Overlengthening of the radial column in radial head replacement: a review of the literature and presentation of a classification system

verfasst von: K. Wegmann, M. Hackl, T. Leschinger, K. J. Burkhart, L. P. Müller

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 9/2021

Abstract

Background

Radial head arthroplasty is a common procedure in elbow surgery. It has been shown to be of benefit for the patients, but there also are relevant complications that should be prevented if possible. One significant complication is overlengthening of the radial head prosthesis. In overlengthening, the head of the prosthesis overextends the physiological level of the native radial head and leads to overcompression in the radiohumeral joint. Rapid erosion and arthritic changes may then impede the clinical outcome. The incidence of overlengthening is not precisely known, but estimations range to up to 20% of all implanted prostheses.

Methods

The present review discusses the available body of literature on overlengthening and lines out a classification system that may be used to guide treatment algorithms. The classification is based on the personal experiences of the author during their clinical practice.

Results

In low-grade overlengthening (type I) conservative treatment can be an option. In Types II–IV usually revision surgery is needed. Depending on the state of the capitulum and joint stability, it is possible re-implant a prosthesis, or rely on implant removal alone.

Discussion

The present review aimed at shedding light into overlengthening as a complication radial head replacement and to help identify and treat it.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Prosthetic implants are commonly in use for the treatment of pathologies of the proximal radius and have been investigated in several clinical studies [14]. The safety of the devices and the positive effects on pain, joint stability and load sharing across the joint and the forearm have been well documented. Despite the good impact the introduction and ongoing engineering of the implants had on patient treatment, specific complications related to radial head replacement are well known. Besides shaft-loosening and even uncoupling of prosthesis components, erosion of the capitulum is a common downside when the radial head is replaced. Erosion of the capitulum may lead to pain and stiffness with impaired functionality of the joint. One reason for the erosion of the capitulum, undoubtedly, is the difference in elasticity modules of the native capitulum and the prosthetic material. The amount of erosion can be greatly increased, however, when the pressure between prosthesis and capitulum exceeds physiological limits. This typically is the case in overlengthening of the proximal radius by the implants. Overlengthening occurs either by imperfect implantation, or if longitudinal stability of the forearm is disturbed. Needless to say, the latter is less frequently present, like in Essex-Lopresti lesions. In the case of increased pressure between the radial column and a radial head prosthesis due to longitudinal instability of the forearm, we use the term dynamic overlengthening. Even if a prosthesis of the radial head is implanted correctly, due to radioulnar longitudinal shifting, the prosthesis will abut against the capitulum and erode it. The second reason for overlengthening as mentioned is imperfect implantation. Either there is insufficient resection of bone of the proximal radius before implanting the device or the length of the implant is chosen as such, that the final height of the construct exceeds the height of the resected bone or the traumatic bony defect. Both scenarios lead to an increase in pressure within in the radio-humeral joint resulting in rapid erosion of the capitulum. This condition usually leads to severely impaired function of the joint [59]. The clinical study of Schnetzke et al. also showed that overlengthening comes along with impaired clinical outcome [10].
Aside from the term “overlengthening”, this pathology is frequently referred to as “overstuffing” in the available literature. But while overlengthening clearly describes the biomechanical problem at the radio-humeral joint, overstuffing may also be present if the diameter of an implant is too large but sitting at the correct height. Thus, we prefer the term overlengthening, as it more clearly separates the two entities.
As overlengthening is not well investigated and its role in the outcome of radial head prosthetic replacement has not been finally clarified, the present review aims to shed light into its pathomechanics, diagnostics and therapeutic options. In addition, a classification system is presented that shall help to stratify the entity and guide its treatment.

Epidemiology

The precise incidence of overlengthening is unknown, as it has not been investigated in larger clinical series. Still, the present literature has recognized it as a detrimental complication of radial head arthroplasty. Besides loosening and component dissociation, van Riet et al. found 11 cases with signs of overlengthening in their case series of 47 radial head prostheses, making up for 23% [8]. The group stated that clinical causes of failure like stiffness and pain might well be associated with mal-implantation like overlengthening. In another retrospective report, Hackl et al. investigated the cases of 466 patients, who had suffered fractures of the radial head and had to undergo revision surgery for the initial treatment [11]. In a subgroup of 76 patients who had been initially treated with prosthetic radial head replacement, the authors found radiological evidence of overlengthening in 16 patients, making up for 21%. In a systematic review on the general topic of radial head prosthetic replacement, Heijink et al. included 30 studies reporting on 727 patients and reported that in 56 patients revision was necessary, with again 20% being revised for overlengthening. Kachooei et al. reviewed 30 studies including an overall of 1017 patients who had a radial head prosthesis implanted [12]. Out of these patients in 122 the implant was removed or revised, with overlengthening being the cause in 13% of the cases. At hand of these studies, a precise estimation of the incidence of overlengthening of radial head prostheses is not reliably possible. But the reported numbers show that today, overlengthening most likely is not a rare condition, being at least responsible for 13–23% of radial revisions. And likely it is an underreported condition, as asymptomatic or undiagnosed patients may be present.

