Erschienen in: European Radiology 1/2014

01.01.2014 | Cardiac

High-pitch dual-source CT coronary angiography with low volumes of contrast medium

verfasst von: Alexander Lembcke, Carsten Schwenke, Patrick A. Hein, Gesine Knobloch, Tahir Durmus, Bernd Hamm, Alexander Huppertz

Erschienen in: European Radiology | Ausgabe 1/2014

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To assess the effect of lower volumes of contrast medium (CM) on image quality in high-pitch dual-source computed tomography coronary angiography (CTCA).

Methods

One-hundred consecutive patients (body weight 65–85 kg, stable heart rate ≤65 bpm, cardiac index ≥2.5 L/min/m2) referred for CTCA were prospectively enrolled. Patients were randomly assigned to one of five groups of different CM volumes (G30, 30 mL; G40, 40 mL; G50, 50 mL; G60, 60 mL; G70, 70 mL; flow rate 5 mL/s each, iodine content 370 mg/mL). Attenuation within the proximal and distal coronary artery segments was analysed.

Results

Mean attenuation for men and women ranged from 345.0 and 399.1 HU in G30 to 478.2 and 571.8 HU in G70. Mean attenuation values were higher in groups with higher CM volumes (P < 0.0001) and higher in women than in men (P < 0.0001). The proportions of segments with attenuation of at least 300 HU in G30, G40, G50, G60 and G70 were 89 %, 95 %, 98 %, 98 % and 99 %. CM volume of 30 mL in women and 40 mL in men proved to be sufficient to guarantee attenuation of at least 300 HU.

Conclusions

In selected patients high-pitch dual-source CTCA can be performed with CM volumes of 40 mL in men or 30 mL in women.

