Performance Evaluation of a Newly Developed MR-Compatible Mobile PET Scanner with Two Detector Layouts

verfasst von: Masao Watanabe, Yuji Nakamoto, Ryusuke Nakamoto, Takayoshi Ishimori, Tsuneo Saga, Kaori Togashi

Erschienen in: Molecular Imaging and Biology | Ausgabe 2/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

A mobile positron emission tomography (PET) scanner called flexible PET (fxPET), designed to fit existing magnetic resonance imaging (MRI) or computed tomography (CT) system, has been developed. The purpose of this study was to evaluate the image quality, lesion detection rate, and quantitative values of fxPET compared with conventional bismuth germanium oxide (BGO)-based PET/CT without time-of-flight capability.

Procedures

Fifty-nine patients underwent whole-body (WB) PET/CT scans approximately 1 h after injection of 2-deoxy-2-[18F]fluoro-D-glucose, followed by the fxPET scans with detectors located above and below the patients (layout A) and with detectors closer to the patients (layout B). Two readers assessed the image quality using a 4-point grade for each layout and reached a consensus. We evaluated the differences and/or correlations between fxPET and WB PET/CT, including the lesion detection rates, the standardized uptake value (SUV), the metabolic tumor volume (MTV), the total lesion glycolysis (TLG), the tumor-to-normal liver ratio (TLR), and the background liver signal-to-noise ratio (SNR).

Results

The image quality of layout B was better than layout A (p < 0.0001). Of 184 lesions, the detection rate of layout B was significantly higher than WB PET/CT (p = 0.041), while the detection rate of layout A was comparable to WB PET/CT. The SUVmax/mean/peak were larger, and the MTVs were smaller in fxPET than WB PET/CT, especially in the lesions smaller than 2 cm (p < 0.01). The SUVmax/mean/peak, the MTVs and the TLGs of fxPET had significant positive correlations with WB PET/CT (p < 0.0001). The TLRs were significantly larger (p < 0.0001), but the background SNRs were significantly lower in fxPET than WB PET/CT (p < 0.05).

