01.04.2013 | Clinical Investigation

Patient Dose During Carotid Artery Stenting With Embolic-Protection Devices: Evaluation With Radiochromic Films and Related Diagnostic Reference Levels According to Factors Influencing the Procedure

verfasst von: Loredana D’Ercole, Pietro Quaretti, Nicola Cionfoli, Catherine Klersy, Milena Bocchiola, Giuseppe Rodolico, Andrea Azzaretti, Francesco Lisciandro, Tommaso Cascella, Federico Zappoli Thyrion

Erschienen in: CardioVascular and Interventional Radiology | Ausgabe 2/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To measure the maximum entrance skin dose (MESD) on patients undergoing carotid artery stenting (CAS) using embolic-protection devices, to analyze the dependence of dose and exposure parameters on anatomical, clinical, and technical factors affecting the procedure complexity, to obtain some local diagnostic reference levels (DRLs), and to evaluate whether overcoming DRLs is related to procedure complexity.

Materials and Methods

MESD were evaluated with radiochromic films in 31 patients (mean age 72 ± 7 years). Five of 33 (15 %) procedures used proximal EPD, and 28 of 33 (85 %) procedures used distal EPD. Local DRLs were derived from the recorded exposure parameters in 93 patients (65 men and 28 women, mean age 73 ± 9 years) undergoing 96 CAS with proximal (33 %) or distal (67 %) EPD. Four bilateral lesions were included.

Results

MESD values (mean 0.96 ± 0.42 Gy) were <2 Gy without relevant dependence on procedure complexity. Local DRL values for kerma area product (KAP), fluoroscopy time (FT), and number of frames (NFR) were 269 Gy cm2, 28 minutes, and 251, respectively. Only simultaneous bilateral treatment was associated with KAP (odds ratio [OR] 10.14, 95 % confidence interval [CI] 1–102.7, p < 0.05) and NFR overexposures (OR 10.8, 95 % CI 1.1–109.5, p < 0.05). Type I aortic arch decreased the risk of FT overexposure (OR 0.4, 95 % CI 0.1–0.9, p = 0.042), and stenosis ≥ 90 % increased the risk of NFR overexposure (OR 2.8, 95 % CI 1.1–7.4, p = 0.040). At multivariable analysis, stenosis ≥ 90 % (OR 2.8, 95 % CI 1.1–7.4, p = 0.040) and bilateral treatment (OR 10.8, 95 % CI 1.1–109.5, p = 0.027) were associated with overexposure for two or more parameters.

