie DP, Yeni YN (2004) The effect of microcomputed tomography scanning and reconstruction voxel size on the accuracy of stereological measurements in human cancellous bone. Bone 35:1375–1382PubMedCrossRef
19.
Zurück zum Zitat Feldkamp LA, Davis LC, Kress JW (1984) Practical cone-beam algorithm. J Opt Soc Am 1:612–619CrossRef Feldkamp LA, Davis LC, Kress JW (1984) Practical cone-beam algorithm. J Opt Soc Am 1:612–619CrossRef
21.
Zurück zum Zitat Borah B, Gross GJ, Dufresne TE, Smith TS, Cockman MD, Chmielewski PA, Lundy MW, Hartke JR, Sod EW (2001) Three-dimensional microimaging (MRmicroI and microCT), finite element modeling, and rapid prototyping provide unique insights into bone architecture in osteoporosis. Anat Rec 265:101–110PubMedCrossRef Borah B, Gross GJ, Dufresne TE, Smith TS, Cockman MD, Chmielewski PA, Lundy MW, Hartke JR, Sod EW (2001) Three-dimensional microimaging (MRmicroI and microCT), finite element modeling, and rapid prototyping provide unique insights into bone architecture in osteoporosis. Anat Rec 265:101–110PubMedCrossRef
22.
Zurück zum Zitat Lin AS, Barrows TH, Cartmell SH, Guldberg RE (2003) Microarchitectural and mechanical characterization of oriented porous polymer scaffolds. Biomaterials 24:481–489PubMedCrossRef Lin AS, Barrows TH, Cartmell SH, Guldberg RE (2003) Microarchitectural and mechanical characterization of oriented porous polymer scaffolds. Biomaterials 24:481–489PubMedCrossRef
23.
Zurück zum Zitat Tabachnick BG, Fidell LS (2001) Using Multivariate Statistics. Allyn and Bacon, Toronto Tabachnick BG, Fidell LS (2001) Using Multivariate Statistics. Allyn and Bacon, Toronto
24.
Zurück zum Zitat Van de Casteele E, Van Dyck D, Sijbers J, Raman E (2004) The effect of beam hardening on resolution in X-ray microtomography. SPIE 5370:2089–2096CrossRef Van de Casteele E, Van Dyck D, Sijbers J, Raman E (2004) The effect of beam hardening on resolution in X-ray microtomography. SPIE 5370:2089–2096CrossRef
25.
Zurück zum Zitat Majumdar S, Newitt D, Jergas M, Gies A, Chiu E, Osman D, Keltner J, Keyak J, Genant H (1995) Evaluation of technical factors affecting the quantification of trabecular bone structure using magnetic resonance imaging. Bone 17:417–430PubMedCrossRef Majumdar S, Newitt D, Jergas M, Gies A, Chiu E, Osman D, Keltner J, Keyak J, Genant H (1995) Evaluation of technical factors affecting the quantification of trabecular bone structure using magnetic resonance imaging. Bone 17:417–430PubMedCrossRef
26.
Zurück zum Zitat Bernhardt R, Scharnweber D, Muller B, Thurner P, Schliephake H, Wyss P, Beckmann F, Goebbels J, Worch H (2004) Comparison of microfocus- and synchrotron X-ray tomography for the analysis of osteointegration around Ti6Al4V implants. Eur Cell Mater 7:42–51PubMed Bernhardt R, Scharnweber D, Muller B, Thurner P, Schliephake H, Wyss P, Beckmann F, Goebbels J, Worch H (2004) Comparison of microfocus- and synchrotron X-ray tomography for the analysis of osteointegration around Ti6Al4V implants. Eur Cell Mater 7:42–51PubMed
27.
Zurück zum Zitat Hwang SN, Wehrli FW (2002) Subvoxel processing: a method for reducing partial volume blurring with application to in vivo MR images of trabecular bone. Magn Reson Med 47:948–957PubMedCrossRef Hwang SN, Wehrli FW (2002) Subvoxel processing: a method for reducing partial volume blurring with application to in vivo MR images of trabecular bone. Magn Reson Med 47:948–957PubMedCrossRef
Metadaten
Titel
Effect of Voxel Size on 3D Micro-CT Analysis of Cortical Bone Porosity
verfasst von
David Cooper
Andrei Turinsky
Christoph Sensen
Benedikt Hallgrimsson
Publikationsdatum
01.03.2007
Erschienen in
Calcified Tissue International / Ausgabe 3/2007
Print ISSN: 0171-967X
Elektronische ISSN: 1432-0827
DOI
https://doi.org/10.1007/s00223-005-0274-6

Weitere Artikel der Ausgabe 3/2007

Calcified Tissue International 3/2007 Zur Ausgabe

Clinical Investigations

Burden of Hip Fracture in Iran

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Positiver FIT: Die Ursache liegt nicht immer im Dickdarm

27.05.2024 Blut im Stuhl Nachrichten

Immunchemischer Stuhltest positiv, Koloskopie negativ – in solchen Fällen kann die Blutungsquelle auch weiter proximal sitzen. Ein Forschungsteam hat nachgesehen, wie häufig und in welchen Lokalisationen das der Fall ist.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

Mehr Lebenszeit mit Abemaciclib bei fortgeschrittenem Brustkrebs?

24.05.2024 Mammakarzinom Nachrichten

In der MONARCHE-3-Studie lebten Frauen mit fortgeschrittenem Hormonrezeptor-positivem, HER2-negativem Brustkrebs länger, wenn sie zusätzlich zu einem nicht steroidalen Aromatasehemmer mit Abemaciclib behandelt wurden; allerdings verfehlte der numerische Zugewinn die statistische Signifikanz.

ADT zur Radiatio nach Prostatektomie: Wenn, dann wohl länger

24.05.2024 Prostatakarzinom Nachrichten

Welchen Nutzen es trägt, wenn die Strahlentherapie nach radikaler Prostatektomie um eine Androgendeprivation ergänzt wird, hat die RADICALS-HD-Studie untersucht. Nun liegen die Ergebnisse vor. Sie sprechen für länger dauernden Hormonentzug.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.