A statistical model that predicts the length from the left subclavian artery to the celiac axis; towards accurate intra aortic balloon sizing

verfasst von: Haralabos Parissis, Alan Soo, Michalis Leotsinidis, Dimitrios Dougenis

Erschienen in: Journal of Cardiothoracic Surgery | Ausgabe 1/2011

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Ideally the length of the Intraaortic balloon membrane (22-27.5 cm) should match to the distance from the left subclavian artery (LSA) to the celiac axis (CA), (LSA - CA). By being able to estimate this distance, better guidance regarding IABP sizing could be recommended.

Methods

Internal aortic lengths and demographic values were collected from a series of 40 cadavers during autopsy. External somatometric measurements were also obtained.
There were 23 males and 17 females. The mean age was 73.1+/-13.11 years, weight 56.75+/-12.51 kg and the height 166+/-9.81 cm.

Results

Multiple regression analysis revealed the following predictor variables (R2 > 0.70) for estimating the length from LSA to CA: height (standardized coefficient (SRC) = 0.37, p = 0.004), age (SRC = 0.35, p < 0.001), sex (SRC = 0.21, p = 0.088) and the distance from the jugular notch to trans-pyloric plane (SRC = 0.61, p < 0.001).
Recommendations: If LSA-CA < 21.9 cm use 34 cc IABP & if LSA-CA > 26.3 cm use 50 cc IABP. However if LSA-CA = 21.9- 26.3 cm use 40 cc, but be aware that it could be "aortic length-balloon membrane length" mismatching.

Conclusions

Routinely, IABP size selection is being dictated by the patient's height. Inevitably, this leads to pitfalls. We reported a mathematical model of accurate intraaortic balloon sizing, which is easy to be applied and has a high predictive value.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1749-8090-6-95) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

HP conceived of the idea, carried out all the measurements and wrote the manuscript. AS helped in organizing the data and also helped in the process of writing the manuscript. ML carried out the statistical analysis. DD supervised the project and offer valuable points and ideas. All authors read and approved the final manuscript.

Introduction

The selection of the intraaortic balloon pump (IABP) in adults has been mainly limited to the use of 40 cc and occasionally 34 cc balloon volume, with a membrane length that varies among manufacturers from 22 to 27.5 cm and an inflated diameter 15 - 18 mm (Table 1).
Table 1
IABP sizes
Volume
Membrane length
Inflated diameter
34 cc
21.9 (cm)
14.7 (cm)
40
26.3
15 -16.2
50
27.5
18.3
Although in a number of clinical cases the 40 cc volume is sufficient, it shall be noted that an increased balloon volume could potentially contributes to morbidity from vascular events and a reduced balloon volume could decrease the beneficial hemodynamic effects for the heart. There is a causal relashionship between the patient's height and balloon trauma. According to Cox et al [1] the balloon size shall be adjusted depending on the patient's height.
The ideal balloon for any patient shall be placed at an ideal position, which equals in length the distance from the left subclavian artery to the celiac axis, 90-95% of the diameter of the descending aorta and equals in volume the blood volume inside the aorta lumen at all given time [2]..
Modern treatment guidelines (Table 2) suggest for patients up to 162 cm high the use of 34 cc balloons, for patients 162-182 cm high a 40 cc balloon and for patients over 182 cm high the use of 50 cc balloons. The problem arises when the length of the balloon membrane is greater than the length from the left subclavian artery to the celiac axis. Renal arteries and visceral circulation may potentially be occluded at each inflation of the balloon.
Table 2
The current recommendations regarding IABP sizing
34 cc
40 cc
50 cc
147-162 cm
162-182 cm
> 182 cm
BSA < 1.8
BSA > 1.8
 
This paper is a pilot study aiming to predict the length of several anatomical lengths in the descending aorta and by extension to lead to the optimum IABP size selection for every individual patient.
The main question aims to identify an equation that can calculate the length from the take off of the left subclavian artery to the origin of the celiac axis. The length of the balloon membrane in association with the distance from the left subclavian artery to the take off of the celiac axis will be used as a possible guide to the optimum sizing of the IABP.

