Skip to main content
Erschienen in: Journal of Cardiovascular Magnetic Resonance 1/2017

Open Access 01.12.2017 | Research

Assessing exercise cardiac reserve using real-time cardiovascular magnetic resonance

verfasst von: Thu-Thao Le, Jennifer Ann Bryant, Alicia Er Ting, Pei Yi Ho, Boyang Su, Raymond Choon Chye Teo, Julian Siong-Jin Gan, Yiu-Cho Chung, Declan P. O’Regan, Stuart A. Cook, Calvin Woon-Loong Chin

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2017

Abstract

Background

Exercise cardiovascular magnetic resonance (ExCMR) has great potential for clinical use but its development has been limited by a lack of compatible equipment and robust real-time imaging techniques. We developed an exCMR protocol using an in-scanner cycle ergometer and assessed its performance in differentiating athletes from non-athletes.

Methods

Free-breathing real-time CMR (1.5T Aera, Siemens) was performed in 11 athletes (5 males; median age 29 [IQR: 28–39] years) and 16 age- and sex-matched healthy volunteers (7 males; median age 26 [interquartile range (IQR): 25–33] years). All participants underwent an in-scanner exercise protocol on a CMR compatible cycle ergometer (Lode BV, the Netherlands), with an initial workload of 25W followed by 25W-increment every minute. In 20 individuals, exercise capacity was also evaluated by cardiopulmonary exercise test (CPET). Scan-rescan reproducibility was assessed in 10 individuals, at least 7 days apart.

Results

The exCMR protocol demonstrated excellent scan-rescan (cardiac index (CI): 0.2 ± 0.5L/min/m2) and inter-observer (ventricular volumes: 1.2 ± 5.3mL) reproducibility. CI derived from exCMR and CPET had excellent correlation (r = 0.83, p < 0.001) and agreement (1.7 ± 1.8L/min/m2). Despite similar values at rest (P = 0.87), athletes had increased exercise CI compared to healthy individuals (at peak exercise: 12.2 [IQR: 10.2–13.5] L/min/m2 versus 8.9 [IQR: 7.5–10.1] L/min/m2, respectively; P < 0.001). Peak exercise CI, where image acquisition lasted 13–17 s, outperformed that at rest (c-statistics = 0.95 [95% confidence interval: 0.87–1.00] versus 0.48 [95% confidence interval: 0.23–0.72], respectively; P < 0.0001 for comparison) in differentiating athletes from healthy volunteers; and had similar performance as VO2max (c-statistics = 0.84 [95% confidence interval = 0.62–1.00]; P = 0.29 for comparison).

Conclusions

We have developed a novel in-scanner exCMR protocol using real-time CMR that is highly reproducible. It may now be developed for clinical use for physiological studies of the heart and circulation.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​s12968-017-0322-1) contains supplementary material, which is available to authorized users.
Abkürzungen
APMHR
Age-predicted maximal heart rate
CI
Cardiac Index
CMR
Cardiovascular magnetic resonance
CO
Cardiac output
CPET
Cardiopulmonary exercise test
EDV
End-diastolic volume
ESV
End-systolic volume
exCMR
Exercise CMR
LV
Left ventricular/left ventricle
SV
Stroke volume
TGRAPPA
Temporal generalized auto-calibrating partially parallel acquisition
VO2
Oxygen consumption
VO2max
Oxygen consumption at peak exercise

Background

Cardiac exercise testing is commonly used to detect underlying cardiovascular abnormalities that are not apparent at rest. Cardiovascular magnetic resonance (CMR) provides accurate assessment of cardiac volumes and function with excellent reproducibility compared to other standard imaging modalities [1, 2] but its application in stress testing has been limited to pharmacological agents.
Up until recently, the lack of suitable CMR-compatible exercise equipment and real-time imaging techniques have precluded accurate cardiac assessment of exercise physiology. Early studies in exercise CMR (exCMR) were performed using either breath-hold procedures that are not physiological or long free-breathing image acquisitions [35]. Although improvements in CMR technology have shortened the duration of exCMR imaging, these studies used a CMR-compatible treadmill [68]. The strengths of using treadmill exCMR are its validated diagnostic [9] and prognostic value and the ability to perform 12-lead electrocardiogram during exercise stress (the Duke score that also carries prognostic value) [10, 11]. However, a major limitation of performing treadmill exCMR is the obvious time delay needed to transfer the patient from the treadmill into the scanner.
In contrast, a supine cycle ergometer attached to the scan table will allow patients to exercise while in the bore. There are many advantages for exCMR particularly as images are acquired during the intermediate stages as well as at peak exercise, thus providing a large added value for statistical analyses of quantitative indices (e.g. stroke volume) and for repeated appreciation of qualitative changes (e.g. wall motion). However, excessive motion during exercise poses a challenge in image acquisition. Ungated real-time cine imaging and retrospective synchronisation of respiratory cycles have been used [5]. However, this translated to increased image acquisition time and complex image post-processing. We propose an approach of acquiring cine images at every stage of exercise during a brief period of exercise cessation to reduce artefacts from excessive motion and ECG-gating.
Using a CMR-compatible cycle ergometer and real-time CMR, we aimed to evaluate the feasibility and reproducibility of our exercise protocol; and to examine its potential to differentiate athletes from healthy volunteers.

Methods

Study population

A total of 11 athletes and 16 age- and sex-matched healthy volunteers were recruited to the study. Healthy volunteers did not have any diagnosed cardiac conditions and cardiovascular risk factors (hypertension, diabetes mellitus and hyperlipidemia). All the athletes competed in national/international events and trained more than 10 h a week in a variety of sports: triathlons (n = 7), long distance running (n = 1), rowing (n = 1), rugby (n = 1) and badminton (n = 1). We used the well-validated General Practice Physical Activity Questionnaire to assess physical activity levels in all participants, and classified them into four categories: inactive, moderately inactive, moderately active and active [12]. A five-point score was used to assess participant’s experience during exCMR (1 = would not do it again to 5 = highly satisfied; 3 = neutral).
The study was conducted in accordance with the Declaration of Helsinki and approved by the Singhealth Centralised Institutional Review Board. Written informed consent was obtained from all individuals.

