Open Access 01.02.2016 | Leitthema

Ultrasound diagnostics of renal artery stenosis

Stenosis criteria, CEUS and recurrent in-stent stenosis

verfasst von: Dr. W. Schäberle, L. Leyerer, W. Schierling, K. Pfister

Erschienen in: Gefässchirurgie | Sonderheft 1/2016

Abstract

Background and purpose

As a non-invasive, side effect-free and cost-effective method, ultrasonography represents the method of choice for the diagnosis of renal artery stenosis. Four different criteria in total, including two direct criteria in peak systolic velocity (PSV) and renal aortic ratio (RAR) and two indirect criteria in resistance index (RI) and acceleration time (AT) for the measurement of relevant renal artery stenosis are described, each demonstrating highly variable accuracy in studies. Furthermore, there is controversy over the degree beyond which stenosis becomes therapeutically relevant and which ultrasound PSV is diagnostically relevant in terms of stenosis grading.

Material and methods

This article gives a critical review based on a selective literature search on measurement methodology and the validity of ultrasound in renal artery stenosis. A critical evaluation of methods and a presentation of measurement principles to establish the most precise measurement method possible compared with the gold standard angiography, as well as an evaluation of the importance of computed tomography angiography (CTA) and magnetic resonance angiography (MRA).

Results and conclusions

The PSV provides high sensitivity and specificity as a direct measurement method in stenosis detection and grading. Most studies found sensitivities and specificities of 85–90 % for > 50 % stenosis at a PSV > 180–200 cm/s in ROC curve analysis. Other methods, such as the ratio of the PSV in the aorta to the PSV in the renal artery (RAR) or indirect criteria, such as side to side differences in RI (dRI) or AT can be additionally used to improve accuracy. Contrast-enhanced ultrasound improves accuracy by means of echo contrast enhancement. Although in the past only high-grade stenosis was considered relevant for treatment, a drop in pressure of > 20 mmHg in > 50 % stenosis (PSV 180 cm/s) is classified as relevant for increased renin secretion. Stenosis in fibromuscular dysplasia can be reliably graded according to the continuity equation. Although the available studies on the grading of in-stent restenosis are the subject of controversy, there is a tendency to assume higher cut-off values for PSV and RAR. Whilst MRA and CTA demonstrate an accuracy of > 90 %, this is at the cost of possible side effects for patients, particularly in the case of pre-existing renal parenchymal damage.
Begleitmaterial
Hinweise

Additional online material

This article includes two additional video sequences on visualizing renal artery stenosis. This supplemental material can be found under: dx.doi.org/10.​1007/​s00772-015-0060-3
The German version of this article was published in Gefässchirurgie (2015) 20:102–111

Introduction

Renal artery stenosis (RAS) is found to be the cause of arterial hypertension in 1–5 % of patients [31] and is largely responsible for renal failure requiring dialysis in 5–15 % of patients [10, 13]. In addition to treatment for hypertension for which, however, there is no significant benefit compared with drug therapy, stent-assisted percutaneous transluminal angioplasty (PTA) is relevant in terms of organ and function preservation in high-grade RAS [18, 36].
Intra-arterial renal artery angiography is established as the gold standard for the diagnosis of RAS. A number of studies have evaluated the value of color-coded duplex ultrasonography (CCDS) for screening purposes [25, 50]. Magnetic resonance angiography (MRA) and computed tomography angiography (CTA) have also become established alongside CCDS. The latter is non-invasive, comparatively cost-effective, widely deployable and permits stenosis grading using hemodynamic measurement parameters; however, study results on stenosis grading are to some extent conflicting.

Measurement methods

A total of four different methodological approaches to diagnosing RAS using CCDS have been evaluated over the last 25 years, 2 of which measure the degree of stenosis according to direct and 2 according to indirect criteria.
Direct criteria:
1.
Peak systolic velocity (PSV) determines the degree of stenosis according to the continuity equation (PSV is inversely proportional to the cross-sectional area affected by stenosis and luminal reduction).
 
2.
The ratio between PSV in the stenosed renal artery and PSV in the aorta (RAR renal aortic ratio) compares the increased intrastenotic flow velocity in the renal arteries with an individual reference value in the aorta. This approach attempts to reduce systemic influencing factors on PSV, such as current blood pressure; however, other factors having a hemodynamic effect on the aorta are difficult to evaluate.
 
Indirect criteria:
1.
Poststenotic Doppler frequency spectra obtained from the renal hilum are evaluated. A reduction in the resistance index (RI) > 0.05 is an indication of ipsilateral RAS.
 
2.
Delayed acceleration time (AT) distal to high-grade RAS, i.e. delay in systolic rise from end diastole up to PSV on spectral analysis.
 
As obtaining PSV in the proximal renal arteries can be challenging, some researchers prefer to do this at the renal hilum (indirect criteria). As is known from other vascular territories, indirect criteria only offer reliable accuracy in the case of high-grade stenosis. Although they demonstrate sufficient specificity in 50–70 % of cases depending on the method, sensitivity is poor at approximately ≤ 70 %. There is no consensus on the best method to detect RAS using CCDS as each method has its advantages and disadvantages; however, the significant variation in cut-off values above which both direct and indirect criteria assume a > 50 % or 60 % RAS is remarkable. Thus, peak flow velocities of 100–220 cm/s are given as the cut-off for 50 % stenosis using the most commonly used parameter, PSV. This achieves at times comparable, at times differing accuracies, a phenomenon that cannot be explained by study design alone.

Examination procedure

When diagnosing stenosis using direct criteria (e.g. PSV and RAR), the examination takes place with the subject in a supine position. By dosing pressure with a transducer (2–5 MHz) it is possible to suppress artifacts from bowel gas and reduce the required penetration depth. The aorta is sought in cross-section from an epigastric approach and followed from the cranial to the peripheral aspect, where the renal arteries are located 1–3 cm distal to the mesenteric artery branch, which can be well localized in cross-section. Two structures that aid renal artery localization are the left renal vein, which courses between the aorta and the superior mesenteric artery to the vena cava. The right renal artery generally courses initially to the right in a ventrolateral direction (approximately 10–11 o’clock position) from the aorta, turns in an arc in a dorsal direction and then continues to run dorsal to the vena cava up to the right renal hilum (Fig. 1a, b). The left renal artery generally follows a lateral to dorsolateral course (approximately 3–5 o’clock position) from the aorta and then runs in its length of only 4–5 cm to the renal hilum.
Atherosclerotic stenosis is generally located at the origin of the renal artery. This can be the focus of the examination. Stenosis caused by fibromuscular dysplasia can be found in the middle third. To diagnose stenosis according to indirect criteria (e.g. AT and side to side differences in RI), the renal arteries are scanned on both sides in a flanking section and the renal arteries are visualized in the hilum. If patients are not overly adipose, the renal artery can also be probed in this way (transhepatic section) up to where it arises from the aorta in a right-sided paramedian section (banana peel view) (Fig. 1c). In order to obtain reliable values despite the relatively error prone measurement method when indirect criteria are used, the mean of multiple measurements (between three and five in total) needs to be taken.

