Myocardial arterial spin labeling perfusion imaging with improved sensitivity

verfasst von: Hung Phi Do, Terrence R Jao, Krishna S Nayak

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2014

Abstract

Background

Myocardial arterial spin labeling (ASL) is a noninvasive MRI based technique that is capable of measuring myocardial blood flow (MBF) in humans. It suffers from poor sensitivity to MBF due to high physiological noise (PN). This study aims to determine if the sensitivity of myocardial ASL to MBF can be improved by reducing image acquisition time, via parallel imaging.

Methods

Myocardial ASL scans were performed in 7 healthy subjects at rest using flow-sensitive alternating inversion recovery (FAIR) tagging and balanced steady state free precession (SSFP) imaging. Sensitivity encoding (SENSE) with a reduction factor of 2 was used to shorten each image acquisition from roughly 300 ms per heartbeat to roughly 150 ms per heartbeat. A paired Student’s t-test was performed to compare measurements of myocardial blood flow (MBF) and physiological noise (PN) from the reference and accelerated methods.

Results

The measured PN (mean ± standard deviation) was 0.20 ± 0.08 ml/g/min for the reference method and 0.08 ± 0.05 ml/g/min for the accelerated method, corresponding to a 60% reduction. PN measured from the accelerated method was found to be significantly lower than that of the reference method (p = 0.0059). There was no significant difference between MBF measured from the accelerated and reference ASL methods (p = 0.7297).

Conclusions

In this study, significant PN reduction was achieved by shortening the acquisition window using parallel imaging with no significant impact on the measured MBF. This indicates an improvement in sensitivity to MBF and may also enable the imaging of subjects with higher heart rates and imaging during systole.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1532-429X-16-15) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing of interests.

Authors’ contributions

HPD designed and implemented the pulse sequences, performed the experiments, performed all data analysis, and drafted the manuscript. TRJ contributed to data analysis methods and revised the manuscript. KSN participated in experiment design, supervised the performance and revised the manuscript. All authors participated in discussion, read and approved the final manuscript.
Abkürzungen
ASL
Arterial spin labeling
FAIR
Flow-sensitive alternating inversion recovery
MBF
Myocardial blood flow
PN
Physiological noise
TN
Thermal noise
tSNR
Temporal signal to noise ratio
PI
Parallel imaging
SENSE
Sensitivity encoding
GRAPPA
Generalized autocalibrating partially parallel acquisition
TD
Trigger delay time
TI
Inversion time or post labeling delay time
RR
R-wave to R-wave interval
FOV
Field of view
ROI
Region of interest.

Background

Arterial spin labeling (ASL) is a quantitative, contrast-free MRI technique for measuring tissue perfusion. It is most commonly used in the brain for the clinical quantification of cerebral blood flow in cerebrovascular disease and neuro-oncology [13]. ASL outside the brain is an immature technology but shows great promise in the heart. Several groups have been able to use ASL to measure myocardial blood flow (MBF) in both animal models [49] and humans [1013]. Myocardial ASL has even been demonstrated to be compatible with pharmacological stress testing and is able to detect clinically relevant increases in MBF with vasodilation [14, 15], making it a potential diagnostic tool for detecting ischemic heart disease.
However, myocardial ASL perfusion imaging faces several challenges. The ASL signal has low sensitivity to blood flow and only produces a 1-8% signal change under normal physiological myocardial blood flows of 0.5-4 ml/g/min [16]. ASL also requires the subtraction of multiple image pairs and is particularly sensitive to imperfect subtraction caused by either respiratory or cardiac motion. In this study, we aim to reduce physiological noise (PN), which refers to temporal fluctuations in the ASL signal. We hypothesize that motion within the acquisition window is a major contributor to PN, which can be reduced by accelerating the image acquisition using parallel imaging (PI). We chose parallel imaging for acceleration because of its widespread clinical adoption over the past decade and its well-understood noise and artifact behavior.
In this study, we compare the performance of an accelerated myocardial ASL method using rate-2 sensitivity encoding (SENSE) [17] with a reference, unaccelerated myocardial ASL method [12] in healthy volunteers at rest.