Pathobiomechanics

The radial head is not only a significant valgus stabilizer, but also important for longitudinal force transmission, from the wrist through the forearm into the distal humerus. In the healthy distal forearm, ~ 70% of the weight picked up by the wrist is transferred into the distal radius [13]. While the forces travel in a proximal direction through the forearm, the interosseus membrane aids in shifting the load from the radius to the ulna. By that shift, the load distribution at the level of the elbow joint is almost equal for the radial and the ulnar joint compartment [13, 14]. This biomechanical relation is based on the complex anatomy of the joint, where the greater sigmoid notch of the ulna and the radial head articulate intimately with their respective counterparts of the distal humerus, the trochlea and the capitulum. In the native joint, the articular surface of the radial head sits at the level of the lateral edge of the coronoid process, allowing it to have medium contact pressures of 1 to about 1.75 mPa, according to the in vitro study of Cohn et al. [15]. Yet, any changes to that balance result in significant increases in the pressure distribution. In their study, Cohn et al. put the specimens under a load of 4.5 kg, each via biceps/brachialis- and triceps-loading [15]. Overlengthening of the radial column of more than 2 mm led to a significant increase of the contact pressure between the radial head and the capitulum, according to the authors. These in-vitro results comprehensibly explain the clinically observed changes to the capitulum in a joint with an overlengthened prosthesis, where the implant grinds its way into the capitulum (Fig. 1). However, also the contact mechanics between the cartilage and the radial head implant is of relevance. Implantation of a radial head prosthesis means hemi-prosthetic replacement of the joint. Hence, the mechanical properties of the two joint components differ significantly, in terms of their elasticity modulus. While Cobalt/Chrome (e.g., CRF II, Tornier) has a young’s modulus of 210 GPa and Pyrocarbon (e.g., Mopyc, Tornier) of about 25 GPa, articular cartilage has a young’s modulus of 0.5–0.9 MPa and is with that several thousand times lower. By that, erosion of the capitulum is likely to happen anyhow, as a harder material will always grind into the softer material. The study of Burkhart et al. showed a high number of capitellar erosions at an average of 8 years after radial head replacement, even in absence of overlengthening [16]. We know from several biomechanical studies that, besides the differing young’s modulus, compared to the physiological state the contact area between the native capitulum and the radial side is significantly reduced, when commonly available radial head prostheses are implanted [1721]. As a logical consequence, reduced contact area and a discrepancy in young’s modulus can well lead to abrasion, with its clinical consequences. However, if these two factors are combined with increased load, as is the fact in overlengthening, peak contact pressure per area will increase drastically.

Diagnostics and classification

Clinical signs

Patients may present in varying intervals after radial head replacement and report pain or discomfort, concentrated over the lateral side of the elbow. In severe overlengthening, with closure of the medial joint space, even medial sided pain may be present. Also, crepitus is often noticed. The investigator can reproduce crepitus by rotating the forearm, therewith exerting rotational stress with the prosthesis on the capitulum. This may also lead to provocation of pain; hence the manoeuvre should be performed carefully and with respect to the patient’s symptoms. Both, crepitus and pain can be increased by contraction of the forearm-muscles, e.g. by firm fist-closure. This muscle contraction leads to an increase of the compressive force between capitulum and prosthesis with likely enhancement of the symptoms.
The patients often also present with restricted range of motion in the elbow. Extension/flexion and pronation/supination may be limited in varying degrees. Several factors contribute to the problem. First, as overlengthening often results in erosion of the capitulum, it contributes to joint inflammation. Inflammatory processes usually already are present because of the surgical intervention with implantation of the prosthesis. The inflammation often ends up in scarring of the joint and therewith stiffness. As a second effect of overlengthening, the increased pressure at the capitulum is able to limit rotation by the mechanical effect of increased friction. Same is possible for a flexion deficit, when the overlengthened implant abuts at the ventral aspect of the capitulum in flexion, leading to a mechanical block. This entity was already observed and reported by Birkedahl et al., who stated that the loss of the radio-capitellar joint gap leads to a closure and an abutment in flexion [5]. As a third effect, the pain itself may prevent patients from moving adequately and from proper rehab, by that resulting in stiffness, too.
But the clinical effects of overlengthening are not solely confined to the elbow itself. By the direct biomechanical cohesions between the elbow and the wrist, it is reasonable that overlengthening of the radial column affects the distal forearm with potential negative clinical consequences. With significant radial lengthening the patients might have pain due to the altered radio-ulnar variance. Clinically this may lead to tensioning pain throughout the forearm and constant pain over the DRUJ. Also shooting pain over the DRUJ or wrist is reported by the patients.