Significance of color-coded duplex sonography in the detection of renal artery stenosis

Direct criteria

The significance of CCDS needs to be assessed in a differentiated manner relative to the criteria already described. Compared with the gold standard angiography, PSV demonstrates sensitivities of 71–98 % and specificities of 62–98 %, whereby some studies consider > 50 % and others > 60 % as hemodynamically relevant stenosis. The PSV cut-off ranges from 100 to 220 cm/s (Fig. 2). It should be noted here that older studies [2, 11, 16, 38] tended to set lower peak velocities (below 150 cm/s). These studies were not carried out using CCDS but using a combination of B-mode and Doppler sonography. The lack of color-coded vascular course (flow jet) has shown, also in our own experience, the risk of angle misalignment and hence measurement errors in the PSV approach (Fig. 3).
Studies conducted using CCDS (mostly after 1993) generally set the cut-off at 180–200 cm/s [8, 19, 24, 27, 40, 42]. As peak velocities are determined using receiver operating characteristics (ROC) curves, they also always represent an evaluation performed by the author to assess sensitivity or specificity, positive predictive value (PPV) or negative predictive value (NPV). Thus, a recent paper [1] reported a sensitivity, specificity, PPV, NPV and overall accuracy (OA) of 89 %, 54 %, 56 %, 88 % and 68 %, respectively, for a PSV of 200 cm/s. A PSV of 285 cm/s is set as the ideal cutoff for 60 % stenosis. Staub [42] described a PSV of 180 cm/s for 50 % stenosis with a sensitivity, specificity, PPV, NPV and OA of 96 %, 69 %, 81 %, 93 % and 85 %, respectively. A PSV of 200 cm/s yielded 92 %, 81 %, 87 %, 88 % and 87 %, respectively, while a PSV of 250 cm/s 78 %, 92 %, 93 %, 75 % and 84 %, respectively. This results in an ideal cut-off of 200 cm/s. Selecting higher PSV as cut-off values inevitably resulted in lower sensitivity and greater specificity in ROC curves compared with angiography and the converse in the case of a lower PSV. Other causes for differing PSV as peak velocities include the examination method, angle measurement errors (in particular due to the tortuous course of the proximal right renal artery) and the collective investigated (e.g. greater vessel wall rigidity and chronic renal parenchymal damage). Virtually none of the studies addressed systemic factors that influence PSV, such as current blood pressure and vessel wall rigidity. The PSV cut-off values are also strongly influenced by the little discussed problem of adequate stenosis grading in the reference method, angiography. The renal artery is difficult to visualize in the two planes necessary for appropriate grading (only oblique planes possible). Renal artery branch stenosis in particular can be challenging to visualize. Ultrasound PSV grading is generally compared with angiography solely in an anteroposterior plane. Although angiographic diagnosis offers good accuracy, there is poor concordance in stenosis grading between radiologists [49]. For this reason, an own study [38] used X-ray densitometry as an additional reference method and achieved a sensitivity of 86 % and a specificity of 83 % at a PSV cut-off of 140 cm/s. In addition, a good correlation (R = 0.84) between PSV and X-ray densitometry in stenosis grading of stenosed renal arteries before and after PTA was seen.
Particularly in the case of eccentric stenosis, significant discrepancies are observed between duplex ultrasonography and angiography, the latter showing a far lower hemodynamic effectiveness at the same angiographic diameter reduction compared with concentric stenosis (50 % diameter reduction in concentric stenosis = 75 % area reduction and 50 % diameter reduction in eccentric stenosis = 50 % area reduction). On duplex ultrasonography, the hemodynamic effectiveness of a stenosis is measured as an expression of area reduction. Thus, the PSV in concentric stenosis can be up to twice as high at the same diameter reduction (angiographic data) compared with eccentric stenosis [37].
An RAR of > 3.5 indicates an over 60 % RAS with a sensitivity of 84–91 % and a specificity of 95–97 % [16, 17, 22, 47]. Recent studies were unable to confirm this accuracy, working instead with sensitivities of 73–84 %, specificities of 72–81 % and accuracies of 76–78 % [1, 42]. End-diastolic peak velocity is sometimes also given as a stenosis criterion; however, as a parameter it depends heavily on heart rate and peripheral resistance and the findings, particularly in patients with early renal parenchymal damage cannot be used with sufficient accuracy as increased peripheral resistance causes reduced end-diastolic volume (EDV) early on.
There are currently no studies to validate CCDS specifically in stenosis caused by fibromuscular dysplasia. The main problem lies in the difficulty associated with visualizing the middle third on the left side as a result of gas artifacts from the colon. Comparing spectra and RI at the renal artery branch and at the hilum can be helpful here (see Fig. 5b). If visualization of the middle third is possible, stenosis grading can be reliably performed via the ratios (PSV ratio) between intrastenotic PSV and PSV in the first third (prestenotic) using the continuity equation (Fig. 4 and video clip 1). A PSV ratio of > 2 indicates a stenosis of > 50 % and > 4 of > 75 % (for concentric stenosis), offering greater grading reliability, as known from peripheral artery stenosis, compared with absolute PSV values.

Indirect criteria

Spectral analysis of vessels in other arterial territories reveals that indirect criteria only show measurement relevant lesions in the presence of high-grade stenosis. Thus, it is not surprising that a dRI of > 0.05 (Figs. 5 and 6) as the cut-off for 50 % stenosis has a sensitivity of 42 % and a specificity of 91 % (PPV 69 % and NPV 77 %). This poor sensitivity, even for > 70 % stenosis, has also been confirmed by Zeller [53] with a sensitivity of 77 % but a specificity of 99 %, as well as by Ripollés [33] with a sensitivity of only 50 % and a specificity of 90 % (PPV 69 % and NPV 92 %). Furthermore, Ripollés [33] made the interesting finding that a dRI > 0.05 could be used only in patients aged < 50 years with a sensitivity of 90 % and a specificity of 99 %. Poststenotic Doppler frequency spectra depend heavily on vessel rigidity and parenchyma function. The typical poststenotic changes (significantly reduced PSV in relation to EDV and prolonged AT) are not as marked in older patients with arteriosclerosis and renal parenchymal damage. On the other hand, a different resistance index (dRI) due to different degrees of renal parenchymal damage on the two sides results in measurement errors. The AT also confirms this with poor sensitivities (around 50 %) and good specificities (around 95 %) [8, 27] in < 80 % RAS.