Conclusion

Skin doses are not problematic in CAS with EPD because these procedures rarely lead to doses >2 Gy.
Literatur
1.
Zurück zum Zitat Yadav JS, Wholey MH, Kuntz RE et al (2004) Protected carotid-artery stenting versus endarterectomy in high-risk patients. N Engl J Med 351:1493–1501PubMedCrossRef Yadav JS, Wholey MH, Kuntz RE et al (2004) Protected carotid-artery stenting versus endarterectomy in high-risk patients. N Engl J Med 351:1493–1501PubMedCrossRef
2.
Zurück zum Zitat Mas JL, Chatellier G, Beyssen B et al (2006) Endarterectomy versus stenting in patients with symptomatic severe carotid stenosis. N Engl J Med 355:1660–1671PubMedCrossRef Mas JL, Chatellier G, Beyssen B et al (2006) Endarterectomy versus stenting in patients with symptomatic severe carotid stenosis. N Engl J Med 355:1660–1671PubMedCrossRef
3.
Zurück zum Zitat Ringleb PA, Allenberg J, Berger J et al (2006) 30 day results from the SPACE trial of stent-protected angioplasty versus carotid endarterectomy in symptomatic patients: A randomised non-inferiority trial. Lancet 368:1239–1247PubMedCrossRef Ringleb PA, Allenberg J, Berger J et al (2006) 30 day results from the SPACE trial of stent-protected angioplasty versus carotid endarterectomy in symptomatic patients: A randomised non-inferiority trial. Lancet 368:1239–1247PubMedCrossRef
4.
Zurück zum Zitat Ederle J, Dobson J, Featherstone RL et al (2010) Carotid artery stenting compared with endarterectomy in patients with symptomatic carotid stenosis (International Carotid Stenting Study): An interim analysis of a randomised controlled trial. Lancet 375:985–997PubMedCrossRef Ederle J, Dobson J, Featherstone RL et al (2010) Carotid artery stenting compared with endarterectomy in patients with symptomatic carotid stenosis (International Carotid Stenting Study): An interim analysis of a randomised controlled trial. Lancet 375:985–997PubMedCrossRef
5.
Zurück zum Zitat Brott TG, Hobson RW, George H et al (2010) Stenting versus endarterectomy for treatment of carotid-artery stenosis. N Engl J Med 363(1):11–23PubMedCrossRef Brott TG, Hobson RW, George H et al (2010) Stenting versus endarterectomy for treatment of carotid-artery stenosis. N Engl J Med 363(1):11–23PubMedCrossRef
6.
Zurück zum Zitat Biasi GM, Froio A, Diethrich EB et al (2004) Carotid plaque echolucency increases the risk of stroke in carotid stenting: The imaging in carotid angioplasty and risk of stroke (ICAROS) study. Circulation 110:756–762PubMedCrossRef Biasi GM, Froio A, Diethrich EB et al (2004) Carotid plaque echolucency increases the risk of stroke in carotid stenting: The imaging in carotid angioplasty and risk of stroke (ICAROS) study. Circulation 110:756–762PubMedCrossRef
7.
Zurück zum Zitat Faggioli GL, Ferri M, Freyrie A et al (2007) Aortic arch anomalies are associated with increased risk of neurological events in carotid stent procedures. Eur J Vasc Endovasc Surg 33(4):436–441PubMedCrossRef Faggioli GL, Ferri M, Freyrie A et al (2007) Aortic arch anomalies are associated with increased risk of neurological events in carotid stent procedures. Eur J Vasc Endovasc Surg 33(4):436–441PubMedCrossRef
8.
Zurück zum Zitat Chiam PT, Roubin GS, Ivyer SS et al (2008) Carotid artery stenting in elderly patients: Importance of case selection. Catheter Cardiovasc Interv 72:318–324PubMedCrossRef Chiam PT, Roubin GS, Ivyer SS et al (2008) Carotid artery stenting in elderly patients: Importance of case selection. Catheter Cardiovasc Interv 72:318–324PubMedCrossRef
9.
Zurück zum Zitat GAFCHROMIC® XR Type R Radiochromic Dosimetry Film. Characteristic performance data. International Specialty Products. Available at: http://www.ispcorp.com. Accessed GAFCHROMIC® XR Type R Radiochromic Dosimetry Film. Characteristic performance data. International Specialty Products. Available at: http://​www.​ispcorp.​com. Accessed
10.
Zurück zum Zitat Mantovani L, D’Ercole L, Lisciandro F et al (2006) Radiochromic films for improved evaluation of patient dose in liver interventions. J Vasc Interv Radiol 17(5):855–862PubMedCrossRef Mantovani L, D’Ercole L, Lisciandro F et al (2006) Radiochromic films for improved evaluation of patient dose in liver interventions. J Vasc Interv Radiol 17(5):855–862PubMedCrossRef
