Magnetic tracking in the operation room using the da Vinci® telemanipulator is feasible

verfasst von: H. G. Kenngott, I. Wegner, J. Neuhaus, F. Nickel, L. Fischer, T. Gehrig, H. P. Meinzer, B. P. Müller-Stich

Erschienen in: Journal of Robotic Surgery | Ausgabe 1/2013

Abstract

In recent years, robotic assistance for surgical procedures has grown on a worldwide scale, particularly for use in more complex operations. Such operations usually require meticulous handling of tissue, involve a narrow working space and limit the surgeon’s sense of orientation in the human body. Improvement in both tissue handling and working within a narrow working space might be achieved through the use of robotic assistance. Soft tissue navigation might improve orientation by visualizing important target and risk structures intraoperatively, thereby possibly improving patient outcome. Prerequisites for navigation are its integration into the surgical workflow and accurate localization of both the instruments and patient. Magnetic tracking allows for good integration but is susceptible to distortion through metal or electro-magnetic interference, which may be caused by the operation table or a robotic system. We have investigated whether magnetic tracking can be used in combination with the da Vinci® (DV) telemanipulator in terms of stability and precision. We used a common magnetic tracking system (Aurora®, NDI Inc.) with the DV in a typical operation setup. Magnetic field distortion was evaluated using a measuring facility, with the following reference system: without any metal (R), operation table alone (T), DV in standby (D) and DV in motion (Dm). The maximum error of the entire tracking volume for R, T, D and Dm was 9.9, 32.8, 37.9 and 37.2 mm, respectively. Limiting the tracking volume to 190 mm (from cranial to caudal) resulted in a maximum error of 4.0, 8.3, 8.5 and 8.9 mm, respectively. When used in the operation room, magnetic tracking shows high errors, mainly due to the operation table. The target area should be limited to increase accuracy, which is possible for most surgical applications. The use of the da Vinci® telemanipulator only slightly aggravates the distortion and can thus be used in combination with magnetic tracking systems.

Introduction

The major benefits of laparoscopy for the patient are a reduced need for analgesics, reduced surgical trauma, improved cosmesis and faster recovery [1]. This accounts for the worldwide acceptance of laparoscopy in the clinical setting, with its applications having expanded from simple ablative procedures to more complex operations, such as prostatectomy, gastric bypass and liver and rectal surgery [25]. Under these circumstances, the shortcomings of conventional laparoscopy, especially the limited degrees of freedom, two-dimensional view, restricted ergonomics for the surgeon and the lack of the wrist gear, have become more evident [6, 7]. Minimally invasive robotic systems, such as the da Vinci® surgical system (DV) (Intuitive Surgical, Sunnyvale, CA), theoretically provide an ideal approach to address this challenge [6, 7]. However, orientation in a narrow space, the identification and attribution of structures and the estimation of the instrument’s position in relation to the position of risk and target structures remain difficult [8].
A navigation system, similar to those used in neuro- and maxillofacial surgery, might be a possible solution to these problems [1113]. However, navigation in soft tissue surgery is relatively more complex due to organ deformation and changing organ positions caused by tissue elasticity as well as cardiac and respiratory movements [9]. There is a need for updated information on the position of patient risk and target structures to build a navigation system that shows enough precision and accuracy for soft tissue surgery.
A tracking system is required to assess the spatial position of these structures. The most widely used tracking systems are magnetic and optic systems. While optic tracking systems provide high accuracy and precision, they require a direct line of sight from the sensor to the camera [10]. Their usability, high accuracy and precision in minimally invasive robotic surgery have already been proven [11]. Electromagnetic tracking systems are also highly accurate and precise, and they do not require a direct line of sight. However, they are susceptible to interference from ferromagnetic materials and electromagnetic fields [12]. Since DV effectors and instruments consist of metal, wire ropes and servo motors, its influence on magnetic tracking and its usability in the operating room are not known. The aim of the study reported here was to evaluate whether the DV can be used in combination with an electromagnetic navigation system to provide sufficient accuracy to visualize risk and target structures intraoperatively.

