Open Access 01.12.2020 | Case report

Pitfalls of anesthetic management with the Impella® 5.0 device: a case series

verfasst von: Naoshi Hotta, Akito Tsukinaga, Kenji Yoshitani, Yoshihiko Ohnishi

Erschienen in: JA Clinical Reports | Ausgabe 1/2020

Abstract

Background

Impella® is an antegrade left ventricular assist device with a pump catheter in the left ventricle. We report three cases in which we experienced some pitfalls with circulatory management during Impella placement due to new-onset aortic insufficiency (AI) associated with device placement or the limited maximum flow rate.

Case presentation

Three patients developed new-onset AI due to Impella placement. In a patient, the total assisted flow rate was relatively low because of his large body size. In the other patients, in whom the Impella device was used in combination with percutaneous cardiopulmonary support or venoarterial extracorporeal membrane oxygenation (ECMELLA), total flow was maintained at a sufficient level.

Conclusions

New-onset of AI after Impella placement and its limited flow rate are considered to be pitfalls in circulatory management. Management with ECMELLA is considered to be effective during the acute phase when patients have decreased cardiac function.
Hinweise
Trademark ® is used on the first use.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Background

Impella (Abiomed, Danvers, MA) is an antegrade left ventricular assist device (LVAD) with an intravascular microaxial blood pump that delivers blood from the left ventricle to the ascending aorta. It can be quickly placed with less invasiveness because placement does not require open-heart surgery. It can be used solely as an LVAD (LV-IMPELLA), a device for decompression of the left ventricle in combination with percutaneous cardiopulmonary support (PCPS) or venoarterial extracorporeal membrane oxygenation (VA ECMO) (ECMELLA) [1]. Clinical use of Impella was started in Europe in 2004 and in the USA in 2008, and it has been used in more than 50,000 patients all over the world. It has been available in the clinic since September 2016, and its use is spreading for patients with cardiogenic shock resistant to circulatory support by IABP and PCPS in Japan.
Among adverse events related to Impella, aortic insufficiency (AI) may impair hemodynamic conditions and should be carefully observed [2]. We report hemodynamic disturbance AI after insertion of Impella and limited flow rate of it in three cases.

Case presentation

Table 1 shows the details of the three cases in which we found pitfalls with Impella 5.0 during anesthetic management. The primary diseases were ischemic cardiac myopathy, fulminant myocarditis, and dilated-phase of hypertrophic cardiomyopathy. PCPS and intra-aortic balloon pumping (IABP) were used in all patients for circulatory assistance. Due to insufficient organ perfusion and inadequate unloading of the failing left ventricle causing subsequent pulmonary congestion, Impella 5.0 was indicated. Impella 5.0 was placed under general anesthesia with intubation in all cases. Before entering the operating room, all these patients had been in sedative condition and intubated. Anesthesia was maintained with propofol (4–6 mg/kg/h), remifentanil (0.4–0.6 μg/kg/min), and rocuronium (0.4–0.5 mg/kg/h). Monitoring included electrocardiography, pulse oximetry, and transesophageal echocardiography. Data from an intra-arterial line, central venous catheter, and pulmonary artery catheter were also monitored. Under careful monitoring of hemodynamic conditions, IABP was ceased, and the catheter was withdrawn, followed by insertion of Impella 5.0 via subclavian artery. Patient 1 was switched from PCPS and IABP to LV-IMPELLA, and patients 2 and 3 were switched to ECMELLA with remaining PCPS. In three cases, the flow rate of Impella was not enough for viscera. Patients 2 and 3 required PCPS, but not patient 1. Then they weaned from PCPS within several days and switched to LV-IMPELLA.
Table 1.
Circulatory parameters of the four patients
Patient
Age (years) /sex
BSA (m2)
Primary
disease
Preoperative
EF (%)
Preoperative
concomitant VAD
ABP
PAP
CVP
upon arrival in the operating room (after cessation of IABP) (mmHg)
Approach
vessel
Assisted flow rate (L/min-1)
CCO
(L/min) and CCI (L/min/m2)
monitored with pulmonary artery catheter
Postoperative
concomitant VAD
Total flow rate*1 (L/min)
Total flow rate/BSA (L/min/m2)
Inotropic agent and vasoconstrictor dose*2 (μg/kg/min)
Mixed venous oxygen saturation (%)
ABP
PAP
CVP on postoperative day 1 (mmHg)
Duration of Impella 5.0 placement (days)
Outcome
1
58
M
1.7
ICM
11
PCPS
IABP
94/58 (67)
32/19 (35)
7
RSCA
P-9
5.0
3.5
2.0
None
3.5
2.1
DOA 6
DOB 6
NAD 0.3
65%
64/59 (61)
44/26 (28)
5
8
Implantable LVAD
2
43
F
1.5
FMC
5-10
PCPS
IABP
71/66 (66)
28/24 (26)
8
AsAo
P-9
4.5
3.1
2.1
PCPS
2.2 L/min
5.3
3.5
DOA 3
DOB 6
86%
86/74 (77)
19/11 (14)
10
17
Implantable LVAD
3
54
M
1.8
dHCM
10-15
PCPS
IABP
70/52 (63)
42/28 (31)
12
RSCA
P-9
4.8
3.7
2.1
PCPS
2.1 L/min
5.8
3.2
DOA 3
NAD 0.1
70%
82/78 (79)
47/24 (33)
12
27
Implantable LVAD
ICM ischemic cardiomyopathy, AMI acute myocardial infarction, FMC fulminant myocarditis, dHCM dilated phase of hypertrophic cardiomyopathy, BSA body mass index, PCPS percutaneous cardiopulmonary support, IABP intra-aortic balloon pumping, ECMO extracorporeal membrane oxygenation, ABP arterial blood pressure, PAP pulmonary arterial pressure, CVP central venous pressure, RSCA right subclavian artery, AsAo ascending aorta, CCO continuous cardiac output, CCI continuous cardiac index, AI aortic insufficiency, DOA dopamine, DOB dobutamine, NA noradrenaline
*1Total flow rate is based on the sum of CCO determined with the thermodilution technique using the pulmonary artery catheter and PCPS flow rate
*2Maintenance dose after leaving the operating room