Exercise CMR protocol

Exercise was performed using a programmable supine ergometer (Lode BV, Netherlands) fitted onto the CMR scanner table (1.5T MAGNETOM Aera, Siemens, Erlangen, Germany) (Figs. 1 and 2). Images were acquired using the 60-channel cardiac coils (30 anterior and 30 posterior elements).
After obtaining the baseline images, the participants were asked to cycle at an initial workload of 25W, with cadence maintained at least 70 rpm for 1 min. Workload was increased by 25W every minute until exhaustion. Free-breathing imaging was performed at the end of every stage during a brief period of stopping exercise. This is to avoid poor ECG signal and excessive motion artefacts that might result during exercise. Blood pressure and heart rate were recorded at every stage of exercise.
The imaging sequences and parameters were as follows:
  • Balanced steady-state free precession (bSSFP): standard long axis (vertical, horizontal long axis, sagittal LV outflow tract) and short axis cines (extending from the base to apex) were performed in all patients before initiating exercise (8mm thick and 2mm gap; TE 1.2 ms; TR 3 ms; 280–320 mm field of view; 13 segments per phase; acquired matrix size 205 × 256 pixels; acceleration factor of 2; acquired voxel size 1.6 × 1.3 × 8.0 mm; 30 phases per cardiac cycle).
  • Real-time bSSFP short axis cines: prospective ECG-gated free-breathing image acquisition of 10–13 short axis cine slices, extending from the base to apex was performed at each stage of the exercise (8 mm thick and 2 mm gap; TE 0.99 ms; TR 2.3 ms; 225 × 300 mm field of view; phase FOV 75%; acquired matrix size 68 × 128 pixels; phase resolution 71%; acceleration factor of 4; acquired voxel size 3.3 × 2.3 × 8.0 mm), temporal resolution 39.1 ms. Two different acquisition duration per slice were used in image acquisition: 1500 ms for heart rate less than 80 bpm (scan duration between 17 and 22 s); and 1200 ms for heart rate greater than 80 bpm (scan duration between 13 and 17 s). The rationale of using different acquisition durations was to obtain sufficient frames per cardiac cycle to capture end-diastole and end-systole at the different exercise stages. Using this approach, we were able to capture at least 2 cardiac cycles per exercise stage (Additional file 1: Video S1, Additional file 2: Video S2 and Additional files 3: Video S3).
    Additional file 1: Video S1.
    Breath-hold cine images at baseline. (MOV 857 KB)
    Additional file 2: Video S2.
    Real-time cine images at baseline. (MOV 1010 KB)
    Additional file 3: Video S3.
    Real-time cine images at peak exercise. (MOV 780 KB)

Image analysis

Left ventricular (LV) endocardial borders in short axis cine images, at end diastole and end systole, were manually contoured at baseline and at each exercise stage (CVI42, Circle Cardiovascular Imaging Inc., Canada). Stroke volume (SV) was measured as the difference between maximum volume (end-diastolic volume, EDV) and minimum volume (end-systolic volume, ESV) across the cardiac cycles (Fig. 3). Additional steps to manually select and compose the end-diastolic and end-systolic phases may be required if there was misalignment of phases due to ECG mis-triggering. Cardiac output (CO) was calculated as: CO = SV x HR. All measured volumes and cardiac output were indexed to body surface area (DuBois formula).

Reproducibility and validation

In 10 individuals, a repeat scan using the same exercise protocol was performed at least 7 days from the first scan to assess scan-rescan reproducibility.
A total of 20 individuals (healthy volunteers, n=10 and athletes, n=10) underwent additional cardiopulmonary exercise test (CPET), the gold standard of assessing exercise capacity. CPET was performed on an upright cycle ergometer (Vmax Encore E229C, USA) with a similar incremental protocol beginning at 50W and 25W increments every minute until exhaustion. Oxygen consumption at baseline (VO2) and at peak exercise (VO2max) were measured (Viasys Healthcare Cardiosoft, version 20). Cardiac output can be estimated from VO2 as: \( \mathrm{C}\mathrm{O}=\frac{{100\times \mathrm{V}\mathrm{O}}_2}{5.721+\left(0.1047\times \%{\mathrm{VO}}_{2 \max}\right)} \) [13].

Statistical analysis

All continuous variables were assessed for normal distribution and presented as mean±standard deviation or median [interquartile range (IQR)], as appropriate. Heart rate was normalized to the percent change between rest (0%) and peak heart rate (100%).
Pearson’s correlation and linear regression were used to assess the association between exercise capacities measured from exCMR and from CPET. Fixed and proportional biases with 95% limits of agreement between these two techniques were assessed using the Bland-Altman analysis. The performance of exCMR was assessed using the c statistics for discrimination [area under the receiver operating curve].
All statistical analyses were performed using GraphPad Prism 7 (GraphPad Software, Inc., San Diego, CA) and SPSS version 24 (IBM Corp., Armonk, NY). A 2-sided P <0.05 was considered statistically significant.