Treatment-oriented stenosis grading

Of the different stenosis criteria used a PSV > 180− 200 cm/s shows the best accuracy for the diagnosis and grading of RAS. To improve the inadequate sensitivities and specificities still seen in some studies, a number of authors recommend combining measurement results from a variety of (direct and indirect) stenosis criteria. In the study conducted by Staub et al. combining PSV and RAR yielded sensitivities and specificities of approximately 90 % [8, 24, 42] using a PSV of > 180 cm/s or 200 cm/s and an RAR of 3.5. Abu Rahma [1] described the combination of a PSV of > 285 cm/s and an RAR of > 3.5 as a viable criterion in renal artery diagnostics with a sensitivity of only 60 % but a specificity of 94 %, whereby this comparison was made with an angiographic stenosis of over 60 %. According to the ROC curve in the same study, a lower cut-off and the combination of a PSV of > 285 cm/s and an RAR of > 3.5 inevitably resulted in a significantly better sensitivity of 73 % but a worse specificity of 81 % (Table 1); however, the use of several stenosis criteria in the clinical routine is overly complex and time-consuming. For this reason, stenosis detection should be performed primarily by determining PSV and should only be complemented by RAR and dRI in doubtful situations or borderline findings.
Table 1
Color-coded duplex ultrasonography in the diagnosis of renal artery stenosis (RAS). Angiography-controlled studies on accuracy in hemodynamically relevant RAS. Various criteria are combined (direct and indirect stenosis criteria) to increase accuracy in RAS diagnosis
Author
Number (N) (stenosis degree)
Method/stenosis criterion
Sensitivity
Specificity
Reference method
Zeller et al. [53]
69 (> 70 %)
RAR > 3.5
100 %
60 %
Angiography
dRI > 0.5
77.5 %
99 %
Angiography
RAR > 3.5 and dRI > 0.05
76 %
97%
Angiography
Krumme et al. [24]
135 (> 50 %)
PSV > 200 and dRI > 0.05
89 %
92 %
Angiography
Motew et al. [27]
41 (> 60 %)
PSV > 180 cm/s
94 %
88 %
Angiography
AT > 58ms
58 %
96 %
Angiography
  
Conclusion: combination of methods recommended
 
Ripollés et al. [33]
60 (> 75 %)
AT > 80 ms
89 %
99 %
Angiography
Age < 50 years
AT > 80 ms
100 %
100 %
Angiography
Age > 50 years
AT > 80 ms
75 %
97 %
Angiography
Age < 50 years
dRI > 0.05
90 %
93 %
Angiography
Ag > 50 years
dRI > 0.05
0 %
100 %
Angiography
  
Conclusion: dRI and AT only useful as stenosis criterion in patients aged < 50 years
Rademacher et al. [32]
226 (> 50 %)
PSV
> 180 cm/s and PSV hilum
< 1/4 PSV max. (Stenose)
AT
> 70 ms
96 %
98 %
Angiography
Souza de Oliveira et al. [41]
60 (> 50 %)
PSV > 150 cm/s
83.3 %
89.3 %
Angiography
Conkbayir et al. [8]
50 (> 60 %)
PSV > 180 cm/s
89 %
88 %
Angiography
RAR > 3.0
86 %
97 %
Angiography
AT > 70 ms
48 %
93 %
Angiography
PSV > 180 cm/s or RAR > 3.0
92 %
88 %
Angiography
  
PSV > 180 cm/s or RAR > 3.0 or AT > 70 ms
87 %
86 %
Angiography
  
Conclusion: combination of methods recommended
 
Kawarada et al. [20]
94
 60%
PSV > 219
89 %
89 %
Angiography, transstenotic pressure gradient
Straub et al. [42]
49 (> 50 %)
PSV > 200
92 %
81 %
Angiographic stenosis degree, pressure gradient
  
RAR > 3.0
83 %
91 %
Angiography
dRI > 0.05
31 %
97 %
Angiography
 
49 (> 70 %)
PSV > 250 cm/s
89 %
70 %
Angiography, angiographic stenosis degree, intra-arterial pressure measurement over stenosis
RAR > 3.5
84 %
72 %
dRI > 0.05
42 %
91 %
  
PSV recommended possibly in combination with RAR (and dRI) to increase specificity
Solar et al. [40]
94
 60 %
PSV > 180
85 %
84 %
Angiography
Abu Rahma et al. [1]
313
 60%
PSV180
91 %
41 %
Angiography
  
PSV > 285
67 %
90 %
Angiography
  
RAR > 3.5
72 %
81 %
Angiography
  
PSV
 180 +
RAR
 3.5
73 %
81 %
Angiography
  
PSV
 285 +
RAR
 3.5
60 %
94 %
Angiography
Schäberle et al. [40]
91 (> 50 %)
PSV > 140 cm/s
86 %
83 %
X-ray densitometry, angiography
AT acceleration time, dRI side to side differences in resistance index, PSV peak systolic velocity, RAR renal aortic ratio.
The use of CCDS with PSV, as well as the other criteria, shows greater accuracy for the detection of high-grade stenosis. It has hitherto been assumed that only stenosis > 70 % causes a relevant poststenotic pressure drop and increased renin release as a counterregulatory action via the renin-angiotensin system. Thus, only high-grade RAS are considered as requiring treatment [15, 26, 28, 48]. It is important to reliably identify these RAS for a treatment-oriented diagnosis. The discussion on the PSV cut-off at 50 % stenosis is more academic in nature than anything else; however, even at > 50 % RSA on angiography (Grosse et al. 2001) [42] measurements of the intra-arterial systolic pressure gradient show a mean pressure gradient of > 22 mm Hg. In Staub’s study [42] a PSV of > 200 cm/s (correlating with 50 % stenosis on angiography) showed a mean systolic pressure gradient of 23 mmHg and, hence, hemodynamically relevant stenosis accompanied by a regulatory counteraction [9]; however, it should be pointed out that poststenotic pressure was measured with the catheter lying across the stenosis (additional luminal narrowing).
The method for determining the pressure drop, as validated using the PSV in iliac stenosis [45] across the stenosis using the simplified Bernoulli equation (drop in pressure dP = 4 × intrastenotic PSV2) is only reliable in high-grade stenosis, as the prestenotic PSV is considered negligible in this context. In the case of renal artery branch stenosis, the prestenotic PSV in the aorta cannot be used. The poststenotic PSV [dP = 4 × (intrastenotic PSV2– poststenotic PSV2)] occasionally used instead [46] of the prestenotic PSV (in the Bernoulli equation) is inaccurate and neglects inertial and frictional losses over the stenosis.