New-onset aortic insufficiency due to Impella 5.0 placement

All patients developed new-onset aortic insufficiency and decreased flow rate after placement of Impella. In Patient 1, expected flow displayed on the Impella controller was 5.0 L/min with the maximum assistance level with Impella; however, continuous cardiac output (CCO) determined with the thermodilution technique using a pulmonary artery catheter was 3.5 L/min. TEE revealed newly developed AI after placement of Impella (Fig. 1a, b), suggesting that the decreased cardiac output by 1.5 L/min would correspond to the volume of AI. We suspected that AI was caused by incomplete leaflet coaptation resulting from non-perpendicular placement of Impella to the aortic valve. However, severity of AI was not decreased after adjusting its direction. Similarly, in patients 2 and 3, CCO was over 1 L/min lower than the expected flow displayed on the Impella controller, suggesting aortic insufficiency.

Limited flow rate of the Impella 5.0 device

As shown in Table 1, the dose of noradrenaline was increased in all patients when they experienced a decrease in total assisted flow rate calculated as the sum of CCO and the flow of PCPS. In particular, the dose of noradrenaline in patient 1 was as high as 0.3 μg/kg/min.
In patient 1, pulse pressure was not observed due to an extremely low left ventricular ejection fraction (LVEF, 11%) even with the assisted flow of 5.0 L/min (maximum assistance level) with Impella. Furthermore, due to large body surface area (BSA, 1.7 m2) and new-onset aortic insufficiency, continuous cardiac index (CCI) remained at 2.0 L/min/m2, which was near the lower limit of normal. Large dose of noradrenaline (0.3 μg/kg/min) was required to maintain systemic blood pressure.
In patients 2 and 3, CCI remained at 2.1 L/min/m2, which was near the lower limit of normal, even though the Impella 5.0 assistance level was set to a maximum of P9. However, total flow rate was sufficient for organ perfusion from the point of absolute flow rate (5.3 L and 5.8 L) and flow rate corrected for body surface area (3.5 L/min/m2 and 3.2 L/min/m2) in patients 2 and 3, respectively, because the Impella 5.0 device was used in combination with PCPS as ECMELLA during the acute phase. Therefore, these patients did not require a dose of vasoconstrictors as high as in patient 1.