Materials and methods

To evaluate the effect of the DV and the operation table on the NDI Aurora® V1 (NDI Inc, Waterloo, Ontario, Canada) electromagnetic tracking system, here referred to as EMT, a metal-free measuring facility had to be built to minimize other possible sources of interference. In addition, the data acquisition setup was chosen to be as close to the real operation setup as possible to evaluate the usability of the EMT in a realistic environment.

Development of the measuring facility

A completely metal-free measuring facility was built to minimize interference with the EMT. This consisted of 12 polyvinyl chloride (PVC) rigid tubes that were arranged in such a way that the DV instruments and camera could be used in between them. The measuring facility walls were built out of medium-density fiberboard (diameter 22 mm). The PVC rigid tubes were 500 mm long and had a wall thickness and inner diameter of 3 and 5 mm, respectively. The pipes were arranged to cover the complete tracking volume of the EMT. The cubic tracking volume featuring a volume of 500 × 500 × 500 mm, according to the manufacturer of the EMT, was chosen due to known increased distortion in the periphery of the alternative dome volume. The pipes (500 mm) were arranged along the x-axis of the EMT volume (Fig. 1). A computed tomography (CT) (512 × 512 pixel, layer thickness 0.6 mm) was performed to generate a digital representation of the measuring facility.

Tracking system

The EMT was used with one field generator and one sensor. The field generator was a rectangle box that created an electromagnetic field, with a cube size tracking volume of 500 × 500 × 500 mm, according to the manufacturer. The sender was arranged on the right side of the operating table (Fig. 2). The EMT was used in combination with a 5DOF sensor (type Mednetix-5DK) embedded in a flexible cord. The outer diameter of the cord was adjusted to fit the inner diameter of the pipes using Leukosilk® tape.

Setup in the operating room

The DV was arranged in the operating room in the same way a typical operation setup is used for nissen fundoplication, with the DV over the head of the patient. All medical instruments and devices were positioned as if for a regular abdominal operation (Fig. 2) [13]. The operating table used was an Alphamaquet® 1150 (Maquet, Rastatt, Germany) system and consisted of several parts with ferromagnetic characteristics (Fig. 3).

Data acquisition and processing

Stable conditions were not expected during navigated surgery. Thus, the sensor was constantly pulled orthogonally to the field generator along the path forced by the PVC tube, simulating the instruments in use. The measurement of distorted positions was achieved by pulling the sensor from one end of the pipe to the other. In this way, the shift in the y- and z-axes of the tracking coordinate space was addressed. The shift in the x-axis was not captured in this setup. A shift was detected if the tracked position left the centerline of the corresponding pipe. The distance between the centerline and the tracked coordinate was used to describe distortion at each point along the x-axis. Two measurements for each tube and setup were acquired by slowly and continuously pulling the sensor through each pipe. We performed seven measurements: a reference measurement without the table, the DV and other metallic objects in close proximity (R), respectively, one with the measuring facility directly on the table (Tl), one with the measuring facility lifted 50 mm (T), one with the DV in active mode but not in motion (D), and one with the DV in motion (Dm) (Fig. 4). The DV measurements were only performed with a 50 mm-lifted measuring facility due to obvious high distortion when the measuring facility was not lifted.
For data processing the centerline of each pipe was interactively defined. This was accomplished by processing the CT scan DICOM file of the measuring facility and precisely defining sample points inside each pipe at five equivalent positions, using MITK (Medical Imaging Interaction Toolkit; DKFZ, German Cancer Research Center, Heidelberg, Germany), thereby creating four line segments. Registration between the CT coordinate and tracking coordinate space was done by rigid point-to-point registration. The middle of the tubes were taken for registration. The attribution of each measured track sample to the corresponding line segment was first determined for each x coordinate, and then the offset vector between the matched track sample and the matched line segment in orthogonal direction was calculated. If two measurements were found for the same x coordinate, the larger offset vector was used. To ensure precision, each evaluation setup was also assessed by measuring three fixed positions of each track: head, middle and feet. Orientations were not assessed.