Imaging

The basis of imaging is plain x-rays. The elbow must be filmed in two orthogonal planes, a.p. and strictly lateral. Lateral images are best achieved with the patient sitting and the elbow placed on the table, when the shoulder is abducted at exactly 90°. Therefore, the X-ray table or film must be lifted, but then strictly lateral images are achieved in a reliable fashion. On lateral images, besides overlengthening, joint stability can be evaluated. Joint subluxation can be detected by the so-called “drop-sign” (Fig. 8, black arrow) in the humeroulnar joint cavity. It points towards postero-lateral instability of the ulno-humeral joint, with the displacement of the radial head past the diameter of the capitellum in lateral radiographs (Fig. 8, white circle/line). Also, the ulna “drops” away from the trochlea of the humerus, indicating ulno-humeral instability. When looking for overlengthening, the X-rays should initially be investigated regarding the radial head prosthesis itself. It is of relevance if the implant is firmly fixed or if there are signs of loosening (Fig. 2). An important sign of loosening is osteolysis around the implant and cortical thinning at the proximal radius, which can happen by chronic movement of the shaft of the prosthesis in the radius. Fractures of the shaft by bone loss due to loosening may be present but may also be subtle and should not be overlooked. Also, the remaining length of the bonestock of the proximal radius is of interest for revision surgery.
Resorption must be evaluated, or bone loss due to previous surgeries. Bone loss can be identified by using the biceps tuberosity as landmark. By evaluating the remaining bone proximal to the tuberosity, one can estimate an amount of bone loss. Further, one should look for already implanted bone anchors, whose removal might lead to a loss of bone substance at the distal humerus and would therewith lead to problems during revision surgery when augmenting the ligamentous stabilizers. Such information is of great relevance when planning potential revision surgery.
Further, there are direct signs of overlengthening to be drawn from the plain X-rays. The investigator should look for signs of proper articulation of the prosthesis with its bony counterparts, as the radial head implant should be aligned centrally under the capitulum in both the a.p. and the lateral plain. In severe overlengthening, the radial head implant may even be pushed out the posterior aspect of the joint (Fig. 3).
The images also help in identifying a loss of parallelism of the ulno-humeral joint line. This loss of parallelism, with a widening in the lateral aspect of the ulno-humeral joint and a closure of the medial aspect, is indicative of an overlengthening of the radial column. The work of Frank et al. showed in a cadaver study that parallelism of the ulno-humeral joint line was lost in overlengthening of the radial column [22]. This most likely results from the overlengthened prosthesis pushing against the capitulum, unhinging the ulno-humeral joint. However, the authors found that a visible difference in the parallelism of the ulno-humeral joint line in X-rays was only present in overlengthening of 6 mm or more. Further studies have taken on the understanding of overlengthening in imaging and how to identify it reliably. Athwal et al. published a method using X-ray imaging of the contra-lateral elbow to compare the ulno-humeral alignment in the a.p.-view to check for overlengthening in mono-axial radial head prostheses [23]. With their method, according to the authors, it is possible to quantify the magnitude of overlengthening; in this paper the authors used the Evolve prosthesis (Wright Medical). Another group reported on the usefulness of the ulnar variance, to detect overlengthening [24]. Moon et al. could show that increasing radial length by overlengthening of a radial head implant leads to a decrease of ulnar variance. So, if in question, X-ray imaging of both wrists may add some information. However, in our own study on the influence of overlengthening on ulnar variance and loss of parallelism, we could not replicate these results [25]. Several other authors have done anatomical studies to define anatomical landmarks that can be used on X-ray to identify or quantify overlengthening [26, 27]. Doornberg et al. investigated native CT-scans of 17 elbows to analys**e the anatomic relation of the radial head to the lateral edge of the coronoid [26]. They found that the radial head is levelled in average at 0.9 mm above the lateral edge of the coronoid. The authors recommend placing a prosthetic implant at the level of, or only slightly above the lateral edge of the coronoid to avoid overlengthening of the radial column. Another sign of overlengthening is abrasion of the cartilage, the subchondral bone or even the bone stock of the capitulum, as a result of the severely increased contact pressure. Even if not present initially, the abrasion of the capitulum may take place over time (Fig. 4) and may therewith limit surgical options for revision in the future.

Classification

We propose the following classification system, to be able to differentiate between minor to extensive overlengthening and to take chronic changes of the capitulum into account.
The classification is based on the native anatomical relation of the radial head to the lateral edge of the coronoid. Therefore, direct measurement of that relation is necessary, identifying whether the radial head prosthesis overreaches the lateral edge of the coronoid and to quantify it in millimetres. To achieve this measurement, imaging with an orthogonal view on the PRUJ is needed. In the case of X-ray imaging, extension deficit, which is often present in patients after radial head prosthesis surgery may obscure adequate imaging. Therefore, we recommend inclining the X-ray tube according to the extension deficit, to bring the central-beam of the X-ray tube in an orthogonal direction onto the forearm and not onto the distal humerus. This will result in an a.p. view on the PRUJ. In CT-imaging, 3D-reconstructions help finding the adequate plane independent of elbow position. At hand of such an orthogonal view, the height difference of the radial head prosthesis surface and of the lateral edge of the coronoid is quantified (Figs. 5, 6, 7).
Type
Description
Subtype
A
B
I
Overlengthening of up to 2 mm
Without erosion of the subchondral bone of the capitulum
With erosion of the subchondral bone of the capitulum
II
Overlengthening of more than 2 mm, with stable articulation
III
Overlengthening with subluxation or dislocation of the elbow
IV
Dynamic overlengthening due to longitudinal instability of the forearm
Type I (A) and (B)
Overlengthening of up to 2 mm without (A) or with (B) signs of erosion of the capitellar cartilage in imaging studies (Fig. 6).
Type II (A) and (B)
Overlengthening of more than 2 mm, with a stable joint, without (A) or with (B) signs of erosion of the capitulum in imaging studies (Fig. 7).
Type III (A) and (B)
Overlengthening with subluxation or dislocation of the elbow, without (A) or with (B) erosion of the subchondral bone of the capitulum (Fig. 8).
Type IV (A) and (B)
Dynamic overlengthening due to longitudinal instability of the forearm, without (A) or with (B) erosion of the subchondral bone of the capitulum (Fig. 9).

