Open Access 30.06.2020 | Scientific Article

Perfusion in hand arthritis on dynamic contrast-enhanced computed tomography: a randomized prospective study using MRI as a standard of reference

verfasst von: Sevtap Tugce Ulas, Kay Geert Hermann, Marcus R. Makowski, Robert Biesen, Fabian Proft, Ralph Schilling, Torsten Diekhoff

Erschienen in: Skeletal Radiology | Ausgabe 1/2021

Abstract

Objective

To evaluate the performance of dynamic contrast-enhanced CT (DCE-CT) in detecting and quantitatively assessing perfusion parameters in patients with arthritis of the hand compared with dynamic contrast-enhanced MRI (DCE-MRI) as a standard of reference.

Materials and methods

In this IRB-approved randomized prospective single-centre study, 36 consecutive patients with suspected rheumatoid arthritis underwent DCE-CT (320-row, tube voltage 80 kVp, tube current 8.25 mAs) and DCE-MRI (1.5 T) of the hand. Perfusion maps were calculated separately for mean transit time (MTT), time to peak (TTP), relative blood volume (rBV), and relative blood flow (rBF) using four different decomposition techniques. Region of interest (ROI) analysis was performed in metacarpophalangeal joints II–V and in the wrist. Pairs of perfusion parameters in DCE-CT and DCE-MRI were compared using a two-tailed t test for paired samples and interpreted for effect size (Cohen’s d). According to the Rheumatoid Arthritis Magnetic Resonance Imaging Score (RAMRIS) scoring results, differentiation of synovitis-positive and synovitis-negative joints with both modalities was assessed with the independent t test.

Results

The two modalities yielded similar perfusion parameters. Identified differences had small effects (d 0.01–0.4). DCE-CT additionally differentiates inflamed and noninflamed joints based on rBF and rBV but tends to underestimate these parameters in severe inflammation. The total dose-length product (DLP) was 48 mGy*cm with an estimated effective dose of 0.038 mSv.

Conclusion

DCE-CT is a promising imaging technique in arthritis. In patients with a contraindication to MRI or when MRI is not available, DCE-CT is a suitable alternative to detect and assess arthritis.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

The early detection of inflammatory changes such as synovitis is crucial to prevent the development of structural changes and bone destruction such as erosions [1]. Qualitative evaluation of inflammatory activity is performed using the American College of Rheumatology/European League Against Rheumatism collaborative initiative (ACR/EULAR) criteria [2]. Dynamic contrast-enhanced imaging techniques enable the assessment of synovial enhancement over time in patients with rheumatoid arthritis [35]. As a result, a comprehensive evaluation of inflammatory activity in rheumatic joint diseases is possible [68].
With dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI), it has become possible to evaluate synovitis and disease activity with sufficient accuracy [9] to quantify the response to therapy in research studies. Thus far, dynamic contrast-enhanced imaging has not been used widely to assess patients with arthritis in clinical practice. One reason is that sophisticated imaging technology and special software tools are required to derive and assess perfusion parameters. Ultrasonography can be offered as an alternative imaging test [10], but the results are often dependent on the experience of the examiner and difficult to quantify [11]. The first studies investigated the capability of computed tomography (CT) in the detection of active inflammation [12, 13]. Dynamic contrast-enhanced computed tomography (DCE-CT) has the potential to provide a differentiated estimation of inflammatory joint lesions through the assessment of perfusion parameters. Furthermore, it offers the superior detection of structural changes intrinsic to CT [1416]. The use of DCE-CT in clinical practice has become feasible due to dose reduction with low-kVp scanning protocols and the introduction of iterative reconstruction algorithms in recent years [17, 18].
The purpose of our study was to evaluate the potential of low-dose DCE-CT in the detection and evaluation of synovitis of the hand in patients with acute arthritis using DCE-MRI as the standard of reference. CT offers a sensitive alternative to MRI with lower costs, wider availability, shorter examination duration, and superior detection of erosions at the cost of radiation exposure.

Materials and methods

Subjects

In our randomized prospective single-centre study, we consecutively investigated 37 patients (according to a calculation of the necessary sample size) with suspected or proven diagnosis of rheumatoid arthritis (RA) and fulfilled the classification criteria of the American College of Rheumatology (ACR) [19]. The patients presented to the rheumatology inpatient or outpatient clinic of our institution from September 2016 to October 2017. Due to the request of the local ethics committee, all patients had to be over 50 years old. Exclusion criteria were contraindications to MRI, e.g. pacemaker or cochlear implants, known allergic reactions to contrast agent as well as patients with reduced kidney function (glomerular filtration rate (GFR) < 60 ml/min/1.73 m2) and hyperthyroidism. The final diagnosis was made by board-certified rheumatologists based on clinical (patient’s history, number of swollen and tender joints) and laboratory findings (C-reactive protein, rheumatoid factors (RF) and anti-citrullinated protein antibodies (ACPA)) as well as imaging findings of our study.
The study was approved by the local ethics committee (EA1/259/15) and the Federal Office for Radiation Protection (Z 5–22,462/2–2016-008). All patients gave their written informed consent.

DCE-CT and DCE-MRI procedures

All patients underwent DCE-CT and 1.5 T DCE-MRI of the clinically dominant hand and wrist. Both scans were performed on the same day with a maximum time interval of 1 h between them. Patients were randomized to DCE-CT first or DCE-MRI first.
Patients were scanned in a prone position with the hand stretched above the head into the scanner and positioning the palm on the table (so-called superman position). The arm was fixed with a sandbag to reduce motion. Contrast agent was administered into the contralateral arm at a flow of 2 ml/s using an intravenous needle (Braunüle®, 1.10 × 33 mm, G20, Vasofix®, B. Braun®; Melsungen, Germany) and an automatic injection device.

