Spatial orientation in pathway surgery

verfasst von: Chunman Fan, Dimitra Dodou, Paul Breedveld, Jenny Dankelman

Erschienen in: Surgical Endoscopy | Ausgabe 9/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

In the field of minimally invasive surgery, specifically in pathway surgery (i.e. minimal invasive procedures carried out transluminally or through instrument-created pathways), spatial disorientation is a common experience to endoscopists. In this article, two effects that may cause spatial disorientation in pathway surgery, ‘control-display compatibility’ and ‘local disorientation’, were studied.

Method

A custom-developed simulator Endo-PaC was developed and used for mimicking pathway surgical scenarios. In Study 1, two ways of control-display alignment, normal mapping and mirrored mapping, were tested in combination with two control devices, thumb control and wrist control, in an orienting task using Endo-PaC. In Study 2, a tethered viewpoint was added to the virtual instrument tip. It was hypothesized that the visible tip would provide a cue of orientating direction in the reference frame during the instrument navigation. In both studies, novice participants were involved, and their performance was evaluated with regard to task time, path length travelled by the virtual tip, time and number of warnings, and subjective workload and personal preference.

Results

In Study 1, normal-thumb and normal-wrist mapping yielded significantly lower means than mirrored-thumb and mirrored-wrist control for all investigated objective and subjective performance measurements. Out of 24 participants, 20 participants preferred normal control mapping. In Study 2, participants performed the task in shorter time and with shorter path length when the tip was visible tip on the monitor using a tethered viewpoint, but with a lower number and time of warnings without a visible tip.