Discussion

Impella 5.0 is an innovative ventricular assist device that can be placed in the heart quickly and with less invasiveness, but our experiences suggested that new-onset AI associated with placement of Impella and its limited flow rate could be pitfalls in circulatory management.

New-onset aortic insufficiency

All three patients developed new-onset AI after Impella insertion. While a previous case report described a case of AI associated with aortic valve injury due to Impella insertion that persisted after removal [3], AI disappeared after removal of Impella, suggesting that the direct cause of AI was incomplete leaflet coaptation associated with mechanical compression by the cannula of Impella, rather than aortic valve injury in our cases.
Normal quantitative evaluation of AI during the use of a ventricular assist device may lead to underestimation because AI volume varies based on the duration of aortic valve closing. In some cases, AI may occur during the whole cardiac cycle [4]. In some patients without concomitant use of PCPS, such as patient 1, Impella placement may have a larger impact on hemodynamics. Therefore, we argue that AI should be considered in the differential diagnosis when a patient presents with low organ perfusion after Impella placement. In addition, when AI is visualized, artifacts associated with the Impella device also need to be taken into account [5]. For Patient 1, precise identification of AI by TEE was difficult because of reverberation artifacts associated with axial blood flow inside of Impella (Fig. 1a). It was clearly identified on the deep transgastric view (Fig. 1b), suggesting the requirement of careful observation for making a precise diagnosis.

Limited flow rate of the Impella 5.0 device

In patient 1, for whom the Impella 5.0 device was used as LV-IMPELLA, the flow rate decreased both absolutely and relatively because new-onset AI and no spontaneous cardiac output passing through the aortic valve were observed during the whole cardiac cycle, including the diastolic and systolic phases, and organ perfusion decreased. A previous study also reported that LV-IMPELLA led to insufficient organ perfusion due to its limited flow rate including renal failure and liver failure [6]. Other LVADs with higher flow rates can compensate for reduced organ perfusion due to AI [7], and the BSA was relatively large. LV-IMPELLA may lead to insufficient organ perfusion in patients with LVEF that is too low to obtain spontaneous cardiac output. The impact of AI on hemodynamics becomes more significant because the maximum flow rate of the Impella 5.0 device is only 5.0 L/min and in patients with large BSA.
In patient 1, we could obtain flow rates that did not cause signs of circulatory failure such as increased lactate levels and decreased mixed venous oxygen saturation, but these flow rates were insufficient for maintaining normal blood pressure under anesthesia when systemic vascular resistance decreased. Therefore, with the use of the Impella 5.0 device, high-dose vasoconstrictor use is considered mandatory in some cases, as was the case with patient 1.
In patients 2 and 3, the Impella 5.0-assisted flow rate decreased because of new-onset AI. However, ECMELLA augmented the total flow rate and led to maintain arterial blood pressure and organ perfusion. Therefore, only a small dose of vasoconstrictors (or none at all) was required to maintain systemic blood pressure. In a previous case study, a patient in the acute phase of fulminant myocarditis with progressive organ failure during LV-IMPELLA use improved organ function after that Impella 5.0 device was used as ECMELLA in combination with PCPS [8]. Management with ECMELLA was considered to be effective when LV-IMPELLA does not improve circulatory failure due to the limited flow rate. In patients 2 and 3, spontaneous cardiac output started after improvement in cardiac function, and AI was limited to the diastolic phase, allowing for switching from ECMELLA to LV-IMPELLA.

Conclusions

New-onset of AI after Impella 5.0 placement and its limited flow rate are considered to be pitfalls in circulatory management. When the anesthesiologists encountered AI after placement of Impella, ECMELLA is considered to be effective during the acute phase until patients improve cardiac function.
Ackowledgements
None
Not applicable
We have obtained written informed consent from individual patients for this case report.

