Optimal pre-TAVR annulus sizing in patients with bicuspid aortic valve: area-derived perimeter by CT is the best-correlated measure with intraoperative sizing

verfasst von: Yuan Wang, Moyang Wang, Guanyuan Song, Wei Wang, Bin Lv, Hao Wang, Yongjian Wu

Erschienen in: European Radiology | Ausgabe 1/2019

Einloggen, um Zugang zu erhalten

Abstract

Objective

To clarify the optimal measurements for patients with bicuspid aortic valve (BAV) preferred for transcatheter aortic valve replacement (TAVR), our study compared intraoperative sizing with five different approaches by transthoracic echocardiography (TTE), three-dimensional transesophageal echocardiography (3DTEE) and computed tomography (CT).

Methods

We enrolled 104 BAV patients prescreened for TAVR but who underwent surgery with direct intraoperative annulus sizing. All five approaches [2DTTE, 3DTEE, area-derived perimeter (CTarea), perimeter-derived diameter (CTperi) and mean diameter (CTmean)] were compared with intraoperative sizing, respectively. Agreements on theoretical valve selections by five methods with those by intraoperative sizing were analyzed.

Results

CTarea showed the highest correlation (r = 0.932) and the best agreement with intraoperative sizing. Agreement for theoretical surgical and TAVR prosthesis selection was found in 84.6% and 74.0% BAVs by CTarea (κ = 0.791, κ = 0.585). CTperi-based prosthesis selection led to overestimation of 26.9% for surgical valves (κ = 0.589) and 36.5% for TAVR valves (κ = 0.425). Good correlations were observed between CT measurements and intraoperative sizing regardless of the predominant site of aortic valve calcification (r = 0.860-0.953).

Conclusion

The CTarea, which demonstrated the optimal approach to annulus sizing and prosthesis choice of BAVs with high eccentricity, should be included into the BAV-specific annulus sizing recommendation. The insufficiency of CTperi lay in overestimation of surgical or TAVR valve selections. Good agreement of 3DTEE sizing proved its superiority in annulus sizing for BAVs unsuitable for CT, but it should be used with caution for patients with a calcified annulus, where partial acoustic shadowing could lead to image inaccuracy.

