Development of a Noise Reduction Filter Algorithm for Pediatric Body Images in Multidetector CT

verfasst von: Eiji Nishimaru, Katsuhiro Ichikawa, Izumi Okita, Yukihiro Tomoshige, Takehiro Kurokawa, Yuko Nakamura, Masayuki Suzuki

Erschienen in: Journal of Imaging Informatics in Medicine | Ausgabe 6/2010

Einloggen, um Zugang zu erhalten

Abstract

Recently, several types of post-processing image filter which was designed to reduce noise allowing a corresponding dose reduction in CT images have been proposed and these were reported to be useful for noise reduction of CT images of adult patients. However, these have not been reported on adaptation for pediatric patients. Because they are not very effective with small (<20 cm) display fields of view, they could not be used for pediatric (e.g., premature babies and infants) body CT images. In order to solve this restriction, we have developed a new noise reduction filter algorithm which can be applicable for pediatric body CT images. This algorithm is based on a three-dimensional post processing, in which output pixel values are calculated by multi-directional, one-dimensional median filters on original volumetric datasets. The processed directions were selected except in in-plane (axial plane) direction, and consequently the in-plane spatial resolution was not affected by the filter. Also, in other directions, the spatial resolutions including slice thickness were almost maintained due to a characteristic of non-linear filtering of the median filter. From the results of phantom studies, the proposed algorithm could reduce standard deviation values as a noise index by up to 30% without affecting the spatial resolution of all directions, and therefore, contrast-to-noise ratio was improved by up to 30%. This newly developed filter algorithm will be useful for the diagnosis and radiation dose reduction of pediatric body CT images.
Literatur
1.
Zurück zum Zitat Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ: Multi-detector row CT systems and image reconstruction techniques. Radiology 235:756–773, 2005CrossRefPubMed Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ: Multi-detector row CT systems and image reconstruction techniques. Radiology 235:756–773, 2005CrossRefPubMed
2.
Zurück zum Zitat Puqliese F, Weustink AC, Van Mieqhem C, Alberghina F, Otsuka M, Meijboom WB, Van Pelt N, Mollet NR, Cademartiri F, Krestin GP, Hunink MG, de Feyter PJ: Dual-source coronary computed tomography angiography for detecting in-stent restenosis. Heart Sep 19, 2007 Puqliese F, Weustink AC, Van Mieqhem C, Alberghina F, Otsuka M, Meijboom WB, Van Pelt N, Mollet NR, Cademartiri F, Krestin GP, Hunink MG, de Feyter PJ: Dual-source coronary computed tomography angiography for detecting in-stent restenosis. Heart Sep 19, 2007
3.
Zurück zum Zitat Frush DP, Donnelly LF: Helical CT in children: technical considerations and body applications. Radiology 209:37–48, 1998PubMed Frush DP, Donnelly LF: Helical CT in children: technical considerations and body applications. Radiology 209:37–48, 1998PubMed
4.
Zurück zum Zitat Brenner DJ: Estimating cancer risks from the pediatric CT: going from the qualitative to the quantitative. Pediatr Radiol 32:228–231, 2002CrossRefPubMed Brenner DJ: Estimating cancer risks from the pediatric CT: going from the qualitative to the quantitative. Pediatr Radiol 32:228–231, 2002CrossRefPubMed
5.
