Improved control over implant anchorage under the use of the femoral neck system for fixation of femoral neck fractures: a technical note

verfasst von: Yonghan Cha, Ji-Ung Song, Jun-Il Yoo, Ki Hoon Park, Jung-Taek Kim, Chan Ho Park, Won-Sik Choy

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2021

Abstract

Background

The depth of bolt in Femoral neck system (FNS, DePuy Synthes, Oberdorf, Switzerland) is difficult to finely control as the length of the bolt is in units of 5 mm. Thus, this study introduces a method to control the depth of FNS bolt in analogue scale in patients with femoral neck fracture.

Methods

By the technique of control of reaming and retraction of bolt, the tip of implant could be positioned close to subchondral bone without harming it. The position of implant tip in four cases in which the introduced technique was applied was compared to that of eight cases where the standard technique was performed.

Results

The average tip-apex distance measured in the cases that underwent surgery using the suggested technique in this study was statistically significantly shorter than that measured in the cases that underwent surgery under manufacturer guidelines.

Conclusion

Even though the bolt of FNS is manufactured in the unit of 5 mm, the technique proposed in this study helps surgeons to adjust the depth of bolt for the fixation of femoral neck fracture using FNS.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
DHS
Dynamic hip screw
FNS
Femoral Neck System
MCS
Multiple cannulated screws
TAD
Tip-apex distance

Background

The surgical treatment of femoral neck fractures can be classified into internal fixation and arthroplasty. In order to determine the surgical method, various factors such as the patient's age, bone density, fracture dislocation, and type of fracture must be considered [1]. Among them, internal fixation is preferred for undisplaced fracture or for relatively young patients, and either closed or open reduction can be performed [24]. Implants for internal fixation in femoral neck fracture include cannulated screws, dynamic hip screw (DHS) with or without antirotation screw, DHS with blade instead of screw or similar implants [57]. Parallel multiple cannulated screws (MCS) are commonly used in relatively young patients with femoral neck fractures, but they are associated with lower construct stiffness and earlier failure under cyclic loadings compared with DHS, as demonstrated in previous biomechanical tests [810]. Although DHS provides more stability to femoral neck fractures than MCS, it requires larger skin incision and more extensive soft tissue dissection [11].
The recently introduced implant, the Femoral Neck System (FNS, DePuy Synthes, Oberdorf, Switzerland) (Fig. 1), has both advantages of above two implants. It requires small incision like MCS and provides angular stability like DHS [9].
The FNS is composed of (1) a plate with a barrel and threaded screw holes which accommodates 1 or 2 locking screws, (2) an antirotation screw, and (3) a bolt that supports the head fragment. The proximal fragment with femoral head is held tight by the bolt and antirotation screw, thus it can slide through the axis of barrel to obtain controlled dynamic compression of the fracture site (Fig. 1).
Comparing an extremely robust metal to a fractured bone, acquisition of a longer moment arm provides more stability after implant fixation [9]. As penetration of implant out of femoral head surface jeopardizes the uninjured joint, surgeons walk a tightrope between violating the joint and gaining stability.
The depth of implant insertion of DHS and multiple cannulated screws is easy to adjust as both use screw mechanism. As the pitch of lag screw of DHS is 3.5 mm, half turn which is the minimum unit of adjustment the operator can make is 1.75 mm. Although the cannulated screws are generally manufactured in the unit of 5 mm, surgeon can control the depth of insertion in analogue scale and has more option with washers to adjust the depth. However, the depth of bolt in FNS is difficult to finely control as the length of the bolt is in units of 5 mm. The manufacturer recommended the subtraction of 5 mm from the measured depth read on the direct measuring device and choose the next shorter bolt size [12]. For example, if the measured depth was 102 mm in measuring device after insertion of tip of central guide wire into the subchondral bone, it is recommended to choose 95 mm bolt. According to this manufacturer`s guideline, the implant is positioned within 20 mm of the tip-apex distance (TAD) after surgery [13]. However, TAD is a criterion applied in fixation of DHS for intertrochanteric fractures to lower failure rates, and it has not been demonstrated whether it can be applied to femoral neck fracture using FNS. Also, it is a concern whether the method will ensure the insertion of bolt into sufficient depth and stable fixation of femoral neck fractures.
Thus, this study introduces a method to control the depth of FNS bolt in millimeter unit in patients with femoral neck fracture.