Dynamic contrast-enhanced CT acquisition protocol

DCE-CT was performed on a 320-row CT scanner (Canon Aquilion ONE Vision, Canon Medical Systems; Otawara, Japan) in volume mode with 16 cm z-axis coverage without table movement. We used a tube voltage of 80 kVp and a tube current of 30 mA with a rotation time of 0.275 s. A total of 10 imaging volumes were acquired. The scans were obtained 0, 10, 20, 30, 40, 50, 60, 90, 120, and 180 s after contrast injection. We administered a body-weight-adapted dose of 1 ml/kg iopromide (Ultravist 370®, Bayer®; Leverkusen, Germany) directly followed by a bolus of 30 ml sodium chloride. Images were reconstructed with a slice thickness of 0.5 mm using a medium soft tissue kernel without beam hardening compensation.

Dynamic contrast-enhanced MRI acquisition protocol

DCE-MRI was performed on a 1.5-T Siemens Magnetom Avanto imager (Siemens Healthcare; Erlangen, Germany) using a small flexible 4-channel coil (Siemens healthcare; Erlangen, Germany) for the hand.
Before contrast agent injection, we acquired the following sequences: T1-weighted spin-echo sequences in coronal planes and coronal T1-weighted turbo-inversion recovery-magnitude sequences (TIRM) (for details see Table 1).
Table 1
DCE-MRI protocol
 
FOV Read [mm]
FOV Phase [%]
Slice thickness [mm]
Number of Slices
TR [msec]
TE [msec]
Resolution Matrix
Flip Angle [degree]
T1
160
100
3
12
401
21
512 × 512
90
TIRM
160
100
3
16
5000
69
256 × 241
150
dVIBE
180
75
3
20
6.5
2.53
192 × 128
17
T1 fs
        
Coronal
200
100
3
14
719
11
512 × 256
150
Axial wrist
140
100
4
14
591
15
320 × 192
90
Axial MCP
140
100
4
12
507
15
320 × 192
90
FOV field of view, TR repetition time, TE echo time, TIRM turbo-inversion recovery-magnitude, SWI susceptibility-weighted imaging, VIBE volumetric interpolated breath-hold examination, T1 T1-weighted imaging, T1 fs fat-saturated contrast-enhanced T1 sequence
For perfusion analysis, we acquired coronal fat-saturated dynamic T1-weighted fast 3D gradient-echo volumetric interpolated breath-hold examination (dVIBE) sequences after contrast medium injection. VIBE sequences were obtained every 10 s over 3 min. Postcontrast images were acquired with fat-saturated T1-weighted spin-echo sequences in coronal and transverse slice orientations.
Contrast medium was diluted in isotonic sodium chloride solution at a ratio of 1:4, and an amount of 1 ml/kg was injected. This resulted in a total contrast agent volume of 0.2 ml/kg body weight of gadolinium-DOTA (Dotarem®; Guerbet, France) and sodium chloride volume of 0.8 ml/kg body weight. Thereafter, a bolus of 30-ml pure sodium chloride solution was administered.

Image reading

Prior to quantitative perfusion analysis, three independent readers (K.G.H. with 18 years, T.D. with 9 years, S.T.U. with 1 year of experience in musculoskeletal imaging) blinded to identifying information, clinical data, and other imaging findings scored the fat-saturated contrast-enhanced T1-weighted images for synovitis according to the Rheumatoid Arthritis Magnetic Resonance Imaging Score (RAMRIS) criteria, defined by the Outcome Measurement in Rheumatology (OMERACT) MRI group [20] to dichotomize the joints in inflamed (RAMRIS synovitis ≥ 1) versus noninflamed (RAMRIS synovitis = 0). The agreement of at least two readers was determined as a reference.

Image preprocessing

All images were pseudonymized with regard to name, age, and sex. Preprocessing for perfusion maps in DCE-MRI and DCE-CT was performed using an Olea sphere (Version 3.0, Olea Medical; La Ciotat, France). It was performed separately using four different decomposition methods: delay-sensitive and non-adaptive standard singular value decomposition (sSVD), delay-insensitive and non-adaptive circular singular value decomposition (cSVD), delay-insensitive and semi-adaptive oscillation index singular value decomposition (oSVD), and delay-insensitive and adaptive Bayesian decomposition. The singular value decomposition algorithms (SVD) have integrated an upstream threshold for the signal variance to reduce the effects of non-physiological oscillations [21]. The Bayesian technique enables a rigorous probabilistic estimation of haemodynamic parameters based on Bayes’ theorem [22]. In validated simulations, the Bayesian decomposition technique outperforms other techniques in terms of accuracy and robustness against noise [23].
Motion correction was applied automatically. To compare perfusion parameters derived from DCE-CT and DCE-MRI, the same ten time points were chosen to generate perfusion maps: 0, 10, 20, 30, 40, 50, 60, 90, 120, and 180 s after contrast agent administration. To calculate the arterial input function (AIF), a region of interest (ROI) was placed manually in the radial artery. The venous output function (VOF) was determined in an ROI manually placed in the vein of the back of the hand. The baseline was defined as the time point at which increasing signal intensity was detectable. Colour-coded perfusion maps with 3-mm slice thickness were calculated for the following parameters: time to peak (TTP), which describes the time to the peak of contrast uptake in the selected ROI, relative blood volume (rBV), relative blood flow (rBF), and mean transit time (MTT). Joints with severe synovitis show increased perfusion parameters, indicated by red colour in the perfusion maps, while blue and black indicate regions with lower perfusion parameters (see Fig. 1).