Results

Reference measurements (R) showed low errors (Fig. 5). Accuracy particularly decreased at the periphery of the tracking volume, far from the field generator. In contrast, the tracking system showed low distortion at the center of the tracking volume (Fig. 6).
The results of the two operation table assessments (Tl, T) showed that the closer the field generator was to the table, the higher the amount of errors. These errors increased dramatically towards the periphery of the tracking volume (Figs. 7, 8).
The use of the DV (D, Dm) did not reveal any additional measurable errors, neither in static position nor during motion (Fig. 8).
For all measurements, variance declined from the periphery to the middle of the tracking volume for both the y- and z-axes, although to a lesser extent along the z-axis. Limiting the tracking volume to 190 mm in the craniocaudal direction led to a maximum error of 10 mm. The maximum errors of the entire tracking volume are shown in Table 1.
Table 1
Maximum errors recorded in the measurements
Tracking volumea
Measurementb
R
Tl
T
D
Dm
Full tracking volume [mm]
9.9
43.4
32.8
37.9
37.2
Limited tracking volume [mm]
4.0
9.9
8.3
8.5
8.9
R, Reference, no metal; Tl, operating table, low measuring facility; T, operating table, high measuring facility (50 mm and upwards); D, Da Vinci® in standby; Dm, Da Vinci® in motion
aLimitation of the tracking volume to ±95 mm on each side of the field generator
bMaximum error is given in millimeter

Discussion

In summary, this study has shown that an EMT can be reliably used with the DV within a real medical environment if the tracking volume is limited. Our results also demonstrate that the DV seems to have had hardly any effect on the accuracy of the magnetic tracking system. In contrast, metallic parts of the operating table distorted the positions delivered by the EMT to a large extent. The result from lifting the measurement setup by 50 mm shows that the EMT was more accurate in areas that were further away from ferromagnetic objects. The largest errors were seen towards the borders of the tracking volume, which can be explained by the decreasing strength of the electromagnetic field the further away it was from the electromagnetic field generator.
In general, detected errors have to be either eliminated or accurately detected in order to ensure precise navigation throughout the tracking volume. Since the amount of errors detected is still fairly high, our results indicate the need for an EMT calibration prior to intervention. This can be done by using the wooden measuring facility presented here as a reference; for example, by logging the coordinates over time and registering them with the virtual description of the wooden measuring facility (centerlines). By means of an iterative closest point algorithm, the best fit can be calculated in order to construct a lookup table for distorted positions and their corresponding correction. The distortion can also be decreased by repositioning the field generator (and the patient) to a position which is less affected by distortion due to ferromagnetic objects. However, this repositioning depends on the intervention performed.
The resulting amount of electromagnetic distortion caused by the operating table can also be decreased by choosing a table with a different material. An operation table constructed with carbon might have less effect on the electromagnetic field of the EMT. Furthermore, it should be noted that the results presented here can be only applied to the EMT tested. We did not evaluate a different Aurora tracking system or a different EMT, but according to theory, similar types of results are expected. NDI Inc. has developed a new type of electromagnetic field generator which resembles a board that can be installed right under the patient directly on top of the operation table. According to the manufacturer, this system shows less distortion due to ferromagnetic material of the operation table, but it has yet to be evaluated with the above-mentioned setup.
Previous work has already shown that navigation is feasible, such as for transhiatal minimally invasive esophagectomy [11]. The biggest challenge of these systems is that they do not provide an intraoperative update of the target position and risk structures. Particularly in esophagectomy, but also in rectal surgery, the iatrogenic manipulation causes a substantial shift of structures. Determining the correct height for the anastomosis in transhiatal esophagectomy or rectal resection requires much surgical experience. To achieve this, EMT sensors could be applied endoscopically in the esophagus and the rectum. Thus, combining the developed system, which is based on optical tracking, with an EMT will enable, for example, the intraoperative update of the position of a tumor in the esophagus or rectum, resulting in higher overall precision. This navigation system could help to find the correct position for anastomosis or affected lymph nodes in esophagectomy, or it could help to distinguish between scar and tumor for tumor recurrence of preoperated rectal cancer.
There are also limitations to this study. To compute the overall volume, we deemed accuracy to be less useful for the EMT because of the strong spatial dependence of its errors. Therefore, we computed the maximum errors which we believe to be clinically more important. Values such as the root mean square deviation have been calculated but are of less use since they involve the entire pipe, whereas here only the positional deviation at certain points of the x-axis was important.
To summarize, a magnetic tracking system can be used reliably in combination with the DV in a limited tracking volume and shows sufficient accuracy for navigation purposes.