11.
Zurück zum Zitat D’Ercole L, Mantovani L, Zappoli Thyrion F et al (2007) A study on maximum skin dose in cerebral embolization procedures with radiochromic films. AJNR Am J Neuroradiol 28:503–507PubMed D’Ercole L, Mantovani L, Zappoli Thyrion F et al (2007) A study on maximum skin dose in cerebral embolization procedures with radiochromic films. AJNR Am J Neuroradiol 28:503–507PubMed
12.
Zurück zum Zitat Delle Canne S, Carosi A, Bufacchi A et al (2006) Use of GAFCHROMIC XR type R films for skin-dose measurements in interventional radiology: Validation of a dosimetric procedure on a sample of patients undergoing interventional cardiology. Physica Medica 22(3):105–110PubMedCrossRef Delle Canne S, Carosi A, Bufacchi A et al (2006) Use of GAFCHROMIC XR type R films for skin-dose measurements in interventional radiology: Validation of a dosimetric procedure on a sample of patients undergoing interventional cardiology. Physica Medica 22(3):105–110PubMedCrossRef
13.
Zurück zum Zitat Faj D, Steiner R, Trifunovic D et al. (2008) Patient dosimetry in interventional cardiology at the university Hospital of Osljek. Radiat Protect Dosim 128(4):485–490CrossRef Faj D, Steiner R, Trifunovic D et al. (2008) Patient dosimetry in interventional cardiology at the university Hospital of Osljek. Radiat Protect Dosim 128(4):485–490CrossRef
14.
Zurück zum Zitat Balter S, Miller DL, Vano E et al (2008) A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys 35(2):673–680PubMedCrossRef Balter S, Miller DL, Vano E et al (2008) A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys 35(2):673–680PubMedCrossRef
15.
Zurück zum Zitat Giordano C, D’Ercole L, Gobbi R et al (2010) Coronary angiography and percutaneous transluminal coronary angioplasty procedures: Evaluation of patients’ maximum skin dose using Gafchromic films and a comparison of local levels with reference levels proposed in the literature. Physica Medica 26:224–232PubMedCrossRef Giordano C, D’Ercole L, Gobbi R et al (2010) Coronary angiography and percutaneous transluminal coronary angioplasty procedures: Evaluation of patients’ maximum skin dose using Gafchromic films and a comparison of local levels with reference levels proposed in the literature. Physica Medica 26:224–232PubMedCrossRef
16.
Zurück zum Zitat Ying CK, Kandaiya S (2010) Patient skin dose measurements during coronary interventional procedures using Gafchromic film. J Radiol Prot 30:585–596PubMedCrossRef Ying CK, Kandaiya S (2010) Patient skin dose measurements during coronary interventional procedures using Gafchromic film. J Radiol Prot 30:585–596PubMedCrossRef
17.
Zurück zum Zitat D’Ercole L, Azzaretti A, Zappoli Thyrion F et al (2010) Measurement of patient skin dose in vertebroplasty using radiochromic dosimetry film. Spine 35:1304–1306PubMed D’Ercole L, Azzaretti A, Zappoli Thyrion F et al (2010) Measurement of patient skin dose in vertebroplasty using radiochromic dosimetry film. Spine 35:1304–1306PubMed
18.
Zurück zum Zitat International Commission on Radiological Protection (2000) ICRP Publication 85. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30:7–67 International Commission on Radiological Protection (2000) ICRP Publication 85. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30:7–67
19.
Zurück zum Zitat European Commission (2000) Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiat Protect 109:4–25 European Commission (2000) Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiat Protect 109:4–25
20.
Zurück zum Zitat Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part II: Skin dose. J Vasc Interv Radiol 14:977–990PubMedCrossRef Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part II: Skin dose. J Vasc Interv Radiol 14:977–990PubMedCrossRef
21.
Zurück zum Zitat Fletcher DW, Miller DL, Balter S et al (2002) Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. J Vasc Interv Radiol 13:391–397PubMedCrossRef Fletcher DW, Miller DL, Balter S et al (2002) Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. J Vasc Interv Radiol 13:391–397PubMedCrossRef