Image post-processing and regions of interest

ROI analysis of the computed perfusion maps was performed using Osirix (Version 7.0, Pixmeo SARL; Bernex, Switzerland). Image post-processing was conducted in a coronal orientation. According to the established RAMRIS criteria [20], the ROIs were placed in MCP II, MCP III, MCP IV, MCP V, radioulnar, radiocarpal, and intercarpal joints. ROIs were placed in the region with the highest contrast uptake in inflamed joints. In noninflamed joints, ROIs were placed on the radial side of the joints and on the radial mediocarpal joints for the intercarpal region. All ROIs were circular, and an area of 10 mm2 was used for evaluation. The same ROI positioning was used for all four decomposition techniques and the two imaging modalities. The mean value of each measurement in each ROI was noted.

Radiation exposure

The radiation exposure (estimated effective dose) of DCE-CT examinations was calculated using the overall dose-length product (DLP) and a conversion coefficient of 0.0008 [mSv × mGy−1 × cm−1].

Statistical analysis

Statistical analysis was performed using R (RStudio, Version 1.1.463). The results of ROI analysis were collected and summarized for the respective perfusion parameters. The mean and standard deviation were calculated for each perfusion parameter and decomposition technique separately. Negative values of ROI analysis due to artefacts were excluded from statistical analysis. To identify significant differences between perfusion parameter pairs in DCE-CT and DCE-MRI, we performed a two-tailed paired t test. The results of the paired t test were interpreted based on the effect size using Cohen’s d with the following classification: no effect < 0.1, small 0.2–0.4, medium 0.5–0.7, large 0.8–≥ 1.0 [24]. Furthermore, means, standard deviations, and 95% confidence intervals were used for interpretation of differences. In a second step, an independent two-tailed t test was used to investigate whether DCE-CT is able to differentiate between inflamed and noninflamed joints with a performance similar to that of DCE-MRI (according to scoring results). A p value < 0.05 was considered statistically significant.
For assessment of agreement between DCE-CT and DCE-MRI, Bland-Altman plots of TTP, rBV, rBF, and MTT were created for inflamed joints separately for each of the four decomposition techniques.

Results

Subjects and image reading

The study included 37 patients (27 women; mean age 60.1, 50–77 years) with a mean weight of 77.3 kg (SD 14.3), a mean C-reactive protein (CRP) of 18 mg/dl (SD 42.6), and a mean duration of symptoms of 3.9 years (SD 4.84). In one patient, DCE-MRI could not be performed due to technical problems during the examination and was therefore excluded from further statistical analysis. The study population and RAMRIS scores are presented in Table 2 and Fig. 2. ACPA was positive in 12 patients (33.3%), while this was the case for RF in 8 patients (22.25%). Fourteen patients were never treated, three were treated with NSAIDs, 6 received only corticosteroids, 7 received disease-modifying antirheumatic drugs (DMARDs), and 6 received biologicals. No patient received a corticosteroid injection prior to the examination.
Table 2
Characteristics of the study population
Characteristics
Total
RA
Non-RA
No. of participants
36
24
12
Women
26
17
9
Men
10
7
3
Age (y)
60.4 ± 7.1 (50–77)
60.4 ± 7.2 (50–77)
58.6 ± 5.6 (52–68)
Women
59.2 ± 6.2 (50–74)
58.8 ± 6.6 (50–74)
60.0 ± 5.7 (52–68)
Men
63.4 ± 8.6 (52–77)
67.3 ± 6.9 (56–77)
54.3 ± 3.2 (52–58)
Weight [kg]
77.6 ± 14.4
79.7 ± 15.3
73.8 ± 12.0
Symptom duration (y)
4.2 ± 4.9
3.8 ± 4.7
5.0 ± 5.4
CRP [mg/l]
15.9 ± 40.3
21.9 ± 47.0
1.7 ± 1.6
No. of ACPA-positive
12
11
1
No. of RF-positive
8
8
0
The values are given as follows: means ± standard deviation and range in parentheses. RA rheumatoid arthritis, CRP C-reactive protein, ACPA anti-citrullinated protein antibodies, RF rheumatoid factors, ACPA-positive was defined as ACPA > 20 U/ml and RF-positive as RF > 14 U/ml)
Twenty-one patients underwent DCE-CT before DCE-MRI. According to RAMRIS scores, 50.8% (128/252) of the regions analysed were positive for synovitis (see Fig. 2).

Image preprocessing

In all patients, perfusion maps of TTP, rBV, rBF, and MTT were computed separately for preprocessing with sSVD, cSVD, oSVD, and the Bayesian algorithm. All data were accepted and successfully post-processed by the software. The post-processing and quantitative measurements took approximately 10 min per patient, modality, and reconstruction algorithm. These time, investments add up to approximately 48 h of continuous work for the whole study population. Examples of perfusion maps are shown in Fig. 1. In Fig. 3, boxplots of perfusion parameters of the different RAMRIS gradings are presented.

Image post-processing and regions of interest

Means, standard deviations, t test results, and effect sizes as well as differences in means and standard deviations of differences and 95% confidence intervals of DCE-CT and DCE-MRI for TTP, rBV, rBF, and MTT using the different decomposition techniques are presented in Table 3. Overall, our results revealed no significant differences between perfusion parameters determined with the two modalities, particularly for rBF in sSVD and TTP in oSVD. The observed differences were of small effect (d = 0.01–0.4). The results indicate that DCE-CT is able to distinguish between inflamed and noninflamed joints (Table 4).
Table 3
Results of the paired two-tailed t-test
Decomp.
Param.
MRI-derived mean
MRI-derived SD
CT-derived mean
CT-derived SD
Mean of diff.
SD of diff.
95% CI of diff.
p value
Effect size (Cohen’s d)
sSVD
TTP
116.2
32.7
110.1
32.4
6.1
40.6
0.9 to 11.2
0.019
0.2
 
