The use of uncemented hip arthroplasty prostheses with ceramic articulations are popular, especially in the young, because of a perceived reduction in wear. We highlight a complication of ceramic on polyethylene articulating couples not previously described in the Furlong replacement. Despite widespread metalosis and particulate debris, osteolysis was not initially seen. The contamination compromised subsequent revision.
Introduction
We present two cases of early and catastrophic acetabular failure in the uncemented Furlong total hip replacement (JRI, London, UK). The Furlong has produced some encouraging results, but we would like to sound a note of caution if using a ceramic on polyethylene articulating couple. The patients have provided consent to the publication of their clinical reports.
Case report
A female patient, aged 51, underwent a right total hip replacement for osteoarthritis. An uncemented, HA-coated, Furlong modular replacement was inserted with a 28 mm ceramic head and press-fit Furlong cup. A polyethylene insert of 10 mm thickness was used. This was successful and she returned to work as a secretary, reporting no limitation of activity.
Anzeige
Seven years later she presented to her family doctor complaining of a recurrence of right hip pain. An X-ray was performed and was reported as normal. Her pain persisted and a repeat X-ray was again reported as normal. She was reviewed in our institution 21 months after her recurrence of pain. Examination revealed a good range of movement in all planes, but there was gross audible crepitus. The right leg was shorter than the left. A comparison of the X-ray films is shown in Fig. 1a and b.
×
There were no radiolucencies and no measurable migration of either component was present. There was no osteolysis. On the later film there was very evident failure of the polyethylene liner and migration of the ceramic head.
At revision the femur was found to be solidly fixed. Tissues surrounding the acetabulum were heavily stained with black metalosis. The femoral head had eroded through the polyethylene liner and continued through the roof of the titanium acetabular component to rest on the acetabular bone (Fig. 2). The acetabular prosthesis, despite this, remained securely fixed. There was no suggestion of infection and no abnormality in the original positioning of the components. The acetabular component was revised to a similar 52 mm HA-coated CSF Furlong cup secured with four cancellous screws. The head was replaced with a 28 mm cobalt chrome Furlong head. This patient recently required further acetabular revision 11 years post initial primary arthroplasty. On this occasion, the acetabulum was exchanged for a cemented Ogee (De Puy, Leeds, UK) cup due to persistant pain and evidence of acetabular aseptic loosening. Again the femoral component was solidly fixed at operation and left in situ. According to the images in Figs. 3 and 4, there was evident resorption of the femoral calcar, but this does not appear to have adversely affected femoral stability.
×
×
×
In a similar case, a 58-year-old lady was referred to our clinic complaining of left hip symptoms three years after having undergone bilateral uncemented Furlong replacements for primary hip osteoarthritis. Hydroxyapatite-coated, titanium-threaded cups with 8 mm polyethylene inserts were fitted on the acetabular side and alumina ceramic heads were used in an effort to minimise wear. She made a good recovery and returned to work as a teacher but was urgently referred due to gradually increasing pain in the left groin and an audible squeak. X-rays showed no evidence of acetabular or femoral stem loosening; however, the femoral head lay eccentric in the polyethylene insert (Fig 5).
×
Anzeige
During revision there was moderate metalosis and staining of the tissues. The acetabular and femoral components were solidly fixed in position; however, the ceramic head had eroded through both the lip of the polyethylene liner and the lateral rim of the titanium cup (Fig. 6). There was no evidence of infection or prosthesis loosening. Again this was revised to a similar uncemented Furlong HA-coated acetabular component, and the patient initially appeared to make a good recovery.
×
Discussion
These cases illustrate dramatic abrasion of the polyethylene, cancellous screws and metal acetabulum. In the first example, this occurred in the presence of no apparent prosthesis malpositioning; however, in the later case, a rather steep cup inclination with a cup abduction angle of 52° may have played a contributing role. Neither case showed evidence of defective implant manufacture, nor were there initial radiographic signs of osteolysis, migration or loosening. All components were found to be solidly fixed at the time of revision and the femoral components were left in situ.
The Furlong hip replacement is an uncemented, fully hydroxyapatite-coated arthroplasty system that has attracted attention due to encouraging reported survival [1‐3]. McNally, at a mean follow-up of ten years, found a 97% survival for the acetabular component and 99% for the femoral stem. At 13–15 years, Shetty reported 99% replacement survival, and Tindall et al. also describe 99% survival of the cup at 13 years. Other authors have specifically recommended its use in the young replacement patient, reporting positive results in those aged between 21 and 55 years [4‐6]. Robertson described a ten-year stem survival of 95 and 98% for the acetabular side. There is one note of caution, however, with Sharma reporting disappointing acetabular survival. In 77 cases, 17 cups were revised within seven years. All were revised for aseptic loosening; none showed signs of excessive polyethylene wear. He reported a cup survival of 79% [7].
We report catastrophic failure of the polyethylene acetabular liner in active patients who became symptomatic in the early to intermediate time frame post replacement. This is of particular note due to the initial lack of associated osteolysis and also due to the failure of the titanium shell in conjunction with a ceramic head. Such a failure has not previously been described in the Furlong hip replacement.
Femoral heads in the Furlong system are available in either cobalt chrome or alumina oxide ceramic, and there is a choice of polyethylene or ceramic acetabular liners. The ultrahigh molecular weight polyethylene insert is gamma irradiated in the absence of oxygen and is available in a variety of thicknesses, the minimum of which is 8 mm. There has been an inclination toward the use of ceramics in young, active patients due to their perceived advantages of low friction and chemical inertness. Authors have described a significantly lower polyethylene wear rate when comparing ceramic heads and cobalt chrome [8, 9]. There is, however, contradictory literature [10, 11]. Needham et al. [12] have also recently described a case of catastrophic failure in a ceramic-polyethylene bearing couple.
If the risk of polyethylene wear is recognised early, with modular prostheses such as the Furlong, it may be possible to replace the affected liner [13]. However, if allowed to progress and affect the acetabulum, one faces the prospect of revising components in a field contaminated with abrasive particulate debris. This occurred in both of these cases, and now both of our patients have undergone or are currently awaiting further acatabular revision.
Conflict of interest statement
The authors declare that they have no conflict of interest related to the publication of this article.
Open Access
This article is distributed under the terms of the Creative Commons Attribution 2.0 International License (
https://creativecommons.org/licenses/by/2.0
), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.
In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.
Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.
Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.
Update Orthopädie und Unfallchirurgie
Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.