01.05.2009 | Technical Note

First Magnetic Resonance Imaging-Guided Aortic Stenting and Cava Filter Placement Using a Polyetheretherketone-Based Magnetic Resonance Imaging-Compatible Guidewire in Swine: Proof of Concept

verfasst von: Sebastian Kos, Rolf Huegli, Eugen Hofmann, Harald H. Quick, Hilmar Kuehl, Stephanie Aker, Gernot M. Kaiser, Paul J. A. Borm, Augustinus L. Jacob, Deniz Bilecen

Erschienen in: CardioVascular and Interventional Radiology | Ausgabe 3/2009

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study was to demonstrate feasibility of percutaneous transluminal aortic stenting and cava filter placement under magnetic resonance imaging (MRI) guidance exclusively using a polyetheretherketone (PEEK)-based MRI-compatible guidewire. Percutaneous transluminal aortic stenting and cava filter placement were performed in 3 domestic swine. Procedures were performed under MRI-guidance in an open-bore 1.5-T scanner. The applied 0.035-inch guidewire has a PEEK core reinforced by fibres, floppy tip, hydrophilic coating, and paramagnetic markings for passive visualization. Through an 11F sheath, the guidewire was advanced into the abdominal (swine 1) or thoracic aorta (swine 2), and the stents were deployed. The guidewire was advanced into the inferior vena cava (swine 3), and the cava filter was deployed. Postmortem autopsy was performed. Procedural success, guidewire visibility, pushability, and stent support were qualitatively assessed by consensus. Procedure times were documented. Guidewire guidance into the abdominal and thoracic aortas and the inferior vena cava was successful. Stent deployments were successful in the abdominal (swine 1) and thoracic (swine 2) segments of the descending aorta. Cava filter positioning and deployment was successful. Autopsy documented good stent and filter positioning. Guidewire visibility through applied markers was rated acceptable for aortic stenting and good for venous filter placement. Steerability, pushability, and device support were good. The PEEK-based guidewire allows either percutaneous MRI-guided aortic stenting in the thoracic and abdominal segments of the descending aorta and filter placement in the inferior vena cava with acceptable to good device visibility and offers good steerability, pushability, and device support.
Literatur
1.
Zurück zum Zitat Kos S, Huegli R, Bongartz G et al (2008) MR-guided endovascular interventions: a comprehensive review on techniques and applications. Eur Radiol 18:645–657PubMedCrossRef Kos S, Huegli R, Bongartz G et al (2008) MR-guided endovascular interventions: a comprehensive review on techniques and applications. Eur Radiol 18:645–657PubMedCrossRef
2.
Zurück zum Zitat Bock M, Wacker F (2008) MR-guided intravascular interventions: techniques and applications. J Magn Reson Imaging 27:326–338PubMedCrossRef Bock M, Wacker F (2008) MR-guided intravascular interventions: techniques and applications. J Magn Reson Imaging 27:326–338PubMedCrossRef
3.
Zurück zum Zitat Mahnken A, Chalabi K, Jalali F et al (2004) Magnetic resonance-guided placement of aortic stents grafts: feasibility with real-time magnetic resonance fluoroscopy. J Vasc Interv Radiol 15:189–195PubMed Mahnken A, Chalabi K, Jalali F et al (2004) Magnetic resonance-guided placement of aortic stents grafts: feasibility with real-time magnetic resonance fluoroscopy. J Vasc Interv Radiol 15:189–195PubMed
4.
Zurück zum Zitat Omary RA, Frayne R, Unal O et al (2000) MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study. J Vasc Interv Radiol 11:373–381PubMedCrossRef Omary RA, Frayne R, Unal O et al (2000) MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study. J Vasc Interv Radiol 11:373–381PubMedCrossRef
5.
Zurück zum Zitat Bucker A, Neuerburg J, Adam G et al (2003) MR-guided coil embolisation of renal arteries in an animal model. Rofo 175:271–274PubMed Bucker A, Neuerburg J, Adam G et al (2003) MR-guided coil embolisation of renal arteries in an animal model. Rofo 175:271–274PubMed