Key Points

High-pitch dual-source coronary angiography is feasible with low contrast media volumes.
Traditional injection rules still apply: higher volumes result in higher enhancement.
The patients gender is a co-factor determining the level of contrast enhancement.
Volumes can be reduced down to 3040 mL in selected patients.
Literatur
1.
Zurück zum Zitat Weininger M, Barraza JM, Kemper CA, Kalafut JF, Costello P, Schoepf UJ (2011) Cardiothoracic CT angiography: current contrast medium delivery strategies. AJR Am J Roentgenol 196:W260–W272PubMedCrossRef Weininger M, Barraza JM, Kemper CA, Kalafut JF, Costello P, Schoepf UJ (2011) Cardiothoracic CT angiography: current contrast medium delivery strategies. AJR Am J Roentgenol 196:W260–W272PubMedCrossRef
2.
Zurück zum Zitat Fleischmann D, Kamaya A (2009) Optimal vascular and parenchymal contrast enhancement: the current state of the art. Radiol Clin N Am 47:13–26PubMedCrossRef Fleischmann D, Kamaya A (2009) Optimal vascular and parenchymal contrast enhancement: the current state of the art. Radiol Clin N Am 47:13–26PubMedCrossRef
3.
Zurück zum Zitat Roberts WT, Bax JJ, Davies LC (2008) Cardiac CT and CT coronary angiography: technology and application. Heart 94:781–792PubMedCrossRef Roberts WT, Bax JJ, Davies LC (2008) Cardiac CT and CT coronary angiography: technology and application. Heart 94:781–792PubMedCrossRef
4.
Zurück zum Zitat Cademartiri F, Nieman K, van der Lugt A et al (2004) Intravenous contrast material administration at 16-detector row helical CT coronary angiography: test bolus versus bolus-tracking technique. Radiology 233:817–823PubMedCrossRef Cademartiri F, Nieman K, van der Lugt A et al (2004) Intravenous contrast material administration at 16-detector row helical CT coronary angiography: test bolus versus bolus-tracking technique. Radiology 233:817–823PubMedCrossRef
5.
Zurück zum Zitat Rist C, Nikolaou K, Kirchin MA et al (2006) Contrast bolus optimization for cardiac 16-slice computed tomography: comparison of contrast medium formulations containing 300 and 400 milligrams of iodine per milliliter. Invest Radiol 41:460–467PubMedCrossRef Rist C, Nikolaou K, Kirchin MA et al (2006) Contrast bolus optimization for cardiac 16-slice computed tomography: comparison of contrast medium formulations containing 300 and 400 milligrams of iodine per milliliter. Invest Radiol 41:460–467PubMedCrossRef
6.
Zurück zum Zitat Becker CR, Hong C, Knez A et al (2003) Optimal contrast application for cardiac 4-detector-row computed tomography. Invest Radiol 38:690–694PubMedCrossRef Becker CR, Hong C, Knez A et al (2003) Optimal contrast application for cardiac 4-detector-row computed tomography. Invest Radiol 38:690–694PubMedCrossRef
7.
Zurück zum Zitat Schoepf UJ, Zwerner PL, Savino G, Herzog C, Kerl JM, Costello P (2007) Coronary CT angiography. Radiology 244:48–63PubMedCrossRef Schoepf UJ, Zwerner PL, Savino G, Herzog C, Kerl JM, Costello P (2007) Coronary CT angiography. Radiology 244:48–63PubMedCrossRef
8.
Zurück zum Zitat Hein PA, May J, Rogalla P, Butler C, Hamm B, Lembcke A (2010) Feasibility of contrast material volume reduction in coronary artery imaging using 320-slice volume CT. Eur Radiol 20:1337–1343PubMedCrossRef Hein PA, May J, Rogalla P, Butler C, Hamm B, Lembcke A (2010) Feasibility of contrast material volume reduction in coronary artery imaging using 320-slice volume CT. Eur Radiol 20:1337–1343PubMedCrossRef
9.
Zurück zum Zitat Steigner ML, Mitsouras D, Whitmore AG et al (2010) Iodinated contrast opacification gradients in normal coronary arteries imaged with prospectively ECG-gated single heart beat 320-detector row computed tomography. Circ Cardiovasc Imaging 3:179–186PubMedCentralPubMedCrossRef Steigner ML, Mitsouras D, Whitmore AG et al (2010) Iodinated contrast opacification gradients in normal coronary arteries imaged with prospectively ECG-gated single heart beat 320-detector row computed tomography. Circ Cardiovasc Imaging 3:179–186PubMedCentralPubMedCrossRef
10.