Results

Baseline characteristics

Athletes had larger cardiac volumes and increased LV mass compared to healthy individuals (P<0.05 for all comparisons). In athletes and healthy individuals, there were no sex-related differences in indexed SV and cardiac index (CI) (P>0.05 for all comparisons, Table 1). All athletes exercised more than 10 h a week and considered active based on the questionnaire. On average, healthy volunteers exercised between 1 and 3 h a week and were considered moderately active based on the questionnaire.
Table 1
Baseline characteristics and ventricular measurements
Parameters at baseline
Healthy volunteers (n = 16)
Athletes (n = 11)
P value
Clinical variables
 Age, years
26 [25–33]
29 [28–39]
0.121
 Males, n (%)
7 (44)
5 (45)
0.93
 Systolic blood pressure, mmHg
117 [107–123]
119 [107–127]
0.815
 Diastolic blood pressure, mmHg
67 [59–75]
68 [66–74]
0.446
 Heart rate, beats per minute
68 [59–74]
54 [48–61]
0.026
 Body surface area, m2
1.64 [1.51–1.96]
1.64 [1.50–1.97]
0.394
Cardiovascular variables
 LV mass, g
76 [65–97]
100 [93–122]
0.008
 LV end-diastolic volume, mL
141 [115–169]
169 [145–188]
0.121
 LV end-systolic volume, mL
54 [43–79]
79 [62–83]
0.05
 LV stroke volume, mL
88 [74–104]
101 [80–104]
0.422
 Cardiac output, L/min
5.8 [4.5–6.3]
5.6 [4.0–6.6]
0.610
 RV end-diastolic volume, mL
150 [121–197]
189 [154–208]
0.178
 RV end-systolic volume, mL
69 [45–99]
90 [74–100]
0.231
 Indexed LV mass, g/m2
47 [40–53]
66 [6–74]
<0.001
 Indexed LV end-diastolic volume, mL/m2
82 [75–94]
102 [96–110]
<0.001
 Indexed LV end-systolic volume, mL/m2
33 [28–40]
48 [43–49]
0.001
 Indexed LV stroke volume, mL/m2
52 [48–54]
56 [54–63]
0.003
 Cardiac index, L/min/m2
3.3 [2.8–3.9]
3.4 [2.6–4.1]
0.865
 Indexed RV end-diastolic volume, mL/m2
91 [81–107]
112 [106–122]
0.002
 Indexed RV end-systolic volume, mL/m2
40 [30–53]
56 [45–59]
0.023
Abbreviations: LV left ventricle, RV right ventricle

Exercise CMR protocol

All subjects exercised until they felt they could not continue. Median exercise duration was 7 [IQR: 7–10] minutes in healthy volunteers and 9 [IQR: 8–11] minutes in athletes. The duration of exercise cessation for image acquisition in between each stage ranged between 15 and 25 s, depending on the heart rate. As heart rate increased in the later stages of exercise, the cardiac cycle length became shorter and therefore, shorter acquisition times. The drop in heart rate from the end of exercise to the end of image acquisition was 9±6 bpm in healthy volunteers and 16±9 bpm in athletes. The exCMR was well-tolerated by all participants (median satisfaction score of 4 [IQR: 3–4]). At baseline, LV EDV and ESV measured from the baseline real-time and breath-hold cine images demonstrated excellent agreement (Indexed EDV: mean difference of 0.8±1.8 mL/m2; Indexed ESV: 0.4 ±1.9 mL/m2).
In the intermediate stages of exercise, there were notable differences between athletes and healthy volunteers. In athletes, LV EDV increased at about 75% peak heart rate and remained elevated at peak exercise despite increasing heart rate; whilst in healthy volunteers, LV EDV peaked earlier (about 50% peak heart rate) and decreased subsequently (Fig. 4). LV ESV decreased at every stage of exercise in athletes and healthy volunteers (Figs. 4 and 5). This accounted for the different SV profiles: indexed SV peaked at 75% peak heart rates in both athletes and healthy volunteers. Whilst indexed SV remained elevated at peak exercise in athletes, it decreased in healthy volunteers (Fig. 4). The chronotropic response in the later stages of exercise differed between athletes and less fit healthy volunteers: increased heart rate augment CI in athletes, but not in healthy volunteers (Fig. 4).
At peak exercise, heart rate increased by 238±39% and 264±32% in healthy volunteers and athletes, respectively. This corresponded to 83±6% and 78±7% of age-predicted maximal heart rate in healthy individuals and athletes, respectively. Despite similar CI at rest (3.3 [IQR:2.8–3.9] L/min/m2 versus 3.4 [IQR: 2.6–4.1] L/min/m2; P=0.87), athletes had increased CI compared to healthy volunteers at peak exercise (12.2 [IQR: 10.2–13.5] L/min/m2 versus 8.9 [IQR: 7.5–10.1] L/min/m2; P<0.001; Table 2). Similar to baseline values, there were no sex-related differences in CI at peak exercise in athletes and healthy volunteers (P>0.05 for all comparisons). Therefore, we combined both sexes in the subsequent analysis. Unlike at rest, CI at peak exercise demonstrated excellent ability in differentiating athletes from healthy volunteers (c-statistics=0.48 [95% confidence interval: 0.23–0.72] versus 0.95 [95% confidence interval: 0.87 to 1.00]; P<0.0001 for comparison).
Table 2
Peak exercise comparison between healthy volunteers and athletes
Parameters at peak exercise
Healthy volunteers (n = 16)
Athletes (n = 11)
P value
Clinical variables
 Heart rate, bpm
156 [150–165]
145 [135–158]
0.071
 Systolic blood pressure, mmHg
128 [103–192]
150 [120–165]
0.640
 Diastolic blood pressure, mmHg
90 [59–134]
68 [40–109]
0.379
 Maximum exercise power, W
125 [107–150]
200 [175–225]
0.011
Cardiovascular variables
 LV end-diastolic volume, mL
135 [104–152]
171 [150-208]
0.005
 LV end-systolic volume, mL
32 [26–47]
42 [36–46]
0.178
 LV stroke volume, mL
96 [73–121]
129 [115–158]
0.003
 Cardiac output, L/min
15 [12–19]
20 [16–23]
0.023
 Indexed LV end-diastolic volume, ml/m2
75 [67–90]
112 [96–122]
<0.001
 Indexed LV end-systolic volume, ml/m2
20 [16–24]
26 [22–28]
0.044
 Indexed stroke volume, ml/m2
57 [47–69]
85 [75–94]
<0.001
 Cardiac index, L/min/m2
8.9 [7.5–10.1]
12.2 [10.2–13.5]
<0.001
Abbreviations: LV left ventricle, RV right ventricle