Contrast-enhanced ultrasonography

A study including 120 patients with 38 stenosed renal arteries reported surprisingly good results [6]. Sensitivity, specificity, PPV, NPV and accuracy were reported as being 100 %, 84 %, 0 %, 80 % and 94 %, respectively, for CCDS compared with angiography in the same study. Claudon [7] described a 20 % improvement (from 63.9 % to 83.9 %) in stenosis detection in RAS using contrast-enhanced ultrasonography (CEUS) compared with conventional CCDS. In an already somewhat older study Missouris et al. found an increase in sensitivity from 85 % to 94 % and in specificity from 79 % to 88 % with a 20 dB increase in Doppler intensity following contrast medium administration.

Ultrasound follow-up after stent placement

Using PSV and RAR in in-stent restenosis tend to show higher cut-off values at an equivalent degree of stenosis on angiography compared with native RAS [5, 12]; however, study results are conflicting. The explanation given for this in carotid artery restenosis is stent rigidity and a luminal reduction due to the stent. While Chi [5] set the ideal PSV cut-off for albeit > 70 % stenosis in stented renal arteries at > 395 cm/s and a RAR of > 5.1, Fleming [12] demonstrated a sensitivity, specificity, PPV and accuracy of 73 %, 80 %, 64 % and 77 % for > 60 % stenosis at a PSV of > 180 cm/s, 68 %, 80 %, 63 % and 76 % at > 200 cm/s and 59 %, 95 %, 87 % and 83 % at > 250 cm/s.
According also to these ROC curves, the selection of the ideal PSV cut-off should depend on the objective. If as much restenosis as possible is to be detected, a PSV of 180 cm/s should be selected due to its high sensitivity (73 %). If the focus lies on high-grade stenosis (where only this stenosis is considered relevant in terms of reintervention) a PSV with the highest PPV and specificity should be selected (87 % and 95 %, respectively for a PSV > 250 cm/s).
Controversially, Nolan [30] found similar stenosis velocity criteria for stented renal arteries compared with native stenosis (PSV > 200 cm/s and RAR > 3.5). Singh [34] also obtained the same results (PSV > 225 cm/s and RAR > 3.5 %). Napoli [29] even reduced the values for stented renal arteries (PSV from 180 cm/s to 144 cm/s and RAR from 3.5 % to 2.53 %) in order to improve sensitivity and specificity. Our own attempts to explain these results involved asking the question whether in-stent restenosis in this collective was more eccentric, which exhibits less area reduction and hence less hemodynamic relevance and lower PSV compared with concentric stenosis at equivalent angiographic diameter reduction.
Besides the problems already described for native stenosis in stenosis detection using ultrasound compared with angiography, other weaknesses of poststenting studies lie in the low case numbers in generally retrospective, single center studies. Additional sources of error include a selection bias (angiographic follow-up only in clinically and sonographically proven pathological results), lack of information on the degree to which stenosis could be assessed (ultrasound conditions and angle alignment errors) and the impact of systemic factors on hemodynamics, meaning that some authors [5, 12] themselves questioned the possibility that their study results were a generalization (Fig. 7).

Computed tomography angiography

The advent of multislice CTA and its ability to gather high-speed, thin-slice volume data sets has made it possible, in contrast to earlier technologies, to adequately assess the renal arteries. Although intra-arterial digital subtraction angiography (DSA) is still considered the gold standard, CTA overcomes the limitation whereby vessels (or luminal narrowing) can be visualized only intraluminally and provides images on wall calcification and lumen-narrowing plaque via three-dimensional data sets. Thus, studies conducted in recent years have shown sensitivities of 90–100 % and specificities of 92–98 % [4, 21, 35, 52]. Only the controversially discussed (study design) prospective multicenter Renal Artery Diagnostic Imaging Study in Hypertension (RADISH) has shown sobering results, with a sensitivity of 64 % and specificity of 92 %. Problems arise in the assessment of accessory renal arteries (pole arteries). With sensitivities of 100 % and specificities of 99 %, CTA yielded very good results in the assessment of in-stent stenosis following stent-assisted PTA in a study on 95 renal artery stents [44]. In addition to radiation exposure and possible contrast medium-induced nephropathy, particularly in patients with pre-existing parenchymal damage, the snapshot in time obtained with CTA and its inability to visualize hemodynamics, represents a further disadvantage.

Magnetic resonance angiography

With sensitivities and specificities of 88–100 % [50], contrast-enhanced MRA is well suited as a non-invasive method to diagnose RAS; however, as known from other vascular territories, it overestimates the degree of stenosis by 26–32 % [14, 23, 43]. It also shows good accuracy in the visualization of accessory renal arteries [3, 14, 39, 51] as well as in the assessment of fibromuscular dysplasia [53]; however, the sensitivity was only 68 % for the grading of relevant stenosis in fibromuscular dysplasia.
In addition to a morphological assessment of the renal arteries, MRA also permits a functional assessment of the kidneys (renal parenchymal flow). Besides the known contraindications (i.e. nephrophathy and pacemakers), MRA is also susceptible to artifacts from neighboring metal or gas-containing organs, which can simulate stenosis. Furthermore, the broad application of gadolinium-enhanced MRA is limited due to the risk of nephrotoxicity and fibrosis, particularly in patients with impaired renal function and reduced glomerular filtration rate.

Conclusion

Being a non-invasive, cost-effective and radiation-free examination method with good sensitivities and specificities (approximately 90 %), CCDS is well suited to the diagnosis of RAS. An intrastenotic PSV > 180− 200 cm/s represents the most accurate stenosis criterion for stenosis > 50 %. In the case of inconclusive findings, this criterion can be complemented by RAR or indirect criteria. It is possible to improve accuracies using CEUS. The MRA and CTA, with their good accuracy in the case of doubtful or borderline findings, offer an additional investigation method, albeit at the cost of higher patient exposure. Alternatively, when clinical findings are relevant to treatment, DSA can be performed on stand-by to convert to PTA (Fig. 8).