Methods

Imaging methods

All experiments were performed on a Signa Excite HDxt 3 Tesla scanner (GE Healthcare, Waukesha, WI, USA) with an eight-channel cardiac receiver coil. Myocardial ASL was performed on a single short axis middle ventricular slice using flow-sensitive alternating inversion recovery (FAIR) [18, 19]. The pulse sequence shown in Figure 1 is composed of either a slab selective inversion pulse for control images or a global inversion pulse for tagged images, a fat saturation module, a 19-TR Kaiser-Bessel ramp up of flip angles to minimize transients [20], a snapshot balanced steady state free precession (b-SSFP) image acquisition, and a 19-TR Kaiser-Bessel ramp down of flip angles.
The reference ASL scan was implemented as described previously by Zun et al.[12, 15]. All image acquisitions had 3.2 ms TR, 1.5 ms TE, 50˚ prescribed flip angle, 62.5 kHz receiver bandwidth, and 24–26 cm isotropic field-of-view (FOV). Slice thickness was 10 mm, and the selective inversion slab for control images was 30 mm thick to ensure that the imaging slice was within the inversion slab. The reference ASL scans had a 96 × 96 matrix size resulting in an acquisition window of 307 ms. This was preceded by a 10 ms fat saturation module and a 60 ms Kaiser-Bessel ramp up of flip angles and followed by a 60 ms Kaiser-Bessel ramp down of flip angles. The total imaging window of the reference method was 440 ms.
The accelerated ASL scan used sensitivity encoding (SENSE) [17] with a reduction factor of 2 to shorten the acquisition window to 153 ms, as shown in Figure 1. This was also preceded by a 10 ms fat saturation module and a 60 ms Kaiser-Bessel ramp up, and followed by a 60 ms Kaiser-Bessel ramp down. The total imaging window of the accelerated method was 286 ms. We hypothesized that motion during the long acquisition window was a major source of physiological noise and could be reduced by shortening the acquisition window from 307 ms to 153 ms.
Breath-holds and cardiac triggering were used to minimize respiratory and cardiac motion, respectively. The FAIR labeling inversion pulse was timed to occur at mid-diastole through plethysmograph gating (PG). Mid-diastole was estimated to be at 77% of the RR interval duration [21] and the PG trigger delay was set to this value minus 200 ms to account for circulation time from the R-wave to the fingertip [22]. The center of the acquisition window was set to occur one heartbeat (TI = 1RR) after the FAIR labeling pulse. Control and tagged images were acquired in the same breath-hold.
The ASL protocol was comprised of 7 breath-holds and took roughly between 2–3 minutes. In the first breath-hold, baseline and noise images were acquired in 2 seconds. The baseline and noise images were later used for estimating the blood equilibrium magnetization (M0) and thermal noise (TN), respectively. Six pairs of control and tagged images were acquired for spatial temporal averaging in the following six 12-second breath-holds. There was a 6-second waiting period between control and tagged images within each breath-hold. The protocol was the same for the reference and accelerated scans except the acquisition window was shortening by using rate-2 SENSE in the accelerated scan.

Experimental methods

This study included 7 healthy subjects (6 males, 1 female, age 22 – 29 years, mean age = 25 years). The University of Southern California Institutional Review Board approved the study protocol and informed consent was obtained from all subjects. The scan protocol started with a localization scan and identification of a middle short axis slice [23]. A baseline image, an image without the FAIR labeling pulse, was acquired to ensure the slice was prescribed properly without banding artifacts over the myocardium. If banding artifacts were present on the myocardium, a frequency scout scan was performed. The offset frequency that resulted in no banding artifact in the region of interest (left ventricular myocardium) was recorded and used for subsequent scans. A fully sampled baseline image with no banding artifact was then acquired in a 1-second breath-hold to estimate the coil sensitivity map. The reference ASL scan was performed before the accelerated ASL scan in all subjects.

Data analysis

MBF, PN, and TN were calculated in the same way as in prior work by Zun et al.[12, 15]. All data processing was performed in MATLAB (Mathworks, Natick, MA). After image reconstruction, the myocardium was manually segmented and resampled into polar coordinates using a spatio-temporal averaging filter [24]. MBF quantification was derived from Buxton’s general kinematic model [25].
MBF = C T 2 M 0 · TI · exp TI / T 1 ,
where C, T, and M0 refer to the mean myocardial signal in the control, tagged, and baseline images, TI represents the post labeling delay time and was equal to the R-R interval, and T1 is the longitudinal relaxation time of blood, which was assumed to be 1650 ms [26].
The size of the spatial filter and the number of resampled segments could be freely chosen. Global MBF quantification was performed with a filter size of 2π and a single segment. Regional MBF quantification was performed with a filter size of π/3 and 6 segments [23]. Septal MBF quantification was performed using a filter size of 2π/3 with 3 segments. This measurement was performed in order to compare these data with a prior analysis of physiological noise [12]. A paired Student’s t-test was used to compare MBF and PN between the reference and accelerated ASL methods. Agreement between regional MBF measured from the two methods was also assessed by Bland-Altman plot.