Treatment strategies

Due to the significant changes to joint mechanics and the erosive effects on capitellar cartilage, overlengthening of radial head prostheses often is subject to revision elbow surgery. Several surgical options from revision radial head prosthesis to resection- or interposition arthroplasty are available. As several factors influence the choice of revision technique, we adhere to the presented classification system. Concerning the lower grades of overlengthening, so far it is not clear whether an amount of < 2 mm is of clinical relevance. As we regularly see patients with low-grade overlengthening of up to 2 mm with significant clinical problems, we assort such cases to group I. We believe it is important to acknowledge also low-grade overlengthening as a pathology. We treat patients with Type I overlengthening initially conservatively. The patients are brought to physiotherapy and manual therapy, paired with a trial period of load relief for 6 weeks. If the patient´s pain does not respond to conservative therapy or if stiffness necessitates revision surgery, we perform arthroscopy with detailed analysis of joint mechanics and evaluation of the osteo-cartilaginous condition of the capitulum. If in such cases overlengthening is identified as causative for the clinical symptoms and if the capitulum does not show significant bony erosion (Type IA) (Fig. 10), we see revision radial head prosthesis, levelled at the lateral edge of the coronoid as an appropriate option (Fig. 11).
If erosion of the subchondral bone of the capitulum is found (Type IB) it is an option to remove the prosthetic component and check for valgus stability and for longitudinal stability. If the joint is stable in these regards, one can elect not to re-implant a prosthesis. By that the capitulum is unloaded, and a potential cause for pain and ongoing inflammation is taken from the joint. Another option is to intentionally “underlengthen” the implant. The idea is based on a biomechanical study we performed, to evaluate a beneficial effect of unloading the radio-humeral joint in the case of arthriti [28]. Shortening of the proximal radius by 3 mm leads to a significant unloading of the radio-humeral joint without affecting valgus stability. Maybe, unloading of a radial head prosthesis by deliberately placing it below the lateral edge of the coronoid can help reducing symptoms in patients with erosion of the capitulum. However, if we aim for under-lengthening of a radial head prosthesis, we do not take away more than 2 mm, as further lowering of the implant may lead to conflict with the lower edge of the lesser sigmoid notch. Moreover, the patient has to be informed pre-operatively in a written consent, about the clinical consequences of resection or shortening of the proximal radius. These consequences mainly are the development of tardy valgus- and longitudinal instability, resulting for example in medial elbow pain or wrist pain. If after removal of the radial head implant the elbow is unstable, ligament reconstruction or augmentation is mandatory.
If overlengthening of more than 2 mm is present (Type II A or B) (Fig. 12a, b), we regularly perform revision surgery. In these cases, often the capitulum is already eroded by the increased contact pressure, leading to ongoing inflammation with its typical clinical consequences. Conservative therapy in our experience does not lead to improvement in the case of significant overlengthening. As the capitulum is often already eroded, revision strategies must be evaluated carefully. Implantation of a further radial head prosthesis potentially ends up in recurrence of the symptoms, due to repetitive contact between the prosthesis and the eroded capitulum [29]. Hence, an attractive revision option again can be removal of the prosthesis, while respecting the aforementioned pre-conditions to do so. If sole removal is not an option, one can choose interposition arthroplasty to achieve an abutment between the radial stump and the capitulum. Interposing the anconeus muscle is a viable option, as shown by Rahmi et al. [29]. A limiting factor may be scarring of the anconeus due to previous surgeries, rendering the muscle inadequate to be mobilised and interposed. Also, if due to bony resection or bone-loss the gap between capitulum and the radial stump is substantial, a filling effect can not be achieved with the anconeus. A further option of treatment would be prosthetic replacement of both the radial head and the capitulum [30, 31]. With replacement of an eroded capitulum, the potential pain generator is removed. However, radio-capitellar arthroplasty is not without risks and has been shown to come along with severe complications [32].
In overlengthening of Type III, additionally ligamentous deficiencies are regularly present and should be addressed. As overlengthening in these cases is severe, and the joint is subluxed or even dislocated by the increased contact pressure, erosion of the capitulum is usually always present. Even severe abrasion with significant bone loss can be seen. Surgical revision is warranted in these cases. The first step is removal of the prosthesis. Following that, stability of the joint without the irritating prosthesis should be assessed thoroughly. In overlengthening of Type III, subluxation of the joint happens due to a severely increased length of the proximal radius. By that, the joint is distended and the medial and lateral ligamentous complex is tensioned [22]. Therefore, it is often necessary to address the medial and lateral ligaments. As the cases usually do not present acutely, we prefer to perform ligament reconstruction with autologous tendon grafts, augmented with internal bracing to secure the construct. Patients who do not have sufficient musculature to support joint stability, e.g., adipose arms, we consider additional stabilization with an external fixator for at least 6 weeks after surgery, to protect the repair.
In overlengthening of Type IV, the underlying pathology has to be addressed. As longitudinal instability of the forearm would repetitively lead to increased contact pressure between the prosthesis and the capitulum, relieve of symptoms cannot be achieved by lowering the present implant alone. Type IV usually comes with severe abrasion of the capitulum. To stabilize the proximal push of the radius, we reconstruct the central band of the interosseous membrane. Several surgical techniques have been presented to address the issue of longitudinal instability of the forearm [33, 34]. Therefore, autologous tendons or artificial fiber systems are available. If positive ulnar variance is present, ulnar shortening has to be considered accordingly (Fig. 13). In cases with longitudinal instability, the capitulum often shows severe abrasion or bone loss, due to the high contact pressures. In such cases, revision radial head prosthesis is not recommended and removal of the implant with longitudinal stabilization and shortening of the ulna is a salvage option (Fig. 14).