rBV
24.5
29.1
19.1
20.1
5.4
25.3
2.2 to 8.5
0.001
0.2
 
rBF
43.7
38.9
44.5
53.4
− 0.7
60.0
− 8.2 to 6.8
0.845
0.01
 
MTT
37.5
22.0
34.0
19.9
3.5
23.6
0.5 to 6.5
0.020
0.2
cSVD
TTP
110.3
30.4
103.6
32.4
6.7
38.4
1.9 to 11.5
0.007
0.2
 
rBV
22.4
26.0
18.8
19.9
3.6
23.3
0.7 to 6.5
0.015
0.2
 
rBF
36.4
30.9
32.6
34.8
3.8
43.9
− 1.7 to 9.2
0.176
0.09
 
MTT
28.8
15.0
32.3
20.5
− 3.5
19.2
− 5.9 to − 1.1
0.004
0.2
oSVD
TTP
109.8
36.1
108.6
33.8
1.2
42.5
− 4.0 to 6.5
0.644
0.03
 
rBV
22.7
25.6
19.6
20.9
3.1
23.9
0.1 to 6.1
0.044
0.1
 
rBF
307.3
248.2
1066.6
1480.8
− 759.2
1387.6
− 932.4 to − 586.0
< 0.001
0.5
 
MTT
3.3
2.7
4.0
4.3
− 0.6
5.1
− 1.3 to 0.0005
0.050
0.1
Bay
TTP
48.7
8.9
63.6
12.7
− 14.9
13.5
− 16.6 to − 13.2
< 0.001
1.1
 
rBV
20.3
25.9
23.2
50.3
− 2.9
53.9
− 9.6 to 3.9
0.401
0.05
 
rBF
112.7
92.3
146.3
452.8
− 33.6
459.3
− 90.8 to 23.6
0.248
0.07
 
MTT
17.6
11.9
12.4
9.2
5.2
13.9
3.4 to 6.9
< 0.001
0.4
Decomp. decomposition technique, Param. perfusion parameter, SD standard deviation, diff. difference, 95% CI 95% confidence interval of the difference, sSVD standard singular value decomposition, cSVD circular singular value decomposition, oSVD oscillar singular value decomposition, Bay Bayesian decomposition
Table 4
Results of the independent t-test
Decomp.
Param.
+MRI mean ± SD
-MRI mean ± SD
MRI-derived p value
+CT mean ± SD
-CT mean ± SD
CT-derived p value
sSVD
TTP
126.2 ± 32.0
105.5 ± 29.9
<0.001
116.5 ± 31.7
103.4 ± 31.8
0.001
 
rBV
36.9 ± 34.5
10.9 ± 11.6
<0.001
23.7 ± 22.3
14.2 ± 16.1
<0.001
 
rBF
55.4 ± 45.0
31.0 ± 26.0
<0.001
44.6 ± 33.3
44.4 ± 68.8
0.982
 
MTT
47.4 ± 21.1
26.6 ± 17.4
<0.001
40.7 ± 20.4
26.7 ± 16.5
<0.001
cSVD
TTP
120.9 ± 29.4
99.2 ± 27.4
<0.001
109.5 ± 31.6
97.4 ± 32.1
0.003
 
rBV
34.2 ± 30.5
9.8 ± 10.2
<0.001
23.7 ± 23.1
13.7 ± 14.2
<0.001
 
rBF
48.5 ± 36.1
23.6 ± 16.9
<0.001
31.2 ± 24.1
34.1 ± 43.4
0.522
 
MTT
35.9 ± 14.9
21.3 ± 11.1
<0.001
40.2 ± 21.4
24.0 ± 15.8
<0.001
oSVD
TTP
123.0 ± 35.6
96.0 ± 31.3
<0.001
116.5 ± 34.2
100.3 ± 31.5
<0.001
 
rBV
34.3 ± 29.9
10.3 ± 10.5
<0.001
24.5 ± 23.6
14.4 ± 16.2
<0.001
 
rBF
372.9 ± 254.1
246.5 ± 239.0
<0.001
1397.8 ± 1809.4
716.1 ± 908.8
<0.001
 
MTT
4.3 ± 3.2
2.3 ± 1.4
<0.001
3.9 ± 4.9
4.0 ± 3.7
0.960
Bay
TTP
49.7 ± 8.7
47.7 ± 9.0
0.065
66.0 ± 12.4
61.1 ± 12.6
0.003
 
rBV
27.7 ± 30.9
12.6 ± 16.0
<0.001
22.5 ± 21.5
24.0 ± 68.8
0.818
 
rBF
107.1 ± 70.3
119.2 ± 110.0
0.296
96.5 ± 185.7
198.7 ± 616.7
0.074
 
MTT
20.5 ± 14.1
14.7 ± 7.8
<0.001
14.9 ± 10.2
9.9 ± 7.2
<0.001
Decomp. decomposition technique, Param. perfusion parameter, +MR/+CT results for inflamed joints in MRI und CT, −MRI/-CT results for noninflamed joints, SD standard deviation, sSVD standard singular value decomposition, cSVD circular singular value decomposition, oSVD oscillar singular value decomposition, Bay Bayesian decomposition
Figure 4 presents Bland-Altman plots for TTP, rBV, rBF, and MTT in regions with synovitis for each of the four decomposition techniques, namely, sSVD, oSVD, cSVD, and Bayesian technique. The plots show that joints with severe inflammation tended to be underestimated in DCE-CT.

Radiation exposure

The total DLP was 48 mGy*cm, resulting in an estimated effective dose of 0.038 mSv.