Material and methods

Measurements were carried out from a series of 40 cadavers during autopsy. Twenty-three were male cadavers and 17 female. The mean age was 73.1+/-13.11 years, mean height was 166+/-9.81 cm and mean weight was 56.75+/-12.51 kg.
External somatometric measurements were obtained before autopsy (Figure 1a). With intact aortic length, the internal measurements were carried out in situ (Figure 2a). The distance between the origin of the left subclavian artery and the insertion point of the catheter was measured by calculating the length of a guide wire that was introduced to the aortic lumen, from a point 1 cm below the middle inguinal region to the orifice of the subclavian artery (cross-validated via digital palpation). All other internal measurements were carried out after the anatomical aortic dissection during autopsy (see Table 3, and Table 4).
Table 3
Row data and measurements depicted in Figure 1
Demographics
External Lengths (cm) Diameters
Internal Lengths
Depth
Pt
Hght
Wght
Age
SEX
BSA
Meas1
Meas2
Meas3
Meas4
Meas5
Dthrax
Dren
Dbifrc
LSA-CA
LSA-Ren
LSA-Bifr
LSA-Insr
Dpth
1
168
55
89
m
1.62
26
45
24.25
43.25
49
30
29
25
22.5
27.5
37.5
53
2.5
2
178
71
76
m
1.88
29.5
53.1
27.4
51
58.2
29
28
24
26
27.6
36
52
2.3
3
148
47
84
f
1.38
25.5
44.5
23.25
42.25
49
31
26
21
22.5
25
32.5
50.5
2.2
4
143.5
41
79
f
1.28
21.5
44
19.45
41.95
49.4
28
23
20
19.5
21.4
31
48.5
2.4
5
165
55
89
f
1.60
26
44
24.05
42.55
50.5
27
25
20
23
26
36.5
53.5
2.1
6
163
46
34
f
1.47
26.5
49
24.6
47.6
51
21
20
17
20.5
23.4
32.2
51
2.1
7
153
63
52
f
1.60
27.5
50
25.3
47.8
53
24
21
18
24
27
34
49.5
2.9
8
185
50
77
m
1.67
27
47
24.6
44.6
51.5
32
30
26
27
29.5
41
57.8
1.8
9
165
50
73
m
1.53
28.5
50.7
26.65
48.85
52
23
21
17
23
25
38
51
2.1
10
163
41
56
f
1.40
22
40
20.35
38.35
42.3
18
17
13
20
23
33
47
1.8
11
173
45
83
f
1.52
26.5
47
24.75
45.25
48
23
21
18
26
29
41
56.5
2.0
12
165
66
68
f
1.73
25
43
23.5
41.5
45.6
27
24
19
24.5
28
38.5
52.6
2.5
13
173
63
60
f
1.75
26
46.5
24
44.5
49.5
29
26
21
23
26.5
37.6
53.2
2.4
14
175
64
58
f
1.78
26.5
50.5
24.25
48.25
53.8
28.5
27
20
24.5
27.5
37.7
54.3
2.6
15
170
55
74
f
1.63
26
48
24
46
52.5
26
24.5
19.5
25
28.7
36.6
53.6
2.3
16
175
63
74
m
1.77
27.5
49
25
46.5
53.7
24
22.5
18
22
26.8
38.8
51.5
1.8
17
173
64
94
m
1.76
26.5
46
24.5
44
49.5
25.5
23
17.5
23.5
25.5
36.2
50.7
2.0
18
180
64
76
m
1.82
25
44.5
23
42.5
50.6
24
22.5
18
24
27.8
38.3
56.5
1.6
19
153
54
42
f
1.50
25.5
46
23.75
44.25
49.8
26
24.5
19
20.5
23.3
31.5
47.8
2.1
20
173
84
70
m
1.98
31
52
28.1
49.1
56.2
28
27
19.5
29
31.5
40.7
55.9
3.1
21
179
78
71
m
1.97
29.5
51
27.25
48.75
56.7
29
27.5
20.5
28
32
42.9
58.1
2.3
22
180
98
74
m
2.18
31
51.5
28.5