Zurück zum Zitat Brenner DJ, Elliston CD, Hall EJ, Berdon WE: Estimates of the cancer risks from pediatric CT radiation are not merely theoretical. Comment on point/counterpoint: in X-ray computed tomography, technique factors should be selected appropriate to patient size against the proposition (letter). Med Phys 28:2387–2388, 2001CrossRefPubMed Brenner DJ, Elliston CD, Hall EJ, Berdon WE: Estimates of the cancer risks from pediatric CT radiation are not merely theoretical. Comment on point/counterpoint: in X-ray computed tomography, technique factors should be selected appropriate to patient size against the proposition (letter). Med Phys 28:2387–2388, 2001CrossRefPubMed
6.
Zurück zum Zitat Brody AS, Frush DP, Huda W, Robelt LB: Radiation risk to children from computed tomography. Pediatrics 120:677–682, 2007CrossRefPubMed Brody AS, Frush DP, Huda W, Robelt LB: Radiation risk to children from computed tomography. Pediatrics 120:677–682, 2007CrossRefPubMed
7.
Zurück zum Zitat Laurence K, Yair S, Solange A: Nonlinear filters applied on computerized axial tomography: theory and phantom images. Med Phys 19(4):1057–1064, 1992CrossRef Laurence K, Yair S, Solange A: Nonlinear filters applied on computerized axial tomography: theory and phantom images. Med Phys 19(4):1057–1064, 1992CrossRef
8.
Zurück zum Zitat Michelle H, Cheryl D: Image filtering for improved dose resolution in CT polymer gel dosimetry. Med Phys 31:39–49, 2004CrossRef Michelle H, Cheryl D: Image filtering for improved dose resolution in CT polymer gel dosimetry. Med Phys 31:39–49, 2004CrossRef
9.
Zurück zum Zitat Kalra MK, Maher MM, Blake MA, Lucey BC, Karau K, Toth TL, Avinash G, Halpern E, Saini S: Detection and characterization of lesions on low-radiation-dose abdominal CT images postprocessed with noise reduction filters. Radiology 232:791–797, 2004CrossRefPubMed Kalra MK, Maher MM, Blake MA, Lucey BC, Karau K, Toth TL, Avinash G, Halpern E, Saini S: Detection and characterization of lesions on low-radiation-dose abdominal CT images postprocessed with noise reduction filters. Radiology 232:791–797, 2004CrossRefPubMed
10.
Zurück zum Zitat Okumura M, Ota T, Tsukagoshi S, Katada K: New method of evaluating edge-preserving adaptive filters for computed tomography (CT): digital phantom method. Nippon Hoshasen Gijutsu Gakkai Zasshi 62(7):971–978, 2006PubMed Okumura M, Ota T, Tsukagoshi S, Katada K: New method of evaluating edge-preserving adaptive filters for computed tomography (CT): digital phantom method. Nippon Hoshasen Gijutsu Gakkai Zasshi 62(7):971–978, 2006PubMed
11.
Zurück zum Zitat Sasaki T, Sasaki M, Hanari T, Gakumazawa H, Noshi Y, Okumura M: Improvement in image quality of noncontrast head images in multidetector-row CT by volume helical scanning with a three-dimensional denoising filter. Radiat Med 25:368–372, 2007CrossRefPubMed Sasaki T, Sasaki M, Hanari T, Gakumazawa H, Noshi Y, Okumura M: Improvement in image quality of noncontrast head images in multidetector-row CT by volume helical scanning with a three-dimensional denoising filter. Radiat Med 25:368–372, 2007CrossRefPubMed
12.
Zurück zum Zitat Funama Y, Awai K, Miyazaki O, Nakayama Y, Goto T, Omi Y, Shimonobo T, Liu D, Yamashita Y, Hori S: Improvement of low-contrast detectability in low-dose hepatic multidetector computed tomography using a novel adaptive filter: evaluation with a computer-simulated liver including tumors. Invest Radiol 41(1):1–7, 2006CrossRefPubMed Funama Y, Awai K, Miyazaki O, Nakayama Y, Goto T, Omi Y, Shimonobo T, Liu D, Yamashita Y, Hori S: Improvement of low-contrast detectability in low-dose hepatic multidetector computed tomography using a novel adaptive filter: evaluation with a computer-simulated liver including tumors. Invest Radiol 41(1):1–7, 2006CrossRefPubMed
13.