Discussion

This is the first study investigating the innovative technique of low-dose DCE-CT for the assessment of synovitis of the hand in patients with active arthritis using four different decomposition techniques. We show that DCE-CT is able to differentiate between inflamed and noninflamed joints and between the different grades of the semi-quantitative RAMRIS synovitis scores with no major differences between the four decomposition techniques studied. While DCE-CT tends to underestimate severe inflammation, it provides similar perfusion parameters as DCE-MRI and is able to identify joints with synovitis. In this regard, rBF shows the best agreement between the two modalities. Differences in perfusion parameters identified between the two modalities have small effects.
DCE-CT is more sensitive than MRI or traditional radiography in detecting soft tissue calcification and smaller erosions. Thus, CT allows a broader differential diagnosis. Due to the upcoming ultra-low-dose protocols [25], CT has the potential to be applied in routine care. DCE-CT also allows automated image post-processing and reading. Furthermore, it also offers a gadolinium-free alternative, as MRI contrast media can result in gadolinium deposition with as yet unclear effects [26, 27]. Additionally, CT is faster and more readily available. In our previous study, we showed that patients prefer CT than MRI and feel more comfortable in a CT scanner [13]. For comparability with MRI, CT examinations in our study were performed in the superman position as well. In the clinical setting, CT examinations can be performed with the patient sitting comfortably next to the gantry.
Sonography has several advantages in the detection of synovial inflammation using power-Doppler with a high diagnostic ability in depicting erosions, synovial hypertrophy, and hyperaemia [28]. However, its diagnostic performance varies with the examiner’s experience [11], and the quantification of inflammation and structural lesions for follow-up is limited [29]. For this reason, a comparison of sonography with DCE-CT and DCE-MRI is unfeasible. Modern CT techniques allow the depiction of bone marrow oedema [30]. DCE and dual-energy CT might help to detect osteitis, and the combined use of both techniques may provide further benefits.
Our study was specifically planned to compare DCE-CT and DCE-MRI. We used adapted doses of contrast agent in DCE-CT and DCE-MRI to create comparable conditions. The same software was used to generate perfusion maps in DCE-CT and DCE-MRI and for quantitative ROI analysis. However, our study has some limitations. The different inherent properties of the imaging modalities and contrast agents must be taken into account [31]. We investigated only 36 patients, adhering to our estimate of the necessary sample size. We are aware that, despite showing the absence of significant differences in perfusion parameter means between CT and MRI, our study is not designed to establish equivalence or non-inferiority. Gadolinium retention in patients who underwent MRI first might have influenced the CT enhancement. Using a low-dose CT protocol can both limit the ability to detect subtle bone changes and lead to measurement artefacts that can be minimized by applying higher radiation to improve the contrast-to-noise ratio. According to the new EULAR recommendation, MRI is no longer used for the early diagnosis of RA because of its low specificity [32]. Further studies investigating DCE-CT using an optimized image reconstruction protocol (e.g. artificial intelligence) in larger numbers of patients are needed. In addition, a comparison with 3 T scanners would be of particular interest.
DCE-CT has the potential to assess synovitis. Using four decomposition techniques, we showed that there is overall good agreement between DCE-CT and DCE-MRI, while DCE-CT tends to underestimate some perfusion parameters. In patients with a contraindication to MRI or when MRI is not available, DCE-CT is a suitable alternative to detect and assess arthritis.
In conclusion, low-dose DCE-CT can differentiate inflamed and noninflamed joints in patients with suspected RA. Additionally, DCE-CT produces perfusion parameters comparable with DCE-MRI.

