01.10.2007 | Computer Tomography

Evaluation of dose exposure in 64-slice CT colonography

verfasst von: O. Luz, M. Buchgeister, M. Klabunde, T. Trabold, A. F. Kopp, C. D. Claussen, M. Heuschmid

Erschienen in: European Radiology | Ausgabe 10/2007

Einloggen, um Zugang zu erhalten

Abstract

The radiation exposure of four different 64-slice MDCT-colonography (CTC) protocols was evaluated using an Alderson-Rando phantom. Protocols using 30 mAs (collimation 20 × 1.2mm), 50 mAs (collimation 20 × 1.2 and 64 × 0.6mm) and 80 mAs (20 × 1.2 mm) representing screening low-dose, routine, narrow collimation and oncologic staging setups were measured with an Alderson-Rando phantom (Alderson Research Laboratories Inc.). Scans were performed on a 64-row MDCT (SOMATOM Sensation 64, Siemens) simulating the prone and supine positions with a constant voltage of 120 kV. Dose values (male/female) were 2.5/2.9, 3.8/4.2, 4.2/4.5 and 5.7/6.4 mSv for 30, 50 (20 × 1.2 and 64 × 0.6 mm) and 80 mAs, respectively. Measurements showed an elevated dose for females (11.5% mean; compared to males). Use of narrow collimation combined with 50 mAs resulted in a small increase of dose exposure of 10.5 (male) and 7.1% (female). Gonad doses ranged from 0.9 to 2.6 mSv (male) and from 1.5 to 3.5 mSv (female). In all protocols, the stomach wall, lower colon, urinary bladder and liver were slightly more highly exposed (all <2.3 mSv) than the other organs, and the breast dose was <0.3 mSv in every setup. Values of radiation exposure in 64- and 16-slice CTC differ only marginally when using the narrow collimation. In 64-slice CTC, the use of narrow (64 × 0.6 mm) collimation shows slightly elevated dose values compared to wider (20 × 1.2 mm) collimation.
Literatur
1.
Zurück zum Zitat Kopp AF, Ohnesorge B, Becker C, Schroder S, Heuschmid M, Kuttner A, Kuzo R, Claussen CD (2002) Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT. Radiology 225(1):113–119PubMedCrossRef Kopp AF, Ohnesorge B, Becker C, Schroder S, Heuschmid M, Kuttner A, Kuzo R, Claussen CD (2002) Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT. Radiology 225(1):113–119PubMedCrossRef
2.
Zurück zum Zitat Klingenbeck-Regn K, Schaller S, Flohr T, Ohnesorge B, Kopp AF, Baum U (1999) Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol 31(2):110–124PubMedCrossRef Klingenbeck-Regn K, Schaller S, Flohr T, Ohnesorge B, Kopp AF, Baum U (1999) Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol 31(2):110–124PubMedCrossRef
3.
Zurück zum Zitat Pickhardt PJ, Choi JR, Hwang I, Butler JA, Puckett ML, Hildebrandt HA, Wong RK, Nugent PA, Mysliwiec PA, Schindler WR (2003) Computed tomographic virtual colonoscopy to screen for colorectal neoplasia in asymptomatic adults. N Engl J Med 349:2191–2200PubMedCrossRef Pickhardt PJ, Choi JR, Hwang I, Butler JA, Puckett ML, Hildebrandt HA, Wong RK, Nugent PA, Mysliwiec PA, Schindler WR (2003) Computed tomographic virtual colonoscopy to screen for colorectal neoplasia in asymptomatic adults. N Engl J Med 349:2191–2200PubMedCrossRef
4.
Zurück zum Zitat Fletcher J, Johnson C, Welch T, MacCarty R, Ahlquist D, Reed J, Harmsen W, Wilson L (2000) Optimization of CT colonography technique: prospective trial in 180 patients. Radiology 216:704–711PubMed Fletcher J, Johnson C, Welch T, MacCarty R, Ahlquist D, Reed J, Harmsen W, Wilson L (2000) Optimization of CT colonography technique: prospective trial in 180 patients. Radiology 216:704–711PubMed
5.
Zurück zum Zitat Luboldt W, Höpffner N, Holzer K, Jacobi V, Willenbrock K, Wöste G, Bechstein W, Vogl T (2003) Early detection of colorectal tumors: CT or MRI? Radiologe 43:136–150PubMedCrossRef Luboldt W, Höpffner N, Holzer K, Jacobi V, Willenbrock K, Wöste G, Bechstein W, Vogl T (2003) Early detection of colorectal tumors: CT or MRI? Radiologe 43:136–150PubMedCrossRef
7.