Discussion

Correct sizing and lengthening of prosthetic components is a known primary technical goal in any musculoskeletal replacement surgery. Oversizing, Overstuffing and Overlengthening are well-identified complications in Knee, Hip and Shoulder arthroplasty [35, 36]. The radio-humeral joint is one of intricate anatomy and biomechanics, too. Several clinical and biomechanical studies have reported the problem of overlengthening in radial head arthroplasty and have given strategies for identification and treatment. Van Glabbeek et al. have found significant alterations of elbow biomechanics with overlengthening of the radial column in six human cadaveric specimens [6]. The authors concluded that radial head prostheses would need to be implanted with high accuracy and reproducibility. In a further biomechanical study, van Glabbeek et al. investigated the effect of overlengthening of a radial head prosthesis on the rotational position of the ulna, compared to the native state [7]. The group showed that overlengthening of more than 2.5 mm led to varus malposition of the ulna, together with external rotation. Both reports clearly recommend implantation of radial head prostheses at the correct height. Concerning the radial head, revision of painful overlengthening to a total elbow prosthesis is not a favourable option, as total elbow arthroplasty shows high complication rates and limited survival time in younger trauma patients, which are typically affected by overlenghtening [3739]. As we stated, clinical series on overlengthening of radial head prostheses are rare and no reliable standards for the treatment are available. The presented descriptive classification system aims to raise the awareness of the clinical entity of radial head prosthesis overlengthening. The paper has several limitations, as the presented classification has not yet been evaluated at hand of a clinical series. Hence, it is not clear, except from very little literature, whether resection of radial head prostheses for painful overlengthening alone or combined with interposition does offer good long-term results. Nor it is clear, whether radio-capitellar arthroplasty shows a reasonable rate of complications and favourable long-term outcomes, to solve radial overlengthening with painful capitellar erosion. Also, it is not known whether deliberately reducing the length of the radial head implant below the recommended level of the lateral edge of the coronoid may reduce symptoms of capitellar erosion by a radial head implant. Our biomechanical study on the shortening of the radial head in case of symptomatic radio-humeral arthritis supports at least the idea. [28] It showed no negative biomechanical effects of gentle shortening of the proximal radius regarding stability, but it did reduce the contact pressure. Accordingly, the treatment recommendations given in the present paper are based only on the experiences of our clinical practice. Hence, it is of importance to further shed light into the clinical entity and to improve the knowledge concerning the topic. To the best of our knowledge, this is the first review describing a classification system on overlengthening of radial head prostheses.