Reproducibility of exCMR and comparison with CPET

The exCMR protocol demonstrated excellent scan-rescan reproducibility with no difference in CI between the two scans performed at least 7 days apart (0.2±0.5 L/min/m2, Fig. 6). Moreover, we observed excellent inter-observer variability in the assessment of cardiac volumes (LVEDV: 2.8±5.2 mL; LVESV: –0.5±5.2 mL).
Of the 20 individuals who underwent both exercise tests, VO2max was significantly higher in athletes compared to healthy individuals (50.7 [IQR: 39.4 to 56.9] mL/kg/min versus 29.8 [IQR: 28.3 to 34.0] mL/kg/min, respectively; P<0.001). Despite similar CI achieved (1.7±1.8 L/min/m2; Fig. 7), the maximal workload on the CMR supine ergometer was significantly lower compared to upright cycle CPET (Athletes: 200 [IQR: 175–225] W versus 237 [IQR: 213–300] W, P=0.007; healthy volunteers: 125 [IQR: 100–150] W versus 188 [IQR: 144–206] W, P=0.007). All participants exercised to the point they could not continue: the maximal exCMR workload was equivalent to the CPET workload at 80 [IQR: 70–86]% VO2max, much higher than the maximal workload defined at 60% VO2max in other studies [3, 5]. The CI from exCMR correlated very well with both CPET-derived CI (r=0.83, P<0.001; Fig. 7) and CPET VO2max (r=0.64, P=0.003). Compared to VO2max, exCMR-derived CI at peak exercise demonstrated similar ability to differentiate healthy volunteers from athletes (c-statistics=0.84 [95% confidence interval: 0.62–1.00] versus 0.95 [95% confidence interval: 0.87 to 1.00], respectively; P=0.292 for comparison).

Discussion

We have developed an exCMR protocol with excellent inter-observer and scan-rescan reproducibility. We have also observed excellent correlations and agreement in exercise capacity between exCMR and the gold standard CPET. Using the exercise protocol, we were able to characterise exercise physiology at every stage; and demonstrated excellent ability in differentiating athletes from healthy volunteers (c-statistics=0.95 [95% CI: 0.87 to 1.00]; P<0.0001).
ExCMR requires appropriate exercise equipment, rapid and robust real-time image acquisition to accommodate free-breathing exercise protocols. A previous study tested the feasibility of treadmill placed outside the MR scanner room and used breath-hold imaging protocol [14]. This took 60–90 s to complete post-exercise imaging because of the time needed to transfer patients from the treadmill to the MR scanner and image acquisition. Subsequent studies adopted a modified treadmill in the MR scanner room and free-breathing acquisition protocols, with some improvement in post-exercise imaging time [68, 15]. Although the acquisition time was faster compared to previous studies, imaging was only carried out at maximal exercise and not at every stage (which is a potential strength in supine bike protocols).
The recent use of in-scanner cycle ergometer has eliminated any delay in transferring patients into the scanner after exercise, but excessive motion poses a challenge in image acquisition during exercise [5]. Ungated real-time cine imaging and retrospective synchronisation of respiratory cycles have been used to reduce motion, at the expense of increased image acquisition time and complex image post-processing [5]. Although phase-contrast imaging can reduce image acquisition time, it is not able to assess wall motion abnormalities in myocardial ischemia [4, 16, 17].
We demonstrated rapid acquisition of free-breathing peak-exercise cine images within 13 to 17 s and superior spatiotemporal resolution (spatial resolution: 3.3 × 2.3 mm; temporal resolution < 40ms). In our exercise protocol, we used two different acquisition duration (1500ms and 1200ms per slice for heart rates less than and more than 80, respectively) to ensure at least 2 cardiac cycles that would adequately identify the end diastolic and systolic phases. Importantly, the faster acquisition did not compromise image quality. Moreover, the increased number of coil elements may reduce artefacts that may affect the accurate assessment of cardiac volumes and function; and wall motion abnormalities. Indeed, we were able to achieve excellent scan-rescan and inter-observer variability; and did not observe any difference in ventricular volumes when compared using standard breath-hold and our free-breathing protocols.
CPET is widely accepted as the most reliable and objective test for assessing cardiac reserve in a variety of cardiac conditions, such as heart failure and distinguishing physiologic from pathologic left ventricular hypertrophy [18, 19]. We used similar exercise protocols for both the in-scanner cycle and upright cycle in CPET. All participants were given clear instructions to exercise until they were not able to continue anymore to ensure maximal exercise capacity was achieved. Of note, the maximum workload on the supine cycle ergometer was lower compared to upright cycle CPET. It is perhaps not surprising that the maximal workload would vary across different exercise stress modalities because of different cardiac responses. Despite different workload attained, supine and upright cycling demonstrated similar VO2max [20, 21], but lower than treadmill CPET [11, 22]. Moreover, we observed excellent correlation and agreement in cardiac index between supine exCMR and upright cycle CPET, supporting the validity of the study.
During exCMR, we observed differences in exercise physiology between athletes and healthy individuals. The rate of increase of heart rate in response to exercise was similar in both groups (Fig. 5). Whilst athletes had augmented diastolic filling (increased LV EDV) and improved contractility (decreased LV ESV) throughout the range of exercise intensity tested, healthy volunteers attained peak diastolic filling and contractility earlier (at about 50% peak heart rate). At higher exercise intensity, increased chronotropic response further augment cardiac output in athletes but not in the less fit healthy volunteers. In healthy volunteers, the rapid heart rates reduced diastolic filling and therefore, stroke volumes at peak exercise. There are some uncertainties in the mechanisms associated with cardiac output augmentation in athletes and non-athletes [2327]. Our study adds novelty by demonstrating mechanistic differences in exercise profiles between athletes and healthy volunteers using the same exercise protocol.

Clinical implications

The study highlighted the potential of further extending the clinical applications of CMR to assess cardiac reserve. In addition to CMR being the gold standard for assessing left ventricular mass and cardiac volumes, it is the only imaging modality that can detect myocardial fibrosis non-invasively. The combination of these techniques in a single imaging modality offers valuable diagnostic insights in individuals with cardiac pathologies, which not achievable with CPET. The ability to assess cardiac physiology and function at every stage of exercise provides a unique opportunity to characterise and differentiate the exercise profiles between individuals. In this study, we have demonstrated one such potential application: peak exercise CI significantly outperformed CI at rest and had similar performance as VO2max in differentiating athletes from healthy volunteers. This technique holds promise in distinguishing physiologic from pathologic myocardial biology in patients with LV dilatation, hypertrophy or mildly impaired systolic ejection fraction [28, 29]. The excellent reproducibility adds strength to use exCMR for serial assessments.