Key Points

The area-derived perimeter by CT is the optimal approach to annulus sizing of BAVs.
The perimeter-derived approach is prone to overestimation of BAVs.
3DTEE showed its superiority in annulus sizing for BAVs unsuitable for CT, but it should be used with caution in patients with a calcified annulus.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Mack MJ, Leon MB, Smith CR et al (2015) 5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trial. Lancet 385:2477–2484CrossRef Mack MJ, Leon MB, Smith CR et al (2015) 5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trial. Lancet 385:2477–2484CrossRef
2.
Zurück zum Zitat Perlman GY, Blanke P, Dvir D et al (2016) Bicuspid aortic valve stenosis: favorable early outcomes with a next-generation transcatheter heart valve in a multicenter study. JACC Cardiovasc Interv 9:817–824CrossRef Perlman GY, Blanke P, Dvir D et al (2016) Bicuspid aortic valve stenosis: favorable early outcomes with a next-generation transcatheter heart valve in a multicenter study. JACC Cardiovasc Interv 9:817–824CrossRef
3.
Zurück zum Zitat Pontone G, Andreini D, Bartorelli AL et al (2012) Aortic annulus area assessment by multidetector computed tomography for predicting paravalvular regurgitation in patients undergoing balloon-expandable transcatheter aortic valve implantation: a comparison with transthoracic and transesophageal echocardiography. Am Heart J 164:576–584CrossRef Pontone G, Andreini D, Bartorelli AL et al (2012) Aortic annulus area assessment by multidetector computed tomography for predicting paravalvular regurgitation in patients undergoing balloon-expandable transcatheter aortic valve implantation: a comparison with transthoracic and transesophageal echocardiography. Am Heart J 164:576–584CrossRef
4.
Zurück zum Zitat Nishimura RA, Otto CM, Bonow RO et al (2014) 2014 AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Thorac Cardiovasc Surg 148:e1–e132CrossRef Nishimura RA, Otto CM, Bonow RO et al (2014) 2014 AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Thorac Cardiovasc Surg 148:e1–e132CrossRef
5.
Zurück zum Zitat Otto CM, Kumbhani DJ, Alexander KP et al (2017) 2017 ACC Expert Consensus Decision Pathway for Transcatheter Aortic Valve Replacement in the Management of Adults with Aortic Stenosis: A Report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents. J Am Coll Cardiol 69:1313–1346CrossRef Otto CM, Kumbhani DJ, Alexander KP et al (2017) 2017 ACC Expert Consensus Decision Pathway for Transcatheter Aortic Valve Replacement in the Management of Adults with Aortic Stenosis: A Report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents. J Am Coll Cardiol 69:1313–1346CrossRef
6.
Zurück zum Zitat Wijesinghe N, Ye J, Rodés-Cabau J et al (2010) Transcatheter aortic valve implantation in patients with bicuspid aortic valve stenosis. JACC Cardiovasc Interv 3:1122–1125CrossRef Wijesinghe N, Ye J, Rodés-Cabau J et al (2010) Transcatheter aortic valve implantation in patients with bicuspid aortic valve stenosis. JACC Cardiovasc Interv 3:1122–1125CrossRef
7.
Zurück zum Zitat Hayashida K, Bouvier E, Lefèvre T et al (2013) Transcatheter aortic valve implantation for patients with severe bicuspid aortic valve stenosis. Circ Cardiovasc Interv 6:284–291CrossRef Hayashida K, Bouvier E, Lefèvre T et al (2013) Transcatheter aortic valve implantation for patients with severe bicuspid aortic valve stenosis. Circ Cardiovasc Interv 6:284–291CrossRef
8.
Zurück zum Zitat Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80CrossRef Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80CrossRef
9.
Zurück zum Zitat Jilaihawi H, Chen M, Webb J et al (2016) A bicuspid aortic valve imaging classification for the TAVR Era. JACC Cardiovasc Imaging 9:1145–1158CrossRef Jilaihawi H, Chen M, Webb J et al (2016) A bicuspid aortic valve imaging classification for the TAVR Era. JACC Cardiovasc Imaging 9:1145–1158CrossRef
10.
Zurück zum Zitat Kerl JM, Ravenel JG, Nguyen SA et al (2008) Right heart: split-bolus injection of diluted contrast medium for visualization at coronary CT angiography. Radiology 247:356–364CrossRef Kerl JM, Ravenel JG, Nguyen SA et al (2008) Right heart: split-bolus injection of diluted contrast medium for visualization at coronary CT angiography. Radiology 247:356–364CrossRef
11.