Acknowledgements

The authors thank Mrs. Bettina Herwig for language editing. Ralph Schilling kindly provided statistical advice for this manuscript.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflicts of interest. The study was supported by Canon Medical Systems. However, the company did not influence or authorize the study results.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent was obtained from all individual participants included in the study.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
2.
Zurück zum Zitat Kay J, Upchurch KS. ACR/EULAR 2010 rheumatoid arthritis classification criteria. Rheumatology (Oxford). 2012;51(Suppl 6):vi5–9.CrossRef Kay J, Upchurch KS. ACR/EULAR 2010 rheumatoid arthritis classification criteria. Rheumatology (Oxford). 2012;51(Suppl 6):vi5–9.CrossRef
3.
Zurück zum Zitat Axelsen MB, Poggenborg RP, Stoltenberg M, Kubassova O, Boesen M, Horslev-Petersen K, et al. Reliability and responsiveness of dynamic contrast-enhanced magnetic resonance imaging in rheumatoid arthritis. Scand J Rheumatol. 2013;42(2):115–22.PubMedCrossRef Axelsen MB, Poggenborg RP, Stoltenberg M, Kubassova O, Boesen M, Horslev-Petersen K, et al. Reliability and responsiveness of dynamic contrast-enhanced magnetic resonance imaging in rheumatoid arthritis. Scand J Rheumatol. 2013;42(2):115–22.PubMedCrossRef
4.
Zurück zum Zitat Wojciechowski W, Tabor Z, Urbanik A. Assessing synovitis based on dynamic gadolinium-enhanced MRI and EULAR-OMERACT scores of the wrist in patients with rheumatoid arthritis. Clin Exp Rheumatol. 2013;31(6):850–6.PubMed Wojciechowski W, Tabor Z, Urbanik A. Assessing synovitis based on dynamic gadolinium-enhanced MRI and EULAR-OMERACT scores of the wrist in patients with rheumatoid arthritis. Clin Exp Rheumatol. 2013;31(6):850–6.PubMed
5.
Zurück zum Zitat Deniffel D, Boutelier T, Labani A, Ohana M, Pfeiffer D, Roy C. Computed tomography perfusion measurements in renal lesions obtained by Bayesian estimation, advanced singular-value decomposition deconvolution, maximum slope, and Patlak models: intermodel agreement and diagnostic accuracy of tumor classification. Investig Radiol. 2018;53(8):477–85.CrossRef Deniffel D, Boutelier T, Labani A, Ohana M, Pfeiffer D, Roy C. Computed tomography perfusion measurements in renal lesions obtained by Bayesian estimation, advanced singular-value decomposition deconvolution, maximum slope, and Patlak models: intermodel agreement and diagnostic accuracy of tumor classification. Investig Radiol. 2018;53(8):477–85.CrossRef
6.
Zurück zum Zitat Axelsen MB, Ejbjerg BJ, Hetland ML, Skjodt H, Majgaard O, Lauridsen UB, et al. Differentiation between early rheumatoid arthritis patients and healthy persons by conventional and dynamic contrast-enhanced magnetic resonance imaging. Scand J Rheumatol. 2014;43(2):109–18.PubMedCrossRef Axelsen MB, Ejbjerg BJ, Hetland ML, Skjodt H, Majgaard O, Lauridsen UB, et al. Differentiation between early rheumatoid arthritis patients and healthy persons by conventional and dynamic contrast-enhanced magnetic resonance imaging. Scand J Rheumatol. 2014;43(2):109–18.PubMedCrossRef
7.
Zurück zum Zitat Navalho M, Resende C, Rodrigues AM, Gaspar A, Fonseca JE, Canhao H, et al. Dynamic contrast-enhanced 3-T magnetic resonance imaging: a method for quantifying disease activity in early polyarthritis. Skelet Radiol. 2012;41(1):51–9.CrossRef Navalho M, Resende C, Rodrigues AM, Gaspar A, Fonseca JE, Canhao H, et al. Dynamic contrast-enhanced 3-T magnetic resonance imaging: a method for quantifying disease activity in early polyarthritis. Skelet Radiol. 2012;41(1):51–9.CrossRef
8.
Zurück zum Zitat van de Sande MG, van der Leij C, Lavini C, Wijbrandts CA, Maas M, Tak PP. Characteristics of synovial inflammation in early arthritis analysed by pixel-by-pixel time-intensity curve shape analysis. Rheumatology (Oxford). 2012;51(7):1240–5.CrossRef van de Sande MG, van der Leij C, Lavini C, Wijbrandts CA, Maas M, Tak PP. Characteristics of synovial inflammation in early arthritis analysed by pixel-by-pixel time-intensity curve shape analysis. Rheumatology (Oxford). 2012;51(7):1240–5.CrossRef
9.
Zurück zum Zitat Waterton JC, Ho M, Nordenmark LH, Jenkins M, DiCarlo J, Guillard G, et al. Repeatability and response to therapy of dynamic contrast-enhanced magnetic resonance imaging biomarkers in rheumatoid arthritis in a large multicentre trial setting. Eur Radiol. 2017;27(9):3662–8.PubMedPubMedCentralCrossRef Waterton JC, Ho M, Nordenmark LH, Jenkins M, DiCarlo J, Guillard G, et al. Repeatability and response to therapy of dynamic contrast-enhanced magnetic resonance imaging biomarkers in rheumatoid arthritis in a large multicentre trial setting. Eur Radiol. 2017;27(9):3662–8.PubMedPubMedCentralCrossRef
10.
Zurück zum Zitat Kawashiri SY, Suzuki T, Nakashima Y, Horai Y, Okada A, Nishino A, et al. Synovial inflammation assessed by ultrasonography correlates with MRI-proven osteitis in patients with rheumatoid arthritis. Rheumatology (Oxford). 2014;53(8):1452–6.CrossRef Kawashiri SY, Suzuki T, Nakashima Y, Horai Y, Okada A, Nishino A, et al. Synovial inflammation assessed by ultrasonography correlates with MRI-proven osteitis in patients with rheumatoid arthritis. Rheumatology (Oxford). 2014;53(8):1452–6.CrossRef
11.
Zurück zum Zitat Hoving JL, Buchbinder R, Hall S, Lawler G, Coombs P, McNealy S, et al. A comparison of magnetic resonance imaging, sonography, and radiography of the hand in patients with early rheumatoid arthritis. J Rheumatol. 2004;31(4):663–75.PubMed Hoving JL, Buchbinder R, Hall S, Lawler G, Coombs P, McNealy S, et al. A comparison of magnetic resonance imaging, sonography, and radiography of the hand in patients with early rheumatoid arthritis. J Rheumatol. 2004;31(4):663–75.PubMed