Conclusions

The fxPET system yielded reasonable image quality and quantitative accuracy. Bringing the detectors closer to the patient yielded improved results.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Nensa F, Beiderwellen K, Heusch P, Wetter A (2014) Clinical applications of PET/MRI: current status and future perspectives. Diagn Interv Radiol 20:438–447CrossRef Nensa F, Beiderwellen K, Heusch P, Wetter A (2014) Clinical applications of PET/MRI: current status and future perspectives. Diagn Interv Radiol 20:438–447CrossRef
2.
Zurück zum Zitat Nakamoto R, Nakamoto Y, Ishimori T, Fushimi Y, Kido A, Togashi K (2018) Comparison of PET/CT with sequential PET/MRI using an MRI-compatible mobile PET system. J Nucl Med 59:846–851CrossRef Nakamoto R, Nakamoto Y, Ishimori T, Fushimi Y, Kido A, Togashi K (2018) Comparison of PET/CT with sequential PET/MRI using an MRI-compatible mobile PET system. J Nucl Med 59:846–851CrossRef
3.
Zurück zum Zitat Kobayashi T, Kitamura K (2012) Design considerations for a partial-ring, multi-modal compatible whole-body TOF PET scanner: flexible PET. in Conf Rec IEEE NSS/MIC 2807–2812 Kobayashi T, Kitamura K (2012) Design considerations for a partial-ring, multi-modal compatible whole-body TOF PET scanner: flexible PET. in Conf Rec IEEE NSS/MIC 2807–2812
4.
Zurück zum Zitat Nakazawa M, Ohi J, Furumiya T, et al. (2012) PET data acquisition (DAQ) system having scalability for the number of detector. in Conf Rec IEEE NSS/MIC 2475–2478 Nakazawa M, Ohi J, Furumiya T, et al. (2012) PET data acquisition (DAQ) system having scalability for the number of detector. in Conf Rec IEEE NSS/MIC 2475–2478
5.
Zurück zum Zitat Furuta M, Satoh M, Ohi J, et al. (2013) Development of a proof of concept system for multi-modal compatible PET: flexible PET. in Conf. Rec. IEEE NSS/MIC 1-4 Furuta M, Satoh M, Ohi J, et al. (2013) Development of a proof of concept system for multi-modal compatible PET: flexible PET. in Conf. Rec. IEEE NSS/MIC 1-4
6.
Zurück zum Zitat Yamakawa Y, Kobayashi T, Furuta M, et al. (2014) Development of a dual-head mobile DOI-TOF PET system having multi-modality compatibility. in Conf. Rec. IEEE NSS/MIC 1-3 Yamakawa Y, Kobayashi T, Furuta M, et al. (2014) Development of a dual-head mobile DOI-TOF PET system having multi-modality compatibility. in Conf. Rec. IEEE NSS/MIC 1-3
7.
Zurück zum Zitat Tanaka E, Kudo H (2010) Optimal relaxation parameters of DRAMA (dynamic RAMLA) aiming at one pass image reconstruction for 3D PET. Phys Med Biol 55:2917–2939CrossRef Tanaka E, Kudo H (2010) Optimal relaxation parameters of DRAMA (dynamic RAMLA) aiming at one pass image reconstruction for 3D PET. Phys Med Biol 55:2917–2939CrossRef
8.
Zurück zum Zitat Watson CC (2007) Extension of single scatter simulation to scatter correlation of time-of-flight PET. IEEE Trans on Nucl Sci 54:1679–1686CrossRef Watson CC (2007) Extension of single scatter simulation to scatter correlation of time-of-flight PET. IEEE Trans on Nucl Sci 54:1679–1686CrossRef
9.
Zurück zum Zitat Drzezga A, Souvatzoglou M, Eiber M, Beer AJ, Furst S, Martinez-Moller A, Nekolla SG, Ziegler S, Ganter C, Rummeny EJ, Schwaiger M (2012) First clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses. J Nucl Med 53:845–855CrossRef Drzezga A, Souvatzoglou M, Eiber M, Beer AJ, Furst S, Martinez-Moller A, Nekolla SG, Ziegler S, Ganter C, Rummeny EJ, Schwaiger M (2012) First clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses. J Nucl Med 53:845–855CrossRef
10.
Zurück zum Zitat Wahl RL, Jacene H, Kasamon Y, Lodge MA (2009) From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 50(Suppl 1):122S–150SCrossRef Wahl RL, Jacene H, Kasamon Y, Lodge MA (2009) From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 50(Suppl 1):122S–150SCrossRef
11.