Technical note

The technique to be introduced in this study can be applied to all fixations of femoral neck fracture using FNS, and all the indications and contraindications of this technique follows those indicated on the surgical manual.

Operative technique

After reduction of the femoral neck fracture, the central guide wire is placed into the surgeon’s targeted position under the guidance of image intensifier (Fig. 2) (Table 1). After the depth of guide wire embedded in the subchondral bone is measured, the next longer construct size is selected. For example, should the depth guide indicate a measurement between 97 and 98 mm, then a 100 mm-length bolt would be considered (in comparison to the manufacturer’s guideline’s recommendation of a 90 mm-length bolt).
Table 1
Comparison between the surgical procedure presented by the manufacturer and the surgical procedure presented by the author
Manufacturer`s guideline
Author`s method
1. Central guide wire insertion
1. Central guide wire insertion
2. depth measuring: 98mm
2. depth measuring: 98mm
3. size of bolt: 90mm
3. size of bolt: 100mm
(one size longer than the measured depth)
4. reaming of 90mm reamer
4. 2 step reaming
first 90mm reamer
(same with manufacturer`s guideline)
second 100mm reamer to target depth under image
intensifier (same with selected bolt size)
5. plate, bolt, insertion handle assembled by black screw
5. plate, bolt, insertion handle assembled by black screw
6. insertion of bolt
6. insertion of bolt
7. some untightening of black screw and make contact between plate and femoral cortex
7. insertion of distal locking screw
8. insertion of distal locking screw
9. some retraction of bolt treated by retightening of black screw and tapping by impactor
8. insertion of antirotation screw
10. insertion of antirotation screw
9. fracture site compression
11. fracture site compression
10. removal of insertion handle
12. removal of insertion handle
The operator can control the depth of reaming by dividing the reaming procedure into two steps (as to the manufacturer`s guideline recommendation: single 90 mm reaming). Continuing the former example, the reamer of the first reaming step should be set at 90 mm. The first reaming is to countersink the lateral cortex of femur. By sliding the reamer-component over the drill bit into the remarked numbers, the limit of reamer tip advancement can be controlled by the remarked numbers. Assembling the reamer to set the depth of reaming for 10 mm less than the implant size, the surgeon can achieve countersinking for the barrel of the plate without penetration of articular surface.
To make room for a deeper insertion of the 100 mm-length bolt to the targeted position after the first reaming using a 90 mm-length bolt, the final advancement of reamer tip is adjusted finely on the second reaming step. After the reamer was re-assembled to indicating number on the reamer as the same number of the implant size (Continuing the former example, the reamer of the second reaming step should be set at 100 mm.), the reamer without power tool is pushed manually back into the pre-reamed hole to touch the blind end of reamed hole with reamer tip. The manual advancement and rotation in counterclockwise of the reamer without power tool ease the reinsertion of reamer without unintended harm of osteoporotic trabeculae. After reaching the end of blind pipe, the reamer is assembled to the power tool and image intensifier is used to estimate the remaining distance to reach the target. Under the guidance of image intensification, the reamer is advanced to the target carefully.
The plate, bolt, insertion handle, and insert are assembled by tightening of black screw as the surgical manual. After then, FNS is inserted. The cylinder impactor can be used to tap manually the plate on to the bone. When the tip of bolt reaches the target depth, plate does not touch the lateral surface of femoral cortex as the length of bolt is longer than the depth of reaming (Fig. 3). By untightening of the black screw, bolt is retracted backward (Fig. 4). After retraction of bolt, plate could contact the lateral surface of femoral cortex with tapping. After removal of central guide wire, distal locking screw is inserted, taking care not to insert eccentrically into the subtrochanteric area (Fig. 5). During this procedure, the bolt could be retracted more than intended. If black screw is re-tightened, the bolt protracts to reach the base of reamed bone (Fig. 6). After the surgeon confirms that the tip of bolt is positioned on the target and the plate is set on the lateral cortex with the image intensifier, the anti-rotation screw is inserted (Fig. 7). Inter-fragmentary compression is up to surgeon’s decision. Intraoperative radiographs of the case shown in Fig. 7B demonstrate clinical application of our technique (Fig. 8).
From September 2019 to September 2020, investigation for radiographs was conducted on 14 patients who were hospitalized for femoral neck fracture at our hospital and underwent surgery using FNS. All radiographs were retrieved from a picture archiving communication system of M-view (version 5.4.10.38, Infinitt Healthcare Co., Seoul, South Korea). All radiographs were calibrated before their evaluation using the known diameter of the bolt (10 mm). In both hip AP with 15˚ internal rotation view and translateral view of affected hip, TAD was measured for the tip of bolt [13].