49
55.8
28.5
26.5
18.5
27
31.8
43.7
59.6
3.3
23
178
55
73
m
1.69
27.5
48.5
24.75
45.75
53.9
24
23
17
25.5
29.7
41.8
58.7
2.1
24
163
64
83
m
1.69
28.5
49
25.75
46.25
52.8
27
25.5
18.5
23
26.9
35.8
52.6
2.6
25
163
43
75
f
1.43
25.5
44.5
23.85
42.85
48.2
23
21
16.5
23
26.5
35.7
50.8
1.7
26
158
40
57
m
1.35
24.5
42
22.75
40.25
45.6
22
20.5
14.5
20
23.6
30.9
46.6
1.8
27
178
72
69
m
1.89
27
46.5
24.75
44.25
51.2
24
21
15.5
24
28.3
39
56.8
1.9
28
163
56
63
m
1.60
26
45
24.25
43.25
49.2
23.5
22
16
23
26.8
36.1
50.4
2.0
29
170
46
84
m
1.51
25.5
43.5
23.25
41.25
47.4
25
23.5
17
24
28.1
38.8
53.9
1.6
30
180
57
73
m
1.73
27
49
24.25
46.25
54.6
27
25
18.5
25
29.6
42.3
57.7
1.7
31
155
55
76
f
1.53
24.5
45.5
22.25
43.25
48.7
24.5
22.5
19
23
26.2
33.1
48.3
1.9
32
156
54
71
f
1.52
24
43.5
21.75
41.25
46.5
23.5
22.5
17.5
22
25.3
32.4
49.1
1.8
33
153
42
93
f
1.35
21.5
42.4
19.75
40.65
46.2
22
20.5
15.5
21.5
24.8
31.3
47.9
1.5
34
165
55
87
m
1.60
26.5
49.5
24
47
52.9
27.5
25
19.5
26
28.5
37.2
51.8
1.8
35
175
66
73
m
1.80
28
47
25.25
44.25
52.1
26
24.5
18
24
28.9
38.1
54.1
1.5
36
163
40
90
m
1.38
26.5
44
24.25
41.75
48.2
24.5
22
17.5
26
28.1
37.4
50.9
1.3
37
163
62
88
m
1.67
25
46.5
22.75
44.25
49.6
28.5
27
20
24
27.8
36.9
51.6
1.9
38
165
49
69
f
1.52
24.5
43.5
22.75
41.75
47.7
26
24.5
18.5
23
27.1
36.1
50.2
1.7
39
165
47
80
m
1.50
25
44
23
42
48.4
25.5
23.5
16.5
23
26.9
35.3
51.1
1.5
40
158
50
67
m
1.49
25
45.5
22.75
43.25
48
26
24.5
17.5
22
25.6
34.2
49.9
1.7
(The Trans-pyloric plane is located halfway between the jugular notch and the upper border of the pubic symphysis);
Table 4
Definitions of the variables analysed in table 3 (eg. Measurements, Diameters, lengths between various aortic branches)
Measurement 1
from the jugular notch to the trans-pyloric plane
LSA-CA
from the left subclavian take off to the celiac axis
Measurement 2
from the jugular notch to the anterior superior iliac spine
LSA-Renal
from the left subclavian take off to the renal artery
Measurement 3
from the middle point between the jugular
notch and the sternal angle to the
trans-pyloric plane
LSA-Bifur
from the left subclavian take off
to the aortic bifurcation
Measurement 4
from the middle point between the jugular notch and the sternal angle to the anterior superior iliac spine.
LSA-Inser
from the left subclavian artery take off to the IAB insertion 2 cm below the mid-inguinal ligament
Measurement 5
from the jugular notch to the middle point in the line between the anterior superior iliac spine and symphysis pubis
Depth
the depth of the femoral artery from the surface of the skin, 2 cm below the mid-inguinal ligament
Dthorax
Aortic diameter (at the middle of the thorax)
  