Zurück zum Zitat McDonnel MJ: Box-filtering techniques. Comput Graph Image Process 17:65–70, 1981CrossRef McDonnel MJ: Box-filtering techniques. Comput Graph Image Process 17:65–70, 1981CrossRef
14.
Zurück zum Zitat Itoh K, Ichioka Y, Minami T: Nearest-neighbor median filter. Appl Opt 27(16):3445–3450, 1988CrossRefPubMed Itoh K, Ichioka Y, Minami T: Nearest-neighbor median filter. Appl Opt 27(16):3445–3450, 1988CrossRefPubMed
15.
Zurück zum Zitat Nodes TA, Gallagher NC: The output distribution of median type filters. IEEE Trans Commun 32(5):532–541, 1984CrossRef Nodes TA, Gallagher NC: The output distribution of median type filters. IEEE Trans Commun 32(5):532–541, 1984CrossRef
18.
Zurück zum Zitat Riederer SJ, Pelc NJ, Chesler DA: The noise power spectrum in computed X-ray tomography. Phys Med Biol 23(3):446–454, 1978CrossRefPubMed Riederer SJ, Pelc NJ, Chesler DA: The noise power spectrum in computed X-ray tomography. Phys Med Biol 23(3):446–454, 1978CrossRefPubMed
19.
Zurück zum Zitat Harpen MD: A computer simulation of wavelet noise reduction in computed tomography. Med Phys 26(8):1600–1606, 1999CrossRefPubMed Harpen MD: A computer simulation of wavelet noise reduction in computed tomography. Med Phys 26(8):1600–1606, 1999CrossRefPubMed
20.
Zurück zum Zitat Ichikawa K: Fundamentals and applications of image evaluation in digital age—image evaluation in computed tomography. Nippon Hoshasen Gijutsu Gakkai Zasshi 58(1):37–40, 2002 (in Japanese)PubMed Ichikawa K: Fundamentals and applications of image evaluation in digital age—image evaluation in computed tomography. Nippon Hoshasen Gijutsu Gakkai Zasshi 58(1):37–40, 2002 (in Japanese)PubMed
21.
Zurück zum Zitat Verdun FR, Denys A, Valley JF, Schnyder P, Meuli RA: Detection of low-contrast objects: experimental comparison of single- and multi-detector row CT with a phantom. Radiology 223:426–431, 2002CrossRefPubMed Verdun FR, Denys A, Valley JF, Schnyder P, Meuli RA: Detection of low-contrast objects: experimental comparison of single- and multi-detector row CT with a phantom. Radiology 223:426–431, 2002CrossRefPubMed
22.
Zurück zum Zitat Tanikake M, Shimizu T, Narabayashi I, Matsuki M, Masuda K, Yamamoto K, Uesugi Y, Yoshikawa S: Three-dimensional CT angiography of the hepatic artery: use of multi-detector raw helical CT and a contrast agent. Radiology 227:883–889, 2003CrossRefPubMed Tanikake M, Shimizu T, Narabayashi I, Matsuki M, Masuda K, Yamamoto K, Uesugi Y, Yoshikawa S: Three-dimensional CT angiography of the hepatic artery: use of multi-detector raw helical CT and a contrast agent. Radiology 227:883–889, 2003CrossRefPubMed
23.
Zurück zum Zitat Uchida M, Ishibashi M, Abe T, Nishimura H, Hayabuchi N: Three-dimensional imaging of liver tumors using helical CT during intravenous injection of contrast medium. J Comput Assist Tomogr 23:435–440, 1999CrossRefPubMed Uchida M, Ishibashi M, Abe T, Nishimura H, Hayabuchi N: Three-dimensional imaging of liver tumors using helical CT during intravenous injection of contrast medium. J Comput Assist Tomogr 23:435–440, 1999CrossRefPubMed
Metadaten
Titel
Development of a Noise Reduction Filter Algorithm for Pediatric Body Images in Multidetector CT
verfasst von
Eiji Nishimaru
Katsuhiro Ichikawa
Izumi Okita
Yukihiro Tomoshige
Takehiro Kurokawa
Yuko Nakamura
Masayuki Suzuki
Publikationsdatum
01.12.2010
Verlag
Springer-Verlag
Erschienen in
Journal of Imaging Informatics in Medicine / Ausgabe 6/2010
Print ISSN: 2948-2925
Elektronische ISSN: 2948-2933
DOI
https://doi.org/10.1007/s10278-009-9218-4

Weitere Artikel der Ausgabe 6/2010

Journal of Digital Imaging 6/2010 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.