Acknowledgments

This research project was carried out with the support of the Research Training Group 1126 “Soft-tissue Surgery: New Computer-based Methods for the Future Workplace”, funded by the German Research Foundation.

Conflict of interest

None.
Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License ( https://​creativecommons.​org/​licenses/​by/​2.​0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Urologie

Kombi-Abonnement

Mit e.Med Urologie erhalten Sie Zugang zu den urologischen CME-Fortbildungen und Premium-Inhalten der urologischen Fachzeitschriften.

Weitere Produktempfehlungen anzeigen
Literatur
2.
Zurück zum Zitat Sauerland S, Angrisani L, Belachew M, Chevallier JM, Favretti F, Finer N, Fingerhut A, Garcia Caballero M, Guisado Macias JA, Mittermair R, Morino M, Msika S, Rubino F, Tacchino R, Weiner R, Neugebauer EA (2005) Obesity surgery: evidence-based guidelines of the European association for endoscopic surgery (EAES). Surg Endosc 19(2):200–221. doi:10.1007/s00464-004-9194-1 PubMedCrossRef Sauerland S, Angrisani L, Belachew M, Chevallier JM, Favretti F, Finer N, Fingerhut A, Garcia Caballero M, Guisado Macias JA, Mittermair R, Morino M, Msika S, Rubino F, Tacchino R, Weiner R, Neugebauer EA (2005) Obesity surgery: evidence-based guidelines of the European association for endoscopic surgery (EAES). Surg Endosc 19(2):200–221. doi:10.​1007/​s00464-004-9194-1 PubMedCrossRef
3.
Zurück zum Zitat Gumbs AA, Tsai TJ, Hoffman JP (2012) Initial experience with laparoscopic hepatic resection at a comprehensive cancer center. Surg Endosc 26:480–487. doi:10.1007/s00464-011-1904-x Gumbs AA, Tsai TJ, Hoffman JP (2012) Initial experience with laparoscopic hepatic resection at a comprehensive cancer center. Surg Endosc 26:480–487. doi:10.​1007/​s00464-011-1904-x
4.
Zurück zum Zitat Row D, Weiser MR (2010) An update on laparoscopic resection for rectal cancer. Cancer Control 17(1):16–24PubMed Row D, Weiser MR (2010) An update on laparoscopic resection for rectal cancer. Cancer Control 17(1):16–24PubMed
5.
Zurück zum Zitat Coelho RF, Rocco B, Patel MB, Orvieto MA, Chauhan S, Ficarra V, Melegari S, Palmer KJ, Patel VR (2010) Retropubic, laparoscopic, and robot-assisted radical prostatectomy: a critical review of outcomes reported by high-volume centers. J Endourol 24(12):2003–2015. doi:10.1089/end.2010.0295 PubMedCrossRef Coelho RF, Rocco B, Patel MB, Orvieto MA, Chauhan S, Ficarra V, Melegari S, Palmer KJ, Patel VR (2010) Retropubic, laparoscopic, and robot-assisted radical prostatectomy: a critical review of outcomes reported by high-volume centers. J Endourol 24(12):2003–2015. doi:10.​1089/​end.​2010.​0295 PubMedCrossRef
7.
