Steady-state multifocal visual evoked potential (ssmfVEP) using dartboard stimulation as a possible tool for objective visual field assessment

verfasst von: Folkert K. Horn, Franziska Selle, Bettina Hohberger, Jan Kremers

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To investigate whether a conventional, monitor-based multifocal visual evoked potential (mfVEP) system can be used to record steady-state mfVEP (ssmfVEP) in healthy subjects and to study the effects of temporal frequency, electrode configuration and alpha waves.

Methods

Multifocal pattern reversal VEP measurements were performed at 58 dartboard fields using VEP recording equipment. The responses were measured using m-sequences with four pattern reversals per m-step. Temporal frequencies were varied between 6 and 15 Hz. Recordings were obtained from nine normal subjects with a cross-shaped, four-electrode device (two additional channels were derived). Spectral analyses were performed on the responses at all locations. The signal to noise ratio (SNR) was computed for each response using the signal amplitude at the reversal frequency and the noise at the neighbouring frequencies.

Results

Most responses in the ssmfVEP were significantly above noise. The SNR was largest for an 8.6-Hz reversal frequency. The individual alpha electroencephalogram (EEG) did not strongly influence the results. The percentage of the records in which each of the 6 channels had the largest SNR was between 10.0 and 25.2 %.

Conclusion

Our results in normal subjects indicate that reliable mfVEP responses can be achieved by steady-state stimulation using a conventional dartboard stimulator and multi-channel electrode device. The ssmfVEP may be useful for objective visual field assessment as spectrum analysis can be used for automated evaluation of responses. The optimal reversal frequency is 8.6 Hz. Alpha waves have only a minor influence on the analysis. Future studies must include comparisons with conventional mfVEP and psychophysical visual field tests.
Literatur
1.
Zurück zum Zitat Baseler HA, Sutter EE, Klein SA, Carney T (1994) The topography of visual evoked response properties across the visual field. Electroencephalogr Clin Neurophysiol 90(1):65–81PubMedCrossRef Baseler HA, Sutter EE, Klein SA, Carney T (1994) The topography of visual evoked response properties across the visual field. Electroencephalogr Clin Neurophysiol 90(1):65–81PubMedCrossRef
2.
Zurück zum Zitat Klistorner A, Graham SL (2000) Objective perimetry in glaucoma. Ophthalmology 107(12):2283–2299PubMedCrossRef Klistorner A, Graham SL (2000) Objective perimetry in glaucoma. Ophthalmology 107(12):2283–2299PubMedCrossRef
3.
Zurück zum Zitat Hood DC, Zhang X, Greenstein VC, Kangovi S, Odel JG, Liebmann JM, Ritch R (2000) An interocular comparison of the multifocal VEP: a possible technique for detecting local damage to the optic nerve. Invest Ophthalmol Vis Sci 41(6):1580–1587PubMed Hood DC, Zhang X, Greenstein VC, Kangovi S, Odel JG, Liebmann JM, Ritch R (2000) An interocular comparison of the multifocal VEP: a possible technique for detecting local damage to the optic nerve. Invest Ophthalmol Vis Sci 41(6):1580–1587PubMed
4.
Zurück zum Zitat Hood DC, Odel JG, Zhang X (2000) Tracking the recovery of local optic nerve function after optic neuritis: a multifocal VEP study. Invest Ophthalmol Vis Sci 41(12):4032–4038 Hood DC, Odel JG, Zhang X (2000) Tracking the recovery of local optic nerve function after optic neuritis: a multifocal VEP study. Invest Ophthalmol Vis Sci 41(12):4032–4038
5.
Zurück zum Zitat Graham SL, Klistorner AI, Goldberg I (2005) Clinical application of objective perimetry using multifocal visual evoked potentials in glaucoma practice. Arch Ophthalmol 123(6):729–739PubMedCrossRef Graham SL, Klistorner AI, Goldberg I (2005) Clinical application of objective perimetry using multifocal visual evoked potentials in glaucoma practice. Arch Ophthalmol 123(6):729–739PubMedCrossRef
6.