12.
Zurück zum Zitat Fukuda T, Umezawa Y, Tojo S, Yonenaga T, Asahina A, Nakagawa H, et al. Initial experience of using dual-energy CT with an iodine overlay image for hand psoriatic arthritis: comparison study with contrast-enhanced MR imaging. Radiology. 2017;161671. Fukuda T, Umezawa Y, Tojo S, Yonenaga T, Asahina A, Nakagawa H, et al. Initial experience of using dual-energy CT with an iodine overlay image for hand psoriatic arthritis: comparison study with contrast-enhanced MR imaging. Radiology. 2017;161671.
13.
Zurück zum Zitat Diekhoff T, Ulas ST, Poddubnyy D, Schneider U, Hermann S, Biesen R, et al. Ultra-low-dose CT detects synovitis in patients with suspected rheumatoid arthritis. Ann Rheum Dis. 2019;78(1):31–5.PubMedCrossRef Diekhoff T, Ulas ST, Poddubnyy D, Schneider U, Hermann S, Biesen R, et al. Ultra-low-dose CT detects synovitis in patients with suspected rheumatoid arthritis. Ann Rheum Dis. 2019;78(1):31–5.PubMedCrossRef
14.
Zurück zum Zitat Dohn UM, Ejbjerg BJ, Hasselquist M, Narvestad E, Moller J, Thomsen HS, et al. Detection of bone erosions in rheumatoid arthritis wrist joints with magnetic resonance imaging, computed tomography and radiography. Arthritis Res Ther. 2008;10(1):R25.PubMedPubMedCentralCrossRef Dohn UM, Ejbjerg BJ, Hasselquist M, Narvestad E, Moller J, Thomsen HS, et al. Detection of bone erosions in rheumatoid arthritis wrist joints with magnetic resonance imaging, computed tomography and radiography. Arthritis Res Ther. 2008;10(1):R25.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Topfer D, Finzel S, Museyko O, Schett G, Engelke K. Segmentation and quantification of bone erosions in high-resolution peripheral quantitative computed tomography datasets of the metacarpophalangeal joints of patients with rheumatoid arthritis. Rheumatology (Oxford). 2014;53(1):65–71.CrossRef Topfer D, Finzel S, Museyko O, Schett G, Engelke K. Segmentation and quantification of bone erosions in high-resolution peripheral quantitative computed tomography datasets of the metacarpophalangeal joints of patients with rheumatoid arthritis. Rheumatology (Oxford). 2014;53(1):65–71.CrossRef
16.
Zurück zum Zitat de Bucourt M, Scheurig-Munkler C, Feist E, Juran R, Diekhoff T, Rogalla P, et al. Cyst-like lesions in finger joints detected by conventional radiography: comparison with 320-row multidetector computed tomography. Arthritis Rheum. 2012;64(4):1283–90.PubMedCrossRef de Bucourt M, Scheurig-Munkler C, Feist E, Juran R, Diekhoff T, Rogalla P, et al. Cyst-like lesions in finger joints detected by conventional radiography: comparison with 320-row multidetector computed tomography. Arthritis Rheum. 2012;64(4):1283–90.PubMedCrossRef
17.
Zurück zum Zitat Vardhanabhuti V, Riordan RD, Mitchell GR, Hyde C, Roobottom CA. Image comparative assessment using iterative reconstructions: clinical comparison of low-dose abdominal/pelvic computed tomography between adaptive statistical, model-based iterative reconstructions and traditional filtered back projection in 65 patients. Investig Radiol. 2014;49(4):209–16.CrossRef Vardhanabhuti V, Riordan RD, Mitchell GR, Hyde C, Roobottom CA. Image comparative assessment using iterative reconstructions: clinical comparison of low-dose abdominal/pelvic computed tomography between adaptive statistical, model-based iterative reconstructions and traditional filtered back projection in 65 patients. Investig Radiol. 2014;49(4):209–16.CrossRef
18.
Zurück zum Zitat Gervaise A, Osemont B, Lecocq S, Noel A, Micard E, Felblinger J, et al. CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol. 2012;22(2):295–301.PubMedCrossRef Gervaise A, Osemont B, Lecocq S, Noel A, Micard E, Felblinger J, et al. CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol. 2012;22(2):295–301.PubMedCrossRef
19.
Zurück zum Zitat Aletaha D, Neogi T, Silman AJ, Funovits J, Felson DT, Bingham CO 3rd, et al. 2010 rheumatoid arthritis classification criteria: an American College of Rheumatology/European league against rheumatism collaborative initiative. Ann Rheum Dis. 2010;69(9):1580–8.PubMedCrossRef Aletaha D, Neogi T, Silman AJ, Funovits J, Felson DT, Bingham CO 3rd, et al. 2010 rheumatoid arthritis classification criteria: an American College of Rheumatology/European league against rheumatism collaborative initiative. Ann Rheum Dis. 2010;69(9):1580–8.PubMedCrossRef
20.
Zurück zum Zitat Ostergaard M, Edmonds J, McQueen F, Peterfy C, Lassere M, Ejbjerg B, et al. An introduction to the EULAR-OMERACT rheumatoid arthritis MRI reference image atlas. Ann Rheum Dis. 2005;64(Suppl 1):i3–7.PubMedPubMedCentralCrossRef Ostergaard M, Edmonds J, McQueen F, Peterfy C, Lassere M, Ejbjerg B, et al. An introduction to the EULAR-OMERACT rheumatoid arthritis MRI reference image atlas. Ann Rheum Dis. 2005;64(Suppl 1):i3–7.PubMedPubMedCentralCrossRef
21.
Zurück zum Zitat Kudo K, Boutelier T, Pautot F, Honjo K, Hu JQ, Wang HB, et al. Bayesian analysis of perfusion-weighted imaging to predict infarct volume: comparison with singular value decomposition. Magn Reson Med Sci. 2014;13(1):45–50.PubMedCrossRef Kudo K, Boutelier T, Pautot F, Honjo K, Hu JQ, Wang HB, et al. Bayesian analysis of perfusion-weighted imaging to predict infarct volume: comparison with singular value decomposition. Magn Reson Med Sci. 2014;13(1):45–50.PubMedCrossRef
