Erschienen in: European Radiology 6/2007

01.06.2007 | Cardiac

Visual and automatic grading of coronary artery stenoses with 64-slice CT angiography in reference to invasive angiography

verfasst von: Stephanie Busch, Thorsten R. C. Johnson, Konstantin Nikolaou, Franz von Ziegler, Andreas Knez, Maximilian F. Reiser, Christoph R. Becker

Erschienen in: European Radiology | Ausgabe 6/2007

Einloggen, um Zugang zu erhalten

Abstract

The aim of this study was to assess the performance of a software tool for quantitative coronary artery analysis of computed tomography coronary angiography (CT-QCA) in comparison with invasive coronary angiography with quantitative analysis (CAG-QCA) as standard of reference. Two radiologists reviewed the CT angiography data sets (Siemens Sensation 64) of 25 patients, grading coronary artery stenoses visually and with a software tool (Circulation, Siemens). Twenty-three data sets with sufficient image quality were included in the final analysis. CAG revealed a total of 30 wall irregularities and 28 stenoses, of which 17 were graded as moderate and nine as hemodynamically significant. CT-QCA showed a better agreement to CAG-QCA, with a systematic overestimation of the degree of stenosis of 6.1% and limits of agreement of +36.1% and −23.9; the correlation coefficient was 0.82 (p < 0.0001). Using CT-QCA, sensitivity, specificity, and positive and negative predictive value were 89%, 100%, 89%, and 100%, respectively, for significant area stenoses greater than 75%. The positive predictive value for the visual assessment amounted to 53%. Interobserver variability between CT-QCA and visual assessment showed a kappa value of 0.72. In conclusion, software-supported CT-QCA makes it possible to quantify significant coronary artery stenoses automatically, with good agreement to CAG-QCA.
Literatur
1.
Zurück zum Zitat Greuter MJ, Dorgelo J, Tukker WG, Oudkerk M (2005) Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit. Eur Radiol 15:995–1007PubMedCrossRef Greuter MJ, Dorgelo J, Tukker WG, Oudkerk M (2005) Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit. Eur Radiol 15:995–1007PubMedCrossRef
2.
Zurück zum Zitat Horiguchi J, Fukuda H, Yamamoto H, Hirai N, Alam F, Kakizawa H, Hieda M, Tachikake T, Marukawa K, Ito K (2006) The impact of motion artifacts on the reproducibility of repeated coronary artery calcium measurements. Eur Radiol DOI 10.1007/s00330-006-0278-2 Horiguchi J, Fukuda H, Yamamoto H, Hirai N, Alam F, Kakizawa H, Hieda M, Tachikake T, Marukawa K, Ito K (2006) The impact of motion artifacts on the reproducibility of repeated coronary artery calcium measurements. Eur Radiol DOI 10.​1007/​s00330-006-0278-2
3.
Zurück zum Zitat Knez A, Becker C, Ohnesorge B, Haberl R, Reiser M, Steinbeck G (2000) Noninvasive detection of coronary artery stenosis by multislice helical computed tomography. Circulation 101:E221–E222PubMed Knez A, Becker C, Ohnesorge B, Haberl R, Reiser M, Steinbeck G (2000) Noninvasive detection of coronary artery stenosis by multislice helical computed tomography. Circulation 101:E221–E222PubMed
4.
Zurück zum Zitat Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, Grasruck M, Stierstorfer K, Krauss B, Raupach R, Primak AN, Kuttner A, Achenbach S, Becker C, Kopp A, Ohnesorge BM (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, Grasruck M, Stierstorfer K, Krauss B, Raupach R, Primak AN, Kuttner A, Achenbach S, Becker C, Kopp A, Ohnesorge BM (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef
5.
Zurück zum Zitat Johnson TR, Nikolaou K, Wintersperger BJ, Leber AW, von Ziegler F, Rist C, Buhmann S, Knez A, Reiser MF, Becker CR (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef Johnson TR, Nikolaou K, Wintersperger BJ, Leber AW, von Ziegler F, Rist C, Buhmann S, Knez A, Reiser MF, Becker CR (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef
6.
Zurück zum Zitat Knez A, Becker CR, Leber A, Ohnesorge B, Becker A, White C, Haberl R, Reiser MF, Steinbeck G (2001) Usefulness of multislice spiral computed tomography angiography for determination of coronary artery stenoses. Am J Cardiol 88:1191–1194PubMedCrossRef Knez A, Becker CR, Leber A, Ohnesorge B, Becker A, White C, Haberl R, Reiser MF, Steinbeck G (2001) Usefulness of multislice spiral computed tomography angiography for determination of coronary artery stenoses. Am J Cardiol 88:1191–1194PubMedCrossRef
7.
Zurück zum Zitat Leber AW, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef Leber AW, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef
8.
Zurück zum Zitat Achenbach S, Ropers D, Hoffmann U, MacNeill B, Baum U, Pohle K, Brady TJ, Pomerantsev E, Ludwig J, Flachskampf FA, Wicky S, Jang IK, Daniel WG (2004) Assessment of coronary remodeling in stenotic and nonstenotic coronary atherosclerotic lesions by multidetector spiral computed tomography. J Am Coll Cardiol 43:842–847PubMedCrossRef Achenbach S, Ropers D, Hoffmann U, MacNeill B, Baum U, Pohle K, Brady TJ, Pomerantsev E, Ludwig J, Flachskampf FA, Wicky S, Jang IK, Daniel WG (2004) Assessment of coronary remodeling in stenotic and nonstenotic coronary atherosclerotic lesions by multidetector spiral computed tomography. J Am Coll Cardiol 43:842–847PubMedCrossRef
9.
Zurück zum Zitat Ferencik M, Nomura CH, Maurovich-Horvat P, Hoffmann U, Pena AJ, Cury RC, Abbara S, Nieman K, Fatima U, Achenbach S, Brady TJ (2006) Quantitative parameters of image quality in 64-slice computed tomography angiography of the coronary arteries. Eur J Radiol 57:373–379PubMedCrossRef Ferencik M, Nomura CH, Maurovich-Horvat P, Hoffmann U, Pena AJ, Cury RC, Abbara S, Nieman K, Fatima U, Achenbach S, Brady TJ (2006) Quantitative parameters of image quality in 64-slice computed tomography angiography of the coronary arteries. Eur J Radiol 57:373–379PubMedCrossRef
10.
Zurück zum Zitat Cury RC, Pomerantsev EV, Ferencik M, Hoffmann U, Nieman K, Moselewski F, Abbara S, Jang IK, Brady TJ, Achenbach S (2005) Comparison of the degree of coronary stenoses by multidetector computed tomography versus by quantitative coronary angiography. Am J Cardiol 96:784–787PubMedCrossRef Cury RC, Pomerantsev EV, Ferencik M, Hoffmann U, Nieman K, Moselewski F, Abbara S, Jang IK, Brady TJ, Achenbach S (2005) Comparison of the degree of coronary stenoses by multidetector computed tomography versus by quantitative coronary angiography. Am J Cardiol 96:784–787PubMedCrossRef
11.
Zurück zum Zitat Gerber TC, Breen JF, Kuzo RS, Kantor B, Williamson EE, Safford RE, Morin RL (2006) Computed tomographic angiography of the coronary arteries: techniques and applications. Semin Ultrasound CT MR 27:42–55PubMedCrossRef Gerber TC, Breen JF, Kuzo RS, Kantor B, Williamson EE, Safford RE, Morin RL (2006) Computed tomographic angiography of the coronary arteries: techniques and applications. Semin Ultrasound CT MR 27:42–55PubMedCrossRef
12.
Zurück zum Zitat Kyriakou Y, Kachelriess M, Knaup M, Krause JU, Kalender WA (2006) Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner. Eur Radiol 16:1206–1215PubMedCrossRef Kyriakou Y, Kachelriess M, Knaup M, Krause JU, Kalender WA (2006) Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner. Eur Radiol 16:1206–1215PubMedCrossRef
13.
Zurück zum Zitat Austen WG, Edwards JE, Frye RL, Gensini GG, Gott VL, Griffith LS, McGoon DC, Murphy ML, Roe BB (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMed Austen WG, Edwards JE, Frye RL, Gensini GG, Gott VL, Griffith LS, McGoon DC, Murphy ML, Roe BB (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMed
14.