Results

The average TAD measured in the cases that performed surgery with the manufacturer's surgical technique (10 patients) was 16.3 mm (range 14.6 to 18.1 mm), and the average TAD measured in the cases who performed surgery with the technique suggested in this study (4 patients) was 7.2 mm (range 6.1 to 12.2 mm). On the Mann–Whitney U test, TAD of the group that underwent surgery with the technique suggested in this study was significantly shorter than that of the group with the method of manufacturer’s guide (p < 0.05).

Discussion

Previous comparative biomechanical studies between FNS and other implants in cadaveric femoral neck fracture showed that the FNS can provide superior stability to the femur neck fractures [9, 14]. Stoffel et al. performed a biomechanical test after insertion of DHS, MCS and FNS in a cadaveric femoral neck fracture model with posteromedial bone defect [9]. The results of their study showed that FNS provided superior stability compared to MCS and had comparable stability to that of DHS. Schopper et al. performed a biomechanical test by inserting FNS and Hansson pins in the cadaveric model of Pauwels type II [14]. They reported that FNS was more resistant in varus deformation and less sensitive to variations in implant placement compared to Hansson pins. And, they insisted that FNS with diverging lag screw has superiority in femoral neck fracture in terms of less neck shortening. In both studies, the length of the bolt during FNS fixation was determined to make the TAD be within 20 mm which is originally from the evaluation method of DHS or proximal femoral nails in intertrochanteric fractures [13]. Surgery according to the surgical guide of manufacturer also makes the TAD within 20 mm. However, although FNS is morphologically similar to DHS, it does not seem to be proven whether it is reasonable to apply the same evaluation method for the position of FNS in femoral neck fractures which has a shorter length of proximal fragment than that of intertrochanteric fractures.
With MCS, the length as well as the diameter and position of the screw are important [15, 16]. Lindequist reported that the screws must be located within 3 mm of the femoral head cortex to achieve cortical support [16]. They recommended the insertion of screws with enough length to increase mechanical stability. Rau et al. showed that the insertion of the DHS screw into the subchondral region of the femoral head provided satisfactory clinical results of the femoral neck fractures [17]. They also reported that the depth of the DHS screw is an important factor related to the unsatisfactory results. Therefore, subchondral implant anchorage in femoral head by a longer implant plays an important role in stabilizing the femoral neck fractures. However, it is difficult to achieve an ideal depth of FNS bolt following the manufacturer's surgical guide alone. We believe the technique of this study can help to overcome the limitation of the implant manufactured in 5 mm unit and additionally contribute to increase fixation stability.
One concern of the technique is the disengagement between insertion handle and FNS while inserting the assembly of implant and instruments. Theoretically, the disengagement before the insertion of distal locking screw and antirotation screw can block right insertion of screws. However, as complete disengagement requires complete clearance of the insert from the plate and the insertion handle holds the plate tight, the implant cannot disengage from the insertion handle while using our technique. Also, the thread of black screw is on the track of the thread on the insertion handle even in a loose manner, the connecting part to the bolt of the insert block the plate slit form the disengagement (Figs. 9 and 10).
The limitation of this study is that it has not been proven that a difference in fixation of several millimeters causes a clinical difference in treatment of femoral neck fracture using FNS, and the number of cases is small. Also, there was a lack of consideration for other factors that could influence clinical results, such as femoral neck shortening. Further research on this is needed to delineate the clinical difference.