Zurück zum Zitat Kaltwasser C, Horn FK, Kremers J, Juenemann A, Bergua A (2011) Objective visual field determination in forensic ophthalmology with an optimized 4-channel multifocal VEP perimetry system: a case report of a patient with retinitis pigmentosa. Doc Ophthalmol 123(2):121–125PubMedCrossRef Kaltwasser C, Horn FK, Kremers J, Juenemann A, Bergua A (2011) Objective visual field determination in forensic ophthalmology with an optimized 4-channel multifocal VEP perimetry system: a case report of a patient with retinitis pigmentosa. Doc Ophthalmol 123(2):121–125PubMedCrossRef
7.
Zurück zum Zitat Alshowaeir D, Yannikas C, Garrick R, Van Der Walt A, Graham SL, Fraser C, Klistorner A (2015) Multifocal VEP assessment of optic neuritis evolution. Clin Neurophysiol 126:1617–1623PubMedCrossRef Alshowaeir D, Yannikas C, Garrick R, Van Der Walt A, Graham SL, Fraser C, Klistorner A (2015) Multifocal VEP assessment of optic neuritis evolution. Clin Neurophysiol 126:1617–1623PubMedCrossRef
8.
Zurück zum Zitat Horn FK, Kaltwasser C, Junemann AG, Kremers J, Tornow RP (2012) Objective perimetry using a four-channel multifocal VEP system: correlation with conventional perimetry and thickness of the retinal nerve fibre layer. Br J Ophthalmol 96(4):554–559PubMedCrossRef Horn FK, Kaltwasser C, Junemann AG, Kremers J, Tornow RP (2012) Objective perimetry using a four-channel multifocal VEP system: correlation with conventional perimetry and thickness of the retinal nerve fibre layer. Br J Ophthalmol 96(4):554–559PubMedCrossRef
9.
Zurück zum Zitat Punjabi OS, Stamper RL, Bostrom AG, Han Y, Lin SC (2008) Topographic comparison of the visual function on multifocal visual evoked potentials with optic nerve structure on Heidelberg retinal tomography. Ophthalmology 115(3):440–446PubMedCrossRef Punjabi OS, Stamper RL, Bostrom AG, Han Y, Lin SC (2008) Topographic comparison of the visual function on multifocal visual evoked potentials with optic nerve structure on Heidelberg retinal tomography. Ophthalmology 115(3):440–446PubMedCrossRef
10.
Zurück zum Zitat Laron M, Cheng H, Zhang B, Schiffman JS, Tang RA, Frishman LJ (2010) Comparison of multifocal visual evoked potential, standard automated perimetry and optical coherence tomography in assessing visual pathway in multiple sclerosis patients. Mult Scler 16(4):412–426PubMedPubMedCentralCrossRef Laron M, Cheng H, Zhang B, Schiffman JS, Tang RA, Frishman LJ (2010) Comparison of multifocal visual evoked potential, standard automated perimetry and optical coherence tomography in assessing visual pathway in multiple sclerosis patients. Mult Scler 16(4):412–426PubMedPubMedCentralCrossRef
11.
Zurück zum Zitat Klistorner AI, Graham SL, Grigg JR, Billson FA (1998) Multifocal topographic visual evoked potential: improving objective detection of local visual field defects. Invest Ophthalmol Vis Sci 39(6):937–950PubMed Klistorner AI, Graham SL, Grigg JR, Billson FA (1998) Multifocal topographic visual evoked potential: improving objective detection of local visual field defects. Invest Ophthalmol Vis Sci 39(6):937–950PubMed
12.
Zurück zum Zitat Hood DC, Zhang X, Hong JE, Chen CS (2002) Quantifying the benefits of additional channels of multifocal VEP recording. Doc Ophthalmol 104(3):303–320 Hood DC, Zhang X, Hong JE, Chen CS (2002) Quantifying the benefits of additional channels of multifocal VEP recording. Doc Ophthalmol 104(3):303–320
13.
Zurück zum Zitat Hood DC, Greenstein VC (2003) Multifocal VEP and ganglion cell damage: applications and limitations for the study of glaucoma. Prog Retin Eye Res 22(2):201–251PubMedCrossRef Hood DC, Greenstein VC (2003) Multifocal VEP and ganglion cell damage: applications and limitations for the study of glaucoma. Prog Retin Eye Res 22(2):201–251PubMedCrossRef
14.
Zurück zum Zitat Mousa MF, Cubbidge RP, Al-Mansouri F, Bener A (2013) The role of hemifield sector analysis in multifocal visual evoked potential objective perimetry in the early detection of glaucomatous visual field defects. Clin Ophthalmol 7:843–858 Mousa MF, Cubbidge RP, Al-Mansouri F, Bener A (2013) The role of hemifield sector analysis in multifocal visual evoked potential objective perimetry in the early detection of glaucomatous visual field defects. Clin Ophthalmol 7:843–858
