A multi-center phantom study comparing image resolution from three state-of-the-art SPECT-CT systems

verfasst von: Tyler Hughes, BSc, Sergey Shcherbinin, PhD, Anna Celler, PhD

Erschienen in: Journal of Nuclear Cardiology | Ausgabe 6/2009

Einloggen, um Zugang zu erhalten

Abstract

Background

Ongoing advancements in SPECT-CT technology raise important questions regarding the differences in performance between various cameras and their respective image-processing algorithms. Our study seeks the answer to this question via measurement of phantom myocardial wall thickness (MWT) on images obtained from three state-of-the-art cameras.

Methods

A thorax phantom with an insert modeling a healthy heart was scanned using cardiac acquisition protocols with Philips’ Precedence (PP), GE’s Infinia Hawkeye (IH), and Siemens’ Symbia-T6 (SS). Processing was performed using advanced reconstruction techniques available on the cameras and our own independent software. The MWT measurement was used as a figure of merit in performance evaluation.

Results

When using 50% threshold, MWTs measured for the data acquired using PP, IH, and SS and reconstructed with independent standardized software were 8.5 ± 1.2 mm, 7.7 ± 1.2 mm, and 9.3 ± 0.9 mm, respectively; and 9.3 ± 0.5 mm, 19.2 ± 0.8 mm and 18.4 ± 1.1 mm when using the manufacturers’ own reconstructions, respectively. Thresholds optimized for each image (ranging from 32% to 68%) produced much more uniform results.