Drenal
Aortic diameter (at a renal level)
  
Dbifurc
Aortic diameter (bifurcation level)
  
Our Institutional Review Board approved this study. Individual consents were obtained for all cadaveric examinations and entry into the database.

Statistical Processing

Statistical analysis was performed with SPSS 17.0 statistical software (SPSS Inc., Chicago, IL, USA). Unpaired student t-test was applied to compare the various measurements between the two genders. Pearson r correlation coefficients were calculated to identify association between variables. Multiple linear regression models were used to predict the sizes of the aorta. The validation of the prediction for the balloon sizing was performed using discriminant analysis. The Cohen's Kappa test was employed to assess the agreement between the selections of the balloon size based on the results of the categorization by using the discriminant analysis model or using the "traditional" categorization (sizing as it is derived by the height of patients). The statistical significant level was set at 0.05

Results

There have been differences between the two genders with regard to the aortic lengths, with greater differences observed in the distance from the left subclavian artery to the aortic bifurcation. However there have been no differences in the comparison of the internal diameters, between the two genders. (Table 5)
Table 5
Gender comparison of external measurements, aortic lengths and diameters
 
male (n = 23)
female (n = 17)
 
mean
S.D
mean
S.D
Height(cm)*
171.09
7.98
160.97
9.20
Weight*
60.30
13.93
51.94
8.51
BSA*
1.70
0.20
1.28
0.14
BMI
20.50
3.93
20.10
3.25
Age
76.57
9.15
68.41
16.24
Measur1*
27.11
1.88
24.97
1.80
Measur2*
47.38
3.03
45.41
2.83
Measur3*
24.83
1.72
23.04
1.76
Measur4
45.10
2.88
43.53
2.75
Measur5*
51.61
3.31
48.92
2.88
Dthorax
26.24
2.53
25.15
3.25
Drenal
24.52
2.67
22.94
2.62
Dbifurc
18.70
2.88
18.38
2.06
LSA-CA*
24.41
2.14
22.68
1.84
LSA-Rena*
27.99
2.11
25.81
2.12
LSA-Bifur*
38.13
3.00
34.75
2.95
LSA-Inser*
53.57
3.43
50.84
2.71
Depth
2.01
0.49
2.12
0.37
*p < 0.05 (t-test)
The Table 6 and Figure 2 shows the rates for both genders, depending on the distribution of height. One can see that the majority of males is 162-182 cm high, whereas females are almost equally divided between (< 162 cm) and (162-182 cm).
Table 6
Height distribution between genders
Height
male (%)
fem (%)
< 162 cm
8.7
41.2
162-182 cm
87.0
58.8
> 182 cm
4.3
 
With regards to the differences in the aortic diameter above and below the renal arteries in patients < 162 cm, 162-182 cm and > 182 cm, a weak association is observed (r = 0.29, p = 0.07). In addition, this difference in the aortic diameter shows no variance among the three groups or the two genders.
The prediction of intra-aortic diameters was poor (R2 < 0.30). However the prediction of the internal measurements by using multiple regression analysis was satisfactory (R2 > 0.70) as per Table 7.
Table 7
Multiple regression model
 
Predictor variables
URC
SRC
p*
LSA-CA
Height
0.08
0.37
0.004
R2 = 0.71
Age
0.06
0.35
< 0.001
 
ŦGender
0.91
0.21
0.088
 
Measurement 1
0.63
0.61
< 0.001
 
Constant
-12.05
-
0.021
LSA-RENAL
Height
0.11
0.45
< 0.001
R2 = 0.70
Age
0.04
0.24
0.001
 
Measurement 1
0.48
0.43
< 0.001
 
Constant
-6.90
-
0.077
LSA-BIF
Height
0.23
0.66
< 0.001
R2 = 0.81
Age
0.05
0.19
0.017
 
Measurement 1
0.46
0.28
0.005
 
Constant
-16.99
-
< 0.001
LSA-INS
Height
0.25
0.71
< 0.001
R2 = 0.71
Measurement 5
0.23
0.23
0.029
 
Constant
-0.63
-
0.029
(R2 > 0.70)
URC: Unstandardized coefficient
SRC: Standardized coefficient
*Only the coefficients with a p < 0.10 are presented
Ŧmale = 1, female = 2
The selection of the balloon size in the traditional methods, depending on the height, shows relatively low consistency (although statistically significant) to the size that would have been selected if the actual length of the internal dimensions of the descending aorta was known. (Consistency test K = 0.38, p = 0.003).
On the contrary, the selection of the balloon size based on the results of the categorization by using the discriminant analysis model, shows impressive consistency with the sizes that would have been selected based on the actual aortic length (K = 1, p < 0.001).
In order to apply the results of the regression model to predict the length from the left Subclavian artery take off to the Celiac axis (LSA-CA) the following equation is formed:
https://static-content.springer.com/image/art%3A10.1186%2F1749-8090-6-95/MediaObjects/13019_2011_Article_459_Equa_HTML.gif
If LSA-CA < 21.9 cm use 34 cc IABP, LSA-CA = 21.9- 26.3 cm use 40 cc (may be "aortic length-balloon membrane length" mismatching) & if LSA-CA > 26.3 cm use 50 cc IABP.
As far as the depth of the femoral artery is concerned at the inguinal plane, there are no differences between the two genders (t Test) but there is an increased association between weight and depth (r = 0.49, p = 0.002).