Results

Representative control images acquired from the reference method are shown in Figure 2a; those acquired from the accelerated method are shown in Figure 2b. There was no visible residual aliasing in the SENSE accelerated images. Motion artifacts were seen in several images acquired from the reference method, as indicated by red arrows whereas SENSE accelerated images showed significantly less motion artifacts. It is worth noting that the per-pixel SNR of baseline images was 126 and 90 for the reference and accelerated acquisitions, respectively. The SNR of baseline images was high in both cases, and this is not expected to be a limitation of the approach.
Figure 3 shows global MBF measured from the reference (dark gray) and accelerated (light gray) methods. Error bars reflect one standard deviation of measured physiological noise. The mean and standard deviation of PN across all subjects was 0.20 ± 0.08 ml/g/min from the reference method and 0.08 ± 0.05 ml/g/min for the accelerated method, corresponding to a 60% reduction. PN measured from the accelerated method was found to be significantly lower than that from the reference method (p = 0.0059). The mean and standard deviation of thermal noise (TN) across all subjects was 0.0175 ± 0.0040 ml/g/min for the reference method and 0.0241 ± 0.0045 ml/g/min for the accelerated method, and the average standard deviation increase was 39% (expected based on the shortened readout time and g-factor losses). The mean and standard deviation of global MBF measured with the reference method were 1.20 ± 0.44 ml/g/min, and measured with the accelerated ASL method were 1.24 ± 0.25 ml/g/min. There was no significant difference between global MBF measured from the two techniques (p = 0.7297). The measured global MBF range of 0.79-2.03 ml/g/min for the reference method and 0.86-1.52 ml/g/min for the accelerated method are consistent with PET literature values 0.73-2.43 ml/g/min [27].
Table 1 compares the accelerated and reference ASL methods. A significant difference was found between global, septal, and regional PN between the two methods with p-values of 0.0059, 0.0002, and 0.0001, respectively. There was no significant difference between global, septal, and regional MBF with p-values of 0.7297, 0.6806, and 0.5353, respectively. Septal MBF and septal PN values for the reference method were comparable to results reported in Ref. [12], which were septal MBF = 1.36 ± 0.38 ml/g/min and septal PN = 0.23 ± 0.12 ml/g/min
Table 1
Comparison between the reference and accelerated ASL methods
 
Global
Septal
Regional
 
MBF
PN
MBF
PN
MBF
PN
Reference method
1.20 ± 0.44
0.20 ± 0.08
1.23 ± 0.56*
0.28 ± 0.11*
1.22 ± 0.68
0.34 ± 0.21
Accelerated method
1.24 ± 0.25
0.08 ± 0.05
1.28 ± 0.44
0.17 ± 0.06
1.28 ± 0.46
0.21 ± 0.11
p-value
0.7297
0.0059
0.6806
0.0002
0.5353
0.0001
Values are reported as mean ± standard deviation across all subjects in ml/g/min. Values indicated by (*) are comparable to the results reported in Ref. [12].
Figure 4 shows a Bland-Altman plot comparing regional MBF (6 segments per subject) measured from the reference and accelerated method. There was no significant bias in regional MBF measurements using the accelerated method.