Compliance with ethical standards

Conflict of interest

LP Müller is a paid consultant for Tornier Inc. and Arthrex Inc. and receives research support by Tornier Inc. and Arthrex Inc.; K Wegmann is a paid consultant for Arthrex Inc. and receives research support by Tornier Inc. and Arthrex Inc.; M Hackl, T Leschinger receive research support from Tornier Inc. and Arthrex Inc.; KJ Burkhart is a paid consultant for Zimmer Biomet.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.
Not applicable.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Doornberg JN, Parisien R, van Duijn PJ, Ring D (2007) Radial head arthroplasty with a modular metal spacer to treat acute traumatic elbow instability. J Bone Joint Surg Am 89(5):1075–1080CrossRef Doornberg JN, Parisien R, van Duijn PJ, Ring D (2007) Radial head arthroplasty with a modular metal spacer to treat acute traumatic elbow instability. J Bone Joint Surg Am 89(5):1075–1080CrossRef
2.
Zurück zum Zitat Grewal R, MacDermid JC, Faber KJ, Drosdowech DS, King GJ (2006) Comminuted radial head fractures treated with a modular metallic radial head arthroplasty. Study of outcomes. J Bone Joint Surg Am 88(10):2192–2200PubMed Grewal R, MacDermid JC, Faber KJ, Drosdowech DS, King GJ (2006) Comminuted radial head fractures treated with a modular metallic radial head arthroplasty. Study of outcomes. J Bone Joint Surg Am 88(10):2192–2200PubMed
3.
Zurück zum Zitat Marsh JP, Grewal R, Faber KJ, Drosdowech DS, Athwal GS, King GJ (2016) Radial head fractures treated with modular metallic radial head replacement: outcomes at a mean follow-up of eight years. J Bone Joint Surg Am 98(7):527–535CrossRef Marsh JP, Grewal R, Faber KJ, Drosdowech DS, Athwal GS, King GJ (2016) Radial head fractures treated with modular metallic radial head replacement: outcomes at a mean follow-up of eight years. J Bone Joint Surg Am 98(7):527–535CrossRef
4.
Zurück zum Zitat Pugh DM, McKee MD (2002) The “terrible triad” of the elbow. Tech Hand UpExtremSurg 6(1):21–29CrossRef Pugh DM, McKee MD (2002) The “terrible triad” of the elbow. Tech Hand UpExtremSurg 6(1):21–29CrossRef
5.
Zurück zum Zitat Birkedal JP, Deal DN, Ruch DS (2004) Loss of flexion after radial head replacement. J Shoulder Elbow Surg 13(2):208–213CrossRef Birkedal JP, Deal DN, Ruch DS (2004) Loss of flexion after radial head replacement. J Shoulder Elbow Surg 13(2):208–213CrossRef
6.
Zurück zum Zitat Van Glabbeek F, Van Riet RP, Baumfeld JA, Neale PG, O’Driscoll SW, Morrey BF et al (2004) Detrimental effects of overstuffing or understuffing with a radial head replacement in the medial collateral-ligament deficient elbow. J Bone Joint Surg Am 86(12):2629–2635CrossRef Van Glabbeek F, Van Riet RP, Baumfeld JA, Neale PG, O’Driscoll SW, Morrey BF et al (2004) Detrimental effects of overstuffing or understuffing with a radial head replacement in the medial collateral-ligament deficient elbow. J Bone Joint Surg Am 86(12):2629–2635CrossRef
7.
Zurück zum Zitat Van Glabbeek F, van Riet RP, Baumfeld JA, Neale PG, O’Driscoll SW, Morrey BF et al (2005) The kinematic importance of radial neck length in radial head replacement. Med EngPhys 27(4):336–342CrossRef Van Glabbeek F, van Riet RP, Baumfeld JA, Neale PG, O’Driscoll SW, Morrey BF et al (2005) The kinematic importance of radial neck length in radial head replacement. Med EngPhys 27(4):336–342CrossRef
8.
Zurück zum Zitat van Riet RP, Sanchez-Sotelo J, Morrey BF (2010) Failure of metal radial head replacement. J Bone Joint Surg Br 92(5):661–667CrossRef van Riet RP, Sanchez-Sotelo J, Morrey BF (2010) Failure of metal radial head replacement. J Bone Joint Surg Br 92(5):661–667CrossRef
9.
Zurück zum Zitat Van Riet RP, Van Glabbeek F, Verborgt O, Gielen J (2004) Capitellar erosion caused by a metal radial head prosthesis. A case report. J Bone Joint Surg Am 86(5):1061–1064CrossRef Van Riet RP, Van Glabbeek F, Verborgt O, Gielen J (2004) Capitellar erosion caused by a metal radial head prosthesis. A case report. J Bone Joint Surg Am 86(5):1061–1064CrossRef
10.
Zurück zum Zitat Schnetzke M, Aytac S, Deuss M, Studier-Fischer S, Swartman B, Muenzberg M et al (2014) Radial head prosthesis in complex elbow dislocations: effect of oversizing and comparison with ORIF. IntOrthop 38(11):2295–2301 Schnetzke M, Aytac S, Deuss M, Studier-Fischer S, Swartman B, Muenzberg M et al (2014) Radial head prosthesis in complex elbow dislocations: effect of oversizing and comparison with ORIF. IntOrthop 38(11):2295–2301
11.
Zurück zum Zitat Hackl M, Wegmann K, Hollinger B, El-Zayat BF, Seybold D, Guhring T et al (2019) Surgical revision of radial head fractures: a multicenter retrospective analysis of 466 cases. J Shoulder Elbow Surg 28(8):1457–1467CrossRef Hackl M, Wegmann K, Hollinger B, El-Zayat BF, Seybold D, Guhring T et al (2019) Surgical revision of radial head fractures: a multicenter retrospective analysis of 466 cases. J Shoulder Elbow Surg 28(8):1457–1467CrossRef
12.
Zurück zum Zitat Kachooei AR, Baradaran A, Ebrahimzadeh MH, van Dijk CN, Chen N (2018) The rate of radial head prosthesis removal or revision: a systematic review and meta-analysis. J Hand Surg Am 43(1):39-53e1CrossRef Kachooei AR, Baradaran A, Ebrahimzadeh MH, van Dijk CN, Chen N (2018) The rate of radial head prosthesis removal or revision: a systematic review and meta-analysis. J Hand Surg Am 43(1):39-53e1CrossRef
13.
Zurück zum Zitat Birkbeck DP, Failla JM, Hoshaw SJ, Fyhrie DP, Schaffler M (1997) Theinterosseous membrane affects load distribution in the forearm. J Hand Surg Am 22(6):975–980CrossRef Birkbeck DP, Failla JM, Hoshaw SJ, Fyhrie DP, Schaffler M (1997) Theinterosseous membrane affects load distribution in the forearm. J Hand Surg Am 22(6):975–980CrossRef