6.
Zurück zum Zitat Eggebrecht H, Kuhl H, Kaiser GM et al (2006) Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. Eur Heart J 27:613–620PubMedCrossRef Eggebrecht H, Kuhl H, Kaiser GM et al (2006) Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. Eur Heart J 27:613–620PubMedCrossRef
7.
Zurück zum Zitat Nitz WR, Oppelt A, Renz W et al (2001) On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging 13:105–114PubMedCrossRef Nitz WR, Oppelt A, Renz W et al (2001) On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging 13:105–114PubMedCrossRef
8.
Zurück zum Zitat Omary R, Gehl J, Schirf B et al (2006) MR imaging versus conventional X-ray fluoroscopy-guided renal angioplasty in swine: prospective randomized comparison. Radiology 238:489–496PubMedCrossRef Omary R, Gehl J, Schirf B et al (2006) MR imaging versus conventional X-ray fluoroscopy-guided renal angioplasty in swine: prospective randomized comparison. Radiology 238:489–496PubMedCrossRef
9.
Zurück zum Zitat Konings MK, Bartels LW, Smits HF et al (2000) Heating around intravascular guidewires by resonating RF waves. J Magn Reson Imaging 12:79–85PubMedCrossRef Konings MK, Bartels LW, Smits HF et al (2000) Heating around intravascular guidewires by resonating RF waves. J Magn Reson Imaging 12:79–85PubMedCrossRef
10.
Zurück zum Zitat Buecker A, Spuentrup E, Schmitz-Rode T et al (2004) Use of a nonmetallic guide wire for magnetic resonance-guided coronary artery catheterization. Invest Radiol 39:656–660PubMedCrossRef Buecker A, Spuentrup E, Schmitz-Rode T et al (2004) Use of a nonmetallic guide wire for magnetic resonance-guided coronary artery catheterization. Invest Radiol 39:656–660PubMedCrossRef
11.
Zurück zum Zitat Mekle R, Hofmann E, Scheffler K et al (2006) A polymer-based MR-compatible guidewire: a study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). J Magn Reson Imaging 23:145–155PubMedCrossRef Mekle R, Hofmann E, Scheffler K et al (2006) A polymer-based MR-compatible guidewire: a study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). J Magn Reson Imaging 23:145–155PubMedCrossRef
12.
Zurück zum Zitat Kaiser G, Breuckmann F, Aker S et al (2007) Anesthesia for cardiovascular interventions and magnetic resonance imaging in pigs. J Am Assoc Lab Anim Sci 46:30–33PubMed Kaiser G, Breuckmann F, Aker S et al (2007) Anesthesia for cardiovascular interventions and magnetic resonance imaging in pigs. J Am Assoc Lab Anim Sci 46:30–33PubMed
13.
Zurück zum Zitat Neuerburg J, Bucker A, Adam G et al (1997) Cava filter placement under MRI control. Experimental in vitro and in vivo studies. Rofo 167:418–422PubMed Neuerburg J, Bucker A, Adam G et al (1997) Cava filter placement under MRI control. Experimental in vitro and in vivo studies. Rofo 167:418–422PubMed
14.
Zurück zum Zitat Shih MC, Rogers WJ, Hagspiel KD (2006) Real-time magnetic resonance-guided placement of retrievable inferior vena cava filters: comparison with fluoroscopic guidance with use of in vitro and animal models. J Vasc Interv Radiol 17:327–333PubMedCrossRef Shih MC, Rogers WJ, Hagspiel KD (2006) Real-time magnetic resonance-guided placement of retrievable inferior vena cava filters: comparison with fluoroscopic guidance with use of in vitro and animal models. J Vasc Interv Radiol 17:327–333PubMedCrossRef
15.
Zurück zum Zitat Bucker A, Neuerburg JM, Adam GB et al (2001) Real-time MR guidance for inferior vena cava filter placement in an animal model. J Vasc Interv Radiol 12:753–756PubMedCrossRef Bucker A, Neuerburg JM, Adam GB et al (2001) Real-time MR guidance for inferior vena cava filter placement in an animal model. J Vasc Interv Radiol 12:753–756PubMedCrossRef
16.
