01.07.2015 | Physics

Effective dose and organ doses estimation taking tube current modulation into account with a commercial software package

verfasst von: X. Lopez-Rendon, H. Bosmans, R. Oyen, F. Zanca

Erschienen in: European Radiology | Ausgabe 7/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the effect of including tube current modulation (TCM) versus using the average mAs in estimating organ and effective dose (E) using commercial software.

Method

Forty adult patients (24 females, 16 males) with normal BMI underwent chest/abdomen computed tomography (CT) performed with TCM at 120 kVp, reference mAs of 110 (chest) and 200 (abdomen). Doses to fully irradiated organs (breasts, lungs, stomach, liver and ovaries) and E were calculated using two versions of a dosimetry software: v.2.0, which uses the average mAs, and v.2.2, which accounts for TCM by implementing a gender-specific mAs profile. Student’s t-test was used to assess statistically significant differences between organ doses calculated with the two versions.

Results

A statistically significant difference (p < 0.001) was found for E on chest and abdomen CT, with E being lower by 4.2 % when TCM is considered. Similarly, organ doses were also significantly lower (p < 0.001): 13.7 % for breasts, 7.3 % for lungs, 9.1 % for the liver and 8.5 % for the stomach. Only the dose to the ovaries was higher with TCM (11.5 %).

Conclusion

When TCM is used, for the stylized phantom, the doses to lungs, breasts, stomach and liver decreased while the dose to the ovaries increased.