Compliance with ethical guidelines

Conflict of interest

W. Schäberle, L. Leyerer, W. Schierling and K. Pfister state that there are no conflicts of interest.
This article does not include studies on humans or animals.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Gefässchirurgie

Print-Titel

Themenschwerpunkte zu aktuellen Entwicklungen in der vaskulären und endovaskulären Gefäßmedizin. Vermittlung von relevantem Hintergrundwissen und Bewertung wissenschaftlicher Ergebnisse. Konkrete Handlungsempfehlungen.

Anhänge

Caption Electronic Supplementary Material

Video clip 1
Stenosis of 50–60 % in the middle third of the left renal artery due to fibromuscular dysplasia. Stenosis grading based on the continuity equation: PSV ratio where PSV intrastenotic/PSV prestenotic (at the origin of the renal artery) = 220 cm/s / 80 cm/s = 2.7. The video clip shows the intrastenotic PSV (in the middle third of the renal artery) and the prestenotic PSV (proximal third of the renal artery). Artery movement due to breathing complicates the measurement.
Video clip 2
High-grade renal artery recurrent in-stent stenosis after PTA and stent with PSV > 5 m/s. Images from conventional duplex ultrasound (B-mode image allows better visualization of the stent course) and color duplex ultrasound are both shown in an attempt to optimally display the stenosis jet for determination of the maximum PSV and therefore for adequate grading of the recurrent stenosis. Artery movement due to breathing complicates the measurement (the "sample volume" moves out of the vessel lumen/stent in a breathing-dependent manner). The stent (hyperechogenic, wire mesh pattern) protrudes slightly into the aortic lumen.
Literatur
1.
Zurück zum Zitat Abu Rahma AF, Srivastava M, Mousa AY (2012) Critical analysis of renal duplex ultrasound parameters in detecting significant renal artery stenosis. J Vasc Surg 56:1052–1060CrossRef Abu Rahma AF, Srivastava M, Mousa AY (2012) Critical analysis of renal duplex ultrasound parameters in detecting significant renal artery stenosis. J Vasc Surg 56:1052–1060CrossRef
2.
Zurück zum Zitat Avasthi PS, Voyles WF, Greene ER (1984) Noninvasive diagnosis of renal artery stenosis by echo- Doppler velocimetry. Kidney Int 25:824–829CrossRefPubMed Avasthi PS, Voyles WF, Greene ER (1984) Noninvasive diagnosis of renal artery stenosis by echo- Doppler velocimetry. Kidney Int 25:824–829CrossRefPubMed
3.
Zurück zum Zitat Bakker J, Beek FJ, Beutler JJ (1998) Renal artery stenosis and accessory renal arteries: accuracy of detection and visualization with gadolinium-enhanced breath-hold MR angiography. Radiology 207:497–504CrossRefPubMed Bakker J, Beek FJ, Beutler JJ (1998) Renal artery stenosis and accessory renal arteries: accuracy of detection and visualization with gadolinium-enhanced breath-hold MR angiography. Radiology 207:497–504CrossRefPubMed
4.
Zurück zum Zitat Beregi JP, Louvegny S, Ceugnart L (1997) Helical xray computed tomography of renal arteries. Apropos of 300 patients. J Radiol 78:549–556PubMed Beregi JP, Louvegny S, Ceugnart L (1997) Helical xray computed tomography of renal arteries. Apropos of 300 patients. J Radiol 78:549–556PubMed
5.
Zurück zum Zitat Chi YW, White CJ, Thornton S, Milani RV (2009) Ultrasound velocity criteria for renal in-stent restenosis. J Vasc Surg 50:119–123CrossRefPubMed Chi YW, White CJ, Thornton S, Milani RV (2009) Ultrasound velocity criteria for renal in-stent restenosis. J Vasc Surg 50:119–123CrossRefPubMed
6.
Zurück zum Zitat Ciccone MM, Cortese F, Fiorella A, Scicchitano P, Cito F, Quistelli G, Pertosa G, D’Agostino R, Guida P, Favale S (2011) The clinical role of contrast-enhanced ultrasound in the evaluation of renal artery stenosis and diagnostic superiority as compared to traditional echo-color-Doppler flow imaging. Int Angiol 30:135–139PubMed Ciccone MM, Cortese F, Fiorella A, Scicchitano P, Cito F, Quistelli G, Pertosa G, D’Agostino R, Guida P, Favale S (2011) The clinical role of contrast-enhanced ultrasound in the evaluation of renal artery stenosis and diagnostic superiority as compared to traditional echo-color-Doppler flow imaging. Int Angiol 30:135–139PubMed
7.
Zurück zum Zitat Claudon M, Plouin PF, Baxter GM, Rohban T, Devos DM (2002) Renal arteries in patients at risk of renal arterial stenosis: multicenter evaluation of the echo-enhancer SH U 508A at color and spectral Doppler US. Levovist Renal Artery Stenosis Study Group. Radiology 214:739–746CrossRef Claudon M, Plouin PF, Baxter GM, Rohban T, Devos DM (2002) Renal arteries in patients at risk of renal arterial stenosis: multicenter evaluation of the echo-enhancer SH U 508A at color and spectral Doppler US. Levovist Renal Artery Stenosis Study Group. Radiology 214:739–746CrossRef
8.
Zurück zum Zitat Conkbayir I, Yucesoy C, Edguer T (2003) Doppler sonography in renal artery stenosis. An evaluation of intrarenal and extrarenal imaging parameters. Clin Imaging 27:256–260CrossRefPubMed Conkbayir I, Yucesoy C, Edguer T (2003) Doppler sonography in renal artery stenosis. An evaluation of intrarenal and extrarenal imaging parameters. Clin Imaging 27:256–260CrossRefPubMed
9.
Zurück zum Zitat De Bruyne B, Manoharan G, Pijls NH (2006) Assessment of renal stenosis severity by pressure gradient measurements. J Am Coll Cardiol 48:1851–1855CrossRefPubMed De Bruyne B, Manoharan G, Pijls NH (2006) Assessment of renal stenosis severity by pressure gradient measurements. J Am Coll Cardiol 48:1851–1855CrossRefPubMed
10.
11.