Discussion

This study demonstrates that PN in myocardial ASL can be significantly reduced by shortening the acquisition window using SENSE, which supports our hypothesis that motion during image acquisition was a dominant source of PN. Figure 2 shows that motion artifacts from control images in the reference method are significantly smaller after shortening the image acquisition window to roughly 150 ms per cardiac cycle. We suspect that motion during the acquisition window causes inconsistencies amongst reconstructed pairs of control and tagged images, which leads to temporal fluctuations in measured MBF for the reference method. Shortening the acquisition window using parallel imaging effectively decreased these likelihood of such inconsistencies, which in turn reduced PN by roughly 60%. The PN reduction corresponds to a 158% increase in temporal SNR (tSNR). An increased tSNR directly translates to improved sensitivity to MBF. Furthermore, MBF measured using the accelerated method was comparable to those from the reference method. There was no significant difference found between global MBF (p = 0.7297) and regional MBF (p = 0.5353) measurements from the two methods.
It may seem counterintuitive that the use of parallel imaging could increase tSNR, when it is known that SNR decreases. Noise analysis from this study revealed that measured TN increased from 0.0175 ± 0.0040 ml/g/min to 0.0241 ± 0.0045 ml/g/min after accelerating by rate 2. While PN decreased from 0.20 ± 0.08 ml/g/min to 0.08 ± 0.05 ml/g/min after acceleration. There was an overall reduction in total noise because the increase in TN from the shortened acquisition window and g-factor losses was much smaller than the decrease in PN from more consistent reconstructions of control and tagged image pairs. This suggests that it may be possible to reduce noise further by pushing the acceleration until the incremental decreases in PN are matched by equal increases in TN.
Although this study only included healthy subjects at rest, one would expect the accelerated ASL method to also improve sensitivity of myocardial ASL in patient cohorts at both at rest and stress. Zun et al.[15] reported that global PN from 16 “normal” patients was 0.64 and 1.36 at rest and stress, respectively, which is roughly 3 times higher than that measured from healthy subjects, suggesting an even greater need for PN reduction.
Shortening the acquisition window using SENSE may have the additional benefits of allowing imaging of subjects with high heart rates (>90 beats per minute) and image during systole (<100 ms acquisition window). In subjects with high heart rates, diastole is shorter and faster image acquisition may help reduce cardiac motion artifacts. Likewise, systole is shorter than diastole and may also benefit from the same type of accelerated image acquisition. In addition, the use of a shorter acquisition window per slice may allow the spatial coverage to be extended to 2 or more slices per heartbeat, following a single FAIR label.
One important drawback of SENSE is that when the FOV is smaller than the signal-producing region, SENSE is unable to completely unwrap all aliasing and can lead to residual artifacts at the center of the reconstructed images. Generalized autocalibrating partially parallel acquisition (GRAPPA) [28] is a natural alternative and has been shown to be more robust when the FOV is smaller than the object [29]. Rapid imaging methods based on partial Fourier and/or constrained reconstruction may also be appropriate and remain to be explored.
Parallel imaging has been extensively used in first pass CMR perfusion imaging to improve spatio-temporal resolution and spatial coverage [30]. The primary side effect associated with its use is reduced SNR based on the shortened readout time and g-factor losses. Compared to first-pass methods, cardiac ASL methods (based on apparent-T1 or signal subtraction) are limited by temporal SNR. This is a fundamental difference. Many T1-based methods and signal subtraction methods rely on long image acquisition windows, on the order of 30% of the RR interval, and it is conceivable that these methods will experience an increase in temporal SNR by using parallel imaging, as was demonstrated in this study.
There are several inherent assumptions associated with this particular ASL method. First, we assume a stable heart rate within each breath-hold. Heart rate variability within a breath hold may result in control and tagged images being acquired at slightly different cardiac phase. Second, we use baseline image intensity as a surrogate for M0 and assume the blood tissue partition coefficient to be one. In brain ASL, M0 is commonly estimated by scaling local tissue with the blood tissue partition coefficient, where the local tissue signal is acquired from a proton density weighted scan. Third, we assume perfect inversion efficiency for all inversion pulses. Based on our own in-vivo measurements the efficiency is consistently above 94%. Note that the inversion slab thickness of 3 cm does not introduce quantification errors when using a 1 R-R labeling delay, because this does not leave enough time for unlabeled blood within the LV blood pool to perfuse the myocardium. The inversion thickness would, however, require compensation when using longer labeling delays [8].

Conclusions

By using parallel imaging to shorten the acquisition window of human myocardial ASL scans, we have demonstrated a significant reduction in physiological noise, with no significant change in measured MBF. The reduction in PN provides more temporally consistent measurements of MBF and improves the sensitivity of myocardial ASL to MBF.
Written informed consent was obtained from all seven subjects for the publication of this report and any accompanying images.

Acknowledgements

We gratefully acknowledge grant support from the Wallace H. Coulter Foundation and the American Heart Association, and research support from GE Healthcare. HPD has received support from a Merit Fellowship from the USC Danna and David Dornsife College of Letters, Arts and Sciences. TRJ has received support from a USC Provost’s Fellowship.
Grant Sponsor: Wallace H. Coulter Foundation, American Heart Association
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The authors declare that they have no competing of interests.