14.
Zurück zum Zitat Pfaeffle HJ, Fischer KJ, Manson TT, Tomaino MM, Woo SL, Herndon JH (2000) Role of the forearm interosseous ligament: is it more than just longitudinal load transfer? J Hand Surg Am 25(4):683–688CrossRef Pfaeffle HJ, Fischer KJ, Manson TT, Tomaino MM, Woo SL, Herndon JH (2000) Role of the forearm interosseous ligament: is it more than just longitudinal load transfer? J Hand Surg Am 25(4):683–688CrossRef
15.
Zurück zum Zitat Cohn M, Glait SA, Sapienza A, Kwon YW (2014) Radiocapitellar joint contact pressures following radial head arthroplasty. J Hand Surg Am 39(8):1566–1571CrossRef Cohn M, Glait SA, Sapienza A, Kwon YW (2014) Radiocapitellar joint contact pressures following radial head arthroplasty. J Hand Surg Am 39(8):1566–1571CrossRef
16.
Zurück zum Zitat Burkhart KJ, Mattyasovszky SG, Runkel M, Schwarz C, Kuchle R, Hessmann MH et al (2010) Mid- to long-term results after bipolar radial head arthroplasty. J Shoulder Elbow Surg 19(7):965–972CrossRef Burkhart KJ, Mattyasovszky SG, Runkel M, Schwarz C, Kuchle R, Hessmann MH et al (2010) Mid- to long-term results after bipolar radial head arthroplasty. J Shoulder Elbow Surg 19(7):965–972CrossRef
17.
Zurück zum Zitat Wegmann K, Hain MK, Ries C, Neiss WF, Muller LP, Burkhart KJ (2015) Do the radial head prosthesis components fit with the anatomical structures of the proximal radioulnar joint? SurgRadiolAnat 37(7):743–747 Wegmann K, Hain MK, Ries C, Neiss WF, Muller LP, Burkhart KJ (2015) Do the radial head prosthesis components fit with the anatomical structures of the proximal radioulnar joint? SurgRadiolAnat 37(7):743–747
18.
Zurück zum Zitat Liew VS, Cooper IC, Ferreira LM, Johnson JA, King GJ (2003) The effect of metallic radial head arthroplasty on radiocapitellar joint contact area. ClinBiomech (Bristol, Avon) 18(2):115–118CrossRef Liew VS, Cooper IC, Ferreira LM, Johnson JA, King GJ (2003) The effect of metallic radial head arthroplasty on radiocapitellar joint contact area. ClinBiomech (Bristol, Avon) 18(2):115–118CrossRef
19.
Zurück zum Zitat Langohr GD, Willing R, Medley JB, King GJ, Johnson JA (2016) The effect of radial head hemiarthroplasty geometry on proximal radioulnar joint contact mechanics. J Hand Surg Am 41(7):745–752CrossRef Langohr GD, Willing R, Medley JB, King GJ, Johnson JA (2016) The effect of radial head hemiarthroplasty geometry on proximal radioulnar joint contact mechanics. J Hand Surg Am 41(7):745–752CrossRef
20.
Zurück zum Zitat Lanting BA, Ferreira LM, Johnson JA, King GJ, Athwal GS (2015) Radial head implant diameter: a biomechanical assessment of the forgotten dimension. ClinBiomech (Bristol, Avon) 30(5):444–447CrossRef Lanting BA, Ferreira LM, Johnson JA, King GJ, Athwal GS (2015) Radial head implant diameter: a biomechanical assessment of the forgotten dimension. ClinBiomech (Bristol, Avon) 30(5):444–447CrossRef
21.
Zurück zum Zitat Moungondo F, El Kazzi W, van Riet R, Feipel V, Rooze M, Schuind F (2010) Radiocapitellar joint contacts after bipolar radial head arthroplasty. J Shoulder Elbow Surg 19(2):230–235CrossRef Moungondo F, El Kazzi W, van Riet R, Feipel V, Rooze M, Schuind F (2010) Radiocapitellar joint contacts after bipolar radial head arthroplasty. J Shoulder Elbow Surg 19(2):230–235CrossRef
22.
Zurück zum Zitat Frank SG, Grewal R, Johnson J, Faber KJ, King GJ, Athwal GS (2009) Determination of correct implant size in radial head arthroplasty to avoid overlengthening. J Bone Joint Surg Am 91(7):1738–1746CrossRef Frank SG, Grewal R, Johnson J, Faber KJ, King GJ, Athwal GS (2009) Determination of correct implant size in radial head arthroplasty to avoid overlengthening. J Bone Joint Surg Am 91(7):1738–1746CrossRef
23.
Zurück zum Zitat Athwal GS, Rouleau DM, MacDermid JC, King GJ (2011) Contralateral elbow radiographs can reliably diagnose radial head implant overlengthening. J Bone Joint Surg Am 93(14):1339–1346CrossRef Athwal GS, Rouleau DM, MacDermid JC, King GJ (2011) Contralateral elbow radiographs can reliably diagnose radial head implant overlengthening. J Bone Joint Surg Am 93(14):1339–1346CrossRef
24.
Zurück zum Zitat Moon JG, Hong JH, Bither N, Shon WY (2014) Can ulnar variance be used to detect overstuffing after radial head arthroplasty? ClinOrthopRelat Res 472(2):727–731 Moon JG, Hong JH, Bither N, Shon WY (2014) Can ulnar variance be used to detect overstuffing after radial head arthroplasty? ClinOrthopRelat Res 472(2):727–731
25.
Zurück zum Zitat Wegmann K, Zhang W, Strauss V, Ries C, Muller LP, Burkhart KJ (2016) Is the distal radioulnar variance useful for identification of radial head prosthesis overlengthening? A cadaver study. Eur J Med Res 21:12CrossRef Wegmann K, Zhang W, Strauss V, Ries C, Muller LP, Burkhart KJ (2016) Is the distal radioulnar variance useful for identification of radial head prosthesis overlengthening? A cadaver study. Eur J Med Res 21:12CrossRef
26.
Zurück zum Zitat Doornberg JN, Linzel DS, Zurakowski D, Ring D (2006) Reference points for radial head prosthesis size. J Hand Surg Am 31(1):53–57CrossRef Doornberg JN, Linzel DS, Zurakowski D, Ring D (2006) Reference points for radial head prosthesis size. J Hand Surg Am 31(1):53–57CrossRef
27.
Zurück zum Zitat van Riet RP, van Glabbeek F, de Weerdt W, Oemar J, Bortier H (2007) Validation of the lesser sigmoid notch of the ulna as a reference point for accurate placement of a prosthesis for the head of the radius: a cadaver study. J Bone Joint Surg Br 89(3):413–416CrossRef van Riet RP, van Glabbeek F, de Weerdt W, Oemar J, Bortier H (2007) Validation of the lesser sigmoid notch of the ulna as a reference point for accurate placement of a prosthesis for the head of the radius: a cadaver study. J Bone Joint Surg Br 89(3):413–416CrossRef
