Accurate estimation of global and regional cardiac function by retrospectively gated multidetector row computed tomography
Comparison with cine magnetic resonance imaging
verfasst von:
Bénédicte Belge, Emmanuel Coche, Agnès Pasquet, Jean-Louis J. Vanoverschelde, Bernhard L. Gerber
Erschienen in:
European Radiology
|
Ausgabe 7/2006
Einloggen, um Zugang zu erhalten
Abstract
Retrospective reconstruction of ECG-gated images at different parts of the cardiac cycle allows the assessment of cardiac function by multi-detector row CT (MDCT) at the time of non-invasive coronary imaging. We compared the accuracy of such measurements by MDCT to cine magnetic resonance (MR). Forty patients underwent the assessment of global and regional cardiac function by 16-slice MDCT and cine MR. Left ventricular (LV) end-diastolic and end-systolic volumes estimated by MDCT (134±51 and 67±56 ml) were similar to those by MR (137±57 and 70±60 ml, respectively; both P=NS) and strongly correlated (r=0.92 and r=0.95, respectively; both P<0.001). Consequently, LV ejection fractions by MDCT and MR were also similar (55±21 vs. 56±21%; P=NS) and highly correlated (r=0.95; P<0.001). Regional end-diastolic and end-systolic wall thicknesses by MDCT were highly correlated (r=0.84 and r=0.92, respectively; both P<0.001), but significantly lower than by MR (8.3±1.8 vs. 8.8±1.9 mm and 12.7±3.4 vs. 13.3±3.5 mm, respectively; both P<0.001). Values of regional wall thickening by MDCT and MR were similar (54±30 vs. 51±31%; P=NS) and also correlated well (r=0.91; P<0.001). Retrospectively gated MDCT can accurately estimate LV volumes, EF and regional LV wall thickening compared to cine MR.