22.
Zurück zum Zitat Sasaki M, Kudo K, Boutelier T, Pautot F, Christensen S, Uwano I, et al. Assessment of the accuracy of a Bayesian estimation algorithm for perfusion CT by using a digital phantom. Neuroradiology. 2013;55(10):1197–203.PubMedCrossRef Sasaki M, Kudo K, Boutelier T, Pautot F, Christensen S, Uwano I, et al. Assessment of the accuracy of a Bayesian estimation algorithm for perfusion CT by using a digital phantom. Neuroradiology. 2013;55(10):1197–203.PubMedCrossRef
23.
Zurück zum Zitat Kudo K, Christensen S, Sasaki M, Ostergaard L, Shirato H, Ogasawara K, et al. Accuracy and reliability assessment of CT and MR perfusion analysis software using a digital phantom. Radiology. 2013;267(1):201–11.PubMedPubMedCentralCrossRef Kudo K, Christensen S, Sasaki M, Ostergaard L, Shirato H, Ogasawara K, et al. Accuracy and reliability assessment of CT and MR perfusion analysis software using a digital phantom. Radiology. 2013;267(1):201–11.PubMedPubMedCentralCrossRef
24.
Zurück zum Zitat Cohen J. Statistical power analysis for the behaviors science.(2nd). New Jersey: Laurence Erlbaum Associates, Publishers, Hillsdale; 1988. Cohen J. Statistical power analysis for the behaviors science.(2nd). New Jersey: Laurence Erlbaum Associates, Publishers, Hillsdale; 1988.
25.
Zurück zum Zitat Lin S, Lin M, Lau KK. Image quality comparison between model-based iterative reconstruction and adaptive statistical iterative reconstruction chest computed tomography in cystic fibrosis patients. J Med Imaging Radiat Oncol. 2019;63(5):602–9.PubMedCrossRef Lin S, Lin M, Lau KK. Image quality comparison between model-based iterative reconstruction and adaptive statistical iterative reconstruction chest computed tomography in cystic fibrosis patients. J Med Imaging Radiat Oncol. 2019;63(5):602–9.PubMedCrossRef
26.
Zurück zum Zitat Ramalho J, Semelka RC, AlObaidy M, Ramalho M, Nunes RH, Castillo M. Signal intensity change on unenhanced T1-weighted images in dentate nucleus following gadobenate dimeglumine in patients with and without previous multiple administrations of gadodiamide. Eur Radiol. 2016;26(11):4080–8.PubMedCrossRef Ramalho J, Semelka RC, AlObaidy M, Ramalho M, Nunes RH, Castillo M. Signal intensity change on unenhanced T1-weighted images in dentate nucleus following gadobenate dimeglumine in patients with and without previous multiple administrations of gadodiamide. Eur Radiol. 2016;26(11):4080–8.PubMedCrossRef
27.
Zurück zum Zitat Bae S, Lee HJ, Han K, Park YW, Choi YS, Ahn SS, et al. Gadolinium deposition in the brain: association with various GBCAs using a generalized additive model. Eur Radiol. 2017;27(8):3353–61.PubMedCrossRef Bae S, Lee HJ, Han K, Park YW, Choi YS, Ahn SS, et al. Gadolinium deposition in the brain: association with various GBCAs using a generalized additive model. Eur Radiol. 2017;27(8):3353–61.PubMedCrossRef
28.
Zurück zum Zitat Moller I, Janta I, Backhaus M, Ohrndorf S, Bong DA, Martinoli C, et al. The 2017 EULAR standardised procedures for ultrasound imaging in rheumatology. Ann Rheum Dis. 2017;76(12):1974–9.PubMedCrossRef Moller I, Janta I, Backhaus M, Ohrndorf S, Bong DA, Martinoli C, et al. The 2017 EULAR standardised procedures for ultrasound imaging in rheumatology. Ann Rheum Dis. 2017;76(12):1974–9.PubMedCrossRef
29.
Zurück zum Zitat Caporali R, Smolen JS. Back to the future: forget ultrasound and focus on clinical assessment in rheumatoid arthritis management. Ann Rheum Dis. 2018;77(1):18–20.PubMedCrossRef Caporali R, Smolen JS. Back to the future: forget ultrasound and focus on clinical assessment in rheumatoid arthritis management. Ann Rheum Dis. 2018;77(1):18–20.PubMedCrossRef
30.
Zurück zum Zitat Jans L, De Kock I, Herregods N, Verstraete K, Van den Bosch F, Carron P, et al. Dual-energy CT: a new imaging modality for bone marrow oedema in rheumatoid arthritis. Ann Rheum Dis. 2018;77(6):958–60.PubMed Jans L, De Kock I, Herregods N, Verstraete K, Van den Bosch F, Carron P, et al. Dual-energy CT: a new imaging modality for bone marrow oedema in rheumatoid arthritis. Ann Rheum Dis. 2018;77(6):958–60.PubMed
31.
Zurück zum Zitat Ng CS, Waterton JC, Kundra V, Brammer D, Ravoori M, Han L, et al. Reproducibility and comparison of DCE-MRI and DCE-CT perfusion parameters in a rat tumor model. Technol Cancer Res Treat. 2012;11(3):279–88.PubMedCrossRef Ng CS, Waterton JC, Kundra V, Brammer D, Ravoori M, Han L, et al. Reproducibility and comparison of DCE-MRI and DCE-CT perfusion parameters in a rat tumor model. Technol Cancer Res Treat. 2012;11(3):279–88.PubMedCrossRef
32.
Zurück zum Zitat Combe B, Landewe R, Daien CI, Hua C, Aletaha D, Alvaro-Gracia JM, et al. 2016 update of the EULAR recommendations for the management of early arthritis. Ann Rheum Dis. 2017;76(6):948–59.PubMedCrossRef Combe B, Landewe R, Daien CI, Hua C, Aletaha D, Alvaro-Gracia JM, et al. 2016 update of the EULAR recommendations for the management of early arthritis. Ann Rheum Dis. 2017;76(6):948–59.PubMedCrossRef
Metadaten
Titel
Perfusion in hand arthritis on dynamic contrast-enhanced computed tomography: a randomized prospective study using MRI as a standard of reference
verfasst von
Sevtap Tugce Ulas
Kay Geert Hermann
Marcus R. Makowski
Robert Biesen
Fabian Proft
Ralph Schilling
Torsten Diekhoff
Publikationsdatum
30.06.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 1/2021
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-020-03526-5

Weitere Artikel der Ausgabe 1/2021

Skeletal Radiology 1/2021 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.