Conclusion

In conclusion, even though the bolt of FNS is manufactured in the unit of 5 mm, the technique proposed in this study helps surgeons to adjust the depth of bolt for the fixation of femoral neck fracture using FNS.

Declarations

We received Institutional Review Board approval (IRB No. EMC 2020–09-009) of Daejeon Eulji Medical Center. An explanation of the implant and surgical technique to be used for fixation of the femoral neck fracture was performed with a written informed consent from the patient and family before surgery. In addition, the implant, technique, and study design presented in this study were approved by IRB.
As the present study did not include any identifying images or other personal or clinical details of participants, the IRB waived the requirement for patients’ informed consent to publish.

Competing interests

The authors declare that they have no conflict of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
3.
Zurück zum Zitat Hossain S, Paton RW. The management of intracapsular fractures of the proximal femur. J Bone Joint Surg Brit. 2001;83(4):618.CrossRef Hossain S, Paton RW. The management of intracapsular fractures of the proximal femur. J Bone Joint Surg Brit. 2001;83(4):618.CrossRef
4.
Zurück zum Zitat Maiya S, Khan T, Grimer RJ, Carter SR, Tillman RM. The management of intracapsular fractures of the proximal femur. J Bone Joint Surg Brit. 2001;83(4):618–9.CrossRef Maiya S, Khan T, Grimer RJ, Carter SR, Tillman RM. The management of intracapsular fractures of the proximal femur. J Bone Joint Surg Brit. 2001;83(4):618–9.CrossRef
9.
Zurück zum Zitat Stoffel K, Zderic I, Gras F, Sommer C, Eberli U, Mueller D, Oswald M, Gueorguiev B. Biomechanical evaluation of the femoral neck system in unstable pauwels III femoral neck fractures: a comparison with the dynamic hip screw and cannulated screws. J Orthopaedic Trauma. 2017;31(3):131–7. https://doi.org/10.1097/BOT.0000000000000739.CrossRef Stoffel K, Zderic I, Gras F, Sommer C, Eberli U, Mueller D, Oswald M, Gueorguiev B. Biomechanical evaluation of the femoral neck system in unstable pauwels III femoral neck fractures: a comparison with the dynamic hip screw and cannulated screws. J Orthopaedic Trauma. 2017;31(3):131–7. https://​doi.​org/​10.​1097/​BOT.​0000000000000739​.CrossRef
14.
17.
Zurück zum Zitat Rau FD, Manoli A 2nd, Morawa LG. Treatment of femoral neck fractures with the sliding compression screw. Clin Orthopaed Related Res. 1982;163:137–140s. Rau FD, Manoli A 2nd, Morawa LG. Treatment of femoral neck fractures with the sliding compression screw. Clin Orthopaed Related Res. 1982;163:137–140s.
Metadaten
Titel
Improved control over implant anchorage under the use of the femoral neck system for fixation of femoral neck fractures: a technical note
verfasst von
Yonghan Cha
Ji-Ung Song
Jun-Il Yoo
Ki Hoon Park
Jung-Taek Kim
Chan Ho Park
Won-Sik Choy
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2021
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-021-04497-x

Weitere Artikel der Ausgabe 1/2021

BMC Musculoskeletal Disorders 1/2021 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.