15.
Zurück zum Zitat Lindenberg T, Peters A, Horn FK, Lausen B, Korth M (2004) Diagnostic value of multifocal VEP using cross-validation and noise reduction in glaucoma research. Graefes Arch Clin Exp Ophthalmol 242(5):361–367PubMedCrossRef Lindenberg T, Peters A, Horn FK, Lausen B, Korth M (2004) Diagnostic value of multifocal VEP using cross-validation and noise reduction in glaucoma research. Graefes Arch Clin Exp Ophthalmol 242(5):361–367PubMedCrossRef
16.
Zurück zum Zitat Zhang X, Hood DC, Chen CS, Hong JE (2002) A signal-to-noise analysis of multifocal VEP responses: an objective definition for poor records. Doc Ophthalmol 104(3):287–302PubMedCrossRef Zhang X, Hood DC, Chen CS, Hong JE (2002) A signal-to-noise analysis of multifocal VEP responses: an objective definition for poor records. Doc Ophthalmol 104(3):287–302PubMedCrossRef
17.
Zurück zum Zitat Regan D (1989) Human brain electrophysiology: evoked potentials and evoked magnetic fields in science and medicine. Elsevier, New York Regan D (1989) Human brain electrophysiology: evoked potentials and evoked magnetic fields in science and medicine. Elsevier, New York
18.
Zurück zum Zitat Burkitt GR, Silberstein RB, Cadusch PJ, Wood AW (2000) Steady-state visual evoked potentials and travelling waves. Clin Neurophysiol 111(2):246–258PubMedCrossRef Burkitt GR, Silberstein RB, Cadusch PJ, Wood AW (2000) Steady-state visual evoked potentials and travelling waves. Clin Neurophysiol 111(2):246–258PubMedCrossRef
19.
Zurück zum Zitat Bakardjian H, Tanaka T, Cichocki A (2010) Optimization of SSVEP brain responses with application to eight-command brain-computer interface. Neurosci Lett 469(1):34–38PubMedCrossRef Bakardjian H, Tanaka T, Cichocki A (2010) Optimization of SSVEP brain responses with application to eight-command brain-computer interface. Neurosci Lett 469(1):34–38PubMedCrossRef
20.
Zurück zum Zitat Vialatte FB, Maurice M, Dauwels J, Cichocki A (2010) Steady-state visually evoked potentials: focus on essential paradigms and future perspectives. Prog Neurobiol 90(4):418–438PubMedCrossRef Vialatte FB, Maurice M, Dauwels J, Cichocki A (2010) Steady-state visually evoked potentials: focus on essential paradigms and future perspectives. Prog Neurobiol 90(4):418–438PubMedCrossRef
21.
Zurück zum Zitat Mandel C, Luth T, Laue T, Rofer T, Graser A, Krieg-Bruckner B (2009) Navigating a smart wheelchair with a brain–computer interface interpreting steady-state visual evoked potentials. Proc Conf Rec IEEE/RSJ Int Conf Intell Robots Syst, St Louis, MO: 1118–1125 Mandel C, Luth T, Laue T, Rofer T, Graser A, Krieg-Bruckner B (2009) Navigating a smart wheelchair with a brain–computer interface interpreting steady-state visual evoked potentials. Proc Conf Rec IEEE/RSJ Int Conf Intell Robots Syst, St Louis, MO: 1118–1125
22.
Zurück zum Zitat Abdullah SN, Vaegan BMY, Maddess T (2012) Contrast-response functions of the multifocal steady-state VEP (MSV). Clin Neurophysiol 123(9):1865–1871PubMedCrossRef Abdullah SN, Vaegan BMY, Maddess T (2012) Contrast-response functions of the multifocal steady-state VEP (MSV). Clin Neurophysiol 123(9):1865–1871PubMedCrossRef
23.
Zurück zum Zitat Herrmann CS (2001) Human EEG responses to 1–100 Hz flicker: resonance phenomena in visual cortex and their potential correlation to cognitive phenomena. Exp Brain Res 137(3–4):346–353PubMedCrossRef Herrmann CS (2001) Human EEG responses to 1–100 Hz flicker: resonance phenomena in visual cortex and their potential correlation to cognitive phenomena. Exp Brain Res 137(3–4):346–353PubMedCrossRef
24.
Zurück zum Zitat Kaltwasser C, Horn FK, Kremers J, Juenemann A (2009) A comparison of the suitability of cathode ray tube (CRT) and liquid crystal display (LCD) monitors as visual stimulators in mfERG diagnostics. Doc Ophthalmol 118(3):179–189PubMedCrossRef Kaltwasser C, Horn FK, Kremers J, Juenemann A (2009) A comparison of the suitability of cathode ray tube (CRT) and liquid crystal display (LCD) monitors as visual stimulators in mfERG diagnostics. Doc Ophthalmol 118(3):179–189PubMedCrossRef