Zurück zum Zitat Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, Verzijlbergen FJ, Barrington SF, Pike LC, Weber WA, Stroobants S, Delbeke D, Donohoe KJ, Holbrook S, Graham MM, Testanera G, Hoekstra OS, Zijlstra J, Visser E, Hoekstra CJ, Pruim J, Willemsen A, Arends B, Kotzerke J, Bockisch A, Beyer T, Chiti A, Krause BJ, European Association of Nuclear Medicine (EANM) (2015) FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging 42:328–354CrossRef Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, Verzijlbergen FJ, Barrington SF, Pike LC, Weber WA, Stroobants S, Delbeke D, Donohoe KJ, Holbrook S, Graham MM, Testanera G, Hoekstra OS, Zijlstra J, Visser E, Hoekstra CJ, Pruim J, Willemsen A, Arends B, Kotzerke J, Bockisch A, Beyer T, Chiti A, Krause BJ, European Association of Nuclear Medicine (EANM) (2015) FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging 42:328–354CrossRef
12.
Zurück zum Zitat Murphy JD, Chisholm KM, Daly ME, Wiegner EA, Truong D, Iagaru A, Maxim PG, Loo BW Jr, Graves EE, Kaplan MJ, Kong C, le QT (2011) Correlation between metabolic tumor volume and pathologic tumor volume in squamous cell carcinoma of oral cavity. Radiother Oncol 101:356–361CrossRef Murphy JD, Chisholm KM, Daly ME, Wiegner EA, Truong D, Iagaru A, Maxim PG, Loo BW Jr, Graves EE, Kaplan MJ, Kong C, le QT (2011) Correlation between metabolic tumor volume and pathologic tumor volume in squamous cell carcinoma of oral cavity. Radiother Oncol 101:356–361CrossRef
13.
Zurück zum Zitat Schwartz LH, Bogaerts J, Ford R, Shankar L, Therasse P, Gwyther S, Eisenhauer EA (2009) Evaluation of lymph nodes with RECIST 1.1. Eur J Cancer 45:261–267CrossRef Schwartz LH, Bogaerts J, Ford R, Shankar L, Therasse P, Gwyther S, Eisenhauer EA (2009) Evaluation of lymph nodes with RECIST 1.1. Eur J Cancer 45:261–267CrossRef
14.
Zurück zum Zitat Soret M, Bacharach SL, Buvat I (2007) Partial-volume effect in PET tumor imaging. J Nucl Med 48:932–945CrossRef Soret M, Bacharach SL, Buvat I (2007) Partial-volume effect in PET tumor imaging. J Nucl Med 48:932–945CrossRef
15.
Zurück zum Zitat Parvizi N, Franklin JM, McGowan DR et al (2015) Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colorectal liver metastases? Eur J Radiol 84:1873–1878CrossRef Parvizi N, Franklin JM, McGowan DR et al (2015) Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colorectal liver metastases? Eur J Radiol 84:1873–1878CrossRef
16.
Zurück zum Zitat Teoh EJ, McGowan DR, Bradley KM et al (2016) Novel penalised likelihood reconstruction of PET in the assessment of histologically verified small pulmonary nodules. Eur Radiol 26:576–584CrossRef Teoh EJ, McGowan DR, Bradley KM et al (2016) Novel penalised likelihood reconstruction of PET in the assessment of histologically verified small pulmonary nodules. Eur Radiol 26:576–584CrossRef
17.
Zurück zum Zitat van der Vos CS, Koopman D, Rijnsdorp S, Arends AJ, Boellaard R, van Dalen J, Lubberink M, Willemsen ATM, Visser EP (2017) Quantification, improvement, and harmonization of small lesion detection with state-of-art PET. Eur J Nucl Med Mol Imaging 44(Suppl 1):4–16PubMedPubMedCentral van der Vos CS, Koopman D, Rijnsdorp S, Arends AJ, Boellaard R, van Dalen J, Lubberink M, Willemsen ATM, Visser EP (2017) Quantification, improvement, and harmonization of small lesion detection with state-of-art PET. Eur J Nucl Med Mol Imaging 44(Suppl 1):4–16PubMedPubMedCentral
18.
Zurück zum Zitat Surti S, Karp JS (2016) Advances in time-of-flight PET. Phys Med 32:12–22CrossRef Surti S, Karp JS (2016) Advances in time-of-flight PET. Phys Med 32:12–22CrossRef
19.
Zurück zum Zitat Conti M (2011) Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging 38:1147–1157CrossRef Conti M (2011) Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging 38:1147–1157CrossRef
20.
Zurück zum Zitat Kadrmas DJ, Casey ME, Conti M, Jakoby BW, Lois C, Townsend DW (2009) Impact of time-of-flight on PET tumor detection. J Nucl Med 50:1315–1323CrossRef Kadrmas DJ, Casey ME, Conti M, Jakoby BW, Lois C, Townsend DW (2009) Impact of time-of-flight on PET tumor detection. J Nucl Med 50:1315–1323CrossRef