Authors’ contributions

HPD designed and implemented the pulse sequences, performed the experiments, performed all data analysis, and drafted the manuscript. TRJ contributed to data analysis methods and revised the manuscript. KSN participated in experiment design, supervised the performance and revised the manuscript. All authors participated in discussion, read and approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Literatur
1.
Zurück zum Zitat Calamante F, Thomas DL, Pell GS, Wiersma J, Turner R: Measuring cerebral blood flow using magnetic resonance imaging techniques. J of Cereb Blood Flow & Metab. 1999, 19 (7): 701-35.CrossRef Calamante F, Thomas DL, Pell GS, Wiersma J, Turner R: Measuring cerebral blood flow using magnetic resonance imaging techniques. J of Cereb Blood Flow & Metab. 1999, 19 (7): 701-35.CrossRef
2.
Zurück zum Zitat Detre JA, Wang J, Wang Z, Rao H: Arterial spin-labeled perfusion MRI in basic and clinical neuroscience. Curr Opin Neurol. 2009, 22 (4): 348-55. 10.1097/WCO.0b013e32832d9505.CrossRefPubMed Detre JA, Wang J, Wang Z, Rao H: Arterial spin-labeled perfusion MRI in basic and clinical neuroscience. Curr Opin Neurol. 2009, 22 (4): 348-55. 10.1097/WCO.0b013e32832d9505.CrossRefPubMed
3.
Zurück zum Zitat Watts JM, Whitlow CT, Maldjian JA: Clinical applications of arterial spin labeling. NMR Biomed. 2013, 26: 892-900. 10.1002/nbm.2904.CrossRefPubMed Watts JM, Whitlow CT, Maldjian JA: Clinical applications of arterial spin labeling. NMR Biomed. 2013, 26: 892-900. 10.1002/nbm.2904.CrossRefPubMed
4.
Zurück zum Zitat Belle V, Kahler E, Waller C, Rommel E, Voll S, Hiller K, Bauer WR, Haase A: In Vivo quantitative mapping of cardiac perfusion in rats using a noninvasive MR spin-labeling method. J Magn Reson Imaging. 1998, 8 (6): 1240-5. 10.1002/jmri.1880080610.CrossRefPubMed Belle V, Kahler E, Waller C, Rommel E, Voll S, Hiller K, Bauer WR, Haase A: In Vivo quantitative mapping of cardiac perfusion in rats using a noninvasive MR spin-labeling method. J Magn Reson Imaging. 1998, 8 (6): 1240-5. 10.1002/jmri.1880080610.CrossRefPubMed
5.
Zurück zum Zitat Waller C, Kahler E, Hiller K, Hu K, Nahrendorf M, Voll S, Haase A, Ertl G, Bauer WR: Myocardial perfusion and intracapillary blood volume in rats at rest and with coronary dilatation: MR imaging in vivo with use of a spin-labeling technique. Radiology. 2000, 215 (1): 189-97. 10.1148/radiology.215.1.r00ap07189.CrossRefPubMed Waller C, Kahler E, Hiller K, Hu K, Nahrendorf M, Voll S, Haase A, Ertl G, Bauer WR: Myocardial perfusion and intracapillary blood volume in rats at rest and with coronary dilatation: MR imaging in vivo with use of a spin-labeling technique. Radiology. 2000, 215 (1): 189-97. 10.1148/radiology.215.1.r00ap07189.CrossRefPubMed
6.
Zurück zum Zitat Streif JU, Nahrendorf M, Hiller K, Waller C, Wiesmann F, Rommel E, Haase A, Bauer WR: In vivo assessment of absolute perfusion and intracapillary blood volume in the murine myocardium by spin labeling magnetic resonance imaging. Magn Reson Med. 2005, 53 (3): 584-92. 10.1002/mrm.20327.CrossRefPubMed Streif JU, Nahrendorf M, Hiller K, Waller C, Wiesmann F, Rommel E, Haase A, Bauer WR: In vivo assessment of absolute perfusion and intracapillary blood volume in the murine myocardium by spin labeling magnetic resonance imaging. Magn Reson Med. 2005, 53 (3): 584-92. 10.1002/mrm.20327.CrossRefPubMed
7.