Conclusion

The results of our studies show that eliminating the visual-display misalignment would greatly improve novice participants’ performance, reduce the training time and their cognitive workload. A visible tip on the monitor would provide strong direction cue and shorten the performance time, but might introduce collision errors to novices and therefore requires longer training time.
Literatur
1.
Zurück zum Zitat Rattner D, Kalloo A (2006) ASGE/SAGES ASGE/SAGES working group on natural orifice translumenal endoscopic surgery. Surg Endosc 20(2):329–333CrossRefPubMed Rattner D, Kalloo A (2006) ASGE/SAGES ASGE/SAGES working group on natural orifice translumenal endoscopic surgery. Surg Endosc 20(2):329–333CrossRefPubMed
2.
Zurück zum Zitat Shahinian HK, Kabil MS, Jarrahy R (2008) Endoscopic skull base surgery: a comprehensive guide with illustrative cases. Humana Press, Totowa Shahinian HK, Kabil MS, Jarrahy R (2008) Endoscopic skull base surgery: a comprehensive guide with illustrative cases. Humana Press, Totowa
3.
Zurück zum Zitat Cao CG, Milgram P (2000) Disorientation in minimal access surgery: a case study. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 169–172 Cao CG, Milgram P (2000) Disorientation in minimal access surgery: a case study. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 169–172
4.
Zurück zum Zitat Worringham CJ, Beringer DB (1989) Operator orientation and compatibility in visual-motor task performance. Ergonomics 32(4):387–399CrossRefPubMed Worringham CJ, Beringer DB (1989) Operator orientation and compatibility in visual-motor task performance. Ergonomics 32(4):387–399CrossRefPubMed
5.
Zurück zum Zitat Chan WH, Chan AHS (2007) Strength and reversibility of movement stereotypes for lever control and circular display. Int J Ind Ergon 37(3):233–244CrossRef Chan WH, Chan AHS (2007) Strength and reversibility of movement stereotypes for lever control and circular display. Int J Ind Ergon 37(3):233–244CrossRef
6.
Zurück zum Zitat Hollands JG, Wickens CD (1999) Engineering psychology and human performance. Prentice Hall, Upper Saddle River Hollands JG, Wickens CD (1999) Engineering psychology and human performance. Prentice Hall, Upper Saddle River
7.
Zurück zum Zitat Smith RL, Stuart MA (1989) The effects of spatially displaced visual feedback on remote manipulator performance. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 1430–1434 Smith RL, Stuart MA (1989) The effects of spatially displaced visual feedback on remote manipulator performance. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 1430–1434
8.
Zurück zum Zitat Brooks TL, Ince I (1992) Operator vision aids for telerobotic assembly and servicing in space. In: Proceedings of the IEEE International Conference on Robotics and Automation, pp 886–891 Brooks TL, Ince I (1992) Operator vision aids for telerobotic assembly and servicing in space. In: Proceedings of the IEEE International Conference on Robotics and Automation, pp 886–891
9.
Zurück zum Zitat Gray SV, Wilson JR, Syan CS (1992) Human control of robot motion: orientation, perception and compatibility. In: Human-robot interaction. Taylor & Francis, London, pp 48–64 Gray SV, Wilson JR, Syan CS (1992) Human control of robot motion: orientation, perception and compatibility. In: Human-robot interaction. Taylor & Francis, London, pp 48–64
10.
Zurück zum Zitat Macedo JA, Kaber DB, Endsley MR, Powanusorn P, Myung S (1998) The effect of automated compensation for incongruent axes on teleoperator performance. J Hum Factors Ergon Soc 40:541–553CrossRef Macedo JA, Kaber DB, Endsley MR, Powanusorn P, Myung S (1998) The effect of automated compensation for incongruent axes on teleoperator performance. J Hum Factors Ergon Soc 40:541–553CrossRef
11.
Zurück zum Zitat McKee GT, Schenker PS (1993) Nonanthropomorphic viewing for teleoperation. In: Optical tools for manufacturing and advanced automation. International Society for Optics and Photonics, pp 282–291 McKee GT, Schenker PS (1993) Nonanthropomorphic viewing for teleoperation. In: Optical tools for manufacturing and advanced automation. International Society for Optics and Photonics, pp 282–291
12.