Zurück zum Zitat Maeso S, Reza M, Mayol JA, Blasco JA, Guerra M, Andradas E, Plana MN (2010) Efficacy of the Da Vinci surgical system in abdominal surgery compared with that of laparoscopy: a systematic review and meta-analysis. Ann Surg 252(2):254–262. doi:10.1097/SLA.0b013e3181e6239e PubMedCrossRef Maeso S, Reza M, Mayol JA, Blasco JA, Guerra M, Andradas E, Plana MN (2010) Efficacy of the Da Vinci surgical system in abdominal surgery compared with that of laparoscopy: a systematic review and meta-analysis. Ann Surg 252(2):254–262. doi:10.​1097/​SLA.​0b013e3181e6239e​ PubMedCrossRef
8.
10.
Zurück zum Zitat West JB, Maurer CR Jr (2004) Designing optically tracked instruments for image-guided surgery. IEEE Trans Med Imag 23(5):533–545CrossRef West JB, Maurer CR Jr (2004) Designing optically tracked instruments for image-guided surgery. IEEE Trans Med Imag 23(5):533–545CrossRef
11.
Zurück zum Zitat Kenngott HG, Neuhaus J, Muller-Stich BP, Wolf I, Vetter M, Meinzer HP, Koninger J, Buchler MW, Gutt CN (2008) Development of a navigation system for minimally invasive esophagectomy. Surg Endosc 22(8):1858–1865. doi:10.1007/s00464-007-9723-9 PubMedCrossRef Kenngott HG, Neuhaus J, Muller-Stich BP, Wolf I, Vetter M, Meinzer HP, Koninger J, Buchler MW, Gutt CN (2008) Development of a navigation system for minimally invasive esophagectomy. Surg Endosc 22(8):1858–1865. doi:10.​1007/​s00464-007-9723-9 PubMedCrossRef
12.
Zurück zum Zitat Frantz DD, Wiles AD, Leis SE, Kirsch SR (2003) Accuracy assessment protocols for electromagnetic tracking systems. Phys Med Biol 48:2241–2251PubMedCrossRef Frantz DD, Wiles AD, Leis SE, Kirsch SR (2003) Accuracy assessment protocols for electromagnetic tracking systems. Phys Med Biol 48:2241–2251PubMedCrossRef
13.
Zurück zum Zitat Muller-Stich BP, Reiter MA, Wente MN, Bintintan VV, Koninger J, Buchler MW, Gutt CN (2007) Robot-assisted versus conventional laparoscopic fundoplication: short-term outcome of a pilot randomized controlled trial. Surg Endosc 21(10):1800–1805. doi:10.1007/s00464-007-9268-y PubMedCrossRef Muller-Stich BP, Reiter MA, Wente MN, Bintintan VV, Koninger J, Buchler MW, Gutt CN (2007) Robot-assisted versus conventional laparoscopic fundoplication: short-term outcome of a pilot randomized controlled trial. Surg Endosc 21(10):1800–1805. doi:10.​1007/​s00464-007-9268-y PubMedCrossRef
Metadaten
Titel
Magnetic tracking in the operation room using the da Vinci® telemanipulator is feasible
verfasst von
H. G. Kenngott
I. Wegner
J. Neuhaus
F. Nickel
L. Fischer
T. Gehrig
H. P. Meinzer
B. P. Müller-Stich
Publikationsdatum
01.03.2013
Verlag
Springer-Verlag
Erschienen in
Journal of Robotic Surgery / Ausgabe 1/2013
Print ISSN: 1863-2483
Elektronische ISSN: 1863-2491
DOI
https://doi.org/10.1007/s11701-012-0347-2

Weitere Artikel der Ausgabe 1/2013

Journal of Robotic Surgery 1/2013 Zur Ausgabe

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Deutlich weniger Infektionen: Wundprotektoren schützen!

08.05.2024 Postoperative Wundinfektion Nachrichten

Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.

Chirurginnen und Chirurgen sind stark suizidgefährdet

07.05.2024 Suizid Nachrichten

Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.