Zurück zum Zitat Schwenke C, Busse R (2007) Analysis of differences in proportions from clustered data with multiple measurements in diagnostic studies. Methods Inf Med 46:548–552PubMed Schwenke C, Busse R (2007) Analysis of differences in proportions from clustered data with multiple measurements in diagnostic studies. Methods Inf Med 46:548–552PubMed
11.
Zurück zum Zitat Yamamuro M, Tadamura E, Kanao S et al (2007) Coronary angiography by 64-detector row computed tomography using low dose of contrast material with saline chaser: influence of total injection volume on vessel attenuation. J Comput Assist Tomography 31:272–280CrossRef Yamamuro M, Tadamura E, Kanao S et al (2007) Coronary angiography by 64-detector row computed tomography using low dose of contrast material with saline chaser: influence of total injection volume on vessel attenuation. J Comput Assist Tomography 31:272–280CrossRef
12.
Zurück zum Zitat Cademartiri F, Mollet NR, van der Lugt A et al (2005) Intravenous contrast material administration at helical 16-detector row CT coronary angiography: effect of iodine concentration on vascular attenuation. Radiology 236:661–665PubMedCrossRef Cademartiri F, Mollet NR, van der Lugt A et al (2005) Intravenous contrast material administration at helical 16-detector row CT coronary angiography: effect of iodine concentration on vascular attenuation. Radiology 236:661–665PubMedCrossRef
13.
Zurück zum Zitat Bae KT, Seeck BA, Hildeboldt CF et al (2008) Contrast enhancement in cardiovascular MDCT: effect of body weight, height, body surface area, body mass index, and obesity. AJR Am J Roentgenol 190:777–784PubMedCrossRef Bae KT, Seeck BA, Hildeboldt CF et al (2008) Contrast enhancement in cardiovascular MDCT: effect of body weight, height, body surface area, body mass index, and obesity. AJR Am J Roentgenol 190:777–784PubMedCrossRef
14.
Zurück zum Zitat From AM, Bartholmai BJ, Williams AW et al (2008) Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc 83:1095–1100PubMedCrossRef From AM, Bartholmai BJ, Williams AW et al (2008) Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc 83:1095–1100PubMedCrossRef
15.
Zurück zum Zitat Toprak O (2007) Conflicting and new risk factors for contrast induced nephropathy. J Urol 178:2277–2283PubMedCrossRef Toprak O (2007) Conflicting and new risk factors for contrast induced nephropathy. J Urol 178:2277–2283PubMedCrossRef
16.
Zurück zum Zitat Thomsen HS (2008) Current evidence on prevention and management of contrast-induced nephropathy. Eur Radiol 17(Suppl 6):33–37 Thomsen HS (2008) Current evidence on prevention and management of contrast-induced nephropathy. Eur Radiol 17(Suppl 6):33–37
17.
Zurück zum Zitat Thomsen HS, Morcos SK, Barrett BJ (2008) Contrast-induced nephropathy: the wheel has turned 360 degrees. Acta Radiol 49:646–657PubMedCrossRef Thomsen HS, Morcos SK, Barrett BJ (2008) Contrast-induced nephropathy: the wheel has turned 360 degrees. Acta Radiol 49:646–657PubMedCrossRef
18.
19.
Zurück zum Zitat Rhee CM, Bhan I, Alexander EK, Brunelli SM (2012) Association between iodinated contrast media exposure and incident hyperthyroidism and hypothyroidism. Arch Intern Med 172:153–159PubMedCrossRef Rhee CM, Bhan I, Alexander EK, Brunelli SM (2012) Association between iodinated contrast media exposure and incident hyperthyroidism and hypothyroidism. Arch Intern Med 172:153–159PubMedCrossRef
20.
Zurück zum Zitat Dewey M, Hoffmann B, Hamm (2007) CT coronary angiography using 16 and 64 simultaneous detector rows: intraindividual comparison. Röfo 179:581–586PubMed Dewey M, Hoffmann B, Hamm (2007) CT coronary angiography using 16 and 64 simultaneous detector rows: intraindividual comparison. Röfo 179:581–586PubMed
21.
Zurück zum Zitat Burgstahler C, Reimann C, Brodoefel H et al (2009) Quantitative parameters to compare image quality of non-invasive coronary angiography with 16-slice, 64-slice and dual-source computed tomography. Eur Radiol 19:584–590PubMedCrossRef Burgstahler C, Reimann C, Brodoefel H et al (2009) Quantitative parameters to compare image quality of non-invasive coronary angiography with 16-slice, 64-slice and dual-source computed tomography. Eur Radiol 19:584–590PubMedCrossRef