Discussion

The efficiency of the IABP depends on multiple factors, which include: The ability of the ejection system to trace electrical stimulation events and transfer the gas during every cardiac cycle, the patient's hemodynamic status, the anatomical position of the balloon and lastly the volume of the selected balloon.
The variables that affect the size and duration of the IABP inflation with gas and consequently the diastolic increase in pressure, were first published 20 years ago, in an innovative study by Weber et al [3]. They reported that the arterial pressure has direct impact on the balloon performance and that the beneficial effects of the IABP reduces when arterial pressure increases.
The heart rate also affects the balloon functioning. The faster the heart rate, the less the balloon inflation time and the diastolic augmentation.
The relation between the aortic pressure and the volume of the aorta reflects the aortic elasticity and the reduced aortic elasticity means reduced diastolic augmentation.
The closer the IABP to the aortic valve, the higher is the incremented diastolic augmentation [4].
The effect of the IABP diameter to the patient's hemodynamics was studied by Weber, Janicki, and Walker [3]. They concluded that for any given arterial pressure or aortic size, the greatest augmentation in mean aortic diastolic pressure was achieved with complete occlusion. It is obvious that in 100% occlusion of the aortic lumen, a lesion to the aortic wall and erythrocytes could possibly occur and thus the ideal convergence is thought to be approximetly 90 to 95%.
The balloon volume shall be at least 50% of the stroke volume. In a hypothetical patient with poor cardiac index, the rate of the diastolic incrementation to support an insufficient heart is limited because of the fact that the stroke volume in the given patient is low and thus the ideal balloon volume shall be low. In the course of improvement of the hypothetical s patient s hemodynamics and the more the stroke volume improves, which means that the volume of blood transferred within the aorta at every stroke increases, then the balloon volume of the IABP, necessary to maintain the same level of diastolic augmentation, increments. This fact may suggest that during a period of time from the balloon implantation till the successful removal of the balloon, the performance of corrective interventions to the balloon volume by intervals, based on the patient's clinical status, may be the proper method in order to provide ideal diastolic augmentation, both for the insufficient heart and the heart that shows improvement.
To summarize up to now, the optimum balloon sizing shall be accompanied by adjustments of the balloon size and volume, which shall correspond to the different hemodynamic scenarios.
The equation to calculate the ideal balloon size for every individual patient, shall take into account the patient's hemodynamic state (Arterial pressure, Heart supply, pulses/min) as well as the measurements of the aortic length and diameter from the subclavian artery take off to the celiac axis.
Wiekel et al [5] after studying the aortographies of 169 patients, concluded that the corrected mid-thoracic diameters varied in size from 16 to 30 mm with 90% of the patients having a diameter over 19 mm. This has been possibly taken into consideration by the manufacturers and thus the balloon diastolic diameter is between 14.7 mm and 18.5 mm.
We reached the same conclusions in our measurements regarding the mid-thoracic diameters.
Paulin et al [6] suggested that the aorta's size is related to the patient's height, age and weight. After studying a series of aortic diameters, he suggested that the ascending aorta has a relatively fixed lumen, which varies from 22 to 38 mm in adults. The descending aorta has a slightly smaller lumen than the ascending aorta. In addition, the aorta's diameter narrows down after the renal artery take off.
In our measurements regarding the correlation between the difference of the aortic diameter above and below the renal arteries and the patients' height, a weak association resulted. In addition, this difference in the aortic diameter shows no variance between the two genders.