Limitations

The CMR protocol required a brief period of stopping exercise at the end of every stage in order to minimise excessive motion and ECG artefacts during image acquisition. This resulted in a small drop in heart rate (9±6 bpm in controls and 16±9 bpm in athletes). However, the effect may be less significant in patients with cardiac pathologies or individuals who are older and less physically-active. The maximal heart rate achieved using this exCMR protocol was less than the recommended 85% age-predicted maximal heart rate (APMHR) commonly used in defining an adequate treadmill stress test [11, 30]. This may affect the diagnostic accuracy in assessing myocardial ischemia. However, it is well described that heart rates achieved with supine cycling is lower than exercise treadmill [22, 31, 32]; and previous echo studies have demonstrated similar diagnostic accuracies between supine and treadmill exercise stress, despite a large proportion of patients not achieving the 85% APMHR [3133]. It is conceivable that supine exercise stress may not be suitable in some individuals (particularly those who are not accustomed to cycling) because of the postural effects on musculoskeletal fatigue.

Conclusions

These data demonstrate the feasibility, accuracy and reproducibility of an in-scanner exercise protocol using a CMR-compatible cycle ergometer. Future studies will examine its clinical utility in a variety of cardiac pathologies.

Acknowledgements

The authors would like to thank the physiotherapists of the Changi Sports Medicine Centre, Changi General Hospital and the radiographers of the National Heart Centre Singapore for conducting the tests and Mr Chan Cheow Hiong and Mrs Agnes Low for their financial donation to the study.

Funding

Duke-NUS Goh Cardiovascular Research Award, National Medical Research Council Transition Award (NMRC/TA/0034/2015), Philanthropy Fund (Mr Chan Cheow Hiong and Mrs Agnes Low).

Availability of data and materials

The datasets during and/or analysed during the current study available from the corresponding author on reasonable request.