Zurück zum Zitat Gallagher AG, McClure N, McGuigan J, Ritchie K, Sheehy NP (1998) An ergonomic analysis of the fulcrum effect in the acquisition of endoscopic skills. Endoscopy 30:617–620CrossRefPubMed Gallagher AG, McClure N, McGuigan J, Ritchie K, Sheehy NP (1998) An ergonomic analysis of the fulcrum effect in the acquisition of endoscopic skills. Endoscopy 30:617–620CrossRefPubMed
13.
Zurück zum Zitat Hills JW, Green PS, Jensen JF, Gorfu Y, Shah AS (1994) Telepresence surgery demonstration system. In: Proceedings of the IEEE International Conference on Robotics and Automation, pp 2302–2307 Hills JW, Green PS, Jensen JF, Gorfu Y, Shah AS (1994) Telepresence surgery demonstration system. In: Proceedings of the IEEE International Conference on Robotics and Automation, pp 2302–2307
14.
Zurück zum Zitat Fan C, Dodou D, Breedveld P (2012) Review of manual control methods for handheld manoeuvrable instruments. Minim Invasive Ther Allied Technol 22:127–135CrossRefPubMed Fan C, Dodou D, Breedveld P (2012) Review of manual control methods for handheld manoeuvrable instruments. Minim Invasive Ther Allied Technol 22:127–135CrossRefPubMed
15.
Zurück zum Zitat Hart SG, Staveland LE (1988) Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Hum Ment Workload 1:139–183CrossRef Hart SG, Staveland LE (1988) Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Hum Ment Workload 1:139–183CrossRef
16.
Zurück zum Zitat De Winter JCF (2014) Controversy in human factors constructs and the explosive use of the NASA TLX: a measurement perspective. Cogn Technol Work 16:289–297 De Winter JCF (2014) Controversy in human factors constructs and the explosive use of the NASA TLX: a measurement perspective. Cogn Technol Work 16:289–297
17.
Zurück zum Zitat Breedveld P (2010) Steerable laparoscopic cable-ring forceps. J Med Devices 4(2):027518-1CrossRef Breedveld P (2010) Steerable laparoscopic cable-ring forceps. J Med Devices 4(2):027518-1CrossRef
18.
Zurück zum Zitat Wickens CD, Vincow M, Yeh M (2005) Design application of visual spatial thinking: the importance of frame of reference. In: Shah P, Miyake A (eds) The Cambridge handbook of visuospatial thinking, Cambridge University Press, pp 383–425 Wickens CD, Vincow M, Yeh M (2005) Design application of visual spatial thinking: the importance of frame of reference. In: Shah P, Miyake A (eds) The Cambridge handbook of visuospatial thinking, Cambridge University Press, pp 383–425
19.
Zurück zum Zitat Winckens CD, Hollands JG, Parasuraman R, Banbury S (2012) Engineering psychology and human performance, 4th edn. Pearson, Boston Winckens CD, Hollands JG, Parasuraman R, Banbury S (2012) Engineering psychology and human performance, 4th edn. Pearson, Boston
20.
Zurück zum Zitat McCormick EP, Wickens CD, Banks R, Yeh M (1998) Frame of reference effects on scientific visualization subtasks. J Hum Factors Ergon Soc 40(3):443–451CrossRef McCormick EP, Wickens CD, Banks R, Yeh M (1998) Frame of reference effects on scientific visualization subtasks. J Hum Factors Ergon Soc 40(3):443–451CrossRef
21.
Zurück zum Zitat Wang W, Milgram P (2001) Dynamic viewpoint tethering for navigation in large-scale virtual environments. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 1862–1866 Wang W, Milgram P (2001) Dynamic viewpoint tethering for navigation in large-scale virtual environments. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 1862–1866
22.
Zurück zum Zitat Wang W, Milgram P (2003) Effects of viewpoint displacement on navigational performance in virtual environments. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 139–143 Wang W, Milgram P (2003) Effects of viewpoint displacement on navigational performance in virtual environments. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 139–143
23.
Zurück zum Zitat Golledge RG (1999) Human wayfinding and cognitive maps. In: Golledge RG (ed) Wayfinding behaviour: cognitive mapping and other spatial processes, Johns Hopkins Press, Baltimore, pp 5–45 Golledge RG (1999) Human wayfinding and cognitive maps. In: Golledge RG (ed) Wayfinding behaviour: cognitive mapping and other spatial processes, Johns Hopkins Press, Baltimore, pp 5–45
24.