Zurück zum Zitat Vandsburger MH, Janiczek RL, Xu Y, French BA, Meyer CH, Kramer CM, Epstein FH: Improved arterial spin labeling after myocardial infarction in mice using cardiac and respiratory gated Look-Locker imaging with fuzzy C-means clustering. Magn Reson Med. 2010, 63 (3): 648-57. 10.1002/mrm.22280.PubMedCentralCrossRefPubMed Vandsburger MH, Janiczek RL, Xu Y, French BA, Meyer CH, Kramer CM, Epstein FH: Improved arterial spin labeling after myocardial infarction in mice using cardiac and respiratory gated Look-Locker imaging with fuzzy C-means clustering. Magn Reson Med. 2010, 63 (3): 648-57. 10.1002/mrm.22280.PubMedCentralCrossRefPubMed
8.
Zurück zum Zitat Campbell-Washburn AE, Price AN, Wells JA, Thomas DL, Ordidge RJ, Lythgoe MF: Cardiac arterial spin labeling using segmented ECG-gated look-locker FAIR: variability and repeatability in preclinical studies. Magn Reson Med. 2013, 69 (1): 238-47. 10.1002/mrm.24243.CrossRefPubMed Campbell-Washburn AE, Price AN, Wells JA, Thomas DL, Ordidge RJ, Lythgoe MF: Cardiac arterial spin labeling using segmented ECG-gated look-locker FAIR: variability and repeatability in preclinical studies. Magn Reson Med. 2013, 69 (1): 238-47. 10.1002/mrm.24243.CrossRefPubMed
9.
Zurück zum Zitat Kober F, Iltis I, Izquierdo M, Desrois M, Ibarrola D, Cozzone PJ, Bernard M: High-resolution myocardial perfusion mapping in small animals in vivo by spin-labeling gradient-echo imaging. Magn Reson Med. 2004, 51 (1): 62-7. 10.1002/mrm.10676.CrossRefPubMed Kober F, Iltis I, Izquierdo M, Desrois M, Ibarrola D, Cozzone PJ, Bernard M: High-resolution myocardial perfusion mapping in small animals in vivo by spin-labeling gradient-echo imaging. Magn Reson Med. 2004, 51 (1): 62-7. 10.1002/mrm.10676.CrossRefPubMed
10.
Zurück zum Zitat Poncelet B, Koelling T, Schmidt C, Kwong K, Reese T, Ledden P, Kantor H, Brady T, Weisskoff R: Measurement of human myocardial perfusion by double-gated flow alternating inversion recovery EPI. Magn Reson Med. 1999, 41 (3): 510-9. 10.1002/(SICI)1522-2594(199903)41:3<510::AID-MRM13>3.0.CO;2-G.CrossRefPubMed Poncelet B, Koelling T, Schmidt C, Kwong K, Reese T, Ledden P, Kantor H, Brady T, Weisskoff R: Measurement of human myocardial perfusion by double-gated flow alternating inversion recovery EPI. Magn Reson Med. 1999, 41 (3): 510-9. 10.1002/(SICI)1522-2594(199903)41:3<510::AID-MRM13>3.0.CO;2-G.CrossRefPubMed
11.
Zurück zum Zitat Northrup BE, McCommis KS, Zhang H, Ray S, Woodard PK, Gropler RJ, Zheng J: Resting myocardial perfusion quantification with CMR arterial spin labeling at 1.5 T and 3.0 T. J Cardiovasc Magn Reson. 2008, 10: 53-10.1186/1532-429X-10-53.PubMedCentralCrossRefPubMed Northrup BE, McCommis KS, Zhang H, Ray S, Woodard PK, Gropler RJ, Zheng J: Resting myocardial perfusion quantification with CMR arterial spin labeling at 1.5 T and 3.0 T. J Cardiovasc Magn Reson. 2008, 10: 53-10.1186/1532-429X-10-53.PubMedCentralCrossRefPubMed
12.
Zurück zum Zitat Zun Z, Wong EC, Nayak KS: Assessment of myocardial blood flow (MBF) in humans using arterial spin labeling (ASL): Feasibility and noise analysis. Magn Reson Med. 2009, 62 (4): 975-83. 10.1002/mrm.22088.CrossRefPubMed Zun Z, Wong EC, Nayak KS: Assessment of myocardial blood flow (MBF) in humans using arterial spin labeling (ASL): Feasibility and noise analysis. Magn Reson Med. 2009, 62 (4): 975-83. 10.1002/mrm.22088.CrossRefPubMed
13.
Zurück zum Zitat Wang DJJ, Bi X, Avants BB, Meng T, Zuehlsdorff S, Detre JA: Estimation of perfusion and arterial transit time in myocardium using free-breathing myocardial arterial spin labeling with navigator-echo. Magn Reson Med. 2010, 64 (5): 1289-95. 10.1002/mrm.22630.PubMedCentralCrossRefPubMed Wang DJJ, Bi X, Avants BB, Meng T, Zuehlsdorff S, Detre JA: Estimation of perfusion and arterial transit time in myocardium using free-breathing myocardial arterial spin labeling with navigator-echo. Magn Reson Med. 2010, 64 (5): 1289-95. 10.1002/mrm.22630.PubMedCentralCrossRefPubMed