Authors’ contributions

TTL conceived the study design, analysed the images, interpreted the data and drafted the manuscript. JAB optimized imaging sequence, acquired the scans and analysed the images and contributed to the intellectual content of the manuscript. AET acquired the scans for all the patients in the study, involved in the acquisition of data and contributed to the intellectual content of the manuscript. PYH involved in the acquisition of data and contributed to the intellectual content of the manuscript. BS analysed the data, developed in-house add-in program for analysis and contributed to the intellectual content of the manuscript. RCCT involved in the design of cardiopulmonary test, acquired the data and contributed to the intellectual content of the manuscript. JSJG optimized imaging sequence and contributed to the intellectual content of the manuscript. YCC optimized imaging sequence and contributed to the intellectual content of the manuscript. DOR interpreted the data and involved in the critical revision of the manuscript. SC conceived the study design, interpreted the data and involved in the critical revision of the manuscript. CCWL conceived the study design, interpreted the data and drafted the manuscript. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.
Not applicable.
written informed consent was obtained from all participants. The study protocol was approved by Singhealth Centralised Institutional Review Board, reference number 2016/2256 and 2016/2050.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Sechtem U, Pflugfelder PW, Gould RG, Cassidy MM, Higgins CB. Measurement of right and left ventricular volumes in healthy individuals with cine MR imaging. Radiology. 1987;163(3):697–702.CrossRefPubMed Sechtem U, Pflugfelder PW, Gould RG, Cassidy MM, Higgins CB. Measurement of right and left ventricular volumes in healthy individuals with cine MR imaging. Radiology. 1987;163(3):697–702.CrossRefPubMed
2.
Zurück zum Zitat Grothues F, Smith GC, Moon JC, Bellenger NG, Collins P, Klein HU, Pennell DJ. Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol. 2002;90(1):29–34.CrossRefPubMed Grothues F, Smith GC, Moon JC, Bellenger NG, Collins P, Klein HU, Pennell DJ. Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol. 2002;90(1):29–34.CrossRefPubMed
3.
Zurück zum Zitat Roest AA, Kunz P, Lamb HJ, Helbing WA, van der Wall EE, de Roos A. Biventricular response to supine physical exercise in young adults assessed with ultrafast magnetic resonance imaging. Am J Cardiol. 2001;87(5):601–5.CrossRefPubMed Roest AA, Kunz P, Lamb HJ, Helbing WA, van der Wall EE, de Roos A. Biventricular response to supine physical exercise in young adults assessed with ultrafast magnetic resonance imaging. Am J Cardiol. 2001;87(5):601–5.CrossRefPubMed
4.
Zurück zum Zitat Pieles GE, Szantho G, Rodrigues JC, Lawton CB, Stuart AG, Bucciarelli-Ducci C, Turner MS, Williams CA, Tulloh RM, Hamilton MC. Adaptations of aortic and pulmonary artery flow parameters measured by phase-contrast magnetic resonance angiography during supine aerobic exercise. Eur J Appl Physiol. 2014;114(5):1013–23.CrossRefPubMed Pieles GE, Szantho G, Rodrigues JC, Lawton CB, Stuart AG, Bucciarelli-Ducci C, Turner MS, Williams CA, Tulloh RM, Hamilton MC. Adaptations of aortic and pulmonary artery flow parameters measured by phase-contrast magnetic resonance angiography during supine aerobic exercise. Eur J Appl Physiol. 2014;114(5):1013–23.CrossRefPubMed
5.
Zurück zum Zitat La Gerche A, Claessen G, Van de Bruaene A, Pattyn N, Van Cleemput J, Gewillig M, Bogaert J, Dymarkowski S, Claus P, Heidbuchel H. Cardiac MRI: a new gold standard for ventricular volume quantification during high-intensity exercise. Circ Cardiovasc Imaging. 2013;6(2):329–38.CrossRefPubMed La Gerche A, Claessen G, Van de Bruaene A, Pattyn N, Van Cleemput J, Gewillig M, Bogaert J, Dymarkowski S, Claus P, Heidbuchel H. Cardiac MRI: a new gold standard for ventricular volume quantification during high-intensity exercise. Circ Cardiovasc Imaging. 2013;6(2):329–38.CrossRefPubMed
6.
Zurück zum Zitat Jekic M, Foster EL, Ballinger MR, Raman SV, Simonetti OP. Cardiac function and myocardial perfusion immediately following maximal treadmill exercise inside the MRI room. J Cardiovasc Magn Reson. 2008;10. Jekic M, Foster EL, Ballinger MR, Raman SV, Simonetti OP. Cardiac function and myocardial perfusion immediately following maximal treadmill exercise inside the MRI room. J Cardiovasc Magn Reson. 2008;10.
7.
Zurück zum Zitat Raman SV, Dickerson JA, Jekic M, Foster EL, Pennell ML, McCarthy B, Simonetti OP. Real-time cine and myocardial perfusion with treadmill exercise stress cardiovascular magnetic resonance in patients referred for stress SPECT. J Cardiovasc Magn Reson. 2010;12(1):1–9.CrossRef Raman SV, Dickerson JA, Jekic M, Foster EL, Pennell ML, McCarthy B, Simonetti OP. Real-time cine and myocardial perfusion with treadmill exercise stress cardiovascular magnetic resonance in patients referred for stress SPECT. J Cardiovasc Magn Reson. 2010;12(1):1–9.CrossRef
8.
Zurück zum Zitat Thavendiranathan P, Dickerson JA, Scandling D, Balasubramanian V, Pennell ML, Hinton A, Raman SV, Simonetti OP. Comparison of treadmill exercise stress cardiac MRI to stress echocardiography in healthy volunteers for adequacy of left ventricular endocardial wall visualization: A pilot study. J Magn Reson Imaging. 2014;39(5):1146–52.CrossRefPubMed Thavendiranathan P, Dickerson JA, Scandling D, Balasubramanian V, Pennell ML, Hinton A, Raman SV, Simonetti OP. Comparison of treadmill exercise stress cardiac MRI to stress echocardiography in healthy volunteers for adequacy of left ventricular endocardial wall visualization: A pilot study. J Magn Reson Imaging. 2014;39(5):1146–52.CrossRefPubMed
9.
Zurück zum Zitat Raman SV, Dickerson JA, Mazur W, Wong TC, E BS, Min JK, Scandling D, Bartone C, Craft JT, Thavendiranathan P, et al. Diagnostic Performance of Treadmill Exercise Cardiac Magnetic Resonance: The Prospective, Multicenter Exercise CMR’s Accuracy for Cardiovascular Stress Testing (EXACT) Trial. J Am Heart Assoc. 2016;5(8):e003811. Raman SV, Dickerson JA, Mazur W, Wong TC, E BS, Min JK, Scandling D, Bartone C, Craft JT, Thavendiranathan P, et al. Diagnostic Performance of Treadmill Exercise Cardiac Magnetic Resonance: The Prospective, Multicenter Exercise CMR’s Accuracy for Cardiovascular Stress Testing (EXACT) Trial. J Am Heart Assoc. 2016;5(8):e003811.