The balloon distal edge shall be "proximal" to the renal arteries' take off, so as to avoid occluding the renal arterial circulation. Swartz et al [7] studied the arterial dysfunction of the kidneys as a result of the position of the IABP on the take off of the renal arteries. He concluded that such a position of the balloon causes renal ischemia.
In our paper it became apparent that the female population consists of smaller dimensions of skeleton. The mean LSA-CA distance was found to be 22.68+/-1.84. In such a population, even a 34 cc IAB would lie below the take off of the renal arteries, with subsequent «aortic-device mismatch» and negative consequences.
During our study the traditional balloon sizing, based only on the patients height, shows relatively low consistency to the size that would have been selected if the actual length of the internal dimensions of the descending aorta was known. A report by Rastan et al [8] using CT scans identified IABP malpositions to be a common finding. Anatomic to balloon length mishmatch was found in 68.2% of the cases, with subsequently severe adverse effects.
Yosioka et al [9] reported that aortic-balloon mismatch could cause abdominal arterial branch obstruction. However, clinical reports of intra-abdominal ischemia due to anatomic-to-device length mismatch are limited [10, 11]. Furthermore, Cho et al [12] has published in a CT study a 29% rate of renal artery compromise in patient heights of 163 to 183 cm when using a 40-mL balloon size, especially in patients with a small stature. In our study, we confirm by enlarge, these findings: there is only a varying correlation between height and aortic length, which is influenced by age, sex, and possibly ethnic differences [12].
According to Alexander et al [13] the rate of the patients over 70 years old that undergo heart interventions, has raised from 7-9% in the 1980's to over 30% in the 2000's. The number of women in need of cardiac interventions has increased from 31% in 1982 to over 40% today. Moreover, a 3-5% of the patients currently live their eighth decade. Sisto et al [14] mention that an increased number of patients living their eighth decade and who are in need of an IABP support.
Taking into account that that mean age of patients that need a heart operation rises (and the age increases the risk of aortic atherosclerotic disease) and the increasing rate of the women patients population (having smaller aortic diameter) which need an IABP, one would conclude that a more meticulous size selection is required under such clinical conditions, otherwise the aortic environment could be hiding potential risks. Moreover, Wolvek et al [15] reported that 35.7% of female patients fall into the category of less than 62 in. where the aortic diameter narrows down from 20 mm over the renal arteries to 13 mm below the renal arteries. In such patients, a large IABP size is problematic since 5-6 cm of the IABP fills the "normally narrowed" part of the abdominal aorta (aortic-device mismatch).
The population that we studied was 72.8% male and 27.2% female. The average height of women was 160.97 cm. A 41.2% were less than 162 cm high. Contrary to other studies, our measurements did not reveal a great aortic physiological narrowing of the aorta below the renal arteries.
The aortic measurements in our study are related to the height of the patients and other somatometric measurements (age, sex, external measurement of the thoracic cage). Neither weight nor BSA affect the aortic length or diameter.
In our measurements, the depth of the femoral artery at the inguinal region did not differ between the two genders, though it was proportionate to the body weight.
The possible limitations of the study are related to the relative small number of subjects studied, the fact that most aortas were not atheromatic, so no corkscrew effects or angulations were present and lastly the relative small number of patients over 182 cm high.
Concluding, the guidelines according to which the appropriate size of the intra-aortic balloons are selected, are based on the estmation of the LSA-CA taking into consideration the patients height, age, sex and the easily obtainable distance from the jugular notch to «halfway between the jugular notch and the upper border of the pubic symphysis»; so if LSA-CA is < 21.09 cm use 34 cc IABP, 21.9-26.3 cm use 40 cc, but prepare for «aortic-device» mismatching and if > 26.3 cm use 50 cc.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

