The diagnostic accuracy of estimated continuous cardiac output compared with transthoracic echocardiography

verfasst von: Marc-Olivier Fischer, MD, Xavier Balaire, MD, Charles Le Mauff de Kergal, MD, Clément Boisselier, MD, Jean-Louis Gérard, MD, Jean-Luc Hanouz, MD, PhD, Jean-Luc Fellahi, MD, PhD

Erschienen in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Ausgabe 1/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Estimated continuous cardiac output (esCCO) is a new and noninvasive cardiac output (CO) monitoring device using pulse wave transit time. The aim of this study was to assess rapid changes in CO using esCCO (ΔCOesCCO) without invasive calibration and to compare the results with those using transthoracic Doppler echocardiography (ΔCOTTE).

Methods

Fifty-four consecutive patients were enrolled in this study following elective cardiac surgery. The COesCCO and COTTE were collected during four consecutive steps: 1) at baseline, 2) during passive leg raising (PLR), 3) at return to baseline, and 4) after a fluid challenge. The relationship between ΔCOesCCO and ΔCOTTE induced by PLR and a fluid challenge was assessed and a polar plot analysis was performed. Relationship, Bland-Altman analysis, and percentage error for absolute values of COesCCO and COTTE were also performed.

Results

Twenty-four patients were excluded from the analysis. No correlation was found between ΔCOesCCO and ΔCOTTE during PLR (r = 0.07; P = 0.732; n = 30) and after a fluid challenge (r = 0.24; P = 0.394; n = 14). The polar plot analysis showed that 21 data points (87%) of significant changes in CO were above the 30° radial sector lines and confirmed that esCCO was unable to track changes in CO. A weak positive relationship was found between absolute values of COesCCO and COTTE (r = 0.28; P = 0.004). Bias, precision, and limits of agreement were 0.25 L·min−1, 2.4 L·min−1, and −4.4 to 4.9 L·min−1, respectively. The percentage error was 80%.