Key Points

Estimated dose to the ovaries increased with TCM.
Estimated dose to lungs, breasts, stomach and liver decreased with TCM.
A unique but gender-specific mAs profile resulted in a radiation dose shift.
Even for normal size patients there is a variety in mAs profiles.
Literatur
1.
Zurück zum Zitat Brenner DJ (2010) Should we be concerned about the rapid increase in CT usage? Rev Environ Health 25:63–67PubMedCrossRef Brenner DJ (2010) Should we be concerned about the rapid increase in CT usage? Rev Environ Health 25:63–67PubMedCrossRef
2.
Zurück zum Zitat Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 6736:1–7 Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 6736:1–7
3.
Zurück zum Zitat Valentin J (2007) The 2007 Recommendations of the International Commission on Radiological Protection. Annals of the ICRP Valentin J (2007) The 2007 Recommendations of the International Commission on Radiological Protection. Annals of the ICRP
5.
Zurück zum Zitat Zanca F, Demeter M, Oyen R, Bosmans H (2012) Excess radiation and organ dose in chest and abdominal CT due to CT acquisition beyond expected anatomical boundaries. Eur Radiol 22:779–788PubMedCrossRef Zanca F, Demeter M, Oyen R, Bosmans H (2012) Excess radiation and organ dose in chest and abdominal CT due to CT acquisition beyond expected anatomical boundaries. Eur Radiol 22:779–788PubMedCrossRef
6.
Zurück zum Zitat Munk RD, Strohm PC, Saueressig U et al (2009) Effective dose estimation in whole-body multislice CT in paediatric trauma patients. Pediatr Radiol 39:245–252PubMedCrossRef Munk RD, Strohm PC, Saueressig U et al (2009) Effective dose estimation in whole-body multislice CT in paediatric trauma patients. Pediatr Radiol 39:245–252PubMedCrossRef
7.
Zurück zum Zitat Jahnen A, Kohler S, Hermen J et al (2011) Automatic computed tomography patient dose calculation using DICOM header metadata. Radiat Prot Dosim 147:317–320CrossRef Jahnen A, Kohler S, Hermen J et al (2011) Automatic computed tomography patient dose calculation using DICOM header metadata. Radiat Prot Dosim 147:317–320CrossRef
8.
Zurück zum Zitat Van der Molen AJ, Joemai RMS, Geleijns J (2012) Performance of longitudinal and volumetric tube current modulation in a 64-slice CT with different choices of acquisition and reconstruction parameters. Phys Med 28:319–326PubMedCrossRef Van der Molen AJ, Joemai RMS, Geleijns J (2012) Performance of longitudinal and volumetric tube current modulation in a 64-slice CT with different choices of acquisition and reconstruction parameters. Phys Med 28:319–326PubMedCrossRef
9.
Zurück zum Zitat Vollmar SV, Kalender WA (2008) Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 18:1674–1682PubMedCrossRef Vollmar SV, Kalender WA (2008) Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 18:1674–1682PubMedCrossRef
10.
Zurück zum Zitat Angel E, Yaghmai N, Jude CM et al (2009) Dose to Radiosensitive Organs During Routine Chest CT: effects of tube current modulation. AJR Am J Roentgenol 193:1340–1345PubMedCentralPubMedCrossRef Angel E, Yaghmai N, Jude CM et al (2009) Dose to Radiosensitive Organs During Routine Chest CT: effects of tube current modulation. AJR Am J Roentgenol 193:1340–1345PubMedCentralPubMedCrossRef
11.
Zurück zum Zitat Kalra M, Maher M, Toth T et al (2004) Comparion of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology 232:347–353PubMedCrossRef Kalra M, Maher M, Toth T et al (2004) Comparion of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology 232:347–353PubMedCrossRef
12.
Zurück zum Zitat Rizzo S, Kalra M, Schmidt B et al (2005) Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelivs. AJR Am J Roentgenol 186:673–679CrossRef Rizzo S, Kalra M, Schmidt B et al (2005) Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelivs. AJR Am J Roentgenol 186:673–679CrossRef
13.
Zurück zum Zitat Angel E, Yaghmai N, Jude CM et al (2009) Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT. Phys Med Biol 54:497–512PubMedCentralPubMedCrossRef Angel E, Yaghmai N, Jude CM et al (2009) Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT. Phys Med Biol 54:497–512PubMedCentralPubMedCrossRef
14.
Zurück zum Zitat Stamm G, Nagel HD (2011) CT-Expo V 2.0 A tool for dose evaluation in Computed Tomography Stamm G, Nagel HD (2011) CT-Expo V 2.0 A tool for dose evaluation in Computed Tomography
15.
Zurück zum Zitat Stamm G, Nagel HD (2013) CT-Expo V 2.2 A tool for dose evaluation in computed tomography. User’s guide Stamm G, Nagel HD (2013) CT-Expo V 2.2 A tool for dose evaluation in computed tomography. User’s guide
16.
Zurück zum Zitat Kramer R, Zankl M, Williams G, Drexler G (1982) GSF Report S-88 Part I Kramer R, Zankl M, Williams G, Drexler G (1982) GSF Report S-88 Part I
17.
Zurück zum Zitat Larke FJ, Kruger RL, Cagnon CH et al (2011) Estimated radiation dose associated with low-dose chest CT of average-size participants in the National Lung Screening Trial. AJR Am J Roentgenol 197:1165–1169PubMedCrossRef Larke FJ, Kruger RL, Cagnon CH et al (2011) Estimated radiation dose associated with low-dose chest CT of average-size participants in the National Lung Screening Trial. AJR Am J Roentgenol 197:1165–1169PubMedCrossRef
18.
Zurück zum Zitat Gunn MLD, Kohr JR (2010) State of the art: technologies for computed tomography dose reduction. Emerg Radiol 17:209–218PubMedCrossRef Gunn MLD, Kohr JR (2010) State of the art: technologies for computed tomography dose reduction. Emerg Radiol 17:209–218PubMedCrossRef
19.
Zurück zum Zitat Söderberg M, Gunnarsson M (2010) Automatic exposure control in computed tomography–an evaluation of systems from different manufacturers. Acta Radiol 51:625–634PubMedCrossRef Söderberg M, Gunnarsson M (2010) Automatic exposure control in computed tomography–an evaluation of systems from different manufacturers. Acta Radiol 51:625–634PubMedCrossRef
20.
Zurück zum Zitat Tsalafoutas IA, Varsamidis A, Thalassinou S, Efstathopoulos EP (2013) Utilizing a simple CT dosimetry phantom for the comprehension of the operational characteristics of CT AEC systems. Med Phys 40:111918PubMedCrossRef Tsalafoutas IA, Varsamidis A, Thalassinou S, Efstathopoulos EP (2013) Utilizing a simple CT dosimetry phantom for the comprehension of the operational characteristics of CT AEC systems. Med Phys 40:111918PubMedCrossRef
Metadaten
Titel
Effective dose and organ doses estimation taking tube current modulation into account with a commercial software package
verfasst von
X. Lopez-Rendon
H. Bosmans
R. Oyen
F. Zanca
Publikationsdatum
01.07.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 7/2015
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-015-3623-5

Weitere Artikel der Ausgabe 7/2015

European Radiology 7/2015 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.