28.
Zurück zum Zitat Hackl M, Wegmann K, Kahmann SL, Heinze N, Staat M, Neiss WF et al (2017) Radial shortening osteotomy reduces radiocapitellar contact pressures while preserving valgus stability of the elbow. Knee Surg Sports TraumatolArthrosc 25(7):2280–2288CrossRef Hackl M, Wegmann K, Kahmann SL, Heinze N, Staat M, Neiss WF et al (2017) Radial shortening osteotomy reduces radiocapitellar contact pressures while preserving valgus stability of the elbow. Knee Surg Sports TraumatolArthrosc 25(7):2280–2288CrossRef
29.
Zurück zum Zitat Rahmi H, Neumann JA, Klein C, van Eck CF, Lee BK, Itamura J (2018) Clinical outcomes of anconeus interposition arthroplasty after radial head resection in native and prosthetic radial heads. J Shoulder Elbow Surg 27(6S):S29–S34CrossRef Rahmi H, Neumann JA, Klein C, van Eck CF, Lee BK, Itamura J (2018) Clinical outcomes of anconeus interposition arthroplasty after radial head resection in native and prosthetic radial heads. J Shoulder Elbow Surg 27(6S):S29–S34CrossRef
30.
Zurück zum Zitat Giannicola G, Calella P, Bigazzi P, Mantovani A, Spinello P, Cinotti G (2019) Midterm results of radiocapitellararthroplasty of the elbow: a multicentre prospective study on two different implants. Bone Joint J 101-B(11):1362–1369CrossRef Giannicola G, Calella P, Bigazzi P, Mantovani A, Spinello P, Cinotti G (2019) Midterm results of radiocapitellararthroplasty of the elbow: a multicentre prospective study on two different implants. Bone Joint J 101-B(11):1362–1369CrossRef
31.
Zurück zum Zitat Heijink A, Morrey BF, Eygendaal D (2014) Radiocapitellar prosthetic arthroplasty: a report of 6 cases and review of the literature. J Shoulder Elbow Surg 23(6):843–849CrossRef Heijink A, Morrey BF, Eygendaal D (2014) Radiocapitellar prosthetic arthroplasty: a report of 6 cases and review of the literature. J Shoulder Elbow Surg 23(6):843–849CrossRef
32.
Zurück zum Zitat Hackl M, Burkhart KJ, Wegmann K, Hollinger B, Lichtenberg S, Muller LP (2015) From radial head to radiocapitellar to total elbow replacement: a case report. Int J Surg Case Rep 15:35–38CrossRef Hackl M, Burkhart KJ, Wegmann K, Hollinger B, Lichtenberg S, Muller LP (2015) From radial head to radiocapitellar to total elbow replacement: a case report. Int J Surg Case Rep 15:35–38CrossRef
33.
Zurück zum Zitat Sabo MT, Watts AC (2012) Reconstructing the interosseous membrane: a technique using synthetic graft and endobuttons. Tech Hand UpExtremSurg 16(4):187–193CrossRef Sabo MT, Watts AC (2012) Reconstructing the interosseous membrane: a technique using synthetic graft and endobuttons. Tech Hand UpExtremSurg 16(4):187–193CrossRef
34.
Zurück zum Zitat Adams JE (2017) Forearm instability: anatomy, biomechanics, and treatment options. J Hand Surg Am 42(1):47–52CrossRef Adams JE (2017) Forearm instability: anatomy, biomechanics, and treatment options. J Hand Surg Am 42(1):47–52CrossRef
35.
Zurück zum Zitat Geervliet PC, Willems JH, Sierevelt IN, Visser CPJ, van Noort A (2019) Overstuffing in resurfacing hemiarthroplasty is a potential risk for failure. J OrthopSurg Res 14(1):474 Geervliet PC, Willems JH, Sierevelt IN, Visser CPJ, van Noort A (2019) Overstuffing in resurfacing hemiarthroplasty is a potential risk for failure. J OrthopSurg Res 14(1):474
36.
Zurück zum Zitat Kemp MA, Metcalfe AJ, Sayers A, Wylde V, Eldridge JD, Blom AW (2018) Does overstuffing of the patellofemoral joint in total knee arthroplasty have a significant effect on postoperative outcomes? Knee 25(5):874–881CrossRef Kemp MA, Metcalfe AJ, Sayers A, Wylde V, Eldridge JD, Blom AW (2018) Does overstuffing of the patellofemoral joint in total knee arthroplasty have a significant effect on postoperative outcomes? Knee 25(5):874–881CrossRef
37.
Zurück zum Zitat Geurts EJ, Viveen J, van Riet RP, Kodde IF, Eygendaal D (2019) Outcomes after revision total elbow arthroplasty: a systematic review. J Shoulder Elbow Surg 28(2):381–386CrossRef Geurts EJ, Viveen J, van Riet RP, Kodde IF, Eygendaal D (2019) Outcomes after revision total elbow arthroplasty: a systematic review. J Shoulder Elbow Surg 28(2):381–386CrossRef
38.
Zurück zum Zitat Prkic A, Welsink C, The B, van den Bekerom MPJ, Eygendaal D (2017) Why does total elbow arthroplasty fail today? A systematic review of recent literature. Arch Orthop Trauma Surg 137(6):761–769CrossRef Prkic A, Welsink C, The B, van den Bekerom MPJ, Eygendaal D (2017) Why does total elbow arthroplasty fail today? A systematic review of recent literature. Arch Orthop Trauma Surg 137(6):761–769CrossRef
39.
Zurück zum Zitat Welsink CL, Lambers KTA, van Deurzen DFP, Eygendaal D, van den Bekerom MPJ (2017) Total elbow arthroplasty: a systematic review. JBJS Rev 5(7):e4CrossRef Welsink CL, Lambers KTA, van Deurzen DFP, Eygendaal D, van den Bekerom MPJ (2017) Total elbow arthroplasty: a systematic review. JBJS Rev 5(7):e4CrossRef
Metadaten
Titel
Overlengthening of the radial column in radial head replacement: a review of the literature and presentation of a classification system
verfasst von
K. Wegmann
M. Hackl
T. Leschinger
K. J. Burkhart
L. P. Müller
Publikationsdatum
14.10.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 9/2021
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-020-03619-9

Weitere Artikel der Ausgabe 9/2021

Archives of Orthopaedic and Trauma Surgery 9/2021 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.