25.
Zurück zum Zitat Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98(3):207–232PubMedCrossRef Meigen T, Bach M (1999) On the statistical significance of electrophysiological steady-state responses. Doc Ophthalmol 98(3):207–232PubMedCrossRef
26.
Zurück zum Zitat Norcia AM, Tyler CW, Hamer RD, Wesemann W (1989) Measurement of spatial contrast sensitivity with the swept contrast VEP. Vis Res 29(5):627–637PubMedCrossRef Norcia AM, Tyler CW, Hamer RD, Wesemann W (1989) Measurement of spatial contrast sensitivity with the swept contrast VEP. Vis Res 29(5):627–637PubMedCrossRef
27.
Zurück zum Zitat Lin FC, Zao JK, Tu KC, Wang Y, Huang YP, Chuang CW, Kuo HY, Chien YY, Chou CC, Jung TP (2012) SNR analysis of high-frequency steady-state visual evoked potentials from the foveal and extrafoveal regions of human retina. Conf Proc IEEE Eng Med Biol Soc 2012:1810–1814PubMed Lin FC, Zao JK, Tu KC, Wang Y, Huang YP, Chuang CW, Kuo HY, Chien YY, Chou CC, Jung TP (2012) SNR analysis of high-frequency steady-state visual evoked potentials from the foveal and extrafoveal regions of human retina. Conf Proc IEEE Eng Med Biol Soc 2012:1810–1814PubMed
28.
Zurück zum Zitat Vaegan, Elia MA, Zheng J (2007) Optimising steady-state sweep VEPs to stimuli like the central FDT target: comparison to psychophysical thresholds in early glaucoma detection. In: Marmor M, ed XLV ISCEV international symposium—scientific programme, hotel Taj Krishna, Banjara Hills, Hyderabad, India, August 25–29, 2007 Doc Ophthalmol; 115(suppl):12 Vaegan, Elia MA, Zheng J (2007) Optimising steady-state sweep VEPs to stimuli like the central FDT target: comparison to psychophysical thresholds in early glaucoma detection. In: Marmor M, ed XLV ISCEV international symposium—scientific programme, hotel Taj Krishna, Banjara Hills, Hyderabad, India, August 25–29, 2007 Doc Ophthalmol; 115(suppl):12
29.
Zurück zum Zitat Vaegan RAMA, Sanderson GF (2008) Glaucoma affects steady-state VEP contrast thresholds before psychophysics. Optom Vis Sci 85(7):547–558PubMedCrossRef Vaegan RAMA, Sanderson GF (2008) Glaucoma affects steady-state VEP contrast thresholds before psychophysics. Optom Vis Sci 85(7):547–558PubMedCrossRef
30.
Zurück zum Zitat Birca A, Carmant L, Lortie A, Lassonde M (2006) Interaction between the flash evoked SSVEPs and the spontaneous EEG activity in children and adults. Clin Neurophysiol 117(2):279–288PubMedCrossRef Birca A, Carmant L, Lortie A, Lassonde M (2006) Interaction between the flash evoked SSVEPs and the spontaneous EEG activity in children and adults. Clin Neurophysiol 117(2):279–288PubMedCrossRef
31.
Zurück zum Zitat Klimesch W (1999) EEG alpha and theta oscillations reflect cognitive and memory performance: a review and analysis. Brain Res Brain Res Rev 29(2–3):169–195PubMedCrossRef Klimesch W (1999) EEG alpha and theta oscillations reflect cognitive and memory performance: a review and analysis. Brain Res Brain Res Rev 29(2–3):169–195PubMedCrossRef
32.
Zurück zum Zitat Garn H, Waser M, Lechner M, Dorfer M, Grossegger D (2012) Robust, automatic real-time monitoring of the time course of the individual alpha frequency in the time and frequency domain. Conf Proc IEEE Eng Med Biol Soc 2012:2227–2231PubMed Garn H, Waser M, Lechner M, Dorfer M, Grossegger D (2012) Robust, automatic real-time monitoring of the time course of the individual alpha frequency in the time and frequency domain. Conf Proc IEEE Eng Med Biol Soc 2012:2227–2231PubMed
33.
Zurück zum Zitat Klistorner AI, Graham SL (1999) Multifocal pattern VEP perimetry: analysis of sectoral waveforms. Doc Ophthalmol 98(2):183–196PubMedCrossRef Klistorner AI, Graham SL (1999) Multifocal pattern VEP perimetry: analysis of sectoral waveforms. Doc Ophthalmol 98(2):183–196PubMedCrossRef
34.