22.
Zurück zum Zitat Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part I: Overall measures of dose. J Vasc Interv Radiol 14:711–727PubMedCrossRef Miller DL, Balter S, Cole PE et al (2003) Radiation dose in interventional radiology procedures: The RAD-IR study part I: Overall measures of dose. J Vasc Interv Radiol 14:711–727PubMedCrossRef
23.
Zurück zum Zitat Bor D, Türkay T, Olgar T et al (2006) Variations of patient doses in interventional examinations at different angiographic units. Cardiovasc Interv Radiol 29:797–806CrossRef Bor D, Türkay T, Olgar T et al (2006) Variations of patient doses in interventional examinations at different angiographic units. Cardiovasc Interv Radiol 29:797–806CrossRef
24.
Zurück zum Zitat European Union (1997) Council Directive 97/43 Euratom of 30 June 1997, Official Journal of the European Communities, 1997; Legislation L. 180 European Union (1997) Council Directive 97/43 Euratom of 30 June 1997, Official Journal of the European Communities, 1997; Legislation L. 180
25.
Zurück zum Zitat Gianfelice D, Lepanto L, Perreault P et al (2000) Effect of the learning process on procedure times and radiation exposure for CT fluoroscopy-guided percutaneous biopsy procedures. J Vasc Interv Radiol 11:1217–1221PubMedCrossRef Gianfelice D, Lepanto L, Perreault P et al (2000) Effect of the learning process on procedure times and radiation exposure for CT fluoroscopy-guided percutaneous biopsy procedures. J Vasc Interv Radiol 11:1217–1221PubMedCrossRef
26.
Zurück zum Zitat Andrews RT, Brown PH (2000) Uterine arterial embolization: Factors influencing patient radiation exposure. Radiology 217:713–722PubMed Andrews RT, Brown PH (2000) Uterine arterial embolization: Factors influencing patient radiation exposure. Radiology 217:713–722PubMed
27.
Zurück zum Zitat Watson LE, Riggs MW, Bourland PD (1997) Radiation exposure during cardiology fellowship training. Health Phys 73:690–693PubMedCrossRef Watson LE, Riggs MW, Bourland PD (1997) Radiation exposure during cardiology fellowship training. Health Phys 73:690–693PubMedCrossRef
28.
Zurück zum Zitat Verdun FR, Aroua A, Trueb R et al (2005) Diagnostic and interventional radiology: A strategy to introduce reference dose level taking into account the national practice. Radiat Prot Dosimetry 114(1–3):188–191PubMedCrossRef Verdun FR, Aroua A, Trueb R et al (2005) Diagnostic and interventional radiology: A strategy to introduce reference dose level taking into account the national practice. Radiat Prot Dosimetry 114(1–3):188–191PubMedCrossRef
29.
Zurück zum Zitat Aroua A, Rickli H, Stauffer JC et al (2007) How to set up and apply reference levels in fluoroscopy at a national level? Eur Radiol 17:1621–1633PubMedCrossRef Aroua A, Rickli H, Stauffer JC et al (2007) How to set up and apply reference levels in fluoroscopy at a national level? Eur Radiol 17:1621–1633PubMedCrossRef
30.
Zurück zum Zitat Stratis AI, Anthopoulos PL, Gavaliatsis IP et al (2009) Patient dose in cardiac radiology. Hellenic J Cardiol 50:17–25PubMed Stratis AI, Anthopoulos PL, Gavaliatsis IP et al (2009) Patient dose in cardiac radiology. Hellenic J Cardiol 50:17–25PubMed
Metadaten
Titel
Patient Dose During Carotid Artery Stenting With Embolic-Protection Devices: Evaluation With Radiochromic Films and Related Diagnostic Reference Levels According to Factors Influencing the Procedure
verfasst von
Loredana D’Ercole
Pietro Quaretti
Nicola Cionfoli
Catherine Klersy
Milena Bocchiola
Giuseppe Rodolico
Andrea Azzaretti
Francesco Lisciandro
Tommaso Cascella
Federico Zappoli Thyrion
Publikationsdatum
01.04.2013
Verlag
Springer-Verlag
Erschienen in
CardioVascular and Interventional Radiology / Ausgabe 2/2013
Print ISSN: 0174-1551
Elektronische ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-012-0392-2

Weitere Artikel der Ausgabe 2/2013

CardioVascular and Interventional Radiology 2/2013 Zur Ausgabe

CIRSE Standards of Practice Guidelines

Below-the-knee Interventions

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.