10.
Zurück zum Zitat Lauer MS, Francis GS, Okin PM, Pashkow FJ, Snader CE, Marwick TH. Impaired chronotropic response to exercise stress testing as a predictor of mortality. JAMA. 1999;281(6):524–9.CrossRefPubMed Lauer MS, Francis GS, Okin PM, Pashkow FJ, Snader CE, Marwick TH. Impaired chronotropic response to exercise stress testing as a predictor of mortality. JAMA. 1999;281(6):524–9.CrossRefPubMed
11.
Zurück zum Zitat Fletcher GF, Balady GJ, Amsterdam EA, Chaitman B, Eckel R, Fleg J, Froelicher VF, Leon AS, Pina IL, Rodney R, et al. Exercise standards for testing and training: a statement for healthcare professionals from the American Heart Association. Circulation. 2001;104(14):1694–740.CrossRefPubMed Fletcher GF, Balady GJ, Amsterdam EA, Chaitman B, Eckel R, Fleg J, Froelicher VF, Leon AS, Pina IL, Rodney R, et al. Exercise standards for testing and training: a statement for healthcare professionals from the American Heart Association. Circulation. 2001;104(14):1694–740.CrossRefPubMed
12.
Zurück zum Zitat Khaw KT, Jakes R, Bingham S, Welch A, Luben R, Day N, Wareham N. Work and leisure time physical activity assessed using a simple, pragmatic, validated questionnaire and incident cardiovascular disease and all-cause mortality in men and women: The European Prospective Investigation into Cancer in Norfolk prospective population study. Int J Epidemiol. 2006;35(4):1034–43.CrossRefPubMed Khaw KT, Jakes R, Bingham S, Welch A, Luben R, Day N, Wareham N. Work and leisure time physical activity assessed using a simple, pragmatic, validated questionnaire and incident cardiovascular disease and all-cause mortality in men and women: The European Prospective Investigation into Cancer in Norfolk prospective population study. Int J Epidemiol. 2006;35(4):1034–43.CrossRefPubMed
13.
Zurück zum Zitat Stringer WW, Hansen JE, Wasserman K. Cardiac output estimated noninvasively from oxygen uptake during exercise. J Appl Physiol. 1997;82(3):908–12.PubMed Stringer WW, Hansen JE, Wasserman K. Cardiac output estimated noninvasively from oxygen uptake during exercise. J Appl Physiol. 1997;82(3):908–12.PubMed
14.
Zurück zum Zitat Rerkpattanapipat P, Gandhi SK, Darty SN, Williams RT, Davis AD, Mazur W, Clark HP, Little WC, Link KM, Hamilton CA, et al. Feasibility to detect severe coronary artery stenoses with upright treadmill exercise magnetic resonance imaging. Am J Cardiol. 2003;92. Rerkpattanapipat P, Gandhi SK, Darty SN, Williams RT, Davis AD, Mazur W, Clark HP, Little WC, Link KM, Hamilton CA, et al. Feasibility to detect severe coronary artery stenoses with upright treadmill exercise magnetic resonance imaging. Am J Cardiol. 2003;92.
15.
Zurück zum Zitat Lafountain RA, da Silveira JS, Varghese J, Mihai G, Scandling D, Craft J, Swain CB, Franco V, Raman SV, Devor ST, et al. Cardiopulmonary exercise testing in the MRI environment. Physiol Meas. 2016;37(4):N11–25.CrossRefPubMed Lafountain RA, da Silveira JS, Varghese J, Mihai G, Scandling D, Craft J, Swain CB, Franco V, Raman SV, Devor ST, et al. Cardiopulmonary exercise testing in the MRI environment. Physiol Meas. 2016;37(4):N11–25.CrossRefPubMed
16.
Zurück zum Zitat Barber NJ, Ako EO, Kowalik GT, Steeden JA, Pandya B, Muthurangu V. MR augmented cardiopulmonary exercise testing-a novel approach to assessing cardiovascular function. Physiol Meas. 2015;36(5):N85–94.CrossRefPubMed Barber NJ, Ako EO, Kowalik GT, Steeden JA, Pandya B, Muthurangu V. MR augmented cardiopulmonary exercise testing-a novel approach to assessing cardiovascular function. Physiol Meas. 2015;36(5):N85–94.CrossRefPubMed
17.
Zurück zum Zitat Mohiaddin RH, Gatehouse PD, Firmin DN. Exercise-related changes in aortic flow measured with spiral echo-planar MR velocity mapping. J Magn Reson Imaging. 1995;5(2):159–63.CrossRefPubMed Mohiaddin RH, Gatehouse PD, Firmin DN. Exercise-related changes in aortic flow measured with spiral echo-planar MR velocity mapping. J Magn Reson Imaging. 1995;5(2):159–63.CrossRefPubMed
18.
Zurück zum Zitat Guazzi M, Arena R, Halle M, Piepoli MF, Myers J, Lavie CJ. 2016 Focused Update: Clinical Recommendations for Cardiopulmonary Exercise Testing Data Assessment in Specific Patient Populations. Circulation. 2016;133(24):e694–711.CrossRefPubMed Guazzi M, Arena R, Halle M, Piepoli MF, Myers J, Lavie CJ. 2016 Focused Update: Clinical Recommendations for Cardiopulmonary Exercise Testing Data Assessment in Specific Patient Populations. Circulation. 2016;133(24):e694–711.CrossRefPubMed
19.
Zurück zum Zitat Sharma S, Elliott PM, Whyte G, Mahon N, Virdee MS, Mist B, McKenna WJ. Utility of metabolic exercise testing in distinguishing hypertrophic cardiomyopathy from physiologic left ventricular hypertrophy in athletes. J Am Coll Cardiol. 2000;36(3):864–70.CrossRefPubMed Sharma S, Elliott PM, Whyte G, Mahon N, Virdee MS, Mist B, McKenna WJ. Utility of metabolic exercise testing in distinguishing hypertrophic cardiomyopathy from physiologic left ventricular hypertrophy in athletes. J Am Coll Cardiol. 2000;36(3):864–70.CrossRefPubMed
20.
Zurück zum Zitat Leyk D, Essfeld D, Hoffmann U, Wunderlich HG, Baum K, Stegemann J. Postural effect on cardiac output, oxygen uptake and lactate during cycle exercise of varying intensity. Eur J Appl Physiol Occup Physiol. 1994;68(1):30–5.CrossRefPubMed Leyk D, Essfeld D, Hoffmann U, Wunderlich HG, Baum K, Stegemann J. Postural effect on cardiac output, oxygen uptake and lactate during cycle exercise of varying intensity. Eur J Appl Physiol Occup Physiol. 1994;68(1):30–5.CrossRefPubMed
21.
Zurück zum Zitat Quinn TJ, Smith SW, Vroman NB, Kertzer R, Olney WB. Physiologic responses of cardiac patients to supine, recumbent, and upright cycle ergometry. Arch Phys Med Rehabil. 1995;76(3):257–61.CrossRefPubMed Quinn TJ, Smith SW, Vroman NB, Kertzer R, Olney WB. Physiologic responses of cardiac patients to supine, recumbent, and upright cycle ergometry. Arch Phys Med Rehabil. 1995;76(3):257–61.CrossRefPubMed
22.
Zurück zum Zitat Proctor DN, Sinning WE, Bredle DL, Joyner MJ. Cardiovascular and peak VO2 responses to supine exercise: effects of age and training status. Med Sci Sports Exerc. 1996;28(7):892–9.CrossRefPubMed Proctor DN, Sinning WE, Bredle DL, Joyner MJ. Cardiovascular and peak VO2 responses to supine exercise: effects of age and training status. Med Sci Sports Exerc. 1996;28(7):892–9.CrossRefPubMed