Zurück zum Zitat Cotton PB, Williams CB (2008) Practical gastrointestinal endoscopy: the fundamentals. Wiley, BlackwellCrossRef Cotton PB, Williams CB (2008) Practical gastrointestinal endoscopy: the fundamentals. Wiley, BlackwellCrossRef
25.
Zurück zum Zitat Cirocco WC, Rusin LC (1996) Fluoroscopy, a valuable ally during difficult colonoscopy. Surg Endosc 10(11):1080–1084CrossRefPubMed Cirocco WC, Rusin LC (1996) Fluoroscopy, a valuable ally during difficult colonoscopy. Surg Endosc 10(11):1080–1084CrossRefPubMed
26.
Zurück zum Zitat Cao CGL (2001) Designing for spatial orientation in endoscopic environments. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 1259–1263 Cao CGL (2001) Designing for spatial orientation in endoscopic environments. In: Proceedings of the Human Factors and Ergonomics Society Annual Meeting, SAGE Publications, pp 1259–1263
27.
Zurück zum Zitat Marvik RM, Langø T, Tangen GA, Andersen JO, Kaspersen JH, Ystgaard B et al (2004) Laparoscopic navigation pointer for three-dimensional image-guided surgery. Surg Endosc Other Interv Tech 18(8):1242–1248CrossRef Marvik RM, Langø T, Tangen GA, Andersen JO, Kaspersen JH, Ystgaard B et al (2004) Laparoscopic navigation pointer for three-dimensional image-guided surgery. Surg Endosc Other Interv Tech 18(8):1242–1248CrossRef
28.
Zurück zum Zitat Saunders BP, Bell GD, Williams CB, Bladen JS, Anderson AP (1995) First clinical results with a real time, electronic imager as an aid to colonoscopy. Gut 36(6):913–917PubMedCentralCrossRefPubMed Saunders BP, Bell GD, Williams CB, Bladen JS, Anderson AP (1995) First clinical results with a real time, electronic imager as an aid to colonoscopy. Gut 36(6):913–917PubMedCentralCrossRefPubMed
29.
Zurück zum Zitat Taylor HA, Brunye TT, Taylor ST (2008) Spatial mental representation: implications for navigation system design. Rev Hum Factors Ergon 4(1):1–40CrossRef Taylor HA, Brunye TT, Taylor ST (2008) Spatial mental representation: implications for navigation system design. Rev Hum Factors Ergon 4(1):1–40CrossRef
30.
Zurück zum Zitat Haluck RS, Webster RW, Snyder AJ, Melkonian MG, Mohler BJ, Dise ML, Lefever A (2003) A virtual reality surgical trainer for navigation in laparoscopic surgery. Med Meets Virtual Real 105(2):119–131 Haluck RS, Webster RW, Snyder AJ, Melkonian MG, Mohler BJ, Dise ML, Lefever A (2003) A virtual reality surgical trainer for navigation in laparoscopic surgery. Med Meets Virtual Real 105(2):119–131
31.
Zurück zum Zitat Carrozza MC, Arena A, Accoto D, Menciassi A, Dario P (2003) A SMA-actuated miniature pressure regulator for a miniature robot for colonoscopy. Sens Actuators A 105(2):119–131CrossRef Carrozza MC, Arena A, Accoto D, Menciassi A, Dario P (2003) A SMA-actuated miniature pressure regulator for a miniature robot for colonoscopy. Sens Actuators A 105(2):119–131CrossRef
32.
Zurück zum Zitat Ng WS, Phee S, Seow C, Davies BL (2000) Development of a robotic colonoscope. Dig Endosc 12(2):131–135CrossRef Ng WS, Phee S, Seow C, Davies BL (2000) Development of a robotic colonoscope. Dig Endosc 12(2):131–135CrossRef
33.
Zurück zum Zitat Zazzarini CC, Pansini A, Cerveri P, Zaltieri R, Lavizzari D (2011) Design of a robotic endoscope for mini invasive surgery. In: ASME 2011 International Design Engineering Technical Conferences and Computers and Information in Engineering Conference. American Society of Mechanical Engineers, pp 771–777 Zazzarini CC, Pansini A, Cerveri P, Zaltieri R, Lavizzari D (2011) Design of a robotic endoscope for mini invasive surgery. In: ASME 2011 International Design Engineering Technical Conferences and Computers and Information in Engineering Conference. American Society of Mechanical Engineers, pp 771–777
34.
Zurück zum Zitat Inoue H, Okabe S, Takeshita K, Muraoka Y, Yoneshima H, Endo M (1993) Colonoscopy using a transparent plastic cap. Gastroenterol Endosc 35:378–381 Inoue H, Okabe S, Takeshita K, Muraoka Y, Yoneshima H, Endo M (1993) Colonoscopy using a transparent plastic cap. Gastroenterol Endosc 35:378–381
35.
Zurück zum Zitat Tsumura T, Torii A, Fujita S, Takeda J, Hikita H, Nishikawa H, Ochi J, Miura K (2003) Usefulness of oblique transparent cylinders in facilitating colonoscopy. Dig Endosc 15(2):121–124CrossRef Tsumura T, Torii A, Fujita S, Takeda J, Hikita H, Nishikawa H, Ochi J, Miura K (2003) Usefulness of oblique transparent cylinders in facilitating colonoscopy. Dig Endosc 15(2):121–124CrossRef