Zurück zum Zitat Fortune B, Hood DC (2003) Conventional pattern-reversal VEPs are not equivalent to summed multifocal VEPs. Investig Ophthalmol Vis Sci 44(3):1364–1375CrossRef Fortune B, Hood DC (2003) Conventional pattern-reversal VEPs are not equivalent to summed multifocal VEPs. Investig Ophthalmol Vis Sci 44(3):1364–1375CrossRef
35.
Zurück zum Zitat Bradnam MS, Montgomery DM, Evans AL, Keating D, McClure EA, Damato BE, McFadzean R (1996) Objective detection of hemifield and quadrantic field defects by visual evoked cortical potentials. Br J Ophthalmol 80(4):297–303PubMedPubMedCentralCrossRef Bradnam MS, Montgomery DM, Evans AL, Keating D, McClure EA, Damato BE, McFadzean R (1996) Objective detection of hemifield and quadrantic field defects by visual evoked cortical potentials. Br J Ophthalmol 80(4):297–303PubMedPubMedCentralCrossRef
36.
Zurück zum Zitat Bach M, Meigen T (1999) Do’s and don’ts in fourier analysis of steady-state potentials. Doc Ophthalmol 99(1):69–82PubMedCrossRef Bach M, Meigen T (1999) Do’s and don’ts in fourier analysis of steady-state potentials. Doc Ophthalmol 99(1):69–82PubMedCrossRef
37.
Zurück zum Zitat Chen CS, Hood DC, Zhang X, Karam EZ, Liebmann JM, Ritch R, Thienprasiddhi P, Greenstein VC (2003) Repeat reliability of the multifocal visual evoked potential in normal and glaucomatous eyes. J Glaucoma 12(5):399–408PubMedCrossRef Chen CS, Hood DC, Zhang X, Karam EZ, Liebmann JM, Ritch R, Thienprasiddhi P, Greenstein VC (2003) Repeat reliability of the multifocal visual evoked potential in normal and glaucomatous eyes. J Glaucoma 12(5):399–408PubMedCrossRef
38.
Zurück zum Zitat Fortune B, Demirel S, Zhang X, Hood DC, Johnson CA (2006) Repeatability of normal multifocal VEP: implications for detecting progression. J Glaucoma 15(2):131–141PubMedCrossRef Fortune B, Demirel S, Zhang X, Hood DC, Johnson CA (2006) Repeatability of normal multifocal VEP: implications for detecting progression. J Glaucoma 15(2):131–141PubMedCrossRef
39.
Zurück zum Zitat Punjabi OS, Stamper RL, Bostrom AG, Lin SC (2008) Repeatability of the multifocal visual evoked potentials in a clinical glaucoma setting. Can J Ophthalmol 43(4):435–440PubMedCrossRef Punjabi OS, Stamper RL, Bostrom AG, Lin SC (2008) Repeatability of the multifocal visual evoked potentials in a clinical glaucoma setting. Can J Ophthalmol 43(4):435–440PubMedCrossRef
40.
Zurück zum Zitat Perlstein WM, Cole MA, Larson M, Kelly K, Seignourel P, Keil A (2003) Steadystate visual evoked potentials reveal frontally-mediated working memory activity in humans. Neurosci Lett 342:191–195PubMedCrossRef Perlstein WM, Cole MA, Larson M, Kelly K, Seignourel P, Keil A (2003) Steadystate visual evoked potentials reveal frontally-mediated working memory activity in humans. Neurosci Lett 342:191–195PubMedCrossRef
41.
Zurück zum Zitat Gray M, Kemp AH, Silberstein RB, Nathan PJ (2003) Cortical neurophysiology of anticipatory anxiety: an investigation utilizing steady-state probe topography (SSPT). Neuroimage 20:975–986PubMedCrossRef Gray M, Kemp AH, Silberstein RB, Nathan PJ (2003) Cortical neurophysiology of anticipatory anxiety: an investigation utilizing steady-state probe topography (SSPT). Neuroimage 20:975–986PubMedCrossRef
Metadaten
Titel
Steady-state multifocal visual evoked potential (ssmfVEP) using dartboard stimulation as a possible tool for objective visual field assessment
verfasst von
Folkert K. Horn
Franziska Selle
Bettina Hohberger
Jan Kremers
Publikationsdatum
01.02.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 2/2016
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-015-3214-9

Weitere Artikel der Ausgabe 2/2016

Graefe's Archive for Clinical and Experimental Ophthalmology 2/2016 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

CME: Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Wird ein Glaukom chirurgisch behandelt, ist die anschließende Wundheilung von entscheidender Bedeutung. In diesem CME-Kurs lernen Sie, welche Pathomechanismen der Vernarbung zugrunde liegen, wie perioperativ therapiert und Operationsversagen frühzeitig erkannt werden kann.

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.