HP conceived of the idea, carried out all the measurements and wrote the manuscript. AS helped in organizing the data and also helped in the process of writing the manuscript. ML carried out the statistical analysis. DD supervised the project and offer valuable points and ideas. All authors read and approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.
Literatur
1.
Zurück zum Zitat Cox P, Kellett M, Goran S, Morton J, Wolvek S: Plaque abrasion and Intra-aortic Balloon Leak. Chest. 1995, 108 (6): 1495-98. 10.1378/chest.108.6.1495.CrossRefPubMed Cox P, Kellett M, Goran S, Morton J, Wolvek S: Plaque abrasion and Intra-aortic Balloon Leak. Chest. 1995, 108 (6): 1495-98. 10.1378/chest.108.6.1495.CrossRefPubMed
2.
Zurück zum Zitat Kantrowitz A: Origins of intraaortic balloon pumping. Ann Thorac Surg. 1990, 50 (4): 672-4. 10.1016/0003-4975(90)90220-Z.CrossRefPubMed Kantrowitz A: Origins of intraaortic balloon pumping. Ann Thorac Surg. 1990, 50 (4): 672-4. 10.1016/0003-4975(90)90220-Z.CrossRefPubMed
3.
Zurück zum Zitat Weber KT, Janicki JS: Intra-aortic balloon counterpulsation. A review of physiological principles, clinical results and device safety. Ann Thorac Surg. 1974, 17 (6): 602-36. 10.1016/S0003-4975(10)65706-2.CrossRefPubMed Weber KT, Janicki JS: Intra-aortic balloon counterpulsation. A review of physiological principles, clinical results and device safety. Ann Thorac Surg. 1974, 17 (6): 602-36. 10.1016/S0003-4975(10)65706-2.CrossRefPubMed
4.
Zurück zum Zitat Gundel WD, Brown BG, Gott VL: Coronary collateral flow studies during variable aortic root pressure waveforms. J Appl Physiol. 1970, 29 (5): 579-86.PubMed Gundel WD, Brown BG, Gott VL: Coronary collateral flow studies during variable aortic root pressure waveforms. J Appl Physiol. 1970, 29 (5): 579-86.PubMed
5.
Zurück zum Zitat Weikel AM, Jones RT, Dinsmore R, Petschek HE: Size limits and pumping effectiveness of intra-aortic balloons. Ann Thorac Surg. 1971, 12 (1): 45-53. 10.1016/S0003-4975(10)65092-8.CrossRefPubMed Weikel AM, Jones RT, Dinsmore R, Petschek HE: Size limits and pumping effectiveness of intra-aortic balloons. Ann Thorac Surg. 1971, 12 (1): 45-53. 10.1016/S0003-4975(10)65092-8.CrossRefPubMed
6.
Zurück zum Zitat Paulin S: Aortography. Cardiac catheterisation, angiography and intervention, Philadelphia. Edited by: Grossman W, Baim D. 1991, Lea & Febiger Paulin S: Aortography. Cardiac catheterisation, angiography and intervention, Philadelphia. Edited by: Grossman W, Baim D. 1991, Lea & Febiger
7.
Zurück zum Zitat Swartz MT, Sakamoto T, Arai H, Reedy JE, Salenas L, Yuda T: Effects of intraaortic balloon position on renal artery blood flow. Ann Thorac Surg. 1992, 53 (4): 604-10. 10.1016/0003-4975(92)90318-X.CrossRefPubMed Swartz MT, Sakamoto T, Arai H, Reedy JE, Salenas L, Yuda T: Effects of intraaortic balloon position on renal artery blood flow. Ann Thorac Surg. 1992, 53 (4): 604-10. 10.1016/0003-4975(92)90318-X.CrossRefPubMed
8.
Zurück zum Zitat Rastan A, Tillmann E, Subramanian S, Lehmkuhl L, Funkat A, Leontyev S: Visceral Arterial Compromise During Intra-Aortic Balloon Counterpulsation Therapy. Circulation. 2010, 122 (11 suppl): S92-S99.CrossRefPubMed Rastan A, Tillmann E, Subramanian S, Lehmkuhl L, Funkat A, Leontyev S: Visceral Arterial Compromise During Intra-Aortic Balloon Counterpulsation Therapy. Circulation. 2010, 122 (11 suppl): S92-S99.CrossRefPubMed
9.
Zurück zum Zitat Yoshioka Y, Saitoh K, Kubo E, Kitamura M, Nishida H, Endoh M: A new IABP balloon for Japanese to prevent the abdominal complication. Kyobu Geka. 1987, 40 (9): 736-742.PubMed Yoshioka Y, Saitoh K, Kubo E, Kitamura M, Nishida H, Endoh M: A new IABP balloon for Japanese to prevent the abdominal complication. Kyobu Geka. 1987, 40 (9): 736-742.PubMed
10.