Zurück zum Zitat Ferretti G, Salomone A, Castagno PL (1988) Renovascular hypertension: a non-invasiv duplex scanning screening. Int Angiol 7:219–223PubMed Ferretti G, Salomone A, Castagno PL (1988) Renovascular hypertension: a non-invasiv duplex scanning screening. Int Angiol 7:219–223PubMed
12.
Zurück zum Zitat Fleming SH, Ross PD, Timothy EC (2010) Accuracy of duplex sonography scans after renal artery stenting. J Vasc Surg 52:953–958CrossRefPubMed Fleming SH, Ross PD, Timothy EC (2010) Accuracy of duplex sonography scans after renal artery stenting. J Vasc Surg 52:953–958CrossRefPubMed
13.
Zurück zum Zitat Garovic VD, Textor SC (2005) Renovascular hypertension and ischemic nephropathy. Circulation 112:1362–1374CrossRefPubMed Garovic VD, Textor SC (2005) Renovascular hypertension and ischemic nephropathy. Circulation 112:1362–1374CrossRefPubMed
14.
Zurück zum Zitat Glifeather M, Yoon HC, Siegelman ES (1999) Renal artery stenosis: evaluation with conventional angiography versus gadolinium-enhanced MR angiography. Radiology 210:367–372CrossRef Glifeather M, Yoon HC, Siegelman ES (1999) Renal artery stenosis: evaluation with conventional angiography versus gadolinium-enhanced MR angiography. Radiology 210:367–372CrossRef
15.
Zurück zum Zitat Guo Z, Fenster A (1996) Three-dimensional power Doppler imaging: a phantom study to quantify vessel stenosis. Ultrasound Med Biol 22:1059–1069CrossRefPubMed Guo Z, Fenster A (1996) Three-dimensional power Doppler imaging: a phantom study to quantify vessel stenosis. Ultrasound Med Biol 22:1059–1069CrossRefPubMed
16.
Zurück zum Zitat Hansen KJ, Tribble RW, Reavis SW (1990) Renal duplex sonography: evaluation of clinical utility. J VascSurg 12:227–236 Hansen KJ, Tribble RW, Reavis SW (1990) Renal duplex sonography: evaluation of clinical utility. J VascSurg 12:227–236
17.
Zurück zum Zitat Hawkins PG, McKnoulty LM, Gordon RD (1989) Noninvasive renal artery duplex ultrasound and computerized nuclear renography to screen for and follow progress in renal artery stenosis. J Hypertens Suppl 7:184–185CrossRef Hawkins PG, McKnoulty LM, Gordon RD (1989) Noninvasive renal artery duplex ultrasound and computerized nuclear renography to screen for and follow progress in renal artery stenosis. J Hypertens Suppl 7:184–185CrossRef
18.
Zurück zum Zitat Hirsch AT, Haskal ZJ, Hertzer NR (2006) ACC/AHA 2005 partice guidelines for the management of patients with peripheral arterial disease (lower extremity, renal, mesenteric, and abdominal aortic): a collaborative report from the American Association for Vascular Surgery/Society for Vascular Surgery. Society for Cardiovascular Angiography and Interventions, Society for Vascular Medicine and Biology, Society of Interventional Radiology, and the ACC/AHA Task Force on Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Peripheral Arterial Disease). Circulation 113:e463–e654CrossRefPubMed Hirsch AT, Haskal ZJ, Hertzer NR (2006) ACC/AHA 2005 partice guidelines for the management of patients with peripheral arterial disease (lower extremity, renal, mesenteric, and abdominal aortic): a collaborative report from the American Association for Vascular Surgery/Society for Vascular Surgery. Society for Cardiovascular Angiography and Interventions, Society for Vascular Medicine and Biology, Society of Interventional Radiology, and the ACC/AHA Task Force on Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Peripheral Arterial Disease). Circulation 113:e463–e654CrossRefPubMed
19.
Zurück zum Zitat Karasch T, Strauss AL, Grün B, Worringer M, Neuerburg-Heusler D, Roth FJ, Rieger H (1993) Farbkodierte Duplexsonographie in der Diagnostik von Nierenarterienstenosen. Dtsch Med Wochschr 118:1429–1436CrossRef Karasch T, Strauss AL, Grün B, Worringer M, Neuerburg-Heusler D, Roth FJ, Rieger H (1993) Farbkodierte Duplexsonographie in der Diagnostik von Nierenarterienstenosen. Dtsch Med Wochschr 118:1429–1436CrossRef
20.
Zurück zum Zitat Kawarada O, Yokoi Y, Yakemoto K et al (2006) The performance of renal ultrasonography for the detection of hemodynamically significant renal artery stenosis. Catheter Cardiovasc Interv 68:311–318CrossRefPubMed Kawarada O, Yokoi Y, Yakemoto K et al (2006) The performance of renal ultrasonography for the detection of hemodynamically significant renal artery stenosis. Catheter Cardiovasc Interv 68:311–318CrossRefPubMed
21.
Zurück zum Zitat Kim TS, Chung JW, Park JH (1998) Renal artery evaluation: comparison of spiral CT angiographyto intra-arterial DSA. J Vasc Interv Radiol 9:553–559CrossRefPubMed Kim TS, Chung JW, Park JH (1998) Renal artery evaluation: comparison of spiral CT angiographyto intra-arterial DSA. J Vasc Interv Radiol 9:553–559CrossRefPubMed
22.
Zurück zum Zitat Kohler TR, Zierler RE, Martin RL (1986) Noninvasive diagnosis of renal artery stenosis by ultrasonic duplex scanning. J Vasc Surg 4:450–456CrossRefPubMed Kohler TR, Zierler RE, Martin RL (1986) Noninvasive diagnosis of renal artery stenosis by ultrasonic duplex scanning. J Vasc Surg 4:450–456CrossRefPubMed
23.
Zurück zum Zitat Krinsky G, Rofsky N, Giangola G (1996) Gadolinium-enhanced three-dimensional MR angiography of acquired arch vessel disease. AJR Am J Roentgenol 167:981–987CrossRefPubMed Krinsky G, Rofsky N, Giangola G (1996) Gadolinium-enhanced three-dimensional MR angiography of acquired arch vessel disease. AJR Am J Roentgenol 167:981–987CrossRefPubMed
24.
Zurück zum Zitat Krumme B, Blum U, Schwertfeger E (1996) Diagnosis of renovascular disease by intra- and extrarenal Doppler scanning. Kidney Int 50:1288–1292CrossRefPubMed Krumme B, Blum U, Schwertfeger E (1996) Diagnosis of renovascular disease by intra- and extrarenal Doppler scanning. Kidney Int 50:1288–1292CrossRefPubMed