23.
Zurück zum Zitat Holverda S, Gan CT, Marcus JT, Postmus PE, Boonstra A, Vonk-Noordegraaf A. Impaired stroke volume response to exercise in pulmonary arterial hypertension. J Am Coll Cardiol. 2006;47(8):1732–3.CrossRefPubMed Holverda S, Gan CT, Marcus JT, Postmus PE, Boonstra A, Vonk-Noordegraaf A. Impaired stroke volume response to exercise in pulmonary arterial hypertension. J Am Coll Cardiol. 2006;47(8):1732–3.CrossRefPubMed
24.
Zurück zum Zitat Crawford MH, Petru MA, Rabinowitz C. Effect of isotonic exercise training on left ventricular volume during upright exercise. Circulation. 1985;72(6):1237–43.CrossRefPubMed Crawford MH, Petru MA, Rabinowitz C. Effect of isotonic exercise training on left ventricular volume during upright exercise. Circulation. 1985;72(6):1237–43.CrossRefPubMed
25.
Zurück zum Zitat Ginzton LE, Conant R, Brizendine M, Laks MM. Effect of long-term high intensity aerobic training on left ventricular volume during maximal upright exercise. J Am Coll Cardiol. 1989;14(2):364–71.CrossRefPubMed Ginzton LE, Conant R, Brizendine M, Laks MM. Effect of long-term high intensity aerobic training on left ventricular volume during maximal upright exercise. J Am Coll Cardiol. 1989;14(2):364–71.CrossRefPubMed
26.
Zurück zum Zitat Rowland T. Echocardiography and circulatory response to progressive endurance exercise. Sports Med. 2008;38(7):541–51.CrossRefPubMed Rowland T. Echocardiography and circulatory response to progressive endurance exercise. Sports Med. 2008;38(7):541–51.CrossRefPubMed
27.
Zurück zum Zitat Gledhill N, Cox D, Jamnik R. Endurance athletes’ stroke volume does not plateau: major advantage is diastolic function. Med Sci Sports Exerc. 1994;26(9):1116–21.CrossRefPubMed Gledhill N, Cox D, Jamnik R. Endurance athletes’ stroke volume does not plateau: major advantage is diastolic function. Med Sci Sports Exerc. 1994;26(9):1116–21.CrossRefPubMed
28.
Zurück zum Zitat Sharma S, Elliott P, Whyte G, Jones S, Mahon N, Whipp B, McKenna WJ. Utility of cardiopulmonary exercise in the assessment of clinical determinants of functional capacity in hypertrophic cardiomyopathy. Am J Cardiol. 2000;86(2):162–8.CrossRefPubMed Sharma S, Elliott P, Whyte G, Jones S, Mahon N, Whipp B, McKenna WJ. Utility of cardiopulmonary exercise in the assessment of clinical determinants of functional capacity in hypertrophic cardiomyopathy. Am J Cardiol. 2000;86(2):162–8.CrossRefPubMed
29.
Zurück zum Zitat La Gerche A, Baggish AL, Knuuti J, Prior DL, Sharma S, Heidbuchel H, Thompson PD. Cardiac imaging and stress testing asymptomatic athletes to identify those at risk of sudden cardiac death. JACC Cardiovasc Imaging. 2013;6(9):993–1007.CrossRefPubMed La Gerche A, Baggish AL, Knuuti J, Prior DL, Sharma S, Heidbuchel H, Thompson PD. Cardiac imaging and stress testing asymptomatic athletes to identify those at risk of sudden cardiac death. JACC Cardiovasc Imaging. 2013;6(9):993–1007.CrossRefPubMed
30.
Zurück zum Zitat Gibbons RJ, Balady GJ, Beasley JW, Bricker JT, Duvernoy WF, Froelicher VF, Mark DB, Marwick TH, McCallister BD, Thompson Jr PD, et al. ACC/AHA Guidelines for Exercise Testing. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Exercise Testing). J Am Coll Cardiol. 1997;30(1):260–311.CrossRefPubMed Gibbons RJ, Balady GJ, Beasley JW, Bricker JT, Duvernoy WF, Froelicher VF, Mark DB, Marwick TH, McCallister BD, Thompson Jr PD, et al. ACC/AHA Guidelines for Exercise Testing. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Exercise Testing). J Am Coll Cardiol. 1997;30(1):260–311.CrossRefPubMed
31.
Zurück zum Zitat Badruddin SM, Ahmad A, Mickelson J, Abukhalil J, Winters WL, Nagueh SF, Zoghbi WA. Supine bicycle versus post-treadmill exercise echocardiography in the detection of myocardial ischemia: a randomized single-blind crossover trial. J Am Coll Cardiol. 1999;33(6):1485–90.CrossRefPubMed Badruddin SM, Ahmad A, Mickelson J, Abukhalil J, Winters WL, Nagueh SF, Zoghbi WA. Supine bicycle versus post-treadmill exercise echocardiography in the detection of myocardial ischemia: a randomized single-blind crossover trial. J Am Coll Cardiol. 1999;33(6):1485–90.CrossRefPubMed
32.
Zurück zum Zitat Modesto KM, Rainbird A, Klarich KW, Mahoney DW, Chandrasekaran K, Pellikka PA. Comparison of supine bicycle exercise and treadmill exercise Doppler echocardiography in evaluation of patients with coronary artery disease. Am J Cardiol. 2003;91(10):1245–8.CrossRefPubMed Modesto KM, Rainbird A, Klarich KW, Mahoney DW, Chandrasekaran K, Pellikka PA. Comparison of supine bicycle exercise and treadmill exercise Doppler echocardiography in evaluation of patients with coronary artery disease. Am J Cardiol. 2003;91(10):1245–8.CrossRefPubMed
33.
Zurück zum Zitat Park TH, Tayan N, Takeda K, Jeon HK, Quinones MA, Zoghbi WA. Supine bicycle echocardiography improved diagnostic accuracy and physiologic assessment of coronary artery disease with the incorporation of intermediate stages of exercise. J Am Coll Cardiol. 2007;50(19):1857–63.CrossRefPubMed Park TH, Tayan N, Takeda K, Jeon HK, Quinones MA, Zoghbi WA. Supine bicycle echocardiography improved diagnostic accuracy and physiologic assessment of coronary artery disease with the incorporation of intermediate stages of exercise. J Am Coll Cardiol. 2007;50(19):1857–63.CrossRefPubMed
Metadaten
Titel
Assessing exercise cardiac reserve using real-time cardiovascular magnetic resonance
verfasst von
Thu-Thao Le
Jennifer Ann Bryant
Alicia Er Ting
Pei Yi Ho
Boyang Su
Raymond Choon Chye Teo
Julian Siong-Jin Gan
Yiu-Cho Chung
Declan P. O’Regan
Stuart A. Cook
Calvin Woon-Loong Chin
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2017
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-017-0322-1

Weitere Artikel der Ausgabe 1/2017

Journal of Cardiovascular Magnetic Resonance 1/2017 Zur Ausgabe

PET kann infarktgefährdete Koronararterien entdecken

04.06.2024 Koronare Herzerkrankung Nachrichten

Der Nachweis aktiver Plaques mittels 18F-Natriumfluorid-PET hilft nicht nur, infarktgefährdete Patienten, sondern auch infarktgefährdete Koronararterien zu erkennen. Von einer gezielten Behandlung vulnerabler Plaques ist man trotzdem weit entfernt.

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.