14.
Zurück zum Zitat Wacker C, Fidler F, Dueren C, Hirn S, Jakob P, Ertl G, Haase A, Bauer W: Quantitative assessment of myocardial perfusion with a spin-labeling technique: Preliminary results in patients with coronary artery disease. J Magn Reson Imaging. 2003, 18 (5): 555-60. 10.1002/jmri.10386.CrossRefPubMed Wacker C, Fidler F, Dueren C, Hirn S, Jakob P, Ertl G, Haase A, Bauer W: Quantitative assessment of myocardial perfusion with a spin-labeling technique: Preliminary results in patients with coronary artery disease. J Magn Reson Imaging. 2003, 18 (5): 555-60. 10.1002/jmri.10386.CrossRefPubMed
15.
Zurück zum Zitat Zun Z, Varadarajan P, Pai RG, Wong EC, Nayak KS: Arterial spin labeled CMR detects clinically relevant increase in myocardial blood flow with vasodilation. JACC Cardiovasc Imaging. 2011, 4 (12): 1253-61. 10.1016/j.jcmg.2011.06.023.CrossRefPubMed Zun Z, Varadarajan P, Pai RG, Wong EC, Nayak KS: Arterial spin labeled CMR detects clinically relevant increase in myocardial blood flow with vasodilation. JACC Cardiovasc Imaging. 2011, 4 (12): 1253-61. 10.1016/j.jcmg.2011.06.023.CrossRefPubMed
16.
Zurück zum Zitat Epstein FH, Meyer CH: Myocardial perfusion using arterial spin labeling CMR: promise and challenges. JACC Cardiovasc Imaging. 2011, 4 (12): 1262-4. 10.1016/j.jcmg.2011.08.015.PubMedCentralCrossRefPubMed Epstein FH, Meyer CH: Myocardial perfusion using arterial spin labeling CMR: promise and challenges. JACC Cardiovasc Imaging. 2011, 4 (12): 1262-4. 10.1016/j.jcmg.2011.08.015.PubMedCentralCrossRefPubMed
17.
Zurück zum Zitat Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P: SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999, 42 (5): 952-62. 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S.CrossRefPubMed Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P: SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999, 42 (5): 952-62. 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S.CrossRefPubMed
18.
Zurück zum Zitat Kim S: Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn Reson Med. 1995, 34 (3): 293-301. 10.1002/mrm.1910340303.CrossRefPubMed Kim S: Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn Reson Med. 1995, 34 (3): 293-301. 10.1002/mrm.1910340303.CrossRefPubMed
19.
Zurück zum Zitat Kwong K, Chesler D, Wwisskoff R, Donahue K, Davis T, Ostergaard L, Campbell T, Rosen B: MR perfusion studies with T-1-weighted echo-planar imaging. Magn Reson Med. 1995, 34 (6): 878-87. 10.1002/mrm.1910340613.CrossRefPubMed Kwong K, Chesler D, Wwisskoff R, Donahue K, Davis T, Ostergaard L, Campbell T, Rosen B: MR perfusion studies with T-1-weighted echo-planar imaging. Magn Reson Med. 1995, 34 (6): 878-87. 10.1002/mrm.1910340613.CrossRefPubMed
20.
Zurück zum Zitat Le Roux P: Simplified model and stabilization of SSFP sequences. J Magn Reson. 2003, 163 (1): 23-37. 10.1016/S1090-7807(03)00115-0.CrossRefPubMed Le Roux P: Simplified model and stabilization of SSFP sequences. J Magn Reson. 2003, 163 (1): 23-37. 10.1016/S1090-7807(03)00115-0.CrossRefPubMed
21.
Zurück zum Zitat Otton JM, Phan J, Feneley M, Yu C, Sammel N, McCrohon J: Defining the mid-diastolic imaging period for cardiac CT – lessons from tissue Doppler echocardiography. BMC Med Imaging. 2013, 13 (1): 5-10.1186/1471-2342-13-5.PubMedCentralCrossRefPubMed Otton JM, Phan J, Feneley M, Yu C, Sammel N, McCrohon J: Defining the mid-diastolic imaging period for cardiac CT – lessons from tissue Doppler echocardiography. BMC Med Imaging. 2013, 13 (1): 5-10.1186/1471-2342-13-5.PubMedCentralCrossRefPubMed
22.
