Gamma-irradiation Sterilization in an Inert Environment: A Partial Solution

verfasst von: Daniel J. Berry, MD, Barbara H. Currier, MChE, Michael B. Mayor, MD, John P. Collier, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 7/2012

Einloggen, um Zugang zu erhalten

Abstract

Background

In the mid to late 1990s, to sterilize UHMWPE bearings, manufacturers changed from gamma-irradiation-in-air (gamma-air) sterilization, which initiated oxidation leading to bearing fatigue, to gamma-irradiation sterilization in an inert environment (gamma-inert). The change to gamma-inert sterilization reportedly prevented shelf oxidation before implantation but not in vivo oxidation.

Questions/purposes

We asked: (1) Has the change to gamma-inert sterilization prevented shelf oxidation that led to early in vivo fatigue damage in gamma-air-sterilized tibial inserts? And (2) has the change to gamma-inert sterilization prevented the occurrence of fatigue secondary to in vivo oxidation?

Methods

We rated 183 retrieved gamma-air- and 175 retrieved gamma-inert-sterilized tibial inserts for clinical fatigue damage and analyzed 132 gamma-air- and 174 gamma-inert-sterilized tibial inserts for oxidation by Fourier transform infrared spectroscopy.

Results

Oxidation led to decreased mechanical properties in shelf-aged gamma-air-sterilized tibial inserts. Barrier packaging prevented shelf oxidation in gamma-inert-sterilized tibial inserts. Gamma-air- and gamma-inert-sterilized inserts oxidized in vivo. Fatigue damage (delamination) occurred more frequently in inserts retrieved after longer time in vivo. Longer in vivo time correlated with higher oxidation and more accumulated cycles of use.

Conclusions

Published oxidation projections suggest gamma-inert-sterilized tibial inserts would reach the critical oxidation for the onset of fatigue after 11 to 14 years in vivo. These retrievals appear to follow the projected oxidation trends. Frequency of fatigue damage increased with increasing oxidation.

Clinical Relevance

Fatigue of tibial inserts becomes more likely, especially in active patients, after more than a decade of good clinical performance.
Literatur
1.
Zurück zum Zitat American Society for Testing and Materials. ASTM F2102-06. Standard Guide for Evaluating the Extent of Oxidation in Ultra-high-molecular-weight Polyethylene Fabricated Forms Intended for Surgical Implants. Vol 13.01. West Conshohocken, PA: American Society for Testing and Materials; 2006. American Society for Testing and Materials. ASTM F2102-06. Standard Guide for Evaluating the Extent of Oxidation in Ultra-high-molecular-weight Polyethylene Fabricated Forms Intended for Surgical Implants. Vol 13.01. West Conshohocken, PA: American Society for Testing and Materials; 2006.
2.
Zurück zum Zitat American Society for Testing and Materials. ASTM D638-10. Standard Test Method for Tensile Properties of Plastics. Vol 08.01. West Conshohocken, PA: American Society for Testing and Materials; 2010. American Society for Testing and Materials. ASTM D638-10. Standard Test Method for Tensile Properties of Plastics. Vol 08.01. West Conshohocken, PA: American Society for Testing and Materials; 2010.
3.
Zurück zum Zitat American Society for Testing and Materials. ASTM F648-10a. Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants. Vol 13.01. West Conshohocken, PA: American Society for Testing and Materials; 2010. American Society for Testing and Materials. ASTM F648-10a. Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants. Vol 13.01. West Conshohocken, PA: American Society for Testing and Materials; 2010.
4.
Zurück zum Zitat Bargmann LS, Bargmann BC, Collier JP, Currier BH, Mayor MB. Current sterilization and packaging methods for polyethylene. Clin Orthop Relat Res. 1999;369:49–58.PubMedCrossRef Bargmann LS, Bargmann BC, Collier JP, Currier BH, Mayor MB. Current sterilization and packaging methods for polyethylene. Clin Orthop Relat Res. 1999;369:49–58.PubMedCrossRef
5.
Zurück zum Zitat Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMedCrossRef Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMedCrossRef
6.
Zurück zum Zitat Costa L, Bracco P, Brach Del Prever EM, Kurtz SM, Gallinaro P. Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement components. J Biomed Mater Res B Appl Biomater. 2006;78:20–26.PubMed Costa L, Bracco P, Brach Del Prever EM, Kurtz SM, Gallinaro P. Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement components. J Biomed Mater Res B Appl Biomater. 2006;78:20–26.PubMed
7.
Zurück zum Zitat Currier BH, Currier JH, Collier JP, Mayor MB, Scott RD. Shelf life and in vivo duration: impacts on performance of tibial bearings. Clin Orthop Relat Res. 1997;342:111–122.PubMedCrossRef Currier BH, Currier JH, Collier JP, Mayor MB, Scott RD. Shelf life and in vivo duration: impacts on performance of tibial bearings. Clin Orthop Relat Res. 1997;342:111–122.PubMedCrossRef
8.
Zurück zum Zitat Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved Crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMedCrossRef Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved Crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMedCrossRef
9.
Zurück zum Zitat Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMedCrossRef Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMedCrossRef
10.
Zurück zum Zitat Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef
11.
Zurück zum Zitat Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and cross-linking of UHMWPE for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.PubMedCrossRef Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and cross-linking of UHMWPE for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.PubMedCrossRef
12.
Zurück zum Zitat Medel FJ, Kurtz SM, Hozack WJ, Parvizi J, Purtill JJ, Sharkey PF, MacDonald D, Kraay MJ, Goldberg V, Rimnac CM. Gamma inert sterilization: a solution to polyethylene oxidation? J Bone Joint Surg Am. 2009;91:839–849.PubMedCrossRef Medel FJ, Kurtz SM, Hozack WJ, Parvizi J, Purtill JJ, Sharkey PF, MacDonald D, Kraay MJ, Goldberg V, Rimnac CM. Gamma inert sterilization: a solution to polyethylene oxidation? J Bone Joint Surg Am. 2009;91:839–849.PubMedCrossRef
13.
Zurück zum Zitat Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA. The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am. 2011;93:1588–1596.PubMedCrossRef Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA. The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am. 2011;93:1588–1596.PubMedCrossRef
14.
Zurück zum Zitat Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR, Kasprzak DJ, Surprenant VA. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res. 1995;319:28–40.PubMed Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR, Kasprzak DJ, Surprenant VA. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res. 1995;319:28–40.PubMed
Metadaten
Titel
Gamma-irradiation Sterilization in an Inert Environment: A Partial Solution
verfasst von
Daniel J. Berry, MD
Barbara H. Currier, MChE
Michael B. Mayor, MD
John P. Collier, PhD
Publikationsdatum
01.07.2012
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 7/2012
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-011-2150-1

Weitere Artikel der Ausgabe 7/2012

Clinical Orthopaedics and Related Research® 7/2012 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.