Zurück zum Zitat Moselewski F, Ferencik M, Achenbach S, Abbara S, Cury RC, Booth SL, Jang IK, Brady TJ, Hoffmann U (2006) Threshold-dependent variability of coronary artery calcification measurements – implications for contrast-enhanced multi-detector row-computed tomography. Eur J Radiol 57:390–395PubMedCrossRef Moselewski F, Ferencik M, Achenbach S, Abbara S, Cury RC, Booth SL, Jang IK, Brady TJ, Hoffmann U (2006) Threshold-dependent variability of coronary artery calcification measurements – implications for contrast-enhanced multi-detector row-computed tomography. Eur J Radiol 57:390–395PubMedCrossRef
15.
Zurück zum Zitat Mollet NR, Cademartiri F, Nieman K, Saia F, Lemos PA, McFadden EP, Pattynama PM, Serruys PW, Krestin GP, de Feyter PJ (2004) Multislice spiral computed tomography coronary angiography in patients with stable angina pectoris. J Am Coll Cardiol 43:2265–2270PubMedCrossRef Mollet NR, Cademartiri F, Nieman K, Saia F, Lemos PA, McFadden EP, Pattynama PM, Serruys PW, Krestin GP, de Feyter PJ (2004) Multislice spiral computed tomography coronary angiography in patients with stable angina pectoris. J Am Coll Cardiol 43:2265–2270PubMedCrossRef
16.
Zurück zum Zitat Herrington DM, Siebes M, Sokol DK, Siu CO, Walford GD (1993) Variability in measures of coronary lumen dimensions using quantitative coronary angiography. J Am Coll Cardiol 22:1068–1074PubMedCrossRef Herrington DM, Siebes M, Sokol DK, Siu CO, Walford GD (1993) Variability in measures of coronary lumen dimensions using quantitative coronary angiography. J Am Coll Cardiol 22:1068–1074PubMedCrossRef
17.
Zurück zum Zitat Achenbach S, Moselewski F, Ropers D, Ferencik M, Hoffmann U, MacNeill B, Pohle K, Baum U, Anders K, Jang IK, Daniel WG, Brady TJ (2004) Detection of calcified and noncalcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography: a segment-based comparison with intravascular ultrasound. Circulation 109:14–17PubMedCrossRef Achenbach S, Moselewski F, Ropers D, Ferencik M, Hoffmann U, MacNeill B, Pohle K, Baum U, Anders K, Jang IK, Daniel WG, Brady TJ (2004) Detection of calcified and noncalcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography: a segment-based comparison with intravascular ultrasound. Circulation 109:14–17PubMedCrossRef
18.
Zurück zum Zitat Mahnken AH, Wildberger JE, Sinha AM, Dedden K, Stanzel S, Hoffmann R, Schmitz-Rode T, Gunther RW (2003) Value of 3D-volume rendering in the assessment of coronary arteries with retrospectively ECG-gated multislice spiral CT. Acta Radiol 44:302–309PubMedCrossRef Mahnken AH, Wildberger JE, Sinha AM, Dedden K, Stanzel S, Hoffmann R, Schmitz-Rode T, Gunther RW (2003) Value of 3D-volume rendering in the assessment of coronary arteries with retrospectively ECG-gated multislice spiral CT. Acta Radiol 44:302–309PubMedCrossRef
19.
Zurück zum Zitat Escaned J, Baptista J, Di Mario C, Haase J, Ozaki Y, Linker DT, de Feyter PJ, Roelandt JR, Serruys PW (1996) Significance of automated stenosis detection during quantitative angiography. Insights gained from intracoronary ultrasound imaging. Circulation 94:966–972PubMed Escaned J, Baptista J, Di Mario C, Haase J, Ozaki Y, Linker DT, de Feyter PJ, Roelandt JR, Serruys PW (1996) Significance of automated stenosis detection during quantitative angiography. Insights gained from intracoronary ultrasound imaging. Circulation 94:966–972PubMed
20.
Zurück zum Zitat Swallow RA, Court IA, Calver AL, Curzen NP (2006) The limitations of coronary angiography: identification of a critical coronary stenosis using intravascular ultrasound. Int J Cardiol 106:123–125PubMedCrossRef Swallow RA, Court IA, Calver AL, Curzen NP (2006) The limitations of coronary angiography: identification of a critical coronary stenosis using intravascular ultrasound. Int J Cardiol 106:123–125PubMedCrossRef
21.