21.
Zurück zum Zitat Laffon E, Barret O, Marthan R, Ducassou D (2009) Is the physical decay correction of the 18F-FDG input function in dynamic PET imaging justified? J Nucl Med Technol 37:111–113CrossRef Laffon E, Barret O, Marthan R, Ducassou D (2009) Is the physical decay correction of the 18F-FDG input function in dynamic PET imaging justified? J Nucl Med Technol 37:111–113CrossRef
22.
Zurück zum Zitat Schmall JP, Karp JS, Werner M, Surti S (2016) Parallax error in long-axial field-of-view PET scanners-a simulation study. Phys Med Biol 61:5443–5455CrossRef Schmall JP, Karp JS, Werner M, Surti S (2016) Parallax error in long-axial field-of-view PET scanners-a simulation study. Phys Med Biol 61:5443–5455CrossRef
23.
Zurück zum Zitat Nakazawa M, Ohi J, Tonami H, et al. (2010) Development of a prototype DOI-TOF-PET scanner. in Conf. Rec. IEEE NSS/MIC 2077-2080 Nakazawa M, Ohi J, Tonami H, et al. (2010) Development of a prototype DOI-TOF-PET scanner. in Conf. Rec. IEEE NSS/MIC 2077-2080
24.
Zurück zum Zitat Kubota K, Itoh M, Ozaki K, Ono S, Tashiro M, Yamaguchi K, Akaizawa T, Yamada K, Fukuda H (2001) Advantage of delayed whole-body FDG-PET imaging for tumour detection. Eur J Nucl Med 28:696–703CrossRef Kubota K, Itoh M, Ozaki K, Ono S, Tashiro M, Yamaguchi K, Akaizawa T, Yamada K, Fukuda H (2001) Advantage of delayed whole-body FDG-PET imaging for tumour detection. Eur J Nucl Med 28:696–703CrossRef
25.
Zurück zum Zitat Murray I, Kalemis A, Glennon J, Hasan S, Quraishi S, Beyer T, Avril N (2010) Time-of-flight PET/CT using low-activity protocols: potential implications for cancer therapy monitoring. Eur J Nucl Med Mol Imaging 37:1643–1653CrossRef Murray I, Kalemis A, Glennon J, Hasan S, Quraishi S, Beyer T, Avril N (2010) Time-of-flight PET/CT using low-activity protocols: potential implications for cancer therapy monitoring. Eur J Nucl Med Mol Imaging 37:1643–1653CrossRef
26.
Zurück zum Zitat Laffon E, Adhoute X, de Clermont H, Marthan R (2011) Is liver SUV stable over time in 18F-FDG PET imaging? J Nucl Med Technol 39:258–263CrossRef Laffon E, Adhoute X, de Clermont H, Marthan R (2011) Is liver SUV stable over time in 18F-FDG PET imaging? J Nucl Med Technol 39:258–263CrossRef
27.
Zurück zum Zitat Al-Faham Z, Jolepalem P, Rydberg J et al (2016) Optimizing 18F-FDG uptake time before imaging improves the accuracy of PET/CT in liver lesions. J Nucl Med Technol 44:70–72CrossRef Al-Faham Z, Jolepalem P, Rydberg J et al (2016) Optimizing 18F-FDG uptake time before imaging improves the accuracy of PET/CT in liver lesions. J Nucl Med Technol 44:70–72CrossRef
28.
Zurück zum Zitat Surti S, Karp JS (2008) Design considerations for a limited angle, dedicated breast, TOF PET scanner. Phys Med Biol 53:2911–2921CrossRef Surti S, Karp JS (2008) Design considerations for a limited angle, dedicated breast, TOF PET scanner. Phys Med Biol 53:2911–2921CrossRef
Metadaten
Titel
Performance Evaluation of a Newly Developed MR-Compatible Mobile PET Scanner with Two Detector Layouts
verfasst von
Masao Watanabe
Yuji Nakamoto
Ryusuke Nakamoto
Takayoshi Ishimori
Tsuneo Saga
Kaori Togashi
Publikationsdatum
20.06.2019
Verlag
Springer International Publishing
Erschienen in
Molecular Imaging and Biology / Ausgabe 2/2020
Print ISSN: 1536-1632
Elektronische ISSN: 1860-2002
DOI
https://doi.org/10.1007/s11307-019-01384-9

Weitere Artikel der Ausgabe 2/2020

Molecular Imaging and Biology 2/2020 Zur Ausgabe

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Wie toxische Männlichkeit der Gesundheit von Männern schadet

08.04.2024 Andrologie Nachrichten

Stark, erfolgreich, allzeit belastbar – das sind Erwartungen, die Jungen und Männer von der Gesellschaft spüren. Das kann sie „toxisch“ werden lassen – und letztlich sogar der Gesundheit schaden, mahnt Dr. Dirk Sander von der Deutschen Aidshilfe.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.