Competing interests

The authors declare that they have no competing interests
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Tschöpe C, Van Linthout S, Klein O, Mairinger T, Krackhardt F, Potapov EV, et al. Mechanical unloading by fulminant myocarditis : LV-IMPELLA, ECMELLA, BI-PELLA, and PROPELLA concepts. J Cardiovasc Transl Res. 2019;12(2):116–23.CrossRef Tschöpe C, Van Linthout S, Klein O, Mairinger T, Krackhardt F, Potapov EV, et al. Mechanical unloading by fulminant myocarditis : LV-IMPELLA, ECMELLA, BI-PELLA, and PROPELLA concepts. J Cardiovasc Transl Res. 2019;12(2):116–23.CrossRef
2.
Zurück zum Zitat Dimas VV, Morray BH, Kim DW, Almond CS, Shahanavaz S, Tume SC, et al. A multicenter study of the impella device for mechanical support of the systemic circulation in pediatric and adolescent patients. Catheter Cardiovasc Interv. 2017;90(1):124–9.CrossRef Dimas VV, Morray BH, Kim DW, Almond CS, Shahanavaz S, Tume SC, et al. A multicenter study of the impella device for mechanical support of the systemic circulation in pediatric and adolescent patients. Catheter Cardiovasc Interv. 2017;90(1):124–9.CrossRef
3.
Zurück zum Zitat R Chandola, R Cusimano, M Osten EH. Severe aortic insufficiency secondary to 5 L Impella device placement. J Card Surg. 2012;27:400–402. R Chandola, R Cusimano, M Osten EH. Severe aortic insufficiency secondary to 5 L Impella device placement. J Card Surg. 2012;27:400–402.
4.
Zurück zum Zitat Stainback RF, Estep JD, Agler DA, Birks EJ, Bremer M, Hung J, et al. Echocardiography in the management of patients with left ventricular assist devices: recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr [Internet]. 2015;28(8):853–909. Available from: https://doi.org/10.1016/j.echo.2015.05.008. Stainback RF, Estep JD, Agler DA, Birks EJ, Bremer M, Hung J, et al. Echocardiography in the management of patients with left ventricular assist devices: recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr [Internet]. 2015;28(8):853–909. Available from: https://​doi.​org/​10.​1016/​j.​echo.​2015.​05.​008.
5.
Zurück zum Zitat Hong E, Naseem T. Color doppler artifact masking iatrogenic aortic valve injury related to an Impella device. J Cardiothorac Vasc Anesth. 2019 Jun 1;33(6):1584–7.CrossRef Hong E, Naseem T. Color doppler artifact masking iatrogenic aortic valve injury related to an Impella device. J Cardiothorac Vasc Anesth. 2019 Jun 1;33(6):1584–7.CrossRef
6.
Zurück zum Zitat Pieri M, Contri R, Winterton D, Montorfano M, Colombo A, Zangrillo A, et al. The contemporary role of Impella in a comprehensive mechanical circulatory support program: a single institutional experience. BMC Cardiovasc Disord [Internet]. 2015;15(1). Available from: https://doi.org/10.1186/s12872-015-0119-9. Pieri M, Contri R, Winterton D, Montorfano M, Colombo A, Zangrillo A, et al. The contemporary role of Impella in a comprehensive mechanical circulatory support program: a single institutional experience. BMC Cardiovasc Disord [Internet]. 2015;15(1). Available from: https://​doi.​org/​10.​1186/​s12872-015-0119-9.
8.
Zurück zum Zitat Narain S, Paparcuri G, Fuhrman TM, Silverman RB, Peruzzi WT. Novel combination of Impella and extra corporeal membrane oxygenation as a bridge to full recovery in fulminant myocarditis. Case Reports Crit Care. 2012;2012:1–3.CrossRef Narain S, Paparcuri G, Fuhrman TM, Silverman RB, Peruzzi WT. Novel combination of Impella and extra corporeal membrane oxygenation as a bridge to full recovery in fulminant myocarditis. Case Reports Crit Care. 2012;2012:1–3.CrossRef
Metadaten
Titel
Pitfalls of anesthetic management with the Impella® 5.0 device: a case series
verfasst von
Naoshi Hotta
Akito Tsukinaga
Kenji Yoshitani
Yoshihiko Ohnishi
Publikationsdatum
01.12.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
JA Clinical Reports / Ausgabe 1/2020
Elektronische ISSN: 2363-9024
DOI
https://doi.org/10.1186/s40981-020-00324-9

Weitere Artikel der Ausgabe 1/2020

JA Clinical Reports 1/2020 Zur Ausgabe

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.