36.
Zurück zum Zitat Dafnis GM (2000) Technical considerations and patient comfort in total colonoscopy with and without a transparent cap: initial experiences from a pilot study. Endoscopy 32(5):381–384CrossRefPubMed Dafnis GM (2000) Technical considerations and patient comfort in total colonoscopy with and without a transparent cap: initial experiences from a pilot study. Endoscopy 32(5):381–384CrossRefPubMed
37.
Zurück zum Zitat Kondo S, Yamaji Y, Watabe H, Yamada A, Sugimoto T, Ohta M et al (2007) A randomized controlled trial evaluating the usefulness of a transparent hood attached to the tip of the colonoscope. Am J Gastroenterol 102(1):75–81CrossRefPubMed Kondo S, Yamaji Y, Watabe H, Yamada A, Sugimoto T, Ohta M et al (2007) A randomized controlled trial evaluating the usefulness of a transparent hood attached to the tip of the colonoscope. Am J Gastroenterol 102(1):75–81CrossRefPubMed
38.
Zurück zum Zitat Yap CK, Ng HS (2001) Cap-fitted gastroscopy improves visualization and targeting of lesions. Gastrointest Endosc 53(1):93–95CrossRefPubMed Yap CK, Ng HS (2001) Cap-fitted gastroscopy improves visualization and targeting of lesions. Gastrointest Endosc 53(1):93–95CrossRefPubMed
39.
Zurück zum Zitat Urita Y, Nishino M, Ariki H, Ozaki M, Naruki Y, Otsuka S (1997) A transparent hood simplifies magnifying observation of the colonic mucosa by colonoscopy. Gastrointest Endosc 46(2):170–172CrossRefPubMed Urita Y, Nishino M, Ariki H, Ozaki M, Naruki Y, Otsuka S (1997) A transparent hood simplifies magnifying observation of the colonic mucosa by colonoscopy. Gastrointest Endosc 46(2):170–172CrossRefPubMed
40.
Zurück zum Zitat Lee YT, Hui AJ, Wong VWS, Hung LCT, Sung JJY (2006) Improved colonoscopy success rate with a distally attached mucosectomy cap. Endoscopy 38(7):739–742CrossRefPubMed Lee YT, Hui AJ, Wong VWS, Hung LCT, Sung JJY (2006) Improved colonoscopy success rate with a distally attached mucosectomy cap. Endoscopy 38(7):739–742CrossRefPubMed
41.
Zurück zum Zitat Tada M, Inoue H, Yabata E, Okabe S, Endo M (1997) Feasibility of the transparent cap-fitted colonoscope for screening and mucosal resection. Dis Colon Rectum 40(5):618–621CrossRefPubMed Tada M, Inoue H, Yabata E, Okabe S, Endo M (1997) Feasibility of the transparent cap-fitted colonoscope for screening and mucosal resection. Dis Colon Rectum 40(5):618–621CrossRefPubMed
42.
Zurück zum Zitat Matsushita M, Hajiro K, Okazaki K, Takakuwa H, Tominaga M (1998) Efficacy of total colonoscopy with a transparent cap in comparison with colonoscopy without the cap. Endoscopy 30(5):444–447CrossRefPubMed Matsushita M, Hajiro K, Okazaki K, Takakuwa H, Tominaga M (1998) Efficacy of total colonoscopy with a transparent cap in comparison with colonoscopy without the cap. Endoscopy 30(5):444–447CrossRefPubMed
43.
Zurück zum Zitat Harada Y, Hirasawa D, Fujita N, Noda Y, Kobayashi G, Kazuhiko I, Yonechi M et al (2009) Impact of a transparent hood on the performance of total colonoscopy: a randomized controlled trial. Gastrointest Endosc 69(3):637–644CrossRefPubMed Harada Y, Hirasawa D, Fujita N, Noda Y, Kobayashi G, Kazuhiko I, Yonechi M et al (2009) Impact of a transparent hood on the performance of total colonoscopy: a randomized controlled trial. Gastrointest Endosc 69(3):637–644CrossRefPubMed
Metadaten
Titel
Spatial orientation in pathway surgery
verfasst von
Chunman Fan
Dimitra Dodou
Paul Breedveld
Jenny Dankelman
Publikationsdatum
01.09.2015
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 9/2015
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-014-3993-9

Weitere Artikel der Ausgabe 9/2015

Surgical Endoscopy 9/2015 Zur Ausgabe

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Deutlich weniger Infektionen: Wundprotektoren schützen!

08.05.2024 Postoperative Wundinfektion Nachrichten

Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.

Chirurginnen und Chirurgen sind stark suizidgefährdet

07.05.2024 Suizid Nachrichten

Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.