Zurück zum Zitat Shin H, Yozu R, Sumida T, Kawada S: Acute ischemic hepatic failure resulting from intraaortic balloon pump malposition. Eur J Cardiothorac Surg. 2000, 17 (4): 492-494. 10.1016/S1010-7940(00)00366-3.CrossRefPubMed Shin H, Yozu R, Sumida T, Kawada S: Acute ischemic hepatic failure resulting from intraaortic balloon pump malposition. Eur J Cardiothorac Surg. 2000, 17 (4): 492-494. 10.1016/S1010-7940(00)00366-3.CrossRefPubMed
11.
Zurück zum Zitat Haga M, Sasajima T, Asada H, Azuma N, Morimoto N, Kubo Y: Extensive bowel infarction caused by intraaortic balloon pumping. Nippon Kyobu Geka Gakkai Zasshi. 1993, 41 (1): 93-96.PubMed Haga M, Sasajima T, Asada H, Azuma N, Morimoto N, Kubo Y: Extensive bowel infarction caused by intraaortic balloon pumping. Nippon Kyobu Geka Gakkai Zasshi. 1993, 41 (1): 93-96.PubMed
12.
Zurück zum Zitat Cho YS, Lim C, Han MJ, Chun EJ, Choi SI, Park KH: Should we consider the ethnic difference in selecting size of intraaortic balloon by commercial guideline?. ASAIO J. 2009, 55 (5): 519-522. 10.1097/MAT.0b013e3181ac6a35.CrossRefPubMed Cho YS, Lim C, Han MJ, Chun EJ, Choi SI, Park KH: Should we consider the ethnic difference in selecting size of intraaortic balloon by commercial guideline?. ASAIO J. 2009, 55 (5): 519-522. 10.1097/MAT.0b013e3181ac6a35.CrossRefPubMed
13.
Zurück zum Zitat Alexander KP, Anstrom KJ, Muhlbaier LH, Grosswald RD, Smith PK, Jones RH, Peterson ED: Outcomes of cardiac surgery in patients > or = 80 years: results from the National Cardiovascular Network. J Am Coll Cardiol. 2000, 35 (3): 731-738. 10.1016/S0735-1097(99)00606-3.CrossRefPubMed Alexander KP, Anstrom KJ, Muhlbaier LH, Grosswald RD, Smith PK, Jones RH, Peterson ED: Outcomes of cardiac surgery in patients > or = 80 years: results from the National Cardiovascular Network. J Am Coll Cardiol. 2000, 35 (3): 731-738. 10.1016/S0735-1097(99)00606-3.CrossRefPubMed
14.
Zurück zum Zitat Sisto DA, Hoffman DM, Fernandes S, Frater RW: Is Use of Intraaortic Balloon Pump in Octogenarians justified?. Ann Thorac Surg. 1992, 54 (3): 507-11. 10.1016/0003-4975(92)90444-9.CrossRefPubMed Sisto DA, Hoffman DM, Fernandes S, Frater RW: Is Use of Intraaortic Balloon Pump in Octogenarians justified?. Ann Thorac Surg. 1992, 54 (3): 507-11. 10.1016/0003-4975(92)90444-9.CrossRefPubMed
15.
Zurück zum Zitat Wolvek S: The hostile environment of the aging human aorta and the smaller patient-their implications for the intrapaortic balloon. Perfusion. 1994, 9 (2): 87-94. 10.1177/026765919400900202.CrossRefPubMed Wolvek S: The hostile environment of the aging human aorta and the smaller patient-their implications for the intrapaortic balloon. Perfusion. 1994, 9 (2): 87-94. 10.1177/026765919400900202.CrossRefPubMed
Metadaten
Titel
A statistical model that predicts the length from the left subclavian artery to the celiac axis; towards accurate intra aortic balloon sizing
verfasst von
Haralabos Parissis
Alan Soo
Michalis Leotsinidis
Dimitrios Dougenis
Publikationsdatum
01.12.2011
Verlag
BioMed Central
Erschienen in
Journal of Cardiothoracic Surgery / Ausgabe 1/2011
Elektronische ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-6-95

Weitere Artikel der Ausgabe 1/2011

Journal of Cardiothoracic Surgery 1/2011 Zur Ausgabe

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Real-World-Daten sprechen eher für Dupilumab als für Op.

14.05.2024 Rhinosinusitis Nachrichten

Zur Behandlung schwerer Formen der chronischen Rhinosinusitis mit Nasenpolypen (CRSwNP) stehen seit Kurzem verschiedene Behandlungsmethoden zur Verfügung, darunter Biologika, wie Dupilumab, und die endoskopische Sinuschirurgie (ESS). Beim Vergleich der beiden Therapieoptionen war Dupilumab leicht im Vorteil.

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.