22.
Zurück zum Zitat Seifarth H, Puesken M, Kalafut JF et al (2009) Introduction of an individually optimized protocol for the injection of contrast medium for coronary CT angiography. Eur Radiol 19:2373–2382PubMedCrossRef Seifarth H, Puesken M, Kalafut JF et al (2009) Introduction of an individually optimized protocol for the injection of contrast medium for coronary CT angiography. Eur Radiol 19:2373–2382PubMedCrossRef
23.
Zurück zum Zitat Nakaura T, Awai K, Yanaga Y et al (2011) Low-dose contrast protocols using the test bolus technique for 64-detector computed tomography coronary angiography. Jpn J Radiol 29:457–465PubMedCrossRef Nakaura T, Awai K, Yanaga Y et al (2011) Low-dose contrast protocols using the test bolus technique for 64-detector computed tomography coronary angiography. Jpn J Radiol 29:457–465PubMedCrossRef
24.
Zurück zum Zitat Tatsugami F, Kanamoto T, Nakai G et al (2010) Reduction of the total injection volume of contrast material with a short injection duration in 64-detector row CT coronary angiography. Br J Radiol 83:35–39PubMedCentralPubMedCrossRef Tatsugami F, Kanamoto T, Nakai G et al (2010) Reduction of the total injection volume of contrast material with a short injection duration in 64-detector row CT coronary angiography. Br J Radiol 83:35–39PubMedCentralPubMedCrossRef
25.
Zurück zum Zitat Tatsugami F, Matsuki M, Inada Y et al (2010) Feasibility of low-volume injections of contrast material with a body weight-adapted iodine-dose protocol in 320-detector row coronary CT angiography. Acad Radiol 17:207–211PubMedCrossRef Tatsugami F, Matsuki M, Inada Y et al (2010) Feasibility of low-volume injections of contrast material with a body weight-adapted iodine-dose protocol in 320-detector row coronary CT angiography. Acad Radiol 17:207–211PubMedCrossRef
26.
Zurück zum Zitat Yang WJ, Chen KM, Liu B et al (2012) Contrast media volume optimization in high-pitch dual-source CT coronary angiography: feasibility study. Int J Cardiovasc Imaging 29:245–252PubMedCrossRef Yang WJ, Chen KM, Liu B et al (2012) Contrast media volume optimization in high-pitch dual-source CT coronary angiography: feasibility study. Int J Cardiovasc Imaging 29:245–252PubMedCrossRef
27.
Zurück zum Zitat Kumamaru KK, Steigner ML, Soga S et al (2011) Coronary enhancement for prospective ECG-gated single R-R axial 320-MDCT angiography: comparison of 60- and 80-mL iopamidol 370 injection. AJR Am J Roentgenol 197:844–850PubMedCrossRef Kumamaru KK, Steigner ML, Soga S et al (2011) Coronary enhancement for prospective ECG-gated single R-R axial 320-MDCT angiography: comparison of 60- and 80-mL iopamidol 370 injection. AJR Am J Roentgenol 197:844–850PubMedCrossRef
28.
Zurück zum Zitat Zhu X, Zhu Y, Xu H, Yang G, Tang L, Xu Y (2013) Dual-source CT coronary angiography involving injection protocol with iodine load tailored to patient body weight and body mass index: estimation of optimal contrast material dose. Acta Radiol 54:149–155PubMedCrossRef Zhu X, Zhu Y, Xu H, Yang G, Tang L, Xu Y (2013) Dual-source CT coronary angiography involving injection protocol with iodine load tailored to patient body weight and body mass index: estimation of optimal contrast material dose. Acta Radiol 54:149–155PubMedCrossRef
29.
Zurück zum Zitat Kang EJ, Lee J, Lee KN, Kown H, Ha DH, Kim RB (2013) An initial randomised study assessing free-breathing CCTA using 320-detector CT. Eur Radiol 23:1199–1209PubMedCrossRef Kang EJ, Lee J, Lee KN, Kown H, Ha DH, Kim RB (2013) An initial randomised study assessing free-breathing CCTA using 320-detector CT. Eur Radiol 23:1199–1209PubMedCrossRef
Metadaten
Titel
High-pitch dual-source CT coronary angiography with low volumes of contrast medium
verfasst von
Alexander Lembcke
Carsten Schwenke
Patrick A. Hein
Gesine Knobloch
Tahir Durmus
Bernd Hamm
Alexander Huppertz
Publikationsdatum
01.01.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 1/2014
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-013-2988-6

Weitere Artikel der Ausgabe 1/2014

European Radiology 1/2014 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.