Zurück zum Zitat Allen J, Murray A: Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes. J Hum Hypertens. 2002, 16 (10): 711-7. 10.1038/sj.jhh.1001478.CrossRefPubMed Allen J, Murray A: Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes. J Hum Hypertens. 2002, 16 (10): 711-7. 10.1038/sj.jhh.1001478.CrossRefPubMed
23.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS: AHA scientific statement. Circulation. 2002, 105: 539-42. 10.1161/hc0402.102975.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS: AHA scientific statement. Circulation. 2002, 105: 539-42. 10.1161/hc0402.102975.CrossRefPubMed
24.
Zurück zum Zitat Jao T, Zun Z, Varadarajan P, Pai RG, Nayak KS: Mapping of myocardial ASL perfusion and perfusion reserve data. Proceedings of the 19th Annual Meeting of the International Society for Magnetic Resonance in Medicine. May 2011, Montreal, Quebec, Canada, 1339- Jao T, Zun Z, Varadarajan P, Pai RG, Nayak KS: Mapping of myocardial ASL perfusion and perfusion reserve data. Proceedings of the 19th Annual Meeting of the International Society for Magnetic Resonance in Medicine. May 2011, Montreal, Quebec, Canada, 1339-
25.
Zurück zum Zitat Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR: A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med. 1998, 40 (3): 383-96. 10.1002/mrm.1910400308.CrossRefPubMed Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR: A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med. 1998, 40 (3): 383-96. 10.1002/mrm.1910400308.CrossRefPubMed
26.
Zurück zum Zitat Lu H, Clingman C, Golay X, van Zijl P: Determining the longitudinal relaxation time (T1) of blood at 3.0 Tesla. Magn Reson Med. 2004, 52 (3): 679-82. 10.1002/mrm.20178.CrossRefPubMed Lu H, Clingman C, Golay X, van Zijl P: Determining the longitudinal relaxation time (T1) of blood at 3.0 Tesla. Magn Reson Med. 2004, 52 (3): 679-82. 10.1002/mrm.20178.CrossRefPubMed
27.
Zurück zum Zitat Chareonthaitawee P, Kaufmann P, Rimoldi O, Camici P: Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovasc Res. 2001, 50 (1): 151-61. 10.1016/S0008-6363(01)00202-4.CrossRefPubMed Chareonthaitawee P, Kaufmann P, Rimoldi O, Camici P: Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovasc Res. 2001, 50 (1): 151-61. 10.1016/S0008-6363(01)00202-4.CrossRefPubMed
28.
Zurück zum Zitat Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A: Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002, 47 (6): 1202-10. 10.1002/mrm.10171.CrossRefPubMed Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A: Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002, 47 (6): 1202-10. 10.1002/mrm.10171.CrossRefPubMed
29.
Zurück zum Zitat Griswold MA, Kannengiesser S, Heidemann RM, Wang J, Jakob PM: Field-of-view limitations in parallel imaging. Magn Reson Med. 2004, 52 (5): 1118-26. 10.1002/mrm.20249.CrossRefPubMed Griswold MA, Kannengiesser S, Heidemann RM, Wang J, Jakob PM: Field-of-view limitations in parallel imaging. Magn Reson Med. 2004, 52 (5): 1118-26. 10.1002/mrm.20249.CrossRefPubMed
30.
Zurück zum Zitat Kellman P, Arai AE: Imaging sequences for first pass perfusion-A review. J Cardiovasc Magn Reson. 2007, 9 (3): 525-37. 10.1080/10976640601187604.CrossRefPubMed Kellman P, Arai AE: Imaging sequences for first pass perfusion-A review. J Cardiovasc Magn Reson. 2007, 9 (3): 525-37. 10.1080/10976640601187604.CrossRefPubMed
Metadaten
Titel
Myocardial arterial spin labeling perfusion imaging with improved sensitivity
verfasst von
Hung Phi Do
Terrence R Jao
Krishna S Nayak
Publikationsdatum
01.12.2014
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2014
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-16-15

Weitere Artikel der Ausgabe 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.