Zurück zum Zitat Shin HJ, Shin JK, Chee HK et al (2015) Characteristics of aortic valve dysfunction and ascending aorta dimensions according to bicuspid aortic valve morphology. Eur Radiol 25:2103–2114CrossRef Shin HJ, Shin JK, Chee HK et al (2015) Characteristics of aortic valve dysfunction and ascending aorta dimensions according to bicuspid aortic valve morphology. Eur Radiol 25:2103–2114CrossRef
12.
Zurück zum Zitat Vaquerizo B, Spaziano M, Alali J et al (2016) Three-dimensional echocardiography vs. computed tomography for transcatheter aortic valve replacement sizing. Eur Heart J Cardiovasc Imaging 17:15–23PubMed Vaquerizo B, Spaziano M, Alali J et al (2016) Three-dimensional echocardiography vs. computed tomography for transcatheter aortic valve replacement sizing. Eur Heart J Cardiovasc Imaging 17:15–23PubMed
13.
Zurück zum Zitat Schnell S, Smith DA, Barker AJ et al (2016) Altered aortic shape in bicuspid aortic valve relatives influences blood flow patterns. Eur Heart J Cardiovasc Imaging 17:1239–1247CrossRef Schnell S, Smith DA, Barker AJ et al (2016) Altered aortic shape in bicuspid aortic valve relatives influences blood flow patterns. Eur Heart J Cardiovasc Imaging 17:1239–1247CrossRef
14.
Zurück zum Zitat Pham T, Martin C, Elefteriades J et al (2013) Biomechanical characterization of ascending aortic aneurysm with concomitant bicuspid aortic valve and bovine aortic arch. Acta Biomater 9:7927–7936CrossRef Pham T, Martin C, Elefteriades J et al (2013) Biomechanical characterization of ascending aortic aneurysm with concomitant bicuspid aortic valve and bovine aortic arch. Acta Biomater 9:7927–7936CrossRef
15.
Zurück zum Zitat Zhao ZG, Jilaihawi H, Feng Y et al (2015) Transcatheter aortic valve implantation in bicuspid anatomy. Nat Rev Cardiol 12:123–128CrossRef Zhao ZG, Jilaihawi H, Feng Y et al (2015) Transcatheter aortic valve implantation in bicuspid anatomy. Nat Rev Cardiol 12:123–128CrossRef
16.
Zurück zum Zitat Yano M, Nakamura K, Nagahama H et al (2012) Aortic annulus diameter measurement: what is the best modality? Ann Thorac Cardiovasc Surg 18:115–120CrossRef Yano M, Nakamura K, Nagahama H et al (2012) Aortic annulus diameter measurement: what is the best modality? Ann Thorac Cardiovasc Surg 18:115–120CrossRef
17.
Zurück zum Zitat Tamborini G, Fusini L, Muratori M et al (2014) Feasibility and accuracy of three-dimensional transthoracic echocardiography vs. multidetector computed tomography in the evaluation of aortic valve annulus in patient candidates to transcatheter aortic valve implantation. Eur Heart J Cardiovasc Imaging 15:1316–1323CrossRef Tamborini G, Fusini L, Muratori M et al (2014) Feasibility and accuracy of three-dimensional transthoracic echocardiography vs. multidetector computed tomography in the evaluation of aortic valve annulus in patient candidates to transcatheter aortic valve implantation. Eur Heart J Cardiovasc Imaging 15:1316–1323CrossRef
18.
Zurück zum Zitat Jurencak T, Turek J, Kietselaer BL et al (2015) MDCT evaluation of aortic root and aortic valve prior to TAVI. What is the optimal imaging time point in the cardiac cycle? Eur Radiol 25:1975–1983CrossRef Jurencak T, Turek J, Kietselaer BL et al (2015) MDCT evaluation of aortic root and aortic valve prior to TAVI. What is the optimal imaging time point in the cardiac cycle? Eur Radiol 25:1975–1983CrossRef
19.
Zurück zum Zitat Murphy DT, Blanke P, Alaamri S et al (2016) Dynamism of the aortic annulus: Effect of diastolic versus systolic CT annular measurements on device selection in transcatheter aortic valve replacement (TAVR). J Cardiovasc Comput Tomogr 10:37–43CrossRef Murphy DT, Blanke P, Alaamri S et al (2016) Dynamism of the aortic annulus: Effect of diastolic versus systolic CT annular measurements on device selection in transcatheter aortic valve replacement (TAVR). J Cardiovasc Comput Tomogr 10:37–43CrossRef
20.
Zurück zum Zitat Kempfert J, Van Linden A, Lehmkuhl L et al (2012) Aortic annulus sizing: echocardiographic versus computed tomography derived measurements in comparison with direct surgical sizing. Eur J Cardiothorac Surg 42:627–633CrossRef Kempfert J, Van Linden A, Lehmkuhl L et al (2012) Aortic annulus sizing: echocardiographic versus computed tomography derived measurements in comparison with direct surgical sizing. Eur J Cardiothorac Surg 42:627–633CrossRef
21.