Zurück zum Zitat Neri E, Giusti P, Battolla L, Vagli P, Boraschi P, Lencioni R, Caramella D, Bartolozzi C (2002) Colorectal cancer: role of CT colonography in preoperative evaluation after incomplete colonoscopy. Radiology 223:615–619PubMedCrossRef Neri E, Giusti P, Battolla L, Vagli P, Boraschi P, Lencioni R, Caramella D, Bartolozzi C (2002) Colorectal cancer: role of CT colonography in preoperative evaluation after incomplete colonoscopy. Radiology 223:615–619PubMedCrossRef
8.
Zurück zum Zitat Wessling J, Fischbach R, Meier N, Allkemper T, Klusmeier J, Ludwig K, Heindel W (2003) CT colonography: protocol optimization with multi-detector row CT study in an anthropomorphic colon phantom. Radiology 228:753–759PubMedCrossRef Wessling J, Fischbach R, Meier N, Allkemper T, Klusmeier J, Ludwig K, Heindel W (2003) CT colonography: protocol optimization with multi-detector row CT study in an anthropomorphic colon phantom. Radiology 228:753–759PubMedCrossRef
9.
Zurück zum Zitat Flohr T, Stierstorfer K, Raupach R, Ulzheimer S, Bruder H (2004) Performance evaluation of a 64-slice CT system with z-flying focal spot. Rofo 176(12):1803–1810PubMed Flohr T, Stierstorfer K, Raupach R, Ulzheimer S, Bruder H (2004) Performance evaluation of a 64-slice CT system with z-flying focal spot. Rofo 176(12):1803–1810PubMed
10.
Zurück zum Zitat Iannaccone R, Laghi A, Catalano C, Mangiapane F, Piacentini F, Passariello R (2003) Feasibility of ultra-low-dose multislice CT colonography for the detection of colorectal lesions: preliminary experience. Eur Radiol 13:1297–1302PubMed Iannaccone R, Laghi A, Catalano C, Mangiapane F, Piacentini F, Passariello R (2003) Feasibility of ultra-low-dose multislice CT colonography for the detection of colorectal lesions: preliminary experience. Eur Radiol 13:1297–1302PubMed
11.
Zurück zum Zitat Cohnen M, Vogt C, Aurich V, Beck A, Haussinger D, Modder U (2002) Multi-slice CT-colonography in low-dose technique-preliminary results. Rofo 174(7):835–838PubMed Cohnen M, Vogt C, Aurich V, Beck A, Haussinger D, Modder U (2002) Multi-slice CT-colonography in low-dose technique-preliminary results. Rofo 174(7):835–838PubMed
12.
Zurück zum Zitat Luz O, Schäfer J, Heuschmid M, Dammann F, Vonthein R, Traboldt T, Claussen CD (2004) Evaluation of different 16-row CT colonography protocols using a porcine model. Rofo 176(10):1493–1500PubMed Luz O, Schäfer J, Heuschmid M, Dammann F, Vonthein R, Traboldt T, Claussen CD (2004) Evaluation of different 16-row CT colonography protocols using a porcine model. Rofo 176(10):1493–1500PubMed
13.
Zurück zum Zitat Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposure using reference human phantoms and Monte Carlo methods. Part IV: organ doses from computed tomographic examinations. GSF-Bericht 30/91, GSF-Forschungszentrum für Umwelt und Gesundheit, Oberschleisßheim Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposure using reference human phantoms and Monte Carlo methods. Part IV: organ doses from computed tomographic examinations. GSF-Bericht 30/91, GSF-Forschungszentrum für Umwelt und Gesundheit, Oberschleisßheim
14.
Zurück zum Zitat International Commission on Radiological Protection (ICRP) (1991) Recommendation of the International Commission on Radiological Protection. ICRP Publication 60, Pergamon Press, Oxford International Commission on Radiological Protection (ICRP) (1991) Recommendation of the International Commission on Radiological Protection. ICRP Publication 60, Pergamon Press, Oxford
15.
Zurück zum Zitat Trabold T, Buchgeister M, Küttner A, Heuschmid M, Kopp A, Schröder S, Claussen C (2003) Estimation of radiation exposure in 16-row computed tomography of the heart with retrospective ECG-gating. Rofo 175:1051–1055PubMed Trabold T, Buchgeister M, Küttner A, Heuschmid M, Kopp A, Schröder S, Claussen C (2003) Estimation of radiation exposure in 16-row computed tomography of the heart with retrospective ECG-gating. Rofo 175:1051–1055PubMed
16.
Zurück zum Zitat Luz O, Schäfer J, Heuschmid M, Stüker D, Claussen C (2002) CT-colonography: current status and future directions in consideration of the technical evolution. Radiologe 42:712–721PubMedCrossRef Luz O, Schäfer J, Heuschmid M, Stüker D, Claussen C (2002) CT-colonography: current status and future directions in consideration of the technical evolution. Radiologe 42:712–721PubMedCrossRef