Conclusions

No significant differences were observed between image resolutions when data acquired from different cameras were reconstructed with an independent algorithm. However, different manufacturers’ reconstruction algorithms produced MWTs that differed by up to about 110% when using a set threshold of 50%.
Literatur
1.
Zurück zum Zitat Ficaro EP, Fessler JA, Shreve PD, et al. Simultaneous transmission/emission myocardial perfusion tomography: Diagnostic accuracy of attenuation-corrected 99mTc-sestamibi single-photon emission computed tomography. Circulation 1996;93:463-73.PubMed Ficaro EP, Fessler JA, Shreve PD, et al. Simultaneous transmission/emission myocardial perfusion tomography: Diagnostic accuracy of attenuation-corrected 99mTc-sestamibi single-photon emission computed tomography. Circulation 1996;93:463-73.PubMed
2.
Zurück zum Zitat Heller GV, Bateman TM, Johnson LL, et al. Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging. J Nucl Cardiol 2004;11:273-81.CrossRefPubMed Heller GV, Bateman TM, Johnson LL, et al. Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging. J Nucl Cardiol 2004;11:273-81.CrossRefPubMed
3.
Zurück zum Zitat Myers RW, Haseman MK, Durbin M, et al. Halftime acquisition of MPI with maintained image quality. J Nucl Cardiol 2006;13:S14-5.CrossRef Myers RW, Haseman MK, Durbin M, et al. Halftime acquisition of MPI with maintained image quality. J Nucl Cardiol 2006;13:S14-5.CrossRef
4.
Zurück zum Zitat Frey EC, Gilland KL, Tsui BM. Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Trans Med Imaging 2002;21:1040-50.CrossRefPubMed Frey EC, Gilland KL, Tsui BM. Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Trans Med Imaging 2002;21:1040-50.CrossRefPubMed
5.
Zurück zum Zitat Vija H, Hawman EG, Engdahl JC. Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: Phantom studies. In: 2003 IEEE nuclear science symposium, Medical imaging conference, 2003, Oct 19-26, Portland, USA, p. 2662-6. Vija H, Hawman EG, Engdahl JC. Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: Phantom studies. In: 2003 IEEE nuclear science symposium, Medical imaging conference, 2003, Oct 19-26, Portland, USA, p. 2662-6.
6.
Zurück zum Zitat Borges-Neto S, Pagnanelli R, Shaw K, et al. Clinical validation of a novel wide beam reconstruction method for shortening scan time of gated cardiac SPECT perfusion studies. J Nucl Cardiol 2007;14:555-65.CrossRefPubMed Borges-Neto S, Pagnanelli R, Shaw K, et al. Clinical validation of a novel wide beam reconstruction method for shortening scan time of gated cardiac SPECT perfusion studies. J Nucl Cardiol 2007;14:555-65.CrossRefPubMed
7.
Zurück zum Zitat Blinder S, Celler A, Wells RG, et al. Experimental verification of 3D detector response compensation using the OSEM reconstruction method. In: 2001 IEEE nuclear science symposium conference record, 2001, San Diego, USA, p. 2174-8. Blinder S, Celler A, Wells RG, et al. Experimental verification of 3D detector response compensation using the OSEM reconstruction method. In: 2001 IEEE nuclear science symposium conference record, 2001, San Diego, USA, p. 2174-8.
8.
Zurück zum Zitat Floyd CE, Jaszczak RJ, Harris CC, et al. Energy and spatial distribution of multiple order Compton scattering in SPECT: A Monte Carlo investigation. Phys Med Biol 1984;29:1217-30.CrossRefPubMed Floyd CE, Jaszczak RJ, Harris CC, et al. Energy and spatial distribution of multiple order Compton scattering in SPECT: A Monte Carlo investigation. Phys Med Biol 1984;29:1217-30.CrossRefPubMed
9.
Zurück zum Zitat Frey EC, Tsui MW. A new method for modeling the spatially-variant, object-dependent scatter response function in SPECT. In: Proceeding of 1995 nuclear science symposium, 1996, p. 1082-6. Frey EC, Tsui MW. A new method for modeling the spatially-variant, object-dependent scatter response function in SPECT. In: Proceeding of 1995 nuclear science symposium, 1996, p. 1082-6.
10.
Zurück zum Zitat Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: A simulation study. Phys Med Biol 1997;42:1619-32.CrossRefPubMed Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: A simulation study. Phys Med Biol 1997;42:1619-32.CrossRefPubMed
11.
Zurück zum Zitat Wells RG, Celler A, Harrop R. Analytical calculation of photon distributions of SPECT projections. IEEE Trans Nucl Sci 1998;45:3202-14.CrossRef Wells RG, Celler A, Harrop R. Analytical calculation of photon distributions of SPECT projections. IEEE Trans Nucl Sci 1998;45:3202-14.CrossRef
12.
Zurück zum Zitat Vandervoort E, Celler A, Wells G, et al. Implementation of an analytically based scatter correction in SPECT reconstructions. In: IEEE nuclear science symposium conference record, 2003, Oct 19-25, p. 2647-51. Vandervoort E, Celler A, Wells G, et al. Implementation of an analytically based scatter correction in SPECT reconstructions. In: IEEE nuclear science symposium conference record, 2003, Oct 19-25, p. 2647-51.
13.
Zurück zum Zitat O’Connor MK, Kemp B, Anstett F, et al. A multicenter evaluation of commercial attenuation compensation techniques in cardiac SPECT using phantom models. J Nucl Cardiol 2002;9:361-76.CrossRefPubMed O’Connor MK, Kemp B, Anstett F, et al. A multicenter evaluation of commercial attenuation compensation techniques in cardiac SPECT using phantom models. J Nucl Cardiol 2002;9:361-76.CrossRefPubMed
14.
Zurück zum Zitat Shcherbinin S, Celler A, Belhocine T, et al. Accuracy of quantitative reconstructions in SPECT/CT imaging. Phys Med Biol 2008;53:4595-604.CrossRefPubMed Shcherbinin S, Celler A, Belhocine T, et al. Accuracy of quantitative reconstructions in SPECT/CT imaging. Phys Med Biol 2008;53:4595-604.CrossRefPubMed
15.
Zurück zum Zitat Knoll P, Kotalova D, Zadrazil L, et al. Evaluation of advanced SPECT reconstruction technologies. Eur J Nucl Med Imaging 2008;35:S154-80.CrossRef Knoll P, Kotalova D, Zadrazil L, et al. Evaluation of advanced SPECT reconstruction technologies. Eur J Nucl Med Imaging 2008;35:S154-80.CrossRef
16.
Zurück zum Zitat National Electrical Manufacturers Association. NEMA standards publication NU 1-2007. Performance measurements of gamma cameras. National Electrical Manufacturers Association. NEMA standards publication NU 1-2007. Performance measurements of gamma cameras.
17.
Zurück zum Zitat Ye J, Song X, Zhao Z, et al. Iterative SPECT reconstruction using matched filtering for improved image quality. In: IEEE nuclear science symposium conference record, 2006, Oct 29-Nov 1, p. 2285-7. Ye J, Song X, Zhao Z, et al. Iterative SPECT reconstruction using matched filtering for improved image quality. In: IEEE nuclear science symposium conference record, 2006, Oct 29-Nov 1, p. 2285-7.
18.
19.
Zurück zum Zitat Dixon KL, Baldwin LN, Coquinco B, et al. A robust and versatile method for the quantification of myocardial perfusion defect size. In: IEEE nuclear science symposium and medical imaging conference record, 2003, Oct 19-23, Portland, USA, p. 2707-11. Dixon KL, Baldwin LN, Coquinco B, et al. A robust and versatile method for the quantification of myocardial perfusion defect size. In: IEEE nuclear science symposium and medical imaging conference record, 2003, Oct 19-23, Portland, USA, p. 2707-11.
20.
Zurück zum Zitat Germano G, Kavanagh PB, Waechter P, et al. A new algorithm for the quantitation of myocardial perfusion SPECT. I: Technical principles and reproducibility. J Nucl Med 2000;41:712-9.PubMed Germano G, Kavanagh PB, Waechter P, et al. A new algorithm for the quantitation of myocardial perfusion SPECT. I: Technical principles and reproducibility. J Nucl Med 2000;41:712-9.PubMed
21.
Zurück zum Zitat O’Connor M, Hammell T, Gibbons R. In vitro validation of a simple tomographic technique for estimation of percentage myocardium at risk using methoxyisobutyl isonitrile technetium 99m (sestamibi). Eur J Nucl Med 1990;17:69-76.CrossRefPubMed O’Connor M, Hammell T, Gibbons R. In vitro validation of a simple tomographic technique for estimation of percentage myocardium at risk using methoxyisobutyl isonitrile technetium 99m (sestamibi). Eur J Nucl Med 1990;17:69-76.CrossRefPubMed
22.
Zurück zum Zitat Wu M, Gao G, Sievers R, et al. Pinhole single-photon emission computed tomography for myocardial perfusion imaging of mice. J Am Coll Cardiol 2003;42:576-82.CrossRefPubMed Wu M, Gao G, Sievers R, et al. Pinhole single-photon emission computed tomography for myocardial perfusion imaging of mice. J Am Coll Cardiol 2003;42:576-82.CrossRefPubMed
Metadaten
Titel
A multi-center phantom study comparing image resolution from three state-of-the-art SPECT-CT systems
verfasst von
Tyler Hughes, BSc
Sergey Shcherbinin, PhD
Anna Celler, PhD
Publikationsdatum
01.12.2009
Verlag
Springer-Verlag
Erschienen in
Journal of Nuclear Cardiology / Ausgabe 6/2009
Print ISSN: 1071-3581
Elektronische ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-009-9132-7

Weitere Artikel der Ausgabe 6/2009

Journal of Nuclear Cardiology 6/2009 Zur Ausgabe

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.