Zurück zum Zitat Wang H, Hanna JM, Ganapathi A et al (2015) Comparison of aortic annulus size by transesophageal echocardiography and computed tomography angiography with direct surgical measurement. Am J Cardiol 115:1568–1573CrossRef Wang H, Hanna JM, Ganapathi A et al (2015) Comparison of aortic annulus size by transesophageal echocardiography and computed tomography angiography with direct surgical measurement. Am J Cardiol 115:1568–1573CrossRef
22.
Zurück zum Zitat Salaun E, Zenses AS, Evin M et al (2016) Effect of oversizing and elliptical shape of aortic annulus on transcatheter valve hemodynamics: an in vitro study. Int J Cardiol 208:28–35CrossRef Salaun E, Zenses AS, Evin M et al (2016) Effect of oversizing and elliptical shape of aortic annulus on transcatheter valve hemodynamics: an in vitro study. Int J Cardiol 208:28–35CrossRef
23.
Zurück zum Zitat Kim WK, Meyer A, Mollmann H et al (2016) Cyclic changes in area- and perimeter-derived effective dimensions of the aortic annulus measured with multislice computed tomography and comparison with metric intraoperative sizing. Clin Res Cardiol 105:622–629CrossRef Kim WK, Meyer A, Mollmann H et al (2016) Cyclic changes in area- and perimeter-derived effective dimensions of the aortic annulus measured with multislice computed tomography and comparison with metric intraoperative sizing. Clin Res Cardiol 105:622–629CrossRef
24.
Zurück zum Zitat Rixe J, Schuhbaeck A, Liebetrau C et al (2012) Multi-detector computed tomography is equivalent to trans-oesophageal echocardiography for the assessment of the aortic annulus before transcatheter aortic valve implantation. Eur Radiol 22:2662–2669CrossRef Rixe J, Schuhbaeck A, Liebetrau C et al (2012) Multi-detector computed tomography is equivalent to trans-oesophageal echocardiography for the assessment of the aortic annulus before transcatheter aortic valve implantation. Eur Radiol 22:2662–2669CrossRef
25.
Zurück zum Zitat Podlesnikar T, Prihadi EA, van Rosendael PJ et al (2018) Influence of the quantity of aortic valve calcium on the agreement between automated 3-dimensional transesophageal echocardiography and multidetector row computed tomography for aortic annulus sizing. Am J Cardiol 121:86-93 Podlesnikar T, Prihadi EA, van Rosendael PJ et al (2018) Influence of the quantity of aortic valve calcium on the agreement between automated 3-dimensional transesophageal echocardiography and multidetector row computed tomography for aortic annulus sizing. Am J Cardiol 121:86-93
27.
Zurück zum Zitat Ruile P, Blanke P, Krauss T et al (2016) Pre-procedural assessment of aortic annulus dimensions for transcatheter aortic valve replacement: comparison of a non-contrast 3D MRA protocol with contrast-enhanced cardiac dual-source CT angiography. Eur Heart J Cardiovasc Imaging 17:458–466CrossRef Ruile P, Blanke P, Krauss T et al (2016) Pre-procedural assessment of aortic annulus dimensions for transcatheter aortic valve replacement: comparison of a non-contrast 3D MRA protocol with contrast-enhanced cardiac dual-source CT angiography. Eur Heart J Cardiovasc Imaging 17:458–466CrossRef
28.
Zurück zum Zitat Schmauss D, Schmitz C, Bigdeli AK et al (2012) Three-dimensional printing of models for preoperative planning and simulation of transcatheter valve replacement. Ann Thorac Surg 93:31–33CrossRef Schmauss D, Schmitz C, Bigdeli AK et al (2012) Three-dimensional printing of models for preoperative planning and simulation of transcatheter valve replacement. Ann Thorac Surg 93:31–33CrossRef
29.
Zurück zum Zitat Patsalis PC, Al-Rashid F, Neumann T et al (2013) Preparatory balloon aortic valvuloplasty during transcatheter aortic valve implantation for improved valve sizing. JACC Cardiovasc Interv 6:965–971CrossRef Patsalis PC, Al-Rashid F, Neumann T et al (2013) Preparatory balloon aortic valvuloplasty during transcatheter aortic valve implantation for improved valve sizing. JACC Cardiovasc Interv 6:965–971CrossRef
Metadaten
Titel
Optimal pre-TAVR annulus sizing in patients with bicuspid aortic valve: area-derived perimeter by CT is the best-correlated measure with intraoperative sizing
verfasst von
Yuan Wang
Moyang Wang
Guanyuan Song
Wei Wang
Bin Lv
Hao Wang
Yongjian Wu
Publikationsdatum
20.06.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 1/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5592-y

Weitere Artikel der Ausgabe 1/2019

European Radiology 1/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.