Conclusions

Estimated continuous cardiac output without external calibration seems unable to assess rapid changes in CO following cardiac surgery and was not interchangeable with transthoracic Doppler echocardiography.
Literatur
1.
Zurück zum Zitat Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg 2011; 112: 1392-402.PubMedCrossRef Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg 2011; 112: 1392-402.PubMedCrossRef
2.
Zurück zum Zitat Maguire S, Rinehart J, Vakharia S, Cannesson M. Technical communication: respiratory variation in pulse pressure and plethysmographic waveforms: intraoperative applicability in a North American academic center. Anesth Analg 2011; 112: 94-6.PubMedCrossRef Maguire S, Rinehart J, Vakharia S, Cannesson M. Technical communication: respiratory variation in pulse pressure and plethysmographic waveforms: intraoperative applicability in a North American academic center. Anesth Analg 2011; 112: 94-6.PubMedCrossRef
4.
Zurück zum Zitat Evangelista A, Garcia-Dorado D, Garcia Del Castillo H, Gonzalez-Alujas T, Soler-Soler J. Cardiac index quantification by Doppler ultrasound in patients without left ventricular outflow tract abnormalities. J Am Coll Cardiol 1995; 25: 710-6.PubMedCrossRef Evangelista A, Garcia-Dorado D, Garcia Del Castillo H, Gonzalez-Alujas T, Soler-Soler J. Cardiac index quantification by Doppler ultrasound in patients without left ventricular outflow tract abnormalities. J Am Coll Cardiol 1995; 25: 710-6.PubMedCrossRef
5.
Zurück zum Zitat McLean AS, Needham A, Stewart D, Parkin R. Estimation of cardiac output by noninvasive echocardiographic techniques in the critically ill subject. Anaesth Intensive Care 1997; 25: 250-4.PubMed McLean AS, Needham A, Stewart D, Parkin R. Estimation of cardiac output by noninvasive echocardiographic techniques in the critically ill subject. Anaesth Intensive Care 1997; 25: 250-4.PubMed
6.
Zurück zum Zitat de Waal EE, Wappler F, Buhre WF. Cardiac output monitoring. Curr Opin Anaesthesiol 2009; 22: 71-7.PubMedCrossRef de Waal EE, Wappler F, Buhre WF. Cardiac output monitoring. Curr Opin Anaesthesiol 2009; 22: 71-7.PubMedCrossRef
7.
Zurück zum Zitat Ishihara H, Okawa H, Tanabe K, et al. A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors. J Clin Monit Comput 2004; 18: 313-20.PubMedCrossRef Ishihara H, Okawa H, Tanabe K, et al. A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors. J Clin Monit Comput 2004; 18: 313-20.PubMedCrossRef
8.
Zurück zum Zitat Yamada T, Tsutsui M, Sugo Y, et al. Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output. Anesth Analg 2012; 115: 82-7.PubMedCrossRef Yamada T, Tsutsui M, Sugo Y, et al. Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output. Anesth Analg 2012; 115: 82-7.PubMedCrossRef
9.
Zurück zum Zitat Bataille B, Bertuit M, Mora M, et al. Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth 2012; 109: 879-86.PubMedCrossRef Bataille B, Bertuit M, Mora M, et al. Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth 2012; 109: 879-86.PubMedCrossRef
10.
Zurück zum Zitat Bossuyt PM, Reitsma JB, Bruns DE, et al. The STARD statement for reporting studies of diagnostic accuracy: explanation and elaboration. Ann Intern Med 2003; 138: W1-12.PubMedCrossRef Bossuyt PM, Reitsma JB, Bruns DE, et al. The STARD statement for reporting studies of diagnostic accuracy: explanation and elaboration. Ann Intern Med 2003; 138: W1-12.PubMedCrossRef
11.
Zurück zum Zitat Lewis JF, Kuo LC, Nelson JG, Limacher MC, Quinones MA. Pulsed Doppler echocardiographic determination of stroke volume and cardiac output: clinical validation of two new methods using the apical window. Circulation 1984; 70: 425-31.PubMedCrossRef Lewis JF, Kuo LC, Nelson JG, Limacher MC, Quinones MA. Pulsed Doppler echocardiographic determination of stroke volume and cardiac output: clinical validation of two new methods using the apical window. Circulation 1984; 70: 425-31.PubMedCrossRef
12.
Zurück zum Zitat Monnet X, Rienzo M, Osman D, et al. Passive leg raising predicts fluid responsiveness in the critically ill. Crit Care Med 2006; 34: 1402-7.PubMedCrossRef Monnet X, Rienzo M, Osman D, et al. Passive leg raising predicts fluid responsiveness in the critically ill. Crit Care Med 2006; 34: 1402-7.PubMedCrossRef
13.
Zurück zum Zitat Preau S, Saulnier F, Dewavrin F, Durocher A, Chagnon JL. Passive leg raising is predictive of fluid responsiveness in spontaneously breathing patients with severe sepsis or acute pancreatitis. Crit Care Med 2010; 38: 819-25.PubMedCrossRef Preau S, Saulnier F, Dewavrin F, Durocher A, Chagnon JL. Passive leg raising is predictive of fluid responsiveness in spontaneously breathing patients with severe sepsis or acute pancreatitis. Crit Care Med 2010; 38: 819-25.PubMedCrossRef
14.
Zurück zum Zitat Cavallaro F, Sandroni C, Marano C, et al. Diagnostic accuracy of passive leg raising for prediction of fluid responsiveness in adults: systematic review and meta-analysis of clinical studies. Intensive Care Med 2010; 36: 1475-83.PubMedCrossRef Cavallaro F, Sandroni C, Marano C, et al. Diagnostic accuracy of passive leg raising for prediction of fluid responsiveness in adults: systematic review and meta-analysis of clinical studies. Intensive Care Med 2010; 36: 1475-83.PubMedCrossRef
15.
Zurück zum Zitat Monnet X, Teboul JL. Volume responsiveness. Current Opin Crit Care 2007; 13: 549-53.CrossRef Monnet X, Teboul JL. Volume responsiveness. Current Opin Crit Care 2007; 13: 549-53.CrossRef
16.
Zurück zum Zitat Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007; 17: 571-82.PubMedCrossRef Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007; 17: 571-82.PubMedCrossRef
17.
Zurück zum Zitat Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15: 85-91.PubMedCrossRef Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15: 85-91.PubMedCrossRef
18.
Zurück zum Zitat Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg 2010; 111: 1180-92.PubMedCrossRef Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg 2010; 111: 1180-92.PubMedCrossRef
19.
Zurück zum Zitat Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth 2011; 25: 536-46.PubMedCrossRef Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth 2011; 25: 536-46.PubMedCrossRef
20.
Zurück zum Zitat Bergenzaun L, Gudmundsson P, Ohlin H, et al. Assessing left ventricular systolic function in shock: evaluation of echocardiographic parameters in intensive care. Crit Care 2011; 15: R200.PubMedCrossRef Bergenzaun L, Gudmundsson P, Ohlin H, et al. Assessing left ventricular systolic function in shock: evaluation of echocardiographic parameters in intensive care. Crit Care 2011; 15: R200.PubMedCrossRef
21.
Zurück zum Zitat Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology 2010; 113: 1220-35.PubMedCrossRef Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology 2010; 113: 1220-35.PubMedCrossRef
22.
Zurück zum Zitat Kim SH, Song JG, Park JH, Kim JW, Park YS, Hwang GS. Beat-to-beat tracking of systolic blood pressure using noninvasive pulse transit time during anesthesia induction in hypertensive patients. Anesth Analg 2013; 116: 94-100.PubMedCrossRef Kim SH, Song JG, Park JH, Kim JW, Park YS, Hwang GS. Beat-to-beat tracking of systolic blood pressure using noninvasive pulse transit time during anesthesia induction in hypertensive patients. Anesth Analg 2013; 116: 94-100.PubMedCrossRef
23.
Zurück zum Zitat Le Manach Y, Hofer CK, Lehot JJ, et al. Can changes in arterial pressure be used to detect changes in cardiac output during volume expansion in the perioperative period? Anesthesiology 2012; 117: 1165-74.PubMedCrossRef Le Manach Y, Hofer CK, Lehot JJ, et al. Can changes in arterial pressure be used to detect changes in cardiac output during volume expansion in the perioperative period? Anesthesiology 2012; 117: 1165-74.PubMedCrossRef
24.
Zurück zum Zitat Ishihara H, Sugo Y, Tsutsui M, et al. The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm. J Clin Monit Comput 2012; 26: 465-71.PubMedCentralPubMedCrossRef Ishihara H, Sugo Y, Tsutsui M, et al. The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm. J Clin Monit Comput 2012; 26: 465-71.PubMedCentralPubMedCrossRef
Metadaten
Titel
The diagnostic accuracy of estimated continuous cardiac output compared with transthoracic echocardiography
verfasst von
Marc-Olivier Fischer, MD
Xavier Balaire, MD
Charles Le Mauff de Kergal, MD
Clément Boisselier, MD
Jean-Louis Gérard, MD
Jean-Luc Hanouz, MD, PhD
Jean-Luc Fellahi, MD, PhD
Publikationsdatum
01.01.2014
Verlag
Springer US
Erschienen in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Ausgabe 1/2014
Print ISSN: 0832-610X
Elektronische ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-013-0055-z

Weitere Artikel der Ausgabe 1/2014

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 1/2014 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.