25.
Zurück zum Zitat Leiner T, de Haan MW, Nelemans PJ (2005) Contemporary imaging techniques for the diagnosis of renal artery stenosis. Eur Radiol 15:2219–2229CrossRefPubMed Leiner T, de Haan MW, Nelemans PJ (2005) Contemporary imaging techniques for the diagnosis of renal artery stenosis. Eur Radiol 15:2219–2229CrossRefPubMed
26.
Zurück zum Zitat May AG, Deweese JA, Rob CG (1963) Hemodynamic effects of arterial stenosis. Surgery 53:513–524PubMed May AG, Deweese JA, Rob CG (1963) Hemodynamic effects of arterial stenosis. Surgery 53:513–524PubMed
27.
Zurück zum Zitat Motew SJ, Cherr GS, Craven TE (2000) Renal duplex sonography: main renal artery versus hilaranaylsis. J Vasc Surg 32:462–471CrossRefPubMed Motew SJ, Cherr GS, Craven TE (2000) Renal duplex sonography: main renal artery versus hilaranaylsis. J Vasc Surg 32:462–471CrossRefPubMed
28.
Zurück zum Zitat Muster BR, Williams DM, Prince MR (1998) In vitro model of arterial stenosis: correlation of MR signal dephasing and trans-stenotic pressure gradients. Magn Reson Imaging 16:301–310CrossRef Muster BR, Williams DM, Prince MR (1998) In vitro model of arterial stenosis: correlation of MR signal dephasing and trans-stenotic pressure gradients. Magn Reson Imaging 16:301–310CrossRef
29.
Zurück zum Zitat Napoli V, Pinto S, Bargellini I, Vignali C, cioni R, Petruzzi P (2002) Duplex ultrasonographic study of the renal arteries before and after renal artery stenting. Eur Radiol 12:796–803CrossRefPubMed Napoli V, Pinto S, Bargellini I, Vignali C, cioni R, Petruzzi P (2002) Duplex ultrasonographic study of the renal arteries before and after renal artery stenting. Eur Radiol 12:796–803CrossRefPubMed
30.
Zurück zum Zitat Nolan BW, Schermerhorn ML, Powell RJ, Rowell E (2005) Restenosis in gold-coated renal artery stents. J Vasc Surg 42:40–46CrossRefPubMed Nolan BW, Schermerhorn ML, Powell RJ, Rowell E (2005) Restenosis in gold-coated renal artery stents. J Vasc Surg 42:40–46CrossRefPubMed
32.
Zurück zum Zitat Radermacher J, Chavan A, Schäffer J (2000) Detection of significant renal artery stenosis with color Doppler sonography: combining extrarenal and intrarenal approaches to minimize technical failure. Clin Nephrol 53:333–343PubMed Radermacher J, Chavan A, Schäffer J (2000) Detection of significant renal artery stenosis with color Doppler sonography: combining extrarenal and intrarenal approaches to minimize technical failure. Clin Nephrol 53:333–343PubMed
33.
Zurück zum Zitat Ripolles T, Aliaga R, Morote V (2001) Utility of intrarenal Doppler ultrasound in the diagnosis of renal artery stenosis. Eur J Radiol 40:54–63CrossRefPubMed Ripolles T, Aliaga R, Morote V (2001) Utility of intrarenal Doppler ultrasound in the diagnosis of renal artery stenosis. Eur J Radiol 40:54–63CrossRefPubMed
34.
Zurück zum Zitat Rocha-Singh K, Jaff MR, Lynne Kelly E (2008) RENAISSANCE Trial Investigators. renal artery stenting with noninvasive duplex ultrasound followup: 3-year results from the RENAISSANCE renal stent trial. Catheter Cardiovasc Interv 72:853–862CrossRefPubMed Rocha-Singh K, Jaff MR, Lynne Kelly E (2008) RENAISSANCE Trial Investigators. renal artery stenting with noninvasive duplex ultrasound followup: 3-year results from the RENAISSANCE renal stent trial. Catheter Cardiovasc Interv 72:853–862CrossRefPubMed
35.
Zurück zum Zitat Rountas C, Vlychou M, Vassiou K (2007) Imaging modalities for renal artery stenosis in suspected renovascular hypertension: prospective intraindividual comparison of color Doppler US, CT angiography, GD-enhanced MR angiography and digital subtraction angiography. Ren Fail 29:295–302CrossRefPubMed Rountas C, Vlychou M, Vassiou K (2007) Imaging modalities for renal artery stenosis in suspected renovascular hypertension: prospective intraindividual comparison of color Doppler US, CT angiography, GD-enhanced MR angiography and digital subtraction angiography. Ren Fail 29:295–302CrossRefPubMed
37.
Zurück zum Zitat Schäberle W (2009) Ultraschall in der Gefäßdiagnostik, 3. Aufl. Springer, Heidelberg, S 338, 402–409, 450–455 Schäberle W (2009) Ultraschall in der Gefäßdiagnostik, 3. Aufl. Springer, Heidelberg, S 338, 402–409, 450–455
38.
Zurück zum Zitat Schäberle W, Strauss A, Neuerburg-Heusler D, Roth FJ (1992) Wertigkeit der Duplexsonographie in der Diagnostik der Nierenarterienstenose und ihre Eignung in der Verlaufskontrolle nach Angioplastie (PTA). Ultraschall Med 13:271–276CrossRefPubMed Schäberle W, Strauss A, Neuerburg-Heusler D, Roth FJ (1992) Wertigkeit der Duplexsonographie in der Diagnostik der Nierenarterienstenose und ihre Eignung in der Verlaufskontrolle nach Angioplastie (PTA). Ultraschall Med 13:271–276CrossRefPubMed
39.
Zurück zum Zitat Shetty AN, Bis KG, Kirsch M (2000) Contrast-enhanced breath-hold three-dimensional magnetic resonance angiography in the evaluation of renal arteries: optimization of technique and pitfalls. J Magn Reson Imaging 12:912–923CrossRefPubMed Shetty AN, Bis KG, Kirsch M (2000) Contrast-enhanced breath-hold three-dimensional magnetic resonance angiography in the evaluation of renal arteries: optimization of technique and pitfalls. J Magn Reson Imaging 12:912–923CrossRefPubMed
40.
Zurück zum Zitat Solar M, Zizka J, Krajina A, Michl A, Raupach J, Kizo L, Ryska P, Ceral J (2011) Comparison of duplex ultrasonography and magnetic resonance imaging in the detection of significant renal artery stenosis. Acta Medica 54(1):9–12PubMed Solar M, Zizka J, Krajina A, Michl A, Raupach J, Kizo L, Ryska P, Ceral J (2011) Comparison of duplex ultrasonography and magnetic resonance imaging in the detection of significant renal artery stenosis. Acta Medica 54(1):9–12PubMed