Zurück zum Zitat Martuscelli E, Romagnoli A, D’Eliseo A, Razzini C, Tomassini M, Sperandio M, Simonetti G, Romeo F (2004) Accuracy of thin-slice computed tomography in the detection of coronary stenoses. Eur Heart J 25:1043–1048PubMedCrossRef Martuscelli E, Romagnoli A, D’Eliseo A, Razzini C, Tomassini M, Sperandio M, Simonetti G, Romeo F (2004) Accuracy of thin-slice computed tomography in the detection of coronary stenoses. Eur Heart J 25:1043–1048PubMedCrossRef
22.
Zurück zum Zitat Brown MS, Goldin JG, Rogers S, Kim HJ, Suh RD, McNitt-Gray MF, Shah SK, Truong D, Brown K, Sayre JW, Gjertson DW, Batra P, Aberle DR (2005) Computer-aided lung nodule detection in CT: results of large-scale observer test. Acad Radiol 12:681–686PubMedCrossRef Brown MS, Goldin JG, Rogers S, Kim HJ, Suh RD, McNitt-Gray MF, Shah SK, Truong D, Brown K, Sayre JW, Gjertson DW, Batra P, Aberle DR (2005) Computer-aided lung nodule detection in CT: results of large-scale observer test. Acad Radiol 12:681–686PubMedCrossRef
23.
Zurück zum Zitat Hermiller JB, Cusma JT, Spero LA, Fortin DF, Harding MB, Bashore TM (1992) Quantitative and qualitative coronary angiographic analysis: review of methods, utility, and limitations. Cathet Cardiovasc Diagn 25:110–131PubMedCrossRef Hermiller JB, Cusma JT, Spero LA, Fortin DF, Harding MB, Bashore TM (1992) Quantitative and qualitative coronary angiographic analysis: review of methods, utility, and limitations. Cathet Cardiovasc Diagn 25:110–131PubMedCrossRef
24.
Zurück zum Zitat Cademartiri F, Mollet NR, Runza G, Bruining N, Hamers R, Somers P, Knaapen M, Verheye S, Midiri M, Krestin GP, de Feyter PJ (2005) Influence of intracoronary attenuation on coronary plaque measurements using multislice computed tomography: observations in an ex vivo model of coronary computed tomography angiography. Eur Radiol 15:1426–1431PubMedCrossRef Cademartiri F, Mollet NR, Runza G, Bruining N, Hamers R, Somers P, Knaapen M, Verheye S, Midiri M, Krestin GP, de Feyter PJ (2005) Influence of intracoronary attenuation on coronary plaque measurements using multislice computed tomography: observations in an ex vivo model of coronary computed tomography angiography. Eur Radiol 15:1426–1431PubMedCrossRef
25.
Zurück zum Zitat Nikolaou K, Flohr T, Knez A (2004) Advances in cardiac CT imaging: 64-slice scanner. Int J Cardiovasc Imaging.20:535–540PubMedCrossRef Nikolaou K, Flohr T, Knez A (2004) Advances in cardiac CT imaging: 64-slice scanner. Int J Cardiovasc Imaging.20:535–540PubMedCrossRef
26.
Zurück zum Zitat Nikolaou K, Knez A, Rist C (2006) Accuracy of 64-MDCT in the diagnosis of ischemic heart disease. AJR Am J Roentgenol 187:111–117PubMedCrossRef Nikolaou K, Knez A, Rist C (2006) Accuracy of 64-MDCT in the diagnosis of ischemic heart disease. AJR Am J Roentgenol 187:111–117PubMedCrossRef
27.
Zurück zum Zitat Nikolaou K, Rist C, Wintersperger BJ (2006) Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. AJR Am J Roentgenol 186:1659–1668PubMedCrossRef Nikolaou K, Rist C, Wintersperger BJ (2006) Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. AJR Am J Roentgenol 186:1659–1668PubMedCrossRef
28.
Zurück zum Zitat Nikolaou K, Becker CR, Wintersperger BJ, Rist C (2004) Evaluating multislice computed tomography for imaging coronary atherosclerosis. Radiologe 44(2):130–139PubMedCrossRef Nikolaou K, Becker CR, Wintersperger BJ, Rist C (2004) Evaluating multislice computed tomography for imaging coronary atherosclerosis. Radiologe 44(2):130–139PubMedCrossRef
Metadaten
Titel
Visual and automatic grading of coronary artery stenoses with 64-slice CT angiography in reference to invasive angiography
verfasst von
Stephanie Busch
Thorsten R. C. Johnson
Konstantin Nikolaou
Franz von Ziegler
Andreas Knez
Maximilian F. Reiser
Christoph R. Becker
Publikationsdatum
01.06.2007
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 6/2007
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-006-0512-y

Weitere Artikel der Ausgabe 6/2007

European Radiology 6/2007 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.