17.
Zurück zum Zitat Theocharopoulos N, Perisinakis K, Damilakis J, Karampekios S, Gourtsoyiannis N (2006) Dosimetric characteristics of a 16-slice computed tomography scanner. Eur Radiol 16(11):2575–2585PubMedCrossRef Theocharopoulos N, Perisinakis K, Damilakis J, Karampekios S, Gourtsoyiannis N (2006) Dosimetric characteristics of a 16-slice computed tomography scanner. Eur Radiol 16(11):2575–2585PubMedCrossRef
18.
Zurück zum Zitat Hohl C, Muhlenbruch G, Wildberger JE, Leidecker C, Suss C, Schmidt T, Gunther RW, Mahnken AH (2006) Estimation of radiation exposure in low dose multislice computed tomography of the heart and comparison with a calculation program. Eur Radiol 16(8):1841–1846PubMedCrossRef Hohl C, Muhlenbruch G, Wildberger JE, Leidecker C, Suss C, Schmidt T, Gunther RW, Mahnken AH (2006) Estimation of radiation exposure in low dose multislice computed tomography of the heart and comparison with a calculation program. Eur Radiol 16(8):1841–1846PubMedCrossRef
19.
Zurück zum Zitat Coppenrath E, Meindl T, Herzog P, Kalil R, Mueller-Lisse U, Krenn L, Reiser M, Mueller-Lisse UG (2006) Dose reduction in multidetector CT of the urinary tract. Studies in a phantom model. Eur Radiol 16(9):1982–1989PubMedCrossRef Coppenrath E, Meindl T, Herzog P, Kalil R, Mueller-Lisse U, Krenn L, Reiser M, Mueller-Lisse UG (2006) Dose reduction in multidetector CT of the urinary tract. Studies in a phantom model. Eur Radiol 16(9):1982–1989PubMedCrossRef
20.
Zurück zum Zitat Sosna J, Morrin MM, Copel L, Raptopoulos V, Kruskal JB (2003) Computed tomography colonography (virtual colonoscopy): update on technique, applications and future developments. Surg Technol Int 11:102–110PubMed Sosna J, Morrin MM, Copel L, Raptopoulos V, Kruskal JB (2003) Computed tomography colonography (virtual colonoscopy): update on technique, applications and future developments. Surg Technol Int 11:102–110PubMed
21.
Zurück zum Zitat Mulkens TH, Bellinck P, Baeyaert M, Ghysen D, Van Dijck X, Mussen E, Venstermans C, Termote JL (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237(1):213–223PubMedCrossRef Mulkens TH, Bellinck P, Baeyaert M, Ghysen D, Van Dijck X, Mussen E, Venstermans C, Termote JL (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237(1):213–223PubMedCrossRef
22.
Zurück zum Zitat Tschammler A, Kenn W, Dinkel HP, Sandstede J, Keberle M, Tschammler S (2005) Individualized dosis requirement for spiral CT of the abdomen. Rofo 177(8):1123–1130PubMed Tschammler A, Kenn W, Dinkel HP, Sandstede J, Keberle M, Tschammler S (2005) Individualized dosis requirement for spiral CT of the abdomen. Rofo 177(8):1123–1130PubMed
24.
Zurück zum Zitat Nishizawa K, Matsumoto M, Iwai K, Tonari A, Yoshida T, Takayama M (2002) Dose evaluation and effective dose estimation from multi detector CT. Igaku Butsuri 22(3):152–158PubMed Nishizawa K, Matsumoto M, Iwai K, Tonari A, Yoshida T, Takayama M (2002) Dose evaluation and effective dose estimation from multi detector CT. Igaku Butsuri 22(3):152–158PubMed
25.
Zurück zum Zitat Cohnen M, Poll LJ, Puettmann C, Ewen K, Saleh A, Modder U (2003) Effective doses in standard protocols for multi-slice CT scanning. Eur Radiol 13(5):1148–1153PubMed Cohnen M, Poll LJ, Puettmann C, Ewen K, Saleh A, Modder U (2003) Effective doses in standard protocols for multi-slice CT scanning. Eur Radiol 13(5):1148–1153PubMed
Metadaten
Titel
Evaluation of dose exposure in 64-slice CT colonography
verfasst von
O. Luz
M. Buchgeister
M. Klabunde
T. Trabold
A. F. Kopp
C. D. Claussen
M. Heuschmid
Publikationsdatum
01.10.2007
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 10/2007
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-007-0601-6

Weitere Artikel der Ausgabe 10/2007

European Radiology 10/2007 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.