41.
Zurück zum Zitat Souza de Oliveira IR, Widmann A, Molnar LJ (2000) Colour Doppler ultrasound: a new indes improves the diagnosis of renal artery stenosis. Ultrasound Med Biol 26:41–47CrossRefPubMed Souza de Oliveira IR, Widmann A, Molnar LJ (2000) Colour Doppler ultrasound: a new indes improves the diagnosis of renal artery stenosis. Ultrasound Med Biol 26:41–47CrossRefPubMed
42.
Zurück zum Zitat Staub D, Canevascini R, Huegli RW, Aschwanden M, Thalhammer C, Imfeld S, Singer E, Jacob AL, Jaeger KA (2007) Best duplex-sonographic criteria for the assessment of renal artery stenosis – correlation with intra-arterial pressure gradient. Ultraschall Med 28:45–51CrossRefPubMed Staub D, Canevascini R, Huegli RW, Aschwanden M, Thalhammer C, Imfeld S, Singer E, Jacob AL, Jaeger KA (2007) Best duplex-sonographic criteria for the assessment of renal artery stenosis – correlation with intra-arterial pressure gradient. Ultraschall Med 28:45–51CrossRefPubMed
43.
Zurück zum Zitat Steffens JC, Link J, Graessnerr J (1997) Contrast-enhanced, K-space-centered, breath-hold MR angiography of renal arteries and the abdominal aorta. J Magn Reson Imaging 7:617–622CrossRefPubMed Steffens JC, Link J, Graessnerr J (1997) Contrast-enhanced, K-space-centered, breath-hold MR angiography of renal arteries and the abdominal aorta. J Magn Reson Imaging 7:617–622CrossRefPubMed
44.
Zurück zum Zitat Steinweder C, Schutzenberger W, Fellner F (2009) 64-Detector CT angiography in renal artery stent evaluation: prospective comparison with selective catheter angiography. Radiology 252:299–305CrossRef Steinweder C, Schutzenberger W, Fellner F (2009) 64-Detector CT angiography in renal artery stent evaluation: prospective comparison with selective catheter angiography. Radiology 252:299–305CrossRef
45.
Zurück zum Zitat Strauss AL, Roth FJ, Rieger H (1993). Noninvasive assessment of pressure gradients across iliac artery stenosis: duplex and catheter correlative study. J Ultrasound Med 12:17–22PubMed Strauss AL, Roth FJ, Rieger H (1993). Noninvasive assessment of pressure gradients across iliac artery stenosis: duplex and catheter correlative study. J Ultrasound Med 12:17–22PubMed
46.
Zurück zum Zitat Stock KF (2009) Ultraschalldiagnostik der Nierengefäße und der Transplantatniere. Radiologe 49:1040–1047CrossRefPubMed Stock KF (2009) Ultraschalldiagnostik der Nierengefäße und der Transplantatniere. Radiologe 49:1040–1047CrossRefPubMed
47.
Zurück zum Zitat Taylor DC, Kettler MD, Monetta GL et al (1988) Duplex ultrasound in the diagnosis of renal artery: a prospective evaluation. J Vasc Surg 7:363–769CrossRefPubMed Taylor DC, Kettler MD, Monetta GL et al (1988) Duplex ultrasound in the diagnosis of renal artery: a prospective evaluation. J Vasc Surg 7:363–769CrossRefPubMed
48.
Zurück zum Zitat Textor SC (1994) Renovascular Hypertension. Endocrinol Metabol Clin North Am 23:235–253 Textor SC (1994) Renovascular Hypertension. Endocrinol Metabol Clin North Am 23:235–253
49.
Zurück zum Zitat Van Jaarsveld BC, Pieterman H, van Dijk LC, van Seijen AJ, Derkx FH, Man in’t Veld AJ, Schalekamp MA (1999) Inter-observer variability in the angiographic assessment of renal artery stenosis. DRASTIC study group. Dutch Renal Artery Stenosis Intervention Cooperative. J Hypertense 17(12 Pt 1):1731–1736CrossRef Van Jaarsveld BC, Pieterman H, van Dijk LC, van Seijen AJ, Derkx FH, Man in’t Veld AJ, Schalekamp MA (1999) Inter-observer variability in the angiographic assessment of renal artery stenosis. DRASTIC study group. Dutch Renal Artery Stenosis Intervention Cooperative. J Hypertense 17(12 Pt 1):1731–1736CrossRef
50.
Zurück zum Zitat Vasbinder GB, Nelemans PJ, Kessels AG (2001) Diagnostic tests for renal artery stenosis in patients suspected of having renovascular hypertension: a meta-analysis. Ann Intern Med 135:401–411CrossRefPubMed Vasbinder GB, Nelemans PJ, Kessels AG (2001) Diagnostic tests for renal artery stenosis in patients suspected of having renovascular hypertension: a meta-analysis. Ann Intern Med 135:401–411CrossRefPubMed
51.
Zurück zum Zitat Willoteaux S, Faivre-Pierret M, Moranne O (2006) Fibromuscular dysplasia of the main renal arteries: comparison of contrast-enhanced MR angiography with digital subtraction angiography. Radiology 241:922–929CrossRefPubMed Willoteaux S, Faivre-Pierret M, Moranne O (2006) Fibromuscular dysplasia of the main renal arteries: comparison of contrast-enhanced MR angiography with digital subtraction angiography. Radiology 241:922–929CrossRefPubMed
52.
53.
Zurück zum Zitat Zeller T, Frank U, Späth M (2001) Farbduplexsonographische Darstellbarkeit von Nierenarterien und Erkennung hämodynamisch relevanter Nierenarterienstenosen. Ultraschall Med 22:116–121CrossRefPubMed Zeller T, Frank U, Späth M (2001) Farbduplexsonographische Darstellbarkeit von Nierenarterien und Erkennung hämodynamisch relevanter Nierenarterienstenosen. Ultraschall Med 22:116–121CrossRefPubMed
Metadaten
Titel
Ultrasound diagnostics of renal artery stenosis
Stenosis criteria, CEUS and recurrent in-stent stenosis
verfasst von
Dr. W. Schäberle
L. Leyerer
W. Schierling
K. Pfister
Publikationsdatum
01.02.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Gefässchirurgie / Ausgabe Sonderheft 1/2016
Print ISSN: 0948-7034
Elektronische ISSN: 1434-3932
DOI
https://doi.org/10.1007/s00772-015-0060-3

Weitere Artikel der Sonderheft 1/2016

Gefässchirurgie 1/2016 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.