Zurück zum Zitat Glowinski A, Adam G, Bucker A et al (1997) Catheter visualization using locally induced, actively controlled field inhomogeneities. Magn Reson Med 38:253–258PubMedCrossRef Glowinski A, Adam G, Bucker A et al (1997) Catheter visualization using locally induced, actively controlled field inhomogeneities. Magn Reson Med 38:253–258PubMedCrossRef
17.
Zurück zum Zitat Dumoulin CL, Souza SP, Darrow RD (1993) Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med 29:411–415PubMedCrossRef Dumoulin CL, Souza SP, Darrow RD (1993) Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med 29:411–415PubMedCrossRef
18.
Zurück zum Zitat Quick HH, Kuehl H, Kaiser G et al (2003) Interventional MRA using actively visualized catheters, TrueFISP, and real-time image fusion. Magn Reson Med 49:129–137PubMedCrossRef Quick HH, Kuehl H, Kaiser G et al (2003) Interventional MRA using actively visualized catheters, TrueFISP, and real-time image fusion. Magn Reson Med 49:129–137PubMedCrossRef
19.
Zurück zum Zitat Quick HH, Zenge MO, Kuehl H et al (2005) Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization. Magn Reson Med 53:446–455PubMedCrossRef Quick HH, Zenge MO, Kuehl H et al (2005) Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization. Magn Reson Med 53:446–455PubMedCrossRef
20.
Zurück zum Zitat Ocali O, Atalar E (1997) Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med 37:112–118PubMedCrossRef Ocali O, Atalar E (1997) Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med 37:112–118PubMedCrossRef
21.
Zurück zum Zitat Rauschenberg J, de Oliveira A, Muller S et al (2007) An algorithm for passive marker localization in interventional MRI. Z Med Phys 17:180–189PubMed Rauschenberg J, de Oliveira A, Muller S et al (2007) An algorithm for passive marker localization in interventional MRI. Z Med Phys 17:180–189PubMed
22.
Zurück zum Zitat Shellock FG (2008) Reference manual for magnetic resonance safety, implants and devices: 2008 edition. Biomedical Research Publishing Group, Los Angeles CA, p 422 Shellock FG (2008) Reference manual for magnetic resonance safety, implants and devices: 2008 edition. Biomedical Research Publishing Group, Los Angeles CA, p 422
23.
Zurück zum Zitat Leiner T, Schoenberg S (2007) Current status of renal artery magnetic resonance imaging: theoretical and practical considerations and the potential role of blood-pool contrast agents. Eur Radiol 17(Suppl. 2):B13–B17PubMed Leiner T, Schoenberg S (2007) Current status of renal artery magnetic resonance imaging: theoretical and practical considerations and the potential role of blood-pool contrast agents. Eur Radiol 17(Suppl. 2):B13–B17PubMed
24.
Zurück zum Zitat Meaney J, Goyen M (2007) Recent advances in contrast-enhanced magnetic resonance angiography. Eur Radiol 17(Suppl 2):B2–B6PubMed Meaney J, Goyen M (2007) Recent advances in contrast-enhanced magnetic resonance angiography. Eur Radiol 17(Suppl 2):B2–B6PubMed
25.
Zurück zum Zitat Bremerich J, Bilecen D, Reimer P (2007) MR angiography with blood pool contrast agents. Eur Radiol 17:3017–3024PubMedCrossRef Bremerich J, Bilecen D, Reimer P (2007) MR angiography with blood pool contrast agents. Eur Radiol 17:3017–3024PubMedCrossRef
Metadaten
Titel
First Magnetic Resonance Imaging-Guided Aortic Stenting and Cava Filter Placement Using a Polyetheretherketone-Based Magnetic Resonance Imaging-Compatible Guidewire in Swine: Proof of Concept
verfasst von
Sebastian Kos
Rolf Huegli
Eugen Hofmann
Harald H. Quick
Hilmar Kuehl
Stephanie Aker
Gernot M. Kaiser
Paul J. A. Borm
Augustinus L. Jacob
Deniz Bilecen
Publikationsdatum
01.05.2009
Verlag
Springer-Verlag
Erschienen in
CardioVascular and Interventional Radiology / Ausgabe 3/2009
Print ISSN: 0174-1551
Elektronische ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-008-9483-5

Weitere Artikel der Ausgabe 3/2009

CardioVascular and Interventional Radiology 3/2009 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.